2016 GD02 AUSTRALASIAN ASSOCIATION OF CLINICAL BIOCHEMISTS INC 5/85 Bourke Rd, Alexandria NSW, 2015 Telephone: +61 2 9669 6600 Facsimile: +61 2 9669 6607 Email: [email protected] Guideline Title Recommendations for Use of Point-of-Care (POC) Troponin Assays in Assessment of Acute Coronary Syndrome Document Number: 2016 GD02 Publication date: December 2016 Next review date: December 2021 Owner: Australasian Association of Clinical Biochemists Troponin PoCT Working Party

AACB Guideline cover sheet

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: AACB Guideline cover sheet

2016 GD02

AUSTRALASIAN ASSOCIATION OF CLINICAL BIOCHEMISTS INC

5/85 Bourke Rd, Alexandria NSW, 2015 Telephone: +61 2 9669 6600 Facsimile: +61 2 9669 6607 Email: [email protected]

Guideline Title Recommendations for Use of Point-of-Care (POC) Troponin Assays in

Assessment of Acute Coronary Syndrome Document Number: 2016 GD02 Publication date: December 2016 Next review date: December 2021 Owner: Australasian Association of Clinical Biochemists Troponin PoCT Working Party

Page 2: AACB Guideline cover sheet

December 2016 1

Recommendations for Use of Point-of-Care (POC) Troponin Assays in Assessment of Acute Coronary

Syndrome

Endorsed by:

Australasian Association of Clinical Biochemists (AACB)

Royal College of Pathologists of Australasia (RCPA)

The Integrated Cardiovascular Clinical Network CHSA

New South Wales Health Pathology

Pathology Queensland

Page 3: AACB Guideline cover sheet

December 2016 2

TABLE of CONTENTS ITEM PAGES Preamble 3-4 Executive Summary: 5-8 1. Clinical recommendations 2. Laboratory recommendations

2.1 Analytical 2.2 Preanalytical (sample collection and processing) 2.3 Postanalytical 2.4 Governance

3. Literature references Comprehensive Document: 1. Clinical recommendations 9-10

Table 1 – Causes of elevated cardiac troponins

2. Laboratory recommendations 11-20 2.1 Analytical 2.2 Preanalytical (sample collection and processing) 2.3 Postanalytical

2.3.1 - Test name 2.3.2 - Reporting units 2.3.3 - Education to users 2.4 Governance

2.4.1 - Input from stakeholders 2.4.2 - Audit and review of performance and clinical use

Table 2: Analytical characteristics of commercial Point-of-Care troponin (cTn) assays declared by manufacturers Figure 1: Analytical sensitivity of POC troponin I assay (i-Stat) and sensitive troponin I assays (Beckman Coulter AccuTnI and Abbott Architect cTnI). Figures 2-5: Comparison and diagnostic accuracy of serial troponin measurements by POC and high sensitive TnT assays in patients presenting with symptoms suggestive of ACS. Figures 6A-6E: Precision of high sensitive troponin (hs-cTn) and POC cTn assays determined close to the 99th percentile cTn concentration. Table 3: Long term imprecision for cardiac troponin assays

Appendix 1: Methods for developing the recommendations 21-22 Appendix 2: Local algorithms for Point-of-Care Troponin testing 23-26

I. AQT90 Troponin T Testing Protocol – iCCnet CHSA and NSW Health Pathology II. AQT90 Troponin I Testing Protocol – iCCnet CHSA III. Abbott i-STAT Troponin I Testing Protocol – Pathology Queensland IV. Cobas h232 Troponin T Testing protocol – iCCnet CHSA

Appendix 3: Diagnostic sensitivity and specificity for Point-of-Care (POC) and high sensitive troponin studies from the recent literature 27-30

Appendix 4: Educational materials 31

Appendix 5: Glossary of terms: highly sensitive cTn assays; limit of detection; limit of quantitation;

99th percentile of a reference population 31

Literature references 32-33

Page 4: AACB Guideline cover sheet

December 2016 3

Recommendations for Use of Point-of-Care (POC) Troponin Assays in Assessment of Acute Coronary Syndrome

Endorsed by:

Australasian Association of Clinical Biochemists (AACB) Royal College of Pathologists of Australasia (RCPA)

The Integrated Cardiovascular Clinical Network CHSA New South Wales Health Pathology

Pathology Queensland

PREAMBLE Background and Rationale

Surveys conducted in Europe 1-3 and in Australia 4 have reported on the lack of harmonisation of all aspects of troponin measurement in the routine laboratory and at the point-of-care (POC), including pre-analytical sample collection and haemolysis, decision limits, and optimal sampling times, in addition to guideline recommendations for troponin use in diagnostic pathways for patients presenting with chest pain to the Emergency Department. Harmonised analytical and clinical recommendations are needed in the assessment of acute coronary syndrome (ACS). As an example, laboratories in Australia that are using the Architect high sensitive troponin I assay (hs-TnI) or the Cobas or Elecsys high sensitive troponin T assay (hs-TnT) have agreed upon harmonised decision limits (personal communications, LC, JT, RH). Aims and Scope

This document is based on the currently available evidence and consensus and comments of key stakeholders involved in the management of ACS patients. The recommendations aim to cover both the clinical and laboratory aspects of using POC troponin assays in the assessment of ACS. Clinical recommendations will highlight the importance of understanding: o Differences between laboratory-based and POC troponin assays and their impact on interpretation

of test results and clinical decision making; o Timed serial measurements on patients presenting with symptoms suggestive of ACS to ensure

safety and that acute myocardial infarction (AMI) is not missed; o Interpretation of troponin results in the clinical context; o Importance of clinical acumen when interpreting all troponin results. Laboratory recommendations will highlight: o Limitations of POC troponin assays; o Delineation of troponin assays as to whether they are “clinically usable”; o Assay performance specifications. Target group

The recommendations are intended for: o Clinicians requesting and interpreting troponin tests in patients presenting with symptoms of ACS o Health care professionals, both in clinical and laboratory settings, providing a POC troponin testing

service o Manufacturers of POC troponin devices

Page 5: AACB Guideline cover sheet

December 2016 4

Format and tools for implementation

The recommendations are presented in two formats: 1/ executive summary outlining the key clinical and laboratory recommendations; 2/ comprehensive document explaining the recommendations and the evidence behind them (Appendix 3),

and the process of the development of the document (Appendix 1). The recommendations are supplemented with the below tools to facilitate implementation: o Executive summary of key recommendations for clinical and laboratory staff o Diagnostic algorithms for different POC troponin assays (see Appendix 2 for examples) o Educational materials including a slide presentation of the recommendations and case studies

The educational materials are available on the AACB website at: http://www.aacb.asn.au/resources/recommendations-for-use-of-troponin-poc References

1. Pulkki K, Suvisaari J, Collinson P, Ravkilde J, Stavljenic-Rukavina A, Hammerer-Lercher A, et al. A pilot survey of the use and implementation of cardiac markers in acute coronary syndrome and heart failure across Europe. Clin Chem Lab Med 2009; 47:227-34.

2. Collinson PO, van Dieijen-Visser MP, Pulkki K, Hammerer-Lercher A, Suvisaari J, Ravkilde J, et al. Evidence-based laboratory medicine: how well do laboratories follow recommendations and guidelines? The Cardiac Marker Guideline Uptake in Europe (CARMAGUE). Clin Chem 2012;58:305-6.

3. McKeeman GC, Auld PW. A national survey of troponin testing and recommendations for improved practice. Ann Clin Biochem 2015;52:527-42.

4. Davey RX. Troponin reporting in Australia in 2003. Ann Clin Biochem 2005;42:61-3.

Page 6: AACB Guideline cover sheet

December 2016 5

EXECUTIVE SUMMARY: KEY RECOMMENDATIONS 1. Clinical Recommendations

I. This guideline provides recommendations for the use of POC Troponin test results in the management of ACS patients.

II. The following recommendations take into account the impact of the analytical performance of currently available Tn assays on their clinical use and interpretation. The recommendations regarding use of POC troponin assays are based on laboratory data showing that current POC assays are less analytically sensitive than most laboratory assays (labelled as ‘high sensitivity’ or ‘sensitive’ assays) for troponin.

III. The results of various troponin assays, whether they are available in POC or laboratory formats, are NOT directly comparable and values CANNOT be transferred between assays unless there is evidence of such comparability.

IV. Clinical users of POC and laboratory-based troponin assays need to be aware of the limitations of the actual troponin assay used in the management of their patients in order to avoid adverse effects on patient management and for patient safety.

V. Troponin results must always be interpreted in the clinical context. Biomarkers are only one component of clinical information to be used with ECG(s) and clinical history for risk stratification according to clinical guidelines. 5

− Elevated troponin results alone do not rule in the diagnosis of ACS as elevated troponin can occur in non-ACS conditions (causes of non-ACS troponin elevations are shown in Table 1).

− Normal troponin results (at 0 and 3 hours after admission of patient to the Emergency Department [ED]) do not rule out ACS where there is a high clinical suspicion; in such cases further troponin testing may be indicated at 6 hours or at 12 hours post admission to ED (slowly evolving AMIs may not elevate troponin until >12 hours after admission and unstable angina (UA) will not be detected). 6

− Less sensitive troponin assays may require a longer time period to show a clinically significant elevation in troponin.

VI. Serial troponin measurements should be performed for all patients presenting to an ED with symptoms suggestive of ACS, unless the patient has been reliably ACS symptom-free for 12 hours when a single sample is appropriate.

− The reason a second sample has NOT been tested should be clearly justified and documented. − The baseline should be the time of the first blood draw rather than from time of onset of

symptoms. − Recommended timing of samples using accelerated diagnostic pathways, using laboratory-

based highly sensitive assays, is obtaining the first sample at admission to ED and the second sample at 2-3 hours later.

− In order to safely discharge patients from the ED, it is recommended when using less sensitive (e.g. POC) troponin assays that a sample is collected at 6-8 hours after the first blood draw. Evidence is lacking to use POC troponin results in accelerated diagnostic pathways. 7

VII. Serial testing should be performed using the same troponin assay and platform. We recommend to re-baseline the troponin value for individual patients if transported to a different hospital location where a different troponin assay is used. Example scenarios are: − Baseline Troponin I done at presentation but second location uses troponin T assay, or vice

versa; − Baseline troponin I done by a POC device but second location uses a different laboratory-

based troponin I assay or one with higher analytical sensitivity; − Baseline troponin T done by a POC device but second location uses a different laboratory-

based troponin T assay or one with higher analytical sensitivity.

