200323226 1 Chromatography Metods Practically

Embed Size (px)

Citation preview

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    1/151

    1Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Chapter One

    Aim Separation Techniques

    1-Biological fluids are extremely complex in composition.

    2-Chemical analysis would be impossibleif it were necessary to completelyisolate each substance prior to its measurement.

    3- An optimal method tests for a specific substance in the presence of all

    others, requiring no isolation of the substance under analysis.

    4- A test is specific when none of the other substances present interfere.

    However, virtually all chemical tests are subject to at least some

    interference.

    5-This is one of the most important problems in clinical chemistry. Therefore

    some type of separation procedure is required.

    7-Separation in clinical chemistry usually is based on differences in the

    size, solubility or chargeof the substances involved.

    Introduction

    Chromatography is a separation method in which the analyte is

    contained in a mobile phase and pumped through a stationary phase.

    Sample components interact differently with these two phases and elute

    from the column at different retention times tR. Since the first description

    of chromatography by Russian botanical scientist Mikhail Semenovich

    Tswett is discovery of chromatography. He used a column of powdered

    calcium carbonate to separate green leaf pigments into a series ofcolored bands by allowing a solvent to percolate through the column bed.

    Since these experiments by Tswett many scientists have made substantial

    contributions to the theory and practice of chromatography. Not least among

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    2/151

    2Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    these is A. J. P. Martin who received the Nobel Prize in 1952 for the

    invention of partition chromatography (with R. L. K. Synge) and in the

    same year with A. T. James he introduced the technique of gas-liquid

    chromatography. Chromatography is now an important tool used in allbranches of the chemical and life sciences.

    Chromatographic separations can be described quantitatively with a number

    of parameters including the capacity factor k , the selectivity factor , the

    plate number N or height equivalent of a theoretical plate H and the

    resolution RS. The optimum flow rate of a chromatographic separation can

    be determined with the van Deemter equation. In bioanalytical chemistry,

    chromatography is mainly employed for the isolation and purification of

    proteins. Reversed p hase chromatography can separate biomolecules

    according to their interaction with the hydro phob ic stat ionary phase and

    the hydro phi l ic mobl i le phase. This separation method can be coupled to

    an ESI mass spectrometer. Ion exchange chromatography separates

    molecules depending on their net charge. Aff in i ty chrom atographymakes

    use of molecular recognition between biomolecules; and size exclusion

    chromatography allows for the separation of molecules depending on

    their size.

    1-Definition of Chromatography

    Chromatography is essentially a physical method of separation in

    which the components to be separated are distributed between two phases

    one of which is stationary (stationary phase)while the other (the mobilephase)through it in a definite direction.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    3/151

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    4/151

    4Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    chromatography (GSC), liquidliquid chromatography (LLC), and liquidsolid

    chromatography (LSC),

    4- Main Type of Chromatography

    In general, there are four main types which can be classified asfollows:

    4.1-Liquid-Solid chromatography

    Classical adsorption chromatography(Tswett column)

    Ion-exchange chromatography

    4.2. Gas-Solidchromatography

    4.3. Liquid-Liquidchromatography

    Classicalpartition chromatography

    Paper chromatography

    4.4Gas-Liquidchromatography

    5-Separation techniques

    Technique Property Description

    Precipitation Solubility Some of the substances

    precipitate while the others

    remain dissolved

    Ultra-filtration or Dialysis Molecular size Some of the substances

    pass through a layer or

    sheet of porous material

    while the other substances

    are retainedExtraction Solubility Some of the substances

    dissolve (partition) more in

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    5/151

    5Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    water. While other

    substances dissolve more

    organic solvent in contact

    with the waterThin layer

    Chromatography

    or

    Column Chromatography

    Solubility Some of the substances

    dissolve (partition) more in

    the immobile file of water on

    a solid supporting medium

    (or stick more to the

    exposed areas of the solid

    supporting medium) while

    the other substances

    dissolve more in the

    surrounding film of flowing

    organic solvent

    Gas liquid

    Chromatography

    Solubility Some of the substances

    dissolve more in the

    immobile film of wax or oil-

    like material on a solid

    supporting medium. While

    the others dissolve more in

    surrounding stream of

    flowing gas.Gel filtration

    Chromatography

    Molecular Size Some of the substances

    diffuse into the pores in a

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    6/151

    6Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    porous, solid material while

    others remain outside in the

    surrounding stream of

    flowing waterIon-exchange

    Chromatography

    Electrical

    charge

    Some of the substances are

    bound by immobile charges

    on the solid supporting

    medium while others are not

    bound

    Electrophoresis

    Chromatography

    Electrical

    charge

    The substances with more

    charge move faster and,

    therefore, further.

    Substances with opposite

    charges move in opposite

    directions.

    6-Adsorption chromatography

    Adsorption column chromatography is the oldest form of

    chromatography. Whether two or more substances of a mixture can be

    separated by adsorption chromatography depends on a number of factors.

    Most important is the strength with which each component of mixture is

    adsorbed and its solubility in the solvent used for elution.The degree to

    which a particular substance is adsorbed depends on the type of bondswhich can be formed between the solute molecules and the surface of the

    adsorbent.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    7/151

    7Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    5-Purpose of Chromatography

    Analytical- determine chemical composition of a samplePreparative- purify and collect one or more components of a sample

    Other classification of Chromatographic Methods

    Chromatography is classified according to three ways:

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    8/151

    8Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    1. According to the physical state of the mobile phase:

    Liquid chromatography: This subdivided according to the stationary

    phase into liquid-liquid or liquid-solid chromatography.Gas chromatography: This subdivided according to the stationary

    phase into Gas-liquid or Gas-solid chromatography.

    2. According to the method of contact between the mobile phase and

    stationary phase:

    Column chromatography: the stationary phase is placed in a column

    through which the mobile phase moves under the influence of gravity or

    pressure. The stationary phase is either a solid or a thin, liquid film

    coating on a solid particulate packing material or the columns walls.

    Planar chromatography: the stationary phase coats a flat glass, metal, or

    plastic plate and is placed in a reservoir containing the mobile phase

    which moves by capillary action carrying with it the sample components

    3. According to the chemical or physical mechanism responsible for

    separating the samples constituents.(attractive forces)

    Adsorption chromatography: for polar non-ionic compounds

    Ion Exchange chromatography: for ionic compounds

    Anion: analyte is anion; bonded phase has positive charge

    Cation: analyte is cation; bonded phase has negative charge

    Partition chromatography: based on the relative solubility of analyte in

    mobile and stationary phases Normal: analyte is non-polar organic; stationary phase MORE polar

    than the mobile phase

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    9/151

    9Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Reverse: analyte is polar organic; stationary phase LESS polarthan

    the mobile phase

    Size Exclusion chromatography: stationary phase is a porous matrix.

    6-The Principle of ChromatographyChromatography is a separation method where the analyte is contained

    within a liquid or gaseous mobile phase, which is pumped through a

    stationary phase.

    Usually, one phase is hydrophilic and the other lipophilic. The components of

    the analyte interact differently with these two phases. Depending on their

    polarity, they spend more or less time interacting with the stationary phase

    and are thus retarded to a greater or lesser extend. This leads to the

    separation of the different components present in the sample. Each sample

    component elutes from the stationary phase at a specific time, its retention

    time tR(Fig. 1.1). As the components pass through the detector, their signal

    is recorded and plotted in the form of a chromatogram.

    Chromatographic methods can be classified into

    Gas chromatography (GC) and liquid chromatography (LC) depending on

    the nature of the mobile phase involved.

    Gas chromatographycan be applied only to gaseous or volatile substances

    that are heat-stable. The mobile phase, an inert carrier gas such as

    nitrogen, hydrogen or helium, is pumped through a heated

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    10/151

    10Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Figure1-1

    column. This column can be packed with a silicon oxide based material or

    is coated with a polymeric wax. The sample is vaporised, pumped throughthe column and the analytes are detected in the gas stream as they exit the

    column. Analyte detection can be achieved by either flame ionisation or

    thermal conductivity. GC is not commonly used for the analysis of

    biomolecules since large molecular weight compounds such as peptides and

    proteins are thermally destroyed before evaporation. Smaller molecules such

    as amino acids, fatty acids, peptides and certain carbohydrates can be

    analysed if they are modified chemically to increase their volatility. Some cell

    cultures produce volatile metabolites such as aldehydes, alcohols or

    ketones. These can be analysed readily via GC.

