Upload
vaishnavi-patil
View
218
Download
0
Embed Size (px)
Citation preview
1
VAISHNAVI PATIL1ST M-PHARMMMCP BELGAUM.
THERMO ANALYTICAL METHODS OF ANALYSIS
2
DEFINITION
Differential scanning calorimetry measures the temperatures and heat flows associated with transitions in the materials as a function of time and temperature in a controlled atmosphere
It provides quantitative and qualitative information about physical and chemical changes that involve exothermic and endothermic or changes in heat capacity
3
DSC can measure….
Glass transitions Melting and boiling points Crystallization time and temperature Heats of fusion and reactions Specific heat capacity Oxidative/thermal stability Reaction kinetics Purity
4
5
6
TYPES OF DSCHyper-DSC
Based on principle that high heating rates give large broad transitions.
•Heating rates typically 400-500oC/min
•Need very small sample sizes (~nanograms)
Advantages:
•A quick overview of new sample
•Picking out minute transition
Disadvantages:
•Accuracy: transitions can be shifted by as much as 40oC
•Repeatabiliy: Very sensitive to thermal lag.
7
Modulated DSC
• Composite heating profile: • Determines heat capacity and separates heat flow into that due to
reversible and non-reversible events.
Typically:
Heating rates: 0 - 50C
Modulation:
Period: 60 second
Amplitude: +/-10C
8
Note that temperature is not decreasing during Modulation i.e. no cooling
Modulate +/- 0.42 °C every 40 secondsRamp 4.00 °C/min to 290.00 °C
52
54
56
58
60
62
Mod
ulat
ed T
empe
ratu
re (
°C)
52
54
56
58
60
62
Tem
pera
ture
(°C
)
13.0 13.5 14.0 14.5 15.0
Time (min)
9
REQUIREMENTS for DSC Sample sizes for DSC range from 0.1 to 10mg.
Heating rates between 10 and 20 C = min are ⁰common in pharmaceutical studies.
The sample should be finely divided to provide good thermal contact with the block and the thermocouple.
10
INSTRUMENTATION
It consists of…. Sample holder Furnace/heating device Temperature programmer Recording device Atmosphere control Cooling control
11
SAMPLE ASSEMBLY
Two types of sample assembly
1. Design for temperature above 500 0C
2. Design for temperature below 500 0C
12
1. Design for temperature above 500 0C
In this the sample or reference thermocouple heads fit underneath the metal pans containing sample or reference material. It is surrounded by metallic or ceramic block.
This block ensures even heat distribution. Flushing of the sample assembly is by using
a selected gas atmosphere and also cooling facility is applied for the rapid return to ambient temperature.
13
2. Design for temperature below 500 0C
The disc or plate thermocouples are used as platforms for the sample or reference container pans, which have flat bottom.
This gives a high sensitivity for low mass of sample
14
FURNACE
A uniform hot zone. Linear heating rates. Should of low mass. Designed to preclude electrical interaction
effects (Nichrome wire – upto1000 0C, platinum wire - 1500 0C)
Should have cooling facility for rapid return to ambient temperature.
15
PROGRAMMER
It should be capable of giving a wide range of heating and cooling rates, ex: temperature should rise from 1 to 50 0C /min
16
RECORDER SYSTEM
It contains an amplifier for signals from temperature (T) sensor & differential temperature (∆T) thermocouples.
These signals will be recorded on a twin pen recorder.
It should have range of sensitivity settings, so that peak width as well as height may be varied by the operator.
17
18
INTERPRETATIONS
19
There are three methods to measure the peak areaCut and weigh methodPlanimeter methodIntegrated method
20
Cut and weigh method
This is usually done on a photocopy of the record, so that the original may be preserved.
The area may be obtained by comparison with a square cut from the same paper.
The method is accurate to within +/- 5%
21
Planimeter method
This involves the tracing the perimeter of a peak on the DSC record with a stylus connected to a planimeter arm.
The area will be registered on the planimeter dial.
This method eliminates errors like inaccurate cutting & variations from paper quality.
Gives accurate reading within +/- 1%
22
Integrator Method
It consist of an electronic device that automatically measures the area under a peak accurately. Thus, the number of calories or joules corresponding to particular transition.
It is the most accurate method.
23
ADVANTAGES instruments can be used at very high temperatures
• instruments are highly sensitive
• characteristic transition or reaction temperatures can be accurately
determined
DISADVANTAGES• uncertainty of heats of fusion, transition, or reaction estimations is 20-
50%
24
FACTORS AFFECTING DSC
Sample weight Particle size Heating rate Atmospheric conditions Conditions of sample packing into dishes
25
26
APPLICATIONS
Pharmaceutical analysis are in quantitative identification and purity assessment of raw materials. It can be accomplished by comparing the DSC curves of the sample to that of a reference curve.
Studying the compatibility of drugs and excepients in formulation work.
Impurities may be detected by depression of the melting point ,the appearance of transitions, etc.
27
Study thermal stability of inorganic compounds.
DSC offers a wide spectrum of useful investigations related to reaction kinetics, polymerization, solvent retention, phase- transformation, solid-phase reactions and curing or drying properties of a product.
28
SOURCES OF ERRORS• Calibration
• Contamination
• Sample preparation – how sample is loaded into a pan
• Residual solvents and moisture.
• Thermal lag
• Heating/Cooling rates
• Sample mass
• Processing errors
29
REFERENCE
Pharmaceutical Drug Analysis –Ashutosh Kar
Pharmaceutical Analysis (VOL II) – by Dr. A V Kasture
Thermal methods by-James W. Dodd & Kenneth H. Tonge
30
THANK YOU