147
VAPORIZERS DR. AVADHESH PRATAP RESIDENT DEPARTMENT OF ANAESTHESIOLOGY AND CRITICAL CARE N.S.C.B. MEDICAL COLLEGE JABALPUR (M.P.)

Copy of vaporizers

  • Upload
    imran80

  • View
    9.012

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Copy of vaporizers

VAPORIZERS

DR. AVADHESH PRATAP

RESIDENT DEPARTMENT OF ANAESTHESIOLOGY

AND CRITICAL CARE

N.S.C.B. MEDICAL COLLEGE JABALPUR (M.P.)

Page 2: Copy of vaporizers

WHY WE NEED A VAPORIZER

• Most volatile agents exist as liquid at room temperature and atm. pressure .

• Vaporizers convert this liquid form to vapor phase and add a certain amount of this vapor to anaesthetic circuit

• In precisely determined concentration over a wide range of temp, pressure and carrier gas flow rates

Page 3: Copy of vaporizers

PHYSICS RELATED TO VAPORIZERS ?

• What is vapor ?

• What is vapor pressure ?

• What is a vaporizers ?

Page 4: Copy of vaporizers

DEFINITIONS

• Vapor is the gaseous phase of a substance which is normally a liquid at room temp. and atm. pressure. Eg. water vapor is the vapor form of water

• Gas is substance which exists only in the gaseous state at room temp. and amt. pressure.

Page 5: Copy of vaporizers

WHAT IS CRITICAL TEMPERATURE ?

• For any substance there is a max. temp. at which it can be compressed so as to convert it from a gas to a liquid. This is known as the critical temp. and above this temp. no amount of compression will liquefy it.

• Under this condition the substance is a gas.

• Below that critical temp it is a vapor.

Page 6: Copy of vaporizers
Page 7: Copy of vaporizers

WHAT IS VAPOR PRESSURE• This vapor exerts a pressure on its surroundings –

Vapor pressure .

• Vapor pressure depends only on the temp. and liquid.

• Vapor pressure of an agent determines how much of vapor will be formed from 1 ml of the liquid.

• Since different anaesthestic agents have different vapor pressures the need separate vaporizers.

Page 8: Copy of vaporizers

SATURATED VAPOR PRESSURE ?

Page 9: Copy of vaporizers

• For a particular liquid at a particular temp. there occurs an equilibrium at which the number of molecules leaving the liquid equals the number reentering

• It is the maximum VP at a particular temp. Increases rapidly as boiling point approaches.

Page 10: Copy of vaporizers

BOILING POINT ?

Page 11: Copy of vaporizers

• Go on increasing the temp.

• The V.P. will increase.

• At a point it equals the atm pressure

• B.P. OF A LIQUID IS THE TEMPERATURE AT WHICH ITS VAPOR PRESSURE IS EQUAL TO THE ATM PRESSURE

• Low atm pressure - > Low B.P.

• High SVP -> Low B.P. (Means very volatile)

Page 12: Copy of vaporizers

CONCENTRATION OF VAPOR

May be expressed as

VOLUME %

OR

PARTIAL PRESSURE

Page 13: Copy of vaporizers

PARTIAL PRESSURE• A mixture of gases in a closed container will exert a

pressure on the walls of the container. The part of the pressure exerted by any one gas in the mixture is called its partial pressure

• Depends only on the temperature and nature of liquid

• The highest partial pressure that can be exerted by a gas at a given temperature is its vapor pressure it is an absolute value.

• UPTAKE & DEPT. OF ANAE. α PP.

Page 14: Copy of vaporizers

VOLUMES PERCENT

• No of units of volume of a gas in relation to a total of 100 units of volume for the total gas mixture

• It expresses the relative ratio of gas molecules in a mixture.

• Most commonly used.PARTIAL PRESSURE/TOTAL PRESSURE

= VOLUMES PERCENT/100

Page 15: Copy of vaporizers

LATENT HEAT OF VAPORIZATION ?

• The number of calories required to convert 1gm of liquid into a vapor or number of calories required to convert 1 ml of liquid into a vapor.

• In a vaporizer as the liquid agent vaporizes -> heat is lost -> temp. drops -> V.P. drops -> less volume of agent is available for the carrier gas to take away -> decreases output

Page 16: Copy of vaporizers

SPECIFIC HEAT ?• It is the amount of heat required to raise the temp of

1gm or 1ml of a substance through 1 degree 0C • Significance• LIQUID ANAESTHETIC AGENT – Should have a

low specific heat so as to facilitate vaporization • VAPORISER CONSTRUCTION MATERIAL-

Should have high specific heat; acts as a heat sink; provides a more stable temp.

