View
293
Download
6
Category
Preview:
Citation preview
CastleLoc –P Cervical Plate SystemLow Profile Cervical Plate System
Designed for Patient Comfort
Surgical Technique
2LKSTD-A011(Rev.1508)
Introduction CastleLoc –P Cervical Plate System 4
Implant
Instrument List 5
Surgical Technique
Step 1 - Patient Position and Surgical Approach 9
Step 2 - Measuring Size 10
Step 3 - Plate Contouring 11
Step 4 - Plate Positioning 12
Step 5 - Plate Fixation 13
Step 6 - Pathway Guide for Screw Insertion (Optional) 14
Step 7.1 - Select Drill Guide 15
Step 7.2 - Drill Screw Holes 16
Step 8 - Insert Plate Screws 17
Step 9 - Locking 18
Implant Removal 19
Order Information 20
Instruction For Use 24
Table of Contents
3LKSTD-A011(Rev.1508)
The CastleLoc-P Anterior Cervical Plate System implants consist of plates and plate screws. All
components are manufactured from titanium alloy (Ti-6Al-4V). The plates are available in lengths
ranges from 17-97mm to accommodate one to four segments of fixation. The system also
includes 3.5mm and 4.0mm titanium alloy screws which are available in lengths ranging from
12mm-18mm.
INDICATIONS
The CastleLoc-P Anterior Cervical Plate System is intended for anterior screw fixation to the
cervical spine (C2-C7). The System is indicated for use in the immobilization and stabilization of
the spine as an adjunct to fusions in patients with:
Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the
disc confirmed by patient history and radiographic studies),
Spondylolisthesis,
Trauma (i.e. fractures or dislocations),
Tumors,
Deformity (defined as kyphosis, lordosis, or scoliosis),
Pseudoarthrosis,
Failed previous fusion,
Spinal stenosis.
CastleLoc –P Cervical Plate System
Introduction
4LKSTD-A011(Rev.1508)
CastleLoc –P Cervical Plate System
Plate
• Low profile with proven strength
• Lowest profile: 1.7mm
1-4 levels
One-step locking
• Rotational anti-pull out clip secures screws to the plate, eliminating the need for anadditional locking or blocking mechanism
• The one-step locking allows quicker construction.
Implant
5
3.5mm 4.0mm
<Fixed angle>
3.5mm 4.0mm
<Variable angle>
Screw
• Color coded
• Fixed / Variable type
• Screw lengths from 12 mm – 18 mm
in 2mm increments
• 3.5 mm / 4.0 mm diameters
• Self-drilling / Self-tapping
LKSTD-A011(Rev.1508)
<Self -dri l l ing> <Self -tapping>
Fixed ScrewCephalad / Caudal angulation Medial / Lateral angulation
Variable ScrewCephalad / Caudal angulation Medial / Lateral angulation
Awl
CP02-0101
Plate Bender
CP02-0701
Caliper
CP02-0702
Drill Guide Fixed
CP02-0201
Drill Guide Variable
CP02-0202
Pin Holder
CP02-0704
Instrument
6
Plate Holder
CP02-0703
Prefixation Pin
CP02-0705
Punch Awl
CP02-0102
(Optional)
LKSTD-A011(Rev.1508)
Quick Coupling Handle
CP02-0501
Lock Remover
CP02-0801
Screw Driver
CP02-0601
Instrument
7
Drill Bit 12mm
CP02-0312
Drill Bit 14mm
CP02-0314
Lock Driver
CP02-0602
Container (Cover)
CP02-0001
Container (Base)
CP02-0002
LKSTD-A011(Rev.1508)
Upper Tray
CP02-0003
Lower Tray
CP02-0004
Trays with Instrument
8LKSTD-A011(Rev.1508)
Step 1 – Patient Position and Surgical
Approach
Place patient in supine posit ion. Either a r ight or lef t
approach can be used considering patient condit ion and
surgeons preference. Place a thick pad under the
patient’s neck for cervical spine reduction. Prepare
surgical site. Implantat ion of cervical plate follows
dissection, removal of disc and osteophytes for bone
graf t insert ion.
Figure 1
Surgical Technique
LKSTD-A011(Rev.1508) 9
Step 2 – Measuring Size
Plate size is measured with a caliper af ter bone graf t
and decompression is completed.