Page 7: AACB Guideline cover sheet

December 2016 6

At the second location collect a new baseline sample and a second sample at 2-3 hours (if testing with a highly sensitive troponin assay) and 6-8 hours (if testing with a less analytically sensitive assay) unless the patient has been symptom-free for >12 hours, when a single sample is appropriate.

VII. In general quantitative troponin delta change has not been determined for use with POC assays as POC assays are less precise at low troponin concentration (see local algorithms for examples).

VIII. When a laboratory is used, results should be available as soon as possible, with a goal of within 60 minutes after the sample is received in the laboratory. Otherwise POC testing could be considered although the laboratory-based troponin assay is the preferred testing option when available. 8

2. Laboratory Recommendations

2.1 Analytical

I. It is the laboratory’s responsibility to inform clinicians about the limitations of POC troponin vs. laboratory-based troponin assays in terms of analytical performance which impacts their clinical utility. − All current troponin POC assays are analytically less sensitive for the detection of the troponin

molecule. − Troponin is measurable above the limit of detection in <10% of the healthy population with

POC devices. − POC assays are less analytically sensitive at earlier time points (<6 hours after the onset of

symptoms) compared with highly sensitive and sensitive assays.

II. POC troponin assays are “clinically usable” based on a scorecard performance approach. 9

III. Both laboratory-based and POC troponin assays may be affected by analytical interferences including heterophilic antibodies, immunoglobulin complexes, fibrin clots, etc. These may interfere with troponin measurement in some assays and cause false-positive or false-negative values.

IV. POC troponin methods should be validated for clinical concordance, or at least compared against

the local laboratory-based sensitive or highly sensitive troponin assay where patients may be referred. In this way POC results can be assessed against laboratory results, but not used for monitoring unless the differences are larger than the variation between assays.

V. Analytical performance specifications for POC troponins should ideally be driven by data coming from direct studies investigating the impact of analytical performance of the test on clinical outcomes related to the management of patients at risk of ACS; or, for practical reasons, from indirect outcome studies investigating the impact of analytical performance on clinical classification or decisions related to the management of ACS patients.

2.2 Preanalytical (sample collection and processing)

I. Troponin testing should only be performed by operators who have attained the required competency standards and where the POC system is accredited within a quality framework.

II. False negative and false positive troponin results due to micro-clots and haemolysis are relatively common in Emergency Department samples. 10 Educational examples (Appendix 4) are available at the AACB website at: http://www.aacb.asn.au/resources/recommendations-for-use-of-troponin-poc

Page 8: AACB Guideline cover sheet

December 2016 7

III. A general approach to reducing micro-clots is: Mix blood tubes immediately after collection to avoid formation of micro-clots that may not be visible on testing but which may cause incorrect troponin values.

IV. Avoid haemolysed samples by: − Use of trained collectors − Do not collect blood from the site of insertion of an intravenous cannula − Do not use very fine bore needles − Do not collect blood into syringe and then transfer it into a vacuum tube.

2.3 Postanalytical

2.3.1 Test name:

I. The naming of generations of troponin assays has been inconsistent and largely market-driven. Highly sensitive assays measure the same protein as the sensitive assays, and even more sensitive cardiac troponin assays will certainly be developed, thus making any ‘high sensitivity’ designation obsolete.

II. For the purpose of reporting of results, if the analytical performance of a new assay is significantly different from a previous troponin assay the test name should indicate this, e.g. ‘Troponin T – Highly sensitive’.

III. We recommend there is a general consensus in Australia regarding troponin test names for the same manufacturer’s assay to allow for harmonised reporting of troponin in the electronic medical record.

IV. Troponin results released from POC devices should be clearly differentiated from troponin results reported from the laboratory using an assay with different analytical sensitivity (e.g. ‘Troponin T – POC’ vs ‘Troponin T – Highly sensitive’).

2.3.2 Reporting units:

I. The international consensus is that troponin should be reported in ng/L unitage and expressed as whole numbers. 11 Highly sensitive troponin assays generally use units of ng/L, whereas results from less sensitive assays have usually been reported in µg/L to two or more decimal points. As a consequence, the reporting of these more sensitive assays produces results that may be a thousand-fold different compared to results by other less sensitive assays, and thus has the potential for confusion and medical error, e.g. 4000 ng/L is 4.000 µg/L and 4 ng/L is 0.004 µg/L.

II. To improve patient safety we recommend that the clinical and laboratory communities standardise on troponin units of measurement in Australia namely “ng/L” and report values in whole numbers. This will allow for harmonised reporting of troponin in the electronic medical record.

2.3.3 Education to users about assay: 12

I. Various healthcare professionals request and measure troponin. It is essential that there is a comprehensive education process communicated to all users.

II. Supporting information detailing troponin assay utility and comparing the local assay to others in use should be provided.

III. General educational materials including a slide presentation of the recommendations and case studies are available on the AACB website at: http://www.aacb.asn.au/resources/recommendations-for-use-of-troponin-poc

Page 9: AACB Guideline cover sheet

December 2016 8

2.4 Governance

2.4.1 Input from stakeholders: 12

I. Introduction of a more sensitive troponin assay requires close communication between the laboratory, Emergency Department, cardiologists, and other relevant clinicians. Ideally decisions regarding assay performance, clinical cut-off, and turnaround time of results should be decided on by all the stakeholders.

II. Discussion regarding the likelihood of the impact of a more sensitive assay on the increase in patients with elevated troponin and patient classifications is essential prior to assay implementation.

2.4.2 Audit and review of assay performance and clinical use: 12

I. There should be a review process to ensure ongoing performance and correct laboratory use of troponin assays.

II. Review by the site doing the assay is required to determine its performance, e.g. imprecision close to the troponin cut-off concentration.

III. Laboratories should collaborate with the clinical team and regularly review and audit the clinical use of the assay.

3. Literature references

5. Thygesen K, Alpert JS, Jaffe AS, Simoons ML, Chaitman BR, White HD: the Writing Group on behalf of the Joint ESC/ACCF/AHA/WHF Task Force for the Universal Definition of Myocardial Infarction. Third Universal Definition of Myocardial Infarction. Circulation 2012;126:2020-35.

6. Hammarsten O, Fu ML, Sigurjonsdottir R, Petzold M, Said L, Landin-Wilhelmsen K, et al. Troponin T percentiles from a random population sample, emergency room patients and patients with myocardial infarction. Clin Chem 2012; 58:628-37.

7. Aldous S, Richards AM, George PM, Cullen L, Parsonage WA, Flaws D, et al. Comparison of new point-of-care troponin assay with high sensitivity troponin in diagnosing myocardial infarction. Int J Cardiol 2014;177:182-6.

8. Anderson JL, Adams CD, Antman EM, Bridges CR, Califf RM, Casey DE Jr, et al. ACC/AHA 2007 guidelines for the management of patients with unstable angina/non-ST-elevation myocardial infarction. J Am Coll Cardiol 2007;50:e1-e157.

9. Apple FS. A new season for cardiac troponin assays: it’s time to keep a scorecard. Clin Chem 2009; 55:1303-6.

10. Jones GRD, Panteghini M. Pre-analytical factors affecting troponin measurement. In: Tate J, Johnson R, Jaffe A, Panteghini M, editors. Laboratory and Clinical Issues Affecting the Measurement and Reporting of Cardiac Troponins: A Guide for Clinical Laboratories. Alexandria: AACB; 2012. p. 63-7.

11. Barth JH, Panteghini M, Bunk DM, Christenson RH, Katrukha A, Noble JE, et al. Recommendation to harmonize the units for reporting cardiac troponin results. Clin Chim Acta 2014;432:166.

12. Jones GRD, Finckh A, Wilson S. Introducing highly sensitive troponin into routine use – a worked example. In: Tate J, Johnson R, Jaffe A, Panteghini M, editors. Laboratory and Clinical Issues Affecting the Measurement and Reporting of Cardiac Troponins: A Guide for Clinical Laboratories. Alexandria: AACB; 2012. p. 97-102.

Page 10: AACB Guideline cover sheet

December 2016 9

COMPREHENSIVE DOCUMENT 1. Clinical Recommendations

I. This guideline provides recommendations for the use of POC Troponin test results in the management of ACS patients.

II. The following recommendations take into account the impact of the analytical performance of currently available Tn assays on their clinical use and interpretation. The recommendations regarding use of POC troponin assays are based on laboratory data showing that current POC assays are less analytically sensitive than most laboratory assays (labelled as ‘high sensitivity’ or ‘sensitive’ assays) for troponin.

III. There are many different assays available both for POC and laboratory-based troponin testing. Importantly, the results of various assays are NOT directly comparable and values CANNOT be transferred between assays unless there is evidence of such comparability.

IV. Clinical users of POC and laboratory-based troponin assays need to be aware of the limitations of the actual troponin assay used in the management of their patients in order to avoid adverse effects on patient management and for patient safety.

V. Troponin results must always be interpreted in the clinical context. Biomarkers are only one component of clinical information to be used with ECG(s) and clinical history for risk stratification according to clinical guidelines. 5

− Elevated troponin results alone do not rule in the diagnosis of ACS as elevated troponin can occur in non-ACS conditions (causes of non-ACS troponin elevations are shown in Table 1).

− Normal troponin results (at 0 and 3 hours after admission of patient to the Emergency Department [ED]) do not rule out ACS where there is a high clinical suspicion; in such cases further troponin testing may be indicated at 6 hours or at 12 hours post admission to ED

(slowly evolving AMIs may not elevate troponin until >12 hours after admission and unstable angina (UA) will not be detected). 6

− Less sensitive troponin assays may require a longer time period to show a clinically significant elevation in troponin.

VI. Serial troponin measurements should be performed for all patients presenting to an ED with symptoms suggestive of ACS, unless the patient has been reliably ACS symptom-free for 12 hours when a single sample is appropriate.

− The reason a second sample has NOT been tested should be clearly justified and documented. − The baseline should be the time of the first blood draw rather than from time of onset of

symptoms. − Recommended timing of samples using accelerated diagnostic pathways, using laboratory-

based highly sensitive assays, is obtaining the first sample at admission to ED and the second sample at 2-3 hours later.

