50
XD-502 Operator’s Manual 204 Technology Park Lane, Fuquay-Varina, NC 27526 USA Tel: (919) 557-7300 Web: www.apexinst.com

XD-502 Operator’s Manual · 2019. 8. 9. · XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console Apex Instruments, Inc. 4 Apex Instruments, Inc. 5 Warranty Policy

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

  • XD-502Operator’s Manual

    204 Technology Park Lane,Fuquay-Varina, NC 27526 USA

    Tel: (919) 557-7300Web: www.apexinst.com

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.2

    Apex Instruments, Inc.3

    204 Technology Park LaneFuquay-Varina, NC 27526

    www.apexinst.com(919) 557-7300(800) 882-3214

    For Support or Service contact:

    Technical Services [email protected]

    (877) 726-3919

    XD-502 Operator’s Manual - SF v2.1 Revision Date: 5.2.19

    Apex Instruments Inc. makes no representations or warranties with respect to this manual and, to the maximum extent permitted by law, expressly limits its liability for breach of any warranty that may be implied to the replacement of this manual with another. Furthermore, Apex Instruments reserves the right to revise this publication at any time without incurring an obligation to notify any person of the revision.

    The information provided in this documentation contains general descriptions and/or technical characteristics of the performance of the products contained herein. This documentation is not intended as a substitute for and is not to be used for determining suitability or reliability of these products for specific user applications. It is the duty of any such user or integrator to perform the appropriate and complete risk analysis, evaluation and testing of the products with respect to the relevant specific application or use thereof. Neither Apex Instruments nor any of its affiliates or subsidiaries shall be responsible or liable for misuse of the information that is contained herein. If you have any suggestions for improvements or amendments or have found errors in this publication, please notify us.

    All pertinent state, regional, and local safety regulations must be observed when installing and using this product. For reasons of safety and to help ensure compliance with documented system data, only the manufacturer shall perform repairs to components.

    Failure to observe this information can result in injury or equipment damage.

    Copyright © 2019 by Apex Instruments, Inc.

    All rights reserved. No part of this publication may be reproduced, distributed, or transmitted in any form or by any means, including photocopying, recording, or other electronic or mechanical methods, without the prior written permission of the publisher.

    Disclaimer

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.4

    Apex Instruments, Inc.5

    Warranty Policy

    Product Limited Warranty Policy Apex Instruments, Inc. warrants products to be free from defects in materials and workmanship for a period of one year from date of delivery. During the warranty period, we will repair or replace, at our option, any component that fails due to a defect in materials or workmanship under normal use excluding installation. Items not manufactured by Apex Instruments are subject to original manufactures warranty. All returns must be shipped freight prepaid at the purchasers’ expense.

    Repair Services Limited Warranty Policy Apex Instruments, Inc. warrants repairs to be free from defects in materials for a period of one year and workmanship for a period of 90 days from date of delivery. During the warranty period, we will repair any item Apex Instruments repaired or replaced that fails due to a defect in materials or workmanship under normal use.

    What is Not CoveredAccidents, misuse, abuse and modifications. All forms of corrosion or other environmental conditions including lightning, natural disasters, improper installation and installation of an incorrect part. Normal wear and tear, used or salvage parts installed and shipping damage.

    ContaminationCustomer is responsible for decontamination of all equipment returned to Apex Instruments. Apex Instruments reserves the right to return or decontaminate any equipment deemed unsafe or contaminated at the customers expense. Additional Important Information This limited warranty does not cover labor cost or incidental, indirect, special or consequential damages such as, but not limited to, psychical injuries or property damage, loss of time, loss of use of the component, inconvenience, rental equip-ment charges, or accommodations resulting from a defect in or failure of the component. You must be the original pur-chaser of the component, warranty is not transferable.

    This limited warranty represents the exclusive remedy for any component defect or failure and the total liability of Apex Instruments for any component it warrants.

    Apex Instruments makes no other warranties, express or implied, including any warranties of merchantability, condition of any kind or fitness for a particular purpose.

    Some states do not allow disclaimers of implied warranties. Therefore, all implied warranties that may apply to the compo-nent are limited to the duration of this written warranty. Some states also do not allow limitations on how long an implied warranty lasts or the exclusion or limitation of incidental, special or consequential damages, so the above information may not apply. This warranty provides specific rights, and there may be other rights which vary from state to state. Apex Instruments, Inc does not authorize any entity to vary the terms, conditions or exclusions of this warranty. If any provision of the warranty should become invalid or unenforceable because of any laws, the remaining terms and provisions of the warranty shall remain in full force and effect. To the extent allowed under local law, the remedies provided in this Limited Warranty Policy are the purchaser’s sole and exclusive remedies. Apex Instruments, Inc., 204 Technology Park Lane, Fuquay-Varina, NC 27526. Visit us on the web at www.apexinst.com.

    Acceptance of Terms and Conditions Apex Instruments manufactures and sells its products under the foregoing terms and conditions and will not be bound to terms and conditions stated by any purchaser in any offer, acceptance or other contractual document. Placing an order or accepting delivery of products constitutes acceptance of terms and conditions.

    Table of Contents

    Disclaimer .........................................................................................................................3Warranty Policy ................................................................................................................41.0 Preface ........................................................................................................................8

    1.1 Description of the User .................................................................................................... 8

    1.2 Purpose of this manual .................................................................................................... 8

    1.3 Explanation of Safety Warnings ...................................................................................... 9

    1.4 Retaining Instructions ...................................................................................................... 9

    1.5 Obtaining Documentation and Information .................................................................... 101.5.1 Internet ...........................................................................................................................101.5.4 Support and Service.......................................................................................................10

    2.0 Description of the Product ...................................................................................... 112.1 Intended Use and Reasonably Foreseeable Misuse ......................................................11

    2.3 System Overview .......................................................................................................... 12

    2.4 Technical Specifications ................................................................................................ 14

    2.5 Product Compliance ...................................................................................................... 15

    2.6 Console Components .................................................................................................... 16

    3.0 Safety Instructions ...................................................................................................183.1 How to Use the Product Safely ..................................................................................... 18

    3.1.3 Safety Information related to the intended use ..............................................................18

    4.0 Audit and Leak Checks with the XD-502 ................................................................194.1 Transducer Calibration Check ....................................................................................... 19

    4.2 Calibration Checks: Pressure Transducers ................................................................... 20

    4.3 Console Leak Check Procedure (Vacuum Side) ........................................................... 24

    4.4 Pre-Calibration Leak Test for Consoles with a Diaphragm Pump ................................. 26

    4.5 Thermocouple and Barometric Pressure Calibration Checks ....................................... 27

    4.6 Setting the Fuji Temperature Controller ......................................................................... 28

    5.0 Preparation ...............................................................................................................295.1 What to do when you receive your console ................................................................... 29

    5.1.1 Unpack and inspect .......................................................................................................295.1.2 Leak Checks ..................................................................................................................295.1.3 Calibration Audits ...........................................................................................................29

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.6

    Apex Instruments, Inc.7

    5.1.4 Test Plan and Methods ...................................................................................................29

    5.2 How to Transport and Store the Console ...................................................................... 295.2.1 Dimensions ....................................................................................................................295.2.2 Lifting and handling ........................................................................................................29

    6.0 Using the Console ...................................................................................................306.1 User Interface Design .................................................................................................... 31

    6.1.1 User Interface Buttons ...................................................................................................326.1.2 Button Functions Explained ...........................................................................................336.1.3 User Interface - Button Function Chart ..........................................................................346.1.4 User Interface - Screen Hierarchy..................................................................................356.1.5 User Display Screen Navigation ....................................................................................36

    6.2 (UI) Sampling Test Operations with the User Interface ................................................. 38UI.1 Pre-Test Preparation........................................................................................................38UI.2 On Site Preparation .........................................................................................................41UI.3 Sample Test Operations ..................................................................................................45

    6.3 Software Design ............................................................................................................ 486.3.1 Main Screen ...................................................................................................................496.3.2 Monitor Screen ...............................................................................................................566.3.3 Job Information Screen ..................................................................................................576.3.4 Storage Screen ..............................................................................................................586.3.5 Utility Screen ..................................................................................................................59

    6.4 (SW) Sampling Test Operations with the Software ....................................................... 65SW.1 Pre-Test Preparation......................................................................................................65SW.2 On Site Preparation .......................................................................................................71SW.3 Sample Test Operations ................................................................................................79SW.4 Exporting Data ...............................................................................................................84

    7.0 Maintenance .............................................................................................................857.1 User Inspection and Maintenance of Console ............................................................... 85

    7.1.2 Pitot Tube Lines .............................................................................................................857.1.3 Orifice Meter Lines .........................................................................................................857.1.4 Timer ..............................................................................................................................857.1.5 Temperature Controller ..................................................................................................857.1.6 User Interface Display ....................................................................................................85

    7.1.7 Dry Gas Meter ................................................................................................................867.1.8 Console In-Line Filter .....................................................................................................877.1.9 Thermocouple Plugs ......................................................................................................87

    7.2 Manufacturer Support for your Product ......................................................................... 887.2.1 Technical Services .........................................................................................................887.2.2 Calibration Services .......................................................................................................88

    7.3 Identifying and Solving Common Console Problems .................................................... 89

    8.0 Appendix ...................................................................................................................908.1 Console Maintenance Front Panel Parts Chart ............................................................. 90

    8.2 Console Maintenance Interior Parts Chart .................................................................... 91

    8.3 Console Replacement Parts .......................................................................................... 92

    8.4 Plumbing Diagram ......................................................................................................... 94

    8.5 Electrical Diagram ......................................................................................................... 96

    9.0 References and Related Documentation ...............................................................989.1 References .................................................................................................................... 98

    9.2 Related Documentation ................................................................................................. 98

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.8

    Apex Instruments, Inc.9

    1.0 Preface

    1.1 Description of the UserIsokinetic source sampling is intended to be performed by technicians who have been trained in source sampling methods. Personnel conducting source sampling are expected to understand the theory of isokinetic source sampling, including basic gas laws and chemistry.

