Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
Variable Angle LCP 1st MTP FusionPlates 2.4/2.7. Part of the Variable AngleLCP Forefoot /Midfoot System 2.4 /2.7.
Technique Guide
Synthes 1 Synthes 1
Table of Contents
Introduction
Controlled Compression Technique
Screw Insertion Techniques
Surgical Technique
Product Information
Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 2
Indications 4
5
9
Approach and Preparation 10
Implantation 13
Implant Removal 25
Screws 26
Plates 28
Instruments 30
Synthes Biomaterials Overview 36
Image intensifier control
WarningThis description alone does not provide sufficient background for direct use ofthe product. Instruction by a surgeon experienced in handling this product ishighly recommended.
Reprocessing, Care and Maintenance of Synthes InstrumentsFor general guidelines, function control and dismantling of multi-part instruments,please contact your local sales representative or refer to:www.synthes.com/reprocessing
2 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Variable Angle LCP 1st MTP FusionPlates 2.4/2.7. Part of the Variable AngleLCP Forefoot/Midfoot System 2.4 /2.7.
Features and Benefits
Oblong combi-hole for plate positioning and compression
Anatomic, low profile designedspecifically for 1st MTP arthrodesis
Fusion line marking indicates the fusion site to ensure proper plate placement
Compression wire holes allow pre -liminary fixation of the plate to the bone
Plantar side undercut for contouringaround medial phalangeal eminence
Compression featureCompression holes used with compres-sion wires and forceps allow for tactilecompression up to 4 mm.
Variable angleScrew holes allow up to 15° off-axisscrew angulation in all directions.
Minimized soft tissue irritationLow profile plates with rounded edgesand highly polished surface minimizesoft tissue irritation.
Synthes 3
Small and medium plates– Small and medium plates are offered
in 3 different dorsiflexion angles:0°, 5°, 10°
– Left and right plates
Large plates– Large plates are offered with 5° dor-
sisiflexion angle– Left and right plates
Revision plates– Revision plates are offered with 0°
dorsiflexion and an extra variable an-gle locking hole to secure a boneblock
– Left and right plates
Proximal and distal reamers– Used to prepare articulating joint
surfaces for fusion – specifically be-tween metatarsals and phalanges
– Cannulated for use over a Kirschnerwire for better control
– 14–24 mm reamers accept 1.6 mmKirschner wire
– Proximal reamer is concave-shapedand forms the sphere on the jointsurface
– Distal reamer is convex-shaped andforms the pocket in the joint surface
Indications
The 1st MTP Fusion Plate of the Variable Angle LCP Forefoot/Midfoot System 2.4 /2.7 is indicated for deforma-tions of the 1st metatarsophalangeal (MTP) joint (HalluxRigidus) and fractures, nonunions and replantations of thefirst metatarsal bone, particularly in osteopenic bone.
4 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Spherical stop
Compressionslot
Compressionwire hole
Synthes 5
The plates included in the Variable Angle LCP Forefoot / Mid-foot System 2.4 / 2.7 aid in reconstructive foot surgery by allowing controlled compression with the use of compressionwires and compression forceps.
Compression feature– Allows for up to 4 mm of compression– Tactile compression– Designed within the plate to minimize additional soft
tissue dissection– Allows for final screw fixation after compression is
achieved
Compression wires– 1.6 mm diameter, 150 mm overall length– Seven thread lengths: 10, 15, 20, 25, 30, 35 and 40 mm– Stop feature allows for quick and easy preliminary plate
fixation, which eliminates the need for plate holding forceps or another hand to hold the plate to the bone
– Spherical stop:– Designed to be seated on top of compression wire
holes and compression slots, and inside variable angleLCP holes
– Allows for off-axis wire insertion while maintaining thecompression feature capability
– Material: Cobalt chromium alloy that is stiffer than con-ventional stainless steel
Compression forceps– Spherical shaped recess matches the spherical stops on
the compression wires, which ensures the forceps graspthe stops regardless of the angle in which the wires wereinserted
– Locking ratcheting mechanism holds compression duringinsertion of fixation screws
– Simple lightweight design does not require holding duringinsertion of fixation screws
Controlled Compression Technique
1Position plate
Place the plate on the bone, ensuring that the plate is placedappropriately according to the specific procedure.
