Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
A Review of the IEXUndulatorUndulator
Intermediate Energy X-ray
By Mark Jaski
Historyy
The project started with a request if we could do something similar
h i l l i i d lto the circular polarizing undulator (CPU) that is installed in sector 4.
Several possible undulator designs di dwere discussed
– Apple II (all permanent magnet)
– Electromagnetic/Permanent tmagnet
– Electromagnetic
The device had to be i i diThe CPU in Sector 4 quasiperiodic.
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
2
Optimization
OPERA Optimizer was used to optimize the pole geometry
Optimization was done on a ½ period model
The currents were chosen such that the Bx and By coils were limited tothe Bx and By coils were limited to 40 and 45 watts respectively. This constraint keeps the coil temperatures down.
The geometry was modified, dimensions were changed, and the Bx and By effective fields were calculated.
This process was repeated over and over again until the maximum Bx+By field was obtained.
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
3
Optimization Model DimensionsTable 3: Optimized dimensions.
I.D. Description Optimal Value unit
A By coil height max 6.3 cm B Bx coil height max 5.034 cm C By pole width 1.823 cmD* Conductor size max 0.3264 cm E* Conductor size min 0.3231 cmF Bx pole gap 1.021 cm G Bx pole ball 3.561 cm H Bx toenail chamfer 0.404 cmJ Bx pole toe 0.852 cm K Bx tip chamfer z 0.094 cm L Bx tip chamfer x 1.176 cmM Bx toe tip chamfer x 0.13 cm N Bx toe tip chamfer y 0.072 cm P Bx cut z 0.131 cm Q Bx base height 2.057 cmR Bx radius 0.5 cm S Bx chamfer 0.276 cm T Bx angle 15 5 °T Bx angle 15.5 U By tip offset 0.037 cm V By cut z 0.341 cmW By base height 5.417 cm X By radius 0.5 cm Y By chamfer 0.351 cm Z By angle 9 ° AA* C d t di 0 081AA* Conductor radius 0.081 cmBB* Conductor insulation 0.049 cm CC* Inner coil fiberglass 0.046 cmDD* Outer coil fiberglass 0.036 cm
EE* Conductor allowance 0.0076 cm
FF Bx pole height 2.218 cm
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
4
GG Bx pole heel width 2.973 cm * Not shown in figure 2
Field Plots
Bx(x,0.0,z) By(x,0.0,z)
0.151.5
Normalized Bx Multipoles r = 3mm (Bx peak = 3767 Gauss) BxBS_3BS 5 0.04
0.051.5
Normalized By Multipoles r = 3mm (By peak = 5456 Gauss) ByB_3B 5
-0.05
0
0.05
0.1
-0.5
0
0.5
1
0 2 4 6 8 10 12
oles
Nor
mal
ized
e M
ain
Fiel
d
orm
aliz
ed B
x Fi
eld
BS_5
-0.02
-0.01
0
0.01
0.02
0.03
-0.5
0
0.5
1
0 2 4 6 8 10 12
oles
Nor
mal
ized
he M
ain
Fiel
d
orm
aliz
ed B
y Fi
eld
B_5
-0.15
-0.1
-1.5
-1
Mul
tipo
to thN
o
z (cm)
(a)-0.05
-0.04
-0.03
-1.5
-1
Mul
tipo
to thN
o
z (cm)
(b)
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
5
Bx Field By Field
IEX Selected ParametersPeriod 12.5 cmPeriod 12.5 cmGap 10.5 mmPeriods per device (including end poles) 38 PeriodsMinimum Photon Energy 250 eVRequired vertical effective field 4510 GaussCurrent density in the copper conductor2 4.7 A/mm2
C t 47 6 A
General
Current 47.6 ATurns per coil1 62 turnsAmpere-turns1, 2 2951 Ampere-turnsWatts per coil1, 2 44.9 WattsTotal number of coils 152 EachT t l 2 6630 W tt
Horizontal Linear
Polarization
Total power2 6630 WattsMaximum temperature of coils 100 °CMinimum Photon Energy 440 eVRequired horizontal effective field 3310 GaussCurrent density in the copper conductor2 4.9 A/mm2
Current 50.3 ACurrent 50.3 ATurns per coil1 46 turnsAmpere-turns1, 2 2314 Ampere-turnsWatts per coil1, 2 40.2 WattsTotal number of coils 304 EachTotal power2 11,868 Watts
Vertical Linear
Polarization
Total power 11,868 WattsMaximum temperature of coils 100 °CMinimum Photon Energy 440 eVRequired horizontal and vertical effective field 2340 GaussCurrent at vertical effective field 20.7 ACurrent at horizontal effective field 34.2 A
1E d il ll
Circular Polarization