Instruments
CP02-0702 Caliper
Figure 2
Surgical Technique
10LKSTD-A011(Rev.1508)
Step 3 – Plate Contouring
Contour plate with a plate bender according to the
lordotic curvature. The CaslteLoc Anterior Cervical
Plate is provided with a pre-machined lordotic curve.
I f required, the plate may be contoured to increase the
amount of lordotic curvature (Figure 3) or decrease
the amount of lordotic curvature (Figure 4) by using
the Plate Bender. A gradual bend should be made
over the entire length of the plate and abrupt changes
in curvature should be avoided.
Instruments
CP02-0701 Plate bender
<Increase lordotic curvature> <Decrease lordotic curvature>
Note : Repeated bending may weaken the plate.
Figure 4Figure 3
Surgical Technique
LKSTD-A011(Rev.1508) 11
Step 4 – Plate Positioning
Place the plate on the medial l ine on vertebral body
using a Plate Holder.
Instruments
CP02-0703 Plate Holder
Figure 5
Surgical Technique
LKSTD-A011(Rev.1508) 12
Step 5 – Plate Fixation
Prefixation Pin can be placed in the plate to provide temporary
fixation while drilling and placing bone screws.
Pin Holder allows easy insertion into the bone. (Figure 6) Place
one or two temporary Prefixation Pins in the screw holes. Once
seated, a Prefixation Pin may be disengaged from the Pin Holder
by applying upward pressure on the locking sleeve.
Instruments
CP02-0705 Pref ixation Pin
CP02-0704 Pin Holder
Figure 6
Surgical Technique
LKSTD-A011(Rev.1508) 13
Step 6 – Pathway Guide for Screw
Insertion (Optional)
Pierce the cort ical bone with Awl through the holes of
cervical plate to guide the pathway of the screw.
Instruments
CP02-0101 Awl
CP02-0102 Punch Awl (available upon request)
Figure 7
Note : Punch Awl is available upon request
Surgical Technique
Figure 8
14LKSTD-A011(Rev.1508)
Step 7.1 – Select Drill Guide
Choose either f ixed angle or variable angle Dril l Guide.
The color band on the Dril l Guide corresponds to the
color of the screws associated with each guide.
Instruments
CP02-0202 Dril l Guide Variable
CP02-0201 Dril l Guide Fixed
Figure 9
Surgical Technique
<Variable angle>
3.5mm 4.0mm3.5mm 4.0mm
<Fixed angle>
LKSTD-A011(Rev.1508) 15
Step 7.2 – Drill Screw Holes
Attach a Dril l Bit to a Quick Coupling Handle. Dril l
through cort ical bone by insert the dri l l bit into the barrel
of the Dril l Guide.
Instruments
CP02-0202 Dril l Guide Variable
CP02-0201 Dril l Guide Fixed
CP02-0501 Quick Coupling Handle
CP02-0312 Dril l Bit 12mm
CP02-0314 Dril l Bit 14mm(Optional)
Note : The stopper on the Dril l Bits is the indicator of
appropriate depth.
Figure 10
Surgical Technique
LKSTD-A011(Rev.1508) 16
Step 8 – Insert Plate Screws
Plate screws are available in four conf igurat ions. Select
appropriate length screws and insert the screws using
the Screw Driver.
Repeat for all screw holes af ter removing the Pref ixation
Pins.
Instruments
CP02-0601 Screw Driver
Note : 3.5 mm
should be used at
f irst so that 4.0mm
screw can be used
as an emergent
measure.
Final plate screw
t ightening should be
done slowly. Care
should be taken not
to over t ighten plate
screws.
Fluoroscopic
imaging should be
considered to
confirm plate screw
depth and
orientat ion to ensure
important structures
are not at r isk.
Refer to the ordering
information (page 20)
of implant for
selection of plate
screw type Figure 11
Surgical Technique
LKSTD-A011(Rev.1508) 17
Step 9 – Locking
All screws should be secured to the vertebral bodies
before beginning the locking procedure.
Rotate the lock with Lock Driver unti l the lock stops at
the convex surface.
Repeat for all screw holes af ter insert all of the screws.
Instruments
CP02-0602 Lock Driver
Surgical Technique
<Unlock> <Lock>
LKSTD-A011(Rev.1508) 18
Figure 12 Figure 13
Implant Removal(Optional)
Use the Lock Driver to unlock.
In case of the Lock Driver is not working properly, l i f t the
lock on the plate using the Lock Remover.
