A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

Embed Size (px)

Citation preview

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    1/221

    A HIGH-SENSITIVITY FLEXIBLE-EXCITATION ELECTRICAL

    CAPACITANCE TOMOGRAPHY SYSTEM

    A thesis submitted to the University of Manchester

    Institute of Science and Technology for the degree of

    Doctor of Philosophy

    1997

    JOSE CARLOS GAMIO ROFFE

    Department of Electrical Engineering and Electronics

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    2/221

    LIST OF CONTENTS

    ABSTRACT .............................................. vii

    DECLARATION ............................................ ix

    ACKNOWLEDGEMENTS ........................................ x

    CHAPTER 1: Introduction ................................. 1

    1.1 OVERVIEW OF ELECTRICAL CAPACITANCE TOMOGRAPHY ...... 1

    1.2 MAIN AREAS OF IMPROVEMENT .......................... 3

    1.3 AIMS AND OBJECTIVES ................................ 6

    1.4 ORGANISATION OF THIS THESIS ........................ 6

    CHAPTER 2: Theory of electrical capacitance

    Tomography (ECT) ............................. 8

    2.1 THE ECT SENSOR AS A SYSTEM OF CHARGED

    CONDUCTORS ......................................... 8

    2.2 ECT MEASUREMENT STRATEGIES ........................ 12

    2.2.1 THE SINGLE-ELECTRODE EXCITATION METHOD ......... 12

    2.2.2 QUALITATIVE IMAGE RECONSTRUCTION FOR

    SINGLE-ELECTRODE EXCITATION: THE LINEAR

    BACK-PROJECTION (LBP) ALGORITHM ................ 14

    2.2.2.1 THE SENSITIVITY MAPS ....................... 14

    2.2.2.2 THE NORMALISED MEASUREMENTS ................ 16

    2.2.2.3 THE WEIGHTED BACK-PROJECTION OPERATION ..... 17

    2.2.3 MULTIPLE-ELECTRODE EXCITATION METHODS .......... 17

    2.2.3.1 RECONSTRUCTION WITH MULTIPLE-ELECTRODE

    EXCITATION ................................. 22

    i

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    3/221

    2.2.3.2 OPTIMAL MULTIPLE-ELECTRODE EXCITATION

    PATTERNS ................................... 25

    2.3 ECT TRANSDUCERS ................................... 28

    2.3.1 THE CHARGE-DISCHARGE CAPACITANCE TRANSDUCER

    CIRCUIT ........................................ 29

    2.3.2 AC-BASED ECT TRANSDUCERS ....................... 32

    2.4 DISCUSSION AND CONCLUSIONS ........................ 35

    2.4.1 SINGLE- VS. MULTIPLE-ELECTRODE EXCITATION ...... 35

    2.4.2 CHARGE-DISCHARGE VS. AC-BASED TRANSDUCERS ...... 36

    CHAPTER 3: Finite-element simulation of single-

    electrode and parallel-field excitation ..... 38

    3.1 INTRODUCTION ...................................... 39

    3.2 FINITE-ELEMENT MODELLING OF ECT SENSORS............ 39

    3.3 SINGLE-ELECTRODE EXCITATION ....................... 44

    3.4 IDEAL SENSITIVITY MAPS .......................... 45

    3.5 PARALLEL-FIELD EXCITATION ......................... 47

    3.5.1 PARALLEL-FIELD GENERATION ...................... 47

    3.5.2 SENSITIVITY MAPS FOR PARALLEL-FIELD

    EXCITATION ..................................... 50

    3.6 COMPARISON OF IMAGES OBTAINED USING SINGLE-

    ELECTRODE AND PARALLEL-FIELD EXCITATION ........... 52

    3.7 DETERMINATION OF MUTUAL CAPACITANCES FROM

    PARALLEL-FIELD CHARGE MEASUREMENTS ................ 56

    3.8 CONCLUSIONS ....................................... 59

    CHAPTER 4: Design of an AC-based multiple-

    excitation ECT system ....................... 62

    4.1 INTRODUCTION ...................................... 62

    ii

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    4/221

    4.2 DESIRED SYSTEM CHARACTERISTICS .................... 64

    4.3 GENERAL DESIGN STRATEGY ........................... 65

    4.4 THE BASIC DETECTOR CIRCUIT ........................ 67

    4.4.1 TRANSFER FUNCTION WITH FREQUENCY-DEPENDENT

    OP-AMP GAIN .................................... 68

    4.4.2 COMMENTS ON NOISE PERFORMANCE .................. 73

    4.4.3 DETAILED CIRCUIT DESIGN ........................ 74

    4.4.4 FINAL FREQUENCY RESPONSE ....................... 76

    4.4.5 EFFECT OF CHANGES IN STRAY CAPACITANCE ......... 77

    4.5 DESIGN OF THE MULTI-EXCITATION ECT TRANSDUCER

    CHANNEL ........................................... 80

    4.5.1 NOMINAL MEASUREMENT RANGES AND SENSITIVITIES ... 84

    4.5.2 EFFECT OF CMOS SWITCH ON RESISTANCE .......... 85

    4.6 REFERENCE SINE-WAVE GENERATOR ..................... 91

    4.7 SIGNAL CONDITIONING AND DATA CONVERSION SECTION ... 93

    4.8 SYSTEM INTEGRATION AND CONSTRUCTION ............... 95

    4.9 CALIBRATION OF SYSTEM ELECTRONICS ................ 101

    4.9.1 CALIBRATION OF THE EXCITATION VOLTAGE

    SOURCES ....................................... 101

    4.9.2 DETECTOR CALIBRATION .......................... 103

    4.10 EXPERIMENTAL EVALUATION OF THE SYSTEM ........... 106

    4.10.1 CAPACITANCE MEASUREMENT COMPARATIVE TEST ..... 107

    4.10.2 LINEARITY EVALUATION ......................... 108

    4.10.3 INTRINSIC NOISE LEVEL TEST ................... 110

    4.10.4 DYNAMIC PERFORMANCE OF THE SYSTEM ............ 112

    4.11 IMAGE SAMPLES ................................... 115

    4.12 CONCLUSIONS ..................................... 118

    iii

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    5/221

    CHAPTER 5: Comparative experimental evaluation of

    single- and multiple-electrode excitation

    methods..................................... 120

    5.1 INTRODUCTION ..................................... 120

    5.2 EXPERIMENTS WITH OPTIMUM MULTI-ELECTRODE

    EXCITATION ....................................... 121

    5.2.1 DETERMINATION OF THE OPTIMUM EXCITATION

    VECTORS ....................................... 121

    5.2.2 APPLYING THE OPTIMUM EXCITATION VECTORS ....... 122

    5.2.3 EXPERIMENTAL PROCEDURE ........................ 127

    5.3 EXPERIMENT RESULTS AND DISCUSSION ................ 130

    5.3.1 EXPERIMENT No. 1 .............................. 131

    5.3.2 EXPERIMENT No. 2 .............................. 136

    5.3.3 EXPERIMENT No. 3 .............................. 139

    5.3.4 EXPERIMENT No. 4 .............................. 142

    5.3.5 EXPERIMENT No. 5 .............................. 145

    5.3.6 EXPERIMENT No. 6 .............................. 148

    5.4 CONCLUSIONS ...................................... 151

    CHAPTER 6: Iterative linear back-projection image-

    reconstruction techniques .................. 153

    6.1 INTRODUCTION ..................................... 153

    6.2 PREVIOUS WORK .................................... 153

    6.3 A RECONSTRUCTION ALGORITHM INSPIRED ON FEED-

    BACK CONTROL ..................................... 156

    6.3.1 FIRST APPROACH: ITERATIVE LBP RECONSTRUCTION .. 156

    6.3.2 AN ALGORITHM BASED ON A CONTROL-SYSTEM

    ANALOGY ....................................... 163

    iv

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    6/221

    6.4 EXPERIMENTAL EVALUATION OF THE

    FEED-BACK

    ALGORITHM ........................................ 169

    6.4.1 RESULTS OF IMAGE RECONSTRUCTION TESTS ......... 169

    6.4.2 DISCUSSION .................................... 178

    6.5 CONCLUSIONS ...................................... 179

    CHAPTER 7: Conclusions and further work ............... 180

    7.1 CONCLUSIONS ...................................... 180

    7.2 RECOMMENDATIONS FOR FUTURE WORK .................. 182

    REFERENCES ............................................ 184

    APPENDIX A: System design details and circuit

    diagrams .................................. 194

    APPENDIX B: Computer programs ......................... 201

    B.0 General ......................................... 201

    B.1 ECT system monitoring and control programs ...... 201

    B.2 Programs used in the optimum excitationexperiments ..................................... 204

    B.3 Simulation programs ............................. 205

    B.3.1 Programs to calculate the sensitivity mapsfor single-electrode excitation ............... 206

    B.3.2 Programs to calculate the sensitivity mapsfor parallel-field excitation ................. 206

    B.3.3 Programs to simulate single-electrode-excitation measurements ....................... 207

    B.3.4 Programs to simulate parallel-field-excitation measurements ....................... 208

    B.3.5 Programs to perform LBP image reconstruction .. 208

    B.3.6 Iterative LBP image reconstruction program .... 209

    v

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    7/221

    APPENDIX C: Papers produced as a result of this

    work ...................................... 210

    vi

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    8/221

    ABSTRACT

    Several ways of improving the performance of electrical

    capacitance tomography (ECT) systems are presented and

    evaluated, including the use of alternative excitation

    schemes, more sensitive and less noisy electronics, and more

    accurate image reconstruction algorithms.