    In liquid chromatography, the sample is dissolved and pumped through a

    column containing the stationary phase. LC is more versatile than GC as it is

    not restricted to volatile and heat-stable samples; the sample only has to

    dissolve completely in the mobile phase. Common detection methods are

    UV spectroscopy, measurement of refractive index, fluorescence, electrical

    conductivity and mass spectrometry.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    11/151

    11Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Modes of operation can be classified as normal and reversed phase

    chromatography.

    In normal phase chromatography, the stationary p hase consists of a

    hydrophilic material such as silica particles and the mobile phase is ahydrophobic organic solvent such as hexane.

    In reversed phase chromatography, on the other hand, the stationary

    phase is hydrophobic and the mobile phase is a mixture of polar solvents, for

    example water and acetonitrile. Biomolecules are generally soluble in polar

    solvents; hence, reversed phase chromatography is the method of choice for

    amino acids, peptides, proteins, nucleic acids and carbohydrates.

    7-Comparison of classical and bioanalytical chemistry

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    12/151

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    13/151

    13Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    The optimisation of chromatography is aimed towards completely

    separating all of the components of a sample in the shortest possible time.

    This can, for example be achieved by modifying the composition of the

    mobile phase, choosing a different stationary phase or by changing the flowrate. A typical chromatogram is depicted in (Fig. 1.2).

    Fig. 1.2.Definition of retention time, tR, and peak width, w.

    The sample is injected into the chromatographic column at t = 0 s.

    Substances that are not retarded by the stationary phase leave the column

    at zero retent ion t im e, t0, corresponding to the flow rate of the mobile

    phase. Compounds A and B are retarded by the stationary phase and leavethe column at their retention times tR(A) and tR(B), respectively. The peak

    width, w, is defined as the intersection of the tangents on each side of the

    peak with the baseline. These basic parameters, retention time and peak

    width, can be used to derive a number of other parameters that express the

    quality of the achieved chromatographic separation. In the following

    paragraphs, a brief summary of the most important parameters of

    chromatographic theory are discussed.

    The capacity factor k' (equation1. 1)describes the velocity of the analyte

    relative to the velocity of the mobile phase. Each compound spends a

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    14/151

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    15/151

    15Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    (equations 1.3)

    (equations 1.4)

    The parameters that influence band broadening can be approximated by the

    van Deemter equation (equation 1.5) which is valid for gas and liquid

    chromatography as well as capillary electrophoresis

    (equations 1.5)

    In this simplified equation, the height of theoretical plates, H, is given as a

    sum of three terms. The first term, A,describes the influence of the column

    packing on band broadening. This so-called Eddy diffusion is constantfor

    a given column and independent of the flow rate. The second term, B/u,

    describes the diffusion in or opposed to the direction of flow. This

    longitudinal diffusion is inversely proportional to the flowrate u. The third

    term, Cu, describes the resistance to mass transfer between the stationary

    and mobile phase which is directly proportional to the flow rate. By plotting H

    as a function of u, the optimum flow ratefor a chromatographic separation

    can be determined (Fig. 1.3).

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    16/151

    16Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Fig. 1.3. A van Deemter plot for the determination of the optimum flow rate.

    The ultimate goal of a separation is to achieve a high resolution, Rs,

    (equations 1.6 and 1.7). If Rs = 1.5, then peaks of identical area overlap by

    only 0.3 %, an Rs = 1equals a peak overlap of 4 %. Peak resolution can be

    optimized by increasing the selectivity and minimizing band broadening.

    Resolution

    (equations 1.6)

    valid for < 1.2

    (equations 1.7)

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    17/151

    17Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    As can be seen from equation 1.7, the capacity factor k' has a great

    influence on the resolution. Usually the components in the sample have a

    wide variety of k values. If conditions are optimised such that the firstcompounds to elute have k' values between the optimum of 1 and 5, then the

    other compounds with higher k valueselute much later and show excessive

    band broadening. If, on the other hand, conditions are optimised for the later

    eluting compounds, then the resolution will be poor for the compounds that

    elute first. This general elutionproblem can be overcome by decreasing k'

    during the separation. In LC, the composition of the mobile phase can be

    changed during the separation. This is called a gradient elut ion as

    opposed to an isocratic elution, where the composition of the mobi le phase

    remains unchanged during the separation process. In GC, a temperature

    gradient can be applied during separation rather than operating under

    isothermic conditions. Generally, the first step in trying to achieve a good

    separation of the sample mixture is to choose a stationary phase with which

    the analyte can interact. Then, the composition and gradient of the mobile

    phase can be chosen to optimise the capacity factor and resolution.

    Chromatographic theory as outlined in the above paragraphs can be

    applied to the analysis of smaller molecules such as amino acids,

    peptides and short biopolymers. Care has to be taken for larger

    biomolecules such as high molecular weight proteins. These often show

    different behavior and the theory can only be applied to a limited extent.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    18/151

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    19/151

    19Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    hydroxyl groups of the silica particle with silanes containing non-polar

    hydrocarbon chains. Any chain length from ethyl silane (C2) to n-octadecyl

    silane (ODS) (C18) is used, although octyl silane (C8) and ODS are the most

    commonly employed chain lengths. For analytical separations, the particlesize is typically 5 m or smaller. In preparative liquid chromatography, where

    the goal is to isolate a compound of interest for further analysis or

    investigation, larger particles with a higher capacity and larger column

    diameters are used. The pore size of the silica particles is usually about 10

    nm, resulting in a very large surface area, as much as 100 to 400 m2/g.

    This gives the analytes ample opportunity to interact with the stationary

    phase whilst flowing through the separation column.

    The mobile phase is based on a polar solvent system consisting of an

    aqueous buffer and acetonitrile or methanol. Gradient elution is often

    employed to increase resolution and shorten separation times. This is

    achieved by increasing the organic solvent and thus decreasing the mobile

    phase polarity and the retention of less polar analytes during the separation

    process. Solvents can be classified according to their elution strength and

    polarity (Fig. 1.5).

    Buffer systems based on ammonium acetate, phosphate or hydrogen

    carbonate are usually added at concentrations of about 20 mMto adjust the

    pH of the mobile phase to values between 2 and 8. Ion pairing reagents can

    be used at low concentrations, typically 0.1%, to increase the hydrophobicity

    of charged analytes. They

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    20/151

    20Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Fig. 1.5. Solvents ordered according to polarity and elution speed of the

    analytes.

    Fig. 1.6.Instrumental setup of an HPLC gradient system.

    form ion-pair complexes with the analyte. Anionic ion pairing reagents such

    as trifluoroacetic acid (TFA) bind to positively charged analytes,

    whereas cationic ion pairing reagents such as tetra alkyl ammonium salts

    can be used to bind to negatively charged analytes. These complexes are

    retarded more by the stationary phase and are thus easier to separate than

    the largely unretained charged analytes alone.

    In modern chromatography, the separation columns are tightly packed with

    small particles of about 15m in diameter. To achieve ambient flow rates in

    these columns, high pressures of up to 300400 bars must be generated. A

    typical instrumental setup for this high pressure or high performance liquid

    chromatography (HPLC) is shown in Fig. 1.6.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    21/151

    21Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Computer controlled pumps move the mobile phase through the system.

    Aqueous solvent A and organic solvent B are mixed to the desired

    composition. In the case of gradient elution, the composition is gradually

    altered during the separation.Sample volumes are injected with either a manual loop and valve system or

    automatically via an auto sampler. Depending on the column dimensions

    sample volumes can be as low as several nL and as high as a mL. Often the

    column is situated inside an oven which is thermostatically regulated to

    maintain a constant temperature. After eluting from the column, the analytes

    pass through the detector.

    UV detection using a fixed wavelength could be performed at = 210 nm

    for peptides and = 254 nm or = 280 nm for proteins. More expensive

    instruments have diode array detectors (DAD) which can take several

    whole spectra per second and allow for more unambiguous identification.

