Waters Driver Pack 2020 Release 1
Release Notes
715006967Version 00
Copyright © Waters Corporation 2020All rights reserved
General information
Copyright notice© 2020 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND INIRELAND. ALL RIGHTS RESERVED. THIS DOCUMENT OR PARTS THEREOF MAY NOT BEREPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHER.
The information in this document is subject to change without notice and should not be construedas a commitment by Waters Corporation. Waters Corporation assumes no responsibility for anyerrors that may appear in this document. This document is believed to be complete and accurateat the time of publication. In no event shall Waters Corporation be liable for incidental orconsequential damages in connection with, or arising from, its use. For the most recent revisionof this document, consult the Waters website (www.waters.com).
TrademarksACQUITY™ is a trademark of Waters Corporation.
ACQUITY UPC2™ is a trademark of Waters Corporation.
ACQUITY UPLC™ is a trademark of Waters Corporation.
APC™ is a trademark of Waters Corporation.
Empower™ is a trademark of Waters Corporation.
LAC/E™ is a trademark of Waters Corporation.
MassLynx™ is a trademark of Waters Corporation.
nanoACQUITY™ is a trademark of Waters Corporation.
SAT/IN™ is a trademark of Waters Corporation.
THE SCIENCE OF WHAT'S POSSIBLE™ is a trademark of Waters Corporation.
Waters™ is a trademark of Waters Corporation.
Windows® is a registered trademark of Microsoft Corporation in the US and/or other countries.
Xevo™ is a trademark of Waters Corporation.
All other trademarks are property of their respective owners.
December 10, 2020, 715006967 Ver. 00Page ii
Table of contents
General information........................................................................................................ii
Copyright notice...................................................................................................................................... ii
Trademarks ............................................................................................................................................ii
Waters Driver Pack 2020 Release 1...............................................................................5
What's new in this Driver Pack 2020 R1 release ...................................................................................6Windows support............................................................................................................................. 6ACQUITY UPC2 system Windows 10 updates................................................................................ 6ACQUITY UPC2 system startup improvements...............................................................................6UPC2 Binary Solvent Manager (ccBSM) updates............................................................................6Instrument drivers updated in DP 2020 R1......................................................................................7Instrument drivers added to the DP 2020 R1 media........................................................................7Supported chromatography data software.......................................................................................8
General information and support details................................................................................................ 9Contact Waters Technical Service................................................................................................... 9DP 2020 R1 installation media.........................................................................................................9Waters recommendation on choosing a Typical installation.......................................................... 10Backwards compatibility.................................................................................................................10Driver pack compatibility with Waters enterprise systems............................................................. 102465 Electrochemical Detector support.........................................................................................10WPC support..................................................................................................................................10Anti-virus considerations................................................................................................................10Compliance recommendations.......................................................................................................11
Instrument drivers.................................................................................................................................13Solvent managers .........................................................................................................................13Sample managers..........................................................................................................................16Column modules............................................................................................................................18Fraction collector............................................................................................................................25Accessories....................................................................................................................................26
Fixed issues in this release.................................................................................................................. 27Solvent managers..........................................................................................................................27Fraction collector............................................................................................................................27
Known issues in this release................................................................................................................ 28General.......................................................................................................................................... 28
December 10, 2020, 715006967 Ver. 00Page iii
Detectors........................................................................................................................................28Sample managers..........................................................................................................................28Solvent managers..........................................................................................................................29Column manager........................................................................................................................... 30Third-party chromatography data software defects....................................................................... 30
December 10, 2020, 715006967 Ver. 00Page iv
Waters Driver Pack 2020 Release 1
Driver Pack 2020 Release 1 (DP 2020 R1) supports instrument drivers for deployment ofinstrument control software (ICS) and firmware required to control Waters instruments. Theinstrument drivers and firmware in DP 2020 R1 support these analytical systems:
• Alliance HPLC System
• ACQUITY UPLC System
• ACQUITY UPLC H-Class Series System
• ACQUITY UPLC H-Class Bio Series System
• ACQUITY UPLC I-Class Series System
• ACQUITY Arc Systems for HPLC and UHPLC
• ACQUITY Arc Bio Systems
• ACQUITY Advanced Polymer Chromatography (APC) System
• ACQUITY UPC2 System
Additionally, this release supports Empower 3, MassLynx with compatible SCNs, the stand-aloneconsole, and third-party chromatography data software (CDS). For details, see Supportedchromatography data software (Page 8).
DP 2020 R1 supports the following operating systems:
• Windows 10 (64-bit) Enterprise and Professional editions
• Windows Server 2008 R2 Standard edition
• Windows Server 2012 R2 Standard edition
• Windows Server 2016 Standard edition
Requirement: DP 2020 R1 uses Microsoft .NET Framework version 3.5. Ensure thatMicrosoft .NET Framework version 3.5 is installed and enabled.