    ! DANGER

    In addition, all technicians should have adequate general safety training as well as site specific training to identify, abate, and prevent job related hazards including site specific training.

    1.2 Purpose of this manual

    The purpose of this manual is to provide a basic understanding of the Apex Instruments Isokinetic XD-502 Source Sampling Console. The XD-502 is applicable for use with a variety of US EPA Isokinetic Sampling Methods.

    Additionally, this manual provides the users with a reasonable amount of reference information on system configuration, calibration procedures, maintenance and troubleshooting referencing the specific product and the US EPA Regulations.

    For more information about isokinetic sampling, see our reference materials (Isokinetic Sampling Handbook or the 450 Sampling Student Manual).

    Isokinetic Sampling

    The first four source test reference methods are applied before conducting Method 5. If prior data is available or reasonably good assumptions can be made, not all of the methods need to be conducted. If you look at the isokinetic rate equation and the nozzle diameter equation, you can see that there are a number of parameters that must be followed before one can proceed with isokinetic sampling.

    US EPA Method Descriptions:

    Method 1 Determination of Sampling Location and Traverse PointsMethod 2 Determination of Stack Gas Velocity and Volumetric Flow-ratesMethod 3 Determination of Dry Molecular Weight and Percent Excess AirMethod 4 Determination of Moisture ContentMethod 5 Determination of Particulate Matter Emissions from Stationary Sources

    Detailed information on method sampling may be found through the US EPA website - please visit http://www.epa.gov/emc-tests-methods/ for complete method descriptions.

    1.3 Explanation of Safety Warnings

    ! DANGER

    Danger indicates a hazard with a high level of risk which, if not avoided, will result in death or serious injury

    ! WARNING

    Warning indicates a hazard with a medium level of risk which, if not avoided, could result in death or serious injury.

    ! CAUTION

    Caution indicates a hazard with a low level of risk which, if not avoided, could result in minor or moderate injury.

    NOTICE

    Indicates information considered important, but not hazard-related.

    1.4 Retaining InstructionsRead and understand this manual and its safety instructions before using this product. Failure to do so can result in serious injury or improper execution of sampling procedures.

    Follow all relevant instructions. This will avoid fire, explosions, electric shocks or other hazards that may result in damage to property and/or severe or fatal injuries.

    The product shall only be used by persons who have fully read and understand the contents of this user manual.

    Ensure that each person who uses the product has read these warnings and instructions and follows them.

    Keep all safety information and instructions for future reference and pass them on to subsequent users of the product.

    The manufacturer is not liable for cases of material damage or personal injury caused by incorrect handling or non-compliance with the safety instructions. In such cases, the warranty will be voided.

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.10

    Apex Instruments, Inc.11

    1.5 Obtaining Documentation and Information

    1.5.1 Internet

    Detailed product information, stack sampling guides and EPA regulation references are available on our website:

    https://www.apexinst.com

    1.5.4 Support and Service

    Our knowledgeable service staff includes skilled industry professionals, stack testers and technicians ready to help you with your specific service needs.

    From basic troubleshooting to full equipment overhauls and repairs, our technical service team is available to offer support to the user.

    Toll Free (877) 726-3919 or (919) 346-5754

    E-mail: [email protected].

    2.1 Intended Use and Reasonably Foreseeable Misuse

    2.1.1 Isokinetic Sampling

    The Sampling Console is the operator’s control station. The XD-502 can be used via the user interface or the software to execute sampling operations for EPA regulation or EPA equivalent regulation compliance.

    2.1.2 Versatile Use

    You can easily adapt the basic Method 5 sampling train to test for many other gaseous and particulate emissions from stationary sources. Adapting basic test methods allows you to expand testing to include parameters of interest such as metals, polychlorinated biphenyls (PCBs), dioxins/furans, polycyclic aromatic hydrocarbons (PAHs), particle size distributions and an ever-increasing group of other pollutants.

    2.0 Description of the Product

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.12

    Apex Instruments, Inc.13

    2.3 System OverviewNOTE: This manual covers the sampling console part of the train. The rest of the console is sold separately.

    The first step to effective sampling is to familiarize yourself with the standard equipment. To illustrate the necessary components of source sampling, we’ve included an overview of the six main components of the Apex Instruments Source Sampling Train.

    The main components of an Isokinetic Sampling system are: probe assembly, filter oven, impinger case, umbilical, sampling pump, and the console.

    2.3.1 Probe Assembly

    Includes a probe sheath, probe liner, tube heater, Type-S pitot tubes, nozzle, stack and probe Type K thermocouples. The probe nozzle draws in sample from the source effluent and the probe sheath protects the heated probe liner.

    2.3.2 Heated Filter Oven (Hot Box)

    The insulated, heated filter oven contains the filter assembly to collect particulate from the gas sample while preventing condensation of moisture within the sample.

    2.3.3 Impinger Case (Cold Box)

    The impinger serves the purpose of preventing the moisture from entering the console. The case includes a durable, foam-insulated, drainable reservoir with a pre-punched foam insert for positioning and holding the impinger bottles in place. The ice lowers the sample gas temperature which causes condensation inside the impingers. Moisture and other condensables are trapped for gravimetric or chemical analysis. The sample gas exits the last impinger and passes through a check valve designed to eliminate back pressure.

    2.3.4 Umbilical Cable

    Connects the impinger case and heated filter oven assembly to the source sampling console and includes the sample line, two pitot lines, 5 type K thermocouples, a power output line, and a spare orsat line.

    2.3.5 Sampling Console

    The sampling console houses the dry gas meter, internal sampling pump, temperature controllers, differential pressure measurement devices, flow adjustment valves and a user interface display.

    Some sampling trains use an external pump assembly to pull the sample. An external pump is not necessary when a console, like the XD-502, uses an internal sampling pump to pull the sample.

    Sample gas enters the console through the umbilical, passes through a filter to remove any foreign matter, then through the pump and dry gas meter, and exits the system through a calibrated orifice tube. The calibrated orifice tube is connected to the differential pressure measurement device to allow the user a

    ***The other components of the Sampling System are sold separately.***

    reference of sample gas flow. The ∆P reading is then compared to the ∆H and flow is adjusted to maintain an isokinetic sampling rate.

    The console is housed in a strong and durable polyethylene case with a carrying strap on top.

    Probe Assembly

    Heated Filter Oven

    Impinger Case

    Umbilical Cable

    Sampling Console

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.14

    Apex Instruments, Inc.15

    2.4 Technical SpecificationsDry Gas Meter: Model SK25EX, with 100 CPR Quadrature Encoder, 0.7L/rev., 41 LPM max.

    Sample Pump: Internal Double Head Diaphragm Pump, 24VDC, 18 LPM @ 15” Hg and 70 LPM free flow.

    Display: 4x20 Character Back-lit Transflective Liquid Crystal Display with digital operation of Timer, ∆P, ∆H, DGM Volume, Vacuum, Barometric Pressure, and up to 8 Temperature Channels.

    Display Control: Six-button Long-Life Membrane Keypad for display operation and standard operations.

    Flow Indicator: Precision Machined Stainless Steel Orifice with Pressure Transducer, Range 0 - 5” H2O (1245 Pa) resolution of 0.01” (1 Pa).

    Temperature Measurement: 8 channel individually isolated Type-K thermocouple channels, °C/°F selectable, -200°C to 1372°C range. (-328°F to 2502°F).

    Digital Pressure Transducers: Ultra high resolution temperature-compensated, miniaturized,ΔP +/-1.0” H2O (+/- 248.8 Pa) Range Bi-Directional with 0.001” (1 Pa) Resolution, ΔP +/-10.0” H2O (+/- 2488 Pa) Range Bi-Directional with 0.001” (1 Pa) Resolution, ΔH +/-5.0” H2O (+/-1245 Pa) Range Bi-Directional with 0.01” (1 Pa) Resolution.

    Vacuum Measurement: PCB-mounted Sensor.

