1
TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2016 Waters Corporation INTRODUCTION Biotherapeutics are increasingly becoming a significant part of the pharmaceutical arsenal as more and more companies work towards using them individually or in combination with other large or small molecules as drugs of choice. This trend manifests itself in the growing number of bioanalytical laboratories across the globe incorporating technological and scientific expertise required to deal with the complexities such analyses bring. The diversity within biotherapeutics ranges from small linear or cyclic peptides, all the way to complex monoclonal antibodies and antibody drug conjugates. The instruments of choice for performing these types of analysis are largely tandem quadrupole mass spectrometers, like the Waters Xevo TQ-S. Because of the nature of biotherapeutics, one of the biggest analytical challenges for scientists is to be able to distinguish and separate out analyte of interest from a very complex matrix background. The resultant increase in signal to noise allows for lower LLOQ’s. Sample preparation tools significantly help with this problem. Tandem quadrupole instruments, by principle, further help in attaining this goal by better focusing the ions of interest and removing the neutrals and other matrix ions. Biotherapeutic peptides and monoclonal antibodies were tested using the Xevo TQ-XS systems. Vancomycin and Infliximab were chosen as candidate molecules to represent some of the common classes of biotherapeutics currently in clinic. IMPROVEMENTS IN SENSITIVITY FOR BIOTHERAPEUTICS USING A XEVO TQ-XS TANDEM QUADRUPOLE MASS SPECTROMETER Nikunj Tanna 1 , Kerri Smith 1 , Leonard Dillon 2 , Mark Roberts 2 , Kelly Doering 1 , Anthony Marcello 1 and Mark Wrona 1 1 Waters Corporation, Milford, MA, USA, 2 Waters Corporation, Wilmslow, USA Figure 8. Stepwave device with a second stage segmented quadrupole METHODS Vancomycin Vancomycin in a branched, tricyclic non-ribosomal peptide antibiotic which was discovered in 1953 and has been used for over 60 years in clinical practice to treat multiple types of infections and inflammation. Vancomycin purchased from Sigma Aldrich was spiked in buffer to obtain a calibration curve from 100 pg/mL to 1 μg/mL. A ACQUITY BEH C18 column was used with generic LC conditions as shown below. The MS was optimized for the Vancomycin transition. MS Conditions: Capillary Voltage—3 kV Cone Voltage—50 V Source Temperature—150°C Desolvation Temperature—400°C Cone Gas Flow—150 L/Hr Desolvation Gas Flow—800 L/Hr Transition—724.8—>144.13 LC Conditions: RESULTS Significant improvements in analyte area counts were observed for the molecules tested across the entire concentration range. For Vancomycin, a >10 fold increase in analyte area counts was observed across the range of 100pg/mL to 1 μg/mL. The signal to noise also increased by >5 across the entire concentration range. A chromatographic comparison between the Xevo TQS and Xevo TQ-XS is shown in the figure below. CONCLUSION The new Xevo TQ-XS tandem quadrupole Mass Spectrometer has shown significant improvements in analyte area counts as well as signal to noise compared to the current best in class Xevo TQ-S. These improvements are seen across a wide variety of small and large biomolecules in buffer as well as in matrix. DISCUSSION Quantitation of proteins using the surrogate peptide approach is rapidly becoming a key part of most bioanalytical laboratories have traditionally focused on performing small molecule quantification. Protein/ peptide quantification brings with it unique challenges which are different from those typically experienced with a small molecule assay. Scientists are having to learn to navigate these challenges quickly. Advancements in instrumentation technologies continue to enhance the quality and speed of data generated. The Xevo TQ-XS instrument has a newly designed Stepwave ionguide. The Stepwave has a second stage featuring a segmented quadrupole design for a focused ion beam and increased ion transmission. Infliximab Infliximab is a chimeric humanized mouse monoclonal antibody therapeutic developed against TNF-α and is indicated for Crohn’s disease, ulcerative colitis, psoriasis, psoriatic arthritis, ankylosing spondylitis and rheumatoid arthritis. Intact Infliximab was spiked into rat serum to generate a calibration curve and QC samples. All samples and QC’s were then digested using the Waters ProteinWorks digest kits. Briefly, the samples were denatured, reduced, alkylated and digested as per the protocol provided in the kit. The protocol has been optimized for use across biological therapeutics or biomarkers in a biological matrix. The simplified protocol reduces method development time significantly and provides scientists with an uncomplicated, kitted approach to digestion which can be very easily standardized and transferred across laboratories. Generic LC and MS conditions were used and the same set of samples were injected into the Waters Xevo TQ-S and the Xevo TQ-XS. SINS peptide was chosen as the primary quantitation peptide. Waters mAb was added as the internal standard and a signature peptide from the Waters mAb was chosen as the internal standard peptide. Figure 7. Average fold change in Area Counts for Infliximab and internal standard peptides Figure 4. Fold change in area counts for Vancomycin across different concentrations Figure 1. Xevo TQ-XS Tandem Quadrupole Mass Spectrometer Figure 5. Average fold change in Area counts and Signal to Noise for Vancomycin Figure 3. Schematic structure of Infliximab Figure 2. Molecular structure of Vancomycin Figure 6. Chromatogram overlay for the same sample injected on Xevo TQ-S and Xevo TQ-XS Time 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 1.70 1.80 1.90 2.00 2.10 2.20 2.30 2.40 2.50 2.60 2.70 2.80 2.90 3.00 3.10 3.20 3.30 3.40 3.50 3.60 3.70 3.80 3.90 4.00 4.10 4.20 4.30 4.40 % 0 100 5 ng-mL MRM of 3 Channels ES+ TIC (Vancomycin) 7.09e4 0.99 Xevo TQ-XS Xevo TQS Concentration – 5ng/mL Injection Volume - 10μL Significant increase in area counts observed on Xevo TQ-XS compared to Xevo TQS For digested Infliximab, data from the most sensitive (SINS) peptide is shown below along with the digested internal standard peptide. On average, there was an increase in the analyte area counts for these peptides between 2-3 fold form a complex digested biological matrix. The ProteinWorks kit used for digestion yielded a simple protocol with a shortened sample preparation time of around 5 hours including reduction, alkylation and digestion Time Flow Rate (μL) %A %B Curve Initial 0.600 95 5 6 0.50 0.600 90 10 6 3.50 0.600 20 80 6 3.60 0.600 5 95 6 4.00 0.600 5 95 6 4.10 0.600 95 5 6 4.50 0.600 95 5 6 0 0.5 1 1.5 2 2.5 3 3.5 Int Std SINS Average Fold Change Average Fold Change Xevo TQ-XS v Xevo TQ-S Proto-type Xevo TQ-XS Xevo TQ-S 0 2 4 6 8 10 12 14 Area Counts Signal to Noise Average Fold Change Average Fold Change Xevo TQ-XS v Xevo TQ-S Proto-type Xevo TQ-XS Xevo TQ-S 0 5 10 15 100 1,000 10,000 100,000 1,000,000 Fold Change Concentration (pg/mL) Fold Change in Area Counts Xevo TQ-XS/Xevo TQ-S Proto-type Xevo TQ-XS Xevo TQ-S