Restrictions:• The following operating and data systems are not supported in this release: Windows 7,
Windows XP, Windows 8.1, Windows Server 2003, and Empower version 2154 and earlierversions.
• The 2420 ELS detector will not operate after you install DP 2020 R1. The solution is to use adedicated Empower workstation using version 1.40 of the ICS.
• The ACQUITY UPLC M-Class System, ACQUITY UPLC M-Class System with HDXTechnology, ionKey/MS Systems, and nanoACQUITY UPLC Systems work with MassLynxsoftware and third-party chromatography software only. Empower does not support thesesystems.
December 10, 2020, 715006967 Ver. 00Page 5
See also:• For detailed installation instructions, see the Waters Driver Pack 2020 Release 1 Installation
and Configuration Guide .
• For system configuration information, see the appropriate chromatography data software'sinstallation guide, for example, the Empower 3 Installation, Configuration, and UpgradeGuide. For MassLynx, see the MassLynx 4.2 online Help and the appropriate SCN releasenotes.
What's new in this Driver Pack 2020 R1 releaseFor a list of specific issues that were fixed in DP 2020 R1, see Fixed issues in this release (Page27).
Windows supportThe DP 2020 R1 driver pack is compatible with the Windows 10, 64-bit, Enterprise orProfessional edition operating system only. Windows 7 is not supported with DP 2020 R1.However, DP 2020 R1 is compatible with LAC/E modules running Windows 7 with driver packsreleased prior to DP 2020 R1.
ACQUITY UPC2 system Windows 10 updatesThis release adds Windows 10 support for ACQUITY UPC2 systems.
ACQUITY UPC2 system startup improvementsThis release improves system startup as follows:
• All modules installed in the ACQUITY UPC2 system now display in the System Startupwindow.
• Users can set the Convergence Chromatography Manager's Automatic Back PressureRegulator (ABPR) pressure from the System Startup window using the Equilibrate toMethod tab.
UPC2 Binary Solvent Manager (ccBSM) updates• Adds improved chiller management.
• Easily power-on and power-off the ccBSM pump head chillers.
• Set the pump head chillers to power-off after a specified idle time.
• Displays ACQUITY UPC2 delta system pressure on the control panel.
December 10, 2020, 715006967 Ver. 00Page 6
• Users can assess the solvent manager back pressure stability of the ACQUITY UPC2
system.
• ACQUITY UPC2 system delta pressure continuously updates over the span of oneminute.
Instrument drivers updated in DP 2020 R1The following table provides details on the updated instrument drivers for this release. SeeInstrument Drivers (Page 13) for a complete list of the current firmware and drivers.
Table 1: Software changes in this release
Modulename
Part number Systems ICS name inDeploymentManager
NewICSversion
Firmwareupdated
ACQUITYUPC2 BinarySolventManager(ccBSM)
186015036 ACQUITYUPC2
UPC2 Binary SolventManager
1.75 No
ACQUITYUPC2
ConvergenceManager(ccM)
186015037 ACQUITYUPC2
ConvergenceManager
1.75 No
ACQUITY ArcQuaternarySolventManager-R(QSM-R), ArcBioQuaternarySolventManager-R(bioQSM-R)
186017000,186017011
ACQUITY Arcand ACQUITYArc Bio
Quaternary SolventManager-R
1.75 No
ACQUITY ELS 186015027 All ACQUITYUPLC
ELS Detector 1.75 No
Instrument drivers added to the DP 2020 R1 mediaThis release adds instrument drivers for the modules shown in the following table.
Note: See Instrument Drivers (Page 13) for additional details including part numbers, Waterssystems that use the modules, and checksums where applicable.
December 10, 2020, 715006967 Ver. 00Page 7
Table 2: Instrument drivers now available on the media
Waters module name ICS version
25X25 Quaternary Gradient Module (QGM) 1.10
2432 Conductivity Detector (CD) 1.65
Fraction Collector (WFC) III 1.10
Supported chromatography data softwareDP 2020 R1 supports the following software:
• Empower 3 Feature Release 4 and Feature Release 5
• MassLynx software with compatible SCNs (see MassLynx supported mass spectrometers(Page 8) or MassLynx supported MSs with M-Class (Page 9))
• Stand-alone console
DP 2020 R1 focuses on enhancements for the ACQUITY UPC2 system. The system supportsthird-party chromatography data software (CDS). DP 2020 R1 and ACQUITY UPC2 were notevaluated with third party CDS.
Note: It is expected that any Waters ICS changes remain compatible with third party CDS. Referto the appropriate documentation for known issues specific to third party CDS.
MassLynx supported mass spectrometersThe following table lists the mass spectrometers tested with this release. Waters expects thatother mass spectrometers that were not tested are compatible with this release. For example,they function as expected after you install the appropriate Software Change Note (SCN).