− In order to safely discharge patients from the ED, it is recommended when using less sensitive (e.g. POC) troponin assays that a sample is collected at 6-8 hours after the first blood draw. Few POC troponin assays have been studied and reported as part of an accelerated, integrated assessment process that includes a combination of clinical assessment, ECGs and POC troponin. 7

VII. Serial testing should be performed using the same troponin assay and platform. We recommend to re-baseline the troponin value for individual patients if transported to a different hospital location where a different troponin assay is used. Example scenarios are: − Baseline troponin I done at presentation but second location uses troponin T assay, or vice

versa; − Baseline troponin I done by a POC device but second location uses a different laboratory-

based troponin I assay or one with higher analytical sensitivity;

Page 11: AACB Guideline cover sheet

December 2016 10

− Baseline troponin T done by a POC device but second location uses a different laboratory-based troponin T assay or one with higher analytical sensitivity.

At the second location collect a new baseline sample and a second sample at 2-3 hours (if testing with a highly sensitive troponin assay) and 6-8 hours (if testing with a less analytically sensitive assay) unless the patient has been symptom-free for >12 hours, when a single sample is appropriate.

VIII. In general quantitative troponin delta change has not been determined for use with POC assays as

POC assays are less precise at low troponin concentration (see local algorithms for examples).

XI. When a laboratory is used, results should be available as soon as possible, with a goal of within 60 minutes after the sample is received in the laboratory. Otherwise POC testing could be considered although the laboratory-based troponin assay is the preferred testing option when available.8

X. Local algorithms for use of POC troponin testing are shown in Appendix 2. Table 1: Causes of elevated cardiac troponins

• Acute Myocardial Infarction • Tachy or bradyarrhythmias • Aortic dissection or severe aortic valve

disease • Severe hypo or hypertension, e.g.

haemorrhagic shock, hypertensive emergency

• Acute or chronic heart failure • Hypertrophic cardiomyopathy • Coronary vasculitis, e.g. SLE, Kawasaki

syndrome • Coronary artery spasm, e.g. cocaine • Severe pulmonary embolism or pulmonary

hypertension • Dialysis dependent renal failure

• Cardiac Contusion or surgery • Rhabdomyolysis with cardiac

involvement • Myocarditis, severe sepsis • Cardiotoxic agents, e.g.

anthracyclines, CO poisoning • Severe burns affecting >30% body

surface • Severe acute neurological conditions,

e.g. stroke, trauma • Infiltrative diseases, e.g. amyloidosis,

sarcoidosis • Extreme exertion, e.g. marathon

running • Frequent defibrillator shocks

Page 12: AACB Guideline cover sheet

December 2016 11

2. Laboratory Recommendations

2.1 Analytical

I. It is the laboratory’s responsibility to inform clinicians the limitations of POC troponin vs. laboratory-based troponin assays in terms of analytical performance which impacts their clinical utility. − All current troponin POC assays are analytically less sensitive for the detection of the troponin

molecule and troponin is measurable above the limit of detection in <10% of the healthy population with POC devices. 9

− Studies using the i-STAT POC assay have shown that by lowering the decision cut-off to half of that of the 99th percentile troponin concentration of the reference population distribution that more AMIs are detected 13 (example of analytical sensitivity of i-STAT cTnI POC vs. AccuTnI and Architect cTnI lab-based assays is shown in Figure 1). 14

− Current POC assays are less diagnostically sensitive at earlier time points (<6 hours after the onset of symptoms) compared with highly sensitive and sensitive assays (see Figures 2-5).

− The increased analytical sensitivity of highly sensitive assays allows for the second sample to be collected at an earlier time point.

II. Table 2 shows the analytical characteristics of commercial POC troponin assays declared by the manufacturer. Generally POC troponin assays are “clinically usable” based on a scorecard performance approach related to the management of patients at risk of ACS. 9 This means that imprecision at the 99th percentile troponin concentration is >10 % CV and ≤20 % CV. For example:

− I-STAT POC Troponin I assay is clinically usable at a cut-off 0.04 µg/L with an imprecision of approx. 20 % CV reported (Pathology Queensland local data collated over several months).

− AQT90 Troponin T assay is also “clinically usable” at a cut-off of 0.02 µg/L cTnT with an imprecision of 11 % CV (NSW Health Pathology local data).

III. All troponin assays should be verified for analytical imprecision using the manufacturer’s QC and ideally, also, a low concentration QC sample or patient pool at close to the 99th percentile cTn concentration. Figures 6A-E show the analytical imprecision that can be obtained close to the 99th percentile cTn concentration for the highly sensitive hs-TnT and hs-TnI assays vs. POC troponin assays.

− Note that the laboratory-based highly sensitive Roche hs-TnT (Fig. 6A) and Abbott Architect hs-TnI (Fig. 6B) assays have lower imprecision than the three POC troponin assays, AQT90 Troponin I (Fig. 6C), AQT90 Troponin T (Fig. 6D) and i-STAT cTnI (Fig. 6E).

− Regarding imprecision of highly sensitive troponin assays over a multiple number of analysers, Saenger et al. reported hs-TnT precision on 11 Roche Elecsys and Cobas analysers over 21 days at 14 ng/L (2SD: 3 ng/L)15; Kavsak et al. reported hs-TnI precision for 4 Architect analysers over 12 months at 5 ng/L (2SD: 2 ng/L), and at 36 ng/L (2SD: 6 ng/L) [Table 3]. 16

− For POC troponin assays less long term precision information is available at concentrations close to or below the 99th percentile concentration. Local precision data using one AQT90 analyser are: cTnT at 44 ng/L (2SD: 7 ng/L) and cTnI at 36 ng/L (2SD: 7 ng/L) [Table 3]. Using 7 i-STAT analysers and 12 cartridge reagent lots over 14 weeks, the QC with a mean cTnI of 0.41 µg/L gave 2SD of 0.06 µg/L.

IV. Both laboratory-based and POC troponin assays may be affected by analytical interferences including heterophilic antibodies, immunoglobulin complexes, fibrin clots, etc. These may interfere with troponin measurement in some assays and cause false-positive or false-negative values.

V. POC troponin methods should be validated for clinical concordance, or at least compared against the local laboratory-based sensitive or highly sensitive troponin assay where patients may be referred. In this way POC results can be assessed against laboratory results, but not used for monitoring unless the differences are larger than the variation between assays.

Page 13: AACB Guideline cover sheet

December 2016 12

VI. Analytical performance specifications for POC troponins should ideally be driven by data coming from direct studies investigating the impact of analytical performance of the test on clinical outcomes related to the management of patients at risk of ACS; or from indirect outcome studies investigating the impact of analytical performance on clinical classification or decisions related to the management of ACS patients.

VII. Appendix 3 shows diagnostic sensitivity and specificity for troponin POC and high sensitive studies reported in recent literature.

2.2 Preanalytical (sample collection and processing)

I. Troponin testing should only be performed by operators who have attained the required competency standards and where the POC system is accredited within a quality framework.

I. False negative and false positive troponin results due to micro-clots and haemolysis are relatively common in Emergency Department samples. 10

− Note false positives are reported to occur in some POC studies. Ter Avesta et al. reported that 3/261 samples tested by AQT90 Troponin T assay were false positive. 17 Local studies confirm that false positive results occur on AQT 90, i-STAT and h232 analysers at a frequency of an average of 2 per 300 samples most likely due to pre-analytical sample handling (local data from NSW Health, Pathology Queensland, iCCnet SA).

− Refer to Appendix 4 for case studies which are available at the AACB website at: http://www.aacb.asn.au/resources/recommendations-for-use-of-troponin-poc

II. A general approach to reducing micro-clots is: Mix blood tubes immediately after collection to avoid formation of micro-clots that may not be visible on testing but which may cause incorrect troponin values.

III. Avoid haemolysed samples by:

− Use of trained collectors − Do not collect blood from the site of insertion of an intravenous cannula − Do not use very fine bore needles − Do not collect blood into syringe and then transfer it into a vacuum tube.

2.3 Postanalytical

2.3.1 Test name:

I. The naming of generations of troponin assays has been inconsistent and largely market-driven. Highly sensitive assays measure the same protein as the sensitive assays, and even more sensitive cardiac troponin assays will certainly be developed, thus making any ‘high sensitivity’ designation obsolete.

II. For the purpose of reporting of results, if the analytical performance of a new assay is significantly different from a previous troponin assay the test name should indicate this, e.g. ‘Troponin T – Highly sensitive’.

III. We recommend there is a general consensus in Australia regarding troponin test names for the same manufacturer’s assay to allow for harmonised reporting of troponin in the electronic medical record.

IV. Troponin results released from POC devices should be clearly differentiated from troponin results reported from the laboratory using an assay with different analytical sensitivity (e.g. ‘Troponin T – POC’ vs ‘Troponin T – Highly sensitive’).

Page 14: AACB Guideline cover sheet

December 2016 13

2.3.2 Reporting units:

I. The international consensus is that troponin should be reported in ng/L (or pg/mL) unitage and expressed as whole numbers. 11 In Australasia highly sensitive troponin assays generally use units of ng/L, whereas results from less sensitive assays have usually been reported in µg/L to two or more decimal points. As a consequence, the reporting of these more sensitive assays produces results that may be a thousand-fold different compared to results by other less sensitive assays, and thus has the potential for confusion and medical error, e.g. 4000 ng/L is 4.000 µg/L and 4 ng/L is 0.004 µg/L.

II. To improve patient safety we recommend that the clinical and laboratory communities standardise on troponin units of measurement in Australia namely “ng/L” and report values in whole numbers. This will allow for harmonised reporting of troponin in the electronic medical record.

2.3.3 Education to users about assay: 12

I. Various healthcare professionals request and measure troponin. It is essential that there is a comprehensive education process communicated to all users.

II. Supporting information detailing troponin assay utility and comparing the local assay to others in use should be provided.

III. General educational materials including a slide presentation of the recommendations and case studies are available on the AACB website at: http://www.aacb.asn.au/resources/recommendations-for-use-of-troponin-poc

2.4 Governance

2.4.1 Input from stakeholders: 12

I. Introduction of a more sensitive troponin assay requires close communication between the laboratory, Emergency Department, cardiologists, and other relevant clinicians. Ideally decisions regarding assay performance, clinical cut-off, and turnaround time of results should be decided on by all the stakeholders.

II. Discussion regarding the likelihood of the impact of a more sensitive assay on the increase in patients with elevated troponin and patient classifications is essential prior to assay implementation.

2.4.2 Audit and review of assay performance and clinical use: 12

I. There should be a review process to ensure ongoing performance and correct laboratory use of troponin assays.

II. Review by the site doing the assay is required to determine its performance, e.g. imprecision close to the troponin cut-off concentration.

