7th Lecture. Dimitar Stefanov. Recapping. Three types electrodes are used for sensing of EMG signals: indwelling (intramuscular) electrodes (single fiber electrodes, monopolar electrodes, concentric electrodes) Wire electrodes surface electrodes – non-invasive recordings. - PowerPoint PPT Presentation

Citation preview

Page 1: 7th Lecture

7th Lecture7th LectureDimitar Stefanov

Page 2: 7th Lecture

RecappingRecappingThree types electrodes are used for sensing of EMG signals:1. indwelling (intramuscular) electrodes (single fiber electrodes, monopolar

electrodes, concentric electrodes)2. Wire electrodes3. surface electrodes – non-invasive recordings

Potential of surface electrode (V)

Differential voltage waveform

Page 3: 7th Lecture

Velocity of propagation of the m.a.p. – 4 m/s

There is a delay between the EMG and muscle contraction (30-80 milliseconds).

In case of isometric muscle tension, a linear dependency between the muscle tension and the rectified EMG output is observed.

Important parameters of the EMG amplifiers:1. Gain and dynamic range2. Input impedance3. Frequency response4. Common mode rejection.

EMG signal:• contains certain level of noises• has specific spectral density function.

Fatigue –(1) If we assume that the EMG is stimulation rate remains constant then the muscle tension deceases in case of fatigue. (2) The shape of the m.a.p. is altered in case of fatigue. (3) tremor occurs.

Page 4: 7th Lecture

Problem with the electrodes: polarization• The electric conductivity of the body involves ions as charge carrier.• Electrodes can be considered as electrical conductors in contact with the aqueous ionic solutions of the body.• The interaction between electrons in the electrodes and ions in the body can affect the EMG signal

Half-cell potential (HCP) is called the potential difference between the metal of the electrode and the bulk of the electrolyte.

•HCP depends on the ionic concentration

•HCP can be measured when no electric current flows between an electrode and the electrolyte

Page 5: 7th Lecture

Problem with the electrodes: polarization

Polarization – arises in case when current flows between the electrode and the solution.

Perfectly polarizable electrodes – no actual current crosses the electrode- electrolyte interfaceNonpolarized electrodes – allow the current to pass freely in electrode-electrolyte interface.

Silver – silver chloride electrode (Ag/AgCl)– it possesses characteristics which are similar to a perfect nonpolarizable electrode.

Page 6: 7th Lecture

AgCl film

insulated lead wire

Ag metal

Ag lead wiresintered Ag and AgCl(Ag and AgCl powder mechanically pressed)

greater mechanical stability

Silver – silver chloride electrodes

Low noise electrodes

Page 7: 7th Lecture

Equivalent circuit of a biopotential electrode

Ehc – half-cell potentialRd and Cd – represent the impedance associated with the electrode-electrolyte interfaceRs – series resistance.

Biopotential electrode impedance as a function of frequency

Page 8: 7th Lecture

EMG amplifiersEMG amplifiers

Amplifier gain – the ratio of the output voltage to the input voltage

• Surface EMG electrodes - maximum amplitude of 5 mV peak-to-peak• Indwelling electrodes – amplitude of up to 10 mV• Single m.a.p. electrodes – amplitude of 100 V

Amplitudes of the EMG signal :

Noise level of the amplifier is the amplitude of the higher frequency random signal on the output of the amplifier when the electrodes are shorten together.

Noise level of the amplifier should not exceed 50 V, (preferably 20V).

Page 9: 7th Lecture

Input impedance of an amplifier of biosignalsInput impedance of an amplifier of biosignalsThe resistance of the electrode-skin interface depends on:

thickness of the skin layer, the cleaning of the skin prior to the attachment of the electrodes, the area of the electrode surface, temperature.

Electrode paste – decreases the resistance between the electrode and the skin.

Page 10: 7th Lecture

Input impedance of an amplifier of biosignalsInput impedance of an amplifier of biosignals

The capacitance between the electrode and the skin causes frequency distortions.

EMG amplifiers should possess high input resistance

Page 11: 7th Lecture

Frequency response of the EMG amplifier

Frequency bandwidth Frequency bandwidth

All frequencies present in the EMG should be amplified at one and the same level.

Bandwidth – the difference between upper cutoff frequency f2 and the lower cutoff frequency f1.The gain of the amplifier at f1 and f2 is 0.707 from the gain of the gain in the mid-frequency region (half-power).

Amplifier gain:

Example: linear gain 1000, or 60 dB; gain at the cutoff frequencies – 57 dB (3dB less than that at the mid-frequencies).