Mass spectrometers supported Systems
QDa • Alliance HPLC
• ACQUITY UPLC
• ACQUITY UPLC H-Class Series
• ACQUITY UPLC I-Class Series
• ACQUITY Arc
• ACQUITY Arc Bio
• ACQUITY UPC2
SQD2 • ACQUITY UPLC
• ACQUITY UPLC H-Class Series
December 10, 2020, 715006967 Ver. 00Page 8
Mass spectrometers supported Systems
• ACQUITY UPLC I-Class Series
• ACQUITY UPC2
Mass spectrometers with M-Class systems supported by MassLynxThe following table lists the mass spectrometers and SCNs tested with M-Class systems. Watersexpects that other mass spectrometers not tested are compatible with this release, meaning theywill function as expected after you install the appropriate SCNs. Note that Waters supports theability to “switch” to DP 2019 R2 if it is co-installed. However, you cannot run two different driverpack versions at the same time.
Note: For DP 2020 R1, Mass spectrometers supported (but not tested) include those runningMassLynx 4.2 with Windows 10.
Mass spectrometers tested SCN
Xevo G2-XS QTof 966
Xevo TQ-XS 1007
Xevo TQ-S 986
General information and support detailsThis section provides important details related to Waters driver packs.
Contact Waters Technical ServiceIf you are located in the USA or Canada, report malfunctions or other problems to WatersTechnical Service (800 252-4752). If you are located elsewhere, phone the Waters corporateheadquarters in Milford, Massachusetts (USA), or contact your local Waters subsidiary. TheWaters' Web site includes phone numbers and e-mail addresses for Waters locations worldwide.
DP 2020 R1 installation mediaThe media that ships with DP 2020 R1 includes one DVD. The Analytical DVD provides theDeployment Manager that installs all modules supporting analytical systems. It also includes theICS for MS control in Empower. For detailed instructions, see the Waters Driver Pack 2020Release 1 Installation and Configuration Guide .
See also: For a complete list of supported modules, see Instrument drivers (Page 13).
December 10, 2020, 715006967 Ver. 00Page 9
Waters recommendation on choosing a Typical installationWaters recommends using the Typical installation option. If you have specific questions, contactWaters Technical Service.
Backwards compatibilityWaters recommends that you run instruments at the current firmware version and that theinstrument firmware matches the driver pack. To upgrade, install the new driver pack, and thenupdate the firmware to the latest version. Although not recommended, users can install the newdriver pack without updating the instrument firmware, providing that the firmware version alreadyinstalled is included in a driver pack released within 24 months of the newly installed driver pack.Exceptions are described in the following sections.
Note: You can use DP 2020 R1 with firmware that was available in DP 2019 R1. Any featuresdependent on new firmware are not supported. Waters recommends upgrading to the latestfirmware to enable the expected features.
Driver pack compatibility with Waters enterprise systemsIn Empower, the DP 2020 R1 driver pack supports control of ACQUITY UPLC H-Class Series, I-Class Series, Arc, and UPC2 systems on the same workstation. You can connect up to foursystems to the workstation at the same time.
In MassLynx, the DP 2020 R1 driver pack supports control of ACQUITY UPLC H-Class Series, I-Class Series, M-Class, Arc, and UPC2 systems on the same workstation. You can connect oneLC system to the workstation at a time.
2465 Electrochemical Detector supportDriver pack DP 2020 R1 does not support the 2465 Electrochemical Detector (ECD). Thedetector requires driver pack version 1.00.
WPC supportDriver pack DP 2020 R1 supports WPC v3.30 for ACQUITY UPLC systems with 2D technologyand ACQUITY UPLC M-Class systems.
Requirement: WPC v3.30 does not support existing Prep SFC systems. You must use theSFC-specific WPC v2.40 with Prep SFC systems.
Anti-virus considerationsSome real-time virus scanners mistake normal data acquisition and instrument control for virusactivity, and thus interfere with proper operations. Full-system scans and live updates can benetwork-intensive, disk-intensive, and CPU-intensive, and they can also interfere with normal
December 10, 2020, 715006967 Ver. 00Page 10
data acquisition. Schedule scans and updates for idle times when data acquisition does notoccur.
Certain anti-virus program features like "intrusion prevention" and "tamper protection" can alsointerfere with normal operation. Disable them as well.
Empower installationsFor Empower software installations, exclude the Empower installation folder (usuallyC:\Empower) and its sub-folders.
MassLynx installationsFor MassLynx installations, exclude these folders:
• For 64-bit computers: C:\Program Files (x86)\Waters Instruments, and its sub-folders.
• The MassLynx installation folder, usually C:\Masslynx, and its sub-folders.
Antivirus considerations - Stand-alone installationsFor stand-alone console software installations, exclude C:\Program Files (x86)\WatersInstruments and its sub-folders.
Compliance recommendationsAny time you install, change, or uninstall software or system modules in a regulated environment,Waters recommends that you follow your organization's approved standard operatingprocedures.
