ore Gage
Calibra ion Reference Card
1
. Inspect the bore gage visually as follows:
Check for damage and excessive wear.
Check for worn contact tips on gage andextension rods.
Check for clear bezel, legible dial face.
Inspect for proper function of the bezeladjustment and lock.
Inspect for missing or lose screws.
Check for smoothness of travel by depressingthe bore gage contact point throughout theentire travel range. Repair any gages withrestricted movement prior to calibration.
2. Clean both of the receiver pads and the
mounting surfaces of the bore gage fixtureusing the cloth and ZEP I.D. Red cleaner.
3. Locate the bore gage fixture. With the bridge
facing to the right, place the bore gage fixture
against the left receiver pad shoulder.
Setup
Materials Needed:
MIC TRAC MT-3000 base unit and CPU readout 5/32" hex wrench
Bore gage 30 in/lb torque wrench
Bore gage fixture (TF-VN Block) and cap screws Cloth
Flat face anvil (TF-1F Block) cap screws and washers ZEP I.D. Red cleaner
MIC TRAC MT-3000
1 1999 Gagemaker, Inc.RCBGMT30009-99
7/26/2019 VeeBlockBoreGage CRC Training
2/12
ore Gage
Calibra ion Reference Card
Setup
4. While holding the bore gage fixture against thereceiver pad shoulder, insert the two cap
screws into the holes on either side of the
fixture.
5. While applying pressure toward the receiver
pad shoulder, use a 5/32" hex wrench to
slightly tighten the screws.
6. Use a 30 in/lbs torque wrench to secure the
cap screws.
7. Clean the flat face anvil and place it against the
right receiver pad shoulder.
8. While holding the anvil against the receiver padshoulder, insert the two cap screws withwashers into the holes on either side of theanvil.
9. While applying pressure toward the receiver
pad shoulder, use a 5/32" hex wrench to
slightly tighten the screws.
10. Use a 30 in/lb torque wrench to secure the capscrews.
MIC TRAC MT-3000
2 1999 Gagemaker, Inc.RCBGMT30009-99
7/26/2019 VeeBlockBoreGage CRC Training
3/12
ore Gage
Calibra ion Reference Card
MIC TRAC MT-3000
3 1999 Gagemaker, Inc.RCBGMT30009-99
Setup
11. Loosen the knob on top of the bore gage fixture
to allow the bore gage to slip under the bridge.
Note: Attach an extension rod on the bore gage
before mounting.
12. Insert the bore gage extension under the bridge
and gently tighten the knob.
13. Turn the coarse adjust knob counterclockwise
until the contact point of the bore gage touches
the flat face anvil.
14. Lock the coarse adjust lock.
7/26/2019 VeeBlockBoreGage CRC Training
4/12
ore Gage
Calibra ion Reference Card
Setup
15. Turn the fine adjust knob clockwise until the
bore gage dial travels of a full revolution.
16. Turn the bore gage dial to align the needle with
zero.
17. Press the PLUS/MINUS pad on the CPU panel
to display positive values during calibration.
MIC TRAC MT-3000
4 1999 Gagemaker, Inc.RCBGMT30009-99
7/26/2019 VeeBlockBoreGage CRC Training
5/12
ore Gage
Calibrat on Reference Cardi
o
Calibration
If using a computer withthe MT-3000, go to page 7.
M
aterials Needed:
MIC TRAC MT-3000 base unit and CPU readout 5/32" hex wrench
Seiko printer (optional) Lightweight gage oil
Bore gage Gage Calibration Record
1. Press the INT (internal measurement) pad onthe front panel of the CPU.
2. Press the ZERO pad on the CPU panel twice.The readout displays 0.00000.
3. Turn the fine adjust knob clockwise to the first
calibration value.
Note: If you pass the calibration value, turn the
dial back and approach the value again,
from the same direction. This practice will
increase the accuracy of the calibration.
MIC TRAC MT-3000
5 1999 Gagemaker, Inc.RCBGMT30009-99
7/26/2019 VeeBlockBoreGage CRC Training
6/12
ore Gage
Calibra ion Reference Card
Calibration
4. If you have a Seiko Printer attached to your
MIC TRAC for printing a permanent record of
calibration measurements, press the PRINT
pad on the CPU panel.
5. Record any deviations on the Gage Calibration
Record or in-house calibration report. Continue
with remaining measurements.
6. Remove the bore gage and continue with the
same calibration process for the next bore
gage.
7. After calibrating all bore gages, remove the
bore gage fixture and the flat face anvil from
the MT-3000. Oil the fixture and anvil and
return them to the storage case.
MIC TRAC MT-3000
6 1999 Gagemaker, Inc.RCBGMT30009-99
7/26/2019 VeeBlockBoreGage CRC Training
7/12
ore Gage
Calibrat on Reference Cardi
o
If using a computer withthe MT-3000, begin here.
