53
Internal Quality Control in coagulation lab Dr Ankit Raiyani Haematology Department Sahyadri Speciality Hospital

Internal quality control (IQC) in coagulation lab

Embed Size (px)

Citation preview

Page 1: Internal quality control (IQC) in coagulation lab

Internal Quality Control in coagulation lab

Dr Ankit RaiyaniHaematology Department

Sahyadri Speciality Hospital

Page 2: Internal quality control (IQC) in coagulation lab

Presentation layout• Introduction to Quality Assurance and Quality Control• Definitions of common terms, Accuracy vs. Precision• Preparation/selection of reagents-

• PTT reagent PT reagent MNPT• QC for manual coagulation tests• Quality control procedures- schedule• Internal Quality Control• LJ charts• Westgard rules• Other measures for IQC

Page 3: Internal quality control (IQC) in coagulation lab

Introduction

In the haematology laboratory it is essential to ensure that the right test is carried out on the right specimen and that the correct results are delivered to the appropriate recipient without delay.

Quality assurance (QA) is defined as the overall programme for achieving these objectives.

Quality control (QC) is defined as measures that must be included during each assay run to verify that the test is working properly.

Page 4: Internal quality control (IQC) in coagulation lab

Quality Assurance vs. Quality Control

Quality Assurance Quality ControlAn overall

management plan to guarantee the

integrity of data(The “system”)

A series of analytical

measurements usedto assess thequality of the analytical data(The “tools”)

Page 5: Internal quality control (IQC) in coagulation lab

The Quality Assurance Cycle

• Data and Lab Management

• Safety• Customer

Service

Patient/Client PrepSample Collection

Sample Receipt and Accessioning

Sample TransportQuality Control

Record Keeping

ReportingPersonnel CompetencyTest Evaluations

Testing

Page 6: Internal quality control (IQC) in coagulation lab

Hurdles in Quality Assurance

6

Page 7: Internal quality control (IQC) in coagulation lab

Hurdles in Quality Assurance

7

Pre-analytical

Page 8: Internal quality control (IQC) in coagulation lab

Hurdles in Quality Assurance

8

Analytical

Page 9: Internal quality control (IQC) in coagulation lab

Hurdles in Quality Assurance

9

Post-analytical

Page 10: Internal quality control (IQC) in coagulation lab

Terms relating to quality control in clinical laboratory practicePrecision Can be controlled by replicate tests and by repeated tests on previously measured

specimens

Accuracy Can be checked only by the use of reference materials that have been assayed by reference methods

Controls Are preparations that are used for either internal quality control or external quality assessment. They should not be used as standards

Error The discrepancy between the result of a measurement and the true (or accepted true) value

Random error

An error which varies in an unpredictable manner, in magnitude and sign, when a large number of measurements of the same quantity are made under effectively identical conditions

Systemic error (bias)

An error which, in the course of a number of measurements of the same value of a given quantity, remains constant when measurements are made under the same conditions or varies according to a definite law when conditions change

Page 11: Internal quality control (IQC) in coagulation lab

Accuracy vs. Precision

Accuracy Precision

Meaning How well a measurement agrees with an accepted value

How well a series of measurements agree with each other

Achieved with…

Calibration Quality control

When low… Increases systemic errors

Increases random errors

Page 12: Internal quality control (IQC) in coagulation lab

Calibration

Calibration is a comparison of measuring equipment against a standard instrument of higher accuracy to detect, correlate, adjust, rectify and document the accuracy of the instrument being compared.

Page 13: Internal quality control (IQC) in coagulation lab

Preparation/selection of reagents

Page 14: Internal quality control (IQC) in coagulation lab

Reagent selection for PTT

LA-sensitive reagents- (low-phospholipid reagents)

• Siemens Actin FSL• IL HemosIL aPTT-SP • Stago PTT-LA

Intermediate responsiveness to LA

• Stago STAPTT-LA • IL HemosIl SynthASil

Low responsiveness to LA

• Siemens Actin FS • Stago C.K. Prest

Page 15: Internal quality control (IQC) in coagulation lab

Reagent selection for PTT

Selection of aPTT reagent depends on indication of aPTT--

• used for evaluating the factor VIII, IX, XI deficiency and effect of heparin• sensitive to deficiencies of factors VIII, IX and XI at concentrations of 0.35–0.4 iu/ml.• responsive to unfractionated heparin over the therapeutic range of approximately 0.3–0.7 iu/ml.• low responsiveness to LA

The routine laboratory aPTT

• High responsiveness to LA (silica+ low PL content)• Two phospholipid based clotting assays, one of which may be aPTT based, are required1

LA screening

• Used in labs with mixed sample load, when only one reagent is feasible/desired

Intermediate responsiveness to both factor levels and LA

1PENGO, V., TRIPODI, A., REBER, G., RAND, J. H., ORTEL, T. L., GALLI, M. and DE GROOT, P. G. (2009), Update of the ISTH guidelines for lupus anticoagulant detection. Journal of Thrombosis and Haemostasis, 7: 1737–1740.