6 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Controlled Compression Technique
Synthes 7
2Insert compression wires
Instrument
03.211.410.01– Compression Wire � 1.6 mm, 03.211.440.01 length 150 mm, thread length 10 – 40 mm
Estimate the appropriate thread length needed for the plateand bone combination. Bicortical fixation is recommended.
Using a wire driver, insert the compression wire through thecompression wire hole and through both bone cortices.
Important: To minimize stripping of the bone threads, slowthe insertion once the spherical stop of the wire gets close tothe plate. Slowly control the insertion to achieve good com-pression of the wire to the plate and to the bone. Highpower insertion and stripping of the bone threads can leadto loosening of the compression wires and reduced compres-sion.
Insert the second compression wire into the far side of theslot.
Controlled Compression Technique
3Compress using forceps
Instrument
03.211.400 Compression Forceps for use with Compression Wire
Move the ratcheting switch so the forceps ratchet when closing preventing the spring from opening the forceps.
Place the compression forceps in position, ensuring that thearms are around the compression wire spheres.
Compress by squeezing the handles.
Important: Compression is tactile, but be careful not to overcompress. This may cause the compression wires to strip outof the bone.
When the ratcheting mechanism is in the correct position,compression can be maintained without holding the forceps.This leaves the hands free for image intensifier control ofcompression gap closure and for inserting final fixationscrews.
8 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Synthes 9
The plate holes of the Variable Angle LCP Technology 2.4 / 2.7 accept 2.4 mm and 2.7 mm Variable Angle (VA)Locking Screws.
Screws can be inserted using two different techniques:– Variable angle technique– Pre-defined nominal angle technique
Variable angle techniqueTo drill variable angle holes at a +/-15° deviation from thenominal trajectory of the locking hole, insert the tip of theconical VA-LCP drill sleeve (03.211.003 / 03.110.023) and keyinto the cloverleaf design of the VA-LCP hole.
Note: It is important not to angulate more than 15° fromthe central axis of the screw hole. Overangulation may resultin difficulty while locking the screw and inadequate screw locking.
Pre-defined nominal angle techniqueThe fixed-angle VA-LCP drill sleeve (03.211.004 / 03.110.024)only allows the drill bit to follow the nominal trajectory ofthe locking hole.
Screw Insertion Techniques
Use of funnel-shaped VA-LCP Drill Sleeve
VA-LCP Drill Sleeve, coaxial, for Drill Bits (03.211.004/03.110.024)
VA-LCP drill sleeve, conical, for Drill Bits (03.211.003/03.110.023)
Approach and Preparation
1Approach
Make a dorsomedial incision extending from the proximalinter phalangeal joint of the hallux to the middle of the firstmetatarsal. This incision should be medial to the extensorhallucis longus tendon. Isolate and retract the superficialnerve and vessel. Make a linear capsular incision to exposethe first metatarsal head and proximal phalanx base.
Dissect down to the joint. Bend the phalanx down to exposethe joint.
2Insert Kirschner wire
Instrument
292.560 Kirschner Wire � 1.6 mm with double tip, length 150 mm, Stainless Steel
Insert the 1.6 mm Kirschner wire into the base of the firstmetatarsal. Ensure the wire is centered in the canal.
A chielectomy may be necessary to remove impinging boneand create a flat surface for the plate.
10 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Synthes 11
3Ream proximally
Instruments
03.211.114 Cannulated Proximal Reamer � 14 mm
03.211.116 Cannulated Proximal Reamer � 16 mm
03.211.118 Cannulated Proximal Reamer � 18 mm
03.211.120 Cannulated Proximal Reamer � 20 mm
03.211.122 Cannulated Proximal Reamer � 22 mm
03.211.124 Cannulated Proximal Reamer � 24 mm
Select the appropriate reamer. Always start with a larger sizeand work down to a smaller size.