TIC (Vancomycin) - Waters Corporation › webassets › cms › library › docs › ... · TIC (Vancomycin) 7.09e4 0.99 Xevo TQ-XS Xevo TQS Concentration – 5ng/mL Injection Volume

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: TIC (Vancomycin) - Waters Corporation › webassets › cms › library › docs › ... · TIC (Vancomycin) 7.09e4 0.99 Xevo TQ-XS Xevo TQS Concentration – 5ng/mL Injection Volume

TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2016 Waters Corporation

INTRODUCTION

Biotherapeutics are increasingly becoming a

significant part of the pharmaceutical arsenal as more

and more companies work towards using them

individually or in combination with other large or

small molecules as drugs of choice. This trend

manifests itself in the growing number of bioanalytical

laboratories across the globe incorporating

technological and scientific expertise required to deal

with the complexities such analyses bring. The

diversity within biotherapeutics ranges from small

linear or cyclic peptides, all the way to complex

monoclonal antibodies and antibody drug conjugates.

The instruments of choice for performing these types

of analysis are largely tandem quadrupole mass

spectrometers, like the Waters Xevo TQ-S.

Because of the nature of biotherapeutics, one of the biggest analytical challenges for scientists is to be

able to distinguish and separate out analyte of interest from a very complex matrix background. The resultant increase in signal to noise allows for lower LLOQ’s. Sample preparation tools significantly help with this problem. Tandem quadrupole instruments, by principle, further help in attaining this goal by better focusing the ions of interest and removing the neutrals and other matrix ions.

Biotherapeutic peptides and monoclonal antibodies were tested using the Xevo TQ-XS systems. Vancomycin and Infliximab were chosen as candidate molecules to represent some of the common classes of biotherapeutics currently in clinic.

IMPROVEMENTS IN SENSITIVITY FOR BIOTHERAPEUTICS USING A XEVO TQ-XS TANDEM QUADRUPOLE MASS SPECTROMETER

Nikunj Tanna1 , Kerri Smith1 , Leonard Dillon2 ,

Mark Roberts2 , Kelly Doering1 , Anthony Marcello1 and Mark Wrona1 1Waters Corporation, Milford, MA, USA, 2Waters Corporation, Wilmslow, USA

Figure 8. Stepwave device with a second stage segmented

quadrupole

METHODS

Vancomycin

Vancomycin in a branched, tricyclic non-ribosomal peptide antibiotic which was discovered in 1953 and has been used for over 60 years in clinical practice to

treat multiple types of infections and inflammation. Vancomycin purchased from Sigma Aldrich was spiked in buffer to obtain a calibration curve from 100 pg/mL to 1 µg/mL. A ACQUITY BEH C18 column was used with generic LC conditions as shown below. The MS was optimized for the Vancomycin transition.

