1060. QUALITY AND AUTOMATION IN THE DETERMINATION OF THE ERYTHROCYTE SEDIMENTATION RATE by Paolo Galiano. In the last 5 years many scientific works on the TEST1 system have been published by different authors, many of them Italian, but others from The Netherlands, New Zealand, Korea, Spain. - PowerPoint PPT Presentation

Citation preview

Page 1: 1060

1

1060

Page 2: 1060

2

QUALITY AND AUTOMATION IN THE DETERMINATION OF THE

ERYTHROCYTE SEDIMENTATION RATE

by Paolo Galiano

Page 3: 1060

3

In the last 5 years many scientific works on the TEST1 system have been published by different authors, many of them Italian, but others from The Netherlands, New Zealand, Korea, Spain.

The state of the art of these articles provides a view of the new state of the art of the

Erythrocyte Sedimentation Rate in the world.

Page 4: 1060

4

SCIENTIFIC WORKS AND PUBLICATIONSSCIENTIFIC WORKS AND PUBLICATIONS1. M. Plebani, S. De Toni, M.C. Sanzari, E. Stockreiter, D. Bernardi (Dept. of Laboratory

Medicine, University-Hospital, Padova, Italy), “The TEST 1 Automated System: A New Method for Measuring for Erythrocyte Sedimentation Rate”, American Journal of Clinical Pathology, 1998, 110:334-340.

2. N. Cirilli, Z. Abu Asy, N. Giacchè, F. Bordicchia, S. Paolucci, M. Tocchini (Dept. of Laboratory Medicine, G. Salesi Hospital, Ancona, Italy), “TEST1: Un Nuovo Metodo per la Determinazione della VES”, Biochimica Clinica, Vol. 22, N. 5-6, 1998, p. 339.

3. M. Plebani, S. De Toni, M.C. Sanzari, E. Stockreiter, D. Bernardi, F. Floriani (Dept. of Laboratory Medicine, University-Hospital, Padova, Italy), “Il Sistema Automatizzato TEST1: un Nuovo Metodo per la Determinazione della Velocità di Eritrosedimentazione”, Medicina di Laboratorio, Vol. 6, N. 2, June 1998, pp. 166-172.

4. G. Soffiati (Clinical Chemistry and Hematology Laboratory, San Bortolo Hospital, Vicenza, Italy), “Nuovo Metodo per la Determinazione della Velocità di Eritrosedimentazione (VES)”, August 1998, private communication.

5. L. Germagnoli, S. Lopez-Silva, M. Murone, S. Vazzana, L. Grassini, L. Calloni, V. Gioia (Dept. of Laboratory Medicine, Scientific Institute San Raffaele, Milan, Italy), “Evaluation of the Automatic System TEST1™ for Measurement of the Erythrocyte Sedimentation Rate (ESR)”, issued as scientific poster on Clinical Chemistry, July 1999.

Page 5: 1060

5

SCIENTIFIC WORKS AND PUBLICATIONSSCIENTIFIC WORKS AND PUBLICATIONS6. K. S. Shin, J.S. Kim, B.R. Son (Dept. of Clinical Pathology, College for Medicine Chungbuk

National University, Cheongju, Korea): “Evaluation of the TEST 1 for Measuring Erythrocyte Sedimentation Rate” Journal of Clinical Pathology and Quality Control, Vol. 21, No. 1, 1999, pp. 223-228.

7. John Robert "Erythrocyte Sedimentation. A New Solution to an Old Problem", (Hitech Pathology, Melbourne, Australia) issued on the official publication of the New Zealand Institute of Medical Laboratory Science and Australian Institute of Medical Scientists, South Pacific Congress, Christchurch, New Zealand, 23-27 August 1999, sponsored by Dade Behring Diagnostics.

8. C. Gasparoli, D. Pulè, A. Fusco (Dept. of Laboratory Medicine, Istituto Dermopatico Immacolata – IRCCS, Rome), “Sostituzione di un Metodo Tradizionale per la Misurazione delle VES con il Nuovo Metodo Automatico TEST1”, October 1999, private communication.

9. K. Taylor (Canterbury Health Laboratories, Christchurch, New Zealand), “TEST1 Evaluation Report”, December 1999, private communication.

10.N. de Jonge, I. Sewkaransing, J. Slinger, J.J.M. Rijsdijk (Dept. Clinical Chemistry, Leyenburg Hospital, The Netherlands), “Erythrocyte Sedimentation Rate by Test-1 Analyzer”, Clinical Chemistry, June 2000, 46: 881-882.

Page 6: 1060

6

SCIENTIFIC WORKS AND PUBLICATIONSSCIENTIFIC WORKS AND PUBLICATIONS11.M. Plebani, E. Piva, M.C. Sanzari, G. Servidio (Dept. of Laboratory Medicine,

University-Hospital, Padova, Italy), “Length of Sedimentation Reaction in Undiluted Blood (Erythrocyte Sedimentation Rate): Variations with Sex and Age and Reference Limits”, Clinical Chemistry and Laboratory Medicine, May 2001, 39: 451-454.

12.D. Giavarina, S. Capuzzo, M. Carta, F. Cauduro, G. Soffiati (Clin. Chem. & Hematol. Lab., San Bortolo Hospital, Vicenza, Italy), “Internal Quality Control for Erythrocyte Sedimentation Rate (ESR) measured by TEST-1 Analyzer”, Clinical Chemistry, June 2001, 47: 162.

13.D. Spedding, D. Smith (Dade Behring Diagnostics, New Zealand), “Evaluation of Agreement between the TEST1 and Starrsed Automated ESR Analysers”, November 2001, private communication.