Calibration
M
aterials Needed:
MIC TRAC MT-3000 base unit and CPU readout Bore gage
Computer 5/32" hex wrench
CALSPEX software Lightweight gage oil
ComWin32 software Brother P-Touch Label Printer (optional)
1. Start ComWin32 software by double clickingthe ComWin32 icon. The ComWin RS232Handler window displays.
Note: This software allows the computer to
read the output from the serial port ofthe CPU.
2. Click the down arrow to select the propersetting for the COM port the computer is using.For example, general2 is the setting for COMport 2.
Note: To have measured values from the
calibration recorded in CALSPEX, be surethat the box beside Press Add to ListButton is selected.
3. Minimize ComWin32. This program must berunning while using CALSPEX.
4. Start CALSPEX by double clicking theCALSPEX icon.
5. Click the New Calibration Report icon and the
Pick List window displays.
The gages available in the Pick List are thoseyou have created yourself or gage typesprovided by CALSPEX.
ComWin32 Icon
New Report IconCALSPEX Icon
MIC TRAC MT-3000
7 1999 Gagemaker, Inc.RCBGMT30009-99
7/26/2019 VeeBlockBoreGage CRC Training
8/12
ore Gage
Calibra ion Reference Card
Calibration
MIC TRAC MT-3000
8 1999 Gagemaker, Inc.RCBGMT30009-99
6. Select the bore gage from the Pick List.
7. Click the OK button. The Header Informationwindow displays for entering information aboutthe company whose gage is being calibrated.
8
. Complete the following header information:
Contact - name of person to contactregarding the gage.
Department - department to which the gagebelongs.
PO/Account - purchase order or accountnumber.
Calibration Date - automatically displays,but can be changed by typing over thedisplayed date.
Certified By - name of the person certifyingthe gage.
9. Click the Next button. The Gage Informationwindow displays.
10. Enter the following information about the gage
you are calibrating:
Control Number - unique identification
number for the gage.
Gage Manufacturer
Model Serial Number
Description
Location
11. Click the Next button The window for entering
beginning and ending calibration values
displays.
7/26/2019 VeeBlockBoreGage CRC Training
9/12
ore Gage
Calibra ion Reference Card
Calibration
MIC TRAC MT-3000
9 1999 Gagemaker, Inc.RCBGMT30009-99
12. Type the beginning and ending calibration
values and the increment values for the bore
gage.
13. Click the Next button. The Tolerances window
displays.
14. Enter the tolerances for the bore gage.
15. Click the Next button. The calibration values
table displays.
16. Enter the gage measurements in the Measuredcolumn as follows:
Turn the fine adjust knob clockwise to thefirst calibration value.
Note: If you pass the calibration value, turn thedial back and approach the value again,from the same direction. This practice will
increase the accuracy of the calibration.
Press the PRINT pad on the CPU.
CALSPEX records the value and displays agreen box in the In/Out column if the value iswithin tolerance. A red box displays if the valueis not within tolerance. The Deviation columnshows the deviation of the measured valuefrom the master value.
7/26/2019 VeeBlockBoreGage CRC Training
10/12
ore Gage
Calibra ion Reference Card
Calibration
17. Using each master value, continue measuring
the bore gage until all values are recorded in
CALSPEX.
18. Click the Next button. The Next
Calibration/NIST window displays.
19. Change the Next Calibration Date, if
necessary, by typing over the displayed date.
20. Click the Next button. The Pass/Fail window
displays.
21. Click the Pass this Gage option or Fail this
Gage option to report the condition of the gage.
You can pass a gage even if CALSPEX has
determined that the gage measurementsrecorded are not within tolerance.
22. Click the Next button. The Save As window
displays.
MIC TRAC MT-3000
10 1999 Gagemaker, Inc.RCBGMT30009-99
7/26/2019 VeeBlockBoreGage CRC Training
11/12
ore Gage
Calibra ion Reference Card
Calibration
Note: CALSPEX automatically assigns a filename
to each calibration. DO NOTchange the
filename to avoid problems with the
CALSPEX database.
23. Click the Save button. The Gage Status
window displays.
24. Click the Update button to update the Recall
Database.
Note: If you have a Brother P-Touch Label
Printer for printing calibration stickers,
the Confirm window for printing a
calibration tag displays.
25. If the Confirm window displays, click the Yes
button to print the Calibration Tag. Affix the
Calibration Tag to the bore gage.
MIC TRAC MT-3000
11 1999 Gagemaker, Inc.RCBGMT30009-99
7/26/2019 VeeBlockBoreGage CRC Training
12/12
ore Gage
Calibra ion Reference Card
Calibration
26. The Calibration Summary displays and showsthe calibration information for the bore gage.
27. Remove the bore gage and continue with the
same calibration process for the next bore
gage.
28. After calibrating all bore gages, remove the
bore gage fixture and the flat face anvil from
the MT-3000. Oil the fixture and anvil and
return them to the storage case.
MIC TRAC MT-3000