III. Laboratories should collaborate with the clinical team and regularly review and audit the clinical use of the assay.

Page 15: AACB Guideline cover sheet

December 2016 14

Table 2: Analytical characteristics of commercial Point-of-Care troponin (cTn) assays declared by the manufacturer (modified from IFCC troponin table) 18 Commercial cTn Assays: Company/platform(s)/assay

LoB (µg/L)

LoD (µg/L)

99th % (µg/L)

% CV at 99th per.

10 % CV (µg/L)

Acceptance criterion a

Abbott i-STAT cTnI 0.02 0.08 16.5 0.10 Clinically usable

Alere Triage Cardio 3 cTnI 0.002 0.01 0.02 17.0 0.04 Clinically usable

Radiometer AQT90 TnI 0.0095 0.023 12.3 0.027 Clinically usable

Radiometer AQT90 TnT 0.0080 0.017 15.2 0.026 Clinically usable

Roche Cobas h232 TnT 0.05 NAD b NA NA Clinically usable

Commercial cTn Assays: Company/platform(s)/assay

LoB (ng/L)

LoD (ng/L)

99th % (ng/L)

% CV at 99th per.

10 % CV (ng/L)

Acceptance criterion

Abbott i-STAT cTnI Analyser values are only given in µg/L not ng/L

Alere Triage Cardio 3 cTnI 2 10 22 17.0 37 Clinically usable

Radiometer AQT90 TnI 9.5 23 12.3 27 Clinically usable

Radiometer AQT90 TnT 8 17 15.2 26 Clinically usable

Roche Cobas h232 TnT 50 NAD b NA NA Clinically usable

LoB, limit of blank; LoD, limit of detection determined according to CLSI EP17-A guideline; NAD, the 99th percentile (per.) concentration of the troponin value distribution of a reference population is indeterminate; NA, data are not available; a Acceptance criterion: % CV at 99th percentile: guideline acceptable ≤10 %; clinically usable >10 to ≤20 %; not acceptable >20 % 8; b Cobas h232 TnT next generation assay to be introduced in 2016 has a LoD of 40 ng/L and CV of 10.7 % at 63 ng/L (personal communication, RT, PS). NB: The international consensus is that troponin should be reported in ng/L (or pg/mL) unitage and expressed as whole numbers. This issue requires further harmonisation in Australasia.

Page 16: AACB Guideline cover sheet

December 2016 15

Figure 1: Analytical sensitivity of POC troponin I assay (i-Stat) and sensitive troponin I assays (Beckman Coulter AccuTnI and Abbott Architect cTnI). A new 0.04 µg/L (40 ng/L) i-Stat cut-off is presented by the horizontal lines. The vertical lines present the 99th percentile cut-off values of the comparative methods; N=172 (data are from Pathology Queensland study). 14 Even at the lowered i-Stat cut-off, a number of samples will have a negative i-Stat result and a positive laboratory cTnI (i.e. false negatives by POC testing; also see Schneider et al. 13).

Page 17: AACB Guideline cover sheet

December 2016 16

Figure 2: Serial troponin measurements by POC assays compared with high sensitive hs-TnT assay and diagnostic sensitivity for 806 patients presenting to Chest Pain Centre (South Australia) with symptoms suggestive of ACS. A positive result for AMI was defined as the baseline and/or the second time point giving a result above the 99th percentile of the assay.

Figure 3: Serial troponin measurements by POC assays compared with high sensitive hs-TnT assay and diagnostic specificity for 806 patients presenting to Chest Pain Centre (South Australia) with symptoms suggestive of ACS. A positive result for AMI was defined as the baseline and/or the second time point giving a result above the 99th percentile of the assay. Note that the diagnostic specificity and Positive Predictive Value are higher for hs-TnT when a cTnT value above the 99th percentile combined with a delta cTnT change of ≥7 ng/L is used, in keeping with the Universal Definition of AMI guidelines, which requires there be a rise and/ or fall of cTn over time (see Appendix 3).

Page 18: AACB Guideline cover sheet

December 2016 17

Figure 4: Serial troponin measurements by POC assays compared with high sensitive hs-TnT assay. Negative Predictive Value for 806 patients presenting to Chest Pain Centre (South Australia) with symptoms suggestive of ACS. A positive result for AMI was defined as the baseline and/or the second time point giving a result above the 99th percentile of the assay.

Figure 5: Serial troponin measurements by POC assays compared with high sensitive hsTnT assay Positive Predictive Value for 806 patients presenting to Chest Pain Centre (South Australia) with symptoms suggestive of ACS. A positive result for AMI was defined as the baseline and/or the second time point giving a result above the 99th percentile of the assay. Note that the diagnostic specificity and Positive Predictive Value are higher for hs-TnT when a cTnT value above the 99th percentile combined with a delta cTnT change of ≥7 ng/L is used, in keeping with the Universal Definition of AMI guidelines, which requires there be a rise and/ or fall of cTn over time (see Appendix 3).

Page 19: AACB Guideline cover sheet

December 2016 18

Figures 6A-E: Precision of high sensitive troponin (hs-cTn) and POC cTn assays determined close to the 99th percentile cTn concentration. Laboratory data and manufacturer’s data are plotted vs. the analytical imprecision (2SD) determined from replicate measurements of patient sample pools. Depending on the cTn concentration of pools, linear regression slopes may vary between laboratory and manufacturer’s data; only data close to the 99th percentile cTn concentration are shown. Dotted lines represent the imprecision (2SD) at the 99th percentile determined from the laboratory data. Note that the laboratory-based high sensitive Roche hs-TnT (Fig. 6A) and Abbott Architect hs-TnI (Fig. 6B) assays have lower analytical imprecision than the three POC cTn assays, AQT90 Troponin I (Fig. 6C), AQT90 Troponin T (Fig. 6D) and i-STAT cTnI (Fig. 6E). POC cTn assays when tested by routine laboratories gave comparable precision to that specified by the manufacturer. i.e., they are ‘fit for purpose’.

Roche hs-TnT Assay Abbott hs-TnI Assay Manufacturer’s data: 99th percentile 14.0 ng/L; imprecision (2SD) is 2.8 ng/L. Manufacturer’s data: 99th percentile 26.2 ng/L; imprecision (2SD) is 2.1 ng/L. Laboratory data: imprecision (2SD) at 14.0 ng/L is 2.5 ng/L. Laboratory data: imprecision (2SD) at 26.2 ng/L is 2.3 ng/L. On repeat analysis, 14.0 ng/L cTnT by hs-TnT is from 11.5 to 16.5 ng/L. On repeat analysis, 26.2 ng/L cTnI by hs-TnI is from 23.9 to 28.5 ng/L. Assay is fit for purpose. Assay is fit for purpose.

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

0 5 10 15 20 25 30 35 40 45 50Elecsys hs-TnT (ng/L)

Impr

ecis

ion

2SD

(ng/

L)

Laboratory data

Manufacturer's data

Linear (Laboratory data)

Linear (Manufacturer's data)

A

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

0 5 10 15 20 25 30 35 40 45 50

Architect hs-TnI (ng/L)

Impr

ecis

ion

2SD

(ng/

L)

Laboratory data

Manufacturer's data

Linear (Laboratory data)

Linear (Manufacturer's data)

B

Page 20: AACB Guideline cover sheet

December 2016 19

Radiometer AQT90 Troponin T Assay Radiometer AQT90 Troponin I Assay Abbott i-STAT Troponin I Assay Manufacturer’s data: 99th percentile 17 ng/L; imprecision (2SD) is 5 ng/L. Manufacturer’s data: 99th percentile 23 ng/L; imprecision (2SD) is 6 ng/L. Manufacturer’s data: 99th percentile 0.08 µg/L; imprecision (2SD) is 0.03 µg/L. Laboratory data: imprecision (2SD) at 20 ng/L cTnT is 4 ng/L. Laboratory data: imprecision (2SD) at 23 ng/L is 5 ng/L. Laboratory data: imprecision (2SD) at 0.04 µg/L cTnI is 0.01 ng/L. On repeat analysis, 20 ng/L cTnT is from 16 to 24 ng/L. On repeat analysis, 23 ng/L cTnI is from 18 to 28 ng/L. On repeat analysis, 0.04 µg/L cTnI is from 0.03 to 0.05 µg/L. Assay is fit for purpose. Assay is fit for purpose. Assay is fit for purpose.

Note that the present i-STAT assay reports only in µg/L to 2 decimal places.

0

1

2

3

4

5

6

7

8

9

10

0 10 20 30 40 50

AQT90 TnT (ng/L)

Impr

ecis

ion

2SD

(ng/

L)

Laboratory data

Manufacturer's data

Linear (Laboratory data)

Linear (Manufacturer's data)

C

0

1

2

3

4

5

6

7

8

9

10

0 5 10 15 20 25 30 35 40 45 50AQT90 TnI (ng/L)

Impr

ecis

ion

2SD

(ng/

L)

Laboratory data

Manufacturer's data

Linear (Laboratory data)

Linear (Manufacturer's data)

D

0

0.01

0.02

0.03

0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 0.16 0.18 0.2 0.22 0.24

i-STAT cTnI (µg/L)

Impr

ecis

ion

2SD

(µg/

L)

Laboratory data

Manufacturer's data

Linear (Laboratory data)

Linear (Manufacturer's data)

E

Page 21: AACB Guideline cover sheet

December 2016 20

Table 3. Long term imprecision for cardiac troponin assays

Assay (unit) Manufacturer/analyser(s)

Precision for serum pools or manufacturers’ quality controls (QC) Notes Reference

hs-TnT (ng/L) Roche Elecsys 2010/cobas e 411 /E 170 / cobas e 601

LOW serum pool: CV range at 14.2 ng/L: 2.9% - 10.7% (2SD: 0.8 - 3.0 ng/L) - E 2010 (N=5 analysers) CV: 6.0% - 10.7% (2SD: 1.7-3.0 ng/L) - E 170 (N=3 analysers) CV: 2.9% - 5.8% (2SD: 0.8-1.7 ng/L) - e 411 (N=2 analysers) CV: 5.7% - 9.3% (2SD: 1.6-2.6 ng/L) - e 601 (N=1 analyser) CV: 3.9% (2SD: 1.1 ng/L)

Multi-site study of 8 labs and 11 analysers; up to 3 reagent lots used; study conducted over 21 days with 4 calibrations