    The design of a new electrical capacitance tomography (ECT)

    data acquisition system is presented, having a number of

    improvements over the one previously designed at UMIST. The

    new system uses AC-based instead of charge-discharge

    capacitance transducers, providing an increase in resolution

    from 0.26 to 0.025 fF (peak value of noise level) and a

    ten-fold improvement in stray-immunity. Thanks to the use of

    AC amplifiers before demodulation the problem of drift is

    practically eliminated. Phase-sensitive demodulation is

    employed in order to be able to discriminate between the

    effects of the conductive and capacitive components of the

    unknown admittance. Each channel has its own demodulator thus

    allowing parallel measurement. The latter, coupled with

    high-frequency (500 kHz) operation, results in a potential

    acquisition rate of more than 100 frames per second. Finally,

    excitation signals can be applied to several electrodes at

    the same time, and, thus, the system can be employed to

    investigate the possibility of using multi-electrode

    excitation patterns.

    The use of parallel-field excitation (attempting to mimic

    X-ray tomography) is explored employing finite-element

    simulation techniques and found to be disadvantageous. It is

    shown that, due to fundamental differences in the underlying

    physics, an analogy cannot really be established between

    X-ray tomography and parallel-field excitation ECT. On the

    other hand, the use of optimal or adaptive excitation

    vii

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    9/221

    methods, seeking to maximize the visibility or

    distinguishability of the permittivity distributions imaged,

    were successfully tested against the conventional single-

    electrode excitation method, and results are presented for

    different permittivity distributions.

    Finally, results are presented of an assessment of a new

    iterative image reconstruction algorithm based on the quite

    singular approach of viewing the reconstruction process as a

    feed-back control system.

    viii

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    10/221

    DECLARATION

    No portion of the work referred to in this thesis has been

    submitted in support of an application for another degree or

    qualification of this or any other university, or other

    institute of learning.

    ix

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    11/221

    ACKNOWLEDGEMENTS

    I would like to express my appreciation for their help,

    encouragement and valuable advice to my supervisor Dr. R C

    Waterfall, Professor M S Beck and Dr. W Q Yang.

    I also acknowledge the financial support of the Mexican

    Petroleum Institute and the National Council for Science and

    Technology of Mexico (Conacyt).

    Finally, I thank my wife Sara for her patience and support.

    x

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    12/221

    C H A P T E R 1 :

    I N T R O D U C T I O N

    1.1 OVERVIEW OF ELECTRICAL CAPACITANCE TOMOGRAPHY

    Electrical capacitance tomography (or ECT) is one of a

    relatively new breed of imaging techniques developed for

    industrial process applications, collectively known as

    process tomography [1,2]. The aim of all these methods, which

    started to evolve in the mid 1980s, is to provide a

    non-invasive, non-intrusive means to obtain cross-sectional

    images of the interior of process vessels (figure 1.1), which

    can be used to control and monitor process operations, or as

    a model validation tool in process design.

    Fig. 1.1 Process tomography systemFig. 1.1Fig. 1.1

    1

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    13/221

    Although the use of several tomographic modalities has been

    explored, including ionising radiation, magnetic resonance

    imaging, ultrasonic and optical techniques, electrical

    methods based on impedance measurement have generally proved

    more suitable for process tomography applications, being

    fast, robust and relatively inexpensive. Electrical impedance

    tomography (EIT) can be subdivided into resistance,

    inductance and capacitance tomography, depending on the

    physical quantity being measured.

    Fig. 1.2 Electrical capacitance tomography systemFig. 1.2Fig. 1.2

    Electrical capacitance tomography is aimed at industrial

    processes involving non-conducting materials, or mixtures

    where the continuous phase is non-conducting. In a

    conventional ECT system (figure 1.2), the sensor takes the

    form of a circular array of electrodes placed around an

    2

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    14/221

    insulating pipe and surrounded by a grounded screen (the

    latter not shown in the figure for clarity). The data

    acquisition unit contains capacitance transducers which are

    used to determine the capacitance of all possible electrode

    pair combinations, thus producing n(n-1) measurements, where

    n is the number of electrodes. By means of a computer and a

    suitable algorithm, these data must then be used to

    reconstruct an image of the permittivity distribution inside

    the sensor, which directly reflects the material

    distribution.

    The first attempts to do capacitance-based tomography were

    carried out more or less at the same time (

    1986-88) both at

    UMIST [3-5] and the Morgantown Energy Technology Centre (in

    the USA) [6-8]. Later on, in 1991, the first real-time ECT

    system was developed in a joint project by UMIST, Leeds

    University and Schlumberger Cambridge Research [9,10]. ECT

    has been applied, at an experimental level, to the on-line

    visualisation of gas-oil flows [11,12], as well as imaging

    combustion processes [13-17] and fluidised beds [18-20]. For

    an excellent review article including numerous applications

    and an extensive bibliography (80 references) see [42].

    1.2 MAIN AREAS OF IMPROVEMENT

    In what follows we identify some of the principal areas

    subject to improvement in ECT, which will become the focusand motivation of this work.

    The main challenge encountered in the design of an ECT system

    comes from the fact that the capacitances to be measured are

    extremely small. For example, for a typical 12-electrode

    empty sensor the various inter-electrode capacitances range

    from 10 to 1,000 femtoFarads, and the expected full-scale

    3

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    15/221

    capacitance changes extend from about 5 to 80 femtoFarads,

    for 10 cm long electrodes. Therefore, depending on the

    particular permittivity distribution, the system may have to

    measure accurately capacitance changes of a few tenths of a

    femtoFarad. Furthermore, these very low inter-electrode

    capacitances have to be measured in the presence of stray

    capacitances to ground several order of magnitude larger

    (

    100 picoFarads). There is, then, the need for better

    high-sensitivity low-noise capacitance transducers, in order

    to increase the signal to noise ratio (SNR) and the

    resolution of the measurements. With the use of more

    sensitive electronics, it would be possible to increase the

    spatial and/or axial resolution by employing smaller

    electrodes. Alternatively, the higher sensitivity can be

    exploited for imaging lean flows.

    Another problem area found

    Fig. 1.3 Sensitivity mapFig. 1.3Fig. 1.3

    for opposite electrodes

    i n E C T i s t h a t

    inter-electrode capacitances

    are much more sensitive to

    changes in permittivity near

    the electrodes than to

    changes occurring near the

    centre of the sensor. This

    is reflected on the shape of

    the sensitivity maps, which

    are basically graphs of

    dC/d for a specific pair of

    electrodes. For instance, infigure 1.3 we can see the

    typical sensitivity map for opposite electrodes. It roughly

    defines a channel of sensitivity across the sensor area.

    However, we can see that the sensitivity is not constant

    along this

    channel

    ; the two peaks correspond to the

    electrode positions and the sensitivity decreases towards the

    centre. As a result of this non-uniform sensitivity effect,

    4

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    16/221

    it is much more difficult to

    see

    permittivity changes near

    the centre than elsewhere in the sensor. The idea of somehow

    increasing the sensitivity in the centre to achieve a uniform

    sensing area sounds, therefore, very attractive. In the early

    stages of this work, it was thought that the problem would be

    solved or alleviated if a parallel electric field could be

    created inside the sensor by applying specific voltages to

    the electrodes [21,22], although, after a thorough

    investigation of the matter, it was later found that this was

    not entirely the case [23]. There are, however, other types

    of multiple-electrode excitation that can be used to improve

    the overall sensitivity of EIT systems in general, which have

    been already successfully tried for resistance tomography

    [30,31,33], but not for ECT.

    Let us finally say a word about image reconstruction. To

    perform an accurate quantitative reconstruction of the

    permittivity distribution inside the sensor from the

    capacitance measurements is a very complex task, which

    mathematically belongs to the category of inverse problems

    and normally involves computationally intensive iterative

    procedures based on optimisation. However, by linearising the

    problem, a qualitative reconstruction can be done using a

    simple and fast algorithm known as linear back-projection

    (LBP) [4], adapted from medical X-ray tomography. Because of

    its simplicity and speed, virtually all ECT systems use the

    LBP algorithm. Nevertheless, the quality of the images

    obtained with LBP is rather poor, and there is clearly the

    need for better reconstruction methods that can provide bothaccuracy and speed.