A risk-based review may assist you in a regulated environment to evaluate changes detailed inthe release notes. Using company SOPs, determine if any documentation updates andrequalification of the system modules, chromatographic system, or chromatographic data system(CDS) are required.
Instrument update classification (minor update)Waters considers this update to be a minor change from previous versions. This update isdesigned to not alter the physical specifications of instruments or systems. Waters expects thatmethods run and data generated on instruments before and after the update will be consistent.As with any change, you should evaluate the impact of this update to your instruments orsystems.
Software requalification optionsConsider using the requalification options outlined below to verify software installation and correctoperation:
December 10, 2020, 715006967 Ver. 00Page 11
• To confirm that the new files loaded properly, consider performing a software installationqualification.
• To confirm the operation of the newly loaded software, consider performing an operationalqualification (user or vendor) for the updated software installation.
• To determine if additional testing is required, consider evaluating the changes in the softwarerelease, to assess the risk associated with the installation. Depending on the risk, it may beappropriate to perform existing, updated, or new software tests. These tests may be knownas performance qualification tests, user acceptance tests, verification tests, or validationtests.
To assist you, if required, Waters provides various levels of Qualification (or Compliance)Services and validation consultancy through our Professional Services organization.
When multiple, identical systems are involved, consider a risk-based approach to qualificationactivities.
Instrument requalification options after software or firmware changeConsider using the requalification options outlined below to verify hardware installation andcorrect instrument operation:
• To confirm that the firmware files on the system modules were installed correctly, compare thechecksum values in the product release notes to the checksum values displayed in theconsole.
• To confirm instrument system operation with any newly loaded software, driver, or firmware,consider performing an operational qualification for the updated instrument system.
• To confirm performance, control, and communications of the instrument system, considerconducting a performance qualification (user or vendor) or system suitability test.
Requalification with Waters' Total Assurance PlansThe Waters' Total Assurance Plan (TAP) with System Qualification Option covers upgrades andrequalification of the instrument driver, software, firmware, or hardware in these cases:
• During yearly requalification, as provided in the plan.
• If installing this release is required for operation of a new module or system, wherequalification of the new module or system is covered by the plan.
Requalification of the CDS software and computers after a driver upgrade may or may not beincluded in your TAP.
Review your TAP to determine which services are covered and which are not covered. Forsituations not covered by the plan, Waters can perform the qualification, but additional chargeswill apply.
December 10, 2020, 715006967 Ver. 00Page 12
Instrument driversThe tables in this section list the software and firmware drivers included in DP 2020 R1.
You can determine the currently installed ICS and firmware versions on a system from the Waters Console. From the Help menu, select AboutConsole. In an Empower Enterprise environment, the About Console dialog box shows the ICS versions on the client and the LAC/E module.
Tip: To confirm that the firmware files were installed correctly, compare the checksum values in the table below to the checksum values displayedin the console. You can do this by selecting the module in the system tree, and then selecting Configure > View Module Information.
Solvent managersTable 3: Solvent managers
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
