STARX32: A Complete On-Chip X-Ray Spectroscopy Readout System
with Imaging Capability
Presented by P. Bastia on behalf of the STARX32 team
All rights reserved, 2010, Thales Alenia Space
Page 2
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
CONTEXT
A Industry-University team formed in Italy to search a solution to this technological challenge
In the last decade the Detector technology became mat ure for realization oflarge 2D X-ray detector array in GaAs, CdZnTe and simil ar materials
ORIGIN OF THE PROJECT
To use these arrays in a space instrument a Read Out Integ rated Circuit ismandatory, this ROIC should be low-power, low-noise, h i-speed, highlyintegrated … and in general subjected to all limitation of space environment.
No device with these characteristics existed around 2002 when ESA started a dedicated technological program “Large Format Detector Readout ”
All rights reserved, 2010, Thales Alenia Space
Page 3
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
CONTEXTCONTEXT
TEAM ORGANIZATION
OVERALLPROJECT MANAGEMENT
FRONT-END SECTION DESIGN
ADC & READOUT LOGIC DESIGN
TEST EQUIPMENT DESIGN & MFG
All rights reserved, 2010, Thales Alenia Space
Page 4
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
CONTEXTMILANO plant, formerly:
More than 30 year experience in development of space instruments for x-rays, gamma-rays, cosmic-rays, and scientific payload in general.
IBIS GAMMA RAY IMAGER
AGILEGAMMA RAYOBSERVATORY
ALTEA COSMIC RAY DETECTOR
PLANCK LFIHIGHLY INTEGRATED READOUT ELECTRONICS
All rights reserved, 2010, Thales Alenia Space
Page 5
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
CONTEXT
Politecnico di Milano, Department of Electronics Engi neeringand Information Science
• particular skills in designing ultra low noise front end for X-ray detectors (more than 100 papers on the subject)
University of Pavia, Integrated Microsystem Laborator y
• member EUROPRACTICE
• large experience in mixed-signal ASIC development
UNIVERSITY TEAMS
All rights reserved, 2010, Thales Alenia Space
Page 6
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
REQUIREMENTS REVIEW
All rights reserved, 2010, Thales Alenia Space
Page 7
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
REQUIREMENTS REVIEW
≈ 500 µW/channelPower
consumption
~106 cps total rate capability> 10 kcps/channelEvent rate
system-on-a-chip including all readout logic
and ADC
on-chip≥ 9 bit resolution
ADC
1-10µs selectableShaping peak
time
with Cdet < 400fF and Idet< 2 pA
detector connected
< 30 electrons r.m.s.
ENC
0.5 – 50 keV in GaAs120 – 12000
electronsInput charge
range
bump-bonding required between detector and ASIC
Square array of channels
(300µm pitch)Layout
1024:1 mux-ing required32 x 32Number of channels
RemarksValueParameter
A SET OF CHALLENGING REQUIREMENTS
All rights reserved, 2010, Thales Alenia Space
Page 8
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
REQUIREMENTS REVIEW
2D array layout Bump-Bonding needed
Pixel-Individual analog modules
low ENCLow-noise analog section
control of internal interference
on-chip ADCcomplex conversionmanagement logic
low powerconsumption
300 x 300 µm² cells
high sensitivity !