Undo the plate screws using a Screw Driver
Instruments
CP02-0602 Lock Driver
CP02-0801 Lock Remover
CP02-0601 Plate Screw Driver
Figure 14
Surgical Technique
LKSTD-A011(Rev.1508) 19
1.1) Self-drilling Fixed Type
No. Cat. No.Size
ColorD L
1 6607-3512 3.5 12.0 Purple
2 6607-3514 3.5 14.0 Purple
3 6607-3516 3.5 16.0 Purple
4 6607-3518 3.5 18.0 Purple
5 6607-4012 4.0 12.0 Green
6 6607-4014 4.0 14.0 Green
7 6607-4016 4.0 16.0 Green
8 6607-4018 4.0 18.0 Green
Ordering Information
LKSTD-A011(Rev.1508) 20
1.2) Self-drilling Variable Type
No. Cat. No.Size
ColorD L
9 6608-3512 3.5 12.0 None
10 6608-3514 3.5 14.0 None
11 6608-3516 3.5 16.0 None
12 6608-3518 3.5 18.0 None
13 6608-4012 4.0 12.0 Gold
14 6608-4014 4.0 14.0 Gold
15 6608-4016 4.0 16.0 Gold
16 6608-4018 4.0 18.0 Gold
Screw
21
1.3) Self-tapping Fixed Type
No. Cat. No.Size
ColorD L
17 6601-3512 3.5 12.0 Purple
18 6601-3514 3.5 14.0 Purple
19 6601-3516 3.5 16.0 Purple
20 6601-3518 3.5 18.0 Purple
21 6601-4012 4.0 12.0 Green
22 6601-4014 4.0 14.0 Green
23 6601-4016 4.0 16.0 Green
24 6601-4018 4.0 18.0 Green
1.4) Self-tapping Variable Type
No. Cat. No.Size
ColorD L
25 6602-3512 3.5 12.0 None
26 6602-3514 3.5 14.0 None
27 6602-3516 3.5 16.0 None
28 6602-3518 3.5 18.0 None
29 6602-4012 4.0 12.0 Gold
30 6602-4014 4.0 14.0 Gold
31 6602-4016 4.0 16.0 Gold
32 6602-4018 4.0 18.0 Gold
LKSTD-A011(Rev.1508)
2.1) 1 Level
No. Cat. No.Size
ColorL
1 6711-1719 19.0 Blue
2 6711-1721 21.0 Blue
3 6711-1723 23.0 Blue
4 6711-1725 25.0 Blue
5 6711-1727 27.0 Blue
6 6711-1729 29.0 Blue
7 6711-1731 31.0 Blue
8 6711-1733 33.0 Blue
9 6711-1735 35.0 Blue
Ordering Information
22LKSTD-A011(Rev.1508)
2.2) 2 Level
No. Cat. No.Size
ColorL
10 6712-1735 35.0 Blue
11 6712-1737 37.0 Blue
12 6712-1739 39.0 Blue
13 6712-1741 41.0 Blue
14 6712-1743 43.0 Blue
15 6712-1745 45.0 Blue
16 6712-1747 47.0 Blue
17 6712-1749 49.0 Blue
18 6712-1751 51.0 Blue
2.3) 3 Level
No. Cat. No.Size
ColorL
19 6713-1748 48.0 Blue
20 6713-1751 51.0 Blue
21 6713-1754 54.0 Blue
22 6713-1757 57.0 Blue
23 6713-1760 60.0 Blue
24 6713-1763 63.0 Blue
25 6713-1767 66.0 Blue
26 6713-1769 69.0 Blue
27 6713-1772 72.0 Blue
2.4) 4 Level
No. Cat. No.Size
ColorL
28 6714-1765 65.0 Blue
29 6714-1769 69.0 Blue
30 6714-1773 73.0 Blue
31 6714-1777 77.0 Blue
32 6714-1781 81.0 Blue
33 6714-1785 85.0 Blue
34 6714-1789 89.0 Blue
35 6714-1793 93.0 Blue
36 6714-1797 97.0 Blue
Plate
No. Cat No. Description Q’ty
1 CP02-0101 Awl 1
2 CP02-0102 Punch Awl 1 (Optional)
3 CP02-0201 Drill Guide Fixed 1
4 CP02-0202 Drill Guide Variable 1
5 CP02-0312 Drill Bit 12mm 1
6 CP02-0314 Drill Bit 14mm 1 (Optional)
7 CP02-0501 Quick Coupling Handle 1
8 CP02-0601 Screw Driver 2
9 CP02-0602 Lock Driver 2
10 CP02-0701 Plate Bender 1
11 CP02-0702 Caliper 1
12 CP02-0703 Plate Holder 1
13 CP02-0704 Pin Holder 1
14 CP02-0705 Prefixation Pin 4
15 CP02-0801 Lock Remover 1 (Optional)
16 CP02-0001 Container (Cover) 1
17 CP02-0002 Container (Base) 1
18 CP02-0003 Upper Tray 1
19 CP02-0004 Lower Tray 1
20 CP02-0008 Implant Tray (Cover) 1
21 CP02-0009 Implant Tray (Base) 1
22 CP02-0010 Prefiaxion Pin Container Cover 1
23 CP02-0011 Prefiaxion Pin Container 1
Ordering Information
23LKSTD-A011(Rev.1508)
Instrument
CastleLoc-P Anterior Cervical Plate System
(LKIFU-A011 Rev.2015.04)
IMPORTANT NOTE TO OPERATING SURGEON
CastleLoc-P Anterior Cervical Plate System implants, like other internal fixation devices, have a finite useful life. The patient’s activity level has a significant impact on this useful life. Your patient must be informed that any activity increases the risk of loosening, bending, or breaking the implant components. It is essential to instruct patients about restrictions on their activities in the postoperative period and to examine the patient postoperatively to evaluate the development of the fusion mass and the status ofthe implant components. Even if solid bone fusion occurs, implant components may nevertheless bend, break, or loosen. Therefore,the patient must be made aware that implant components may bend, break, or loosen even though restrictions in activity are followed.