ACQUITY UPLCBinary SolventManager (BSM),ACQUITY UPLCBinary SolventManager (BSM)PLUS
186015001,186015000,186015082
ACQUITY UPLCand ACQUITYUPLC I-ClassSeries, ACQUITYUPLC H-ClassSeries
Binary SolventManager
1.73 Binary SolventManager [V1.72]
0x19251B64
ACQUITY UPC2
Binary SolventManager (ccBSM)
186015036 ACQUITY UPC2 UPC2 BinarySolvent Manager
1.75 ACQUITY UPC2Binary SolventManager
0x1090528B
ACQUITY UPC2
ConvergenceManager (ccM)
186015037 ACQUITY UPC2 ConvergenceManager
1.75 ACQUITYConvergenceManager
0x72815285
December 10, 2020, 715006967 Ver. 00Page 13
Table 3: Solvent managers (continued)
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
ACQUITY UPLCQuaternarySolvent Manager(QSM),bioQSM,QSM PLUS,bioQSM PLUS,QSM-XR PLUS,andbioQSM-XR PLUS
186015018 (QSM),186015041(bioQSM),186015080 (QSMPLUS),186015081(bioQSM PLUS),186015083 (QSM-XR PLUS),186015084(bioQSM-XRPLUS)
ACQUITY UPLCH-Class Seriesand ACQUITYUPLC H-Class BioSeries
QuaternarySolvent Manager
1.73 QuaternarySolvent Manager[V1.72]
0x93EFE762
ACQUITY ArcQuaternarySolvent Manager-R (QSM-R), ArcBio QuaternarySolvent Manager-R (bioQSM-R)
186017000,186017010
ACQUITY Arc andACQUITY Arc Bio
QuaternarySolvent Manager-R
1.75 QuaternarySolvent Manager-R [V1.72]
0x897ACE21
ACQUITY UPLCIsocratic SolventManager (ISM)andACQUITY APCIsocratic SolventManager (p-ISM)
186015019,186015050
All ACQUITYUPLC andACQUITYAdvanced PolymerChromatography
Isocratic SolventManager
1.73 Isocratic SolventManager [V1.72]
0x9AF42CBA
December 10, 2020, 715006967 Ver. 00Page 14
Table 3: Solvent managers (continued)
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
ACQUITY UPLCµBinary SolventManager (µBSM)
186016003 ACQUITY UPLCM-Class
µBinary SolventManager
1.74 microBinarySolvent Manager[V1.56]
0x0AE1E908
ACQUITY UPLCnanoBinarySolvent Manager(nanoBSM)
186016002 nanoACQUITY nanoBinarySolvent Manager
1.74 microBinarySolvent Manager[V1.56]
0x0AE1E908
ACQUITY UPLCµAuxiliary SolventManager (ASM)
186016022 ACQUITY UPLCM-Class
Auxiliary SolventManager
1.74 microBinarySolvent Manager[V1.56]
0x0AE1E908
ACQUITY UPLCAuxiliary SolventManager (ASM)
186016021 nanoACQUITY Auxiliary SolventManager
1.74 microBinarySolvent Manager[V1.56]
0x0AE1E908
2545 QuaternaryGradient Module(QGM)2535 QuaternaryGradient Module2555 QuaternaryGradient Module
186004556,186004557,186255500
HPLC Waters 25X5QuaternaryGradient Module
1.10 Waters 25X25QuaternaryGradient Module
220C2ECD
December 10, 2020, 715006967 Ver. 00Page 15
Sample managersTable 4: Sample managers
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
ACQUITY UPLCSample Manager(SM)
186015005,186015006
ACQUITY UPLC Sample Manager 1.73 Sample Manager[v1.65]
0x42C0E8CB
ACQUITY UPLCSample Manager –Flow ThroughNeedle (SM-FTN,bioSM-FTN, SM-FTN PLUS, andbioSM-FTN PLUS)
186015017,186015040,186015046,186015085 (H-Class PLUS),186015086 (H-Class Bio PLUS),186015088 (I-Class PLUS)
ACQUITY UPLCH-Class Series, H-Class Bio Series,andI-Class Series
Sample Manager-FTN
1.74 Sample Manager-FTN [v1.71]
0x285858D7
ACQUITY UPLCSample Manager –Fixed Loop Series(SM-FL and SM-FLPLUS) andACQUITY UPC2
Sample Manager –Fixed Loop (ccSM-FL)
186015060,186015035,186015087 (SM-FL PLUS)
ACQUITY UPLC I-Class Series andACQUITY UPC2
Sample Manager 1.73 Sample Manager-FL [v1.70]
0x30B5761D
ACQUITY ArcSample Manager –Flow Through
186017001,186017007,186017011
ACQUITY Arc Sample ManagerFTN-R
1.73 Sample ManagerFTN-R [v1.71]
0xA7F6FD36
December 10, 2020, 715006967 Ver. 00Page 16
Table 4: Sample managers (continued)
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
Needle-R, with orwithout heating/cooling (SM FTN-R)ACQUITY Arc BioSample Manager –Flow ThroughNeedle-R (bioSM-FTN-R)
ACQUITY APCSample Manager –Flow ThroughNeedle (SM-pFTN)
186015051 ACQUITYAdvanced PolymerChromatography
Sample Manager-FTN
1.73 Sample ManagerAPC [v1.60]
0x31502C7A
Waters 2707Autosampler, withor without Cooler
186004462 or186004463 (w/Cooler)
HPLC W2707 1.30 W2707Autosampler
0xE020ED1E
ACQUITY UPLCSample Organizer(SO) andACQUITY UPLCSample Organizerfor rotary tray-stylesample managers
186015020,186015021,186015014
All ACQUITYUPLC, H-ClassSeries, H-ClassBio Series, I-ClassSeries, and Arc