Page 17: Copy of vaporizers

THERMAL CAPACITY

SPECIFIC HEAT x MASS

• Amount of heat stored in the vaporizer body

• A vaporizer constructed from a substance

with high thermal capacity -> temp. changes

more slowly and so preferred

Page 18: Copy of vaporizers

THEMAL CONDUCTIVITY• Speed with which heat flows through a substance; Cu >

Al> brass> Steel >> Glass. • Higher the thermal conductivity, better the substance

conducts heat.

Significance• Vap. Constructor material should be able to conduct heat

from surroundings to contained liquid.• Cu has a moderate sp. Heat and high thermal

conductivity – use for construction of vaporizers• More recently use bronze stainless steel

Page 19: Copy of vaporizers

THERMOSTABILIZATION ?

• Utilization of some means to minimize temp. changes.

• Construct vaporizers of materials with high thermal

conductivity and specific heat to minimize temp. changes

when in use.

• Heavy metal parts act as a heat reservoir.

• Wicks to be in contact with the metal part so that heat loss

due to vaporization is quickly replaced.

Page 20: Copy of vaporizers

GAS LAWS

• Boyles law: V α 1/p (at constant temperature )• Charle’s law: V α T (at constant pressure )• DALTON’S LAW OF PARTIAL PRESSURES

• In a mixture of gases the pressure exerted by each gas is the same as that which it would exert if it alone occupied the container

Page 21: Copy of vaporizers

AVOGADRO’S HYPOTHESIS

• Equal volumes of gases at the same temperature and pressure contain equal number of molecules.

• This no. is 6.023 x 1023

• This much no. of particles of any gas at STP will occupy 22.4 Ls

Page 22: Copy of vaporizers

WHAT IS MAC ?

• Typically expressed as vol. % or alveolar gas at 1 atm Eg. MAC of

halothane - 0.75

enflurane - 1.68

isoflurane - 1.2

desflurane - 6.0

sevoflurane - 2.0

Page 23: Copy of vaporizers

WHAT IS MAPP ?• Minimum alveolar partial pressure ( MAPP)

• Expresses MAC in terms of PP (Pmac1)

• MAC of hal is 0.75

pp of hal for 1 MAC is

0.75/100x760 = 5.7 mm Hg

• Pmac1 of des for is

6/100x760 = 45.6 mm Hg

Page 24: Copy of vaporizers

WHAT IS A VAPORIZER ?• A vaporizer is an instrument designed to facilitate

the change of a liquid anaesthetic agent into a vapor

And

Add a controlled amount of this vapor to the FGF• The SVP of most inhalation agents is MUCH more

that is required to produce anesthesia i.e. 32% vs 0.75 or 243 mm Hg vs 5.7 mm Hg for halothane

• Need to dilute this vapor with the carrier gas and deliver a controlled amount of this vapor to the patient

Page 25: Copy of vaporizers

• What amount of vapor produced by 1 gram of its

liquid agent ?• 1 mole of a gas weighs 1 gram molecular weight. • The molecular weight of isoflurane is 185 • Therefore, 1 mole of isoflurane vapor weighs 185 g • As per avogadro’s hypothesis 1 mole of a substance =

22.4 liters of gas • 22.4 L of isoflurane vapour weighs 185 g

Page 26: Copy of vaporizers

density of isoflurane vapor is 185/22.4 = 8.25 g/litre

1 gram = 1000/8.25 ml of vapor• Density of lsoflurane liquid is 1.5 g/ml 1 gram = 1/1.5 ml of liquid

• 1/1.5 ml of liquid = 1000/8.25 ml of vapor

• 1 ml of liquid = 1000/8.25 x 1.5 = 180 ml.

For most modern agents,

• 1 ml of liquid volatile agent yields about 200 ml of vapor

Page 27: Copy of vaporizers

• How much liquid agent does a vaporizer use per hour ?

• The amount of vapor used per min. =

FGF x time x concentration setting

For e.g., to deliver 1% isoflurane at FGF of 2 litres per min for 60 mins. One would need 1/100 x 2000 ml x 60 min = 1200 ml of isoflurane vapor

This woul correspon to 1200/180 = 6.7 ml of liquid isoflurane

Another formula which can be used with most agents is

3 x Fresh gas glow (FGF) (L/min)x volume % ml liquid used per hour

Page 28: Copy of vaporizers

CALSSIFICATION OF VAPORIZERS

Page 29: Copy of vaporizers

METHOD OF REGULATING THE OUTPUT CONCENTRATION

• CONCENTRATION CALIBRATED

• MEASURED FLOW

Page 30: Copy of vaporizers

METHOD OF VAPORIZATION

• FLOW OVER

• BUBBLE THROUGH

• INJECTION

• SIMPLE

Page 31: Copy of vaporizers

• THERMOCOMPENSATION

• SUPPLIED HEAT

TEMPERATURE COMPENSATION

Page 32: Copy of vaporizers

SPECIFICITY

• AGENT SPECIFIC

• MULTIPLE AGENTS

Page 33: Copy of vaporizers

POSITION

• VAPORIZER INSIDE CIRCUIT

• VAPORIZER OUTSIDE CIRCUIT

Page 34: Copy of vaporizers

RESISTANCE

• PLENUM

• LOW RESISTANCE/DRAW OVER

Page 35: Copy of vaporizers

BASIC DESIGN

Page 36: Copy of vaporizers

METHOD OF REGULATING OUTPUT CONSENTRATION CONCENTRATION CALIBRATED

• Total flow form the machine passes through the vaporizer. This is split by a variable resistance proportionationg valve into two:

• One part (usually major) flows through the bypass chamber & the other (usually small) through the vaporizing chamber.