    5

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    17/221

    1.3 AIMS AND OBJECTIVES

    In view of the considerations presented in the previous

    section, the general aims of this work are set as follows:

    a) To investigate ways to increase the sensitivity of ECT

    systems in the central area of the sensor, particularly

    the use of parallel fields and multi-electrode excitation.

    b) To develop new capacitance transducers for ECT use, having

    more sensitivity and lower noise level than those

    currently being employed.

    c) To investigate improved reconstruction algorithms based

    on iterative LBP methods.

    In order to have experimental support, an ECT system will be

    designed and built, incorporating the new high-sensitivity

    transducers developed.

    1.4 ORGANISATION OF THIS THESIS

    In chapter 2, the physical situation occurring in an ECT

    sensor is described, according to the laws of

    electromagnetism. The theoretical implications of using

    various different excitation arrangements are presented, and

    a comparison is made between them, showing the unsuitabilityof parallel fields. Also, the previous work on ECT system

    design in discussed here, and new ways of improving

    capacitance transducer sensitivity using AC-based circuits

    are proposed.

    Chapter 3 presents a critical analysis of the idea of using

    parallel-field excitation as a means to increase the

    6

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    18/221

    sensitivity in the centre of the sensor and reduce the

    non-uniformity of the sensitivity maps, in an attempt to

    mimic X-ray tomography. The analysis is based on finite-

    element simulation using the software package PC-OPERA, and

    includes a comparison between parallel-field and conventional

    single-electrode excitation.

    The design and testing of an AC-based high-sensitivity

    capacitance transducer is presented in chapter 4. The design

    of a multiple-excitation ECT system based on this transducer

    is also described.

    Chapter 5 presents results obtained with an ECT system built

    after the design presented in chapter 4, including the

    experimental evaluation of various single- and multiple-

    electrode excitation methods.

    Chapter 6 discusses the implementation of a new iterative

    reconstruction algorithm based on back-projection.

    Finally, chapter 7 summarises the main achievements of this

    project and gives suggestions about possible areas for future

    work.

    7

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    19/221

    C H A P T E R 2 :

    THEORY OF ELECTRICAL CAPACITANCE TOMOGRAPHY

    2.1 THE ECT SENSOR AS A SYSTEM OF CHARGED CONDUCTORS

    In ECT, a number of electrodes are installed around the pipe

    or vessel to be imaged, surrounded by a grounded screen. This

    is the basic ECT sensor configuration, shown in figure 2.1

    for 12 electrodes. The mutual capacitance (defined below) of

    the different electrode pairs depends on the permittivity

    distribution inside the sensor. When a body is placed in the

    sensor there will be a change in the mutual capacitances. The

    principle of ECT is to measure the change in mutual

    capacitance of the different electrode pair combinations and

    then use these measurements to reconstruct an image of the

    permittivity distribution in the cross section being

    investigated by the sensor. The reconstruction process can be

    done with a computer using a simple algorithm known as linear

    back-projection (LBP) [4].

    Fig. 2.1 Basic 12-electrode ECT sensorFig. 2.1Fig. 2.1

    8

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    20/221

    From a physics point of view, the ECT sensor can be

    considered as a special case of a system of charged

    conductors separated by a dielectric medium [24-27], the

    theory of which was first developed by Maxwell [28]. In our

    particular case, with the sensor electrodes acting as the

    charged conductors, the electrode charges Qi and the electrode

    potentials Vi are related by the following set of linear

    equations for an n-electrode sensor

    Q1= c11V1+c12V2++c1nVn

    Q2= c21V1+c22V2++c2nVn

    (2.1) Qn= cn1V1+cn2V2++cnnVn

    where the coefficients cii are called the self-capacitance of

    electrode i, while the others, cij, with i j, are the mutual

    capacitance of electrodes i and j.

    Writing equation (2.1) in matrix form we have

    Q1 c11 c12 c1n V1 Q2 c21 c22 c2n V2 = (2.2) = Qn cn1 cn2 cnn Vn

    or

    Q = C V (2.3)

    9

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    21/221

    The matrix C is called the capacitance matrix of the system.

    The self and mutual capacitances are sometimes called

    coefficients of capacitance and coefficients of

    (electrostatic) induction respectively, and in the very old

    books they are termed coefficients of capacity and

    coefficients of influence. These coefficients depend only on

    the geometry and the permittivity distribution of the system,

    and have the following important properties:

    a) The self-capacitances are always positive.

    b) The mutual capacitances are always negative.

    c) For every conductor we have

    ci1 + ci2 + + c in 0 (2.4)

    d) For the mutual capacitances

    cij = cji (2.5)

    The matrix C is called the capacitance matrix and completely

    characterises the system of conductors. C is a non-linear

    function of the system geometry and of the permittivity

    distribution in the dielectric medium. In our case the

    geometry is fixed, so any change in C will be due to a change

    in the permittivity distribution.

    An ECT sensor (or any system of charged conductors) can also

    be modelled using a circuit theory approach, as a network of

    component capacitances [29], as illustrated in figure 2.2

    using a 4-electrode sensor. To do this, we re-write equation

    (2.1) in terms of the voltage difference between the various

    electrodes, to put it in the form

    Q1= C1V1+C12(V1-V2)+C13(V1-V3)+

    Q2= C2V2+C21(V2-V1)+C23(V2-V3)+ (2.6)

    10

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    22/221

    or, in compact form, for electrode i of n

    n

    Qi = CiVi + Cij(Vi-Vj) (2.7)j=1

    (i j)

    where the component capacitance between electrode i and

    ground is given by

    Ci = ci1+ci2+ci3++cin (2.8)

    and the inter-electrode component capacitance between

    electrodes i and j by

    Cij = -cij (i j) (2.9)

    Fig. 2.2 Equivalent circuit (based on the componentFig. 2.2Fig. 2.2

    capacitances) of a 4-electrode ECT sensor

    As mentioned earlier, in ECT we are concerned only with the

    inter-electrode capacitances (i.e. the mutual capacitances),

    11

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    23/221

    since this are the ones that depend on the permittivity

    distribution inside the sensing area of the sensor. We are

    not interested in the component capacitances to ground (i.e.

    between the electrodes and the screen) because, due to the

    geometry of the sensor (assuming that the inter-electrode

    gaps are small), they depend mainly on the permittivity

    distribution in the annular region between the electrode ring

    and the outer screen, outside the imaging area. We are not

    interested in the self-capacitances either because they are

    determined by the mutual capacitances and the component

    capacitances to ground (see equation 2.8).

    Consequently, we are only interested in the mutual

    capacitances, only half of which are independent (because of

    their reciprocity relationship). So, we can say that all the

    information about any change in the permittivity distribution

    inside the sensor will be contained in the variations of the

    n(n-1) independent mutual capacitances, which form the lower

    (or upper) triangular part of the capacitance matrix C.

    2.2 ECT MEASUREMENT STRATEGIES

    The value of the self and mutual capacitances can be found by

    applying known potentials to the sensor electrodes and

    measuring the electrode charges. In practice, the

    determination on the electrode charges is normally done

    indirectly by measuring the electrode currents (Q = i d t), andthe excitation potentials are applied to the electrodes in

    the form of a periodic signal of known amplitude.

    2.2.1 THE SINGLE-ELECTRODE EXCITATION METHOD

    All ECT systems reported in the literature so far, including

    those developed earlier at UMIST [10], use single-electrode

    12

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    24/221

    excitation to measure the mutual capacitances, with the

    exception of the Morgantown system, which uses a special

    bipolar excitation technique [8]. Let us consider a

    12-electrode sensor (figure 2.1). Using the single-electrode

    excitation method, the mutual capacitances are determined as

    follows: First an excitation voltage is applied to electrode

    1 while keeping all the others at zero potential and the

    charge on electrodes 2 to 12 is measured. According to

    equation (2.1), these measurements directly represent c2 1 to

    c12 1 . Next, the excitation voltage is applied to electrode 2

    while keeping all the others at zero and the charge on

    electrodes 3 to 12 is measured, representing c3 2 to c12 2 . This

    procedure is repeated, applying voltage to electrode n and

    measuring the charge on electrodes (n+1) to 12, until, as a

    final step, voltage is applied to electrode 11 and the charge

    of electrode 12 is measured. In this way, the 66 independent

    mutual capacitance values corresponding to the lower half of

    the capacitance matrix are determined (the other 66 being

    given by equation (2.5)), requiring 66 electrode charge

    measurements.

    From a hardware design point of view, single-electrode

    excitation has the advantage of requiring only one voltage

    source, which can be switched sequentially to the electrode

    being used as a source.