    High sensitivity can be achieved via fluorescence detection of derivatised

    amino acids and peptides. A more recent developm ent is to couple

    liquid chromatography systems to an elect rospray ion isation m ass

    spectrometer, ESI-MS.

    Mass spectrometry allows universal detection at very high sensitivity and

    also gives structural information about the analyte. However, not all buffers

    commonly employed for liquid chromatography are compatible with mass

    spectrometers.

    In recent years, there has been a trend to develop ever smaller liquidchromatography systems. LC systems on micro and even nanoscales have

    been demonstrated. Shorter and smaller columns with smaller particles offer

    faster analysis times, decreased solvent consumption and require less

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    22/151

    22Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    sample. The differences between preparative, analytical, micro and nano

    LC are summarized in Table 1.2.

    Table 1.2. Differences between preparative, analytical, micro and nano

    liquid chromatography.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    23/151

    23Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Chapter Two

    ADSORPTION CHROMATOGRAPHYIn adsorption chromatography the compounds to be separated are

    adsorbed onto the surface of a solid material. The compounds are desorbed

    from the solid adsorbent by eluting solvent.

    1-Separation of the compounds depends on

    1-The relativebalance between the aff in i ties of the comp ound s

    for the adsorbent and their solubi l i ty in the solvent.

    2-The chemical nature of the subs tances.

    3-The nature of the solvent.

    4-The nature of the adsorbent.

    Solid adsorbents commonly used are alumina, si l ica gel , charcoal

    (act ive carbo n), cel lulose, starch, and suc ros e.

    Solvents commonly usedare hexane, benzene, petroleum ether, diethyl

    ether, chlo rofo rm, methylene chlo r ide, var ious alco hols (ethyl , pro py l ,

    n-bury l and t-buty l alcohols), and various aqueous b uf fers and sal ts,

    some in combination with organic solvents

    Adsorption chromatography is a column that is packed with the adsorbents.

    The adsorbent is prepared and poured into the column with an inert support

    at the bottom. Suitable supports include plast ic disc s, or sheetsof ny lon

    or Tef lon fabr ics.The adsorbent bed must be homogeneous and f ree of bubb les, cracks,

    or sp aces between the adsorbents and the wal ls of the column .

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    24/151

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    25/151

    25Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Figure 2-1:Collection of fractions from a column by an automatic fraction

    a device that accumulates from an elution column the same predetermined

    volume in each of a series of tubes that automatically change position when

    the proper volume has been collected .This may be accomplished in various

    ways. For example, set volume, with a timer, or by counting drops with adrop counter. The latter is frequently used and is usually the most reliable

    and flexible. The fraction collector may be Equipped with a detection cell

    that automatically measures some parameter of the solution going into the

    tubes and may correlated with fraction number and automatically recorded.

    The detection cell is frequently a small spectrophotometer that can measure

    absorbances at a fixed wavelength or at variable wavelengths. Other

    detecting cell use index of refraction, optical rotation, and other properties.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    26/151

    26Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Figure 2-2: Adsorption chrornatography

    A = adsorbent, S=Sample, ES= eluting solvent

    (A) Application of sample to the column.

    (B) Adsorption of sample onto adsorbent.

    (C)Addition of elution solvent.

    (D) and (E) Partial fraction of sample components.

    (F) Complete fractionation of sample. (G) and (H) Separation of all three

    components at various stages on the adsorbents.

    (I) Elution of the first component from the column.

    The substances adsorbed on the column support can be eluted in threeways

    In the simp lest method, a single solvent continuously flows through

    the column until the compounds have been separated and eluted from

    the column

    Stepwise elution, in which twoor mo re di f ferent solvents of fixed

    volume are added in sequence to elute the desired compounds.

    Gradient elut ion, in which the composition of the solvent is

    continuously changing. The latter method is used to effect separations

    that are difficult because of a tendency of component to be eluted in

    broad. Trailing bands when a single solvent is used. Gradient elution

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    27/151

    27Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    frequently provides a means of sharpening the bands, a simple linear

    gradient has two solvents, A and B, in which A is the starting solvent

    and B is the final solvent. Solvent B is allowed to flow into solvent A as

    solvent A flows into the column. The composition of solvent A is, thus,constantly changing as solvent B is constantly being added to A (Fig.

    2-3).

    Gradients other than linear gradients (e.g., exponent ial , concave. or

    convex) may be obtained by introducing a third vessel and varying the

    composition of the solvents in the vessels. These eluting methods are also

    used with other column chromatographic methods.

    2-Activation of adsorbent

    Many adsorbents such as alumin a, si l ica gel , and act ive carbon and

    Mg si l icatecan obtain commercially, but they require activation before use.

    Activation is achieved by heating and there is usually an optimum

    temperature for activation, for e.g. alumina is about 400oC. For reduced

    activityby the controlled addi t ion of w ater, and the subsequent activity is

    Figure 2-3: Gradient elution. Flow of solvent B into solventAWith mixing,

    continuously changing the composition of solvent Aas it flows into column

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    28/151

    28Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    related to the amount of water added. Brookman and Schodder

    established five grades of aluminaGrade I is the most active and the is

    simply alumina heated at about 3500

    C for several hours. Grade IIhas about

    2-3% water, Grade III5-7%, Grade IV9-11 %, Grade Vfilm. (Least active)

    about 15%.

    3-Retention

    The retention is a measure of the speed at which a substance moves in a

    chromatographic system. In continuous development systems like HPLC or

    GC, where the compounds are eluted with the eluent, the retention is usually

    measured as the retention time Rt or tR, the time between injection anddetection. In interrupted development systems like TLC the retention is

    measured as the retention factor Rf, the run length of the compound divided

    by the run length of the eluent front:

    The retention of a compound often differs considerably between

    experiments and laboratories due to variations of the eluent, the stationary

    phase, temperature, and the setup. It is therefore important to compare the

    retention of the test compound to that of one or more standard compounds

    under absolutely identical Conditions.

    4-Plate theory

    The plate theory of chromatography was developed by Martin and

    Synge.The plate theory describes the chromatography system, the mobile

    and stationary phases, as being in equilibrium. The partition coefficient K

    is based on this equilibrium, and is defined by the following equation:

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    29/151

    29Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    K is assumed to be independent of concentration, and can change ifexperimental conditions are changed, for example temperature is increased

    or decreased. As Kincreases, it takes longer for solutes to separate. For a

    column of fixed length and flow, the retention time (tR) and retention

    volume (Vr) can be measured and used to calculate K

    5- Column chromatography

    1. Small plug of wool (or cotton)

    2. Sandto cover "dead volume"

    3. Silica gel, length = 5.5 - 6 inch (Note 1inch=2.54cm).

    4. Tap column on bech (carefully) to remove air bubbles inside

    the column

    5. Add solvent system

    6. Add sandon top of silica

    7. The top of the silica gel should not be allowed to run dry.

    8. Sample is diluted (20-25% solution)

    9. The sample is applied by pipette

    10. Solvent used to pack the column is reused

    11. Walls of column are washed with a few milliliters of eluant

    12. Column is filled with eluant

    13. Flow controller is secured to column and adjusted 2.0 in / min.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    30/151

    30Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    31/151

    31Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Figure 2-5

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    32/151

    32Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Table 2-1: Common adsorbents and the type of compounds

    Solid Suitable for separation of

    Alumina Steriods, vitamins, ester, and alkaloids

    Silica gel Steriods, amino acids, alkaloids

    Carbon Peptides, carbohydrates, amino acid

    Magnesium carbonate Porphyrins

    Magnesium silicate Steriods, ester, glycerides, alkaloids

    Magnesia Similar to alumina.

    Ca(OH)2 Carotenoids.

    CaCO3 Carotenoids and xanthophylls.

    Ca Phosphate Enzymes, protein, and polynucleotide

    Starch Enzymes.

    Sugar Chlorophyll.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    33/151

    33Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Chapter Three

    PAPER CHROMATOGRAPHY

    Paper chromatography is a type of liquid-liquid partitionchromatography that may be performed by ascending or descending

    solvent f low. Each mode has its advantages and disadvantages.