Because of the limitations imposed by anatomic considerations and modern surgical materials, metallic implants cannot be made tolast indefinitely. Their purpose is to provide temporary internal support while the fusion mass is consolidating. These types ofimplants are more likely to fail if no bone graft is used or if a pseudoarthrosis develops.
The surgeon may determine to remove these implants after bone fusion occurs. The possibility of a second surgical procedure mustbe discussed with the patient, and the risks associated with a second surgical procedure must also be discussed. If the implants do break, the decision to remove them must be made by the physician who must consider the condition of the patient and the risksassociated with the presence of the broken implant.
DESCRIPTION
The CastleLoc-P Anterior Cervical Plate System implants consist of plates and plate screws. All components are manufactured fromtitanium alloy (Ti-6Al-4V) per ASTM F136. The plates are available in lengths ranging from 17-93mm to accommodate one to four segments of fixation. The system also includes 3.5mm and 4.0mm titanium alloy screws which are available in lengths ranging from12mm-20mm.
CLEANING AND STERILIZATION
Implants are supplied non-sterile and are for single use only. So all implants used in surgery must be sterilized by the hospital prior to use. Otherwise, Instruments are supplied non-sterile and may be re-used. Instruments must be thoroughly cleaned prior to sterilization. Trained personnel must perform cleaning and mechanical inspection prior to sterilization.
Unless just removed from an unopened L&K BIOMED package, all instruments and implants must be disassembled (if applicable) and cleaned using neutral cleaners before sterilization and introduction into a sterile surgical field or (if applicable) return of the product to L&K BIOMED.
Manual Cleaning procedure
Use the neutral pH enzyme soaking solution that has been prepared.
Completely submerge the instrument in enzyme solution and allow it to soak for 20 minutes(water temperature: 35-45°C). Use a soft-bristled brush to gently clean the device (particular attention shall be given to crevices, lumens, mated surfaces and other hard-to-clean areas) until all visible soil has been removed. Lumens should be cleaned with a long, narrow, soft-bristled brush (i.e. pipe cleaner brush).
Note: Any assembled instruments such as caspa retractor, please disassemble the parts of retractor both blades, legs and body before submerge. And reassemble it before disassemble.
Note: The enzyme solution should be changed on a regular basis in order to ensure its effectiveness.
Remove the device from the enzyme solution and rinse in purified water (from one or any combination of the following processes: ultra-filter, RO,DI and/or distilled) for a minimum of 3 minutes. Thoroughly flush lumens, holes and other difficult to reach areas(water temperature: 35-45°C).
Prepare the neutral pH cleaning (detergent) solution and place in a sonication unit.
Completely submerge device in cleaning solution and sonicate for 10 minutes, preferably at 45-50 kHz.
Rinse instrument in purified water (from one or any combination of the following processes: ultra-filter, RO, DI and/or distilled) thoroughly for at least 3 minutes or until there is no sign of blood or soil in the rinse stream.