Page 16: Internal quality control (IQC) in coagulation lab

Reagent selection for PT (Thromboplastin)Contains tissue factor and coagulant phospholipids

• Manufactured using recombinant human tissue factor produced in Escherichia coli and synthetic phospholipids, • Do not contain any other clotting factors such as prothrombin, factor VII and factor X. • Highly sensitive to factor deficiencies and oral anticoagulant-treated patient plasma samples

Should have ISI value as close to 1 as possible.

Source- human, bovine, rabbit brain/lung, human placenta, recombinant.

Animal/human source- potential hazard of viral, prion and other infections from handling

Recombinant thromboplastins-

Page 17: Internal quality control (IQC) in coagulation lab

What is ISI?

International Sensitivity Index

It is the quantitative measure, (in term of first international reference preparation of thromboplastin, human, combined, coded 67/40,) of the responsiveness of PT system to the defect induced by oral anticoagulants.

Page 18: Internal quality control (IQC) in coagulation lab

Calculation of ISI

Calibration of a test thromboplastin must be against a reference thromboplastin of the same species e.g. human against human, rabbit against rabbit

Tests are performed By using reference thromboplastin and test thromboplastinon 20 normal donors not on anticoagulants and 60 patients who have been on oral anticoagulant treatment for at least 6 weeks and has stable INR.

Prothrombin Times are performed in duplicate for each sample and the mean for each pair of tests derived.

If there is more than a 10% difference in the clotting times between duplicate samples, the tests on that plasma sample should be repeated

Mean of each pair of PT results are plotted on double-log paper with the reference sample on the Y axis and the test plasma on the X-axis.

Page 19: Internal quality control (IQC) in coagulation lab

• A best fit line is drawn with points above the highest recorded PT and the lowest PT (see figure above). The slope of the line is calculated

• Slope= AC/BC

• ISItest = ISIStandard X Slope

• INR= (PT/MNPT)ISI

Page 20: Internal quality control (IQC) in coagulation lab

Preparation of pooled normal plasma and derivation of MNPTCitrated blood samples from at least 20 healthy volunteers (with PT, aPTT in normal range) should be collected. Equal numbers of male and female (not pregnant, not on OC pills) volunteers.

Platelet poor plasma should be prepared. After pooling it should be immediately frozen at -80⁰ C in aliquots.

The mean of PT of individual samples is calculated and taken as mean normal prothrombin time (MNPT).

SD is calculated and reference range is taken as -2 SD to +2 SD.

This pooled plasma is used as control with each batch of samples.

Page 21: Internal quality control (IQC) in coagulation lab

MNPT calculation- 18/6/15

No PT PTT1 11.5 28.62 12.4 27.63 10.5 29.24 10.6 26.25 10.6 32.16 11.7 29.87 11.1 30.78 12.2 31.69 11.9 30.9

10 11.3 31.511 10.6 29.112 12.3 29.713 10 26.314 11.9 28.715 11.6 33.316 11.2 31.417 10.5 28.218 12.5 28.719 10.6 27.820 11.1 27.721 10.9 29.522 11.3 31.923 11.3 29.924 12.5 32

total 272.1 712.4Arithmetic Mean 11.3 29.6SD 0.724 1.899

Range 9.8 to 12.7 25.8 to 33.4

Page 22: Internal quality control (IQC) in coagulation lab

QC for manual coagulation tests

All tests and controls should be performed in duplicates.

• Serviced and Calibrated every 6 months or after break down• Put on sturdy platform to reduce vibrations.• Sample load should be well balanced.

Operator skill

• Temperature- 37⁰ C (checked in all four corners)• Transparent walls for transillumination• Filled with distilled water till indicated level.

Required instruments should be properly calibrated.

• Calibrated. • Disposable tips.

Centrifuge

Waterbaths

Pipette

Page 23: Internal quality control (IQC) in coagulation lab

Calibration of centrifuge

Tachometer is used for calibration of RPM.