14.M. Plebani, E. Piva (Dept. of Laboratory Medicine, University-Hospital, Padova, Italy), “Erythrocyte Sedimentation Rate. Use of Fresh Blood for Quality Control”, American Journal of Clinical Pathology, 2002, 117:621-626.

15.E. Heverin (Galway-Mayo Institute of Technology, Ireland), “Comparison of the Westergren method versus the TEST1 technique for determining the Erythrocyte Sedimentation Rate”, May 2002, private communication.

Page 7: 1060

7

SCIENTIFIC WORKS AND PUBLICATIONSSCIENTIFIC WORKS AND PUBLICATIONS16.P. Napoli, B. Montaruli, S. Plateroti, A. Martini, A. Sacchi, A. Toja. M. Saitta (Analysis

Lab, CIOV, Evangelico Valdese Hospital, Turin, Italy), “Sistema Automatizzato TEST1: Controllo di Qualità Interno per la Determinazione della VES”, Biochimica Clinica, Vol. 26, N. 3, 2002, p. 215.

17.B.H. Lee, J. Choi, M.S. Gee, K.K. Lee, H. Park (Dept. of Laboratory Medicine, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Korea), “Basic Evaluation and Reference Range Assessment of TEST1 for the Automated Erythrocyte Sedimentatioon Rate”, Journal of Clinical Pathology and Quality Control, Vol. 24, No. 1, 2002.

18.D. Giavarina, S. Capuzzo, F. Cauduro, M. Carta, G. Soffiati (Clin. Chem. & Hematol. Lab., San Bortolo Hospital, Vicenza, Italy), “Internal Quality Control for Erythrocyte Sedimentation Rate Measured Test 1 Analyzer”, Clinical Laboratory 2002, 48: 459-462.

19.Romero A., Muñoz M., Ramirez G. (Dept. of Haematology, H.C.U. "Virgen de la Victoria", Málaga & *GIEMSA, School of Medicine, University of Málaga, Spain), "Determination of the Length of Sedimentation Reaction in Blood: a Comparison of the Test1 ESR System with the ICSH Reference Method and the Sedisystem", Clinical Chemistry and Laboratory Medicine 2003, 41 (2).

Page 8: 1060

8

SCIENTIFIC WORKS AND PUBLICATIONSSCIENTIFIC WORKS AND PUBLICATIONS• 20. M. Plebani (Dept. of Laboratory Medicine, University-Hospital, Padova, Italy), “Erythrocyte

Sedimentation Rate: Innovative Techniques for an Obsolete Test?”, Clinical Chemistry and Laboratory Medicine, 2003, 41 (2): 115-116.

• 21. Nicoli M., Lanzoni E., Massocco A., Franceschini C.* (Laboratory of Clinical Chemistry and Haematological Analysis, Ospedale Civile Maggiore, Verona, Italy & *Dasit, Cornaredo, Italy), “Integrated Haematology and Coagulation Laboratory”, Poster, Euromedlab Congress, Barcelona, Spain, 1-5 June 2003.

• 22. P. Galiano, “Quality and Automation in the Determination of the Erythrocyte Sedimentation Rate, Symposium 046, 22nd World Congress of Pathology & Laboratory Medicine, 30 August-2 September 2003, Busan, Korea.

• 23. J.M. Jou, “La VSG: còmo, cuàndo y para qué puede ser ùtil”,  (Hospital Clinic University of Barcelona), AEHH/SETH Hematology Society National Congress, 23-25 October 2003, Santiago de Compostela, Spain.

• 24. B. Olivera Alonso, M. Sirvent Monerris, M.T. Rotella Belda, V. Ballenilla Antón, García Vidal (M Laboratorio Hospital San Vicente y Area Sanitaria 18. Alicante, Spain), “Cambio De Método Para La Determinación De V.S.G.: Repercusiones Sobre La Fase Preanalítica”, Generalitat Valenciana - Conselleria De Sanitat (for Valencia Government – MOH), Spain 2004.

Page 9: 1060

9

Heverin from Galway-Mayo Institute of Technology, in describing the reproducibility of the TEST1, verifies a very important point, “temperature”. To perform a reliable internal quality control it is absolutely needed to keep the samples into the refrigerator and to allow them to reach room temperature the day after before testing them again. I remind you that from NCCLS 4th edition, approved May 2001, ESR cannot be calibrated.

Page 10: 1060

10From E. Heverin (Galway-Mayo Institute of Technology, Ireland), “Comparison of the Westergren method vs the TEST1 technique for determining the Erythrocyte Sedimentation Rate”, May 2002

K3EDTA @ 24hrs 4°C, K2EDTA @ 24hrs 4°C r=0.974• Sta

bilit

y af

ter

24 h

ours

a

t 4°C

Page 11: 1060

11

Test 1 ESR results at 4hrs versus 48hrs at 4°C

From E. Heverin (Galway-Mayo Institute of Technology, Ireland), “Comparison of the Westergren method vs the TEST1 technique for determining the Erythrocyte Sedimentation Rate”, May 2002

K3EDTA @ 48hrs 4°C, K2EDTA @ 48hrs 4°C r=0.964

• Sta

bilit

y af

ter

48 h

ours

a

t 4°C

Page 12: 1060

12

From E. Heverin (Galway-Mayo Institute of Technology, Ireland), “Comparison of the Westergren method vs the TEST1 technique for determining the Erythrocyte Sedimentation Rate”, May 2002

Page 13: 1060

13

From E. Heverin (Galway-Mayo Institute of Technology, Ireland), “Comparison of the Westergren method vs the TEST1 technique for determining the Erythrocyte Sedimentation Rate”, May 2002

Page 14: 1060

14

Now let me make a brief list of the reference subjects described by the National Committee for Clinical Laboratory Standards. All the topics belong to sodium citrate collected samples.