    Ascending c hromatography involves relatively simple and inexpensive

    equipment compared with descending chromatography and usually gives

    more uniform migration with less diffusion of the sample "spots."

    Descending chrom atography, on the other hand is usually faster because

    gravity aids the solvent flow and with substances of relatively low mobility.

    The solvent can run off the paper. Giving a longer path for migration. To

    resolve compounds with low mobility. Ascending chromatography may be

    performed more than once utilizing a multiple-ascent technique.

    For descending chromatography, papers 22 cm wide and 56 cm

    long can be used. To facilitate the flow of solvent from the paper, the bottom

    of the paper is serrated with a pair of pinking shears. Three pencil lines are

    drawn 25 mm apart at the top of the sheet, and small aliquot of the sample

    (10-50 ml) is placed at a marked spot on the third line. The spot is kept as

    small as possible by adding the aliquot in small increments. With drying in

    between. This may be expedited with a hair dryer. Several samples,

    including standards, are placed 15-25 mm apart.

    The paper is then folded along the other two lines and placed in thesolvent trough of the descending tank (Fig. 3-1).This has been equilibrated

    with solvent beforehand to ensure a saturated atmosphere. The paper is

    irrigated with solvent until the solvent reaches the bottom or for a longer

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    34/151

    34Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    period, allowing the solvent to flow off the end of the paper, if necessary. The

    chromatogram is then removed dried and developed to reveal the locations

    of the compounds. (Part II gives methods of locating carbohydrates, amino

    acids. proteins. nucleotides and nucleic acids and lipids.)In ascending chromatography, a paper approximately 25 cm x 25

    cm is used. A pencil line 20-25 mm from the bottom is drawn across the

    paper

    Fig. 3-1 Steps in descending paper chromatography

    and aliquots (10-50l) of the samples and standards are spotted

    approximately 15-25 mm apart along the line. The spots are dried and the

    paper is rolled into a cylinder and stapled so that the ends of the paper are

    not touching (Fig. 3-2). Solvent is poured into the bottom of a

    chromatographic chamber, and the cylinder is placed inside. The chamber is

    closed and solvent is allowed to flow up

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    35/151

    35Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Fig.3-2 Steps in ascending paper chromatography

    The paper by capillary action. The chamber may be a simple wide-mouth,

    screw top, gallon jar or a cylinder with a ground-glass edge and a glass plate

    top. As with descending chromatography, the chamber should be

    equilibrated with solvent beforehand. Contrary to a popular misconception, if

    the chamber has been sealed and is airtight, the paper does not have to be

    removed as soon as the solvent reaches the top. When multiple ascents are

    performed, the paper is removed, thoroughly dried, and returned to the

    chamber for another ascent of solvent.

    The resolved compounds on a paper chromatogram may be detected by

    their color if they are colored, by their fluorescence if they are fluorescent, by

    a color that is produced from a chemical reaction on the paper after sprayingor dipping the chromatogram with various reagents, or by autoradiography if

    the compounds are radioactive. Identification of compounds on a

    chromatogram is usually based on a comparison with authentic compounds

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    36/151

    36Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    (standards). A quantitative comparison may be made by measuring the Rf,

    which is the ratio of the distance the compound migrates to the distance the

    solvent migrates. A better comparison is the ratio of the distance a particular

    compound migrates to the distance a particular standard migrates. Forexample, in the separation of carbohydrates, the standard might be glucose

    and the ratio would be RGlcor for amino acids, the standard might be glycine

    and the ratio would be RGly

    A useful modification is two-dimensional paper chromatography, in

    which the sample is spotted in the lower left-hand corner and irrigated in the

    first dimension with solvent A. The chromatogram is removed from the

    solvent dried, turned 90, and irrigated in the second dimension with solvent

    B, giving a two-

    Fig. 3-3 Two-dimensional paper or thin-layer chromatography

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    37/151

    37Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    dimensional separation (Fig. 3-3). An application of this procedure has

    been developed for the study of enzyme specificity in which a solution of the

    enzyme is sprayed onto the chromatogram between the first irrigation andthe second to see what products are formed by the action of the enzyme on

    the compounds separated in the first dimension.

    Fig.3-4. Elution of compounds from paper chromatograms for preparative

    chromatography or quantitative determination

    eluted with water. To accomplish the elution, tabs of chromatographic paper

    are stapled to the narrow ends of each strip. As shown in Figure 3.4, one

    end is fitted with two pieces of glass (cut microscope slides), which arc held

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    38/151

    38Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    together with rubber bands, and the bottom end is cut tapered, like a pipet

    tip. This assembly is played so that one end lies in a chromatographic trough

    containing water, and the elution of the strip occurs by capillary flow of the

    water down the paper strip into a baker.Usually less than 1 mL of water is sufficient to effect quantitative

    elution, the samples are quantitatively diluted to a specific volume, and a

    chemical analysis is performed for the specific compound separated. This

    technique also may be used as a preparative procedure to obtain small

    quantities of pure compound from a mixture of compounds.

    In an alternate quantitative procedure, the compounds in the sample are

    radioactively labeled and separated in the usual way, and an autoradiogram

    is prepared. The labeled compounds are located on the chromatogram by

    comparing their positions on the autoradiogram. The radioactive compounds

    are cut out and placed into a liquid scintillation cocktail, and the radioactivity

    is determined by heterogeneous liquid scintillation counting

    In paper chromatography, the mobile phase (solvent) carries the

    components of the sample on the stationary phase (filter paper) separating

    them according to the differences in the migration rate (depends on the

    molecular weight , polarity and adsorption ability)

    Components

    For one-dimensional paper chromatography, either ascending or

    descending development can be carried out in simple units. Descendingdevelopment is more often used because it is faster and more suitable for

    long paper sheets.

    The stationary phase (filter paper)

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    39/151

    39Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    The mobile phase (solvent may be in a reservoir)

    Procedures

    1. Make the initial line on the paper.

    2. Apply the solvent alone on the initial line.3. Wait till the solvent migration is stopped, then make the final line.

    4. Spot the sample, and then apply the solvent either in ascending or

    descending or concentric manner.

    5. In case of colored sample: Calculate the rate of flow (Rf) directly then

    compare it with stander in order to know the unknown sample (qualitatively).

    6. In case of the colorless sample: use UV-lamb to detect the spot

    position then determine the (Rf). Rf depends on the temp., solvent, type of

    paper Rf = distance of sample migration / distance of solvent migration

    Appl icat ions

    1. Separation of amino acids

    2. Separation of the plant pigments

    Advantages

    1. Simple

    2. Cheap

    Disadvantages

    1. Time consuming.

    2. Need high quantity of sample.

    3. With weak solvent power.

    4. limited use5. Difficulty detection of spots

    6. Difficulty isolation of separated substances.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    40/151

    40Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Chapter Four

    Thin layer chromatography

    This technique is particularly useful for the separation of very smallamounts of material. The general pr inc ip le involved is similar to that

    involved in column chromatography, i.e. it is primarily adsorption

    chromatography, although other partition effects may also be involved. A

    glass sheet is covered by a uniform thin layer of an adsorbent. Adsorbents

    used in TLC, di f fer f rom colum n adsorb ents. It contains a binding agent

    such as calcium sulphate, which facilitates the adsorbent sticking to the

    glass plate. The plates are prepared by spreading slurry of adsorbent in

    water over them, starting at one end, and moving progressively to the other

    and then dry ing them in an oven at 100-120C. Dryingserves to remove

    the water and to leave a coating of adsorbent on the plate. Equipment is

    available which will ensure the production of an even coating of adsorbent

    over a series of glass plates. The normal thickness of slurry layer used is

    0.25 mm for qualitative analysis, but layers up to 5-10 mm thick may be

    made for preparative work.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    41/151

    41Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    The sample is applied to the plate by micropipette or syringes, as spot2.5

    cm from one endand at least an equal distance from the edge. The solvent

    is removed from the sample by the use of an air blower. All spots should be

    placed on equal distance from the end of the plate.Separation takes place in glass tank which contains the developing solvent

    (mo bi le phase)to a depth of 1.5 cm, this is allowed to stand for at least 1

    hour with a glass plate over the top of the tank to ensure that the

    atmosphere within the tank becomes saturated with solvent vapor.