Page 37: Copy of vaporizers

CONCENTRATION CALIBRATED

• AGENT CONCENTRATION IS CONTROLLED BY A DIAL CALIBRATED IN VOLUMES PERCENT

• Depends on resistance of the 2 pathways (controlled by conc. Control dial & thermocomp. vol.)

• Volume vapourised – typically 200 ml vapour per ml of liquid Anaesthetic

Page 38: Copy of vaporizers

IF WE DIAL A HIGH CONCENTRATION

Page 39: Copy of vaporizers

IF WE DIAL A LOW CONCENTRATION

Page 40: Copy of vaporizers

SPLITTING RATIO• The ratio of the bypass gas to the gas going to

the vaporizing chamber is called the SPLITTING RATIO. Splitting ratio depends on:

The resistance of the two pathways, which inturn depends on the variable orifice of the inlet/outlet.

Temperature of the liquid/carrier gas.Flow rate of gases

Page 41: Copy of vaporizers

MEASURED FLOW VAPORIZERS

• The vaporizer heats the anesthetic agent to a temperature above its boiling point (so it behaves as gas) and this is then metered into the fresh gas flow.

• A measured flow is sent by a separate oxygen flow meter to pass to the vaporizer with the output being at SVP for the anesthetic agent.

• In order to dilute this otherwise lethal concentration, outpur from that flowmeter is combined with gas passing form the main flowmeter

Page 42: Copy of vaporizers

• Operator has to set the flow to the vaporizer and bypass with separate flowmeters

• This means that respective flows have to be calculated for each agent for a given temp and vapour output

• To calculate the vaporizer output, one must know the -Vapor pressure of the agent

- The atmospheric pressure

- The total flow of gases

- The flow of the vaporizer

MEASURED FLOW VAPORIZERS

Page 43: Copy of vaporizers

CALCULATIONS FOR OUTPUT IN MEASURED FOW VAPORISERS

• Set 100 ml/min flow of carrier gas (oxygen) from dedicated flowmeter

• SVP of hal in vap. Chamber is 243 mmHg

• Hal forms 243/760 x 100 = 32% of gas mixture

• Carrier gas will occupy the rest of the vol. i.e. 100-32=68%

• This 68% is occupied by 100ml/min carrier gas

• And 32% hal will be = 100/68x32=47 ml

Page 44: Copy of vaporizers

• Gas exiting is 147 ml with 47 ml hal vapor

• To get a mixture containing 1% hal this 47 ml should be diluted in 4700 ml

• Required carrier gas is 4700-147=4553 ml

• If set 100 ml measured flow to vaporiser usually set 5 l/min flow of carrier gas to get 1% halothane

• Ratio of gas through vaporiser: main gas flow is 100:4600=1:46

• % concentration of agent = 100 x vaporizer output of anaesthetic/total flow

CALCULATIONS FOR OUTPUT IN MEASURED FOW VAPORISERS

Page 45: Copy of vaporizers

VAPORIZATION METHODS

FLOW OVER

• Here the carrier gas passes over the surface of a liquid

• Surface area of vaporization can be increased by using wicks.

Page 46: Copy of vaporizers

FLOW OVER

Page 47: Copy of vaporizers

• Due to capillary action, the anaesthetic agent rises into the wicks. This dramatically increases the surface area of anaesthetic agent exposed to the fresh gas and thereby improves the efficiency of vaporization.

• Carrier gas can be directed using baffles or spiral tracks that lengthen the gas pathways over the liquid. This increases the time and area of contact.

FLOW OVER

Page 48: Copy of vaporizers

BUBBLE THROUGH

Page 49: Copy of vaporizers

• Carrier gas is allowed to bubble through the liquid.

