PIPETTING AND LABORATORY
MATHEMATICS
ANTONIO F. LAUDE JR, RMT MPH
SCHOOL OF MEDICAL TECHNOLOGY
Pipetting Technique using Manual
Pipettes
Develop good technique
Through practice
Improper pipetting – laboratory accidents
Mouth pipetting – potential hazard
Never acceptable
Caustic reagents
Contaminated specimens
Poisonous solutions
Pipetting with Manual Pipettes
1. Check the pipette to ascertain its correct size.
2. Wearing protective gloves.
3. Place the tip of the pipette well below the surface
of the liquid to be pipetted.
4. Use aspirator bulb – pipette until the level of
liquid is well above the calibration mark
5. Quickly cover the suction opening at the top of the
pipette with the index finger
Pipetting with Manual Pipettes
6. Wipe the outside of the pipette dry with a piece of
gauze or tissue to remove excess fluid.
7. Hold the pipette in a vertical position with the
delivery tip against the inside of the original vessel.
Carefully allow the liquid in the pipette to drain by
gravity.
Miniscus it the concave or convex surface of a column
of liquid as seen in a laboratory pipette, buret, or
other measuring device.
Pipetting with Manual Pipettes
8. While still holding the pipette in a vertical position,
touch the tip of the pipette to the inside wall of the
receiving vessel.
Remove index finger, permit free drainage.
TD pipettes – small amount will remain in the delivery
tip.
Pipetting with Manual Pipettes
9. To be certain that the drainage is as complete as
possible, touch the delivery tip of the tip to another
area on the inside wall of the receiving vessel.
10. Remove the pipette from the receiving vessel, and
place it in appropriate place for washing.
Automatic Micropipettors
Micropipettor – most common type of micropipette
used in many laboratories
Allow repeated, accurate, reproducible delivery of
specimens, reagents, and other liquids requiring
measurement in small amounts.
Delivery volume is selected by adjusting the settings
on the device.
0.5 to 500 L – allow volume
Automatic Micropipettors
Contain or deliver from 1 to 500 L
Follow manufacturers instructions
Steps apply for use of a micropipettor
1. Attach the proper tip to the pipettor, and set the
delivery volume.
2. Depress the piston to a stop position on the
pipettor.
3. Place the tip into the solution, and allow the piston
to rise slowly back to its original position.
4. Some tips are wiped with a dry gauze at this step,
and some are not wiped. Follow the
manufacturer’s direction.
Steps apply for use of a micropipettor
5. Place the tip on the wall of the receiving vessel,
and depress the piston, first to stop position where the
liquid is allowed to drain, then to a second stop
position where the full dispensing of the liquid takes
place.
6. Dispose of the tip in the waste disposal receptacle.
Some pipettors automatically eject the used tips, thus
minimizing biohazard exposure.
Reading Meniscus
1. With a clear colorless solution, read the bottom of
the meniscus
2. With colored solutions, read the top fluid column.
3. Reading must be made with eye level to avoid
parallax error.
LABORATORY MATHEMATICS
Minimum numbers of digits needed to express a
particular value in scientific notation without loss
of accuracy.
728.4 contains how many significant figures?
7.284 x 102
0.000532 contains how many significant figures?
5.32 x 10-4
Significant Figures
Is equal to parts per 100 or the amount of solute per 100 total units of solution
It is determined in the same manner regardless of whether it is w/w, v/v or w/v units are used
Percent solution
1. Weight/Volume (w/v) % solutions
The most common type of solution prepared in clinical
laboratory
Refers to the number of grams of solute per 100 mL
of solution
Grams of solute = % solution desired x total volume desired
100
2. Volume/Volume (v/v)% solutions
Used when both solute and solvent are liquid
It refers to the amount of solute in mL in 100mL of
solvent
mL of solute = % solution desired x total volume desired
100
3. Weight/Weight (w/w) % solutions
Refers to the number of grams of solute per 100 gms
of solution
Grams of solute = % solution desired x grams of the total solution
100
When preparing concentrated acid solutions,
always add acid to water.