Saenger et al. 15

hs-TnT (ng/L) Roche Elecsys 2010

LOW QC (N=91): Mean 26.5 (2SD: 3.7) ng/L; CV 7.0% HIGH QC (N=91): Mean 2053 (2SD: 193) ng/L; CV 4.7%

One site; more than one reagent lot and calibrator lot used

ICCnet data

hs-TnI (ng/L) Abbott Architect 4100/ 8200/ 16200 STAT

LOW serum pool (N= 495 - 831): - 4100 (N=1 analyser): Mean 4.7 (2SD: 1.8) ng/L; CV 19% - 8200 (N=1 analyser): Mean 4.2 (2SD: 1.3) ng/L; CV 15% - 16200 (N=2 analysers): Mean 5.0 (2SD: 1.5) ng/L; CV 15% MEDIUM serum pool (N= 498 - 842): - 4100 (N=1 analyser): Mean 36.8 (2SD: 5.9) ng/L; CV 8% - 8200 (N=1 analyser): Mean 36.0 (2SD: 5.8) ng/L; CV 8% - 16200 (N=2 analysers): Mean 36.0 (2SD: 5.0) ng/L; CV 7%

Three sites and 4 analysers; multiple reagent and calibrator lots used; precision data collated over 12 months

Kavsak et al. 16

cTnT (ng/L) Radiometer AQT90 FLEX TnT

LOW QC (N=143): Mean 43.5 (2SD: 6.8) ng/L; CV 7.9% MEDIUM QC (N=144): Mean 643 (2SD: 67) ng/L; CV 5.2% HIGH QC (N=137): Mean 9947 (2SD: 1200) ng/L; CV 6.0%

One site; more than one reagent lot and calibrator lot used

ICCnet data

cTnI (ng/L) Radiometer AQT90 FLEX TnI

LOW QC (N=140): Mean 36.0 (2SD: 7.2) ng/L; CV 10.1% MEDIUM QC (N=132): Mean 303 (2SD: 28) ng/L; CV 4.6% HIGH QC (N=132): Mean 1364 (2SD: 121 ng/L); CV 4.4%

One site; more than one reagent lot and calibrator lot used

ICCnet data

cTnI (µg/L) Abbott i-STAT

QC (N=21): Mean 0.41 (2SD: 0.06) µg/L; CV 7.2% Seven sites; 12 cartridge lots used; QC assayed at each site 3 times over 14 weeks

Pathology Queensland data

Page 22: AACB Guideline cover sheet

December 2016 21

Appendix 1: Methods for developing the recommendations

Membership of the AACB Troponin Point-of-Care Working Party: Rosy Tirimacco, Jill Tate, Paul Simpson, Andrea Rita Horvath, Louise Cullen, Fernando San Gil, Cameron Martin, Gus Koerbin, Phil Tideman The recommendations have been conceived through a consensus process by a group of experts who have included an Emergency Physician (LC), a Cardiologist (PT), a Pathologist (RH), and Scientists (GK, CM, FSG, PS, JT, RT) who routinely analyse and use troponin testing at the POC and in the routine laboratory. There was no IVD Industry involvement in deriving the recommendations. The recommendations were developed after reviewing the relevant evidence available in the literature up to December 2015 (Appendix 3 shows the diagnostic sensitivity and specificity of troponin POC studies from the recent literature) and by examining the results of a recent 2014-2015 clinical trial of patients presenting to the Chest Pain Centre in South Australia. This study investigated the clinical utility of POC troponin (Roche Cobas h232, Radiometer AQT90 Troponin T and Troponin I) versus high sensitivity laboratory troponin (hs-TnT). Strength of recommendations: The local clinical trial data verify the findings of the literature and add strength to the recommendations for serial testing and the timing of samples for current POC assays. The literature and trial data on POC troponin testing strongly support the recommendation that using current POC assays patients presenting with symptoms of ACS cannot be excluded at early time points, unlike the use of high sensitive troponin assays (see Figures 2-5). Next update of document: The next update is due in 3 years (2018) or earlier if new evidence becomes available that warrants review and modification of the recommendations. Conflicts of Interest: LC – Has received research support and speakers fees from Alere, Abbott Diagnostics, Beckman Coulter, Roche, Radiometer Pacific, and Siemens. RH – Roche Diagnostics and Radiometer Australia have partially funded the troponin trial performed in NSW. GK – Has received research support from Abbott Diagnostics. CM – none FSG – none PS – Roche Diagnostics and Radiometer Australia funded the troponin trial performed in SA. JT – none PT - Roche Diagnostics and Radiometer Australia funded the troponin trial performed in SA. RT - Roche Diagnostics and Radiometer Australia funded the troponin trial performed in SA.

Page 23: AACB Guideline cover sheet

December 2016 22

The following organisations and individuals have reviewed and/or commented on the public consultation draft of the recommendations:

­ Australasian Association of Clinical Biochemists (AACB) ­ Royal College of Pathologists of Australasia (RCPA) ­ Australasian College of Emergency Medicine (ACEM) – Council of Advocacy, Practice and

Partnerships ­ The Integrated Cardiovascular Clinical Network CHSA ­ New South Wales Health Pathology ­ Pathology Queensland ­ Dr Grahame Caldwell (Private Pathology) ­ Dr Matthew Bragg (ED Director, Bowral and District Hospital, NSW Cardiac Network) ­ The Cardiac Society of Australia and New Zealand (CSANZ) – Quality Standards Committee

The comments received have been reviewed and answered by the guideline development group (accessible at: http://www.aacb.asn.au/resources/recommendations-for-use-of-troponin-poc) and the final recommendations include all relevant suggestions. The final draft of the recommendations has been endorsed by the above organisations.

DISCLAIMER:

The Australasian Association of Clinical Biochemists (AACB) and the Royal College of Pathologists of Australasia (RCPA) have developed these recommendations as a tool to assist laboratories in the reporting and management of high risk results. Each recommendation includes indicators of the strength of the recommendation (“needs to”, “should” and “may”), which reflect the consensus views of the expert working party. The use of these recommendations is subject to the judgement of individual laboratory practices.

The AACB and RCPA have made all reasonable efforts to ensure the quality of the recommendations. However subject to any warranties, terms or conditions which may be implied by law and which cannot be excluded, the recommendations are provided on an "as is" basis. The AACB and RCPA do not warrant or represent that the recommendations are complete, accurate, error-free, or up-to-date for all possible clinical scenarios or local circumstances. Users are responsible for evaluating the suitability, accuracy, currency, completeness and fitness for purpose of the recommendations. Any decision not to follow these recommendations and the settings where this applies should be documented by the laboratory.

Except as set out in this paragraph, the AACB and RCPA exclude: (i) all warranties, terms and conditions relating in any way to; and (ii) all liability (including for negligence) in respect of any loss or damage (including direct, special, indirect or consequential loss or damage, loss of revenue, loss of expectation, unavailability of systems, loss of data, personal injury or property damage) arising in any way from or in connection with the recommendations or any use thereof. Where any statute implies any term, condition or warranty in connection with the provision or use of the recommendations, and that statute prohibits the exclusion of that term, condition or warranty, then such term, condition or warranty is not excluded. To the extent permitted by law, the AACB and RCPA's liability under or for breach of any such term, condition or warranty is limited to the resupply or replacement of services or goods.

Page 24: AACB Guideline cover sheet

December 2016 23

Appendix 2: Local Algorithms for Point-of-Care Troponin Testing The intent of the local algorithms is to provide information and they are only meant as a guide. The Working Party notes that there are differences between algorithms at present, for example reporting units, decision limits, etc. ICCnet (Country Health South Australia) and NSW Health Pathology

>17 ng /L ≤ 17 ng /L

Time >6 hours: Repeat Troponin T

High - risk ACS

Troponin T on presentation

Causes of elevated cardiac troponins • Acute Myocardial Infarction • Tachy or bradyarrhythmias • Aortic dissection or severe aortic valve disease • Severe hypo or hypertension, eg haemorrhagic shock, hypertensive emergency • Acute or chronic heart failure • Hypertrophi cardiomyopathy • Coronary vasculitis , eg SLE, Kawasaki synd. • Coronary artery spasm, eg cocaine • Severe pulmonary embolism or pulmonary hypertension • Dialysis dependent renal failure

• Cardiac Contusion or surgery • Rhabdomyolysis with cardiac involvement • Myocarditis , severe sepsis • Cardiotoxic agents, eg anthracyclines , CO poisoning • Severe burns affecting >30% body surface • Severe acute neurological conditions, eg stroke, trauma • Infiltrative diseases, eg amyloidosis , sarcoidosis • Extreme exertion, eg marathon running • Frequent defibrillator Shocks

AQT Troponin T Testing Protocol To be used with the Acute Coronary Syndrome Assessment and Treatment Guidelines

>17 ng /L

Clinical Assessment

≤ 17 ng /L

Page 25: AACB Guideline cover sheet

December 2016 24

ICCnet (Country Health South Australia)

>23 ng /L ≤ 23 ng /L

Time >6 hours: Repeat Troponin I

High - risk ACS

Troponin I on presentation

Causes of elevated cardiac troponins • Acute Myocardial Infarction • Tachy or bradyarrhythmias • Aortic dissection or severe aortic valve disease • Severe hypo or hypertension, eg haemorrhagic shock, hypertensive emergency • Acute or chronic heart failure • Hypertrophic cardiomyopathy • Coronary vasculitis , eg SLE, Kawasaki synd. • Coronary artery spasm, eg cocaine • Severe pulmonary embolism or pulmonary hypertension • Dialysis dependent renal failure

• Cardiac Contusion or surgery • Rhabdomyolysis with cardiac involvement • Myocarditis , severe sepsis • Cardiotoxic agents, eg anthracyclines , CO poisoning • Severe burns affecting >30% body surface • Severe acute neurological conditions, eg stroke, trauma • Infiltrative diseases, eg amyloidosis , sarcoidosis • Extreme exertion, eg marathon running • Frequent defibrillator Shocks

AQT Troponin I Testing Protocol To be used with the Acute Coronary Syndrome Assessment and Treat ment Guidelines

>23 ng /L

Clinical Assessment

≤ 23 ng /L

Page 26: AACB Guideline cover sheet

December 2016 25

Pathology Queensland

>0.04 µg/L ≤0.04 µg/L

Repeat i-STAT TnI: Collect second sample at 3-6 hours after presentation (Time 2)

High risk for AMIClinical assessment and further TnI

testing if AMI suspected

TnI on presentation (Time 1)

TnI >0.04 µg/L WITH a typical rise and/or fall of TnI.“Change (Delta TnI)”:

TnI {Time 2 – Time 1} is≥0.02 µg/L at ≤0.10 µg/L, or

>20% at >0.10 µg/L

High-risk stratification

No

Notes about the i-STAT method• At the 0.04 µg/L cut off the i-STAT TnI has a precision of approximately 16.5% CV.• At the 0.08 µg/L cut off the CV is verified 10% by the manufacturer.• At the current time the i-STAT instrument can only report in µg/L.• “Change (Delta TnI)” consistent with an acute change in TnI is: >20% at above 0.10 µg/L TnI and ≥0.02 µg/L at 0.10 µg/L and below (Pretorius CJ et al. Eur Heart J Acute Cardiovasc Care. 2014 Jun;3(2):149-57).