Slide the proximal reamer over the guide wire. Ream to re-move cartilage from the joint.
Important: When preparing the joint, it is possible to remove too much bone, which may result in shortening.Ream only enough to remove the cartilage and exposethe joint surface for fusion.
Approach and Preparation
4Ream distally
Instruments
03.211.014 Cannulated Distal Reamer � 14 mm
03.211.016 Cannulated Distal Reamer � 16 mm
03.211.018 Cannulated Distal Reamer � 18 mm
03.211.020 Cannulated Distal Reamer � 20 mm
03.211.022 Cannulated Distal Reamer � 22 mm
03.211.024 Cannulated Distal Reamer � 24 mm
Bend the toe downwards to expose the phalanx.
Remove the Kirschner wire from the metatarsal. Insert aKirschner wire into the center of the phalanx.
Slide the appropriate distal reamer over the Kirschner wireand ream to remove cartilage from the joint.
Important: When preparing the joint, it is possible to remove too much bone, which may result in shortening.Ream only enough to remove the cartilage and exposethe joint surface for fusion.
Remove the reamer and Kirschner wire.
12 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Synthes 13
Implantation
2Contour plate
Instrument
03.211.005 Bending Pliers for VA Locking Plates
The 1st MTP fusion plates can be contoured to fit the specificanatomy and fixation options.
The bending pliers are designed to protect the variable angleholes during contouring. The feature on the pliers lines upwith the cloverleaf design in the plate. Two pliers are used tocontour the plate.
Important– If possible, bend the plate between the VA holes. Do not
deform the threaded part of the holes or over-bendthe plates during bending as this may adversely affect insertion of VA locking screws.
– Do not repeatedly bend the plates back and forth as thismay weaken the plate.
1Select and position plate
Select the plate that allows for the desired level of correction.
A sterile flat surface can be placed on the bottom of the footto measure the desired amount of toe dorsiflexion. The platemay be contoured slightly to achieve the desired outcome.
Place the plate on the joint surface, aligning the etched lineon the plate with the joint to ensure proper plate position.
14 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Implantation
3Apply reduction and compression
Instruments
03.211.400 Compression Forceps for use with Compression Wire
03.211.410.01 Compression Wire � 1.6 mm, length 150 mm, thread length 10 mm
03.211.415.01 Compression Wire � 1.6 mm, length 150 mm, thread length 15 mm
03.211.420.01 Compression Wire � 1.6 mm, length 150 mm, thread length 20 mm
03.211.425.01 Compression Wire � 1.6 mm, length 150 mm, thread length 25 mm
03.211.430.01 Compression Wire � 1.6 mm, length 150 mm, thread length 30 mm
03.211.435.01 Compression Wire � 1.6 mm, length 150 mm, thread length 35 mm
03.211.440.01 Compression Wire � 1.6 mm, length 150 mm, thread length 40 mm
If required, compression may be achieved as described inthe Controlled Compression Technique section (p. 6) of thistechnique guide.
It is recommended to insert an independent lag screw to aidin stabilization. Insert the screw starting medial and distal tothe joint on the phalanx, crossing the joint and ending in thelateral side of the metatarsal head.
Synthes 15
4Insert independent cortex screw
Instruments – cortex screws 2.7 mm
310.534 Drill Bit � 2.0 mm, with marking, length 110 /85 mm, 2-flute, for Quick Coupling
310.260 Drill Bit � 2.7 mm, length 100 /75 mm, 2-flute, for Quick Coupling
323.260 Universal Drill Guide 2.7
03.111.005 Depth Gauge for Screws � 2.0 to 2.7 mm, measuring range up to 40 mm
03.111.038 Handle with Quick Coupling
314.467 Screwdriver Shaft, Stardrive, T8, self-holding
Instruments – cortex screws 2.4 mm
310.509 Drill Bit � 1.8 mm, with marking, length 110 /85 mm, 2-flute, for Quick Coupling
310.530 Drill Bit � 2.4 mm, length 100/75 mm, 2-flute, for Quick Coupling
323.202 Universal Drill Guide 2.4
03.111.005 Depth Gauge for Screws � 2.0 to 2.7 mm, measuring range up to 40 mm
03.111.038 Handle with Quick Coupling
314.467 Screwdriver Shaft, Stardrive, T8, self-holding
16 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Implantation
Insert an additional independent cortex screw depending onthe corresponding indication and situation. For compressionof the additional independent screw it is recommended touse the lag screw technique.
