Arterial Line Insertion Dr Tarun Bhatnagar

Arterial line insertion

Embed Size (px)

Citation preview

Page 1: Arterial line insertion

Arterial Line Insertion

Dr Tarun Bhatnagar

Page 2: Arterial line insertion

Indications Preparation & Equipments Positioning Insertion Complications Mechanism of Action Troubleshooting

Page 3: Arterial line insertion

1.Continuous, beat-to-beat blood pressure

measurement. - Hemodynamically unstable pts /ICU pts requiring inotropic support - Patients undergoing major surgery

2.Frequent arterial blood gas analysis -pts with respiratory failure on ventilator -severe acid/base disturbance.

Indications

Page 4: Arterial line insertion

Advantages of IBP measurement Continuous blood pressure recording Accurate blood pressure recording even when patients are profoundly hypotensive

vs NIBP which is difficult or inaccurate Real time Visual DisplayDisadvantages of IBP measurement Potential complications Skilled technique reqd Expensive

Page 5: Arterial line insertion

The radial artery has low complication rates compared with other sites. It is a superficial artery which aids

insertion, and also makes it compressible for haemostasis

The ulnar, brachial, axillary, dorsalis pedis, posterial tibial, femoral arteries are alternatives.

Choice of arterial site

Page 6: Arterial line insertion

The Palmar Arch

Page 7: Arterial line insertion

Preparation- Allen’s Test

The idea here is to figure out if the ulnar artery will supply the hand with enough blood, if the radial artery is blocked with an a-line.

Allen’s test is recommended before the insertion of a radial arterial line.

This is used to determine collateral circulation between the ulnar and radial arteries to the hand

If ulnar perfusion is poor and a cannula occludes the radial artery, blood flow to the hand may be reduced.

The test is performed by asking the patient to clench their hand. The ulnar and radial arteries are occluded with digital pressure.

The hand is unclenched and pressure over the ulnar artery is released. If there is good collateral perfusion, the palm should flush in less than 6 seconds.

Page 8: Arterial line insertion

Arterial cannula Made from polytetrafluoroethylene (‘Teflon’) to

minimize the risk of clot formation 20G (pink) cannula - adult patients

22G (blue)- paediatrics 24G (yellow) - neonates and small babies

Larger gauge cannulae increase the risk of thrombosis, smaller cannulae cause damping of the signal.

The cannula is connected to an arterial giving set.

Equipment

Page 9: Arterial line insertion

Arterial cannula

Page 10: Arterial line insertion

Arterial set. - Specialized plastic tubing, short and stiff to reduce resonance, connected to a 500 ml

bag of saline.

Saline bag

-500 ml 0.9% saline pressurized to 300 mmHg using a pressure bag, i.e. a pressure higher than arterial systolic pressure to prevent backflow from the cannula into the giving set. -The arterial set and pressurized saline bag with 2500units Heparin incorporate a continuous slow flushing system of 3–4 ml per hour to keep the line free from clots.

-The arterial set and arterial line should be free from air bubbles. - The line is attached to a transducer. -Do not allow the saline bag to empty

◦ –To maintain patency of arterial cannula.◦ –To prevent air embolism◦ –To maintain accuracy of blood pressure reading◦ –To maintain accuracy of fluid balance chart ◦ –To prevent backflow of blood

Transducer, amplifier and electrical recording equipment. -The transducer is zeroed and placed level with the heart.

Page 11: Arterial line insertion

Tape and/or steri-strips An arm board or towel roll Opsite or Tegaderm cover dressing Local anesthetic (1% or 2% lidocaine ,lidocaine

cream) Suture material for femoral arterial line

placement (2.0 silk) Scissors Monitor cable for transducing arterial waveform. Benzoin solution

Page 12: Arterial line insertion

1. Ensure that all preprocedure steps are taken

2. Assure that pressure tubing with transducer is connected to bedside monitor.

3. Perform the Allen’s test to assure adequate collateral blood flow if using the radial artery.

4. Wash hands and don gloves 5. For the radial artery, the arm is restrained, palm up, with an armboard to hold the wrist dorsiflexed

Procedure

Page 13: Arterial line insertion

Positioning

For the radial artery, the most common insertion, the arm is restrained, palm up, with an armboard to hold the wrist dorsiflexed