• Smaller the bubble, larger will be the surface area.• Methods to break the gas into small bubbles are

agitation & splashing.• Factors influencing the vaporizer output:- Size of the bubble, - Depth of liquid, - Bubbling speed (Velocity of carrier gas flow)

BUBBLE THROUGH

Page 50: Copy of vaporizers

INJECTION VAPORIZERS

• A known amount of liquid agent or pure

vapor is injected into the gas stream to

provide the desired concentration

• Eg. TEC 6 desflurane vaporizer (it is

more ideally classified as a gas-vapor

blender)

Page 51: Copy of vaporizers

TEMPERATURE COMPENSATION

• To maintain a constant output from the vaporizer, mechanisms to compensate for the fluctuations in temperature are to be employed

• Alteration in the splitting ratio (automatic compensation Eg. Bimetallic strip in tec vaporizers, ether filled bellows in penlon vaporizers, EMO

• Computer control – electronic vaporizers • Supplied heat – tec 6 (electrically heated)

Page 52: Copy of vaporizers

TEMPERATURE COMPENSATION

Page 53: Copy of vaporizers

THERMOCOMPENSATION

• Most variable bypass vaporizers compensate for changes in vapor pressure by altering the splitting ratio.

• Done by using a thermosensitive element (BIMETALLIC STRIP) incorporated in the vaporizing chamber or bypass chamber.

• So, the splitting of gas is controlled by 2 valves: • (1) the dial we set (splitting valve)• (2) the temperature compensating valve.

Page 54: Copy of vaporizers

THERMOCOMPENSATION

• In a bimetallic strip, two metals with very different coefficients of thermal expansion are fixed together.

Page 55: Copy of vaporizers

THERMOCOPENSATION • When the temp. of the vaporizing chamber

drops, the bimetallic strip bends and moves away. This reduces the resistance to flow and thus more flow occurs into the vaporizing chamber

Page 56: Copy of vaporizers

SUPPLIED HEAT• An electric heater can be used to

supply heat to a vaporizer and maintain it at a constant temperature

• Eg. Tec 6

Page 57: Copy of vaporizers

RESISTANCE

PLENUM ( Latin = fullness)

• Vaporizers with high resistance which depend on compressed gas driven under pressure are called plenum vaporizers.

• Eg. Boyle bottle, copper kettle, TEC vaporizers

Page 58: Copy of vaporizers

DRAW OVER

• Carrier gas is drawn through the vaporizer either by the patient’s own respiratory efforts, or by a self-inflating bag or manual bellows

• Drawover systems operate at less than, or at ambient pressure

• Flow through the system is intermittent, varying with different phases of inspiration, and ceasing in expiration.

RESISTANCE

Page 59: Copy of vaporizers

DRAW OVER• Have a low internal resistance to gas flow

So that they may be used within the breathing circuit, the gas flow being driven through them by the patient’s breathing.

• They may be used in a non rebreathing DROW-OVER APPARTUS, or as IN-CIRCUIT VAPORZERS in a CIRCLE ABSORBER SYSTEM.

• Eg. Goldman bottle, EMO

Page 60: Copy of vaporizers

FACTORS AFFECTING VAPORIZER OUTPUT

• 1. FLOW THROUGH THE VAPORIZING CHAMBER: Varying the proportion of gas passing through the vaporizing chamber and bypass chamber

• 2. SURFACE AREA OF THE LIQUID GAS INTERFACE: Greater the surface area, more will be the vaporization.

• Bubble through > flow over

Page 61: Copy of vaporizers

FACTORS AFFECTING VAPORIZER OUTPUT

• 3. TEMPERATURE: as temperature increase, output increase

• 4. TIME: Output concentration tend to fall over time

• 5. GAS FLOW RATE: (A) At high flowrates, the gas leaving vaporization chamber is less saturated

(B) Alters the total flow that passes through the vaporization chamber

Page 62: Copy of vaporizers

• 6. CARRIER GAS COMPOSITION:

(a) Changes in viscosity & density may affect the proportion of the total flow passing through the vaporization chamber

(b) N2O dissolves in the flow, thus altering the effective volume passing through the vaporization chamber

• 7. BOILING POINT: Higher the boiling point, less will be the vapor output

FACTORS AFFECTING VAPORIZER OUTPUT

Page 63: Copy of vaporizers

• 8. AMBIENT PRESSURE:

- SVP is solely a function of temp. so if ambient pressure is reduced, the constant SVP becomes a greater proportion of the total pressure

-> output increases.

- Agents with low boiling points are more susceptible to the influence of ambient pressure

FACTORS AFFECTING VAPORIZER OUTPUT

Page 64: Copy of vaporizers

EFFECT OF LOW ATMOSPHERIC PRESSURE

CONCENTRATION CALIBRATED

• High resistance pathway through the vaporizing chamber offers less resistance, under hypobaric conditions and so a slight increase in vapor output occurs.

• Deliver higher conc. if measured in vol. % but deliver same PP so clinically affect unchanged

CP =CP or C=CP/P At 0.5 atm. C=Cx1/.5=2%

Page 65: Copy of vaporizers

EFFECT OF LOW ATMOSPHERIC PRESSURE

MEASURED FLOW

• Here the delivered partial pressure & volume percent increases.

• Amount of increase depends on the barometric pressure & the vapor pressure of the agent.