    Umbilical Connection: Sample Inlet: 1/2” Stainless Steel Instrumental Quick-Connect (alternate sizes available). Pitot Connections: 1/4” Stainless Steel Instrumental Quick-Connects. Power Output Connector: Twist to lock, 120V/220V for Aux, Probe, Filter Power. TC Connections: AUX, STACK, PROBE, OVEN, FILTER, EXIT.

    Power: Input Power: 120V/15A 60Hz, IEC C-14 Inlet. Pump Power: 150W, 24V/6.3A. Display Power Supply: 15W, 12VDC. Optional Input Power: 240V/10A 50Hz, IEC C-14 Inlet.

    Dimensions: H17” x W17” x D12” (43 cm x 43 cm x 30.5 cm).

    Weight: 37 lbs (9.98kg).

    2.5 Product ComplianceThis product complies with all relevant US EPA Sampling Regulations and was designed based on the schematics and sampling practices established by the EPA in the Code of Federal Regulations.

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.16

    Apex Instruments, Inc.17

    2.6 Console Components

    1. Hard Plastic, Lightweight Case

    9. Fine Valve andCoarse Valve

    2. Transflective Backlit User Display

    3. Membrane Switch Buttons

    4. Company Contact Information

    5. Console Calibration Information

    6. Temperature Controllers

    7. Probe and Oven Switches

    8. Main, Pump and AUX Switches

    10. Thermocouple Jacks

    12. Power Supply Receptacle

    13. Power

    14. Main Circuit Breaker Switch

    15. PC Input Connection

    11. Sample and Pitot Line Quick Connects

    Console Components List

    1. Hard plastic, lightweight case

    2. Transflective, back-lit User Display - Displays measurements and sampling data during operation

    3. Membrane Switch Buttons - Used to interact with the user display to control sampling operations

    4. Company Contact Information

    5. Console Calibration Information - The annual calibration results data is displayed via decal in a prominent location on the face

    plate.

    6. Temperature Controllers - 6 easy-access thermocouple jacks for connecting the aux/stack/probe/oven/filter/exit

    7. Probe and Oven Switches

    8. Main, Pump, and AUX Switches

    9. Fine and Coarse Valves - Flow may be adjusted with a ball valve for coarse adjustment and a needle valve for fine

    adjustment.

    10. Thermocouple Jacks

    11. Sample and Pitot Line Quick Connects - Pitot connections for ∆P+ and ∆P- pressures and a sample inlet quick connect.

    12. Power Supply Receptacle

    13. Power Output - Supplies power for probe heater, filter oven, and aux

    14. Main Circuit Breaker Push/Pull Switch - The electronics are protected by a resettable panel mounted circuit breaker.

    15. USB Data Port - Establishes a connection to a device that can use the software to operate the console and export

    data

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.18

    Apex Instruments, Inc.19

    3.1 How to Use the Product Safely

    3.1.3 Safety Information related to the intended use

    Isokinetic source sampling is intended to be performed by technicians who have been trained in source sampling methods. Personnel conducting source sampling are expected to understand the theory of isokinetic source sampling, including basic gas laws and chemistry.

    ! DANGER

    In addition, all technicians should have adequate general safety training to identify, abate, and prevent job related hazards including site specific training.

    Please visit the following link for more information on Stack Sampling Safety Procedures and Protocols: http://www.sesnews.org/

    3.0 Safety Instructions 4.0 Audit and Leak Checks with the XD-502

    4.1 Transducer Calibration CheckThe Transducers on the XD-502 Electronics Package are high resolution, digital ultra low pressure sensors. The operational range of the transducers are 1”, 5” and 10” H2O with an accuracy of 0.25% FSS BFSL (full scale span, best fit straight line). All error compensation is performed internally by an advanced ASIC (application specific integrated circuit) Therefore, no external calculation/calibration is required. When a calibration check is required by EPA method regulations it is recommended that a NIST-traceable, handheld manometer be utilized to verify the transducers are within regulation.

    Check the calibration of the transducer at a minimum of three points approximately representing the range of ∆P values in the stack. The ∆P values, as read by the transducer and NIST device, must agree to within 5 percent for each point for the transducer to be considered within proper calibration. The following pages provide the procedures for conducting Transducer Calibration Checks.

    ATTACH LINE FROMMANOMETER TO LUER “T”

    TO SYRINGE OROTHERPRESSURE

    GENERATINGDEVICE

    1” HO

    10” HO

    5” HO

    XD-502 MAIN BOARDPCB-502MAIN-COMP

    Co

    m 1

    2 3

    4

    5

    6 7

    8

    SW

    ITC

    HE

    S

    8V

    -29V -

    +

    1

    5

    10

    1234

    rP+

    rP-

    Pitot 1 +

    Pitot 1 -

    ATTACH LINE FROMMANOMETER TO LUER “T”

    TO SYRINGE OROTHERPRESSURE

    GENERATINGDEVICE

    1” HO

    10” HO

    5” HO

    XD-502 MAIN BOARDPCB-502MAIN-COMP

    Co

    m 1

    2 3

    4

    5

    6 7

    8

    SW

    ITC

    HE

    S

    8V

    -29V -

    +

    1

    5

    10

    1234

    rP+

    rP-

    Pitot 1 +

    Pitot 1 -

    Air Flow in Diagramin T Valve

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.20

    Apex Instruments, Inc.21

    4.2 Calibration Checks: Pressure Transducers

    NOTICE

    DO NOT EXCEED THE TRANSDUCER PRESSURE RATING OF 100 INCHES OF H2O. DOING SO MAY RESULT IN OPERATIONAL SENSOR FAILURE

    ! CAUTION

    REMOVE THE CONSOLE FROM THE CASE. NOT DOING SO COULD POSE AN ELECTRICAL HAZARD

    See US EPA Method 5 Section 8.4.1 concerning Leak Check Procedures.

    A 20ml syringe, a tee valve and a manometer may be used.

    1. Close the coarse valve to seal off the sample line from the rest of the system. (Figure 4-1)

    2. The dry gas meter (DGM) outlet orifice must be sealed off using the 18mm stopper provided in the audit kit. (Figure 4-2)

    3 Locate the positive and negative ∆H lines on the DGM Orifice. Trace the lines back. (Figure 4-3)

    4. Disconnect (using a 1/2 twist) the ∆H line luer connections and attach the luer Lock “T”’s as shown to both lines. (Figure 4-4)

    5. Reposition the positive ∆H valve to the ALL OPEN position as shown. This position vents the positive side of the ∆H transducer through all three directions. (Figure 4-5)

    6. Reposition the negative ∆H line in the VENT position as shown. This position vents the Negative side of the ∆H transducer to atmosphere. (Figure 4-6)

    to PCB

    to PCB

    to PCB

    to PCB

    to Orifice

    to Orifice

    to Orifice

    to Orifice

    Positive ∆H

    Negative ∆H

    - +

    - +

    to PCB

    to PCB

    to PCB

    to PCB

    to Orifice

    to Orifice

    to Orifice

    to Orifice

    Positive ∆H

    Negative ∆H

    - +

    - +

    to PCB

    to PCB

    to PCB

    to PCB

    to Orifice

    to Orifice

    to Orifice

    to Orifice

    Positive ∆H

    Negative ∆H

    - +

    - +

    to PCB

    to PCB

    to PCB

    to PCB

    to Orifice

    to Orifice

    to Orifice

    to Orifice

    Positive ∆H

    Negative ∆H

    - +

    - +

    to PCB

    to PCB

    to PCB

    to PCB

    to Orifice

    to Orifice

    to Orifice

    to Orifice

    Positive ∆H

    Negative ∆H

    - +

    - +

    Figure 4-1

    Figure 4-2

    Figure 4-3

    Figure 4-4

    Figure 4-5

    Figure 4-6

    1

    2

    3

    4

    5

    6

    7. The ∆H readout should be 0.00. Both sides of the transducers are vented to air. (See Figure 4-7)

    8. Connect a vacuum tee to the +∆H tee and position the lever as shown. (See Figure 4-8)

    9. Pull the plunger on the syringe to the end of its travel within the cylinder or turn the adjustment counterclockwise on the pressure generator to the halfway point. (See Figure 4-9)

    10. Attach the syringe or pressure generator to the vacuum tee as shown. (See Figure 4-10)

    11. Verify positive ∆H reading is 0.00. (See Figure 4-11)

    12. Use a digital manometer for the final steps in this procedure. (See Figure 4-12)

    Figure 4-7

    Figure 4-8

    Figure 4-9

    Figure 4-10

    Figure 4-11

    Figure 4-12

    to PCBto Orifice

    ∆H+

    ∆H+

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    to PCBto Orifice ∆H+

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    to PCBto Orifice

    ∆H+

    ∆H+

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    to PCBto Orifice ∆H+

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    to PCBto Orifice

    ∆H+

    ∆H+

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    to PCBto Orifice ∆H+

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    to PCBto Orifice

    ∆H+

    ∆H+

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    to PCBto Orifice ∆H+

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    T� � � � : 0:00:00PITOT: 0.0ORIFICE: .00TEMPERATURE 25C

    11

    12

    7

    8

    9

    10

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.22

    Apex Instruments, Inc.23

    13. Connect the digital manometer tubing to the ∆H+ tee and to the P1 port on the manometer as shown. (See Figure 4-13)

    14. SLOWLY push the plunger on the syringe to approximately 5ml. The digital manometer reading should match the console ∆H reading. Repeat procedure at 10ml and 15ml. The readings between the ∆H display and the digital manometer should match indicating there are no leaks and the transducers have passed audit. (See Figures 4-14a, 4-14b and 4-14c)

    Figure X413

    Figure 4-14a

    Figure 4-14cFigure 4-14b

    to Orifice to PCB

    to Orifice to PCB

    to Orifice to PCBto Orifice to PCB

    ∆H+

    ∆H+

    ∆H+∆H+

    .05” H2O

    .05” H2O .05” H2O

    13

    14a

    14c14b

    15. Repeat steps 1-14 as needed for the 1” ∆P transducer and the 10” ∆P transducer.

    Each transducer must be checked at 0 (zero), 30% of full range, 60% of full range and 90% of full range for both positive and negative ranges.

