Instron LX-150 Operation System Startup Disconnect switch to “on”. Start workstation Attach Extensometer Run Partner Open test procedure Firmware will download Disengage E-STOP After download, calibrate load cell: Transducer Calibrate Load Cell Calibrate extensometer: Transducer Calibrate extensometer Warm-up system for 15 minutes, minimum. After warm-up, re-calibrate all. Run Test Position ram for specimen length Tighten lower grips. Slack upper grip. Attach specimen upper threads Attach specimen lower threads. Fix upper grip Remove slack in assembly Attach extensometer Check zeros, re-zero if necessary Run procedure On signal, remove extensometer After test is complete, print data Remove broken specimen System Shutdown DISABLE FRAME from menu Engage E-STOP Close partner Open Disconnect switch Remove extensometer

Instron LX-150 Operation

Embed Size (px)

Citation preview

  • Instron LX-150 Operation

    System Startup

    Disconnect switch to on.

    Start workstation

    Attach Extensometer

    Run Partner

    Open test procedure Firmware will download

    Disengage E-STOP

    After download, calibrate load cell:

    Transducer Calibrate Load Cell

    Calibrate extensometer:

    Transducer Calibrate extensometer

    Warm-up system for 15 minutes, minimum. After warm-up, re-calibrate all.

    Run Test

    Position ram for specimen length

    Tighten lower grips. Slack upper grip.

    Attach specimen upper threads Attach specimen lower threads.

    Fix upper grip

    Remove slack in assembly

    Attach extensometer

    Check zeros, re-zero if necessary

    Run procedure

    On signal, remove extensometer

    After test is complete, print data

    Remove broken specimen

    System Shutdown

    DISABLE FRAME from menu

    Engage E-STOP

    Close partner

    Open Disconnect switch

    Remove extensometer