• Closer the vapor pressure is to barometric pressure, greater the effect.

Page 66: Copy of vaporizers

CONCENTRATION CALIBRATED

• ATM PRESSURE

INCREASES -> Density of gas

CHANGES -> More resistance to flow of gas through the vaporizing chamber

-> Decreased vapor O/P (Partial Pressure & Volume Percent)

Effect on partial pressure is less dramatic

EFFECT OF HIGH ATMOSPHERIC PRESSURE

Page 67: Copy of vaporizers

MEASURED FLOW

• Lower concentration in terms of PP/Volume percent

EFFECT OF HIGH ATMOSPHERIC PRESSURE

Page 68: Copy of vaporizers

EFFECT OF INTERMITTENT BACK PRESSURE

• When assisted or controlled ventilation is used, the positive pressure generated during inspiration is transmitted from the breathing system back to the machine & some way may be transmitted to the vaporizers.

• Also seen with the use of oxygen flush

Page 69: Copy of vaporizers

PUMPING EFFECT

• The increase in vaporizer output concentration due to back pressure

Page 70: Copy of vaporizers

PUMPING EFFECT• When the bag is squeezed, pressure is transmitted

back into both, the “by pass” channel and also to the vaporizing chamber. The fresh gas tries to move forward and gets compressed both in the ‘by pass’ channel and the vaporizing chamber. However, the vaporizing chamber volume is much larger than the ‘by pass’ channel volume, an thus, more fresh gas gets compressed into it than into

the ‘by pass’ channel.

Page 71: Copy of vaporizers

PUMPING EFFECT

• This extra fresh gas that enters the vaporizing chamber collects anaesthetic vapor

Page 72: Copy of vaporizers

PUMPING EFFECT• When the positive pressure is suddenly released

(expiration) the previously compressed gases now suddenly expands in all direction.

• Some of the rapidly expanding gas (containing vapor) enter the inlet of the vaporizer and cross over into the ‘by pass’ channel

Page 73: Copy of vaporizers

PUMPING EFFECT• Normally, a vaporizer ‘by pass’ channel does not

have vapor. So his vapor due the pumping effect’ is additional. When this ‘by pass’ vapor flows across to the exit, it meets the vapor from the vaporizing chamber. The addition of the ‘by pass’ vapor to the vapor from the vaporizing chamber raises the final concentration of anaesthetic delivered. i.e. the ‘pumping effect’ increases the delivered concentration of anaesthetic agent.

Page 74: Copy of vaporizers

PUMPING EFFECT

SEEN ESPECIALLY WHEN

-> CARRIER GAS IS LOW

-> AGENT IN VAPORIZING CHAMBER IS LOW

-> DIAL SETTING IS LOW

-> PRESSURE FLUCTUATIONS ARE HIGH & FREQUENT.

Page 75: Copy of vaporizers

PUMPING EFFECT• MECHANISM IN MEASURED FLOW:

Gas flow to these vaporizers become saturated with vapor & is joined by gas from other flowmeter, which dilutes its concentration.

When back pressure is applied, there is retrograde flow of gas so that the diluted gas mixture is forced back into the vaporizer, because this gas is not saturated, it will then pick up anesthetic vapor, the final concentration will be higher

Page 76: Copy of vaporizers

MODIFICATIONS TO MINIMIZE PUMPING EFFECT

• Keep the vaporizing chamber small

Or

Increasing the size of the bypass chamber.

Page 77: Copy of vaporizers

MODIFICATIONS TO MINIMIZE PUMPING EFFECT

• Add long spiral or large diameter tube to lead to the vaporizer chamber. The extra gas forced into this tube & subsequently returned to the bypass does not reach the vaporizing chamber

Page 78: Copy of vaporizers

• Increase resistance to gas flow through the vaporizer

MODIFICATIONS TO MINIMIZE PUMPING EFFECT

Page 79: Copy of vaporizers

• Check valve to prevent backward flow of gas

MODIFICATIONS TO MINIMIZE PUMPING EFFECT

Page 80: Copy of vaporizers

• Exclude wicks from the area where the inlet tube joins the vaporizing chamber.

• Outlet tube may be made longer so that further back before picking up anesthetic vapor.

• Connections of oxygen flush valve line to the common gas outlet be designed to minimize pressure fluctuations that may produce a pumping effect

• Limit pressure transmitted to vaporizer to <10KPa above normal working pressure, conc not to increase > 20%

MODIFICATIONS TO MINIMIZE PUMPING EFFECT

Page 81: Copy of vaporizers

PRESSURIZING EFFECT

• The output of some vaporizer used in conjunction with automatic ventilator has been found to be lower than during free flow to atmosphere.