N/A Controlled by theSample Manager
Sample Organizer[v1.60]
0x96FA9AB0
ACQUITY UPLCnanoSample
186016007 nanoACQUITY µSample Manager- FL
1.74 microSampleManager [V1.56]
0x3BFB8029
December 10, 2020, 715006967 Ver. 00Page 17
Table 4: Sample managers (continued)
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
Manager(nanoSM)
ACQUITY UPLCmicroSampleManager-FL (µSM-FL)
186016012 ACQUITY UPLCM-Class
µSample Manager- FL
1.74 microSampleManager-FL[V1.56]
0x301EDC11
ACQUITY UPLCHDX Manager
186016011 nanoACQUITYUPLC System withHDX Technology,ACQUITY UPLCM-Class Systemwith HDXTechnology
HDX Manager 1.74 HDX Manager[V1.57]
0x3F56CB0D
Column modulesTable 5: Column heaters
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
ACQUITY UPLCColumn Heater
N/A ACQUITY UPLC N/A Controlled by thesample manager
Controlled by thesample manager
N/A
December 10, 2020, 715006967 Ver. 00Page 18
Table 5: Column heaters (continued)
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
ACQUITY UPLCColumn Heater/Cooler (CHC)
186015008 ACQUITY UPLC Column Manager Controlled by thesample manager
Controlled by thesample manager
0x8E9D1DC0
ACQUITY UPLCColumn Heater,Active (CH-A)
186015042 ACQUITY UPLC H-Class Series andH-Class Bio Series,ACQUITY UPLC I-Class Series
N/A Controlled by thesample manager
Controlled by thesample manager
N/A
ACQUITY UPLCHigh-TemperatureColumn Heater(HTCH)
186015010 ACQUITY UPLC N/A Controlled by thesample manager
Controlled by thesample manager
N/A
Table 6: Column managers
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
ACQUITY UPLCColumn Manager
186015007,186015009
ACQUITY UPLC Column Manager 1.73 Column Manager[V1.40.74]
0x8E9D1DC0
ACQUITY UPLCColumn Manager,Active (CM-A)
186015043,186015038
ACQUITY UPLCH-Class Seriesand H-Class BioSeries, ACQUITYUPLC I-ClassSeries,
Column Manager 1.73 Column Manager(Active) [V1.69]
0x17AC4A76
December 10, 2020, 715006967 Ver. 00Page 19
Table 6: Column managers (continued)
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
ACQUITY UPC2
ACQUITY UPLCColumn ManagerAuxiliary (CM-AUX)
186015049,186015039
ACQUITY UPLCH-Class Seriesand H-Class BioSeries
N/A Controlled by theColumn Manager,Active (CM-A)
Controlled by theColumn Manager,Active (CM-A)
N/A
ACQUITY Single-Zone ColumnManager (CM-S)
186015054 ACQUITYAdvanced PolymerChromatography
Column Manager 1.73 Column Manager(Active-S) [V1.62]
0xC1CCF89D
Trap ValveManager (TVM)
186016016 ACQUITY UPLCM-Class
Trap ValveManager
1.74 Trap ValveManager [V1.53]
0x4026CB5B
Table 7: 30-cm Column heaters and managers
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
ACQUITY UPLC30-cm ColumnHeater (30-cm CH)
186017008 ACQUITY Arc N/A Controlled by theSample Manager
Controlled by theSample Manager
N/A
ACQUITY UPLC30-cm ColumnHeater/Cooler (30-cm CHC)
186015011,186017009
All ACQUITY UPLCand ACQUITY Arc
N/A Controlled by theSample Manager
Controlled by theSample Manager
N/A
ACQUITY UPLC30-cm Column
186015045 ACQUITY UPLC H-Class Series and
N/A Controlled by theSample Manager
Controlled by theSample Manager
N/A
December 10, 2020, 715006967 Ver. 00Page 20
Table 7: 30-cm Column heaters and managers (continued)
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
Heater with ActiveSolvent Preheating(CH-30A)
H-Class Bio Series,ACQUITY UPLC I-Class Series, andArc
ACQUITY UPLC30-cm Single-ZoneColumn Manager(CM-30S)
186015056 ACQUITYAdvanced PolymerChromatography
Column Manager 1.73 Column Manager(30S) [V1.62]
0xBED9D41A
Detectors
Table 8: Detectors
Module name Part number Systems ICSname inDeploymentManager
ICSversion
Firmware name inDeployment Manager
Checksum
ACQUITY TUV 186015028 All ACQUITYUPLC
TUVDetector
1.70 TUV Detector (1)[V1.70]
0xB9E50966
ACQUITY TUV, thermallyenhanced
186015031 All ACQUITYUPLC
TUVDetector
1.70 TUV Detector (2)[V1.70]
0xD4B0C7BC
ACQUITY PDA 186015026 All ACQUITYUPLC
PDADetector
1.70 PDA Detector (1)[V1.65] a
0xF7835E17
ACQUITY PDA, thermallyenhanced
186015032 All ACQUITYUPLC
PDADetector
1.70 PDA Detector (2)[V1.70] a
0xDF9B3A2E
December 10, 2020, 715006967 Ver. 00Page 21
Table 8: Detectors (continued)
Module name Part number Systems ICSname inDeploymentManager
ICSversion