    The problem with this method is that, because the mutual

    capacitances are so small, the electrode charges can also be

    very small, and, as a result, the signal-to-noise ratio (SNR)of the measurements tends to be rather poor, even when

    low-noise measuring circuits are used. From equation (2.1),

    it is clear that, if excitation potentials are applied to

    more than one electrode, it is possible to obtain larger

    electrode charges, although they would no longer be a direct

    measure of any particular inter-electrode capacitance.

    13

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    25/221

    2.2.2 QUALITATIVE IMAGE RECONSTRUCTION FOR SINGLE-ELECTRODE

    EXCITATION: THE LINEAR BACK-PROJECTION (LBP) ALGORITHM

    In single-electrode excitation ECT systems, image

    reconstruction for two-component mixtures is done using the

    linear back-projection (LBP) algorithm [3]. The basic idea of

    this qualitative algorithm, which is an adaptation of a

    method used in medical tomography, is to do a weighted

    back-project or smearing of each one of the n(n-1)

    normalised measurements along its sensing zone, given by the

    corresponding sensitivity map.

    2.2.2.1 THE SENSITIVITY MAPS

    Let us consider an n-electrode sensor and an image made of m

    equal-area pixels. For each pair of electrodes i (source) and

    j (detector), a capacitance sensitivity map can be defined by

    Ci j( k ) - C i j emp Qi j( k ) - Q i j empSi j(k)= = (2.10)

    Ci j full - Ci j emp Qi j full - Qi j emp

    where k is the pixel number (from 1 to m), Qi j(k) is the

    charge induced on electrode j by electrode i when the region

    of pixel k is full of high-permittivity material while the

    rest of the sensing area is full of low-permittivity

    material, Qi j full and Qi j emp are the charge induced on electrode

    j by electrode i when the sensor is full of high- and

    low-permittivity material, respectively, and Cij are the

    corresponding mutual capacitances of electrodes i and j under

    the same conditions.

    14

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    26/221

    a) Location of electrodes b) S 13

    c) S15 d) S17

    Figure 2.3 Typical sensitivity maps for single-electrodeFigure 2.3Figure 2.3

    excitation

    The sensitivity maps can be determined experimentally for a

    particular sensor by probing the sensing area using a test

    rod, although this is a very time-consuming task. A more

    practical approach is to use computer simulation techniques

    to model the sensor using the finite-element method (FEM).

    The author chose the latter and, in his work, used PC-OPERA,

    a commercially available FEM software package for

    electromagnetic analysis and simulation. A more detailed

    description of the procedures used to calculate the

    sensitivity maps is given in chapter 3 and in appendix B.

    15

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    27/221

    Regardless of the method used, it is not necessary to

    determine the sensitivity maps for all possible electrode

    pair combinations, since, due to sensor symmetry, all the

    sensitivity maps can be obtained by rotation from (for a

    12-electrode sensor) the following basic set of 6: S12 , S13 ,

    S14 , S15 , S16 and S17.

    Figure 2.3 shows typical sensitivity maps for several

    electrode combinations, which were calculated for imaging oil

    and gas mixtures ( high = 2.1 and low = 1), with m = 313. As

    might be expected, the sensitivity maps show that each

    electrode pair responds mainly to the material lying between

    the electrodes, albeit in a very non-uniform way.

    2.2.2.2 THE NORMALISED MEASUREMENTS

    Prior to back-projection, the measurements obtained with each

    electrode pair are normalised according to

    Ci j meas - Ci j emp Qi j meas - Qi j emp i j= = (2.11)

    Ci j full - Ci j emp Qi j full - Qi j emp

    where i j is the normalised measurement (charge or

    capacitance) corresponding to electrodes i (source) and j

    (detector), Qi j meas is the measured charge induced on electrode

    j by electrode i, Qi j full and Qi j emp are the charge induced on

    electrode j by electrode i when the sensor is full of high-

    and low-permittivity material, respectively, and Cij are the

    corresponding mutual capacitances of electrodes i and j.

    Normalised values are used so that the same software will

    cope with systems having different electrode lengths, which

    produce different absolute measurements but the same

    normalised ones.

    16

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    28/221

    2.2.2.3 THE WEIGHTED BACK-PROJECTION OPERATION

    Mathematically, for an n-electrode sensor and an m-pixel

    image, the LBP algorithm calculates the grey level G(k) for

    each pixel k as

    G(k)

    n 1

    i 1

    n

    j i 1

    i j Si j (k)

    n 1

    i 1

    n

    j i 1

    Si j (k)

    (k 1..m) (2.12)

    where

    i j are the normalised measurements defined by equation(2.11) and Si j are the sensitivity maps defined by equation

    (2.10). The actual back-projection operation occurs in the

    numerator of equation (2.12), while the quantity in the

    denominator serves as a position-dependent weighting factor

    used to compensate for the decrease in sensitivity towards

    the centre of the sensor.

    2.2.3 MULTIPLE-ELECTRODE EXCITATION METHODS

    Although they have not been used in ECT, multiple-electrode

    excitation techniques have been around for quite a while in

    other EIT modes and are especially popular in electrical

    resistance tomography (ERT) work [30,31], where they can be

    found under various names like adaptive , multi-reference ,

    and optimal currents methods.

    If there are n electrodes in an ECT sensor, there is no

    reason why we should not simultaneously apply excitation

    voltages to more than one of them. In fact we can define

    excitation voltage vectors of the form

    V = [V1 , V2 , ..., Vn ] (2.13)

    17

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    29/221

    If we apply this excitation voltage vector to an ECT sensor

    and then measure all the electrode charges, we can form

    another vector

    Q = [Q1 , Q2 , ..., Qn ] (2.14)

    Let us consider two different permittivity distributions

    (p), corresponding to a known reference state such as an

    empty sensor, and (p), corresponding to some unknown

    material distribution, where p denotes a point inside the

    sensor. When we apply a voltage vector V to these two

    permittivity distributions we get two charge vectors Q and

    Q . Let us now define our measurement signal as the change in

    the electrode charges due to the permittivity distribution

    changing from to , i.e.

    Q = Q - Q (2.15)

    We can then define a vector of measured signals m as

    m = Q - Q = [ Q1 , Q2 , ..., Qn ] (2.16)

    Following the ideas of Issacson [30], we define thedistinguishability of with respect to when exciting

    with V as

    Q = m 2 = Q - Q

    2 = [ Q1 , Q2 , ..., Qn ] 2 (2.17)

    18

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    30/221

    additionally, from equation (2.3) we have

    Q = Q - Q

    2

    = (C - C )V 2

    = DV 2

    (2.18)

    where C and C are the capacitance matrices corresponding to

    and , respectively, and D = C - C is called the

    distinguishability matrix.

    Equation (2.17) shows that Q is an indicator of the

    magnitude of the detection signals (defined as the change in

    electrode charge Q). Obviously, the larger the detection

    signals the easier it will be to detect the change in

    permittivity distribution from to . On the other hand,

    equation (2.18) shows that Q depends on the distingui-

    shability matrix (which ultimately depends on and ),

    but, more importantly, also on the excitation vector V.

    In other words, for given permittivity distributions and

    , not all excitation vectors will produce the same signal

    level, and there will be some excitation vectors that are the

    best choice to distinguish between and , in the sense

    that they will produce the largest detection signals, and,

    therefore, the best signal-to-noise ratio (SNR). This is the

    main justification for the use of multiple-electrode

    excitation, since with single-electrode excitation the choice

    of excitation vectors is limited to those having only one

    non-zero element.

    Extending to ECT the result obtained by Issacson in [30], we

    can say that the best voltage vectors to distinguish

    between and

    are the eigenvectors of D having the

    largest eigenvalue. In general, excitation patterns having

    a low spatial frequency will yield measurements which are

    more sensitive to changes in permittivity near the centre,

    19

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    31/221

    while high frequency patterns yield measurements sensitive

    mainly to changes near the electrodes [30].

    Now, although the multiple-electrode excitation method yields

    measurements with the best SNR, there is a price to pay both

    in terms of the system hardware and software. Firstly, the

    hardware becomes more complex and expensive, since it must

    now include n independent voltage sources. Secondly, unlike

    with single-electrode excitation, what is measured is no

    longer capacitance but the response of the sensor (in the

    form of changes in electrode charge Q) to a set of

    excitation vectors Vk, with k = 1, ..., L. Because of this, the

    practical LBP algorithm, which is based on capacitance

    measurements, cannot be used, at least not directly.

    As it will soon become clear, it can be advantageous to

    derive an expression for the measurements Q as a function of

    the inter-electrode voltages, and to re-define the excitation

    vectors in terms of the latter instead of the electrode

    voltages themselves. We shall, for the sake of simplicity,

    use a 4-electrode sensor (figure 2.2) to illustrate these

    concepts, although the same ideas apply fully to sensors with

    any number of electrodes.