MS Conditions: Capillary Voltage—3 kV Cone Voltage—50 V Source Temperature—150°C Desolvation Temperature—400°C Cone Gas Flow—150 L/Hr Desolvation Gas Flow—800 L/Hr Transition—724.8—>144.13

LC Conditions:

RESULTS

Significant improvements in analyte area counts were observed for the molecules tested across the entire concentration range. For Vancomycin, a >10 fold increase in analyte area counts was observed across the range of 100pg/mL to 1 µg/mL. The signal to noise also increased by >5 across the entire concentration range. A chromatographic comparison between the Xevo TQS and Xevo TQ-XS is shown in the figure below.

CONCLUSION

The new Xevo TQ-XS tandem quadrupole Mass Spectrometer has shown significant improvements in analyte area counts as well as signal to noise compared to the current best in class Xevo TQ-S.

These improvements are seen across a wide variety of small and large biomolecules in buffer as well as in matrix.

DISCUSSION Quantitation of proteins using the surrogate peptide approach is rapidly becoming a key part of most bioanalytical laboratories have traditionally focused on performing small molecule quantification. Protein/peptide quantification brings with it unique challenges which are different from those typically experienced with a small molecule assay. Scientists are having to learn to navigate these challenges quickly. Advancements in instrumentation technologies continue to enhance the quality and speed of data generated. The Xevo TQ-XS instrument has a newly designed Stepwave ionguide. The Stepwave has a second stage featuring a segmented quadrupole design for a focused ion beam and increased ion transmission.

Infliximab

Infliximab is a chimeric humanized mouse monoclonal antibody therapeutic developed against TNF-α and is indicated for Crohn’s disease, ulcerative colitis,

psoriasis, psoriatic arthritis, ankylosing spondylitis and rheumatoid arthritis. Intact Infliximab was spiked into rat serum to generate a calibration curve and QC samples. All samples and QC’s were then digested using the Waters ProteinWorks digest kits. Briefly, the samples were denatured, reduced, alkylated and digested as per the protocol provided in the kit. The protocol has been optimized for use across biological therapeutics or biomarkers in a biological matrix. The simplified protocol reduces method development time significantly and provides scientists with an uncomplicated, kitted approach to digestion which can be very easily standardized and transferred across laboratories.

Generic LC and MS conditions were used and the same set of samples were injected into the Waters Xevo TQ-S and the Xevo TQ-XS. SINS peptide was chosen as the primary quantitation peptide. Waters mAb was added as the internal standard and a signature peptide from the Waters mAb was chosen as the internal standard peptide.

Figure 7. Average fold change in Area Counts for Infliximab

and internal standard peptides

Figure 4. Fold change in area counts for Vancomycin across

different concentrations

Figure 1. Xevo TQ-XS Tandem Quadrupole Mass Spectrometer Figure 5. Average fold change in Area counts and Signal to

Noise for Vancomycin

Figure 3. Schematic structure of Infliximab Figure 2. Molecular structure of Vancomycin

Figure 6. Chromatogram overlay for the same sample injected

on Xevo TQ-S and Xevo TQ-XS

Time0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 1.70 1.80 1.90 2.00 2.10 2.20 2.30 2.40 2.50 2.60 2.70 2.80 2.90 3.00 3.10 3.20 3.30 3.40 3.50 3.60 3.70 3.80 3.90 4.00 4.10 4.20 4.30 4.40

%

0

100

5 ng-mL MRM of 3 Channels ES+ TIC (Vancomycin)

7.09e4

0.99

Xevo TQ-XS

Xevo TQS

Concentration – 5ng/mLInjection Volume - 10µL

Significant increase in area counts observed on Xevo TQ-XS compared to Xevo TQS

For digested Infliximab, data from the most sensitive (SINS) peptide is shown below along with the digested internal standard peptide. On average, there was an increase in the analyte area counts for these peptides between 2-3 fold form a complex digested biological matrix. The ProteinWorks kit used for digestion yielded a simple protocol with a shortened sample preparation time of around 5 hours including reduction, alkylation and digestion

Time

Flow

Rate

(µL)

%A %B Curve

Initial 0.600 95 5 6

0.50 0.600 90 10 6

3.50 0.600 20 80 6

3.60 0.600 5 95 6

4.00 0.600 5 95 6

4.10 0.600 95 5 6

4.50 0.600 95 5 6

0

0.5

1

1.5

2

2.5

3

3.5

Int Std SINS

Ave

rage

Fo

ld C

han

ge

Average Fold ChangeXevo TQ-XS v Xevo TQ-S

Proto-type Xevo TQ-XS Xevo TQ-S

0

2

4

6

8

10

12

14

Area Counts Signal to Noise

Ave

rage

Fo

ld C

han

ge

Average Fold ChangeXevo TQ-XS v Xevo TQ-S

Proto-type Xevo TQ-XS Xevo TQ-S

0

5

10

15

100 1,000 10,000 100,000 1,000,000

Fold

Ch

ange

Concentration (pg/mL)

Fold Change in Area CountsXevo TQ-XS/Xevo TQ-S

Proto-type Xevo TQ-XS Xevo TQ-S