Cameron Martin – Pathology Queensland -Statewide Point of Care 07 3646 8426

Causes of elevated cardiac troponins•Acute Myocardial Infarction• Tachy or bradyarrhythmias•Aortic dissection or severe aortic valve disease• Severe hypo or hypertension, eg haemorrhagicshock, hypertensive emergency•Acute or chronic heart failure• Hypertrophic cardiomyopathy• Coronary vasculitis, eg SLE, Kawasaki synd.• Coronary artery spasm, eg cocaine• Severe pulmonary embolism or pulmonary hypertension• Dialysis dependent renal failure

• Cardiac Contusion or surgery• Rhabdomyolysis with cardiac involvement• Myocarditis, severe sepsis• Cardiotoxic agents, eg anthracyclines, CO poisoning• Severe burns affecting >30% body surface• Severe acute neurological conditions, eg stroke, trauma• Infiltrative diseases, eg amyloidosis, sarcoidosis• Extreme exertion, eg marathon running• Frequent defibrillator Shocks

Abbott i-STAT Troponin I Testing ProtocolTo be used with Acute Coronary Syndrome Assessment and Treatment Guidelines

http://qheps.health.qld.gov.au/hssa/pathology/testing/point-of-care.htm

AMI not likely

Yes

Page 27: AACB Guideline cover sheet

December 2016 26

ICCnet (Country Health South Australia)

>100 ng /L 50 - 100 ng /L

Time 8 - 12 hours: Repeat Troponin T

High - risk ACS Clinical Assessment

Troponin T on presentation

<50 ng /L

>100 ng /L <50 ng /L

Causes of elevated cardiac troponins • Acute Myocardial Infarction • Tachy or bradyarrhythmias • Aortic dissection or severe aortic valve disease • Severe hypo or hypertension, eg haemorrhagic shock, hypertensive emergency • Acute or chronic heart failure • Hypertrophic cardiomyopathy • Coronary vasculitis , eg SLE, Kawasaki synd. • Coronary artery spasm, eg cocaine • Severe pulmonary embolism or pulmonary hypertension • Dialysis dependent renal failure

• Cardiac Contusion or surgery • Rhabdomyolysis with cardiac involvement • Myocarditis , severe sepsis • Cardiotoxic agents, eg anthracyclines , CO poisoning • Severe burns affecting >30% body surface • Severe acute neurological conditions, eg stroke, trauma • Infiltrative diseases, eg amyloidosis , sarcoidosis • Extreme exertion, eg marathon running • Frequent defibrillator Shocks

Cobas h232 Troponin T Testing Protocol To be used with the Acute Coronary Syndrome Assessment and Treat ment Guidelines

50 - 100 ng /L

Consider for high - risk ACS. Further

investigation required

Page 28: AACB Guideline cover sheet

December 2016 27

Appendix 3: Diagnostic sensitivity and specificity for Point-of-Care (POC) and high sensitive troponin (cTn) studies from the recent literature

cTn Assay Subject group Sample collection timing

cTn decision limit

Decision criteria Outcome parameter

Study sensitivity (95% CI) (%)

Study specificity (95% CI) (%)

Study PPV (95% CI) (%)

Study NPV (95% CI) (%)

Study

Alere Cardio3 POC cTnI

960 patients presenting to ED, non-traumatic chest pain; 220 (22.9%) AMI (196 NSTEMI)

0h and 2h post presentation

99th perc. = 20 ng/L (no CV given)

Architect cTnI 28 ng/L; 0h & 6-12h; ADP clinical pathway Cardio3 0h vs 2h

MACE 30d 0h: 87.7 (83.6-91.1) 2h: 93.6 (89.9-96.2)

0h: 93.1 (91.9-94.1) 2h: 90.2 (89.0-90.9)

0h: 79.1 (75.3-82.1) 2h: 73.8 (70.9-75.9)

0h: 96.2 (95.0-97.3) 2h: 98.0 (96.7-98.8)

Aldous et al. 7

As above then Cardio3 2h +ECG + TIMI score cut-off of 0

100 (98.0-100)

19.0 (18.4-19.0)

28.1 (27.5-28.1 )

100 (96.7-100)

As above then Cardio3 2h +ECG + TIMI score cut-off of 1

98.3 (95.4-99.4)

37.3 (36.5-37.7)

33.1 (32.2-33.5)

98.6 (96.2-99.5)

Alere Cardio3 POC cTnI

858 (of 1107) patients presenting to ED with suspected ACS. Serial measurements were tested in 18 EDs (MIDAS study - 9.6% AMI)

0h, 90 min, 3h, 6h (680 of 858 patients had complete data across all 4 times)

99th perc. = 0.05 ug/L

cTnI and cTnT assays with varying local decision limits; AHA/ACC criteria for AMI

None given 3h: 84.1 (74.4-91.3) 6h: 87.5 (77.9-93.3)

3h: 93.4 (91.4-95.1) 6h: 92.6 (90.2-94.4)

3h: 57.5 (48.1-66.5) 6h: 58.3 (48.9-67.2)

3h: 98.2 (97.0-99.1) 6h: 98.4 (97.0-99.2)

Diercks et al. 23

Abbott Architect hs-TnI

960 patients presenting to ED, non-traumatic chest pain; 220 (22.9%) AMI (196 NSTEMI)

0h and 2h post presentation

99th perc. = 26.2 ng/L (4.7% CV)

Architect cTnI 28 ng/L; 0h & 6-12h; ADP clinical pathway hs-TnI 0h vs 2h

MACE 30d 0h: 90.0 (86.1-93.1) 2h: 95.0 (91.5-97.3)

0h: 93.9 (92.8-94.8) 2h: 92.5 (91.4-93.1)

0h: 81.5 (77.9-84.3) 2h: 78.9 (76.0-80.7)

0h: 96.9 (95.7-97.9) 2h: 98.4 (97.3-99.1)

Aldous et al. 7

Architect cTnI 28 ng/L; 0h & 6-12h; ADP clinical pathway hs-TnI 2h +ECG + TIMI score cut-off of 0

100 (98.0-100)

19.7 (19.1-19.7)

28.2 (27.7-28.2 )

100 (96.8-100)

Architect cTnI 28 ng/L; 0h & 6-12h; ADP clinical pathway hs-TnI 2h +ECG + TIMI score cut-off of 1

98.7 (96.0-919.7)

38.9 (38.0-39.2)

33.8 (32.9-34.1)

99.0 (96.8-99.7)

Medience Pathfast (POC) cTnI

169 patients presenting with symptoms of ACS; 19 (11.2%) AMI

0h, 3h and 6h post-presentation

99th perc. = 29 ng/L (5% CV)

MI according to 2007 Universal Definition of MI b; clinical symptoms of ischaemia; 1 cTn >99th perc. (Vitros cTnI

None given 0h: 53 (29-76) 3h: 89 (67-100) 6h: 95 (74-100) Max: 95 (74-100)

0h: 86 (79-91) 3h: 82 (84-94) 6h: 82 (75-88) Max: 78 (71-84)

0h: 32 (17-51) 3h: 39 (28-66) 6h: 40 (26-56) Max: 35 (22-50)

0h: 93 (88-97) 3h: 98 (91-98) 6h: 99 (96-100) Max: 99 (95-100)

Palamalai et al. 19

Page 29: AACB Guideline cover sheet

December 2016 28

assay 34 ng/L) Radiometer AQT90 (POC) cTnI

169 patients presenting with symptoms of ACS; 19 (11.2%) AMI

0h, 3h and 6h post-presentation

99th perc. = 23 ng/L (17% CV)

MI according to 2007 Universal Definition of MI b; clinical symptoms of ischaemia; 1 cTn >99th perc. (Vitros cTnI assay 34 ng/L)

None given 0h: 26 (9-51) 3h: 63 (38-84) 6h: 63 (38-84) Max: 68 (43-97)

0h: 93 (87-96) 3h: 91 (86-95) 6h: 91 (85-95) Max: 89 (82-93)

0h: 31 (11-59) 3h: 48 (28-69) 6h: 46 (27-67) Max: 43 (25-63)

0h: 91 (85-95) 3h: 95 (90-98) 6h: 95 (90-98) Max: 96 (91-98)

Palamalai et al. 19

Radiometer AQT90 (POC) cTnI

458 patients referred with chest pain by GP or ambulance service; suspected of ACS; 104 (23%) AMI

0h, 6-9h, 12-24h post-presentation

99th perc. = 39 ng/L (no CV given)

MI according to 2007 Universal Definition of MI b; rise and or fall of cTnT >30 ng/L; clinical symptoms, ECG

None given 1st sample: 58 (47-69) 2nd sample: 85 (75-92)

1st sample: 94 (91-96) 2nd sample: 91 (87-94)

1st sample: 71 (58-81) 2nd sample: 71 (61-80)

1st sample: 90 (86-93) 2nd sample: 96 (93-98)

Hjortshoj et al. 20

Radiometer AQT90 (POC) cTnI

806 patients (39 with AMI)

0h, 2h, 3h, 4h, 6h, 8h and 12h post-presentation

99th perc. = 23 ng/L

MI according to 2007 Universal Definition of MI b; clinical symptoms of ischaemia; hsTnT >14 ng/L; clinical history & further investigations (echocardiography, stress testing, coronary angiography, ultrasound, x-ray)

Followed up at 7d, 30d, & 12mo for CV events, significant bleeding

2h: 89.7 (75.8-97.1) 3h: 89.5 (75.2-97.1) 4h: 94.4 (81.3-99.3) 6h: 97.1 (84.7-99.9) 8h: 100.0 (89.7-100.0) 12h: 100.0 (89.1-100.0)