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
6
1End coils are smaller2At the required effective field
OPERA 4 Period Model For Prototype
No Coils
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
7
Add h C il
OPERA 4 Period Model For Prototype
Add the CoilsTypically end coils are ~¼ - ¾ - 1 turns
Actual TurnsBx 11 - 34 - 46
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
8
By 15 - 42 - 62
OPERA 4 Period Model For Prototype
Add Trim Coils
For Beam SteeringAnd Multi-pole Correction
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
9
And Multi pole Correction
4 Period Prototype. Why only 4 periods?dBB
z
2/
21)(
RMS values
4000
period
dzBzBz
rms
2/
2)(
2000
3000
4000
Bx
1000
0
1000
-40 -30 -20 -10 0 10 20 30 40
eld
(gau
ss) Bx
ByBx_rmsBy_rmsFully Developed
-3000
-2000
-1000
Fie Fully Developed-
Fully Developed+
-4000
z (cm)
2 f ll P i d f f ll d l d fi ld2/2011
Mark Jaski Accelerator Systems Division Magnetic Devices Group10
~2 full Period of fully developed field
General Assembly Exploded ViewVanadiumVanadium permendur poles
By Coils Bx Coils
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
11
Steel Core
Main coil leadsTrim coil leads
Trim coil winding
T i C il
Main coil winding
Trim Coils•Two windings per coil
•Main coil winding
•Trim coil winding
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
12
IEX Prototype 1 (4 periods)
Major Changes Magnetic field roll off was tooMagnetic field roll off was too
large for storage ring injection. The poles will be made wider.
Plating reduced the field by ~1%. The poles are not plated.
The gap was changed from 11.0 mm to 10.5 mm to lower the maximum current to lower themaximum current to lower the temperature rise.
The period was changed from 12.0 cm to 12.5 cm to make the coils larger and lower the field requirements.
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
13
IEX Prototype 2
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
14
Compare Measured Bx Field To Simulated Bx Field
The measured Bx effective field is 3.5% lower than simulated.The measured Bx Peak field is 3.1% lower than simulated.
Bx 51.4 A Eff MaxB i l t d 3331 3709Bx 51.4 A
3000
4000
150
200
Bx simulated 3331 3709Bx Measured 3213 3593
0
1000
2000
ld (G
auss
)
0
50
100
r (G
auss
)
-3000
-2000
-1000
Bx
fiel
-150
-100
-50 Erro
r
Simulated Bx FieldMeasured Bx Field
-4000-400.0 -300.0 -200.0 -100.0 0.0 100.0 200.0 300.0 400.0
z (mm)
-200error
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
15
Similar results for By
Bolt HolesWithout bolt holes
With bolt holes
With bolt holesholes holes holes
Gauss Gauss %Bx max 3719 3690 -0.8Bx eff 3360 3338 -0.7By max 5388 5388 0.0By eff 4590 4589 0.0
• Bolt holes do cause an 0.8% reduction in the Bx field.
• Bolt holes were not modeled• Bolt holes were not modeled because they add more computation time to the analyses.
• IEXP2 has SS bolts• IEXP2 has SS bolts.
• Steel bolts can be used to slightly increase the Bx fields.
• Exchanging steel bolts with SS bolts could be used as a method for tuning.
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
16
Original BH Curve
Use This BH Curve
Modify the VP BH Curve
25000
15000
20000
s)
tentenVP After Machining[1]
10000
B(g
auss VP After Machining[1]
vacoflux 50 820°CVP Jaskivp_adjust_for_IEX.bh
BH curves are dependent on:•Temperature•Heat treat
•Geometry
0
5000•Geometry•Manufacturing process•Chemical composition
00 20 40 60 80 100 120 140 160 180
H(oersted)
[1] "SOME INITIAL RESULTS FROM THE NEW SLAC PERMEAMETER"J K C bb d R A E l
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
17
J. K. Cobb and R. A. Early
Compare Measured Bx Field To Simulated Bx FieldAfter Adjusted BH CurveAfter Adjusted BH Curve
The measured Bx effective field is 1% lower than simulated.The measured Bx Peak field is 0.6% lower than simulated.