Repeat Steps 5 and 6 with freshly prepared cleaning solution.
Dry the instrument with a clean, disposable, absorbent, non-shedding wipe.
Visually inspect the devices under normal room lighting condition to verify all foreign debris has been removed thoroughly clean.
Verify that the devices are visually clean.
Automated cleaning procedure
Automated washer/disinfector systems are not recommended as the sole cleaning method for complex surgical instruments. These instruments should be cleaned following the manual cleaning procedure above. An automated system may be used as a follow-up method but is not required.
CAUTION:
Use of corrosive products and/or instruments including abrasive sponges and metal brushes should be avoided.
Visually inspect the devices under normal room lighting condition to verify all foreign debris has been removed thoroughly clean.
Verify that the instruments are in visually clean.
Instruction For Use
LKSTD-A011(Rev.1508) 24
STERILIZATION
All implants and instruments are supplied NON-STERILE. Prior to use, all implants and instruments should be placed in the instrumentation / implant case which will be either wrapped in an FDA cleared sterilization wrap or placed in a rigid sterilization container and placed in the autoclave for sterilization by the hospital using one of the following recommended cycles.
Implants previously implanted should not be re-used. Inspect visually for damage or contamination by biological residue. If damage or biological residue is observed on the
implant, it must be discarded.
USAGE
The surgeon must be thoroughly knowledgeable not only in the medical and surgical aspects of the implant, but must also be awareof the mechanical and metallurgical limitations of metallic surgical implants. Postoperative care is extremely important. The patient must be instructed in the limitations of the metallic implant and be warned regarding weight bearing and body stresses on theappliance prior to firm bone healing. The patient should be warned that noncompliance with postoperative instructions could lead to failure of the implant and possible need thereafter for additional surgery to remove the device.
INDICATIONS
The CastleLoc-P Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine (C2-C7). The System is indicated for use in the immobilization and stabilization of the spine as an adjunct to fusions in patients with:
degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies),
spondylolisthesis,
trauma (i.e. fractures or dislocations),
tumors,
deformity (defined as kyphosis, lordosis, or scoliosis),
pseudoarthrosis,
failed previous fusion,
spinal stenosis.
POSTOPERATIVE MOBILIZATION
Until maturation of the fusion is confirmed by radiographic examination, external immobilization (such as bracing) may be recommended, based on physician judgment.
Instructions to the patient to reduce stress on the implants are an equally important part of the attempt to avoid the occurrence of clinical problems that may accompany fixation failure.
CONTRAINDICATIONS
Active systemic infection or an infection localized to the site of the proposed implantation.
Severe osteoporosis may prevent adequate fixation of screws and thus preclude the use of this or any other spinal instrumentation system.
Patients who have been shown to be safely and predictably treated without internal fixation.
Open wounds.
METHOD CYCLE TEMPERATURE EXPOSURE TIME
Steam Gravity 270°F(132°C)
15Minutes
(Dry time, 15-30
Minute)
Steam Pre-Vacuum
270°F(132°C)
4Minutes
(Dry time, 20-30
Minute)
275°F(135°C)
3Minutes
(Dry time, 16
Minute)
Instruction For Use
25LKSTD-A011(Rev.1508)
RELATIVE CONTRAINDICATIONS:
Relative contraindications include any entity or condition that totally precludes the possibility of fusion (e.g., cancer, kidney dialysis
or osteopenia), obesity, certain degenerative diseases, and foreign body sensitivity.
WARNINGS
The following are specific warnings, precautions and adverse effects that should be understood by the surgeon and explained to the
patient. These warnings do not include all adverse effects that can occur with surgery in general, but are important considerations
particular to metallic internal fixation devices. General surgical risks should be explained to the patient before surgery.
Potential risks identified with the use of this device system, which may require additional surgery, include: device component fracture,
loss of fixation, nonunion, fracture of the vertebra, neurological injury, and vascular or visceral injury.
1. CORRECT SELECTION OF THE IMPLANT IS EXTREMELY IMPORTANT. The potential for satisfactory fixation is increased by
the selection of the proper size, shape, and design of the implant. While proper selection can help minimize risks, the size and
shape of human bones present limitations on the size, shape and strength of implants. Metallic internal fixation devices cannot
withstand activity levels equal to those placed on normal healthy bone. No implant can be expected to withstand indefinitely the
unsupported stress of full weight bearing.