Mostly seen when

-> High flow

-> Large pressure fluctuations

-> Low dial settings

Page 82: Copy of vaporizers

PRESSURIZING EFFECT

Page 83: Copy of vaporizers

• No of mol.s of the agent picked by carrier gas α density of vapor mol.s in vaporizing chamber α vapor pressure

• Increased pressure in vaporizer chamber compress carrier gas -> more mol.s/mL-> no of mols of the anesthetic agent same ->o/p

• The changes in vaporizer O/P caused by the pumping effect usually are greater in magnitude that those associated with the pressurizing effect

PRESSURIZING EFFECT

Page 84: Copy of vaporizers

1988 ASTM MACHINE STANDARDS FOR VAPORIZERS

1. The effects of variations in ambient temperature and pressure, tilting, back pressure, and input flow rate and gas mixture composition on vaporizer performance must be stated in the accompanying documents.

2. The average delivered concentration from the vaporizer shall not deviated from the set value by more than ± 20% or ± 5% of the maximum setting whichever is greater without back pressure.

Page 85: Copy of vaporizers

3. The average delivered concentration from the vaporizer shall not deviate from the set value by more that + 30% or – 20% or by more than + 7.5% or – 5% of the max. setting whichever is greater, with pressure fluctuations at the common gas outlet of 2 kPa with a total gas flow of 2 L/minute or 5 kPa with a total gas flow of 8 L/minute

4. A system that prevents gas from passing through the vaporizing chamber or reservoir of one vaporizer and the through that of another must be provided

5. The output of the vaporizer shall be less than 0.05% in the “OFF” or “zero” position if the zero position is also the “OFF” position.

Page 86: Copy of vaporizers

6. All vaporizer control knobs must open counterclockwise

7. Either the max. and min. filling levels or the actual usable volume and capacity shall be displayed

8. The vaporizer must be designed so that it cannot be overfilled when in the normal operating position.

9. Vaporizer unsuitable for use in the breathing system must have noninterchangeable proprietary or 23 mm fittings. Conical fittings of 15 mm and 22 mm cannot be used. When 23 mm fittings are used the inlet must be male and the outlet female. The direction of gas flow must be marked.

Page 87: Copy of vaporizers

10. Vaporizers suitable for use in the breathing system must have standard 22-mm fittings or screwthreaded, weight-bearing fittings with the inlet female and the outlet male. The direction of gas flow must be indicated by arrows and the vaporizer marked “for use in the breathing system.

Page 88: Copy of vaporizers

AGENT SPECIFIC FILLING SYSTEMS

Page 89: Copy of vaporizers

• A vaporizer designed for a single agent be fitted with a permanently attached agent specific device to prevent accidental filling with a wrong agent.

• They prevent accidental filling with the wrong agent.

• Reduce air pollution.

• Vaporization chamber

TYPES

KEYED FILLING SYSTEM

SCREW CAPPED FILLING SYSTEM

PIN SAFETY SYSTEM

AGENT SPECIFIC FILLING SYSTEMS

Page 90: Copy of vaporizers

• BOTTLE COLLAR• Attached at the neck of the bottle.• 2 projections, one thicker than the other are there. • This mates with the corresponding indentations on

the bottle adaptor

AGENT SPECIFIC FILLING SYSTEMS

Page 91: Copy of vaporizers

AGENT SPECIFIC FILLING SYSTEMS

Page 92: Copy of vaporizers

• BOTTLE ADAPTOR

AT ONE END IS

1. The bottle connector with a screw thread to match the thread on the bottle

2. Skirt that extends beyond the screw threads

3. Slots that match the projections on the bottle collar

AGENT SPECIFIC FILLING SYSTEMS

Page 93: Copy of vaporizers

AT THE OTHER END IS

1. The male adaptor that fits into the vaporizer filler receptacle.

2. A short length of plastic tubing with 2 inner tubes connects the ends.

3. The tube allows the bottle to be held higher or lower than the vaporizer.

AGENT SPECIFIC FILLING SYSTEMS

Page 94: Copy of vaporizers

• MALE ADAPTER

• Has a groove on one side to prevent the probe from being placed in the incorrect vaporizer

• 2 holes on the other side

• LARGER → for agent to leave/enter

• SMALLER → for air

AGENT SPECIFIC FILLING SYSTEMS

Page 95: Copy of vaporizers

• FILLER RECEPTACLE

(Filler socket/vaporizer filler unit/fill & drain system)

• Must permit the insertion of the intended bottle

• Must have a means of tightening the male adapter, to from a tight seal.

• Must have a means of sealing (plug) the adaptor when bottle adaptor not inserted.

AGENT SPECIFIC FILLING SYSTEMS

Page 96: Copy of vaporizers

• FILLER RECEPTACLE• Single port for filling & drainage.• A valve attached to a knob at the top

controls the opening into the vaporizer • A ball valve in the airline occludes the air

port after the vap is filled to prevent overfilling & flooding of airline with the agent.