    16. Reconnect plumbing in reverse order.

    17. See CFR sections 10.8.1 and 10.8.2 for performing a post-test single-point calibration check.

    ∆P 1”H2O Transducer

    Full Range Values: Reading must be within 3% or +/- .03” H2O 90% = -.90” H2O60% = -.60” H2O30% = -.30” H2O0 (Zero)30% = .30” H2O60% = .60” H2O90% = .90” H2O

    ∆P 10”H2O Transducer

    Full Range Values: Reading must be within 3% or +/- .30” H2O 90% = -9.00” H2O60% = -6.00” H2O30% = -3.00” H2O0 (Zero) = +/- .30” H2O30% = 3.00” H2O60% = 6.00” H2O90% = 9.00” H2O

    Method 2, 2F, 2G Post-test Single-Point Transducer Calibration Checks

    16.2.2.1.1 “Before calibrating the meter box, leak check the system as follows: Fully open the coarse adjust valve, and completely close the by-pass valve. Plug the inlet. Then turn on the pump, and determine whether there is any leakage. The leakage rate shall be zero (i.e., no detectable movement of the DGM dial shall be seen for 1 minute).”

    16.2.2.1.2 “Checkalsoforleakagesinthatportionofthesamplingtrainbetweenthepumpandtheorificemeter.See section 8.4.1 for the procedure; make any corrections, if necessary. If leakage is detected, check for cracked gaskets, loosefittings,wornO-rings,etc.,andmakethenecessaryrepairs.”

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.24

    Apex Instruments, Inc.25

    4.3 Console Leak Check Procedure (Vacuum Side)

    NOTICE

    DO NOT EXCEED THE TRANSDUCER PRESSURE RATING OF 100 INCHES OF H2O. DOING SO MAY RESULT IN OPERATIONAL SENSOR FAILURE

    1. Go to Leak Test screen and position the valves

    - Press Insert Note button on the console and use the up and down arrows to select the Leak Test. - Close the coarse valve by turning the handle right to

    horizontal position. - Increase the fine tune knob fully by turning all the way to

    the right (clockwise).

    2. Block airflow of sample inlet

    - Insert a capped quick-connect into the sample quick-connect 1/2” inlet.

    3. Initiate flow

    - Turn on the pump, then open the coarse valve (twist to vertical position).

    4. Adjust vacuum (if necessary)

    - If your vacuum is below 15” Hg (380mm) slowly increase the fine tune knob until the vacuum reading is at least 15”Hg (380mm).

    1

    2

    3

    4

    5♦

    5. Run the test

    - Start the timer and monitor the DGM flow for 1 minute. If the flow through the DGM exceeds 0.02 CFM (0.00057m3/min or 4% of average sample rate) the leaks must be found and corrected.

    6. Run second test

    - Turn off the pump. Monitor the vacuum for one minute. If the vacuum decreases any amount, then you have a leak somewhere between the plugged QC and the vacuum sensor.

    7. Stop the test

    - Remove the plug from the sample inlet.

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.26

    Apex Instruments, Inc.27

    4.4 Pre-Calibration Leak Test for Consoles with a Diaphragm Pump

    Apex Instruments recommends the following procedure to be followed for conducting a “back-side” leak check:

    10.3.1.1 (Method 5) “Before calibrating the metering system, it is suggested that a leak check be conducted. For metering systems having diaphragm pumps, the normal leak-check procedure will not detect leakages within the pump.

    For these cases the following leak-check procedure is suggested:

    Makea10-minutecalibrationrunat0.00057m³/min(0.020cfm).Attheendoftherun,takethedifferenceofthemeasuredwettestmeterandDGMvolumes.Dividethedifferenceby10(numberofminutesinthecalibration run) to get the leak rate. The leak rate should not exceed 0.00057 m³/min (0.020 cfm).”

    The calibrated critical orifice to be used for this procedure is #09.

    NOTICE

    Italicized text is directly quoted from EPA method regulation documentation (See References and Related Documentation)

    0000 168

    DGM Out

    Vac.

    (Max.) BarometricPressure

    Internal

    0.0

    0.0

    0.0

    0.0

    LPM(CFM)

    Liters(CF)

    0.0

    0.0

    Display/Transducer Module

    P-

    P-

    P+

    P+

    P-

    H+P+

    H-

    1 10 5

    Side View

    To Vacuum Sensor on

    PCB

    Derived from Sensor onPCB Board

    #09Orifice

    Delta H

    0.0

    XD-502 Console

    Delta P

    0.0

    Console Display

    InsertNote

    Select Stop/Start TimerPress to Open LeakTest Screen

    Press to Open InsertNotes Screen

    Return to Main Screen

    Displays briefly

    From Any Main Menu Screen

    Leak Test and Notes Menu

    4.5 Thermocouple and Barometric Pressure Calibration Checks

    The six thermocouple jacks on the XD-502 console face plate require periodic calibration and audit per the EPA Methods.

    Temperatures audited should encompass the range of conditions expected during testing.

    Thermocouple 2: Stack

    10.3.1 (Method 2*)“In-stackThermocouplesshallbecalibratedaftereachfielduseatatemperaturewithin10 percent of the average absolute stack temperature.”

    10.3.2. (Method 2*) “The sensor being calibrated and the reference sensor shall agree to within 1.5%.,

    Thermocouple E: Filter

    6.1.1.7 (Method 5) A temperature sensor capable of measuring temperature towithin ±3 °C (5.4 °F) shall be installed so that the sensing tip of the temperature sensor is indirectcontactwiththesamplegasexitingthefilter.

    Thermocouples 1, 3, 4, 6:Aux, Probe, Oven, Exit

    10.5 (Method 5) “Temperature Sensors. Use the procedure in Section 10.3 of Method 2 to calibrate in-stack temperature sensors. Dial thermometers, such as are used for the DGM and condenser outlet, shall be calibrated against mercury-in-glass thermometers. An alternative mercury-free NIST traceable thermometer maybeusedifthethermometeris,ataminimum,equivalentintermsofperformanceorsuitablyeffectiveforthespecifictemperaturemeasurementapplication.Asanalternative,thefollowingsingle-pointcalibrationprocedure may be used. After each test run series, check the accuracy (and, hence, the calibration) of each thermocouplesystematambienttemperature,oranyothertemperature,withintherangespecifiedbythemanufacturer, using a reference thermometer (either ASTM reference thermometer or a thermometer that has been calibrated against an ASTM reference thermometer). The temperatures of the thermocouple and reference thermometers shall agree to within ±1 °F.”

    10.6 (Method 5) “Barometer. Calibrate against a mercury barometer or NIST-traceable barometer prior to the fieldtest.Alternatively,barometricpressuremaybeobtainedfromaweatherreportthathasbeenadjustedforthe test point (on the stack) elevation.”

    NOTICE

    Italicized text is directly quoted from EPA method regulation documentation (See References and Related Documentation)

    EXIT FILTER OVEN PROBE STACK AUX

    1 2 3 4 5 6

    Tg = ______ ºF + 459.67º = ______ ºC + 273.15º

    Tg = ______ ºF + 459.67º = ______ ºC + 273.15º

  • XD-502 Digital Isokinetic Console XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.28

    Apex Instruments, Inc.29

    To change the Set Value (SV) press once. Current Value is displayed and SV is lit.

    Press the to increase or decrease the displayed SV (set value).

    After the new setting is entered press to lock in the value and return to C1 (control 1) display.

    Repeat procedure to check value if necessary.