2h: 91.1 (88.8-93.1) 3h: 91.1 (88.7-93.1) 4h: 91.0 (88.6-93.1) 6h: 91.6 (89.1-93.7) 8h: 90.3 (87.7-92.7) 12h: 89.8 (86.6-92.5)

2h: 35.0 (25.7-45.2) 3h: 35.8 (26.2-46.3) 4h: 36.6 (26.8-47.2) 6h: 39.8 (29.2-51.1) 8h: 39.5 (29.2-50.7) 12h: 41.6 (30.4-45.4)

2h: 99.4 (98.5-99.8) 3h: 99.4 (98.4-99.8) 4h: 99.7 (98.9-100.0) 6h: 99.8 (99.0-100.0) 8h: 100.0 (99.2-100.0) 12h: 100.0 (99.1-100.0)

South Australian iCCnet Study

Abbott i-STAT (POC) cTnI

169 patients presenting with symptoms of ACS; 19 (11.2%) AMI

0h, 3h and 6h post-presentation

99th perc. = 80 ng/L (16% CV)

MI according to 2007 Universal Definition of MI b; clinical symptoms of ischaemia; 1 cTn >99th perc. (Vitros cTnI assay 34 ng/L)

None given 0h: 32 (13-57) 3h: 68 (43-87) 6h: 68 (43-87) Max: 74 (49-91)

0h: 92 (86-96) 3h: 90 (84-94) 6h: 91 (86-95) Max: 88 (82-93)

0h: 33 (13-59) 3h: 46 (28-66) 6h: 50 (30-70) Max: 44 (26-62)

0h: 91 (86-95) 3h: 96 (91-98) 6h: 96 (91-98) Max: 96 (92-99)

Palamalai et al. 19

Abbott i-STAT (POC) cTnI and Radiometer AQT90 (POC) cTnI

Comparative study of 195 patients presenting to a suburban ED in Melbourne by i-STAT and AQT90 cTnI; 13 (6.7%) AMI

0h for 195 patients; later timed sample for 6 of these patients

Lowered i-STAT cut-off to 50% of 99th perc. to 40 ng/L; Lowered AQT90 cTnI cut-off from 24 to 12 ng/L

Architect cTnI 28 ng/L; clinical symptoms; ECG; Universal Definition of MI b

Clinical audit

i-STAT cTnI: True Positive rate (sensitivity) increased from 34% to 80.9% by changing the cut-off from 80 to 40 ng/L. False Positive rate only increased from 0% to 3.7%. AQT90 cTnI: True Positive rate (sensitivity) increased from 51% to 76.5% by changing the cut-off from 24 to 12 ng/L. False Positive rate only increased from 3.7% to 7.1%. Concluded: used reporting below 99th percentile of normals to improve sensitivity of POC cTn assay and identify additional AMI patients without losing the specificity required to identify appropriate pts for discharge from ED.

Schneider et al. 13

Abbott i-STAT Retrospective Presentation i-STAT 99th Death i-STAT: 43 i-STAT: 85 i-STAT: 33 i-STAT: 89 Venge et al.

Page 30: AACB Guideline cover sheet

December 2016 29

(POC) cTnI & Siemens Stratus CS (POC) cTnI vs. lab-based assays Beckman AccuTnI cTnI and Abbott Architect cTnI

outcomes study Sweden; 851 patients presenting to ED and on whom cTnI requested

sample only evaluated

perc.=80 ng/L Stratus 99th perc.=70 ng/L AccuTnI 99th perc.=40 ng/L Architect 99th perc.=28 ng/L

from CVD and all-cause death; 0-35 months; 131 died

(34-52) Stratus CS: 54 (44-64) AccuTnI: 86 (77-92) Architect: 82 (73-89)

(82-87) Stratus CS: 78 (74-81) AccuTnI: 57 (53-61) Architect: 72 (68-75)

(26-41) Stratus CS: 25 (19-31) AccuTnI: 21 (17-26) Architect: 28 (23-34)

(87-92) Stratus CS: 93 (90-95) AccuTnI: 97 (95-98) Architect: 97 (95-98)

24, a

Roche Elecsys hs-TnT

960 patients presenting to ED, non-traumatic chest pain; 220 (22.9%) AMI (196 NSTEMI)

0h and 2h post presentation

99th perc. = 14 ng/L (<10% CV)

Architect cTnI 28 ng/L; 0h & 6-12h; ADP clinical pathway hs-TnT 0h vs 2h

MACE 30d 0h: 90.0 (85.6-93.3) 0+2h: 94.5 (90.7-97.0)

0h: 81.9 (80.6-82.9) 0+2h: 79.9 (78.8-80.6)

0h: 59.5 (56.5-61.7) 0+2h: 58.3 (55.9-59.8)

0h: 96.5 (95.0-97.7) 0+2h: 98.0 (96.6-98.9)

Aldous et al. 7

Architect cTnI 28 ng/L; 0h & 6-12h; ADP clinical pathway hs-TnT 2h +ECG + TIMI score cut-off of 0

100 (98.0-100)

19.2 (18.5-19.2)

28.1 (27.5-28.1)

100 (96.7-100)

Architect cTnI 28 ng/L; 0h & 6-12h; ADP clinical pathway hs-TnT 2h +ECG + TIMI score cut-off of ≤1

98.7 (96.0-99.7)

36.3 (35.4-36.6)

32.9 (32.0-33.2)

98.9 (96.6-99.7)

Roche hs-TnT 806 patients (39 with AMI)

0h, 2h, 3h, 4h, 6h, 8h and 12h post-presentation

99th perc. = 14 ng/L

MI according to 2007 Universal Definition of MI b; clinical symptoms of ischaemia; hs-TnT >14 ng/L; clinical history & further investigations (echocardiography, stress testing, coronary angiography, ultrasound, x-ray)

Followed up at 7d, 30d, & 12mo for CV events, significant bleeding

2h: 97.4 (86.5-99.9) 3h: 97.4 (86.2-99.9) 4h: 100.0 (90.5-100.0) 6h: 100.0 (90.0-100.0) 8h: 100.0 (90.0-100.0) 12h: 100.0 (89.4-100.0)

2h: 78.1 (74.9-81.1) 3h: 77.2 (73.9-80.3) 4h: 77.3 (73.9-80.5) 6h: 77.1 (74.1-81.0) 8h: 75.7 (71.9-79.3) 12h: 74.3 (70.0-78.3)

2h: 19.2 (14.0-25.4) 3h: 19.3 (13.9-25.6) 4h: 19.9 (14.4-26.4) 6h: 20.8 (15.0-27.8) 8h: 21.2 (15.2-28.2) 12h: 22.3 (15.9-29.9)

2h: 99.8 (99.8-99.0) 3h: 99.8 (98.9-100.0) 4h: 100.0 (99.3-100.0) 6h: 100.0 (99.2-100.0) 8h: 100.0 (99.1-100.0) 12h: 100.0 (98.9-100.0)

South Australian iCCnet Study

99th perc. = 14 ng/L AND delta cTnT of 7 ng/L

MI according to 2007 Universal Definition of MI b; clinical symptoms of ischaemia; hs-TnT >14 ng/L; clinical history & further investigations (echocardiography,

2h: 74.36 (57.9-87.0) 3h: 84.2 (68.7-94.0) 4h: 94.6 (81.8-99.3) 6h: 91.4

2h: 96.45 (94.8-97.7) 3h: 95.45 (93.6-96.9) 4h: 95.3 (93.4-96.8) 6h: 95.3

2h: 52.7 (38.8-66.3) 3h: 50.8 (37.9-63.6) 4h: 53.0 (40.3-65.4) 6h: 53.3

2h: 98.6 (97.4-99.3) 3h: 99.1 (98.0-99.7) 4h: 99.7 (98.9-100.0) 6h: 99.5

Page 31: AACB Guideline cover sheet

December 2016 30

stress testing, coronary angiography, ultrasound, x-ray)

(76.9-98.2) 8h: 88.6 (73.3-96.8) 12h: 93.9 (79.8-99.3)

(93.3-96.9) 8h: 94.0 (91.7-95.9) 12h: 93.3 (90.6-95.4)

(40.0-66.3) 8h: 49.2 (36.4-62.1) 12h: 50.8 (37.7-63.9)

(98.5-99.9) 8h: 99.2 (98.0-99.8) 12h: 99.5 (98.3-99.9)

Roche Cobas h232 (POC) cTnT

Danish paramedic study of 985 patients; 200 (20%) with AMI; median symptom duration of 70 min (35-180)

Pre-hospital sample median time 88 min (58-131)

cTnT >50 ng/L

Paramedics recorded ECG, symptoms of prolonged chest pain, suggestive of AMI

Survival analysis done

39 (32-46)

95 (94-97)

68 (59-77)

86 (84-88)

Stengaard et al. 22

Roche Cobas h232 (POC) cTnT

806 patients (39 with AMI)

0h, 2h, 3h, 4h, 6h, 8h and 12h post-presentation

cTnT decision limit 50 ng/L (no 99th perc. is available)

MI according to 2007 Universal Definition of MI b; clinical symptoms of ischaemia; hs-TnT >14 ng/L; clinical history & further investigations (echocardiography, stress testing, coronary angiography, ultrasound, x-ray)

Followed up at 7d, 30d, & 12mo for CV events, significant bleeding

2h: 79.5 (63.5-90.7) 3h: 89.5 (85.2-97.1) 4h: 91.9 (78.1-98.3) 6h: 88.6 (73.3-96.8) 8h: 88.6 (73.3-96.8) 12h: 97.0 (84.2-99.9)

2h: 96.2 (94.6-97.5) 3h: 95.6 (93.8-97.0) 4h: 95.3 (93.4-96.8) 6h: 94.7 (92.5-96.3) 8h: 95.2 (93.0-96.8) 12h: 92.9 (90.1-95.2)

2h: 52.5 (39.1-65.7) 3h: 53.1 (40.2-65.7) 4h: 52.3 (39.5-64.9) 6h: 49.2 (36.4-62.1) 8h: 54.4 (40.7-67.6) 12h: 50.1 (37.9-63.6)

2h: 98.9 (97.8-99.5) 3h: 99.4 (98.5-99.8) 4h: 99.5 (98.6-99.9) 6h: 99.3 (98.2-99.8) 8h: 99.2 (98.0-99.8) 12h: 99.8 (98.6-100.0)

South Australian iCCnet Study

Radiometer AQT90 (POC) cTnT

806 patients (39 with AMI)