Bx 51.4 A Eff MaxB i l t d 3245 3614Bx 51.4 A
3000
4000
150
200
Bx simulated 3245 3614Bx Measured 3213 3593
0
1000
2000
eld
(Gau
ss)
0
50
100
r (G
auss
)
-3000
-2000
-1000
Bx
fie
-150
-100
-50 Erro
Simulated Bx FieldMeasured Bx Field
-4000-400.0 -300.0 -200.0 -100.0 0.0 100.0 200.0 300.0 400.0
z (mm)
-200error
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
18
Simulated values will drop by ~0.8% if bolt holes are modeled
Measurements of IEXP2 Fix the Skew Sextupole
Aimin X. pointed out a skew sextupole component.
Yes it can be fixed.
Not an active fix.
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
19
Measurements of IEXP2 Fix the Skew SextupoleBX Field Full Current Skew Components Normalized to the Fundimental Field
0.5
1
1.5
d Fi
eld
BS_1
-1
-0.5
0-40 -30 -20 -10 0 10 20 30 40 50
Nor
mal
ized BS_2
BS_3BS_4
-1.5
z (cm)
BX Field Full Current Skew Components Normalized to the Fundimental Field
0.05
0.1
0.15
Fiel
d
BS_1
-0.1
-0.05
0-40 -30 -20 -10 0 10 20 30 40 50
Nor
mal
ized
BS_2BS_3BS_4
2/2011
-0.15
z (cm)
Measurements of IEXP2 Fix the Skew Sextupole
Grinding these four downstream pole tips reduces the skew sextupolereduces the skew sextupolecomponent.
This was tested and showed this on the prototype.p yp
The trim coils will be set so the first and second integrals are zero.
The skew sextupole will be measured and the amount of grinding will be calculated.
Extra pole tips have been Extra pole tips have been ordered.
Provisions for easy replacement of these pole
2/2011
p ptips are provided.
Mark Jaski Accelerator Systems Division Magnetic Devices Group21
Measurements of IEXP2 Fix the Skew Sextupole
int(Bx dz) vs. x
0.0004
0.0005
0.0006
int(Bx dz) vs. x
Measured
simulated
Simulated 0.82 shims and zeroing trimsM d Aft Shi i
0
0.0001
0.0002
0.0003
Bx
(T-m
)
Measured After Shimming
-0.0003
-0.0002
-0.0001
0-0.03 -0.02 -0.01 0 0.01 0.02 0.03
x (m)
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
22
End Coil Multi-pole Field Configurations
Normal Dipole Normal SextupoleNormal QuadrupoleNormal Dipole Normal SextupoleNormal Quadrupole
Skew Dipole Skew OctupoleSkew Quadrupole
2/2011
Preliminary ResultsUS Sextupole
2000
2500
3000
**(2
-n) B_1
B_2
0
500
1000
1500
Inte
grat
ed F
ield
(G-c
m*
B_3B_4BS_1BS_2BS_3BS_4
-500
00 0.5 1 1.5 2 2.5 3 3.5
Bx Amperes
I
US Skew Octupole
80
100
120
140
160
180
G-c
m**
(2-n
) B_1B_2B_3
•US sextupole is clean with no other multipoles.
•US skew octupole also has a
-20
0
20
40
60
80
0 0.5 1 1.5 2 2.5
Inte
grat
ed F
ield
(G B_4BS_1BS_2BS_3BS_4
US skew octupole also has a skew sextupole.
•The flux bridge appears to play a roll in additional multipoles.