For 2.4 mm cortex screws, use the 2.4 universal drill guideand pre-drill the screw hole with the 1.8 mm drill bit. For2.7 mm cortex screws, use the 2.7 universal drill guide andpre-drill the screw hole with the 2.0 mm drill bit.
To drill a gliding hole for compression, use the 2.7 mm drillbit (for 2.7 cortex screw) or the 2.4 mm drill bit (for 2.4 mmcortex screw) with the double drill guide.
Determine the screw length with the depth gauge and insertthe screw.
Synthes 17
5Pre-drill for VA locking screws
Instruments – VA screws 2.7 mm
310.534 Drill Bit � 2.0 mm, with marking, length110 /85 mm, 2-flute, for Quick Coupling
03.211.003 VA-LCP Drill Sleeve 2.7, conical, for Drill Bits � 2.0 mm
03.211.004 VA-LCP Drill Sleeve 2.7, coaxial, for Drill Bits � 2.0 mm
323.260 Universal Drill Guide 2.7
03.111.005 Depth Gauge for Screws � 2.0 to 2.7 mm, measuring range up to 40 mm
Instruments – VA screws 2.4 mm
310.509 Drill Bit � 1.8 mm, with marking, length110 /85 mm, 2-flute, for Quick Coupling
03.110.023 VA-LCP Drill Sleeve 2.4, conical, for Drill Bits � 1.8 mm
03.110.024 VA-LCP Drill Sleeve 2.4, coaxial, for Drill Bits � 1.8 mm
323.202 Universal Drill Guide 2.4
03.111.005 Depth Gauge for Screws � 2.0 to 2.7 mm, measuring range up to 40 mm
Determine the size of screws to be used, 2.4 or 2.7 mm, andwhether they will be inserted at a variable angle (5a) or atthe pre-defined nominal angle (5b).
18 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Implantation
5aPre-drill using variable angle technique
Instruments – VA screws 2.7 mm
310.534 Drill Bit � 2.0 mm, with marking, length 110 / 85 mm, 2-flute, for Quick Coupling
03.211.003 VA-LCP Drill Sleeve 2.7, conical, for Drill Bits � 2.0 mm
03.111.005 Depth Gauge for Screws � 2.0 to 2.7 mm, measuring range up to 40 mm
Instruments – VA screws 2.4 mm
310.509 Drill Bit � 1.8 mm, with marking, length 110 / 85 mm, 2-flute, for Quick Coupling
03.110.023 VA-LCP Drill Sleeve 2.4, conical, for Drill Bits � 1.8 mm
03.111.005 Depth Gauge for Screws � 2.0 to 2.7 mm, measuring range up to 40 mm
Variable angle locking screws allow for manipulation aroundthe independent lag screw.
Insert and lock the VA-LCP drill sleeve tip into the cloverleafdesign of the VA-LCP hole. The cone will self retain in thehole.
Synthes 19
Use the 2.0 mm drill bit (2.7 mm VA screw) or the 1.8 mmdrill bit (2.4 mm VA screw) to drill at the desired angle and tothe desired depth.
The cone of the drill sleeve allows the drill bit to be angledup to 15° around the central axis of the locking hole.
Important: To ensure that the drill guide is locked correctly,do not angle the drill bit in excess of +/-15° from the nomi-nal trajectory of the hole.