Page 12: 7th Lecture

The EMG amplifier should amplify equally all EMG frequency components.

Most of the EMG signals are concentrated in the band between 20 and 200 Hz.

Page 13: 7th Lecture

Recommended range of the EMG amplifiers:• from 10 Hz to 1000 Hz – when the signal is collected with surface electrodes;• from 20 Hz to 2000 Hz – when the signal is collected with indwelling electrodes.

Interferences:Interferences:• Hum from power line (60 Hz in the USA and 50 Hz in Europe)- in the middle of the EMG spectrum• Movement artifacts – their frequency lies in the 0 to 10 Hz range – don’t cause big problems• Noise from low quality cabling systems – interfere with the baseline of the EMG signal; can be eliminated by good low frequency filtering (by setting of f1 to about 20 Hz).

Page 14: 7th Lecture

Influence of the choice of f1 and f2 to the output signal

Common mode rejectionCommon mode rejectionThe human body acts as antenna to pick up any electromagnetic radiation that is present.

Radiation:Radiation: from domestic power lines, fluorescent lighting, and electrical machinery.

Page 15: 7th Lecture

Single-ended amplifierSingle-ended amplifier

Differential amplifierDifferential amplifier

A perfect subtraction never occurs.

Page 16: 7th Lecture

Common mode rejection ratio (CMRR)Common mode rejection ratio (CMRR)

CMRR is measured in dB.

In good quality EMG amplifiers CMRR should be 10,000 (80 dB) or higher.

Page 17: 7th Lecture

Processing of EMGProcessing of EMG

Example:

1. Half of full-wave rectification (absolute value)

2. Linear envelope (low-pass filtering of the rectified signal) –

main decision here is the choice of the low pass filter!

3. Integration of the signal from (2) over the period of the muscle

contraction – area under the curve

4. Integration of the signal from (2) for a fixed time, reset to zero,

and repeating the integration cycle – such scheme represents the

trend of the EMG amplitude with time

5. Integration of the signal from (2) to a present level, reset to

zero, and repeating the integration cycle – represents the level

of the muscle activity (high or low muscle activity).

Page 18: 7th Lecture

Diagram of several common EMG processing systems and the processing results

Page 19: 7th Lecture

Biopotential amplifiersBiopotential amplifiersBasic amplifier requirements:1. The physiological process to be monitored should not be influenced in any way

by the amplifier2. The measured signal should be not distorted3. The amplifier should provide the best possible separation of signal and

interferences4. The amplifier should offer protection of the patient from any hazard and electric

shock5. The amplifier should be protected against damages due to high input voltages.

The input signal to the amplifier consists of 5 components:1. Desired biopotential2. Undesired biopotentials3. A power line interference signal and its harmonics4. Interference signals generated by the tissue-electrode interface5. Noise.

Page 20: 7th Lecture

Block diagram of a biopotential amplifierBlock diagram of a biopotential amplifier

FET transistorsGalvanic decoupling of the patient

Motion artifacts – the contact between the electrode and the tissue changes during the relative motions between the electrodes and the tissue.

Measures for decreasing the motion artifacts:• High input resistance of the amplifier• Usage of non-polarized electrodes (Ag/AgCl)• Reduction of the source impedance by usage of electrode gel.

Artifacts due to electric and magnetic fields – Example.

Page 21: 7th Lecture

Amplitude/frequency characteristics of the Amplitude/frequency characteristics of the bioamplifiers used in different applicationsbioamplifiers used in different applications

Page 22: 7th Lecture

Special circuits which built the biopotential amplifierSpecial circuits which built the biopotential amplifierInstrumentation amplifiers

DCDC instrumentation amplifiers

Page 23: 7th Lecture

AC AC instrumentation amplifiersinstrumentation amplifiersAC amplifiers eliminate the electrode offset potential, permit high gain and permits higher CMRR.

The capacitors between the electrodes and the input stage of the amplifier cause charging effects from the input bias current.

Page 24: 7th Lecture

Isolation amplifierIsolation amplifier

Isolation is realized in the following technologies:• Transformer isolation• Opto-isolation.

Isolation provides a complete galvanic separation between the input stage (patient) and the other part of the measure equipmentmeasure equipment.

Page 25: 7th Lecture

Surge protection of the bioamplifiersSurge protection of the bioamplifiers

Protection of the amplifier from damage due to surge input potentials.

•Diodes•Zener diodes•Gas-discharge tubes

Page 26: 7th Lecture

Input guardingInput guarding

Instrumentation amplifier providing input guarding

Technique for increase both the input impedance of the amplifier of biopotentials and the CMRR

Driven-right-leg circuit reducing common-mode interference.