Page 15: 1060

15

Page 16: 1060

16

Page 17: 1060

17

Page 18: 1060

18

Page 19: 1060

19

Page 20: 1060

20

Page 21: 1060

21

Page 22: 1060

22

Page 23: 1060

23

Page 24: 1060

24

From NCCLS, 4rd ed., Vol. 20, No. 27, December From NCCLS, 4rd ed., Vol. 20, No. 27, December 20002000

Acceptable values comparing EDTA with citrate collected Acceptable values comparing EDTA with citrate collected samplessamples

EDTAEDTAFixedFixed

20 4320 43

CitrateCitrateVariablesVariables

5-17 17-355-17 17-35

202021212222232324242525262627272828--

36363737383839394040414142424343

5-17 5-17 6-176-176-186-186-196-197-197-197-207-208-218-218-218-219-229-22

--13-2913-2913-3013-3014-3014-3014-3114-3115-3215-3215-3315-3316-3416-34

17-3517-35

same

sample

Page 25: 1060

25

Just from one of the points you can read from the original document, I use page 8 to see the limits of all citrate methods and instruments using this type of preservant (sodium citrate). TEST1 has a big advantage: it works with EDTA,considered to be the best blood preservant.In fact, if you see this slide, the variation of citrate is intrinsic to the list shown above offering an acceptable range, but variable against a fixed value offered by EDTA samples.

Page 26: 1060

26

Page 27: 1060

27

Page 28: 1060

28

Page 29: 1060

29

Page 30: 1060

30

t=0t=0 t=1ht=1hWESTERGRENWESTERGREN

t=12 ht=12 hor or

CentrifugationCentrifugation

NormalNormal

PathologicalPathological

ESR – HEMATOCRITESR – HEMATOCRIT

35%35%Normal Normal HematocritHematocritrangerange

ESR 1hESR 1hSAMPLESAMPLE

20% Area requested 20% Area requested for well mixingfor well mixing

THE CLASSICAL ESRTHE CLASSICAL ESR

AgglomerinAgglomerin

Page 31: 1060

31

ESR is a phyisical phenomenon circumscribed on ESR is a phyisical phenomenon circumscribed on time time

From NCCLS: From NCCLS: recognised curve is a recognised curve is a sigmoid type helped sigmoid type helped by the presence of by the presence of specific proteins specific proteins called called agglomerinsagglomerins

Without these agglomerins there is no sedimentation Without these agglomerins there is no sedimentation

Page 32: 1060

32

• Pathologies– leukaemia, rheumatic polimialgy, rheumatoid arthritis, thromboflebitis,

subacute endocarditis, pneumoniae, colangitis, osteomyelitis, SLE, pyelonephritis, post cardiosurgery, acute hearth stroke, lung stroke, linphoma, rheumatic fever, viral hepatitis, glomerular nephritis, infectious mononucleosis, breast cancer, lung cancer, hepatic metastatis, ipernephroma, uremia, acute meningithis, ictus, sarcoidosis, pelvic infections (gonococcus and others), tubercolosis, Crohn’s disease, burns, bone fracture, anemia, arthrosis, gout, cholecystitis, typhus, mumps, acute allergy, ulcerative colitis, pregnancy, postpartum, menstruation, Cytomegalovirus infection, Toxoplasmosis, Rickettsiosis, fever, appendicitis.

• Drugs– contraceptive , A Vitamin– others: B Hepatitis vaccination

3232

ESR values increaseESR values increase

Page 33: 1060

33

• Pathologies– dehydration, hepatic necrosis, Thrichinosis,

polycythemia, fibrinogenopenia and iperfibrinogenemia, cachexy, anticoagulants

• Drugs– salicylates, cortisone, ACTH, Cyclophosphamide

(chemotherapeutic agent), chinin, oxalate

3333

ESR values decreaseESR values decrease

Page 34: 1060

34

Westergren method and use of Sodium Citrate show: (from NCCLS Vol.20 No.27 H2-A4)

- errors in the dilution step - poor temperature control- problems connected to vibrations,verticality,diameter of pipets - test must be run within 4 hours from blood collection- PCV adjustment (<0.35)- No control available

- plastic tubes may interfere with red cells surface charges due to the plastic material

Page 35: 1060

35

VACUTAINER TUBE IN CITRATEVACUTAINER TUBE IN CITRATE

Exceeding level of bloodExceeding level of blood(fresh tubes)(fresh tubes)

NO RATIO RESPECTEDNO RATIO RESPECTED

Correct level of bloodCorrect level of bloodDilution ratio 1:4Dilution ratio 1:4

RATIO RESPECTEDRATIO RESPECTED

insufficient level of insufficient level of bloodblood

(old tubes)(old tubes)NO RATIO RESPECTEDNO RATIO RESPECTED

NONO OKOK NONO

Errors in the dilution phaseErrors in the dilution phase

Correct LevelCorrect Level

Page 36: 1060

36

THE IMPORTANCE OF HEMATOCRITTHE IMPORTANCE OF HEMATOCRITDIFFERENT RESULTS OF THE SAME SAMPLE DILUTEDDIFFERENT RESULTS OF THE SAME SAMPLE DILUTED

WITH AUTOLOGOUS PLASMAWITH AUTOLOGOUS PLASMA

Sedimentation as a function of time with different Hct values artificially obtained adding autologous plasma(Blood, Vol. 70, No 5, Nov. 1987: pp 1572-1576).

