Instron LX-150 Operation
System Startup
Disconnect switch to on.
Start workstation
Attach Extensometer
Run Partner
Open test procedure Firmware will download
Disengage E-STOP
After download, calibrate load cell:
Transducer Calibrate Load Cell
Calibrate extensometer:
Transducer Calibrate extensometer
Warm-up system for 15 minutes, minimum. After warm-up, re-calibrate all.
Run Test
Position ram for specimen length
Tighten lower grips. Slack upper grip.
Attach specimen upper threads Attach specimen lower threads.
Fix upper grip
Remove slack in assembly
Attach extensometer
Check zeros, re-zero if necessary
Run procedure
On signal, remove extensometer
After test is complete, print data
Remove broken specimen
System Shutdown
DISABLE FRAME from menu
Engage E-STOP
Close partner
Open Disconnect switch
Remove extensometer