0h, 2h, 3h, 4h, 6h, 8h and 12h post-presentation

99th perc. = 17 ng/L

MI according to 2007 Universal Definition of MI b; clinical symptoms of ischaemia; hs-TnT >14 ng/L; clinical history & further investigations (echocardiography, stress testing, coronary angiography, ultrasound, x-ray)

Followed up at 7d, 30d, & 12mo for CV events, significant bleeding

2h: 86.6 (69.5-94.1) 3h: 92.1 (78.6-98.3) 4h: 94.4 (81.3-99.3) 6h: 97.1 (84.7-99.9) 8h: 97.1 (84.7-99.9) 12h: 96.9 (83.8-99.9)

2h: 96.9 (95.3-98.0) 3h: 96.5 (94.8-97.7) 4h: 96.5 (94.8-97.8) 6h: 96.3 (94.4-97.6) 8h: 95.1 (92.8-96.7) 12h: 94.6 (92.0-96.5)

2h: 58.9 (45.0-71.9) 3h: 59.3 (45.7-71.9) 4h: 59.6 (45.8-72.4) 6h: 60.0 (45.9-73.0) 8h: 55.9 (42.4-68.6) 12h: 57.4 (43.2-70.8)

2h: 99.2 (98.2-99.7) 3h: 99.5 (98.7-99.9) 4h: 99.7 (98.9-100.0) 6h: 99.8 (99.0-100.0) 8h: 99.8 (98.9-100.0) 12h: 99.8 (98.6-100.0)

South Australian iCCnet Study

ACS, acute coronary syndrome; ADP, Accelerated Diagnostic Pathway; AHA/ACC, American Heart Association/American College of Cardiology; AMI, acute myocardial infarction; ECG, electrocardiograph; MACE, major adverse cardiac events; NPV, Negative Predictive Value; PPV, Positive Predictive Value; ROC AUC – Receiver Operating Curve Area under Curve; TIMI score, The Thrombolysis In Myocardial Infarction score; a Venge P et al. evaluated prognostic performances of assays and found that i-STAT and Stratus CS identified significantly fewer of the deaths than those identified by the lab-based assays. 24 b Universal Definition of MI 2007 guideline: Thygesen K et al. J Am Coll Cardiol 2007;50:2173-95.

Page 32: AACB Guideline cover sheet

December 2016 31

Appendix 4: Educational materials

Educational materials include a slide presentation of the recommendations and case studies and are available on the AACB website at: http://www.aacb.asn.au/resources/recommendations-for-use-of-troponin-poc

Appendix 5: Glossary of terms 18

Highly sensitive (hs) cTn assay – refers to a more analytically sensitive assay than current contemporary cTnI and cTnT assays. Evidence for application of the term to an assay may be:

1) Ability to yield reliable, valid measurements in more than 80% of samples from healthy subjects;

2) Troponin concentration corresponding to a CV of <10% is significantly lower than the 99th percentile value of the healthy reference population; and

3) Clinical studies in chest pain and ACS patients using these assays should show increased prognostic ability over contemporary assays in detection of cardiac events.

Limit of Detection (LoD) – refers to the lowest amount of troponin detected with 99% probability, using an estimation procedure partly based on nonparametric statistics. LoD is the lowest troponin concentration likely to be reliably distinguished from the Limit of Blank (LoB) which is the highest measurement result for a sample that contains no troponin (blank sample).

Limit of Quantitation (LoQ) – for clinical application of troponin the most important assay characteristic is LoQ, defined as the lowest amount of troponin that can be quantitatively determined with stated acceptable (i.e. “clinically meaningful”) imprecision and bias. The allowable analytical performance should be defined and applied to the concentration corresponding to the 99th percentile upper reference limit that, to be clinically usable, cannot be lower than the LoQ of the troponin assay. Necessarily LoQ is ≥ LoD.

The (99th percentile) URL of a healthy reference population – if troponin concentrations from a healthy reference population are ranked in centiles from least to greatest, the upper reference limit (URL) is given by the 99th centile. A sample size of at least 300 individuals per sex- and age-matched reference population should avoid the effect of outliers and achieve a 95% probability that at least 99% of the population will have a troponin concentration representative of health. Different studies using manufacturer’s troponin methods and platforms have shown variability of the 99th percentile concentration depending on the reference population used, the skewness of the distribution and the number of outliers. Clinicians should refer to their local laboratory for information about the analytical performance of the troponin assay and the clinical decision limit in use at their site.

Page 33: AACB Guideline cover sheet

December 2016 32

References

1. Pulkki K, Suvisaari J, Collinson P, Ravkilde J, Stavljenic-Rukavina A, Hammerer-Lercher A, et al. A pilot survey of the use and implementation of cardiac markers in acute coronary syndrome and heart failure across Europe. Clin Chem Lab Med 2009;47:227-34.

2. Collinson PO, van Dieijen-Visser MP, Pulkki K, Hammerer-Lercher A, Suvisaari J, Ravkilde J, et al. Evidence-based laboratory medicine: how well do laboratories follow recommendations and guidelines? The Cardiac Marker Guideline Uptake in Europe (CARMAGUE). Clin Chem 2012;58:305-6.

3. McKeeman GC, Auld PW. A national survey of troponin testing and recommendations for improved practice. Ann Clin Biochem 2015;52:527-42.

4. Davey RX. Troponin reporting in Australia in 2003. Ann Clin Biochem 2005;42:61-3.

5. Thygesen K, Alpert JS, Jaffe AS, Simoons ML, Chaitman BR, White HD: the Writing Group on behalf of the Joint ESC/ACCF/AHA/WHF Task Force for the Universal Definition of Myocardial Infarction. Third Universal Definition of Myocardial Infarction. Circulation 2012;126:2020-35.

6. Hammarsten O, Fu ML, Sigurjonsdottir R, Petzold M, Said L, Landin-Wilhelmsen K, et al. Troponin T percentiles from a random population sample, emergency room patients and patients with myocardial infarction. Clin Chem 2012; 58:628-37.

7. Aldous S, Richards AM, George PM, Cullen L, Parsonage WA, Flaws D, et al. Comparison of new point-of-care troponin assay with high sensitivity troponin in diagnosing myocardial infarction. Int J Cardiol 2014;177:182-6.

8. Anderson JL, Adams CD, Antman EM, Bridges CR, Califf RM, Casey DE Jr, et al. ACC/AHA 2007 guidelines for the management of patients with unstable angina/non-ST-elevation myocardial infarction. J Am Coll Cardiol 2007;50:e1-e157.

9. Apple FS. A new season for cardiac troponin assays: it’s time to keep a scorecard. Clin Chem 2009;55:1303-6.

10. Jones GRD, Panteghini M. Pre-analytical factors affecting troponin measurement. In: Tate J, Johnson R, Jaffe A, Panteghini M, editors. Laboratory and Clinical Issues Affecting the Measurement and Reporting of Cardiac Troponins: A Guide for Clinical Laboratories. Alexandria: AACB; 2012. p. 63-7.

11. Barth JH, Panteghini M, Bunk DM, Christenson RH, Katrukha A, Noble JE, et al. Recommendation to harmonize the units for reporting cardiac troponin results (Letter). Clin Chim Acta 2014;432:166.

12. Jones GRD, Finckh A, Wilson S. Introducing highly sensitive troponin into routine use – a worked example. In: Tate J, Johnson R, Jaffe A, Panteghini M, editors. Laboratory and Clinical Issues Affecting the Measurement and Reporting of Cardiac Troponins: A Guide for Clinical Laboratories. Alexandria: AACB; 2012. p. 97-102.

13. Schneider HG, Ablitt P, Taylor J. Improved sensitivity of point of care troponin I values using reporting to below the 99th percentile of normals. Clin Biochem 2013;46:979-82.

14. Wilgen U, Pretorius CJ, Ungerer JPJ. Support for lowering the 99th percentile cut-off for the point of care Abbott iSTAT cardiac troponin I (letter in response to Schneider HG et al.) Clin Biochem 2013;46:1774-5.

15. Saenger AK, Beyrau R, Braun S, Cooray R, Dolci A, Freidank H, et al. Multicenter analytical evaluation of a high-sensitivity troponin T assay Clin Chim Acta 2011;412:748-54.

Page 34: AACB Guideline cover sheet

December 2016 33

16. Kavsak PA, Don-Wauchope AC, Hill SA, Worster A. Acceptable analytical variation may exceed high-sensitivity cardiac troponin I cutoffs in early rule-out and rule-in acute myocardial infarction algorithms (Letter). Clin Chem 2016;62:887-9.

17. ter Avesta E, Vissera A, Reitsmac B, Breedveldb R, Wolthuisc A. Point-of-care troponin T is inferior to high-sensitivity troponinT for ruling out acute myocardial infarction in the emergency department. Eur J Emerg Med. 2014 Dec 22. [Epub ahead of print] DOI: 10.1097/MEJ.0000000000000225

18. IFCC Troponin Assay Analytical Characteristics table in ng/L units: http://www.ifcc.org/media/276661/IFCC%20Troponin%20Tables%20ng_L%20DRAFT%20Update%20NOVEMBER%202014.pdf (Accessed February 2, 2016)

19. Palamalai V, Murakami MAM, Apple FS. Diagnostic performance of four point of care troponin I assays to rule in and rule out acute myocardial infarction. Clin Biochem 2013;46:1631-5.

20. Hjortshøj S, Venge P, Ravkildeet J. Clinical performance of a new point-of-care cardiac troponin I assay compared to three laboratory assays. Clin Chim Acta 2011;412:370-5.

21. Venge P, Ohberg C, Flodin M, Lindahl B. Early and late outcome prediction of death in the emergency room setting by point-of-care and laboratory assays of cardiac troponin I. Am Heart J 2010;160:835-41.

22. Stengaard C, Sørensen JT, Ladefoged SA, Christensen EF, Lassen JF, Bøtker HE, et al. Quantitative point-of-care troponin T measurement for diagnosis and prognosis in patients with a suspected acute myocardial infarction. Am J Cardiol 2013;112:1361-6.

23. Diercks DB, Peacock WF, Hollander JE, Singer AJ, Birkhahn R, Shapiroet N, al. Diagnostic accuracy of a point-of-care troponin I assay for acute myocardial infarction within 3 hours after presentation in early presenters to the emergency department with chest pain. Am Heart J 2012;163:74-80.

24. Venge P, Lindahl B. Cardiac troponin assay classification by both clinical and analytical performance characteristics: a study on outcome prediction. Clin Chem 2013;59:976-81.