Hct of 26Hct of 26 Hct of 28Hct of 28 Hct of 34Hct of 34 Hct of 45Hct of 45

SED

IMEN

TATI

ON

SE

DIM

ENTA

TIO

N

(mm

)(m

m)

TIME (minutes)TIME (minutes)

002020404060608080

100100120120140140

00 2020 4040 6060 8080 100100 120120 140140 160160 180180 200200

Page 37: 1060

37

LOW INFLUENCE OF HEMATOCRIT ON TEST 1From From Thomas L.Fabry, Thomas L.Fabry, Mechanism of Erythrocyte Aggregation and Mechanism of Erythrocyte Aggregation and

Sedimentation, BloodSedimentation, Blood, Vol. 70, No. 5 (Nov.), 1987: pp 1572-1576., Vol. 70, No. 5 (Nov.), 1987: pp 1572-1576.

Fabry’s Formula to adjust the value Fabry’s Formula to adjust the value of ESRof ESR

Example of adjustment for a sample with hematocrit Example of adjustment for a sample with hematocrit 30.330.3 - Correct value = 69.2

- WG non corrected value = 114- TEST 1 = 6767

Example of adjustment for a sample with Example of adjustment for a sample with hematocrit 26.3hematocrit 26.3 - Correct value = 66.3

- WG non corrected value = 127- TEST 1 = 6767

Page 38: 1060

38

LOW HEMATOCRITLOW HEMATOCRIT

The instrument indicates The instrument indicates with an asterics (*), near with an asterics (*), near the result, the samples the result, the samples with a low hematocrit with a low hematocrit value.value.

In the example reported In the example reported the hematocrit value is the hematocrit value is lowlow

Page 39: 1060

39

Moreover, the commercial quality controls offered and pushed by our competitors, as you can see again, show the great variations of these values, improperly offered under this name: ranges and limits suggested in the figure show how low level is indicated from 1 to14 and high level from 15 to 55.

Page 40: 1060

40

Page 41: 1060

41

Page 42: 1060

42

TEST1 offers much closer reproducibility as many authors have described using internal samples of their labs, tested 24 and 48 hours later. The correlations you see on the data are comparable to normal clinical chemistry controls and, finally, our statistical quality control gives you a method based on statistical data of your population. The cumulative data of any day (white circles) can be compared with the black one, that are the statistical data collected by the instrument during a month and these collected data can be considered as a standard when you have in the black circle values at least from 500-1000 results. The cumulative black circle maintains a revolving number of 6000 samples as a standard.

Page 43: 1060

43

29/01/2003 29/01/2003 29/01/2003 29/01/200312.07.am 12.18am 12.28am 15.38 pm

ESR 2 3 4 238 41 44 4276 79 79 7414 13 13 152 3 3 4

15 16 16 1442 44 40 417 8 7 9

26 25 22 2316 16 16 1711 11 10 1152 50 47 4519 17 20 1859 57 55 5822 22 23 24

TEST 1 REPRODUCIBILITY DURING THE SAME DAYSamples collected on January 28th and processed on January 29th

Samples were properly stored in the refrigerator (4°C) and let them get room temperature taking them out of the refrigerator 30-40 minutes before running the tests performed with Test 1 Instrument. Samples are properly mixed in Test 1, thus aggregated cells are re-suspended in the plasma and ESR is consequently correctly measured.

The 4 columns represent the results obtained in 4 different moments (12,07 am; 12,18 am; 12,28 am; 15,38 pm). Results were obtained alive during the training! The red column is the reference towards which the correlations of columns C, D and E have been calculated.

R = 0,99 in the correlation between column B- column C = R2 0,9944R = 0,99 in the correlation between column B- column D = R2 0,9831R = 0,99 in the correlation between column B- column E = R2 0,988

Page 44: 1060

44

TEST 1 REPRODUCIBILITY DURING THE SAME DAYfig 1

Samples collected on January 28thand processed on January 29th

29/01/2003 29/01/2003 29/01/2003 29/01/200312.07.am 12.18am 12.28am 15.38 pm

ESR 2 3 4 238 41 44 4276 79 79 7414 13 13 152 3 3 4

15 16 16 1442 44 40 417 8 7 9

26 25 22 2316 16 16 1711 11 10 1152 50 47 4519 17 20 1859 57 55 5822 22 23 24

y = 1,0066x + 0,0901R2 = 0,9944

y = 0,9798x + 0,4054R2 = 0,9831

y = 0,9446x + 1,2142R2 = 0,988

0

10

20

30

40

50

60

70

80

90

0 20 40 60 80

R = 0.99

CORRELATION CHARTCORRELATION CHART

Page 45: 1060

45

29/01/2003 30/01/2003 30/01/2003 12.07am 9.21am 9.32am

ESR 2 2 238 36 4176 77 7514 11 122 2 2

15 12 1242 35 397 8 8

26 22 2016 14 1511 9 1052 43 4419 20 2059 56 5522 22 22

REPRODUCIBILITY AFTER 48 HOURSSamples collected on 28/1/2003 and processed on 30/1/2003

SAMPLES HAVE BEEN STORED AT +4°C DURING THE TIME LAPSE BETWEEN 29 AND 30 JANUARY

The following day we took out from the refrigerator the 15 samples tested the day before and let them get room temperature for about 30-40 minutes.

Then we put samples into Test 1 and the working session started. After 3 minutes of mixing cycle the measurement of ESR begun.