MOLARITY
The number of moles of solute per liter of solution.
1 mole of substance equals its gram molecular weight
(gmw)
Gram Molecular Weight (GMW) – is obtained by
adding the atomic weights of the component elements.
Molarity of solution (M) = Grams of solute .
GMW x volume of solution (L)
Sample Problem
Solution
Solution
MOLARITY
To prepare a molar solution:
Grams of solute = Molarity x GMW of the solute x
Volume (liter) desired
To convert % w/v to Molarity (M):
M = %w/v x 10
GMW
MOLARITY
The amount of solute per 1 kilogram of solvent
Expressed in terms of weight/weight or moles per
kilogram (mol/kg)
Molecular Weight (MW) is obtained by adding the
atomic weights of the given compound.
Molality (m) = Grams of solute
MW x kg of solvent
NORMALITY
Is the number of equivalent weight of solute per liter of solution
It has been used in acid-base calculations
Normality (N) = Grams of solute
EW x volume (L)
Equivalent weight (EW) = MW
valence
NORMALITY
To prepare a Normal solution of Solids:
Grams of solute = EW x Normality x Volume (L)
To convert % w/v to Normality (N):
N = % w/v x 10
EW
Relationship between Molarity and Normality:
Normality = Molarity x valence
Molarity = Normality
valence
Sample Problem
Solution
Milliequivalents
The most common way of expressing electrolytes
mEq/L= mg/dL x 10 x valence
MW
Millimoles
Molecular weight in millimoles (mmol/L)
mmol/L= mg/dL x 10
MW
RATIO and DILUTION
Ratio = Volume of solute
Volume of solvent
Dilution = Volume of solute
Volume of solution
Represents the ratio of concentrated or stock
material to the total/ final volume of a solution
DILUTIONS
(volume increases, concentration decreases,
amount of solute remains the same)
DILUTIONS
3 Reasons for Doing Dilution
1. The concentration of material is HIGH to be
accurately measured.
2. Removal of undesirable substances like in PFF
(Protein-free filtrate) preparation
3. Preparation of working standard from stock
solution
Dilution factor
The ration of a concentrated solution to the total
solution volume equals the dilution factor.
Is made by adding the concentrated stock to a
diluent
Example 1
What is the dilution factor needed to make a 100 mEq/L sodium solution from a 3000 mEq/L stock solution? The dilution factor becomes
100 = 1
3000 30
This means that the ratio of stock is 1 part stock made to a total volume of 30.
To make the solution: 1mL of stock is added to 29 mL of diluent.
The sum of the amount of the stock material plus the
amount of diluent must equal the total volume or
dilution fraction denaminator.
The dilution factor may be written as a fraction or
can be expressed as 1:30
1/30 or 1:30 either may be used
Example 2
If in the preceding example, 150 mL of the 100
mEq/L sodium solution was required,
The dilution ratio stock to total volume must be
maintained.
Set up a ratio between the desired total volume
and the dilution factor to determine the amount of
stock needed.
Equation
1 = x
30 150
5/150 = 1/30
To make this solution:
5mL of stock is added to 145 mL of the
appropriate diluent
Making the stock volume to diluent equal to 5/145
Simple Dilutions
The laboratorian must decide on the total volume desired and the amount of stock to be used
Examples:
A 1:10 (1/10) dilution of serum can be achieved by using any of the following approaches:
a. 100 uL of serum and 900 uL of saline
b. 20 uL of serum and 180 uL of saline
c. 1 mL of serum and 9 mL of saline
d. 2 mL of serum and 18 mL of saline
Dilution factor
Is used to determine the concentration of a dilution
or stock material by multiplying the original
concentration by the dilution factor.
When determining the original stock or undiluted
concentration, multiply the concentration of the
dilution by the dilution factor denominator.
Example
A 1:2 dilution of serum with saline had a creatinine
result of 8.6 mg/dL. Calculate the actual serum
creatinine concentration.
Dilution factor ½
Dilution result = 8.6 mg/dL
Because this result represents ½ of the
concentration, the actual serum creatinine value is
2 x 8.6 = 17.2 mg/dL