Page 14: Arterial line insertion

5. Apply anesthetic agent (local lidocaine 1-2% or lidocaine cream). 6. Locate pulsating artery via palpation. 7. Cleanse area selected for arterial line placement. 8. Prepare patient for puncture. 9. Stabilize artery by pulling skin taut. 10. Puncture skin at 45-60 degree angle for radial artery; 90

degrees for femoral artery. 11. Advance catheter when flash of blood is observed in catheter. 12. Connect to pressure I.V. tubing and check for arterial waveform

on bedside monitor. 13. Cleanse area of any blood and allow site to dry. 14. Apply Benzoin to cleansed area and allow to dry and become

“tacky.” 15. Secure arterial line with tape and cover with a Tegaderm

dressing. 16. Secure I.V. tubing to prevent it from being caught and pulling

on arterial catheter. If a femoral arterial line is placed, it should be secured with a suture.

17. Properly dispose of the I.V. sharps and other used materials.

Page 15: Arterial line insertion

Insertion techniques 1. Direct cannulation

2. Transfixation

3. Guidewire (Seldinger) technique

Page 16: Arterial line insertion

Haemorrhage may occur if there are leaks in the system. Connections must be tightly secured and the giving set and line closely observed..

Emboli. Air or thrombo emboli may occur. Care should be taken to aspirate air bubbles

Accidental drug injection may cause severe, irreversible damage to the hand. -No drugs should be injected via an arterial line

- The line should be labelled (in red) to reduce the likelihood of this occurring

Arterial vasospasm Partial occlusion due to large cannula width, multiple attempts

at insertion and long duration of use Permanent total occlusion Sepsis or bacteraemia secondary to infected radial arterial

lines is very rare (0.13%); -local infection is more common. -if the area looks inflamed the line site should be changed.

Complications

Page 17: Arterial line insertion
Page 18: Arterial line insertion

Never inject anything into an arterial cannula or arterial line

–Concentration of a drug into the tissues served by the cannulated artery can result in cell death

–Skin necrosis, severe gangrene, limb ischemia, amputation & permanent disabilities

Page 19: Arterial line insertion

How it works ??? Mechanism of action A transducer is a device that

reads the fluctuations in pressure – it doesn’t matter if it’s arterial, or central venous, or PA

The column of saline in the arterial set transmits the pressure changes to the diaphragm in the transducer

The transducer reads the changing pressure, and changes it into an electrical signal that goes up and down as the pressure does which is displayed as an arterial waveform

The transducer connects to the bedside monitor with a cable, and the wave shows up on the screen, going from left to right

Page 20: Arterial line insertion

The transducer has to sit in a “transducer holder” – this is the white plastic plate that screws onto the rolling pole that holds the whole setup.  

The transducer has to be levelled correctly-to make sure that it’s at the fourth intercostal space, at the mid-axillary line (Phlebostatic axis) 

Make sure there’s no air in the line before you hook it up to the patient – use the flusher to clear bubbles out of the tubing. 

Zero the line to atmospheric pressure properly Choose a screen scale that lets you see the waveform

clearly.

Transducer - A couple of things to remember

Page 21: Arterial line insertion

To ensure accuracy of readings Flush the device & turn it off to patient but

open to atmosphere These exert pressure on transducer This pressure is called zero Zero once per shift or if values are

questionable Ensure flush bag is pumped up

Zeroing

Page 22: Arterial line insertion

Arterial pressure waveform

Once inserted, an arterial waveform trace should be displayed at all times

This confirms that the invasive arterial BP monitoring is set up correctly, and minimizes problems.

Page 23: Arterial line insertion

Typical arterial waveform:

The highest point - systolic pressure, -the lowest is the diastolic. Everybody see the little notch on the diastolic downslope? – there’s one in each beat.

A little after the beginning of diastole – the start of the downward wave – the aortic valve flips closed, generating a little back-pressure bump: called the “dicrotic notch”..

 

Page 24: Arterial line insertion

Is it accurate?

Now we know how the arterial pressure monitoring system works, we need to be able to decide whether or not the trace (and BP in numerical format) is accurate.

Failure to notice this may lead to unnecessary, or missed treatments for our patients.

There are 2 main abnormal tracing problems that can occur once the monitor gain is set correctly.