The results of two running sessions are reported on columns C and D. Column B has been copied and pasted from the previous Excel sheet in order to have an immediate comparison among reproducibility of Test 1 and good quality of results.

R = 0.99 correlation between columns B- columns C R2 0,9832R = 0.99 correlation between columns B- columns D R2 0,9844

Page 46: 1060

46

29/01/2003 30/01/2003 30/01/2003 12.07am 9.21am 9.32am

ESR 2 2 238 36 4176 77 7514 11 122 2 2

15 12 1242 35 397 8 8

26 22 2016 14 1511 9 1052 43 4419 20 2059 56 5522 22 22

y = 0,9557x - 0,9483R2 = 0,9832

y = 0,9519x - 0,3137R2 = 0,9844

0

10

20

30

40

50

60

70

80

90

0 10 20 30 40 50 60 70 80 R = 0.99

CORRELATION CHARTCORRELATION CHART

Page 47: 1060

47

I specified that the value of our control is to be interpreted upon the reproducibility of the correlation value R=0,99 and this is a real reproducibility as a clinical chemistry control, not based on the possible variation of a sample that may also give not so reproducible results the day after. This is not so frequent but if it occurred you would be able to give the correct answer: it is the reproducibility R that creates the stability of control.Moreover, I graphically represented the normal, pathological and myeloma conditions to show the time and curve typology of the myeloma-ESR case. I have intentionally begun a discussion in which, as a first topic, I have reminded the audience that the representation of a myeloma-ESR curve is not even reported by the NCCLS, as this organisation describes only a Sigmoid-like curve to represent a physical condition, in this case the ESR testing, and not a curve that remains plain for many minutes and then precipitates.

Page 48: 1060

48

Storage of blood sample at 4°C for up 24 hours caused a decrease in ESR values obtained with the TEST1: the mean difference was 2.86 ( 95% CI, 2.41-3.31) and 2.28 (95% CI, 1.90-2.65) respectively for two different analyzers with the same samples (n=1.140).

The decreases were of 9% and 11%.

Stability StudyStability Study

Page 49: 1060

49

STATISTICAL QUALITY CONTROL DATA REPORTSTATISTICAL QUALITY CONTROL DATA REPORT

Print out dateScale of values

Value of cumulative average (black circle)Value of daily average O (white circle)

STD S. cumulative standard deviationSTD D. daily standard deviation

N.B. The values printed are only examples not to be considered as absolute but only indicative.

(black circle ) = 6000 data of cumulative values are equivalent to your statistical population, which can be considerd a Standard

For Range 2-120 mm/h and 2-30 mm/h

Page 50: 1060

50

STATISTICAL QUALITY CONTROL DATA REPORTSTATISTICAL QUALITY CONTROL DATA REPORTFor control of population to obtain internal normal range

Page 51: 1060

51

Page 52: 1060

52

STABILITY OF VACUTAINER TUBE IN STABILITY OF VACUTAINER TUBE IN EDTAEDTA

THETHE EDTAEDTA COLLECTED COLLECTED SAMPLE REFRIGERATED AT SAMPLE REFRIGERATED AT +4°C CAN BE TESTED EVEN +4°C CAN BE TESTED EVEN

AFTERAFTER 24 HOURS24 HOURS FROM FROM BLOOD COLLECTIONBLOOD COLLECTION

From Clinical Chemistry, Vol. 47, No. 6, Supplement 2001, p. 162 - Internal quality control for erythrocyte sedimentation rate (ESR) measured by TEST 1 analyzer by D. Giavarina, S. Capuzzo, M. Carta, F. Caoduro, G. Soffiati, Clin. Chem. & Hematol Lab - San Bortolo Hospital: Vicenza, Italy

Page 53: 1060

53

Technical advantages

SAVE of time money

SAVE of waste money (expired tube – tube volume)

SAVE of stock money

Page 54: 1060

54

Quality improvements

The unique method for Lab Accreditation

Page 55: 1060

55

The unique characteristics of ALIFAX ESR analyzers provides a Statistical Internal Quality The unique characteristics of ALIFAX ESR analyzers provides a Statistical Internal Quality Control useful to the certification of the lab.Control useful to the certification of the lab.The new ISO9001/UNI EN ISO 9001 – Ed. 2000 certification is used for the validation of the The new ISO9001/UNI EN ISO 9001 – Ed. 2000 certification is used for the validation of the lab test in which the lab instrumentation, controls and, in general, the analysis system can lab test in which the lab instrumentation, controls and, in general, the analysis system can have the recognition of a value and its maximum validation to guarantee operation have the recognition of a value and its maximum validation to guarantee operation functionality and reliability of the results. functionality and reliability of the results.

TEST1 TH new software offers:TEST1 TH new software offers:• Control of the functionality of reading sensors for each washing Control of the functionality of reading sensors for each washing • Data control by a statistical quality system based on the control validity of Data control by a statistical quality system based on the control validity of the collected data of a populationthe collected data of a population• Control of the daily value relative to a population of 6000 memorized data Control of the daily value relative to a population of 6000 memorized data in the instrumentin the instrument• Control of the single result scanned 1000 times by our system for each Control of the single result scanned 1000 times by our system for each analysisanalysis• Control of sample reproducibility on samples tested the day after Control of sample reproducibility on samples tested the day after • Control of data reproducibility at R value equivalent to 0.98-0.99Control of data reproducibility at R value equivalent to 0.98-0.99• Control for samples improper to the test below the normal hematocrit values Control for samples improper to the test below the normal hematocrit values < 20< 20• Control of data transmission to LIS.Control of data transmission to LIS.

