  • Instron LX-150 Operation

    System Startup

    Disconnect switch to on.

    Start workstation

    Attach Extensometer

    Run Partner

    Open test procedure Firmware will download

    Disengage E-STOP

    After download, calibrate load cell:

    Transducer Calibrate Load Cell

    Calibrate extensometer:

    Transducer Calibrate extensometer

    Warm-up system for 15 minutes, minimum. After warm-up, re-calibrate all.

    Run Test

    Position ram for specimen length

    Tighten lower grips. Slack upper grip.

    Attach specimen upper threads Attach specimen lower threads.

    Fix upper grip

    Remove slack in assembly

    Attach extensometer

    Check zeros, re-zero if necessary

    Run procedure

    On signal, remove extensometer

    After test is complete, print data

    Remove broken specimen

    System Shutdown

    DISABLE FRAME from menu

    Engage E-STOP

    Close partner

    Open Disconnect switch

    Remove extensometer