To achieve the desired angle, verify the drill bit angle anddepth under image intensification. If incorrect, drill at a different angle and verify again under image intensification.
Use the according depth gauge to measure the correct screwlength.
20 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Implantation
5bPre-drill using pre-defined nominal angle technique
Instruments – VA screws 2.7 mm
310.534 Drill Bit � 2.0 mm, with marking, length 110 / 85 mm, 2-flute, for Quick Coupling
03.211.004 VA-LCP Drill Sleeve 2.7, coaxial, for Drill Bits � 2.0 mm
03.111.005 Depth Gauge for Screws � 2.0 to 2.7 mm, measuring range up to 40 mm
Optional instrument
323.260 Universal Drill Guide 2.7
Instruments – VA screws 2.4 mm
310.509 Drill Bit � 1.8 mm, with marking, length110 / 85 mm, 2-flute, for Quick Coupling
03.110.024 VA-LCP Drill Sleeve 2.4, coaxial, for Drill Bits � 1.8 mm
03.111.005 Depth Gauge for Screws � 2.0 to 2.7 mm, measuring range up to 40 mm
Optional instrument
323.202 Universal Drill Guide 2.4
Variable angle locking screws and standard locking screwscan be inserted into the plate at the predefined hole angle orcoaxial.
Synthes 21
Insert and lock the VA-LCP drill sleeve tip into the cloverleafdesign of the VA-LCP hole. The coaxial drill guide will self retain in the hole.
Use the 2.0 mm drill bit (for VA locking screw and standard2.7 mm locking screws) or the 1.8 mm drill bit (for VA lock-ing screw and standard 2.4 mm locking screws) to drill tothe desired depth.
Verify the drill bit depth under image intensification.
Use the according depth gauge to measure the correct screwlength.
22 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Implantation
6Insert VA locking screws
Instruments – VA screws 2.4/2.7 mm
314.467 Screwdriver Shaft, Stardrive, T8, self-holding
311.430 Handle with Quick Coupling, length 110 mmor03.111.038 Handle with Quick Coupling
Insert the correct length variable angle locking screw manu-ally using the screwdriver shaft and handle with quick cou-pling. Insert the screw until the screw head is seated (withlimited force) in the variable angle locking hole.
Important: Do not over-tighten screws. This allows thescrews to be easily removed should they not be in the desired position.
Insert additional screws as needed.
Confirm proper reconstruction, screw placement and screwlength under image intensification.
Synthes 23
7Lock VA locking screws
Instruments – VA screws 2.4/2.7 mm
314.467 Screwdriver Shaft, Stardrive, T8, self-holding
03.110.002 Torque Limiter, 1.2 Nm, with AO / ASIF Quick Coupling
03.110.005 Handle for Torque Limiters 0.4 / 0.8 / 1.2 Nm
Use the 1.2 Nm torque limiting attachment (TLA) to performthe final locking step for variable angle locking screws. Thetorque limiting attachment attaches to the T8 Stardrivescrewdriver shaft and the blue handle of the torque limitingattachment.
After appropriate screw angle and screw length has been finalized, manually insert the screw using the TLA assembly.
Use of the TLA is mandatory for variable angle locking holesto ensure the correct amount of torque is applied when in-serting the screws.
With this final locking step, the screws are securely lockedin the plate to achieve maximum strength of the plate-screwinter face.
Remove the compression wires.
24 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Implantation
8Ensure proper reconstruction
Ensure proper joint reconstruction, screw placement andscrew length under image intensification. Verify that thescrews are not in the soft tissue.
Synthes 25
Instruments – VA screws 2.4/2.7 mm
314.467 Screwdriver Shaft, Stardrive, T8, self-holding
03.111.038 Handle with Quick Coupling
To remove locking screws, first unlock all locking screws before removing them completely. Otherwise, the plate mayrotate and damage the soft tissue.