limited design resources
All rights reserved, 2010, Thales Alenia Space
Page 9
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
ARCHITECTURAL DESIGN
All rights reserved, 2010, Thales Alenia Space
Page 10
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
ARCHITECTURAL DESIGN
OVERALL BLOCK DIAGRAM
• Four quadrant• 256RPCs/quadrant• 16 ADCs/quadrant• Quadrant Trigger Logic• Global Trigger Logic• Acquisition I/F• Configuration I/F• Calibration Network• Probe Network
All rights reserved, 2010, Thales Alenia Space
Page 11
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
RPC DESIGN
READOUT PIXEL CELL
300µm
300µm
AVAILABLE AREA
All rights reserved, 2010, Thales Alenia Space
Page 12
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
RPC DESIGN
•Dedicated preamplifiers supply line
•P-MOS input stage & cascode
•Continuous reset with subthresholdtransistor
•On/Off features
•Calibration input
• 100µm x 100µm occupation
• 120 µW power dissipation
CHARGE PREAMPLIFIER FEATURES
SIMPLIFIED SCHEMATICS
All rights reserved, 2010, Thales Alenia Space
Page 13
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
PZC, SHAPER AND STRETCHER
• Reset circuit non linearitycompensation
• Current-mode shaper
• DC cancellation
• Adjustable peak time1,2,4,5,6,7,9,10 µs
• Equalization gain w.r.t.shaping time
•250 µW power dissipation
RPC DESIGN
SIMPLIFIED BLOCK DIAGRAM
All rights reserved, 2010, Thales Alenia Space
Page 14
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
RPC TRIGGER CIRCUITRY
RPC DESIGN
• Amplitude discriminator
• Fine THR tuning
• Peak discriminator
• Current-mode trigger to periphery
• Reset Logic
• 110µW power dissipation
SIMPLIFIED BLOCK DIAGRAM
All rights reserved, 2010, Thales Alenia Space
Page 15
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
RPC DESIGN
INPUT PAD
RPC LAYOUT
PREAMPLIFIERSHAPER
STRETCHER
OUTPUT BUFFER
TRIGGER & RESET LOGIC
All rights reserved, 2010, Thales Alenia Space
Page 16
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
ADC DESIGN
ADCs
• 16 RPC/ADC
• Wilkinson, 10 bit
• sine/differential clock
• on-chip adjustable ramp slope
• < 2 mW power dissipation
All rights reserved, 2010, Thales Alenia Space
Page 17
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
READOUT LOGIC
QTU/GTU
• Sparse Readout
• Multiple events management
• Global FIFO
• Fast parallel acquisition I/F
All rights reserved, 2010, Thales Alenia Space
Page 18
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
ANCILLARY ON-CHIP FACILITIES
CALIBRATION NETWORK
•three independent nets
•internal or external pulsing
PROBE NETWORK• interesting nodes inspection
ON BOARD DACS• Threshold and preamplifierreset adjustable bytelecommand
All rights reserved, 2010, Thales Alenia Space
Page 19
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
PROTOTYPES
All rights reserved, 2010, Thales Alenia Space
Page 20
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
PROTOTYPES
Four foundry runs:
• 16x16 study prototype #1 (MPW)
• 16x16 study prototype #2 (MPW)
• 32x32 prototype (STARX32 v.1) (dedicated run)
• 32x32 “final” (STARX32 v.2) (dedicated run)
Selected Technology: AMS C35 0.35µm CMOS
All rights reserved, 2010, Thales Alenia Space
Page 21
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TESTING
All rights reserved, 2010, Thales Alenia Space
Page 22
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TESTING
DEDICATED TEST SYSTEM
Device mounted on carrier hybrid
Shielded box with local biasing
ReadOut box ( FPGA )
Fast digital I/O card for data
Standard I/O card forconfiguration
PC-controlled
Control and Analysis SW
All rights reserved, 2010, Thales Alenia Space
Page 23
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TESTING
CARRIER HYBRID
provides a light-tightand protectedenvironment for the device under test
allows easy mating/demating with the test equipment
Hosts some ancillarypassives for filtering/biaspurposes
All rights reserved, 2010, Thales Alenia Space
Page 24
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TESTING
TYPICAL RPC SIGNALS
All rights reserved, 2010, Thales Alenia Space
Page 25
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TESTING
ENC MEASUREMENTS (room-T, no-detector)
0 2 4 6 8 1010
15
20
25
Equ
ival
ent n
oise
cha
rge
[ ele
ctro
ns r
.m.s
. ]
Shaping Time [ µs ]
STAR-X32 v.1
RPC 48-4r.m.s. voltmeter
Minimum ENC:
~ 16 electrons rms at 6µs peak time
All rights reserved, 2010, Thales Alenia Space
Page 26
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TESTING
END-TO-END LINEARITY