New software benefits for LAB ACCREDITATION

5555

Page 56: 1060

56

• Autodiagnostics• Internal or external bar code reader• Interfacing through RS232• QUERY HOST Software for PC LAB management• No dedicated tubes• Compatible with cell counter racks (Bayer, Beckman Coulter, Sysmex, Abbott,

ABX, etc.)• Capacity of 60 samples random batch access• 180 samples/hr• Automatic sample mixing• Blood collection according to the ICSH recommendations• Flow control to check of eventual clots• Minimum Blood Volume > 1 ml• Working volume 150 µL of blood

TECHNICAL CHARACTERISTICS

Page 57: 1060

57

• Cell measure volume 1 µL• Thermostation 37°C • First result after 3 minutes and 20 seconds• Low influence of hematocrit < 35 HCT• No carryover• Autowashing• Safe-control closed cycle system• Automatic reading and print out of the results• Totally safe waste volume tank• Waste production reduced (Test 1=3 L, others=1.000L)• Waste disposal costs reduced• Statistical internal quality control of population for normal values• Safety check card• Electronic calibration for alignment of results

TECHNICAL CHARACTERISTICS

Page 58: 1060

58

What is TEST 1

• The only instrument in the world which can MEASURE the ESR values

• Remember that all other instruments on the market are simply READERS

Page 59: 1060

59

• Test 1 measures the kinetics of blood sedimentation phenomenon

• all the other instruments simply read the final result of this phenomenon, that is why they need more time to give results

Page 60: 1060

60

TEST1TEST1

ESRESRTELEMETRYTELEMETRY

Page 61: 1060

61

CAPACITY OF SEDIMENTATION AND CAPACITY OF SEDIMENTATION AND AGGREGATION AGGREGATION

Aggregation startingt = 0

Aggregation after 1000 scanst = 20 sec.

TEST 1 studies the sedimentation and aggregation capacity TEST 1 studies the sedimentation and aggregation capacity of the blood red cells via optical densityof the blood red cells via optical density

Every sample is read 1000 times in 20 seconds verifying the Every sample is read 1000 times in 20 seconds verifying the aggregation and sedimentation capacity of the blood red aggregation and sedimentation capacity of the blood red

cellscells

Light beambefore

Light beamafter

Page 62: 1060

62

Normal Status

Lag phase

Time (minutes)

Sedi

men

tatio

n (m

m)

0

0 60+

+ +++

+

+

+ +++

+

+

+ +++

+

+

+ +++

++

+ +++

+

+

+ +++

+

+

+ +++

+

+

+ +++

+

+

+ +++

+

+

+ +++

++

+ +++

+

+

+ +++

+

Page 63: 1060

63

Inflammation Status

Lag phase

Precipitation phase

Packaging phase

Time (minutes)

Sedi

men

tatio

n (m

m)

0

70

0 60-

+ ++-

-

Agglomerine

-

+ ++-

-

-

+ ++-

-

-

+ ++-

--

+ ++-

- -

+ ++-

- -

+ ++-

-

-

+ ++-

-

-

+ ++-

--

+ ++-

--

+ ++-

-

-

+ ++-

- -

+ ++-

--

+ ++-

-

-

+ ++-

-

-

+ ++-

-

-

+ ++-

-

-

+ ++-

-

20 seconds

Page 64: 1060

64

MyelomaExample of collapsed sample for ESR

Lag phasePrecipitation phase

Packaging phase

Time (minutes)

Sedi

men

tatio

n (m

m)

0

70

0 60

-

+ ++-

--

+ ++-

-

-

+ ++-

-

-

+ ++-

--

+ ++-

- -

+ ++-

--

+ ++-

--

+ ++-

-

-

+ ++-

-

-

+ ++-

-

-

+ ++-

-

-

+ ++-

--

+ ++-

--

+ ++-

-

-

+ ++-

-

-

+ ++-

--

+ ++-

- -

+ ++-

-

Page 65: 1060

65

Example of a collapsed sample

From 200 mm to the ratio in percentage 100

Westergrenpipet

200 mm

130

70

ESR

Corpuscular

Part 35 Hematocrit or PCV

100 mm

65 ESR

Page 66: 1060

66

MyelomaExample of collapsed sample for ESR

Lag phase Precipitationphase

Packagingphase

Time (minutes)

Sedi

men

tatio

n (m

m) 0

70

0 60

-+ +

+-- -+ +

+--

-+ +

+--

-+ +

+---

+ ++-- -

+ ++--

-+ +

+-- -+ +

+---

+ ++---

+ ++-- -

+ ++--

-+ +

+--

Westergren pipet

200 mm

130

70

ESR

Corpuscular

Part 35 Hematocrit or PCV

100 mm

65 ESR

-+ +

+-- -+ +

+--

-+ +

+--

-+ +

+---

+ ++-- -

+ ++-- -

+ ++-- -

+ ++--

-+ +

+---+ +

+-- -+ +

+---

+ ++--

Page 67: 1060

67

These blood samples without “ Agglomerins” do not express a ESR value with a “sigmoid” tract, but collapse to the level of the Hematocrit. Final point of reading.

Page 68: 1060

68

The mathematical algorithm converts the results The mathematical algorithm converts the results from optical density tofrom optical density to

ERYTHROCYTE SEDIMENTATION RATEERYTHROCYTE SEDIMENTATION RATEof the microvolume analyzedof the microvolume analyzed

fromfromsample optical densitysample optical density

totomm/hr Westergrenmm/hr Westergren

CALCULATION METHODOLOGY OF CALCULATION METHODOLOGY OF RESULTSRESULTS

Page 69: 1060

69

Test-1 ESRexpresses the vitality of aggregation also 12, 24, 48 hr after collection, according to the sigmoid curve of the classic graph, that is the only curve described and represented.