Implant Removal
26 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Variable angle locking screws (VA-LCP) 2.7 mm
0X.211.010– VA Locking Screw Stardrive � 2.7 mm040 (head 2.4), self-tapping, length 10–40 mm
0X.211.042S– VA Locking Screw Stardrive � 2.7 mm 060S (head 2.4), self-tapping, length 42–60 mm, sterile
Threaded, rounded head locks securely into the threaded VA-LCP holes to provide angular stability at angles deter-mined by the surgeon.
Also securely locks into standard locking holes (LCP) of theplate at the pre-defined angle.
Important: For final locking, the 1.2 Nm TLA torque limitingattachment is required.
Screws
All non-sterile screws are also available sterile packed. Add suffix “S” to article number to order sterile product.
X=2: Stainless steelX=4: TAN
Optional:Variable angle locking screws (VA-LCP) 2.4 mm
0X.210.106– VA Locking Screw Stardrive � 2.4 mm,140 self-tapping, length 6–40 mm
0X.210.142S– VA Locking Screw Stardrive � 2.4 mm,160S self-tapping, length 42–60 mm, sterile
Synthes 27
Cortex screws 2.7 mm
X02.870– Cortex Screw Stardrive � 2.7 mm, 900 self-tapping, length 10–40 mm
X02.962S– Cortex Screw Stardrive � 2.7 mm,969S self-tapping, length 42–60 mm, sterile
All non-sterile screws are also available sterile packed. Add suffix “S” to article number to order sterile product.
X=2: Stainless steelX=4: TAN
Cortex screws 2.4 mm
X01.756– Cortex Screw Stardrive � 2.4 mm, 790 self-tapping, length 6–40 mm
0X.210.942S– Cortex Screw Stardrive � 2.4 mm,960S self-tapping, length 42–60 mm, sterile
For use in round or combi-holes.
Optional: Locking head screws 2.4/2.7 mm
X12.806– Locking Screw Stardrive � 2.4 mm,830 self-tapping, length 6–30 mm
X02.206– Locking Screw Stardrive � 2.7 mm 260 (head LCP 2.4), self-tapping, length 6–60 mm
28 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Plates
1st MTP Fusion Plates 2.4/2.7, VA locking, small
Art. No. Length Dorsiflexion Left/Right(mm)
0X.211.230 42 0° Right
0X.211.231 42 0° Left
0X.211.232 42 5° Right
0X.211.233 42 5° Left
0X.211.234 42 10° Right
0X.211.235 42 10° Left
1st MTP Fusion Plates 2.4/2.7, VA locking, medium
Art. No. Length Dorsiflexion Left/Right(mm)
0X.211.236 52 0° Right
0X.211.237 52 0° Left
0X.211.238 52 5° Right
0X.211.239 52 5° Left
0X.211.240 52 10° Right
0X.211.241 52 10° Left
All plates are available non-sterile and sterile packed. Add suffix “S” to article number to order sterile product.
X=2: Stainless steelX=4: TAN
R
L
R
L
Synthes 29
1st MTP Fusion Plates 2.4/2.7, VA locking, large
Art. No. Length Dorsiflexion Left/Right(mm)
0X.211.242 57 5° Right
0X.211.243 57 5° Left
1st MTP Fusion Plates 2.4/2.7, VA locking, for Revision
Art. No. Length Dorsiflexion Left/Right(mm)
0X.211.244 53 0° Right
0X.211.245 53 0° Left
All plates are available non-sterile and sterile packed. Add suffix “S” to article number to order sterile product.
