Continuously monitor water quality in any aquaculture system with the YSI monitoring and control instruments. Whether it is the need for a multiparameter solution or single measurement at one location, the YSI monitors are ideal.

Citation preview

Page 1: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Aquaculture Monitoring & Control Products

Page 2: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Models 5200A, 5400 and 5500D

5200A Multiparameter; DO, pH, Cond, Salinity, pH, ORP with 4 ea 10 amp relays 5200A-AC 5200A-DC

5400 Multichannel (4) Galvanic DO with 8 ea 10 amp relays and 6 ea auxiliary inputs 5400A-AC 5400A-DC

5500D Multichannel (4) optical DO with 8 ea 10 amp relays and 6 ea auxiliary inputs 5500D-01AC 5500D-01-DC 5500D-02AC 5500D-02-DC 5500D-04-AC 5500D-04-DC

Page 3: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

• Extended temperature range backlit graphic display with easy to use menu driven operation

• NEMA 4X, CE, ULC, ETL, IEC, RoHS, IP65, Made in USA

• Communication - RS485 and RS232 standard with optional Ethernet module and wireless radios

• E-mail and SMS alarming via PC connected to Ethernet or direct from instrument using optional Ethernet module

• Non-volatile memory for; 2000 data records, 1250 event relay/alarm records, 1000 entries/change records, 500 GLP records

• Flash based software allows for easy on site upgrades

• User configurable control setting for all parameters including PID function

5200A/5400/5500D Standard Features

Page 4: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

• Multiple timers for photo period management

• Feed Smart™ feed management control software

• Relays and software for dosing and control

• Fully configurable auxiliary (Aux) inputs for additional sensors and process control; level, pressure, flow, CO2, etc

• AC or DC power options; battery back-up function on all AC units

• Multi-instrument, multi-location capability

• Industry leading 2-year warranty

5200A/5400/5500D Standard Features

Page 5: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

5200A/5400/5500D Advantages

• Measurement and control managed at instrument level providing redundancy and no risk of complete network/system failure.

• Fully user configurable and scalable – fits any operation without custom programming and can be easily modified by the user as the needs change. Easy to add new sensors and change settings.

• Tank side display for easy calibration

Page 6: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

5200A/5400/5500D Advantages

• Multiparameter and multiple dissolved oxygen sensor options

• Choice of electrochemical and optical dissolved oxygen sensors

• Model 5500D accepts low maintenance optical dissolved oxygen sensors Extremely stable; no regular calibrations are

required No electrodes to clean or regular membrane

changes More accurate – no flow required for accurate

measurements

Page 7: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

YSI Advanatges vs PLC Systems

• PLCs have one central CPU so failure could be catastrophic and impact the entire monitoring system

• PLCs are located in one location and sensors can be far from CPU. This makes calibration and system checks very difficult.

• PLC software is custom for each system so initial development, changes and expansion require programming and expense.

• PLCs are not specifically designed for water testing applications so sensor options are limited and may have questionable accuracy.

Page 8: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Feed Smart™ Conditional Feed Timer

Manages feed delivery: Up to 100 feedings per day

User defines which days to feed

Decrease feed amount by user defined % based on DO, pH and Temperature values

Increases feed time on a daily basis by user defined FCR

Simple one-step calibration process

Page 9: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Feed Smart™ Conditional Feed Timer

Day 1 Day 2 Day 3 Day 4 Day 5 Day 6 Day 7

Feed % of Biomass (user entered value) 4.00% 4.00% 4.00% 4.00% 4.00% 4.00% 4.00%

Biomass (enter beginning value only) 2500.0 2566.7 2635.1 2705.4 2777.5 2851.6 2927.6

Daily Feed Amount (system calculated) 100.0 102.7 105.4 108.2 111.1 114.1 117.1

Cumulative Feed Amount (system calculated) 100.0 202.7 308.1 416.3 527.4 641.5 758.6

FCR Factor x.x to 1.0 (user entered value) 1.5

Make FCR a decimal (system calculated) 0.67

Increase in Biomass (system calculated) 66.7 68.4 70.3 72.1 74.1 76.0 78.1

Cumulative Biomass Increase (system calculated) 66.7 135.1 205.4 277.5 351.6 427.6 505.7

New Biomass (system calculated) 2566.7 2635.1 2705.4 2777.5 2851.6 2927.6 3005.7

Example Calculation Chart

Page 10: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Event Logging

Page 11: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Parameter Control Software • Set point option - Once High or Low limit is

reached control relay is enabled until set-point is reached

• DO has 2 levels of High and Low control

• Temperature control has a user defined minimum cycle time to protect compressors