Page 70: 1060

70

PATENT

The world patent of Test-1 concerns the mathematic algorithm that expresses the same sigmoid function of the ESR

Page 71: 1060

71

A synthetic control with no active electrical charge and with no agglomerins that induce and activate the sedimentation process is not recognized by Test-1 as these controls give no signals of vitality because they are not part of the “sigmoid sedimentation”, but a mixture of water – sand and not a measure of control.

Page 72: 1060

72

• The Test 1 technology is PATENTED

Page 73: 1060

73

Page 74: 1060

74

Page 75: 1060

75

30 µL of blood30 µL of bloodsuitable for pediatric usesuitable for pediatric use

Page 76: 1060

76

• Micro Test 1 can work with fresh blood, as soon as it has been collected, without need of anticoagulant or preservatives

Page 77: 1060

77

The only instrument that works without preservant

Page 78: 1060

78

Methodology advantages and characteristics

Page 79: 1060

79

• Quality Control • Urgent diagnosis can be fulfilled• Pediatric use • No anticoagulant use- fresh blood for MicroTest1• Test 1 Measures ESR, it is not a Reader Test 1 is an Analyzer !!!!

METHODOLOGY ADVANTAGES

Page 80: 1060

80

METHODOLOGY ADVANTAGES

• Capillary• Microvolume 150 or 30 microliters (T1 or MicroT1)

• Photometric-kinetic• Scanning rate -total scans 1000 for sample in 20 sec.• Not influenced from hematocrit• Rapidity of response correlated to Westergren• Reproducibility• Stability during time (EDTA-48 hours against Na

Citrate 4 hours)

Page 81: 1060

81

==> Time sec.==> Time sec.

During 20 sec. the sample is scanned 1000 timesDuring 20 sec. the sample is scanned 1000 times

Ident. 05483311Ident. 05483311ESR = 118 mm/hrESR = 118 mm/hr

Ident. 05725053Ident. 05725053ESR = 26 mm/hrESR = 26 mm/hr

Ident. 05725044Ident. 05725044ESR = 2 mm/hrESR = 2 mm/hr

==> Time sec.==> Time sec. ==> Time sec.==> Time sec.

Page 82: 1060

82

11 It starts from 0 time to 20 sec. following and measuring the It starts from 0 time to 20 sec. following and measuring the evolutionevolution of the sed rate curves.of the sed rate curves.

22 It measures the optical density related to the concentration It measures the optical density related to the concentration of the of the erythrocytes/aggregates present at the moment of the erythrocytes/aggregates present at the moment of the analysis.analysis.

33 Kinetics following the evolution of the curves with a Kinetics following the evolution of the curves with a frequency of 50frequency of 50 measures per second.measures per second.

44 The capillary system simulates a invivo situation and The capillary system simulates a invivo situation and guarantees guarantees minimal optical paths in blood subject to sedimentation which minimal optical paths in blood subject to sedimentation which enableenable the detection of small variations.the detection of small variations.

METHODOLOGY CHARACTERISTICSMETHODOLOGY CHARACTERISTICS

Page 83: 1060

83

Romero A., Muñoz M., Ramirez G., Dept. of Haematology, H.C.U. "Virgen de la Victoria", Málaga & GIEMSA, School of Medicine, University of Málaga, Spain, ““Determination of the Length of Sedimentation Reaction in Blood: a Comparison of the Test1 ESR System with the ICSH Reference Method and the Sedisystem””..

CORRELATION TEST 1-WESTERGRENCORRELATION TEST 1-WESTERGRENFrom Clinical Chemistry and Laboratory From Clinical Chemistry and Laboratory

Medicine, February 2003, 41(2)Medicine, February 2003, 41(2)

“… “… the correlation coefficient was the correlation coefficient was 0.98…”0.98…”

Page 84: 1060

84

COMPARATIVE SCHEDULE WEIGHT OF WASTE DISPOSAL

COMPANY TYPE OF PRODUCED

WASTE

WASTE PRODUCED FOR EACH TESTING

WASTE PRODUCED FOR 20.000 TESTING

IN KG.

DIESSE DEDICATED TUBE + BLOOD

gr. 12 Kg. 240

BD TUBE WITH BLOOD gr. 10 Kg. 200

ALIFAX ON A COLLECTION TANK

CAPACITY 2.000 TESTING

gr. 0,25 Kg. 5

Page 85: 1060

85

Slides from Customers

Page 86: 1060

86

ESR value RangeCV%TEST 1 13.1 12-14 4.86Westergren 8.0 8.0-109.68

TEST 1 29 27-32 5.30Westergren 25 21-308.45

TEST 1 54.7 51-58 3.37Westergren 57.1 51-605.04

Imprecision of TEST 1 and WestergrenImprecision of TEST 1 and Westergren

Page 87: 1060

87

• It preserves the red blood morphology.• Does not interfere with mechanisms that lead to erythrocyte sedimentation.• Increases specimen stability.• Does not incur problems related to

sample dilution with sodium citrate.

ADVANTAGES OF EDTA AS AN ANTICOAGULANTADVANTAGES OF EDTA AS AN ANTICOAGULANT

Page 88: 1060

88

Specimens anticoagulated with EDTA:• Are suitable for internal quality control programs.• Allow the creation of a unique workstation for measuring ESR and performing other hematological tests (erythrocyte, leukocyte, reticulocyte counts and differential analysis in a single specimen).