    SV

    SV

    C1

    C1

    C2

    C2

    AL1

    AL1

    AL2

    AL2

    SEL

    SEL

    SEL

    SEL

    SEL

    SEL

    °C

    °C

    72

    ProGSVC1C2AL1AL2

    SEL

    °C

    P

    SVC1C2AL1AL2

    SEL

    °C

    Pn-2

    SVC1C2AL1AL2

    SEL

    °C

    5

    SEL

    SVC1C2AL1AL2

    SEL

    °C

    4SVC1C2AL1AL2

    SEL

    °C

    3

    SVC1C2AL1AL2

    SEL

    °C

    38SVC1C2AL1AL2

    SEL

    °C

    32

    SEL

    SV

    SV

    C1

    C1

    C2

    C2

    AL1

    AL1

    AL2

    AL2

    SEL

    SEL

    SEL

    SEL

    SEL

    SEL

    °C

    °C

    72

    ProGSVC1C2AL1AL2

    SEL

    °C

    P

    SVC1C2AL1AL2

    SEL

    °C

    Pn-2

    SVC1C2AL1AL2

    SEL

    °C

    5

    SEL

    SVC1C2AL1AL2

    SEL

    °C

    4SVC1C2AL1AL2

    SEL

    °C

    3

    SVC1C2AL1AL2

    SEL

    °C

    38SVC1C2AL1AL2

    SEL

    °C

    32

    SEL

    SV

    SV

    C1

    C1

    C2

    C2

    AL1

    AL1

    AL2

    AL2

    SEL

    SEL

    SEL

    SEL

    SEL

    SEL

    °C

    °C

    72

    ProGSVC1C2AL1AL2

    SEL

    °C

    P

    SVC1C2AL1AL2

    SEL

    °C

    Pn-2

    SVC1C2AL1AL2

    SEL

    °C

    5

    SEL

    SVC1C2AL1AL2

    SEL

    °C

    4SVC1C2AL1AL2

    SEL

    °C

    3

    SVC1C2AL1AL2

    SEL

    °C

    38SVC1C2AL1AL2

    SEL

    °C

    32

    SEL

    SV

    SV

    C1

    C1

    C2

    C2

    AL1

    AL1

    AL2

    AL2

    SEL

    SEL

    SEL

    SEL

    SEL

    SEL

    °C

    °C

    72

    ProGSVC1C2AL1AL2

    SEL

    °C

    P

    SVC1C2AL1AL2

    SEL

    °C

    Pn-2

    SVC1C2AL1AL2

    SEL

    °C

    5

    SEL

    SVC1C2AL1AL2

    SEL

    °C

    4SVC1C2AL1AL2

    SEL

    °C

    3

    SVC1C2AL1AL2

    SEL

    °C

    38SVC1C2AL1AL2

    SEL

    °C

    32

    SEL

    4.6 Setting the Fuji Temperature Controller

    Set Value Setting 5.1 What to do when you receive your console

    5.1.1 Unpack and inspect

    Unpack your console from its shipping container. Inspect the console. Check your the Packing List to ensure everything has arrived.

    5.1.2 Leak Checks

    Perform your console leak checks as explained in section 4.

    5.1.3 Calibration Audits

    Perform your console calibration audits as explained in section 4.

    5.1.4 Test Plan and Methods

    Begin your sampling operation procedures as directed by the method regulations.

    5.2 How to Transport and Store the Console

    5.2.1 Dimensions

    Height: 17” (43 cm)Width: 17” (43 cm)Depth: 12” (30.5 cm)Weight: 40lbs (16.8 kg)

    5.2.2 Lifting and handling

    ! CAUTION

    Avoid dropping the console and any other forms of collision during transport.

    When lifting, make sure to use the handles on the sides to lift.

    Do not try to lift the console by anything other than the handles or the carrying strap.

    5.0 Preparation

  • User Interface - DesignXD-502 Digital Isokinetic Console

    Apex Instruments, Inc.31

    XD-502 Digital Isokinetic Console

    Apex Instruments, Inc.30

    6.0 Using the Console

    Methods of Operation

    The XD-502 Sampling Console can be operated using two different methods: User Interface Control and Software Control.

    User Interface Console Control

    Operation of the sampling console can be performed using the panel-mounted User Interface. The multifunction membrane switches allow the user to perform all necessary sampling operations and monitor all critical information by interacting with the display.

    Software Console Control

    Operation of the sampling console can be also performed using the Console Software. After the software is installed on a Windows Operating System device and the device is connected to the console, the user is able to perform all the necessary sampling operations and monitor all critical information using the software.

    Certain settings and features that provide extra versatility are only accessible through Software Control. See Section 6.3 for more information.

    LEAK TEST AND NOTES MENU

    1 to 254 Averaging Size

    Select Press to Save

    0 to 50% Change

    Press to Save

    1 to 254 Averaging Size

    Press to Save

    0.0 to 25.0% Change

    Press to Save

    InsertNote

    Return to Damping Menu Screen

    Press to Open SelectedMenu Item Screen

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screen

    MAIN SCREENOPENING SCREEN TEMPERATURE SCREEN TEST INFORMATION SCREEN

    Press to Open SelectedMenu Item Screen

    Press to Open Selected Menu Item Screen

    Select Menu Item

    Select Menu Item

    Select Menu Item

    9999 Maximum Value

    Select Press to Reset Counter

    Return to Test Options Screen

    1 to 15

    Select Press to Overwrite

    Return to Additonal Menu Screen

    Press to Reset TransducersSelect

    Return to Test Options Screen

    Set mm:ss

    Select

    Select

    Press to Save

    Return to Test Options Screen

    Stop/Start Timer

    Press to Open SelectedMenu Item Screen

    Press to Open SelectedMenu Item Screen

    1 to 254 Averaging Size

    Select

    Select

    Press to Save

    0 to 50% Change

    Press to Save

    1 to 254 Averaging Size

    Press to Save

    0.0 to 25.0% Change

    Press to Save

    Return to Damping Menu Screenu

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screenu

    Press to Open Selected Menu Item Screen

    Set mm:ss (5 Second Minimum)

    Select Press to Save

    Return to Test Options Screen

    Select Press to Save

    Return to Additonal Menu Screen

    Metrics, Imperials, Metrics (Pa)

    Timer is ZEROTEST OPTIONS SCREEN ADDITIONAL MENU SCREEN

    TIMER IS ACTIVE

    NO = Back to NEW Run Screen

    YES= Back to Standby Menu

    Return to Main Screen

    Select Select Select

    Return to MarkerScreen

    Displays briefly

    Select Press to Save Marker

    Displays briefly then Returns to Previous Main Menu Screen

    Hold for 3 Seconds to access OPERATION OPTIONS MENU

    Select Menu Item

    ----------------

    ----------------

    ----------------

    ----------------

    --------------------------------

    --------------------------------

    MAIN DISPLAY SCREENS FROM ANY MAIN MENU SCREEN

    DISPLAYS BRIEFLY

    DAMPING MENU SCREEN

    1234

    1234

    Timer is ZERO Timer is ZERO

    5678

    9

    1

    10

    2

    DAMPING MENU SCREEN

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    ACTUAL SLOTLOCATION TBD

    Select

    MenuDGM

    ZeroTimer

    PreviousScreen

    InsertNote

    ¹ Does not include Error Screens 10.01.2018

    LEAK TEST AND NOTES MENU

    1 to 254 Averaging Size

    Select Press to Save

    0 to 50% Change

    Press to Save

    1 to 254 Averaging Size

    Press to Save

    0.0 to 25.0% Change

    Press to Save

    InsertNote

    Return to Damping Menu Screen

    Press to Open SelectedMenu Item Screen

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screen

    MAIN SCREENOPENING SCREEN TEMPERATURE SCREEN TEST INFORMATION SCREEN

    Press to Open SelectedMenu Item Screen

    Press to Open Selected Menu Item Screen

    Select Menu Item

    Select Menu Item

    Select Menu Item

    9999 Maximum Value

    Select Press to Reset Counter

    Return to Test Options Screen

    1 to 15

    Select Press to Overwrite

    Return to Additonal Menu Screen

    Press to Reset TransducersSelect

    Return to Test Options Screen

    Set mm:ss

    Select

    Select

    Press to Save

    Return to Test Options Screen

    Stop/Start Timer

    Press to Open SelectedMenu Item Screen

    Press to Open SelectedMenu Item Screen

    1 to 254 Averaging Size

    Select

    Select

    Press to Save

    0 to 50% Change

    Press to Save

    1 to 254 Averaging Size

    Press to Save

    0.0 to 25.0% Change

    Press to Save

    Return to Damping Menu Screenu

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screenu

    Press to Open Selected Menu Item Screen

    Set mm:ss (5 Second Minimum)

    Select Press to Save

    Return to Test Options Screen

    Select Press to Save

    Return to Additonal Menu Screen

    Metrics, Imperials, Metrics (Pa)

    Timer is ZEROTEST OPTIONS SCREEN ADDITIONAL MENU SCREEN

    TIMER IS ACTIVE

    NO = Back to NEW Run Screen

    YES= Back to Standby Menu

    Return to Main Screen

    Select Select Select

    Return to MarkerScreen

    Displays briefly

    Select Press to Save Marker

    Displays briefly then Returns to Previous Main Menu Screen

    Hold for 3 Seconds to access OPERATION OPTIONS MENU

    Select Menu Item

    ----------------

    ----------------

    ----------------

    ----------------

    --------------------------------

    --------------------------------

    MAIN DISPLAY SCREENS FROM ANY MAIN MENU SCREEN

    DISPLAYS BRIEFLY

    DAMPING MENU SCREEN

    1234

    1234

    Timer is ZERO Timer is ZERO

    5678

    9

    1

    10

    2

    DAMPING MENU SCREEN

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    ACTUAL SLOTLOCATION TBD

    Select

    MenuDGM

    ZeroTimer

    PreviousScreen

    InsertNote

    ¹ Does not include Error Screens 10.01.2018

    6.1 User Interface DesignThe multifunction membrane switches allow the user to perform all necessary sampling operations and monitor all critical information by interacting with the display.