AGENT SPECIFIC FILLING SYSTEMS

Page 97: Copy of vaporizers

• FILLING• Cap of bottle removed

• Adapter screwed to the collar

AGENT SPECIFIC FILLING SYSTEMS

Page 98: Copy of vaporizers

• FILLING• Vaporizer turned off• Plug removed• Bottle with adapter inserted such that the grooves

match; (tube bent such that bottle is below inlet)

AGENT SPECIFIC FILLING SYSTEMS

Page 99: Copy of vaporizers

• FILLING

• Retaining screw tight

• Fill valve (vent) opened

• Bottle held high

• Air from the vaporizer displaced by the liquid moves through the other tube and enters the air space inside the bottle

• Gentle up and down motion may help

AGENT SPECIFIC FILLING SYSTEMS

Page 100: Copy of vaporizers

AGENT SPECIFIC FILLING SYSTEMS

Page 101: Copy of vaporizers

• DRAINING• Adapter attached to bottle → filler plug removed

→ male adaptor inserted → retaining screw tightened → bottle held below the receptacle → drain valve opened → drained → screw loosened → adaptor removed → filler plug inserted

AGENT SPECIFIC FILLING SYSTEMS

Page 102: Copy of vaporizers

1. Difficulty in filling

2. Misalignment of adaptor in filler receptacle

3. Adaptor not sealing at the bottle end

4. Leak in the bottle adaptor

5. Air bubbles.

6. Lost bottle adaptor

7. Failure of keyed system

8. Liquid leaks

AGENT SPECIFIC FILLING SYSTEMS PROBLEMS

Page 103: Copy of vaporizers

10. VAPORIZER TIPPING

If filler receptacle on vaporizer extends beyond the base → it cannot be set upright on a flat

surface → this can cause tipping prevented by setting the vaporizer receptacle at the edge of the

surface or a ring can be fitted at the base that extends below the projection of the filler reeptacle

11.POOR DRAINAGE

12.BROKEN INNER TUBE

AGENT SPECIFIC FILLING SYSTEMS PROBLEMS

Page 104: Copy of vaporizers

LOCATION

• Between the flowmenters & common gas outlet

• Between the common gas outlet & breathing system

• In-system vaporizers

Page 105: Copy of vaporizers

HAZARDS

• INCORRECT AGENT

• Low output or high output

• Misplacing desflurane dangerous

• Gas allowed to flow through it until no agent detected in the outflow

Page 106: Copy of vaporizers

• TIPPING

• Liquid from vaporizing

Chamber→bypass/outlet→high output

• Drained before moving

• OVERFILLING

• In majority, the design of the filling port and agent specific filling systems prevent this

• During filling do not:

Turn the dail ON/Unscrew the bottle adaptor

HAZARDS

Page 107: Copy of vaporizers

• REVERSED FLOW

• Inlet male & outlet female

• Increased output

• CONCENTRATION DIAL IN WRONG POSITION

• CONTAMINATS IN VAPORIZING CAHMBER

• PHYSICAL DAMAGE

• OBSTRUCTION TO FRESH GAS FLOW

• INTERLOCK MALFUNCTION

HAZARDS

Page 108: Copy of vaporizers

INTERLOCK DEVICES

• Applied to control dial of vaporizer so that only one vaporizer can be turned on at a time

• Ensure that

- Only 1 vap. is turned on at a time

- Gas only enters that which is ON at that time

- Trace vapor output is minimized when OFF

- Vap.s are locked onto gas circuit hence correctly seated

Page 109: Copy of vaporizers

INTERLOCK DEVICES

• Selector valve

• Selectatec system

• Back bar devices

• Ohio switch

• Drager lock

Page 110: Copy of vaporizers

ORDER OF VAPORIZERS

• Less potent - upstream

• More potent - downstream

• If equipotent

Low VP - upstream

High VP - downstream

• Also

If explosive - downstream

Trilene - downstream

Easy to clean - downstream

Page 111: Copy of vaporizers

ORDER OF VAPORIZERS

UP STREAM → → → DOWNSTREAM

SEVOFLURANE ENFLURANE ISOFLURANE HALOTHANE DESFLURANE

VP- 157 175 238 243 669

Page 112: Copy of vaporizers

EVOLUTION OF VAPORIZERS

• MORTONS ETHER INHALER

1846, OCT 16

Page 113: Copy of vaporizers

CLOVER PORTABLE REGULATING ETHER INHALER 1877

Page 114: Copy of vaporizers

VERNON HARCOURT CHLOROFORM INHALER 1903

Page 115: Copy of vaporizers

OMBREDANNE ETHER INHALER 1908

Page 116: Copy of vaporizers

OPEN DROP MASK

Page 117: Copy of vaporizers

OPEN DROP MASK

Page 118: Copy of vaporizers

OGSTON INHALER

Page 119: Copy of vaporizers

BOYLES BOTTLE• Parts: (1) vaporizing bottle 300 mL (2)

