19 Molecular Spectroscopy

Embed Size (px)

Citation preview

  • 7/29/2019 19 Molecular Spectroscopy

    1/8

    UV/Vis absorption by organic compounds

    requires that the energy absorbed correspondsto a jump from a populated orbital to an

    unpopulated one.

    !"

    #"

    n

    #*

    !*

    bondingbonding

    non-bonding

    anti-bonding

    anti-bonding

  • 7/29/2019 19 Molecular Spectroscopy

    2/8

    N

    N

    N

    N

    Vapor phase spectrum

    Here we have no

    solvent and minimal

    interaction betweenthe molecules

    Vapor phase

    In hexane

    In water

  • 7/29/2019 19 Molecular Spectroscopy

    3/8

    octahedral tetrahedral square planar

    The type and degree of splitting depends on the

    ligand and geometry of the complex that is formed.

    This splitting of the d orbitals results in a d->d

    transition that is in the UV/Vis range.

    Chromium(III) examples

    Ligand $maxCl- 736

    H2O 573

    NH3 462

    CN- 380

    N

    N

    3 + Fe2+ Fe2+

    N

    N

    3

    1,10-phenanthroline ferroin

    IR radiation is of too low an energy to excite

    electronic transitions.

    Absorption is limited to vibrational and

    rotational levels.

    For liquids and solids, molecular rotation isoften limited so the major type of interaction is

    vibrational.

  • 7/29/2019 19 Molecular Spectroscopy

    4/8

    symmetricalstretching

    asymmetricalstretching

    scissoring

    Due to the large number of vibrational states, IR

    spectra can be very complex.

    2 4 6 8 10 12 14

    wavelength, mC-H

    stretchC=O

    stretch

    C-Hbend

    CH2rock

    The real strength of IR is its ability to identify

    functional groups.

    Functional wavenumber wavelength

    Group (cm-1) (m)

    C-H, aliphatic 3000-2850 3.3-3.5

    C-H, aromatic 3150-3000 3.2-3.3

    O-H 3600-3000 2.8-3.3

    C=O, aldehyde/ketone 1740-1660 5.7-6.0

    -CH2Cl 1300-1200 7.6-8.2850-890 13.2-14

  • 7/29/2019 19 Molecular Spectroscopy

    5/8

    R2NCS

    S- Na+n + M

    n+ R2NCS

    S

    M

    n

    dithiocarbamate

    You want to make your measurement at a $max to

    minimize errors and achieve maximum sensitivity.

    Small error

    Large error

    identical variationsin wavelength.

    The relationship between concentration and

    absorption must be established.

    Your methodshould only beused in thisrange.

    Weve already covered the basics of absorbance

    calculations.

    Lets look at another type of absorbance

    application - using variations in absorbance

    during a titration.

    This is an alternate to using an indicator and

    especially useful for complexometric titrations.

  • 7/29/2019 19 Molecular Spectroscopy

    6/8

    Experimental

    setup

    The absorbance of a solutioncan be monitored during atitration.

    In this example, stirring will

    cause the solution to circulatein the cuvette and theresponse can be measured.

    You could also use a pumpor simply transfer a sample atknown intervals.

    spectrophotometer

    stirbar

    The type of titration curve you obtain is

    dependent on the reaction involved.

    In this example

    - the sample does not absorbat the measured wavelength

    - the titrant will absorb

    equivalence point

    Determine what would cause these types oftitration curves.

    Normal relaxation process

    absorption relaxation

    Large jumps are

    called internal

    conversion.

    The process in

    well understoodbut is temperature

    dependenth%"

    h%"

    Fluorescence process

    absorption fluorescence

  • 7/29/2019 19 Molecular Spectroscopy

    7/8

    Just because a molecule demonstrates

    fluorescence does not mean that there is nointernal conversion. Both processes occur.

    The % of molecules that fluoresce is expressed

    as the quantum yield.

    Q =

    We can consider all species can fluoresce but

    for most, Q is very small.

    rate of fluorescent relaxationtotal relaxation rate

    Fluorescence spectra tend to look like

    overlapping, mirror images of theoriginal absorption spectra.

    The energy range is always lowerbecause some energy is always lost

    to vibrational modes.

    emission

    excitation

    300 350 400

    Remember that the top

    spectra is absorption

    and the bottom isemission.

    Now we not only need to

    deal with the $max for

    absorption but theemission $max.

    Not many compounds undergo significant

    fluorescence.

    Aromatic fused ring structures are best.

    Highly conjugated double bonds will also show

    the effect to a smaller degree.

    Basically, we need a rigid structure with limitedvibrational modes of relaxation.

    Being an emission technique, it can be very

    sensitive and have low detection limits.

    Unfortunately, not many species undergo

    fluorescence on their own.

    We can extend the method by forming complexes

    using fluorescent ligands or adding fluorescent

    groups to other organic molecules.

    OH

    N C|

    8-hydroxyquinoline

    fluorene

  • 7/29/2019 19 Molecular Spectroscopy

    8/8

    These are emission methods.

    Fluorescent intensity = K c

    It best to produce a calibration curve.

    Must consider both the excitation and emission

    $max values.

    Look for the most sensitive combination.

    intensity

    concentration

    Calibration curves are often linear over a 106 range.

    Self-quenching

    region