• PID controller option (proportional-integral-derivative controller) for more accurate control

• Dosing option allows for mixing time after High or Low set point relay is enabled

Page 12: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

DO Control Range Low2

relay on

Range Low1

relay on

Range Low1

and Low2

relays off

4.6

4.8

5

5.2

5.4

5.6

5.8

6

Time

DO

Setpoint

Range Low1

Range Low1

relay on

Range Low1

and Low2

relays off

Range Low2

relay on

Range Low2

Alarm

Page 13: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Alarms

General Alarm

• Ability to configure single relay to control light or siren based on any alarm condition

Alarm Hold Off

• When enabled, acts as a filter for false alarms

• Holds alarm message (page and audible) until alarm condition is present for user determined length of time

• Input in seconds, minutes or

hours

Page 14: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

E-mail and SMS Alarming

[email protected]

Date: Wed 04 Dec 2009 18:10:46

Subject: AutoPoller Alarm Message

YSI5200A Alarm

Tilapia Tank #1

DO Low Alarm: 4.5 mg/L

Alarms

Page 15: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

5402 Serial to Ethernet Device

• Installs in the master instrument

• Connects master directly to Ethernet

• Configure through instrument menu and must have a static IP address

• Order as an accessory

Page 16: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Networking and Communication Options

Page 17: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

1. Data is collected at the instrument 2. Master instrument connected directly to PC 3. PC is connected to Ethernet - alarms are sent via SMS or E-mail 4. Remote users access local PC via PC Anywhere

Internet

Remote Connection

Alarm e-mail or SMS

Fire wall

PC Running AquaManager with alarms from PC

Local Network Ethernet

Alarm

Receipt

5200A/5400/5500D Network

Instrument No. 1

RS-485 Network

Instrument No. 32

Alarm Receipt

Direct RS 232 Connection

Page 18: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

1. Data is collected at the instrument, all configured as masters 2. Instruments connected directly to a radio 3. PC is connected to a radio to communicate with instrument’s

and is connected to Ethernet - alarms are sent via SMS or E-mail 4. Remote users access local PC via PC Anywhere

Internet

Remote Connection

Alarm e-mail or SMS

Fire wall

PC Running AquaManager with alarms from PC

Alarm

Receipt

5200A/5400/5500D

Instrument

No. 1

Radio Radio

Alarm Receipt

Local Network Ethernet

5200A/5400/5500D 5200A/5400/5500D

Connection with Spread

Spectrum Radio

Page 19: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Optional Remote Connection

Fire wall

PC Running AquaManager

1. Data is collected at the instrument 2. Instruments are connected directly to Ethernet with internal

serial to Ethernet device 3. SMS or e-mail alarms are generated at the master node 4. Remote users connect to network with AquaManager or local PC

via PC Anywhere

Internet

Alarm e-mail or SMS

Local Network Ethernet

Alarm Receipt

5200A/5400/5500D Network

Alarming from master device

Instrument No. 1

Instrument No. 32

RS-485 Network

Ethernet Connection

Page 20: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Optional Remote Connection

Fire wall

PC Running AquaManager with alarming from PC

1. Data is collected at the instrument 2. Instruments are connected directly to Ethernet with internal

serial to Ethernet device 3. SMS or e-mail alarms are generated at the local PC which is

running AquaManager 4. Remote users connect to network with AquaManager or local PC

via PC Anywhere

Internet

Alarm e-mail or

SMS S

Local

Network

Ethernet

Alarm Receipt

5200A/5400/5500D Network

Instrument No. 1

Instrument No. 32

RS-485 Network

Local Network Ethernet

Ethernet Connection

Page 21: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

AquaManager Software

Data Management

• Graphing

• Data export

• Manual data entry

• Statistical analysis System Management

• Explorer based network setup and support

• Ethernet communication setup

• E-mail and SMS setup

• Alarming, set-point and control management

Page 22: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Graphing

Page 23: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Set-point and Control Management

Page 24: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

System Management

Page 25: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

Mapping Screen

• Gives a quick visual

of the entire

operation

• Images include

animations

• Easy for user to set

up

Page 26: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

AquaViewer Apple device App

Set up screen #1 Set up screen #2 Network Viewer screen: • Normal • Control in yellow • Alarms in red

Instrument Viewer screen: • Normal in green • Control in yellow • Alarms in red

• iPhone • iPad • iPod Touch (generation #3 or higher)

Page 27: YSI 5500D 5400 and 5200A aquaculture monitors and controllers
Page 28: YSI 5500D 5400 and 5200A aquaculture monitors and controllers

QUESTIONS?

YSI.com/aquaculture