    LEAK TEST AND NOTES MENU

    1 to 254 Averaging Size

    Select Press to Save

    0 to 50% Change

    Press to Save

    1 to 254 Averaging Size

    Press to Save

    0.0 to 25.0% Change

    Press to Save

    InsertNote

    Return to Damping Menu Screen

    Press to Open SelectedMenu Item Screen

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screen

    MAIN SCREENOPENING SCREEN TEMPERATURE SCREEN TEST INFORMATION SCREEN

    Press to Open SelectedMenu Item Screen

    Press to Open Selected Menu Item Screen

    Select Menu Item

    Select Menu Item

    Select Menu Item

    9999 Maximum Value

    Select Press to Reset Counter

    Return to Test Options Screen

    1 to 15

    Select Press to Overwrite

    Return to Additonal Menu Screen

    Press to Reset TransducersSelect

    Return to Test Options Screen

    Set mm:ss

    Select

    Select

    Press to Save

    Return to Test Options Screen

    Stop/Start Timer

    Press to Open SelectedMenu Item Screen

    Press to Open SelectedMenu Item Screen

    1 to 254 Averaging Size

    Select

    Select

    Press to Save

    0 to 50% Change

    Press to Save

    1 to 254 Averaging Size

    Press to Save

    0.0 to 25.0% Change

    Press to Save

    Return to Damping Menu Screenu

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screenu

    Press to Open Selected Menu Item Screen

    Set mm:ss (5 Second Minimum)

    Select Press to Save

    Return to Test Options Screen

    Select Press to Save

    Return to Additonal Menu Screen

    Metrics, Imperials, Metrics (Pa)

    Timer is ZEROTEST OPTIONS SCREEN ADDITIONAL MENU SCREEN

    TIMER IS ACTIVE

    NO = Back to NEW Run Screen

    YES= Back to Standby Menu

    Return to Main Screen

    Select Select Select

    Return to MarkerScreen

    Displays briefly

    Select Press to Save Marker

    Displays briefly then Returns to Previous Main Menu Screen

    Hold for 3 Seconds to access OPERATION OPTIONS MENU

    Select Menu Item

    ----------------

    ----------------

    ----------------

    ----------------

    --------------------------------

    --------------------------------

    MAIN DISPLAY SCREENS FROM ANY MAIN MENU SCREEN

    DISPLAYS BRIEFLY

    DAMPING MENU SCREEN

    1234

    1234

    Timer is ZERO Timer is ZERO

    5678

    9

    1

    10

    2

    DAMPING MENU SCREEN

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    ACTUAL SLOTLOCATION TBD

    Select

    MenuDGM

    ZeroTimer

    PreviousScreen

    InsertNote

    ¹ Does not include Error Screens 10.01.2018

    LEAK TEST AND NOTES MENU

    1 to 254 Averaging Size

    Select Press to Save

    0 to 50% Change

    Press to Save

    1 to 254 Averaging Size

    Press to Save

    0.0 to 25.0% Change

    Press to Save

    InsertNote

    Return to Damping Menu Screen

    Press to Open SelectedMenu Item Screen

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screen

    MAIN SCREENOPENING SCREEN TEMPERATURE SCREEN TEST INFORMATION SCREEN

    Press to Open SelectedMenu Item Screen

    Press to Open Selected Menu Item Screen

    Select Menu Item

    Select Menu Item

    Select Menu Item

    9999 Maximum Value

    Select Press to Reset Counter

    Return to Test Options Screen

    1 to 15

    Select Press to Overwrite

    Return to Additonal Menu Screen

    Press to Reset TransducersSelect

    Return to Test Options Screen

    Set mm:ss

    Select

    Select

    Press to Save

    Return to Test Options Screen

    Stop/Start Timer

    Press to Open SelectedMenu Item Screen

    Press to Open SelectedMenu Item Screen

    1 to 254 Averaging Size

    Select

    Select

    Press to Save

    0 to 50% Change

    Press to Save

    1 to 254 Averaging Size

    Press to Save

    0.0 to 25.0% Change

    Press to Save

    Return to Damping Menu Screenu

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screen

    Select

    Return to Damping Menu Screenu

    Press to Open Selected Menu Item Screen

    Set mm:ss (5 Second Minimum)

    Select Press to Save

    Return to Test Options Screen

    Select Press to Save

    Return to Additonal Menu Screen

    Metrics, Imperials, Metrics (Pa)

    Timer is ZEROTEST OPTIONS SCREEN ADDITIONAL MENU SCREEN

    TIMER IS ACTIVE

    NO = Back to NEW Run Screen

    YES= Back to Standby Menu

    Return to Main Screen

    Select Select Select

    Return to MarkerScreen

    Displays briefly

    Select Press to Save Marker

    Displays briefly then Returns to Previous Main Menu Screen

    Hold for 3 Seconds to access OPERATION OPTIONS MENU

    Select Menu Item

    ----------------

    ----------------

    ----------------

    ----------------

    --------------------------------

    --------------------------------

    MAIN DISPLAY SCREENS FROM ANY MAIN MENU SCREEN

    DISPLAYS BRIEFLY

    DAMPING MENU SCREEN

    1234

    1234

    Timer is ZERO Timer is ZERO

    5678

    9

    1

    10

    2

    DAMPING MENU SCREEN

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    **

    ACTUAL SLOTLOCATION TBD

    Select

    MenuDGM

    ZeroTimer

    PreviousScreen

    InsertNote

    ¹ Does not include Error Screens 10.01.2018

  • XD-502 Digital Isokinetic Console User Interface - Design User Interface - DesignXD-502 Digital Isokinetic Console

    Apex Instruments, Inc.32

    Apex Instruments, Inc.33

    6.1.1 User Interface Buttons

    The User Interface of the Console is where the user can monitor information and execute functions necessary for a sampling operation.

    The User Interface Controls:

    1. Arrow Up and Main Menu Button

    2. Arrow Down and Zero DGM Button

    3. Select and Zero Timer Button

    4. Timer and Previous Screen Button

    5. Insert Note Button

    6. Save Data Button

    7. Contrast Tuner

    The functionality of the user interface buttons varies based on how long they are pressed and the situation in which they are used. See Table 6-1 on the next page for more details.

    Specifically, the factors that change the function of the buttons are:

    - The duration of the button press: INSTANT or 3 SECOND HOLD - The status of the Main Test Timer: NON-ZERO or ZERO - The type of screen that is displayed: Main Navigation, Menu Screens, Leak Test Screen, or Setting

    1

    2 3 4 5

    6

    7

    6.1.2 Button Functions Explained

    1. Arrow Up/Main Menu Button - Primary Function (Simple Press): Moves the indicator/value/main navigation screen up one unit - Secondary Function (Only in Main Navigation - Hold for 3 Seconds): Navigates to Standby Menu if

    Timer is ZERO or Running Test Menu Screen if Timer is NON-ZERO.

    2. Arrow Down Button - Primary Function (Simple Press): Moves the indicator/value/main navigation screen down one unit - Secondary Function (Only in Main Navigation - When Timer NOT ACTIVE Hold for 3 Seconds): Resets

    the volume of the Dry Gas Meter reading - Leak Test Function: Zeroes the leak test DGM volume during a Leak Test

    3. Select and Zero Timer Screen Button - Primary Function (Only in Menu Screens): Enters the selected Menu Screen option - Secondary Function (Only in Main Navigation - Hold for 3 Seconds): Zeroes the Main Test Timer - Leak Test Function: Zeroes the timer during a Leak Test

    4. Timer and Previous Screen Button - Primary Function (Only if a TIMER is on screen - Simple Press): Starts and Stops a timer - Secondary Function (Only in Menu Screens): Navigates the user display back to the previous Main

    Navigation screen the user was on

    5. Insert Note Button - Primary Function (Only in Main Navigation): Navigates to the Insert Note Menu

    6. Save Data Button - Primary Function (Only in Main Navigation): Creates a Save Data Point for Exporting - Leak Test Function: Saves and exits Leak Test Mode

    7. Contrast tuner - Allows the user to use a screwdriver to adjust the degree of contrast of the characters on the screen of

    the display

  • XD-502 Digital Isokinetic Console User Interface - Design User Interface - DesignXD-502 Digital Isokinetic Console

    Apex Instruments, Inc.34

    Apex Instruments, Inc.35

    6.1.3 User Interface - Button Function Chart

    Main Navigation (press once)

    Previous Screen Next Screen N/A Starts/Stops the Timer

    Goes to Note and Leak Test Menu

    Creates a Save Data Point

    Main Navigation (3 Second hold)