Metal top incorporating controls (3) Lever, plunger which is chrome plated (4) Stopper & Retaining chain

• Concentration calibrated

• Flowover or bubble through

• Not temperature compensated

• Multiple agents

• Vaporizer outside circuit

Page 120: Copy of vaporizers

OXFORD MINIATURE VAPORIZER (OMV)

• Variables bypass

• Flowover with steel wicks

• Temperature compensated

• For use by multiple agents

• Vaporizer outside circuit

• Halothane trilene methoxyflura ne ether isoflurane

Page 121: Copy of vaporizers

EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 (EMO)

Page 122: Copy of vaporizers

EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 (EMO)

Page 123: Copy of vaporizers

EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 (EMO)

Page 124: Copy of vaporizers

EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 (EMO)EMO-INTERIOR PARTS

Page 125: Copy of vaporizers

EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 (EMO)

Page 126: Copy of vaporizers

GOLDMAN VAPORIZER

Page 127: Copy of vaporizers

GOLDMAN VAPORIZER

Page 128: Copy of vaporizers

ROWBOTHOM VAPORIZER• Modification of Goldman vaporizer

Page 129: Copy of vaporizers

FOREGGER COPPER KETTLE 1952

Page 130: Copy of vaporizers

FOREGGER COPPER KETTLE 1952

Page 131: Copy of vaporizers

TEC VAPORIZERS• CLASIFICATION (TEC I-V)

1.Variable bypass

2.Flow over with wick

3.Out of system

4.Temp. compensated by automatic flow alteration

5.Conc. calibrated.

6.Agent specific

Page 132: Copy of vaporizers

TEC 5 • CONSTRUCTION• Top control dial• Locking lever and release button• Sight glass-bottom right• Keyed filling device- Filling draining port- Locking lever to secure filler block- Small lever at base allows liquid to be added or drained

- Screw cap with drain plug

Page 133: Copy of vaporizers

TEC 5

Page 134: Copy of vaporizers

TEC 5

Page 135: Copy of vaporizers

TEC 5

• INTERNAL BAFFLE SYSTEMS

• EVALUATION

- Greater accuracy at FGF of 5L/min and dial <3%

- More pumping effect than tec 4

- Accuracy max. 15-35 degree Celsius

Page 136: Copy of vaporizers

TEC 5

• INTERNAL STRUCTURE

• Internal baffle system

- VC lies within the bypass which lies along side of the vap.

- Bimetallic strip in the base

- Before reaching VC - helical IPPV assembly - spiral wick

Page 137: Copy of vaporizers

TEC 5

Page 138: Copy of vaporizers

TEC 5 • EVALUATION• Accurate at FGF 5Lpm & dial settings <3%• Improved key filler• Easier mech - to switch on rotary valve & lock with

one hand• HAZARDS- Pumping effect > tec 4- Large liquid loss if filling part is opened- Overfilling – bottle adaptor loose, vaporizer ON- Reverse flow increases output

Page 139: Copy of vaporizers

TEC 6 • CLASSIFICATION• Conc calibrated• Injection• Thermocompensation

by supplied heat

Or

Electrically heated dual

circuit gas-vapor blender

Page 140: Copy of vaporizers

TEC 6• INT. STRUCTURE

• Electronically powered and controlled

• Des heated to 390 by to heater in the base

• VP 1300mmHg in sump

• FGF restricted at fixed orifice, sensed by differential pressure transducer and adjusts resistor R1(differential pressure transducer)

• Control dial adjusts 2nd resistor R2 (adjusted by control dial)

Page 141: Copy of vaporizers

TEC 6• EFFECT OF BAROMETRIC PRESSURE• Works at absolute pressure-It maintains a constant

output in terms of vol% but pp varies if atm pr decreases-output in pp is also decreased

• Reqd. dial setting=dial setting x 760/ambient pressure• Effect of carrier gas; addition of N2O – less viscosity

– decrease vapor output • Mounting – for rt - side of machine• Bottle; has a spring valve to prevent escape of agent

Page 142: Copy of vaporizers

TEC 7• Accurate through clinical flow range• Easy turning dial to allow left and right hand operation • Smaller graduation for accuracy• 3 filling options - Datex ohmeda easy fil - Funnel fil- Easy fil (for sevolurane)• Non spill system• Prismatic sight glass

Page 143: Copy of vaporizers

ALADIN VAPORIZER

• ALADIN CASSETTE

• Classification

- Conc calibrated

- Flowover

- Automatic thermocompensation

- Use with datex – Engstrom AS/3 ADU

Page 144: Copy of vaporizers

ALADIN VAPORIZER

Page 145: Copy of vaporizers

ALADIN VAPORIZER

Page 146: Copy of vaporizers

ALADIN VAPORIZER

Page 147: Copy of vaporizers

THANK YOU…