    From equation (2.7) we can arrive at the following system of

    equations describing the sensor

    Q1 =

    C1V1 +

    C12 (V1-V2) +

    C13 (V1-V3) +

    C14 (V1-V4) Q2 = C2V2 + C21 (V2-V1) + C23 (V2-V3) + C24 (V2-V4)

    Q3 = C3V3 + C31 (V3-V1) + C32 (V3-V2) + C34 (V3-V4) (2.19)

    Q4 = C4V4 + C41 (V4-V1) + C42 (V4-V2) + C43 (V4-V3)

    Now, because of the geometry of the sensor (assuming that the

    inter-electrode gaps are small) we have Ci 0, i = 1, .., n.

    20

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    32/221

    In other words, the change in the capacitances to ground is

    negligible because they do not depend on the permittivity of

    the material inside the sensing area, but only on that of the

    material in the region between the electrodes and the outer

    screen. Therefore, equation (2.19) becomes

    Q1 = C12 (V1-V2) + C13 (V1-V3) + C14 (V1-V4)

    Q2 = C21 (V2-V1) + C23 (V2-V3) + C24 (V2-V4)

    Q3 = C31 (V3-V1) + C32 (V3-V2) + C34 (V3-V4) (2.20)

    Q4 = C41 (V4-V1) + C42 (V4-V2) + C43 (V4-V3)

    For our 4-electrode sensor (n = 4) we can then define the

    following excitation vectors, of size n2-n:

    U = [ (V1-V2),(V1-V3),(V1-V4),(V2-V1),(V2-V3),(V2-V4),

    (V3-V1),(V3-V2),(V3-V4),(V4-V1),(V4-V2),(V4-V3) ] (2.21)

    In equation (2.20) we can see that our measurement signals,

    considered as the change in charge Qi, do not depend on Vi,

    the actual voltage on the measuring electrode, but only on

    the voltage differences between the electrodes. However, in

    practice the system cannot measure Qi directly, it has to

    measure Qi and Qi separately and from them calculate

    Qi = Qi - Qi

    . From equation (2.7) we see that the voltage Vi

    on the measuring electrode can have a considerable effect on

    its charge Qi, since the capacitances to ground Ci (which in

    an actual system include the capacitance of the cable

    connecting the electrode to the instrument) are much larger

    than the inter-electrode ones, whose effect will be obscured.

    In order to avoid having to measure the electrode charge due

    to Ci, the voltage Vi on the measuring electrode should be set

    21

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    33/221

    to zero. The advantage of defining the measurement signals Q

    and the excitation vectors U in terms of the inter-electrode

    voltages is that, once the most suitable vectors have been

    chosen (according to some criterion), the voltage on the

    measuring electrode can always be set to zero and the rest of

    the electrode voltages adjusted accordingly to satisfy the

    particular excitation vector being applied.

    2.2.3.1 RECONSTRUCTION WITH MULTIPLE-ELECTRODE EXCITATION

    The idea of multiple-electrode excitation brings about the

    important question of how to reconstruct an image from the

    knowledge of the applied excitation vectors and the vector ofmeasurements. For a quantitative reconstruction directly from

    the charge measurements, we have to resort to iterative

    algorithms like the those used in electrical resistance

    tomography (ERT) [32,33], which are normally based on some

    variant of Newton s method.

    The formal development of iterative image reconstruction

    algorithms based on optimisation is a vast and complex task

    that could itself be the subject of another PhD thesis,

    involving a considerable amount of advanced mathematics, and

    is not within the scope of this work. The main objective of

    this work, as far as multiple-electrode excitation is

    concerned, is to design and build an actual system and use it

    to confirm experimentally that multiple excitation can indeed

    improve distinguishability and the SNR of the measurements

    (chapters 4 and 5), a purpose which can be achieved without

    having to produce any images. Nevertheless, without going

    into details, we shall present a general description of some

    of the methods that can be used for iterative image

    reconstruction.

    Let us assume that we apply a set of L = n - 1 linearly

    independent inter-electrode voltage excitation vectors Uk,

    22

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    34/221

    k = 1, .., L. The excitation vectors are applied one by one

    and, for each vector, the change in electrode charge Q is

    measured for every electrode. So, for each excitation vector

    Uk we can form an n-dimensional vector of measurements

    mk like

    the one in equation (2.16). The total number of independent

    measurements is, therefore, N = nL = n (n - 1), and they can be

    stacked in a long vector

    M = [m1 ,m2 , ...,mN ]. (2.22)

    We know that M is a linear function of the excitation

    vectors, which can be stacked in a vector E = [U1, U2, ..., Uk ],

    and a non-linear function of the reference and unknown

    permittivity distributions and , i.e. M = (E, , ).

    As shown in figure 2.4, iterative algorithms start with an

    initial guess of the unknown permittivity distribution, 0 ,

    which is then used to calculate (E, , 0 ) using a

    finite-element model of the sensor. Then the iterative part

    commences, by comparing the simulated measurements with the

    actual ones, the difference between the two being calculated

    as i = (E, , i ) - M 2, i = 0, 1, 2, ...; if i is smaller

    than some specified tolerance then we can consider i as the

    true permittivity distribution, otherwise this error is fed

    to an optimiser (based on some Newton-like formula) that

    produces a new estimated permittivity distribution i+1 ,

    which is used to calculate the next simulated measurements.

    The cycle is repeated until the error is within the accepted

    tolerance. Despite requiring great computing power and being

    relatively slow,iterative algorithms like this produce

    quantitative reconstruction, unlike LBP. One serious problem,

    however, is that the noise level of the measurements can

    severely affect the algorithm s convergence and accuracy. The

    better SNR achieved through multiple-electrode excitation

    would be particularly useful in alleviating this problem.

    23

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    35/221

    Fig. 2.4 General flow diagram of iterative reconstructionFig. 2.4Fig. 2.4

    algorithms

    24

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    36/221

    It really seems rather strange that iterative image

    reconstruction algorithms based on optimisation theory, which

    are mathematically sound and yield quantitative results, have

    not been applied to the specific case of ECT, whereas in ERT

    they are very popular and there has been a lot of research

    into the subject, with the publication of many papers and

    several PhD theses.

    An alternative approach to image reconstruction for

    multiple-electrode excitation ECT systems involves using

    equation (2.20) to recover the inter-electrode capacitance

    changes Cij from the measurements M and the inter-electrode

    excitation voltages E. This can easily be done by solving n

    systems of n - 1 linear equations each. Once we have the

    inter-electrode capacitance changes Cij , which represent the

    equivalent single-electrode excitation measurements, we can

    use the standard LBP algorithm to perform a qualitative image

    reconstruction.

    2.2.3.2 OPTIMAL MULTIPLE-ELECTRODE EXCITATION PATTERNS

    The best inter-electrode voltage vectors U are the ones

    that maximise the changes in electrode charge Q (equation

    (2.20)), and they can be determined from the distingui-

    shability matrix D = C -C .

    Let us see, first, how the D matrix can be determined. For

    i j, we have Dij = cij - cij . For i=j, we have Dii = cii - cii

    , but

    from equation (2.8)

    n

    cii = Ci - cij (2.23)j=1

    (i j)

    therefore

    25

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    37/221

    n n

    Dii = Ci - cij - Ci + cij

    (2.24)j=1 j=1

    (i j) (i j)

    Recalling that the capacitances to ground do not depend on

    the permittivity distribution inside the imaging area, we

    have Ci = Ci , so they cancel out and equation (2.24) becomes

    n n

    Dii = - cij - cij (2.25)

    j=1 j=1

    (i j) (i

    j)

    or

    n

    Dii = - Dij (2.26)j=1

    (i j)

    From the foregoing discussion we can conclude that the matrix

    D is completely determined by the mutual capacitances cij and

    cij (with i j) and does not depend on the self-capacitances

    cii and cii . The mutual capacitances can easily be determined

    using the single-electrode excitation method described in

    section 2.2.1. In a practical situation, cij , the mutual

    capacitances for the reference state (empty sensor), could be

    measured in advance and stored, so that it would only be

    necessary to measure the mutual capacitances corresponding to

    the unknown permittivity distribution

    in order to

    determine D.

    26

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    38/221

    As mentioned earlier, the optimal excitation voltage vectors

    are the eigenvectors of D corresponding to the largest

    eigenvalues [30]. Many methods can be used to obtain the

    eigensystem of the symmetric matrix D, like the QR algorithm,

    singular value decomposition (SVD) and the Jacobi method. The

    SVD method is preferred because it is stable and

    straightforward to obtain [31]. Using SVD we factorise D as

    D = X Y T (2.27)

    where X and Y are orthogonal matrices and is a diagonal

    matrix whose entries are the singular values. The

    eigenvectors of D (i.e. the optimal excitation voltage

    vectors) are given by the columns of X, while its eigenvalues

    are equal to the singular values. The best excitation vector

    is the eigenvector corresponding to the largest eigenvalue.