    Then, the thin layer plate is placed vert ical lyin the tank so that, it stands in

    the solvent with the end bearing the sample in the solvent.

    The cover plate is replaced and separation of the compounds then occurs as

    the solvent travels up the plate. After the solvent had reached the wanted

    level , the run is s topp ed. The chromatographic separation is completed the

    spots of the component substances can be detected by different

    methods:

    1-Many commercially available TLC adsorbents contain a fluorescent

    dye, the plate is examined under UV light, the separated components

    show up as blu e, gr een, blackarea.

    2. Spraying the plate with 50% sulphur ic acid and heat ing so, the

    compounds become charred and show spots

    3. Spraying the plates with speci f ic color reagents will stain up certain

    compounds e.g. ninh yd r in for am ino acid (aa) , ani l ine for aldoses.

    SolventsUniversal TLC System:

    Petroleum ether - ethyl acetate

    Very polar solvent addi t ives:

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    42/151

    42Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Methanol > ethanol > isopropanol

    Moderately po lar addi t ives:

    Acetonitrile > ethyl acetate > chloroform, dichloromethane > diethyl ether >toluene

    Non-polar so lvents:

    Cyclohexane, petroleum ether, hexane, pentane

    TLC Visualization (Detecting the spots)

    Non-dest ruct ive techniqu es:

    1.Ultraviolet lamp. Shows any UV-active spots

    2.Plate can be stained with iodine.

    Bottle containing silica and a few crystals of iodine (especially good for

    unsaturated compounds)

    Destruct ive techniq ues

    Staining Solutions immerse the plate as completely as possible in the stain

    and remove it quickly. Heat carefully with a heating

    Stains Use/Comments

    Anisaldehyde Good general reagent, gives a range of colors

    PMA Good general reagent, gives blue/greenspots

    Vanillin Good general reagent, gives a range of colors

    Ceric sulfate Fairly general reagent, gives a range of colors

    DNP Mainly for aldehydes and ketones, gives orange

    spots

    Permangante Mainly for unsaturated compounds and alcohols,

    gives yellow spots

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    43/151

    43Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Thin-Layer Chromatography of Amino acidsAmino acids may be separated by two-dimensional TLC using either

    si l ica gel or cel lulos eas the separating medium. Two different solvents are

    used for each type of TLC plate and a different type of separation is

    achieved for each type. The amino acids are visualized with two types of

    ninhydrin sprayfor the silica gel and the cellulose gel media.

    Ninhydr in Sprays for amino acid detect ion

    For silica gel TLC: The plate is sprayed with a solution of 300 mg of

    ninhydrin + 3 ml of glacial acetic acid + 100 ml of butyl alcohol and heated

    for 10 minutes at 110C.

    For cellulose TLC:

    The plate is sprayed with a solution of 500 mg of ninhydrin + 350 ml of

    absolute ethanol + 100 ml of glacial acetic acid + 15 ml of 2,4,6-

    trimethylpyridine and heated for 10 minutes at 110C.

    Two-dimensional TLC separation of amino acids.

    On s i l ica gel Gwith

    Solvent I, chlorolorm-17% methanol (v/v)-ammonia (2:2:1, v/v/v) and

    Solvent II, phenol-water (75:25, v/v).

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    44/151

    44Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    on cel lulose MN 300with

    Solvent III, 1-butanol-acetone-diethylamine-water (10:10:2:5,v/v/v/v, pH

    12.0) and

    Solvent IV, 2-propanol-formic acid (99%)-water (40:2:10, v/v/v, pH 2.5)

    Thin-Layer Chromatography of Carbohydrates

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    45/151

    45Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Carbohydrates may be separated on commercial silica gel plates using

    a variety of solvents to achieve specific separations. The results of the

    separation depend on the particular plate used. Whatman K5 silica gel and

    Merck silica gel 60 plates give good results.Solvent for TLC separations of carbohydrates

    Solvent: Acetonitrile-water (35:15, v/v)with four ascents (45 minutes each

    for a 20-cm plate) will separate mono-, di, and trisaccharides

    The visualization of carbohydrates on thin layer si l ica gel plates is

    obtained by spraying with sul fu r ic acid -methanol (1: 3, v/v) followed by

    heat ing for 10 m in utes at 110-120C. Most carbohydrates give black to

    brown spots on a white background.

    Examples of some TLC separation systems

    Compounds Adsorbent Solvent system (v/v)

    Amino acids Silica Gel G 96% Ethanol/water (70/30)

    Butan-1-ol/acetic acids/

    water (80/20/20)

    Mono and di

    saccharides

    Kieselguhr G (sodium

    acetate)

    Kieselguhr G

    (sodium phosphate

    pH5)

    Ethyl acetate/propan-1-ol

    (65/35). Butan-1-ol /

    acetone/phosphate buffer

    pH5 (40/50/10)

    Neutral lipids Silica Gel G Petroleum ether/diethyl

    ether/acetone (90/10/1)

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    46/151

    46Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Cholesterol

    Esters

    Silica Gel G Carbon tetrachloride/

    chloroform (95/5)

    Carotenoids Kieselguhr G Petroleum ether/propan-1-

    ol (99/1)

    Phospholipids Silica Gel G Chloroform/methanol/water

    (65/25/4)

    Advantages of TLC.

    The speedat which separation is achieved. With a volatile solventas

    the mobi le phasethe time involved may be as low as 30 minu tes, but even

    with non-volatile solvents the time involved is rarely longer than 90

    minutes.

    Summ ary for TLC

    Principle

    As in paper chromatography

    Components glass or plastic plate: as a support to the stationary phase

    stationary phase (silica gel, alumina or agar)

    mobile phase solvent system

    Procedures

    (a) Preparing the plate

    1- Prepare a glass plate. for example (20X20)Cm

    2- Dissolve suitable amount of the silica gel in water path.

    3- Spread it on glass plate homogeneously. Then wait till solidification.

    (b) Running the sample

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    47/151

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    48/151

    48Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    2- Limited to non-volatile compounds

    3- Less accurate and less sensitive

    Advantages1- Need small quantity of sample.

    2- With greater solvent power.

    3- easy detection of spots

    4- easy isolation of separated substances

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    49/151

    49Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    The procedure of two-dimensional thin-layer chromatography

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    50/151

    50Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Developing solvent mixtures that have been recommended for two dimensional TLC separation of

    underivatised amino-acids

    Organic component of the solvent continues migrating, thus forming the mobile

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    51/151

    51Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    phase. Therefore, compounds soluble to organic component move faster than

    compounds soluble to aqueous component. -Thus, molecules are separated

    according to their polarities.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    52/151

    52Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Chapter FiveGel filtration

    Biomolecules are purified using chromatography techniques that separate

    them according to differences in their specific properties, as shown in Figure

    5.1.and Table 5.1.

    Property Technique

    Size Gel filtration (GF), also called size

    exclusion

    Charge Ion exchange chromatography (IEX)

    Hydrophobicity Hydrophobic interaction

    chromatography (HIC)

    Reversed phase chromatography

    (RPC)

    Biorecognition (ligand

    specificity)

    Affinity chromatography (AC)

    Table 5.1.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    53/151

    53Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Fig. 5.1Separation principles in chromatography purification.

    Gel filtration has played a key role in the puri f icat ion of enzymes,

    polysacchar ides, nucle ic acids, prote ins and other bio logical

    macromolecules. Gel filtration is the simplest and mildest of all the

    chromatography techniques and separates molecules on the basis of

    differences in size. The technique can be applied in two distinct ways:

    1. Group separations:

    The components of a sample are separated into two major groups

    according to size range. A group separation can be used to remove high orlow molecular weight contaminants(such as phenol red from culture fluids)

    or to desalt and exchange buffers.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    54/151

    54Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    2. High resolution fractionation of biomolecules:

    The components of a sample are separated according to di f ferences

    in their mo lecular size.High resolution fractionation can be used to isolateone or more components, to separate monomers from aggregates, to

    determine m olecular w eightor to perform a molecular weight distribution

    analysis.