Firmware name inDeployment Manager
Checksum
ACQUITY PDAe λ
186015030 All ACQUITYUPLC
PDADetector
1.70 PDA Detector (1)[V1.65] a
0xF7835E17
ACQUITY PDAe λ, thermally enhanced
186015033 All ACQUITYUPLC
PDADetector
1.70 PDA Detector (2)[V1.70] a
0xDF9B3A2E
ACQUITY PDA TS 186015053 All ACQUITYUPLC
PDADetector
1.70 PDA Detector (2)[V1.70] a
0xDF9B3A2E
ACQUITY UPC2 PDA 186015034 All ACQUITYUPLC
PDADetector
1.70 PDA Detector (2)[V1.70] a
0xDF9B3A2E
ACQUITY FLR 186015029 All ACQUITYUPLC
FLRDetector
1.69 FLR Detector[V1.69]
0xD99F06EF
ACQUITY RI 186015070 All ACQUITYUPLC
RIDetector
1.40 RI Detector[1.40]
0xAA39BF4C
ACQUITY ELS 186015027 All ACQUITYUPLC
ELSDetector
1.75 ELS Detector[1.40] b
0x216F9370
2489 UV-Vis 186002487,186248900,186017002
Alliance HPLCand ACQUITYArc
2489UV/VisDetector
1.73 2489 Detector[V1.69]
0xBAB5BE45
2998 PDA 186002998,186299800,186017003
Alliance HPLCand ACQUITYArc
2998 PDADetector
1.73 2998 Detector[V1.70] a
0xED702600
December 10, 2020, 715006967 Ver. 00Page 22
Table 8: Detectors (continued)
Module name Part number Systems ICSname inDeploymentManager
ICSversion
Firmware name inDeployment Manager
Checksum
2475 FLR 186002475,186247500,186017006
Alliance HPLCand ACQUITYArc
2475 FLRDetector
1.73 2475 Detector[V2.20]
0xDBCEAB47
2414 RI 186241400,186241401,186017004
Alliance HPLCand ACQUITYArc
2414 RIDetector
1.73 2414 Detector[V2.10]
0x5038C65E
2424 ELS 186002424,186242400,186017005
Alliance HPLCand ACQUITYArc
2424 ELSDetector
1.73 N/Ac 0x123F2F1E
2432 CD 186243200 Alliance HPLC,ACQUITY Arcand ACQUITYUPLC H-ClassSeries,ACQUITY UPLCH-Class BioSeries Systems
2432ConductivityDetector
1.65 N/A N/A
a. Do not downgrade PDA firmware if it is at version 1.69 or higher. Contact Expert Center.
b. Contact Waters Expert Center for the latest ELS firmware, version 1.65.
c. Contact Waters Expert Center for the latest 2424 ELS firmware, version 1.65.
December 10, 2020, 715006967 Ver. 00Page 23
Table 9: Empower only mass detectors and spectrometers
Module name Part number Systems ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
ACQUITY QDa 186008511,186006511
Alliance HPLC, AllACQUITY UPLC,and ACQUITY Arc
QDa Detector 1.72 N/A N/A
ACQUITY TQ 176001263 All ACQUITY UPLC Waters TQDetector
1.44 N/A N/A
ACQUITY SQ 176000872 All ACQUITY UPLC Waters SQDetector
1.44 N/A N/A
ACQUITY SQD2 186005832 Alliance HPLC, AllACQUITY UPLC,and ACQUITY Arc
Waters SQDetector 2
1.52 N/A N/A
December 10, 2020, 715006967 Ver. 00Page 24
Fraction collectorTable 10: Fraction collector
Module name Part number System ICS name inDeploymentManager
ICS version Firmware namein DeploymentManager
Checksum
Waters FractionManager –Analytical
186006857 HPLC and UPLC Fraction Manager– Analytical
1.66 Fraction Manager– Analytical[V1.60].sft
0x31d1e50e
Waters FractionCollector III
186001878 HPLC and UPLC WFCIII 1.10 W261E (firmwareversion not shownin the DeploymentManager)
N/A
December 10, 2020, 715006967 Ver. 00Page 25
Accessories
Module name Systems ICS name inDeploymentManager
ICS version Checksum
Local ConsoleController
All ACQUITYUPLC
Local ConsoleController
1.60 N/A
ColumnsCalculator
HPLC and UPLC ColumnsCalculator
2.00 N/A
Waters PumpControl Software
Waters ACQUITYsolvent managers,Waters 25X5QuaternaryGradient modules
Waters PumpControl Software
3.30 N/A
e-SAT/IN All ACQUITYUPLC
Waters Empowere-SAT/IN Software
1.3 N/A
DeploymentManager
N/A N/A 3.1.258.5 N/A
December 10, 2020, 715006967 Ver. 00Page 26
Fixed issues in this releaseThis section lists the problems resolved in this release. The numbers identify issues that Waterspersonnel monitor in the legacy system change request tracking tool (marked with SCR-), in the newissue tracking tool (marked with INS-), and customer-reported issues tracking tool (marked with CRI-).
Solvent managers
INS-7432
Instrument or software component: ccBSM
For systems running the Windows 10 operating system with a ccBSM module and no other solventmanager, Stop Flow is always disabled. For systems that include a ccBSM module and another solventmanager, Stop Flow is enabled. However, when you click Stop Flow, the flow stops for the alternatesolvent manager only.