Actual RPM is measured with tachometer at uniform intervals (minimum 5) for 5 times.

Difference between actual and set values is calculated.

If difference is more than +/- 5%, then repeat the test. If repeat reading again shows significant difference corrective measures are required.

Timer and Temperature, when applicable, should also be calibrated.

Page 24: Internal quality control (IQC) in coagulation lab

Calibration of pipettes

Analytical balance is used to weigh pipetted volume in controlled environment. Distilled water is used as substrate.

The pipette is checked 5 times with the maximum volume (nominal volume), the minimum volume or 10% of the maximum volume, whichever is higher.

With multichannel pipettes, both volumes are tested with the two edge channels

Conversion of mass to volume V = (w + e) x Z

• V = Volume (µl), w = Weight (mg), e = Evaporation loss (mg), • Z = Conversion factor for mg/µl conversion (temperature, pressure)

Acceptable accuracy and precision range varies with pipette capacity.

Page 25: Internal quality control (IQC) in coagulation lab

Quality control procedures

Page 26: Internal quality control (IQC) in coagulation lab

Quality control procedures

The procedures that should be included in a quality assurance programme vary with the tests undertaken, the instruments used and (especially if these include a fully automatic system) the size of the laboratory and the numbers of specimens handled.

Some control procedures should be performed daily and other performance checks should be done at appropriate intervals.

The latter is particularly important when there is a change in staff and after maintenance service or repair has been carried out on equipment.

Page 27: Internal quality control (IQC) in coagulation lab

Schedule for quality control procedures

1. Calibration with reference standards

Instruments, pipettes6-month intervals or more frequently if control chart or EQA indicates bias or fluctuation in results and after any repair/service

Others: thermometers, scales, etc Annually

2. Control chart with control materialDaily or more frequently with each batch of specimensDuplicate tests on two or three patients’ samples: if control chart or delta check shows discrepancies

3. Analysis of patients’ resultsCumulative results: following previous tests and if changes in clinical state

4. EQAS performanceAssessment monthly

Page 28: Internal quality control (IQC) in coagulation lab

Components of Quality control

Internal quality control (IQC) • Monitoring the haematology test procedures to ensure continual

evaluation of the reliability of the daily work of the laboratory with validation of tests before reports are released

External quality assessment (EQA) • Evaluation by an outside agency of the between-laboratory and

between-method comparability. It can be organized nationally, regionally or internationally

Page 29: Internal quality control (IQC) in coagulation lab

Internal Quality Control

IQC is based on monitoring the haematology test procedures that are performed in the laboratory and includes measurements on specially prepared materials and repeated measurements on routine specimens, together with daily statistical analysis of the data.

Primarily a demonstration of Precision.

It ensures continual checks that the established reliability of the laboratory’s work does not fluctuate and that reports are validated before they are released

Page 30: Internal quality control (IQC) in coagulation lab

IQC includes

Control charts with tests on control materials

Duplicate tests on a proportion of the specimens

Consistency of mean values of patient data

Correlation check

Page 31: Internal quality control (IQC) in coagulation lab

Control charts

These were first applied in clinical chemistry by Levey and Jennings.

They are now widely used in haematology for both automated and manual procedures.

Samples of the control specimen are included in every batch of patients’ specimens and the results are checked on a control chart.

To check precision, it is not necessary to know the exact value of the control specimen.

Page 32: Internal quality control (IQC) in coagulation lab

ControlsLevel 1 controls- normal controls

Level 2 controls- low abnormal controls

Level 3 controls- high abnormal controls (not used routinely)

Minimum of one level QC at least once a day.

Patient samples >25 per day Employ 2 levels of QC at least once a day.

Patient samples >75 per day Employ 2 levels of QC at least twice a day at appropriate intervals

Controls are intended to simulate random sampling, they must be treated exactly like the patients’ specimens.

Page 33: Internal quality control (IQC) in coagulation lab

Preparation of LJ charts

Mean value and standard deviation (SD) of the control specimen should first be established in the laboratory where the tests are performed.

Using arithmetic graph paper, a horizontal line is drawn to represent the mean (as a base), and on an appropriate scale of quantity and unit, lines representing +2SD and −2SD are drawn above and below the mean.

The results of successive control sample measurements are plotted. If the test is satisfactory, sequential results oscillate about the mean value and <5% of the results fall outside 2SD.