    Gel filtration can also be used to facilitate the refolding of denatured proteins

    by careful control of changing buffer conditions.

    Gel filtration is a robust technique that is well suited to handling biomolecules

    that are sensitive to changes in pH, concentration of metal ions or co-

    factors and harsh environmental conditions. Separations can beperformed

    in the presence of essential ions or cofactors, detergents, urea,

    guanidine hydrochloride, at high or low ionic strength, at 37 C or in the

    cold roomaccording to the requirements of the experiment

    Gel filtration in practice

    Gel filtration separates molecules according to differences in size as

    they pass through a gel filtration medium packed in a column. Unlike ion

    exchange or affinity chromatography, molecules do not bind to the

    chromatography medium so buffer composition does not directly affect

    resolution (the degree of separation between peaks).

    Separation by gel filtrationGel filtration medium is packed into a column to form a packed bed. The

    medium is a porous matr ixin the form of sphericalparticles that have been

    chosen for their chemical and physical stabi l i ty, and inertness (lack of

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    55/151

    55Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    reactivity and adsorptive properties). The packed bed is equilibrated with

    buffer which fills the pores of the matrix and the space in between the

    particles. The liquid inside the pores is sometimes referred to as the

    stat ionary phaseand this liquid is in equilibrium with the liquid outside theparticles, referred to as the mobi le phaseas shown in Figure 2.

    Gel filtration is used in group separation mode to remove small

    molecules from a group of larger moleculesand as a fast, simple solution for

    buffer exchange. Small molecules such as excess salt (desalting) or free

    labels are easily separated. Samples can be prepared for storage or for

    other chromatography techniques and assays. Gel filtration in group

    separation mode is often used in protein pur i f icat ion schemes for

    desalting and buffer exchange

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    56/151

    56Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    57/151

    57Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Fig. 5. 2.Common terms in gel filtration

    Sephadex G-10, G-25 and G-50 are used for group separations. Large

    sample volumes up to 30% of the total column volume (packed bed) can be

    applied at high flow rates using broad, short columns. Figure 3 shows the

    elution profile (chromatogram) of a typical group separation. Large

    mo lecules are eluted in or just af ter the void volum e, Voas they pass

    through the column at the same speed as the flow of buffer. For a well

    packed column the void v olum e is equivalent to approximately 30% of

    the tota l colum n volum e. Small molecules such as salts that have fullaccess to the pores move down the column, but do not separate from each

    other. These molecules usually elute just before one total column volume,

    Vt, of buffer has passed through the column. In this case the proteins are

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    58/151

    58Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    detectedby monitoring theirUV absorbance, usually atA280 nm, and the

    sal ts are detectedby monitoring the cond uct iv i ty of the buf fer.

    Fig. 5.3. Typical chromatogram of a group separation. The UV (protein) and

    conductivity (salt) traces enable pooling of the desalted fractions and

    facilitate optimization of the separation.

    The theoretical elution profile (chromatogram) of a high resolution

    fractionation. Molecules that do not enter the matrix are eluted in the void

    volume, Voas they pass directly through the column at the same speed as

    the flow of buffer. For a well packed column the vo id vo lumeis equivalent to

    approximately 30% of the total column volume (packed bed). Molecules with

    partial access to the pores of the matrix elute from

    the column in order of decreasing size. Small molecules such as salts that

    have full access to the pores move down the column, but do not separate

    from each other. These molecules usually elute just before one total column

    Sample:(His)6 protein eluted from HiTrap

    Chelating HP with

    sodium phosphate 20 mM,

    sodium chloride 0.5 M,

    Imidazole 0.5 M, pH 7.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    59/151

    59Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    volume, Vt, of buffer has passed through the column, Fig. 5.4.

    Fig. 5.4.Theoretical chromatogram of a high resolution fractionation (UV

    absorbance).

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    60/151

    60Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Separation examples

    Fig. 5.5 Cytochrome C, Aprotinin, Gastrin I, Substance P,

    (Gly)6, (Gly)3and Gly

    Media Selection

    Chromatography media for gel filtration are made from p orous

    matr iceschosen for their inertness and chemical and physical stability. The

    size of the pores within a particle and the particle size distribution are

    carefully controlled to produce a variety of media with different selectivities.

    Today'sgel filtration media cover a molecular weightrange from 100 to80 000 000, from peptides to very large proteins and protein complexes.

    Figure.5.7.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    61/151

    61Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Sephacryl is suitable for fast, high recovery separations at laboratory and

    industrial scale

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    62/151

    62Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Sephadexis ideal for rapid group separations such as desalting and buffer

    exchange.

    Sephadexis used at laboratory and production scale, before, between or

    after other chromatography purification steps.Determination molecular weight

    VeV0

    Kav= --------------

    VtV0

    WhereVe= elution volume for the proteinVo= column void volume

    Vt= total bed volume

    On semi logarithmic graph paper, plot the Kav value for each protein

    standard (on the l inear scale) against the corresponding molecular

    weight (on the logar i thmic scale). Draw the straight line which best fits the

    points on the graph. Then, calculate the corresponding Kav for the

    component of interest and determine its molecular weight from the

    calibration curve.

    Sephadex:

    Rapid group separation of high and low molecular weight substances,

    such as desalting, buffer exchange and sample clean up

    Sephadex is prepared by cross-linking dextran with epichlorohydrin.

    Variations in the degree of cross linking create the different Sephadex media

    and influence their degree of swelling and their selectivity for specific

    molecular sizes (Table.5.2).

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    63/151

    63Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Product Fractionation

    range, Mr

    (globularproteins)

    pH stability Bed

    volume

    ml/gdry

    Sephad

    ex

    Particle size,

    wet

    Sephadex G-

    10

    700)from smaller molecules (Mr 30000 frommolecules Mr

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    64/151

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    65/151

    65Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Therefore, The technique chosen must discriminate between the target

    protein and any remaining contaminants

    Gel Filtration = Gel Permeation Chromatography =Size Exclusion Chromatography

    Size exclusion chromatography (SEC), also called gel permeation

    Chromatography (GPC) or gel filtration chromatography (GFC) is a

    technique for separates molecules according to their molecular size. Gel

    particles form the stationary phase of this type of chromatography; the

    mobile phase is the solution of moleculesto be separated and the eluting

    solvent, which most frequently is water or a dilute buffer. The sample is

    applied to the gel, if the molecules are too large for the pores; they never

    enter the gel and move outside the gel bed with the eluting solvent. Thus,

    the very large molecules in a mixture move the fastest through the gel bed

    and the smaller molecules, which can enter the gel pores, are retarded and

    move more slowly through the gel bed. In gel chromatography, molecules

    are, therefore, eluted in order of decreasing molecular size

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    66/151

    66Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Fig.5.11Gel permeation chromatography. Open circles represent porous gel

    molecules: large solid Circles represent molecules too large to enter the gel

    through the pores, and smaller solid circles represent molecules capable of

    entering the gel pores

    Three types of polymers are principally used-dextran, polyacrylamide,

    and agarose

    Dextranis a polysaccharide composed of (-1--->6)-linked glucose residues

    with (-1, 3) branch linkages. It is synthesized from sucroseby an enzyme

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    67/151

    67Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    produced by the bacter ium Leuconostoc mesenteroides B -512F. The

    dextran is cross-linked to various extents by reaction with epichlorohydr in

    to give gel beads with different pore sizes Fig 5.12. Cross-linked dextrans

    are commercially produced by Pharmacia Fine Chemicals, lnc ., (Uppsala,

    Sweden),and sold under the trade name Sephadex. Sephadex gels in th e

    so -cal led G-ser ies, where the G-numbers refer to the amou nt of w ater

    gained wh en the beads are sw el led in water (Table 1) have different

    degrees of cross-linking, hence different pore sizes. This gives gels that

    have capabilities of separating different ranges of molecular weights and

    have different molecular exclusion limits. The exclusion limit is the molecular

    weight of the smallest peptide or globular protein that will not enter the gel

    pore. Sephadex G-10, the highest cross-linked dextran, has a water regain

    of about 1mL/g of dry gel and Sephadex G-200, the lowest cross-linked

    dextran, has a water regain of about 20 mL/g of dry gel. In the swelling

    process, the gels become filled with water.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    68/151

