Carotid stenting-in stent restenosis and
treatment modalities
Jagiellonian University Institute of Cardiology,
John Paul II Hospital Krakow, Poland
Prof. Piotr Pieniazek MD, PhD
Disclosure
Speaker name: PIOTR PIENIAZEK.
I have the following potential conflicts of interest to report:
X Consulting
Employment in industry
Stockholder of a healthcare company
Owner of a healthcare company
Other(s)
I do not have any potential conflict of interest
• comprehensive patient evaluation including non-invasive imaging:
• extra- and intracranial Doppler
• extra- and intracranial CT angio
• brain perfusion (MRI)
• independent neurological consultation (NIHSS, MMSE, Rankin Scale)
• patient-and-lesion-tailored neuroprotection before meshstent era
(Roadsaver & CGuard) - 2015!!!!!!
• independent histologic evaluation of captured material
• coronary angiography prior to CAS (except pts. after CABG or recent PCI)
rigorous follow-up is mandatory !!!!!
The Krakow Program of Stroke Prevention
by Carotid Artery Stenting (from Jan 2001)
4055 CAS procedures
in a single Krakow Center
The largest meta-analysis comparing outcome in different stent
design before meshstents era.
In-stent restenosis after carotid stenting is not a trivial problem as
it still exists.
• Occurs less frequently than in subclavian, vertebral, renal or peripheral arteries.
• Is less frequent than restenosis after carotid endarterectomy.
• 10% according guidelines is not acceptable.
• Treatment modality in-stent restenosis has still growing progress.
• Re- PTA for in-stent restenosis needs always temporary NPD.
• In recurrent in-stent restenosis of label devices need to be use.
• Rigorous US investigation in such patients is mandatory.
Never compare incomparable groups.
Then it's easy to get compromising conclusions!!!
12m. follow-up
Treatment modality of ISR
Patient 78 y. with bilateral ICA stenosis after syncope two months earlier with recurrent -
bilateral RESTENOSIS
LICA ReRe in-stent restenosis: 80% IVUS stenosis; MLA 4.2mm . CAS - 20 Feb. 2018
Viatrac 4.5/20mm STENTYS X position 3.0-3.5/27mm, MAVERIC 5.0/20mm & FilterWire
Diffuse in-stent & edge in-stent restenosis. Standard balloon & DCB5.0/40mm 3 inflation 8 atm for 45 min!! Our standard of treatment for in-stent restenosis
Standard
Balloon
4,5/20mm
DCB
5.0/40mm
Filter
Our current approach with Carotid In-Stent Restenosis. DCB is used after
excellent artery preparation with Stantard, Cutting or NC Balloon. Temporary
NPD use is mandatory.
DEB
Standard
Balloon
NC Balloon
24 Atm
COMPARE : Does Ranger DCB with dose density of
2μg/mm2 have similar efficacy compared to and In.
Pact DCB with dose density of 3.5μg/mm2?
Important knowledge after publication from Dec.2018 which should be considered
Ranger
5mm
In.Pact
5mm
40mm 1301 2553
Nominal dose (ug/device) to treat a typical
lesion.
Lenght
Ranger5mm
1301
All balloons have CE registrations
Treatment modality for in-stent restenosis.
Balloon angioplasty always with bigger diameter.
Cutting Balloon (only whey the standard balloon slides away from
lesion).
Second self expanding stent (stent in stent technique, do not use
Carotid Wallstent).
Drug eluting balloon with Paclitaxel (problem with 45 sec.
inflation).
Self expanding sirolimus coronary stent – STENTYS X position
(withdrawn from the market in 2019 ) but the new DCB is coming.
Surgery should be avoided!!!!
Take home message :
• Restenosis after carotid stenting is not a frequent complication
and should not exceed 5%
• Treatment strategy for in-stent restenosis is changing
• New technological solutions should be taken for consideration
• We use DCB with Paclitaxel as a standard choice for in-stent
restenosis treatment
• For restenosis after CAS or CAE do not use Carotid Wallstent
Thank you
John Paul II Hospital Krakow PL.