(include on-board ADC)
INL < 0.25 %
Range ~1000 to 10000 electrons
Stimulus via calibration network
All rights reserved, 2010, Thales Alenia Space
Page 27
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TESTING
END-TO-END spectrum
(stimulus by on board calibration)
Line details
All rights reserved, 2010, Thales Alenia Space
Page 28
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TESTING
QUICK-LOOK ANALYSIS
• all RPCs found working
• thin individual spectra lines
• gain/offset spread …
All rights reserved, 2010, Thales Alenia Space
Page 29
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TESTING
PEAK WIDTH
noise estimation from peak w idth
0
5
10
15
20
25
0 200 400 600 800 1000
channel ID, 0=(0,0) to 1023=(31,31)
aver
age
peak
wid
th [e
lect
rons
rm
s]
All rights reserved, 2010, Thales Alenia Space
Page 30
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TESTING
PEAK WIDTH
peak w idth distributionmean = 16 electrons rms, sigma = 1.2 electrons rms
0
50
100
150
200
250
300
350
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
peak w idth [electrons rms]
occu
renc
ies
All rights reserved, 2010, Thales Alenia Space
Page 31
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TEST WITH DETECTOR
without detector with detector
• Custom-made 32x32 pixels p-n Silicon detector procured for test in realistic condition
• Bump-Bonded on some samples of STARX-32, one sample put on carrier hybrid
All rights reserved, 2010, Thales Alenia Space
Page 32
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TEST WITH DETECTOR
• bump-bonding process functionally OK (just 1 out of 102 4 missing bond found)
• process possibly critical w.r.t. pixels performance:
FINDINGS
reverse currents much higher than expected,
up to > 50 nA/cm²,
is this due to bump-bonding process ?
All rights reserved, 2010, Thales Alenia Space
Page 33
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TEST WITH DETECTOR
FINDINGS
• inter-pixel cross-talk
some signals from an activeRPC can AC-couple to nearby
RPCs via detector
Additional shielding (metal) above disturbing nodes
implemented in final prototype.
effectiveness to be tested soon
All rights reserved, 2010, Thales Alenia Space
Page 34
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
TEST WITH DETECTOR
All rights reserved, 2010, Thales Alenia Space
Page 35
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
RADIATION TEST
Radiation Test have been performed atESTEC Co-60 Facility
Total Dose = up to 59 kradIntermediate steps every 8 kradTotal Irradiation Time = 3 Days
TID TEST
In view of possible application on spaceborne instruments…
All rights reserved, 2010, Thales Alenia Space
Page 36
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
RADIATION TEST
A dedicated Test Board has been designed and manufactured.
The board has been used both for biasing during irradiation and for measurements.
The board is able to host up to 6 chips together and to stimulate and test up to 5 chip at the same time.
TEST BOARD
All rights reserved, 2010, Thales Alenia Space
Page 37
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
RADIATION TEST
TID TEST RESULTS
• most of RPC parameters do not show significant variations up to 60 krad
• one RPC bias show constant negative trend ~1% at 60 krad (tolerable)
• only marginal drifts on ADCs
• two samples needed re-adjustof the rest voltage above 40 krad
The test results suggest that the AMS 0.35 µm CMOS technology used in the LFDR prototype is radtolerant at least to 60 krad.
All rights reserved, 2010, Thales Alenia Space
Page 38
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
SUMMARY & CONCLUSIONS
All rights reserved, 2010, Thales Alenia Space
Page 39
AMICSA 2010 workshop, 5-7 september 2010, ESA-ESTEC
SUMMARY & CONCLUSIONS
• We presented the latest development for the STARX32 proj ect
• The overall performances are very good, e.g. ENC and p ower dissipation, moreover radiation tolerance (TID) has been proven up to > 50 krad
• As expected for such a challenging design task we are im proving the design through successive prototypes and extended test campai gns
• Control of internal interference is of paramount importanc e and the major reason for the need of multiple prototypes
• Bump-Bonded detector requires carefull “top” shielding o f the front end cell
• The 4-th prototype (STARX32 v.2) has been manufactured in May 2010 and is presently under test, results so far are encouragin g, we are lookingforward to bump-bond a sample with a detector for comple te end-to-end test