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
24
-40
Bx Amperes
p
Possible Multi-pole Fields With Trim Coils
Trim Coils Field per Ampere
1500
700
900
1100
1300
2-n)
) B_1B_2B 3
100
300
500
700
Fiel
d (G
-cm
**(2 B_3
B_4BS_1BS_2BS 3
500
-300
-100
100
ole
ole ad ad ext
oct
ole
ole ad ad ext
oct
F BS_3BS_4
-500
US
_Bx_
dipo
US
_By_
dipo
US_
sk_q
u
US
_qu
US
_se
US_
sk_o
DS
_Bx_
dipo
DS
_By_
dipo
DS_
sk_q
u
DS
_qu
DS
_se
DS_
sk_o
multipole
2/2011
Bx, 1st integral, and Beam Trajectory No Trim Coils
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
26
Bx, 1st integral, and Beam Trajectory With Trim Coils
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
27
Thermal Test (with help from Jeff C.)f i h h b / /
900
100
120
Bx 50.3 A, 18.85V, No fan, with Vacuum chamber, 12/21/10
500
700
60
80
ed by Water (W
)
ature (°C)
By Coils
30040
Heat R
emove
Tempra
Bx Coils
POWER OUT
‐100
100
0
20
0 50 100 150 200 250 300
Time (minutes)
With Vacuum ChamberBx 50.3 Amps18.85 VNo fan949 W tt t t l i
827 W removed by water87% removed by water
Time (minutes)
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
28
949 Watts total inuntil 137 minute markthen power off
Thermal Test (with help from Jeff C.)f i h h b / /
900
100
120
Bx 50.3 A, 17.95V, 10% fan, with Vacuum chamber, 12/21/10
500
700
60
80
ed by Water (W
)
ature (°C)
By Coils
30040
Heat R
emove
Tempra
Bx Coils
POWER OUT
‐100
100
0
20
0 50 100 150 200 250 300
Time (minutes)
With Vacuum ChamberBx 50.3 Amps17.95 V10% fan903 W tt t t l i
578 W removed by water64% removed by water
Time (minutes)
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
29
903 Watts total inuntil 111 minute markthen power off and fan off
Tunnel Heating
Without the fan turned on ~1500 W of heat will leak into the tunnel at full current.
With the fan turned on ~3500 W of heat will leak into the tunnel at full current.
Sometimes the device is on full current. Sometimes the device is off. Sometimes the device operates somewhere in‐between.
The heat into the tunnel is not constant. This make the tunnel temperature difficult to control.
Reversing the fan and sucking the heat from the device into the tunnel air handling system eases controlling the tunnel temperaturehandling system eases controlling the tunnel temperature– Suggested by Marvin K.
This minimizes the heat leaking into the tunnel.
A smoke test with a reversed fan showed adequate good air flow A smoke test with a reversed fan showed adequate good air flow.
A thermal test showed similar coil cooling.
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
30
Quasi-periodicity with an Electromagnet Device
24 Period Model Shown
Quasiperiodic By Field
Quasiperiodic Electron Trajectory
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
31
Quasi-periodicity Suppresses the Higher Harmonics
Flux with and without quasi‐periodicity turned to a 15% field‐strength reduction at the q p y gQP poles for the EM. The 3rd harmonic is reduced by over 90% while the first harmonic is reduced by ~18%.
R. Dejus et al., Spectral Performance Of Circular Polarizing Quasi‐periodic Undulators
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
32
j , p g pFor Soft X‐rays At The Advanced Photon Source, PAC09
IEX Layout Welded aluminum top frames
Linear Roller bearingsfor 7” travel in xAbove pitch roll yaw
6 ‐ 2” dia threaded rod adjustmentsRoll, pitch, yawx +/‐ 0.5, y ‐1.0 +3.0, z +/‐ 0.5
Above pitch, roll, yaw
Steel Second Frame
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
33
Steel Bottom Frame
Modified Vacuum Chamber
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
34
Assembled Position (out position)
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
35
Jaw drops 10 microns (0.00040”) from out to in.
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
36
Fundamental Harmonic 33.8 Hz.
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
37
Load on casters (in position)
A 1400 lb force at the top will cause thiscause this device to tip.
A double wheel caster
4141 lbs2497 lbs
with a 5,000 lb rating will be used.
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
38
Selected Features
200 lb h ~200 lbs to push to get rolling on casters.
34” wide, 194” long, 57.2”34 wide, 194 long, 57.2 tall on casters.
An acme screw will be d d hprovided to move the jaws
in or out of operating position.p
456 coils and 16 power supplies.
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
39
IEX Device
Installation in January 2012
4.8 m long~14,000 lbs
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
40
Laminated Assembly Proposal
Laminated By poles
h h b l f •There have been several requests for a pulsed device.
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
41
Thank You
2/2011Mark Jaski Accelerator Systems Division Magnetic Devices Group
42