Fixed in: DP 2020 R1
INS-7433
Instrument or software component: ccM
When a leak occurs, the Convergence Manager does not stop the flow of the ccBSM module. When anyother module detects a leak, it stops the flow of the ccBSM module.
Fixed in: DP 2020 R1
INS-10011, CRI-520
Instrument or software component: QSM-R
Previously, in Japanese versions of Empower 3 software, the QSM-R status LED remained unlit insteadof showing a green status during normal operation.
Fixed in: DP 2020 R1
Fraction collector
INS-14757, CRI-2082
Instrument or software component: WFC III
In some cases, a WFC III module operating with driver version 1.10 and Windows 10 stopped respondingcausing Empower to display the Chromatographic system lost error message. The issue isresolved. The system now performs collection during a run without losing connection to the instruments.
Fixed in: DP 2020 R1
December 10, 2020, 715006967 Ver. 00Page 27
Known issues in this releaseFor a comprehensive list of all known issues since Driver Pack 4, see Waters Driver Pack Known andFixed Issues, 715005660 Version 5.
General
Opening Help in Windows 10Some Windows Updates prevent the correct Help topic from appearing when you open Help from theConsole and instrument control panels. A message stating that the page cannot be displayed appears.The most recent Windows Updates correct the issue.
If a subsequent Windows Update prevents the correct Help topic from appearing, you can search fortopics by opening the Help window, and then use the Search tab in the left-hand pane.
INS-15650
Instrument or software component: ACQUITY console client
In Empower environments, sometimes the console client becomes unresponsive when the user movesmodules to the Hidden Modules pane in the Arrange Instrument Modules window.
Reported in: DP 2020 R1
Solution: Restart the console from the control panel.
Detectors
INS-14328
Instrument or software component: QDa
The QDa stops producing logs when you restart the QDa while the Run Samples window, the console, orboth are open.
Reported in: DP 2020 R1
Solution: Close, and then open all Run Samples windows and the console.
Sample managers
INS-16545
Instrument or software component: Multiple instruments
During an installation, the Deployment Manager sometimes displays an error that an instrument installfailed.
December 10, 2020, 715006967 Ver. 00Page 28
Reported in: DP 2020 R1
Solution: Initiate another installation of the driver pack. If the device is not displayed on the list ofinstruments to be installed, you can ignore the error. The instrument installed correctly.
Solvent managers
INS-12543
Instrument or software component: ISM
A CO2 pressure drop may cause the ccM to send a Stop Flow severity alarm to the ccBSM through theEthernet cable. However, no Stop Flow severity alarm is sent to the ISM.
Reported in: DP 2020 R1
Solution: Install the contact closure cables between the UPC2 to stop the flow input of the ISM.
INS-13963
Instrument or software component: ccBSM
An over-pressure alarm occurs when a solvent select valve change is programmed together with acolumn selection change between the secondary CM-30S to the primary CM-30S.
Reported in: DP 2020 R1
Solution: Create a method that changes the solvent select valve to the desired setting and leaves thecolumn position unchanged. Set the ccBSM flow rate to 0.00 mL/min in this method. Insert this method asan equilibrate function in the sample list before the Method Set with the intended initial conditions. In thisway, the solvent selection valve position is changed in the first method, and the column position ischanged in the second.
INS-15478
Instrument or software component: ISM
The ISM menu items in the console do not show after running an ISM leak test.
Reported in: DP 2020 R1
Solution: Restart the computer.
INS-15736
Instrument or software component: ccM
When the ccM encounters a Leak detected error, the System Status LED indicator remains greenwhen it should display red. This behavior only occurs when solvent is flowing through the ccBSM.
Reported in: DP 2020 R1
Solution: None.
December 10, 2020, 715006967 Ver. 00Page 29
Column manager
INS-14558
Instrument or software component: CM-30S
An ACQUITY UPC2 system with a CM-30S installed equilibrates, but is unresponsive and does notcontinue to the next injection.
Reported in: DP 2020 R1
Solution: Add an equilibrate line to the sample set before each column switch. Set the run time to 10seconds and the flow to zero.
Third-party chromatography data software defects
INS-1680, INS-14792, INS-14794, INS-14795, INS-14797, INS-14798, INS-14799, INS-14800,INS-14801Users running Chromeleon version 7.2.8 or 7.2.10 may experience this issue when data is collected usingan ACQUITY system. In the Sequence Properties dialog field titles display as heavy, illegible text.Chromatography Studio is missing the title bar, and functionality like close window (right top corner x) andminimize are not available.
Reported in: DP 2020 R1
Solution: After saving the method, restart the Chromeleon console.
Waters known issue identifier Waters module name
INS-1680 PDA
INS-14792 SM-FTN
INS-14794 ISM
INS-14795 SM FTN-R
INS-14797 FLR
INS-14798 30-cm CH
INS-14799 RI
INS-14800 TUV
INS-14801 TVM
December 10, 2020, 715006967 Ver. 00Page 30