Page 25: Arterial line insertion

Dampened trace

Page 26: Arterial line insertion

Dampened: wide, flattened tracing

Dampening occurs due to:◦ air bubbles◦ overly compliant, distensible tubing◦ catheter kinks◦ clots◦ injection ports◦ low flush bag pressure or no fluid in the flush bag◦ Improper scaling◦ Severe hypotension if everything else is ruled

out This type of trace Under estimate SBP, over

estimate DBP

Page 27: Arterial line insertion

Resonant trace

Page 28: Arterial line insertion

Resonant: ‘spiked’ tracing Resonance occurs due to:

◦ long tubing◦ overly stiff, non-compliant tubing◦ increased vascular resistance◦ reverberations in tubing causing harmonics that

distort the trace (i.e. high systolic and low diastolic)

◦ non-fully opened stopcock valve This type of trace Over estimate SBP, under estimate DBP

Page 29: Arterial line insertion

Arterial lines measure systolic BP approximately 5 mmHg higher and the diastolic BP approximately 8 mmHg lower compared to non-invasive BP (NIBP) measurement

Comparison with non-invasive BP

Page 30: Arterial line insertion

Troubleshooting

Page 31: Arterial line insertion

“It takes a year just to learn which way to turn the stopcocks!” This is really true: some stopcocks point to where they’re open, and some point to where they’re closed – it just takes some time to learn which is which.

The trick is remembering which way to turn the stopcock, and avoiding a mess.

Don’t forget to clear the stopcock, recap, and then flush the line.

Keep things nice and sterile. 

How to draw blood samples from a-lines

Page 32: Arterial line insertion

This probably means that the artery being monitored or gone into spasm.

You need to think about things that might make this happen: -Is the patient very cold? -Are his extremities poorly perfused? -Is he on a “shipload” of pressors, making his arterial bed tighten up –- Is he “dry” as well?

Sometimes arteries become unhappy with catheters in them, and you just have to convince the team that the patient needs a new one placed in another site.

What if the a-line has a good tracing on the screen, but I can’t draw blood from it?

 

Page 33: Arterial line insertion

The first thing to think about is:

1.Is the arterial catheter still in place? Yes? Try drawing with a 3cc syringe from the stopcock – if it draws normally, then you’ve got a hardware problem

2.Cables become loose?

3.Did the screen scale get accidentally set to, say, 40, instead of 150 or 200mm of pressure?- you’ll only see a flat line.

4.Is it a transducer setup Failure – try a new setup.   If the line doesn’t draw –

-Is there a clot in the hub?-Try taking the site dressing down – is the catheter kinked going into the patient?- Sometimes art-lines just fail – the artery spasms and won’t open up – time for a new site.

   

What if I lose the trace completely?

Page 34: Arterial line insertion

The routine now is 96 hours – make sure that you label the line setup when you hang it. Obviously, change the line setup if it is contaminated in any way.

 

How often does the transducer setup have to be changed? 

Page 35: Arterial line insertion

Usually this will be pretty obvious: the pulse will diminish, or go away altogether. The hand may look dusky, or be cold, or lose some sensation – remember to assess for coloring, sensation, motion, and capillary refill.

If you think that the a-line is threatening the patient’s hand, let the team know right away, and be ready to set up for another insertion somewhere else if the line is still necessary.

 

How do I know if the patient’s hand is at risk?

Page 36: Arterial line insertion

  Compress the site with a sterile 4x4 for at

least 5 minutes, or longer if the patient is anticoagulated.

Assess the perfusion of the hand. Make sure you put the patient on the non-

invasive cuff at meaningful intervals while you talk to the team about replacing the line

What do I do if the patient pulls out her a-line?

Page 37: Arterial line insertion

This is usually pretty obvious – the patient is hemodynamically stable, needs only one or two blood draws in a day, no more need for ABGs

How do I know when it’s safe to take out my patient’s a-line? 

Page 38: Arterial line insertion

  Disconnect the cable from the monitor which will

automatically turn off the alarms. Take out the sutures in the usual way with a fresh

sterile kit. Have a gauzepiece ready, pull the catheter, and manually

compress the site for at least 3 to 5 minutes. Make sure the patient’s hand is still perfused. Check for hematoma or bleeding, put a compression

dressing on the site (not too tight!), which you can then take off after about an hour.

Recheck the site hourly for a few hours afterwards – a hematoma could still form, and since there isn’t a whole lot of room in a wrist, you’d definitely want to know!

 

How do I remove the line?

Page 39: Arterial line insertion

Any Questions ????