    68Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Fig.5.12. Structure ofepichlorohydr in c ross l inked Dextran

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    69/151

    69Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Table 5.1: Properties of gels used in gel permeation (filtration)

    chromatography

    Gel

    Waterregain

    (mL/g)

    Exclusionlimit

    Maximumhydrostatic

    pressure

    cm H2O

    Maximumflow rate (ml,

    min)

    Sephadex G-10 1.0 700 200 100

    Sephadex G-15 1.5 1500 200 100

    Sephadex G-25 2.5 5000 200 50

    Sephadex G-50 5.0 30000 200 25

    Sephadex G-75 7.5 70000 160 6.4

    Sephadex G-100 10.0 150000 96 4.2

    Sephadex G-150 15.0 300000 36 1.9

    Sephadex G-200 20.0 600000 16 1.0

    Sepharose 6B NA 4 x 10 200 1.2

    Sepharose CL 6B NA 4 x 10 >200 2.5

    Sepharose 4B NA 20 x 106 80 0.96

    Sepharose CL 4B NA 20 x 10 120 2.17

    Sepharose 2B NA 40 x 10 40 0.83

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    70/151

    70Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Sepharose CL 2B NA 40 x 10 50 1.25

    Bio-Gel P-2 1.5 1800 >100 110

    Bio-Gel P-4 2.4 4000 >100 95

    Bio-Gel P-6 3.7 6000 >100 75

    Bio-Gel P-10 4.5 20000 >100 75

    Bio-Gel P-30 5.7 40000 >100 65

    Bio-Gel P-60 7.2 60000 100 30

    Bio-Gel P-100 7.5 100000 100 30

    Bio-Gel P-150 9.2 150000 100 25

    Bio-Gel P-200 14.7 200000 75 11

    Bio-Gel P-300 18.0 400000 60 6

    Bio-Gel A-0.5m NA 500000 >100 3

    Bio-Gel A-1.5m NA 1.5 x 10 >100 2.5

    Bio-Gel A-5m NA 5 x 106 >100 1.5

    Bio-Gel A-15m NA 15 x 106 90 1.5

    Bio-Gel A-50m NA 50 x 10 50 1.0

    Bio-Gel A-150m NA 150 x 10

    6

    30 0.5

    Bio-Gel is a trade name of Bio-Rad Laboratories, Sephadex and

    Sepharose are trade name of Pharmacia Fine Chemical

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    71/151

    71Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Polyacrylamide gels are long polymers of acrylamide cross-linked withN.N'methylene-bisacrylamide (Fig. 5.13).

    Fig.5.13. Structureof cross-linked polyacrylamide

    The gels are commercially produced by BioRad Laboratories, Richmond.

    California, as the Bio-Gel P series. Like the Sephadex G series. the Bio-

    Gels differ in degree of cross-linking and in pore size; the Bio-Gels,

    however. have a wider range of pore sizes than is available in the Sephadex

    G seriesfor the exclusion limits and properties of the different Bio-Gels.

    Agaroseis a gel material with pore sizes larger than cross-linked dextran or

    polyacrylamide. Agarose is the neutral polysaccharide fraction of agar. It is

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    72/151

    72Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    composed of a linear polymer of D-galactopyranose linked ( 1->4) 3,6

    anhydro-L-galactopyranose, which is linked (1-> 3)(Fig.5. 14).

    D-galactose (-1->4) 3, 6-Anhydro-L-galactoseFig.5.14. Structure of the repeating unit of agarose, D-galactopyranose

    linked (-1->4) to 3, 6-anhydro-L-galactopyranose, which is linked (-1-3) tothe next D-galactopyranose residue

    When the polysaccharide is dissolved in boiling water and cooled, it forms a

    gel by forming inter-and intramolecular hydrogen bonds. The pore sizes are

    controlled by the concentration of the agarose. High molecular weight

    materials such as protein aggregates, chromosomal DNA, ribosomes,

    viruses, and cells have been fractionated on agarose gels. Bio-Rad markets

    the agarose Bio-Gel A serieswith different molecular exclusion limits, and

    Pharmacia markets agarose as Sepharose and Sepharose CL. The latter

    is Sepharose cross-linked by reacting with alkaline 2, 3-dibromopropanol

    to give an agarose gelwith increased thermal and chemical stability. Table

    5.1gives the properties of the different Sephadex, Bio-Gel, and Sepharose

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    73/151

    73Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    gels. The separations that may be achieved by gel permeation

    chromatography are based on differences in the molecular sizes of the

    molecules. The method is used for both preparative and analytical purposes.

    The latter has been especially useful in determining the molecular weights of

    proteins. The proteins are chromatographed on a gel column and the elution

    volume of the protein determined. Proteins with known molecular weights

    are also chromatographed and the elution volumes determined. Then, from a

    plot of log molecular weight versus elution volume, the molecular weight of

    an unknown protein may be determined (Fig. 5.15).

    Fig.5.15.Molecular weight determination of proteins by gel permeation chromatography

    using Sephadex G-100 as the gel bed: log molecular weight is plotted versus elution

    volume.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    74/151

    74Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Gel chromatography provides a rapid and mild method of removing salts and

    other small molecules from high molecular weight biomolecules. The sample

    containing the biomolecules and the salt is passed over a gel column whose

    exclusion limit is below the molecular weight of the biomolecules. The

    biomolecules which do not enter the gel emerge in the void volume of the

    column, while the salts enter the gel and are retarded, and therefore are

    removed from the biomolecules.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    75/151

    75Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Summary Gel filtration chromatography

    (Size-exclusion chromatography)

    Principle

    This technique separate proteins according to their size and shape, as theypass through a stationary phase (cross-linked polymer =sephadex) by the

    help of mobile phase (without binding). Larger proteins or molecules, which

    can not penetrate the sephadex pores, move around the sephadex in space

    between them faster than the smaller molecules which may penetrate the

    sephadex pores taking long time to elute from the column.

    Components

    1. Column: as a support to the stationary phase

    2. Stationary phase (pours matrix in the form of spherical particles,

    stable, inert e.g. sephadex or agarose)

    3. Mobile phase (buffer system)

    Procedures

    1. (Loading step): spherical particles of the sephadex are packed into the

    column

    2. (Sampling step): sample is applied to the column

    3. Buffer (mobile phase) and sample move through the column. The

    sample components diffuse in and out of the pores of the matrix (sephadex)

    according to their size.

    4. Larger proteins or molecules move faster than the smaller molecules

    and leave the column first5. Separation completed as the entire buffer volume is passed.

    Applications

    1. Separation of neutral proteins and larger molecules including polymers

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    76/151

    76Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    and biomolecules according to size.

    2. The determination of formula weights.

    Disadvantages1. Limited applications

    2. Low purification

    Advantages

    1. Provides a rapid means for separating larger molecules

    2. Use only one buffer (coast effective)

    3. Do not need elution step because there are no bonds formed.

    Note: Gel Filtration

    Separation based on size

    Molecular sieve chromatography

    Size exclusion chromatography

    Media composed of crosslinkedpolymers

    Pore size of matrix determines degree of interaction

    Larger moleculesare excluded and migrate faster

    Smaller moleculesare included and are retained longer

    Dextran (=Sephadex)

    Agarose (=Sepharose)

    Polyacrylamide choose matrix with desired characteristics

    Size range

    does not interact with solute include 0.15-1 M NaCl in buffer

    Load sample in smallest possible volume

    elute in one column volume

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    77/151

    77Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Practical Considerations

    Sephadex

    Code Range (kDa)

    G-25 1-5, G-50 2-30, G-100 4-150, G-150 5-300, G-200 5-600Applications:

    Purification

    Desalting

    Size determination

    Calculating Size

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    78/151

    78Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Chapter Six

    Ion-exchange chromatography

    Ion-exchange chromatography is a variation of adsorption

    chromatography in which the solid adsorbent has charged groups chemically

    linked to an inert solid. Ions are electrostatically bound to the charged

    groups; these ions may be exchanged for ions in an aqueous solution. Ion

    exchangers are most frequently used in columns to separate molecules

    according to charge. Because charged molecules bind to ion exchangers

    reversibly. Molecules can be bound or eluted by changing the ionic strength

    or pH of the eluting solvent.