    The last eigenvalue is always equal to zero, since the rank

    of D is n - 1, and the corresponding excitation vector is not

    used. If necessary, the electrode voltage vectors thus

    obtained can be re-scaled, in order to fully exploit the

    dynamic range of the voltage sources employed.

    In this way we end up with an optimal set of orthogonal

    n-dimensional electrode-voltage unit vectors Vk ,

    k = 1, .., (n - 1), which will produce the largest measurements

    for a particular permittivity distribution , compared with

    other vectors of unit length. It was shown earlier that,because of the need to always set the measuring electrode

    voltage to zero, it is more convenient to work with

    inter-electrode voltages. The optimal inter-electrode voltage

    vectors Uk can be obtained from Vk using equations (2.13)

    and (2.21).

    27

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    39/221

    2.3 ECT TRANSDUCERS

    The function of an ECT transducer is to measure the charge Q

    on a detection electrode. On a single-electrode excitation

    system, where only two electrodes are involved in each

    measurement, this charge Q is also a measure of the

    capacitance between the two electrodes, given by Cij = - Qi /Vj .

    Among the desired characteristics of an ECT transducer are:

    a) It must be stray-immune. This means that the capacitance

    between the measuring electrode and ground should not have

    an effect on the measurement. This is normally achieved by

    ensuring that the detection electrode potential is

    maintained at zero during measurement (thus making Vi = 0

    in equation (2.7)).

    b) Its range has to match the sensor and the characteristics

    of the materials to be imaged. For example, for a typical

    12-electrode sensor used to image oil ( r = 2.1) and gas

    ( r = 1) mixtures, the inter-electrode capacitances can lie

    anywhere between 10 and 600 femtoFarads approximately.

    c) It should be able to measure small inter-electrode

    capacitance changes in the presence of large standing

    values. Again for the same situation as in (b), we have

    that, depending on the particular electrode pair, the

    full-scale capacitance change can be as low as 15% of the

    standing capacitance. The transducer must have some means

    of balancing these standing capacitances.

    d) The resolution must be high enough, and the noise level

    low enough, to allow the detection of capacitance changes

    of a few tenths of a femtoFarad.

    e) If the system is to be used in fast real-time applications

    like flow imaging, the transducer must have a fast dynamic

    28

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    40/221

    response. Considering a frame rate of, say, 100 frame/s,

    the time available for collecting all the data for one

    frame is 10 ms.

    2.3.1 THE CHARGE-DISCHARGE CAPACITANCE TRANSDUCER CIRCUIT

    The single-electrode excitation ECT system developed at UMIST

    uses capacitance transducers based on the charge transfer

    principle [9,34]. This type of transducer is stray-immune

    (i.e. they are insensitive to the capacitances to ground) and

    its main attractive is its simplicity and relatively low

    cost.

    Fig. 2.5 The charge-discharge capacitance measuring circuitFig. 2.5Fig. 2.5

    The charge transfer transducer circuit is shown in

    figure 2.5. The device works by repeatedly charging and

    discharging the unknown capacitance Cx through the combined

    action of semiconductor switches S1

    to S4. First, S

    1 and S

    2 are

    29

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    41/221

    closed (keeping S3 and S4 open) to charge Cx to voltage Vc, and

    the charging current flows into the current-to-voltage

    converter formed by operational amplifier (op-amp) A1 and its

    feed-back resistor Rf, causing a negative output voltage. Inthe second half of the operating cycle, S

    1 and S

    2 are open

    while S3 and S4 are closed, thus discharging Cx to ground. The

    discharge current flows out of the current-to-voltage

    converter formed by op-amp A2

    and its feed-back resistor Rf,

    producing a positive output voltage. This charge-discharge

    cycle repeats itself at a frequency f (up to 2 MHz) and the

    successive charging and discharging current pulses are

    averaged in the two current detectors, producing two dc

    output voltages:

    V1 = - f V c Rf Cx + e1 (2.28)

    V2 = f Vc Rf Cx + e2 (2.29)

    where e1 and e2 are the output offset voltages of the current

    to voltage converters. Since the detection electrode of Cx is

    always connected to a virtual earth point, any stray

    capacitance to ground Cs1 is always short-circuited and has no

    effect on the measurement. And, because the excitation

    electrode is always being driven by a low-impedance voltage

    source, its stray capacitance Cs2 has no effect either.

    The voltage difference V2 - V1 is taken as the output, giving

    Vo = V2 - V1 = 2 f Vc Rf Cx + e2 - e1 (2.30)

    this has the advantage of doubling the sensitivity and that

    the offset signals e1 and e2 tend to cancel each other.

    30

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    42/221

    Capacitors C at the input of the op-amps ensure a stable

    virtual earth during the fast charge and discharge of Cx.

    The bandwidth of the circuit is set by Rf

    and Cf

    through

    1B = (2.31)

    2 Rf Cx

    and the maximum bandwidth achievable with a particular op-amp

    was calculated by Huang [35] as

    ABmax = 0.1 (2.32)

    T (C+Cf)Rf

    where A is the open-loop gain and T the open-loop time

    constant of the op-amp.

    The main limitation of the charge-discharge transducer is

    that it has a lower signal-to-noise ratio than AC-based

    circuits. For instance, in the system developed at UMIST,

    which uses a sensor with 12 10-cm-long electrodes, the peak

    noise level at the system s output is equivalent to an input

    capacitance 0.26 femtoFarads, with a transducer bandwidth of

    about 10 kHz. This figure essentially sets the resolution of

    the system, and corresponds to a change of 2% in the gas void

    fraction of an oil/gas mixture occurring in the middle of the

    sensor (which is the least sensitive area) [9]. Clearly,

    under these conditions, the measurements will not be too

    reliable or accurate, since the signal-to-noise ratio is

    equal to 1. If we want to use shorter electrodes in order to

    reduce the averaging effect along the axial direction, or

    increase the number of electrodes to, say, 16 in order to

    31

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    43/221

    increase the image spatial resolution, then the electrode

    area would be reduced and more sensitive transducers would be

    required to detect the smaller capacitances produced. The

    same is true for imaging mixtures with low permittivity

    contrast or lean flows.

    Another drawback of the charge-discharge transducer is that

    it is susceptible to the effects of conductance losses, i.e,

    it is not phase-sensitive. This can be a problem in

    applications involving water or in flame imaging, for

    example.

    Finally, this transducer suffers from charge injection caused

    by the feed-through of gate control signals in the switching

    semiconductor devices. This charge injection appears as an

    offset voltage at the output, which is the main component of

    e1 and e2 in equation (2.30). The effect is temperature-

    dependent and, for a temperature change of 15C, it is

    equivalent to an input capacitance of up to 5 femtoFarads

    [9]. These offsets cause a baseline drift and, in order to

    maintain accuracy, they need to be periodically monitored and

    compensated.

    2.3.2 AC-BASED ECT TRANSDUCERS

    The charge-discharge transducer effectively uses square-wave

    excitation. Although this type of excitation signal has the

    advantage of being easily implemented by switching between

    two DC levels, it has been recognized that by employingAC-based circuits, i.e., circuits based on sine-wave

    excitation, a higher signal to noise ratio can be achieved

    [9,35]. The main reason for this is that the use of

    single-frequency excitation allows the use of narrow-band

    filtering techniques based on phase-sensitive demodulators,

    which can greatly reduce the noise bandwidth [45] and also

    provide the means to discriminate between capacitive and

    32

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    44/221

    conductive effects. Additionally, switch-related problems

    like charge injection and the generation of glitches are

    eliminated. Likewise, drift is no longer such a big concern.

    Although there are numerous AC-based methods to measure small

    capacitance values (for a good review of many of them see

    [35]), not all of them lend themselves well to a

    multi-channel application like ECT. In particular, we are

    interested in transducer circuits based on the use of an

    operational amplifier as a current detector [36-40], an area

    relatively new compared with other methods.

    In order to be consistent with the concept of multiple-

    electrode excitation we shall consider the ECT transducer as

    a charge rather than capacitance sensor. Figure 2.6 shows a

    charge detector (or charge amplifier) [41] based on an

    op-amp, which will be used as the basic component for the

    design of a more sensitive ECT transducer in chapter 4.

    Fig. 2.6 Basic charge detectorFig. 2.6Fig. 2.6

    33

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    45/221

    Essentially, the circuit consists of an operational amplifier

    (op-amp) with capacitive feedback. The resistor R provides a

    path for the op-amp s DC bias current to avoid saturation of

    the device due to that current charging Cf

    . It has negligible

    effect on the op-amp output at the frequency of operation.