Page 34: Internal quality control (IQC) in coagulation lab

LJ chart PT / low abnormal May 2015

Page 35: Internal quality control (IQC) in coagulation lab

LJ chart PT / low abnormal June 2015

Page 36: Internal quality control (IQC) in coagulation lab

LJ chart PT / low abnormal June 2015

Reason ??

Page 37: Internal quality control (IQC) in coagulation lab

LJ chart aPTT / NCPNP

Page 38: Internal quality control (IQC) in coagulation lab

LJ chart F VIIIc / NCPNP

Page 39: Internal quality control (IQC) in coagulation lab

Westgard rules

“Multirule Quality Control” developed by Dr. James O. Westgard based on statistical concepts

Uses a combination of decision criteria or control rules

Allows determination of whether an analytical run is “in-control” or “out-of-control”

Page 40: Internal quality control (IQC) in coagulation lab

Westgard multirule system

12S rule

13S rule

22S rule

R4S rule

41S rule

10X rule

Page 41: Internal quality control (IQC) in coagulation lab

Westgard 12S rule

“warning rule”

One of two control results falls outside ±2SD

Alerts tech to possible problems

Not cause for rejecting a run

Must then evaluate the 13S rule

Page 42: Internal quality control (IQC) in coagulation lab

12S Rule = A warning to trigger careful inspection of the control data

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Mean

Day

+1SD

+2SD

+3SD

-1SD

-2SD

-3SD

12S rule violation

Page 43: Internal quality control (IQC) in coagulation lab

Westgard – 13S Rule

If either of the two control results falls outside of ±3SD, rule is violated

Run must be rejected

If 13S not violated, check 22S

Page 44: Internal quality control (IQC) in coagulation lab

13S Rule = Reject the run when a single control measurement exceeds the +3SD or -3SD control limit

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Mean

Day

+1SD

+2SD

+3SD

-1SD

-2SD

-3SD

13S rule violation

Page 45: Internal quality control (IQC) in coagulation lab

Westgard – 22S Rule

2 consecutive control values for the same level fall outside of ±2SD in the same direction, or

Both controls in the same run exceed ±2SD

Patient results cannot be reported

Requires corrective action

Page 46: Internal quality control (IQC) in coagulation lab

22S Rule = Reject the run when 2 consecutive control measurements exceed the same +2SD or -2SD control limit

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Mean

Day

+1SD

+2SD

+3SD

-1SD

-2SD

-3SD

22S rule violation

Page 47: Internal quality control (IQC) in coagulation lab

Westgard – R4S Rule

One control exceeds the mean by –2SD, and the other control exceeds the mean by +2SD

The range between the two results will therefore exceed 4 SD

Random error has occurred, test run must be rejected

Page 48: Internal quality control (IQC) in coagulation lab

R4S Rule = Reject the run when 1 control measurement exceed the +2SD and the other exceeds the -2SD control limit

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Mean

Day

+1SD

+2SD

+3SD

-1SD

-2SD

-3SD

R4S rule violation

Page 49: Internal quality control (IQC) in coagulation lab

Westgard – 41S Rule

Requires control data from previous runs

Four consecutive QC results for one level of control are outside ±1SD, or

Both levels of control have consecutive results that are outside ±1SD

Page 50: Internal quality control (IQC) in coagulation lab

Westgard – 10X Rule

Requires control data from previous runs

Ten consecutive QC results for one level of control are on one side of the mean, or

Both levels of control have five consecutive results that are on the same side of the mean

Page 51: Internal quality control (IQC) in coagulation lab

10x Rule = Reject the run when 10 consecutive control measurements fall on one side of the mean

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Mean

Day

+1SD

+2SD

+3SD

-1SD

-2SD

-3SD

10x rule violation

Page 52: Internal quality control (IQC) in coagulation lab

Alternate ways of checking the precision of routine workDuplicate tests on patients’ samples

• Test of precision can be performed by repeat analysis on samples from two or three patients

Correlation Check

• any unexpected result of a test must be checked to see whether it can be explained on clinical grounds or whether it correlates with other tests

“Delta” check

• Blood count from the same patient checked repeatedly over 2-3 weeks• Values should not change significantly providing the patients condition has not altered significantly

Patient data

• Need enough through put (i.e. >100 samples /day)• MCV, MCHC, MCH should not vary, providing the patient population is stable and samples from different sets of patients are not batched together.• Mean of 20 successive patients can be plotted and any drift in the calibration of the analyzer will be recognized

Page 53: Internal quality control (IQC) in coagulation lab

THANK YOU!!