    Two types of ion exchanger are available: those with chemically

    bound negat ive charges are cal led cat ion exchangers and those with

    chemically bound posi t ive charges are cal led anion exc hangers. The

    charges on the exchangers are balanced by counterions such as chloride

    ions for the anion exchangers and metal ions for the cation exchangers.

    Sometimes buffer ions are the counterions. The molecules in solution which

    are to be adsorbed on the exchangers also have net charges which are

    balanced by counterions. As an example of an ion-exchange process, let us

    say that the molecules to he adsorbed from solution have a negative charge

    (X-), which is counterbalanced by sodium ions (Na+). Such negatively

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    79/151

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    80/151

    80Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Cellulose and cross-linked dextran (Sephadex) are used as the

    solid supports and charged groups such as diethylaminoethyl (DEAE) or

    carboxymethyl (CM) are chemically linked to them to give anion and

    cat ion and the exchangers respect ively. The preparation and commercial

    availability of these materials beginning in the 1960 provided the biochemist

    with powerful tools for separat ion of p roteins and nucle ic acid Figure 2

    presents partial structures of DEAE-cellulose and CMcellulose

    Figure 6.2. Partial structures of diethylaminoethyl-cellulose and carboxymethyl-

    cellulose.

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    81/151

    81Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    The DEAE and CMgroups are shown attached to the C6-hydroxyl group of

    glucose. The DEAE and CM groups are also found attached to the hydroxyl

    groups of C2and C3. The total degree of substitution of the DEAE and CM

    groupsmust be less than one group per five glucose residues to maintain a

    water-insoluble product.

    Table 6.1. Pretreatment steps for DEAE-cellulose and CM -cellulose ion exchangers

    Cellulose First treatment Intermediate

    pH

    Second treatment

    DEAE 0.5 M HCl 4 0.5 M NaOH

    CM 0.5 M NaOH 8 0.5 M HCl

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    82/151

    82Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    The dry ion-exchange celluloses are pretreated with acid and base to swell

    the exchangers so that they become fully accessible to the charged

    macromolecules in solution. The weighed exch anger is sus pended in 15

    volumes (w/v) of the "f i rs t treatment," acid o r alkali depending on th e

    exchanger (Table. 1),and is al low ed to stand at least 30 m inutes but no t

    mo re than 2 hours. The supernatant is decanted and the exchanger is

    washed until the effluent is at the "intermediate pH" The exchanger is stirred

    into 15 volumes of the "second treatment" and allowed to stand for an

    additional 30 minutes. The second treatment is repeated and the exchanger

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    83/151

    83Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    is washed with distilled water until the effluent is close to neutral pH. The

    treated exchanger is placed into the acid component of the buf fer (the pH

    shou ld be less than 4.5) and degassed under vacuum 10 cm Hg

    pressure) with stirring, until bubbling stops The exchanger is then titrated

    with the basic component of the buffer to the desired pH, filtered, and

    suspended in fresh buffer to complete the pretreatment. The exchanger is

    ter) above the

    settled exchanger are removed by decantation. Buffer is added to the

    exchanger so that the final volume of the slurry is l50% of the settled wet

    volume of the exchanger. The column is then packed with the slurry of the

    exchanger, the sample is applied, and elution is performed as described for

    adsorption chromatography.

    Three general methods are used for eluting molecules from the

    exchanger:

    (a)Changing the pH o f the buf fer to a value at which binding is weakened

    (i.e., the pH is lowered for an anion exchanger and raised for a cation

    exchanger),

    (b)Increasing the ionic st rength by increasing the concentration of salt in

    the elution solvent, thereby weakening the electrostatic interactions between

    the adsorbed molecule and the exchanger, and

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    84/151

    84Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    (c)Performin g aff ini ty elut ion. In affinity elution the adsorbed molecule is

    usually a macromolecule that is desorbed from the affinity ligand by adding

    a molecule that is charged and of opp osi te signs to the net chargeon

    the macromolecule and has a specific affinity for the macromolecule. Thus,

    the reduction of the net charge on the macromolecule weakens its

    electrostatic interaction with the exchanger sufficiently to permit the elution of

    the macromolecule from the affinity ligand.

    The stages of anion exchange chromatography.

    An example of the use of ions exchange resins

    Is the purification of Cytochrome C:

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    85/151

    85Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Cytoch rom e C has an isoelectr ic p oint (pI) of 10.05; that is at pH

    10.05 the number of positive charges will equal the number of negative

    charges. A columncontaining a cation exchangerbuffered, at pH 8.5, is

    prepared. This columnhas a full negative charge. Cytochrome Cat pH

    8.5 has a full positive charge. An Impure solution of Cytochrome C at

    pH 8.5 placed on the column, and water is passed through the

    column (the pI of proteins is usually 7.0 or less) but Cytochrome C is

    held firmly by electrostatic attraction to the resin heads. If the eluting

    solvent pH is raised to about 10, the Cytochrome C will now has a

    net zero charge and will pass rapidly through as a pure component

    Summary Ion-exchange chromatography

    Principle

    Ion exchange chromatography separates molecules (proteins) according to

    their differences between the overall charges. The proteins to be separated

    must have a charge opposite to that of stationary phase in order to bind. Ion

    exchange has two types according to the stationary phase charge:

    1. Cation-exchanger: in which the stationary phase is charged

    negatively in order to binds with positive molecules (cations)

    2. Anion-exchanger: in which the stationary phase is charged positively

    in order to binds with negative molecules (anions)

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    86/151

    86Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    A-Cation-exchange chromatography

    Cation-exchange chromatography can be classified as: either strong or

    weak. A strong cation exchanger contains strong acid which stable along

    pH1-14. Whereas, weak cation exchanger contains weak acid which loss its

    charge as the pH decrease below 4-5

    The sample must be charged positive in order to bind with the negative

    matrix (strong or weak acid). H+

    B-Anion-exchange chromatography

    Anion-exchange chromatography can be classified as: either strong or

    weak. A strong anion exchanger contains strong base which stable along

    pH1-14. Whereas, weak anion exchanger contains weak base which loss its

    charge as the pH increase over 9

    The sample must be charged negative in order to bind with the positive

    matrix (strong or weak base).OH-

    Components

    1. The column containing the stationary phase (anion or cation

    exchanger) on suitable matrix

    2. Washing and eluting buffer

    3. pump to withdrew the buffer

    4. Detector

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    87/151

    87Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    Procedures

    Before carry out the process, you must answer two important questions:

    a) What is the sample charge? If +Ve, use cation exchanger. ifVe, use

    anion exchanger

    b) What is the suspected strength of the charge? If weak +Ve, use weak

    cation exchanger, if strong +Ve, use strong cation exchanger, if weak Ve,

    use weak anion exchanger, if strongVe use strong anion exchanger.

    e.g.the sample is weak negative proteins. So we will use anion exchanger

    contain weak base.

    1. (Loading step): the column is packed with the matrix that charged with

    weak positive charge by adding weak base e.g. DEAE- cellulose (stationary

    phase)

    2. (Sampling step): apply the sample in the column: the negatively

    charged proteins bind to positively charged matrix whereas; the positively

    charged proteins flow down to the exterior. Some negative charged

    contaminants can bind to matrix.

    3. (Washing step): apply washing buffer (Tris-HCL) to remove the

    contaminants remaining the target proteins.

    4. (Elution step): now, we need to separate the target proteins from the

    matrix, so we apply an eluting buffer that has the same charge of protein in

  • 8/11/2019 200323226 1 Chromatography Metods Practically

    88/151

    88Chromatography CourseDr Ehab Aboueladab-Lecturer of Biochemistry-Mansoura University-Branch Damietta

    order to substitutes it (ion exchange). Separation can be done also by ion

    exclusion and ion pairing.

    5. (Gradient step): make gradient elution with different buffer till you

    obtain 100% correct proteins. i.e. repeat washing and eluting steps with

    different buffer

    6. (Detection step): after separation carry out detection by electrophoresis

    Applications