    Go to Available

    Menu

    Zero DGM (If timer is INACTIVE) Zeroes the Timer Starts/Stops the Timer

    Goes to Note and Leak Test Menu

    Creates a Save Data Point

    Menu Screens (press once)

    Move Selection Cursor Up

    One

    Move Selection Cursor Down One

    Enters Selected Menu Option or

    Perform Selected Menu Option

    Goes Back to Previous Main Navigation

    ScreenN/A N/A

    Leak Test Screen N/A

    Press and Hold 3 Seconds to Reset the

    Leak Test Dry Gas Meter Volume

    Press and Hold 3 Seconds to zero the

    Leak Test timer

    Start and Stop the Leak Test timer N/A

    Saves and Exits the Leak Test

    Setting Screens (press once)

    Increases Value of

    Setting by One

    Decreases the Value of Setting by One Saves Data

    Goes Back to Previous Menu Screen N/A N/A

    Table 6-1: In-depth table of the functions of each button on the User Interface

    6.1.4 User Interface - Screen Hierarchy

    Main Navigation

    Menu Screens

    Settings Screens

    Running Test Menu

    Pitot ∆P Damping

    ∆P Damping Band

    Orifice ∆H Damping

    ∆H Damping Band

    Main Screen Temperature Screen

    Opening Screen

    Test Information Screen

    Standby Menu

    Pitot ∆P Damping

    Traverse Pt. SAVE

    ∆P Damping Band

    NEW Run

    Orifice ∆H Damping

    ∆H Damping Band

    Reset Marker #

    ZERO Sensors

    Notes and Leak Check

    Leak Test

    Insert Note Option

    Unit Type

  • XD-502 Digital Isokinetic Console User Interface - Design User Interface - DesignXD-502 Digital Isokinetic Console

    Apex Instruments, Inc.36

    Apex Instruments, Inc.37

    LE

    AK

    TE

    ST

    AN

    D N

    OT

    ES

    ME

    NU

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    0 to

    50

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Pre

    ss

    to S

    av

    e

    0.0

    to 2

    5.0

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    Insert

    No

    te

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    dM

    en

    u Ite

    m S

    cre

    en

    Se

    lec

    t

    Retu

    rn to

    Dam

    pin

    g M

    en

    u S

    cre

    en

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    MA

    IN S

    CR

    EE

    NO

    PE

    NIN

    G S

    CR

    EE

    NT

    EM

    PE

    RA

    TU

    RE

    SC

    RE

    EN

    TE

    ST

    INF

    OR

    MA

    TIO

    N S

    CR

    EE

    N

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    dM

    en

    u Ite

    m S

    cre

    en

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    d

    Me

    nu

    Item

    Sc

    ree

    n

    Sele

    ct

    Men

    u Ite

    mS

    ele

    ct

    Men

    u Ite

    mS

    ele

    ct

    Men

    u Ite

    m

    99

    99

    Ma

    xim

    um

    Va

    lue

    Se

    lec

    tP

    res

    s to

    Re

    se

    t Co

    un

    ter

    Re

    turn

    to T

    es

    t Op

    tion

    s S

    cre

    en

    1 to

    15

    Se

    lec

    tP

    res

    s to

    Ov

    erw

    rite

    Re

    turn

    to A

    dd

    iton

    al M

    en

    u S

    cre

    en

    Pre

    ss

    to R

    es

    et T

    ran

    sd

    uc

    ers

    Se

    lec

    t

    Re

    turn

    to T

    es

    t Op

    tion

    s S

    cre

    en

    Se

    t mm

    :ss

    Se

    lec

    t

    Se

    lec

    t

    Pre

    ss

    to S

    av

    e

    Re

    turn

    to T

    es

    t Op

    tion

    s S

    cre

    en

    Sto

    p/S

    tart T

    ime

    r

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    dM

    en

    u Ite

    m S

    cre

    en

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    dM

    en

    u Ite

    m S

    cre

    en

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Se

    lec

    t

    Se

    lec

    t

    Pre

    ss

    to S

    av

    e

    0 to

    50

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Pre

    ss

    to S

    av

    e

    0.0

    to 2

    5.0

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    u

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    u

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    d

    Me

    nu

    Item

    Sc

    ree

    n

    Se

    t mm

    :ss

    (5 S

    ec

    on

    d M

    inim

    um

    )

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    Re

    turn

    to T

    es

    t Op

    tion

    s S

    cre

    en

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    Re

    turn

    to A

    dd

    iton

    al M

    en

    u S

    cre

    en

    Me

    trics

    , Imp

    eria

    ls, M

    etric

    s (P

    a)

    Tim

    er is

    ZE

    RO

    TE

    ST

    OP

    TIO

    NS

    SC

    RE

    EN

    A

    DD

    ITIO

    NA

    L M

    EN

    U S

    CR

    EE

    N

    TIM

    ER

    IS A

    CT

    IVE

    NO

    = B

    ac

    k to

    NE

    W R

    un

    Sc

    ree

    n

    YE

    S=

    Ba

    ck

    to S

    tan

    db

    y M

    en

    u

    Re

    turn

    to M

    ain

    Sc

    ree

    n

    Se

    lec

    tS

    ele

    ct

    Se

    lec

    t

    Re

    turn

    to M

    ark

    erS

    cre

    en

    Dis

    pla

    ys

    brie

    fly

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    Ma

    rke

    r

    Dis

    pla

    ys

    brie

    fly

    the

    n

    Re

    turn

    s to

    Pre

    vio

    us

    M

    ain

    Me

    nu

    Sc

    ree

    n

    Ho

    ld fo

    r 3 S

    ec

    on

    ds

    to a

    cc

    es

    s

    OP

    ER

    AT

    ION

    OP

    TIO

    NS

    ME

    NU

    Sele

    ct

    Men

    u Ite

    m

    ----------------

    ----------------

    ----------------

    ----------------

    ----------------

    ----------------

    ----------------

    ----------------

    MA

    IN D

    ISP

    LA

    Y S

    CR

    EE

    NS

    FR

    OM

    AN

    Y M

    AIN

    ME

    NU

    SC

    RE

    EN

    DIS

    PL

    AY

    S B

    RIE

    FLY

    DA

    MP

    ING

    ME

    NU

    SC

    RE

    EN

    1234

    1234

    Tim

    er is

    ZE

    RO

    Tim

    er is

    ZE

    RO

    56789

    1

    10

    2

    DA

    MP

    ING

    ME

    NU

    SC

    RE

    EN

    **

    **

    **

    **

    **

    **

    * ** *

    **

    ** *

    *

    ***

    *

    AC

    TU

    AL

    S

    LO

    TL

    OC

    AT

    ION

    T

    BD

    Sele

    ct

    Me

    nu

    DG

    MZ

    ero

    Tim

    er

    Pre

    vio

    us

    Sc

    ree

    n

    Insert

    No

    te

    ¹ D

    oe

    s n

    ot in

    clud

    e E

    rror S

    cree

    ns

    10.0

    1.2

    018

    9

    LE

    AK

    TE

    ST

    AN

    D N

    OT

    ES

    ME

    NU

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    0 to

    50

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Pre

    ss

    to S

    av

    e

    0.0

    to 2

    5.0

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    Insert

    No

    te

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    dM

    en

    u Ite

    m S

    cre

    en

    Se

    lec

    t

    Retu

    rn to

    Dam

    pin

    g M

    en

    u S

    cre

    en

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    MA

    IN S

    CR

    EE

    NO

    PE

    NIN

    G S

    CR

    EE

    NT

    EM

    PE

    RA

    TU

    RE

    SC

    RE

    EN

    TE

    ST

    INF

    OR

    MA

    TIO

    N S

    CR

    EE

    N

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    dM

    en

    u Ite

    m S

    cre

    en

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    d

    Me

    nu

    Item

    Sc

    ree

    n

    Sele

    ct

    Men

    u Ite

    mS

    ele

    ct

    Men

    u Ite

    mS

    ele

    ct

    Men

    u Ite

    m

    99

    99

    Ma

    xim

    um

    Va

    lue

    Se

    lec

    tP

    res

    s to

    Re

    se

    t Co

    un

    ter

    Re

    turn

    to T

    es

    t Op

    tion

    s S

    cre

    en

    1 to

    15

    Se

    lec

    tP

    res

    s to

    Ov

    erw

    rite

    Re

    turn

    to A

    dd

    iton

    al M

    en

    u S

    cre

    en

    Pre

    ss

    to R

    es

    et T

    ran

    sd

    uc

    ers

    Se

    lec

    t

    Re

    turn

    to T

    es

    t Op

    tion

    s S

    cre

    en

    Se

    t mm

    :ss

    Se

    lec

    t

    Se

    lec

    t

    Pre

    ss

    to S

    av

    e

    Re

    turn

    to T

    es

    t Op

    tion

    s S

    cre

    en

    Sto

    p/S

    tart T

    ime

    r

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    dM

    en

    u Ite

    m S

    cre

    en

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    dM

    en

    u Ite

    m S

    cre

    en

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Se

    lec

    t

    Se

    lec

    t

    Pre

    ss

    to S

    av

    e

    0 to

    50

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Pre

    ss

    to S

    av

    e

    0.0

    to 2

    5.0

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    u

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    u

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    d

    Me

    nu

    Item

    Sc

    ree

    n

    Se

    t mm

    :ss

    (5 S

    ec

    on

    d M

    inim

    um

    )