    Assuming that amplifier has a high gain at the frequency of

    operation, in the sinusoidal steady state, the circuit is

    described by Vo

    (j ) = - Zf Iin(j ) = j Iin(j )/ Cf where Iin(j )V (j ) I (j ) I (j ) I (j )V (j ) I (j ) I (j ) I (j )

    is the input current. But Iin(t) = - dQ(t)/dt , or in theI (t) Q(t)I (t) Q(t)

    frequency domain Iin

    (j ) = - j Q(j ) where Q(j ) is theI (j ) Q(j ) Q(j )I (j ) Q(j ) Q(j )

    electrode charge. Therefore we have that Vo(j ) = Q(j )/Cf .V (j ) Q(j )V (j ) Q(j )

    Thus, the circuit can be considered as an AC charge to

    voltage (Q-V) converter. Because of the feedback action of

    the op-amp, this circuit has the important advantage of

    keeping the measuring electrode at virtual earth, avoiding

    the appearance of the comparatively large charge due to the

    stray capacitance to ground (C1 in figure 2.6). In other

    words, the circuit is stray immune.

    This simple circuit measures the charge induced on the

    detection electrode. If only one electrode is used for

    excitation with a voltage Vexc (i.e. in the single-electrode

    excitation method) we can get the capacitance between the

    electrode pair from Cx = - Q/Vexc , and the circuit can be used

    as a capacitance transducer with its output given by

    Vo = - (Vexc /Cf ) Cx . The equivalent circuit for this case is

    shown in figure 2.7. The capacitance to ground of the

    detection electrode, CD , has no effect on Vo , since the

    voltage across it is very close to zero, whereas that of the

    excitation electrode, CE , is driven by a low-impedance

    voltage source and so it does not affect Vo either. However,

    if more than one electrode are used simultaneously for

    (multi-electrode) excitation then we cannot say precisely

    which capacitance is being measured, and it is more

    convenient to view the circuit as a charge detector.

    34

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    46/221

    Fig. 2.7 The charge detector as a capacitance meterFig. 2.7Fig. 2.7

    A full analysis of the capabilities and limitations of the

    proposed charge detector circuit will be presented in

    chapter 4, which shows the detailed design of an AC-based

    multiple-electrode excitation ECT system.

    2.4 DISCUSSION AND CONCLUSIONS

    2.4.1 SINGLE- VS. MULTIPLE-ELECTRODE EXCITATION

    It has been shown earlier that multiple-electrode excitation

    using optimal inter-electrode voltage vectors U can produce

    larger detection signals Q and, thus, increase the SNR of

    measurements compared with single-electrode excitation. This,

    35

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    47/221

    however, comes at the cost of more complicated and expensive

    hardware, since n independent voltage sources are required.

    Another point to consider is parallel measurement. In

    single-electrode excitation systems all electrodes but one

    are kept at zero volts, and they can all be put to measure

    simultaneously (i.e. parallel measurement). This is no longer

    achievable with multiple-electrode excitation, because, for

    any particular application of an inter-electrode voltage

    vector U, generally only a few electrodes will be at zero

    potential (i.e. available for measurement) and, therefore,

    each inter-electrode voltage vector will actually have to be

    applied several times, until all electrodes have had a chance

    to be at zero volts and be measured. This, of course, takes

    time and has the effect of slowing down the system.

    Considering the foregoing, it is the view of the author that,

    although multiple-electrode excitation can indeed be very

    useful in special applications where the absolute maximum

    sensitivity is desired, for industrial process applications

    where low-cost and speed of operation are important,

    single-electrode excitation ECT systems, with its straight-

    forward approach to both hardware design and image

    reconstruction (LBP algorithm), would be a more sensible

    choice.

    2.4.2 CHARGE-DISCHARGE VS. AC-BASED TRANSDUCERS

    The use of AC-based ECT transducers result in better noiseperformance, and hence, better resolution and SNR. Later on,

    in chapter 4, a system based on this type of transducer will

    be designed and the results of its experimental evaluation

    presented, showing that a 10-fold improvement in SNR over the

    charge-discharge system is possible in practice. Once again,

    there is a price to pay, and it comes in the form of more

    complex and expensive electronics, mainly due to the fact

    36

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    48/221

    that one demodulator per channel is required in order to have

    parallel measurement capability for fast operation (see

    chapter 4).

    Despite its higher cost, AC-based transducers are a good

    choice in application that require higher resolution and SNR,

    like when working with low-contrast mixtures or with lean

    flows, or in systems where the electrodes are small, either

    because they are short (say less that 10 cm) for better axial

    resolution, or because there are a large number of them (more

    that 12, say 16) for improved spatial resolution.

    37

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    49/221

    C H A P T E R 3 :

    FINITE-ELEMENT SIMULATION OF SINGLE-ELECTRODE AND

    PARALLEL-FIELD EXCITATION

    3.1 INTRODUCTION

    Of the many possible multiple-electrode excitation

    arrangements, so-called parallel-field excitation deserves

    special attention. At the beginning of this work, some ECT

    researchers shared the idea that the problem of low

    sensitivity in the centre of the sensor and the

    non-uniformity of the sensitivity maps (figure 2.3) had

    something to do with the uneven distribution of electric

    force lines which occurs when single-electrode excitation is

    employed [21,22]. This type of excitation results in the

    electric field being very strong near the excitation

    electrode, rapidly weakening as we move away (as shown in

    figure 3.4). It was thought that the ideal situation would

    rather be to have a parallel electric field uniformly

    distributed across the entire sensing area. By so trying to

    mimic X-ray tomography, it was believed that increased

    sensitivity in the central region would be achieved and also

    that the quality of the reconstructed images could be

    improved.

    In this chapter, we show how parallel-field excitation can be

    realised by applying specific excitation voltages to all

    electrodes in the sensor. We present the results of

    simulation experiments carried out to determine what effects

    would the use of parallel-field excitation have, both on the

    shape of the sensitivity maps and on the reconstructed

    38

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    50/221

    images, compared to single-electrode excitation. Finally, we

    consider the question of whether or not parallel field

    excitation can provide a complete set of independent

    measurements required to determine the sensor mutual

    capacitances.

    3.2 FINITE-ELEMENT MODELLING OF ECT SENSORS

    In order to investigate the effects of parallel-field

    excitation, simulation experiments were carried out using

    finite-element (FE) models of the sensor. PC-OPERA, a

    commercially available software package for 2-dimensional

    electromagnetic field analysis based on the finite-element

    method (FEM), was used to perform the simulation. Note that,

    because of the two extra cylindrical guarding electrodes

    (grounded) used in the actual sensor on each side of the

    sensing electrodes in the axial direction (figure 3.1), 2-D

    simulation can be used to model the sensor [4], albeit we are

    restricted to work with 2-dimensional material distributions.

    Fig. 3.1 Side view of ECT sensor showing guard electrodesFig. 3.1Fig. 3.1

    39

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    51/221

    We are going to use the FEM package to solve the following

    basic problem:

    Given a number of conducting bodies with known applied

    potentials placed in a dielectric medium with a known

    permittivity distribution, find the resultant electric

    field distribution and the self and mutual capacitances

    characterising the system

    Essentially what PC-OPERA does is to find the value of the

    electric potential at a large number of points or nodes in a

    mesh of contiguous triangular elements used to represent the

    actual physical system. The potential data can then be used

    to calculate the electric field vectors (by E = -

    V) orEE

    other parameters of interest.

    Given a relative permittivity distribution

    (x,y), PC-OPERA

    finds the potential distribution V(x,y) by numerically

    solving the following partial differential equation (where o

    is the free-space permittivity)

    [ o (x,y) V(x,y)] = 0 (3.1)

    subject to the corresponding Dirichlet conditions (known

    potentials on the boundary).

    Working with PC-OPERA involves the following three steps:

    1) First, in the preprocessing stage, a model of the problem

    is generated using an interactive pre-and-postprocessor

    program. In this phase two files are generated, one with

    extension "MES" which contains the geometry of the mesh,

    and another with extension "OP2" which contains the

    boundary conditions, material properties, etc.

    40

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    52/221

    2) Second, the static analysis program is used to generate

    the solution (i.e. the potential distribution). The input

    to this program are the two files created in the previous

    step, while its output is a solution file with extension

    "ST".

    3) In the third and final step the solution is viewed and the

    parameters of interest (i.e. electrode charge or

    capacitance) are calculated, using once again the pre-and-

    postprocessor program.

    The main parameter of interest, in our case, is the detection

    electrode charge Q, which can be calculated using Gauss Law:

    Q DDDdsss (3.2)

    however, since we are working in two dimensions, we will not

    integrate over a closed surface, but over a closed line

    around the electrode. It will not be a surface integral but

    a line integral, which is evaluated using one of the

    pre-and-postprocessor commands. And, of course, Q then

    represents the charge per unit length.

    Once the charge is known, the capacitance per unit length can

    be easily found (for single-electrode excitation) as

    C = - Q/Vexc , where Vexc is the voltage on the excitation

    electrode.

    PC-OPERA can be run interactively or in an

    off-line

    mode.