2. IMPLANTS CAN BREAK WHEN SUBJECTED TO THE INCREASED LOADING ASSOCIATED WITH DELAYED UNION OR
NONUNION. Internal fixation appliances are load-sharing devices that are used to obtain alignment until normal healing occurs. If
healing is delayed, or does not occur, the implant may eventually break due to metal fatigue. The degree or success of union, loads
produced by weight bearing, and activity levels will, among other conditions, dictate the longevity of the implant. Notches, scratches
or bending of the implant during the course of surgery may also contribute to early failure. Patients should be fully informed of the
risks of implant failure.
3. MIXING METALS CAN CAUSE CORROSION. There are many forms of corrosion damage and several of these occur on metals
surgically implanted in humans. General or uniform corrosion is present on all implanted metals and alloys. The rate of corrosive
attack on metal implant devices is usually very low due to the presence of passive surface films. Dissimilar metals in contact, such
as titanium and stainless steel, accelerates the corrosion process of stainless steel and more rapid attack occurs. The presence of
corrosion often accelerates fatigue fracture of implants. The amount of metal compounds released into the body system will also
increase. Internal fixation devices, such as rods, hooks, wires, etc., that come into contact with other metal objects, must be made
from like or compatible materials.
4. PATIENT SELECTION. In selecting patients for internal fixation devices, the following factors can be extremely important to the
eventual success of the procedure:
A. The patient’s occupation or activity. If the patient is involved in an occupation or activity that includes heavy lifting, muscle
strain, twisting, repetitive bending, stooping, running, substantial walking, or manual labor, he/she should not return to these
activities until the bone is fully healed. Even with full healing, the patient may not be able to return to these activities successfully.
B. A condition of senility, mental illness, alcoholism, or drug abuse. These conditions, among others, may cause the patient to
ignore certain necessary limitations and precautions in the use of the appliance, leading to implant failure or other complications.
C. Certain degenerative diseases. In some cases, the progression of degenerative disease may be so advanced at the time of
implantation that it may substan-tially decrease the expected useful life of the appliance. For such cases, orthopedic devices can
only be considered a delaying technique or temporary remedy.
D. Foreign body sensitivity. The surgeon is advised that no pre-operative test can completely exclude the possibility of sensitivity
or allergic reaction. Patients can develop sensitivity or allergy after implants have been in the body for a period of time.
E. Smoking. Patients who smoke have been observed to experience higher rates of pseudoarthrosis following surgical procedures
where bone graft is used. Additionally, smoking has been shown to cause diffuse degeneration of intervertebral discs. Progressive
degeneration of adjacent segments caused by smoking can lead to late clinical failure (recurring pain) even after successful fusion
and initial clinical improvement.
5. ETC.
•If bony fusion does not occur within an expected period of time, the screws may break due to the high and sustained loading of
these devices. This has been noted in patients with delayed, psuedoarthrosis or non-union and can result in the need to revise the
device(s).
•This system should not be used with components of any other systems or manufacturers.
Based on fatigue testing results, when using this system, the physicians /surgeons should consider the levels of implantation, patient
weight, patient activity level, other patient conditions, etc., which may impact on the performance of this system.
•This system has not been evaluated for safety and compatibility in the MR environment
Instruction For Use
26LKSTD-A011(Rev.1508)
. The CastleLoc-P Anterior Cervical Plate System has not been tested for heating or migration in the MR environment.
PRECAUTIONS
1. SURGICAL IMPLANTS MUST NEVER BE REUSED. An explanted metal implant should never be reimplanted. Even though the device appears undamaged, it may have small defects and internal stress patterns which may lead to early breakage.
2. CORRECT HANDLING OF THE IMPLANT IS EXTREMELY IMPORTANT. Contouring of metal implants should be done only with proper equipment. The operating surgeon should avoid any notching, scratching or reverse bending of the devices when contouring.Alterations will produce defects in surface finish and internal stresses that may become the focal point for eventual breakage of the implant. Do not use implant if damage is suspected.
Excessive torque applied to the screws when seating the plate may cause failure of the bone resulting in stripped threads and/orcompromised screw purchase.
3. BENDING THE CONSTRUCT. Titanium alloy components should never be bent sharply or reverse bent. If a construct is over-contoured it is recommended that a new construct is contoured correctly rather than reverse bending the over-contoured construct.