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    Re

    turn

    to T

    es

    t Op

    tion

    s S

    cre

    en

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    Re

    turn

    to A

    dd

    iton

    al M

    en

    u S

    cre

    en

    Me

    trics

    , Imp

    eria

    ls, M

    etric

    s (P

    a)

    Tim

    er is

    ZE

    RO

    TE

    ST

    OP

    TIO

    NS

    SC

    RE

    EN

    A

    DD

    ITIO

    NA

    L M

    EN

    U S

    CR

    EE

    N

    TIM

    ER

    IS A

    CT

    IVE

    NO

    = B

    ac

    k to

    NE

    W R

    un

    Sc

    ree

    n

    YE

    S=

    Ba

    ck

    to S

    tan

    db

    y M

    en

    u

    Re

    turn

    to M

    ain

    Sc

    ree

    n

    Se

    lec

    tS

    ele

    ct

    Se

    lec

    t

    Re

    turn

    to M

    ark

    erS

    cre

    en

    Dis

    pla

    ys

    brie

    fly

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    Ma

    rke

    r

    Dis

    pla

    ys

    brie

    fly

    the

    n

    Re

    turn

    s to

    Pre

    vio

    us

    M

    ain

    Me

    nu

    Sc

    ree

    n

    Ho

    ld fo

    r 3 S

    ec

    on

    ds

    to a

    cc

    es

    s

    OP

    ER

    AT

    ION

    OP

    TIO

    NS

    ME

    NU

    Sele

    ct

    Men

    u Ite

    m

    ----------------

    ----------------

    ----------------

    ----------------

    ----------------

    ----------------

    ----------------

    ----------------

    MA

    IN D

    ISP

    LA

    Y S

    CR

    EE

    NS

    FR

    OM

    AN

    Y M

    AIN

    ME

    NU

    SC

    RE

    EN

    DIS

    PL

    AY

    S B

    RIE

    FLY

    DA

    MP

    ING

    ME

    NU

    SC

    RE

    EN

    1234

    1234

    Tim

    er is

    ZE

    RO

    Tim

    er is

    ZE

    RO

    56789

    1

    10

    2

    DA

    MP

    ING

    ME

    NU

    SC

    RE

    EN

    **

    **

    **

    **

    **

    **

    * ** *

    **

    ** *

    *

    ***

    *

    AC

    TU

    AL

    S

    LO

    TL

    OC

    AT

    ION

    T

    BD

    Sele

    ct

    Me

    nu

    DG

    MZ

    ero

    Tim

    er

    Pre

    vio

    us

    Sc

    ree

    n

    Insert

    No

    te

    ¹ D

    oe

    s no

    t inclu

    de

    Erro

    r Scre

    en

    s10.0

    1.2

    018

    10

    LE

    AK

    TE

    ST

    AN

    D N

    OT

    ES

    ME

    NU

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    0 to

    50

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Pre

    ss

    to S

    av

    e

    0.0

    to 2

    5.0

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    Insert

    No

    te

    Retu

    rn to

    Dam

    pin

    g M

    en

    u S

    cre

    en

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    dM

    en

    u Ite

    m S

    cre

    en

    Se

    lec

    t

    Retu

    rn to

    Dam

    pin

    g M

    en

    u S

    cre

    en

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    MA

    IN S

    CR

    EE

    NO

    PE

    NIN

    G S

    CR

    EE

    NT

    EM

    PE

    RA

    TU

    RE

    SC

    RE

    EN

    TE

    ST

    INF

    OR

    MA

    TIO

    N S

    CR

    EE

    N

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    dM

    en

    u Ite

    m S

    cre

    en

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    d

    Me

    nu

    Item

    Sc

    ree

    n

    Sele

    ct

    Men

    u Ite

    mS

    ele

    ct

    Men

    u Ite

    mS

    ele

    ct

    Men

    u Ite

    m

    99

    99

    Ma

    xim

    um

    Va

    lue

    Se

    lec

    tP

    res

    s to

    Re

    se

    t Co

    un

    ter

    Re

    turn

    to T

    es

    t Op

    tion

    s S

    cre

    en

    1 to

    15

    Se

    lec

    tP

    res

    s to

    Ov

    erw

    rite

    Re

    turn

    to A

    dd

    iton

    al M

    en

    u S

    cre

    en

    Pre

    ss

    to R

    es

    et T

    ran

    sd

    uc

    ers

    Se

    lec

    t

    Re

    turn

    to T

    es

    t Op

    tion

    s S

    cre

    en

    Se

    t mm

    :ss

    Se

    lec

    t

    Se

    lec

    t

    Pre

    ss

    to S

    av

    e

    Re

    turn

    to T

    es

    t Op

    tion

    s S

    cre

    en

    Sto

    p/S

    tart T

    ime

    r

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    dM

    en

    u Ite

    m S

    cre

    en

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    dM

    en

    u Ite

    m S

    cre

    en

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Se

    lec

    t

    Se

    lec

    t

    Pre

    ss

    to S

    av

    e

    0 to

    50

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Pre

    ss

    to S

    av

    e

    0.0

    to 2

    5.0

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    u

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    Se

    lec

    t

    Re

    turn

    to D

    am

    pin

    g M

    en

    u S

    cre

    en

    u

    Pre

    ss

    to O

    pe

    n S

    ele

    cte

    d

    Me

    nu

    Item

    Sc

    ree

    n

    Se

    t mm

    :ss

    (5 S

    ec

    on

    d M

    inim

    um

    )

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    Re

    turn

    to T

    es

    t Op

    tion

    s S

    cre

    en

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    Re

    turn

    to A

    dd

    iton

    al M

    en

    u S

    cre

    en

    Me

    trics

    , Imp

    eria

    ls, M

    etric

    s (P

    a)

    Tim

    er is

    ZE

    RO

    TE

    ST

    OP

    TIO

    NS

    SC

    RE

    EN

    A

    DD

    ITIO

    NA

    L M

    EN

    U S

    CR

    EE

    N

    TIM

    ER

    IS A

    CT

    IVE

    NO

    = B

    ac

    k to

    NE

    W R

    un

    Sc

    ree

    n

    YE

    S=

    Ba

    ck

    to S

    tan

    db

    y M

    en

    u

    Re

    turn

    to M

    ain

    Sc

    ree

    n

    Se

    lec

    tS

    ele

    ct

    Se

    lec

    t

    Re

    turn

    to M

    ark

    erS

    cre

    en

    Dis

    pla

    ys

    brie

    fly

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    Ma

    rke

    r

    Dis

    pla

    ys

    brie

    fly

    the

    n

    Re

    turn

    s to

    Pre

    vio

    us

    M

    ain

    Me

    nu

    Sc

    ree

    n

    Ho

    ld fo

    r 3 S

    ec

    on

    ds

    to a

    cc

    es

    s

    OP

    ER

    AT

    ION

    OP

    TIO

    NS

    ME

    NU

    Sele

    ct

    Men

    u Ite

    m

    ----------------

    ----------------

    ----------------

    ----------------

    ----------------

    ----------------

    ----------------

    ----------------

    MA

    IN D

    ISP

    LA

    Y S

    CR

    EE

    NS

    FR

    OM

    AN

    Y M

    AIN

    ME

    NU

    SC

    RE

    EN

    DIS

    PL

    AY

    S B

    RIE

    FLY

    DA

    MP

    ING

    ME

    NU

    SC

    RE

    EN

    1234

    1234

    Tim

    er is

    ZE

    RO

    Tim

    er is

    ZE

    RO

    56789

    1

    10

    2

    DA

    MP

    ING

    ME

    NU

    SC

    RE

    EN

    **

    **

    **

    **

    **

    **

    * ** *

    **

    ** *

    *

    ***

    *

    AC

    TU

    AL

    S

    LO

    TL

    OC

    AT

    ION

    T

    BD

    Sele

    ct

    Me

    nu

    DG

    MZ

    ero

    Tim

    er

    Pre

    vio

    us

    Sc

    ree

    n

    Insert

    No

    te

    ¹ D

    oe

    s no

    t inclu

    de

    Erro

    r Scre

    en

    s10.0

    1.2

    018

    LE

    AK

    TE

    ST

    AN

    D N

    OT

    ES

    ME

    NU

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Se

    lec

    tP

    res

    s to

    Sa

    ve

    0 to

    50

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    1 to

    25

    4 A

    ve

    rag

    ing

    Size

    Pre

    ss

    to S

    av

    e

    0.0

    to 2

    5.0

    % C

    ha

    ng

    e

    Pre

    ss

    to S

    av

    e

    Insert

    No

    te

    Retu

    rn to