    Using this option, the three basic steps mentioned earlier

    are automatically executed in sequence. First, a series of

    optional pre-and-postprocessor commands (contained in a

    command input file) are executed on the specified input

    model, then the solution program is run, and finally another

    41

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    53/221

    set of pre-and-postprocessor commands (contained in another

    command input file) is executed on the solution file to

    calculate the parameters of interest.

    For our work, two FE models of a 12-electrode ECT sensor were

    created as shown in figures 3.2 and 3.3. In this type of

    2-dimensional problems the absolute dimensions are

    irrelevant, so we state the geometric characteristics of the

    model in terms of the inner radius R of the pipe (i.e. the

    radius of the imaging area). In this way, the thickness of

    the pipe wall is 0.1R, and the distance between the external

    side of the pipe and the outer screen is 0.2R. The relative

    permittivity of the region between the pipe and the screen is

    set to 1 (air), while that of the pipe is set to 2.5

    (perspex). The electrode angle is 26, with inter-electrode

    gaps of 4.

    a) Model geometry b) FE mesh (5881 elements)

    Figure 3.2 Basic FE sensor modelFigure 3.2Figure 3.2

    42

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    54/221

    a) Model geometry b) FE mesh (9912 elements)

    Figure 3.3 FE sensor model with polar gridFigure 3.3Figure 3.3

    One of the models (figure 3.3) includes a polar grid composed

    of 313 equal-area regions and is used in the determination of

    the sensitivity maps required for the LBP algorithm. The

    permittivity of each one of this regions can be independently

    set to any value, hence allowing the simulation of arbitrary

    material distributions. A polar grid was chosen instead of

    square one because of its particular symmetry, which allows

    the model behaviour to be orientation-independent.

    By setting the proper boundary conditions, the application ofany excitation voltage vector can be simulated.

    43

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    55/221

    3.3 SINGLE-ELECTRODE EXCITATION

    Figure 3.4 shows the simulation results for single-electrode

    excitation. The excitation voltage is applied to electrode

    one and the figure shows the equipotential lines. Clearly,

    with this type of excitation the electric field distribution

    is quite uneven, with the field concentrated near the

    excitation electrode.

    Fig. 3.4 Equipotential lines for single-electrodeFig. 3.4Fig. 3.4

    excitation

    In order to calculate the sensitivity maps (as defined by

    equation (2.10) in chapter 2), a program was written inQuickBasic 4.5, which iteratively runs PC-OPERA off-line

    using the 313-region polar-grid sensor model of figure 3.3

    with the electrode potentials set for single-electrode

    excitation. At the start of the ith iteration, the QuickBasic

    program generates a command input file that will be used by

    PC-OPERA to set the relative permittivity of in-pipe region

    i equal to 2.1 ( oil ) while that of all the others will be set

    44

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    56/221

    to 1. Then the program calls PC-OPERA off-line, which finds

    the solution and calculates the charge on all electrodes

    using another command input file previously written for this

    purpose. The process is repeated 313 times (for each one of

    the in-pipe regions). In this manner, all the sensitivity

    maps associated with electrode 1 are obtained (i.e. S1 1 to

    S1 12 ). The full program can be found in appendix B. The

    sensitivity maps associated with the electrodes 2 to 12 are

    obtained by rotation of those calculated for electrode 1.

    Finally we end up with 144 sensitivity maps, of which only 66

    are really needed for use with the LBP image reconstruction

    algorithm (one for each measurement, see sections 2.2.1 and

    2.2.2 in chapter 2).

    Typical examples of sensitivity maps for single-electrode

    excitation obtained in the way described above are shown in

    figure 2.3 (chapter 2). It can be observed that the detection

    areas of the sensor form clearly defined

    channels

    between

    the detection and excitation electrodes. This is a desirable

    feature in a tomography sensor, since each detector

    looks

    only to a specific area. However, the single-electrode

    excitation ECT sensor departs from the ideal situation in

    that:

    a) The

    channels

    are not straight, and

    b) Their

    height

    is not constant, that is to say, the

    response of the detectors is lower in the middle of the

    channel

    and higher near the electrodes.

    3.4

    IDEAL

    SENSITIVITY MAPS

    In an ideal situation, we would like to see something similar

    to the case of parallel-beam X-ray tomography, where each

    45

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    57/221

    detector is sensitive over a very narrow and straight

    channel

    of constant

    height

    .

    For an ECT sensor, though, the sensitivity

    channels

    could

    not be narrowed too much without having to reduce the

    electrode width to such an extent that measurement signals

    would become undetectable. However, it might seem natural to

    think that if multiple-electrode excitation is used so as to

    create a parallel electric field inside the sensor, at least

    sensitivity maps forming straight and uniform

    channels

    could be obtained, probably something similar to figure 3.5.

    a) Location of electrodes b) Electrode 6

    Figure 3.5Figure 3.5Figure 3.5

    c) Electrode 5

    Ideal

    ECT sensor

    s e n s i t i v i t y m a p s

    intuitively

    expected

    w i t h p a r a l le l - fi e l d

    excitation (field parallel

    to the x axis)

    46

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    58/221

    In order to verify the previous conjecture, in the next

    sections we shall describe how a parallel field can be

    created inside an ECT sensor, and the actual characteristics

    of the resulting sensitivity maps, which were calculated by

    FE simulation, will also be reported.

    3.5 PARALLEL-FIELD EXCITATION

    3.5.1 PARALLEL-FIELD GENERATION

    A parallel field can be approximated inside an n-electrode

    ECT sensor by applying electrode potentials according to

    sin ( i - )

    Vi = E (3.3)n

    MAX [ sin( i - )]

    i = 1

    where, referring to figure 3.6, i indicates the electrode

    number (1 to n), is the angle between the field direction

    and the y axis, E is a voltage constant (determined by

    hardware constraints), and i is the angular position of the

    centre of the ith electrode, given by

    360 i = ( i - 0.5 ) (3.4)

    n

    By using this voltage distribution, the potential difference

    between pairs of electrodes facing each other in the

    direction of the field is made proportional to the separation

    between their centres.

    47

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    59/221

    Fig. 3.6 Parallel electric field inside an ECT sensorFig. 3.6Fig. 3.6

    For example, to produce a parallel field along the y axis

    (

    = 0) on a 12-electrode sensor, with E = 15 V, the voltages

    shown on table 3.1 would have to be used.

    Table 3.1 Electrode potentials for parallel-field

    excitation

    ELECTRODE

    VOLTAGE

    ELECTRODE

    VOLTAGE

    1

    4.02

    7

    -4.02

    2 10.98 8 -10.98

    3

    15.00

    9

    -15.00

    4

    15.00

    10

    -15.00

    5 10.98 11 -10.98

    6 4.02 12 -4.02

    48

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    60/221

    The generation of a parallel field inside the sensor using

    the method described above was confirmed by FE simulation as

    shown on figure 3.7.

    Fig. 3.7 Equipotential lines for parallel-fieldFig. 3.7Fig. 3.7

    excitation using the voltages of Table 3.1

    By shifting the potentials one electrode position, the field

    can be rotated and different

    projections

    (in the sense of

    conventional X-ray computed tomography) can be defined, each

    one associated with a particular direction of the field. For

    a 12-electrode sensor, rotating the field in this way means

    that we can define six different projections (numbered 1

    to 6), corresponding to

    equal to 0, 30, 60, 90, 120

    and 150. Note that the remaining six projections (180,

    210, 240, 270, 300 and 330) do not contribute any

    additional information since they are just a sign-changed

    repetition of the first ones. For every projection, the

    signal from each one of the twelve electrodes can be

    measured, giving a total of 6 12 = 72 measurements.

    49

  • 7/23/2019 A High-sensitivity Flexible-excitation Electrical Capacitance Tomography System - c Gamio Phd Thesis

    61/221

    3.5.2 SENSITIVITY MAPS FOR PARALLEL-FIELD EXCITATION

    The sensitivity map for parallel-field excitation is defined

    as follows, for projection i and electrode j:

    Qi j(k) - Q i j empSi j(k)= (3.5)

    Qi j full - Qi j emp

    were k = 1, .., 313 is the region (or pixel) number, Qi j(k) is

    the charge of electrode j in projection i when region k isfull of high-permittivity material and the rest of the

    sensing area is full of low-permittivity material, while Qi j full

    and Qi j emp are the charge of electrode j in projection i when

    the sensor is full of high- and low-permittivity material,

    respectively.

    The sensitivity maps were calculated for projection 1 using

    the same QuickBasic 4.5 program described in section 3.3,

    which runs of PC-OPERA iteratively, but this time with the

    electrode potentials set for parallel-field according to

    table 3.1, and are shown in figure 3.8. The 12 sensitivity

    maps for projection 1 were then rotated to obtain those for

    projections 2 to 6, giving a total of 72 sensitivity maps.

    Unfortunately, it can be seen in figure 3.8 that the actual

    sensitivity maps for parallel-field excitation bear no

    resemblance whatsoever with the ideal ones of figure 3.5.