4. REMOVAL OF THE IMPLANT AFTER HEALING. If the device is not removed after the completion of its intended use, any of the following complications may occur: (1) Corrosion, with localized tissue reaction or pain; (2) Migration of implant position resulting in injury; (3) Risk of additional injury from postoperative trauma; (4) Bending, loosening, and/or breakage, which could make removal impractical or difficult; (5) Pain, discomfort, or abnormal sensations due to the presence of the device; (6) Possible increased risk of infection; and (7) Bone loss due to stress shielding. The surgeon should carefully weigh the risks versus benefits when decidingwhether to remove the implant. Implant removal should be followed by adequate postoperative management to avoid refracture ordeformity. If the patient is older and has a low activity level, the surgeon may choose not to remove the implant thus eliminating the risks involved in a second surgery.
5. ADEQUATELY INSTRUCT THE PATIENT. Postoperative care and the patient’s ability and willingness to follow instructions are among the most important aspects of successful bone healing. The patient must be made aware of the limitations of the implant, and instructed to limit and restrict physical activities, especially lifting and twisting motions and participating in any type of sports. The patient should understand that a metallic implant is not as strong as normal healthy bone and could loosen, bend and/or break if excessive demands are placed on it, especially in the absence of complete bone healing. Implants displaced or damaged by improper activities may migrate and damage the nerves or blood vessels. An active, debilitated, or demented patient who cannot properly use weight-supporting devices may be particularly at risk during postoperative rehabilitation.
POSSIBLE ADVERSE EFFECTS
1. Nonunion, delayed union.
2. Bending or fracture of implant. Loosening of the implant.
3. Metal sensitivity, or allergic reaction to a foreign body.
4. Infection, early or late.
5. Decrease in bone density due to stress shielding.
6. Pain, discomfort, or abnormal sensations due to the presence of the device.
7. Nerve damage due to surgical trauma or presence of the device. Neurological difficulties including radicular pain, tethering of nerves in scar tissue, muscle weakness, and paraesthesia.
8. Vascular damage could result in catastrophic or fatal bleeding. Malpositioned implants adjacent to large arteries or veins could erode these vessels and cause catastrophic bleeding in the late postoperative period.
9. Dural tears experienced during surgery could result in the need for further surgery for dural repair, a chronic CSF leak or fistula, and possible meningitis.
10. Bursitis.
11. Paralysis.
12. Esophageal perforation, erosion or irritation.
13. Screw back-out, possibly leading to esophageal erosion, implant loosening, and/or reoperation for device removal.
14. Damage to lymphatic vessels and/or lymphatic fluid exudation.
15. Spinal cord impingement or damage.
16. Fracture of bony structures.
17. Degenerative changes or instability in segments adjacent to fused vertebral levels.
18. Death.
Instruction For Use
27LKSTD-A011(Rev.1508)
SHELF-LIFE
This product is not sterilized product. The shelf-life is not applicable to our product.
FURTHER INFORMATION
Recommended directions for use of this system are available at no charge upon request. If further information is needed or required,
please contact L&K BIOMED Co., Ltd..
Manufactured by:
L&K BIOMED Co.,Ltd.
#201, 202, 16-25, Dongbaekjungang-ro 16 beon-gil, Giheung-gu, Yongin-si,
Gyeonggi-do, 446-916, Korea
Tel. 82 - 1600 - 0841 Fax. 82 -70-7813-3355
L&K BIOMED Co.,Ltd.Address: #201, 202, 16-25, Dongbaekjungang-ro 16 beon-gil, Giheung-gu, Yongin-si, Gyeonggi-do, 446-916, Koreawww.lnkbiomed.comTel. 82 - 1600 - 0841 Fax. 82 -70-7813-3355
Instruction For Use
28
SYMBOL TRANSLATION
LOT NUMBERCATALOGNUMBER
QUANTITY
SINGLE
USE ONLYNON -STERILE
MANUFACTURER
See package ins ert forlabeling limitation
Federal Law(USA) restrictsthis device tosale,distribution,or use by or onthe order ofa physician.
“DATE OFMANUFACTURE
Consult
instructionfor use
AUTHORISED REPRESENTATIVE
IN THEEUROPEAN
COMMUNITY
“DO NOT USE
IF PACKAGE IS
DAMAGED
USE BYSTORE AT ROOM
TEMPERATURE
KEEP AWAY
FROM
SUNLIGHT
KEEP DRY
LKSTD-A011(Rev.1508)
Recommended