ADVANTAGES OF EDTA AS AN ANTICOAGULANTADVANTAGES OF EDTA AS AN ANTICOAGULANT

Page 89: 1060

89

Reference limits, 2.5th and 97.5th pervcentiles and their 95% confidence intervals (CI) for SRB

in undiluted EDTA-anticoagulant blood. Variation with age and sex

Age n Sex 2.5 percentiles 97.5 percentiles (year) and 95%CI and 95%CI

0-14 80 W and M 2 (2-2) 34 (26-41)15-50 190 Women 2 (2-2) 37 (36-39)15-50 150 Men 2 (2-2) 28 (20-30)51-70 120 Women 2 (2-3) 39 (38-45)51-70 130 Men 2 (2-2) 37 (31-44)>70 170 W and M 3 (3-3) 46 (45-55)

Clin Chem Lab Med 2001;39(5):451-4.

Page 90: 1060

90

150 µL of 150 µL of bloodblood

TEST 1 THTEST 1 TH

Page 91: 1060

91

ROLLER

Page 92: 1060

92

TEST 1TEST 1THTH works by loading works by loading cell counter racks directly, eg.: cell counter racks directly, eg.: Bayer, Sysmex, Beckman Coulter, Bayer, Sysmex, Beckman Coulter, Abbott, ABX, etc…Abbott, ABX, etc…

Any kind of vacutainer tube Any kind of vacutainer tube available on the market can available on the market can be used to perform the ESR, be used to perform the ESR, e.g.:e.g.:

TEST 1TH RACK

Bect

on

Bect

on

Dick

inso

nDi

ckin

son

Sars

tedt

Sars

tedt

Grei

ner

Grei

ner

Teru

mo

Teru

mo

Sars

tedt

Sars

tedt

Sars

tedt

Sars

tedt L.P.L.P.

Page 93: 1060

93

TUBES LOCK

Tubes lockTubes lock Tubes unlockTubes unlock

Page 94: 1060

94

SAMPLE ID WITH EXTERNAL BAR-CODE READER

Tube identificationTube identificationWith the bar code reader the ID With the bar code reader the ID patient is automatically assigned patient is automatically assigned

to the ESR resultto the ESR result

Tube introduction into the rackTube introduction into the rack

Page 95: 1060

95

Internal Bar Code ReaderInternal Bar Code Reader

AUTOMATIC ID PATIENT WITH QUERY HOST SOFTWARE

Sample Identification with Sample Identification with automatic selection of the tubes automatic selection of the tubes requiring the ESR test both with requiring the ESR test both with dedicated racks.dedicated racks.

Racks used:Racks used:

Page 96: 1060

96

WORKFLOWRack blockingRack blocking

Tubes are blocked Tubes are blocked with an elliptic movementwith an elliptic movement

Rack loadingRack loadingOpen the front door and insert the Open the front door and insert the

rack into the instrument.rack into the instrument.Close the front door and press the Close the front door and press the

START key.START key.

Page 97: 1060

97

RACK ADAPTOR FOR BECKMAN COULTER

Page 98: 1060

98

RACK ADAPTOR FOR BAYER

Page 99: 1060

99

RACK ADAPTOR FOR SYSMEX - ABBOTT - ABX

Page 100: 1060

100

The most recent work made by an Hospital in Verona, Italy, is an important example of TLA (Total Lab Automation), published at Euromedlab, Barcelona 2003, in which the author declares what you can see in the poster (500 tubes of ESR per day not using, as in the past, a dedicated sodium citrate tube).

Page 101: 1060

101

The picture of the lab shows an integration of Sysmex cell counter and 2 TEST1 connected to the LIS. For the purpose of this automation the workflow is now available to perform the ESR test using directly the cell counter racks of Abbott, ABX, Beckman Coulter, Bayer and Sysmex. Thus, TEST1 is loaded with special rack adaptors to contain the original hematology racks with an internal bar code reader selecting the ESR test connected to LIS. I remind you no other author, as far as I can know, has published anything new with the old method and with the old reader systems.

Page 102: 1060

102

INSTRUMENTS IN THE WORLDINSTRUMENTS IN THE WORLDupdated 26.04.04

102102

Page 103: 1060

103

INSTRUMENTS IN THE WORLDINSTRUMENTS IN THE WORLD

103103

TTOOTTAALL

734734

TEST1 MicroTEST1 Roller 20 Country308 75 8 Italy48 42 14 South Africa75 10 3 Spain55 6 Korea21 41 Brasil31 17 2 Turkey17 32 China41 3 3 Belgium40 1 Portugal21 5 Australia20 Germany13 13 1 Colombia12 1 2 Ireland/UK9 Israel5 3 New Zealand6 4 2 Slovenia1 5 Uruguay

9 Costa Rica4 Malaysia

1 3 Guatemala3 Nederlands1 1 India1 1 Hungary

2 Syria1 Norway2 2 U.S.A.

1 Nicaragua1 El Salvador1 Panama1 Greece1 Indonesia1 Singapore1 Vietnam

1 1 Chile1 3 Japan

TEST 1TEST 1

TTOOTTAALL

285285

Micro Micro TEST 1TEST 1

Page 104: 1060

104104104

Sales of TEST 1 tests

819.0002.218.000

3.058.500

4.580.500

7.104.000

10.161.000

0

2.000.000

4.000.000

6.000.000

8.000.000

10.000.000

12.000.000

1998 1999 2000 2001 2002 2003