X=2: Stainless steelX=4: TAN
R
L
R
L
30 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Instruments
311.430 Handle with Quick Coupling, length 110 mm
314.467 Screwdriver Shaft, Stardrive, T8,self-holding
03.110.002 Torque Limiter, 1.2 Nm, with AO / ASIF Quick Coupling
03.110.005 Handle for Torque Limiters 0.4 / 0.8 / 1.2 Nm
03.111.005 Depth Gauge for Screws � 2.0 to 2.7 mm,measuring range up to 40 mm
03.111.038 Handle with Quick Coupling
03.211.001 Holding Pin for VA Locking Plates 2.4 / 2.7
Synthes 31
Instruments for insertion of 2.7 mm screws
310.260 Drill Bit � 2.7 mm, length 100 / 75 mm, 2-flute, for Quick Coupling
310.534 Drill Bit � 2.0 mm, with marking,length 110 / 85 mm, 2-flute, for Quick Coupling
323.260 Universal Drill Guide 2.7
03.211.003 VA-LCP Drill Sleeve 2.7, conical, for Drill Bits � 2.0 mm
03.211.004 VA-LCP Drill Sleeve 2.7, coaxial, for Drill Bits � 2.0 mm
32 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Instruments
Instruments for insertion of 2.4 mm screws
310.530 Drill Bit � 2.4 mm, length 100/75 mm,2-flute, for Quick Coupling
323.202 Universal Drill Guide 2.4
03.110.023 VA-LCP Drill Sleeve 2.4, conical, for Drill Bits � 1.8 mm
03.110.024 VA-LCP Drill Sleeve 2.4, coaxial, for Drill Bits � 1.8 mm
310.509 Drill Bit � 1.8 mm, with marking, length110/85 mm, 2-flute, for Quick Coupling
Synthes 33
03.211.400 Compression Forceps for use withCompression Wire
03.211.410.01 Compression Wire � 1.6 mm, length 150 mm, thread length 10 mm
03.211.420.01 Compression Wire � 1.6 mm, length 150 mm, thread length 20 mm
03.211.415.01 Compression Wire � 1.6 mm, length 150 mm, thread length 15 mm
03.211.425.01 Compression Wire � 1.6 mm, length 150 mm, thread length 25 mm
03.211.440.01 Compression Wire � 1.6 mm, length 150 mm, thread length 40 mm
03.211.435.01 Compression Wire � 1.6 mm, length 150 mm, thread length 35 mm
03.211.430.01 Compression Wire � 1.6 mm, length 150 mm, thread length 30 mm
Instruments for compression
34 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
Cannulated proximal and distal reamers
Art. No. Proximal / Diameter Distal (mm)
03.211.114 Proximal 14
03.211.014 Distal 14
03.211.116 Proximal 16
03.211.016 Distal 16
03.211.118 Proximal 18
03.211.018 Distal 18
03.211.120 Proximal 20
03.211.020 Distal 20
03.211.122 Proximal 22
03.211.022 Distal 22
03.211.124 Proximal 24
03.211.024 Distal 24
Instruments
Additional instrument
03.211.005 Bending Pliers for VA Locking Plates
Synthes 35
Norian SRS
*Facilitated through SynthesOsteoinductive power
Osteoconductive, resorbable, synthetic Enhancing chronOS with biological factors Injectable remodelling
Injectable stability
Furthermore a comprehensive portfolioof allograft products is available in selected countries.
For more detailed information about aspecific product or availability of allo-grafts please contact your local Synthesrepresentative.
DBX*
Synthes Biomaterials Overview
Synthetic and allogenic bone replace-ment materials have the advantage ofuniform quality, unlimited availabilityand absence of potential complicationsat a donor site.
chronOS
Additionally, the application of syntheticand allogenic bone graft substitutes reduces the duration of the surgery.
Synthes offers a wide range of syntheticbiomaterial products in different application forms and with distinct bio-logical properties:
chronOS Perfusion Concept chronOS Inject
36 Synthes Variable Angle LCP 1st MTP Fusion Plates 2.4 /2.7 Technique Guide
0123
Synthes GmbHEimattstrasse 3CH-4436 Oberdorfwww.synthes.com
All technique guides are available as PDF files at www.synthes.com/lit
Ö036.001.234öAAzä
036.
001.
234
vers
ion
AA
re
v. 1
01
/201
230
1011
74
© S
ynth
es, I
nc. o
r its
aff
iliat
es
Subj
ect
to m
odifi
catio
ns
Synt
hes,
LC
P an
d St
ardr
ive
are
trad
emar
ks o
f Sy
nthe
s, In
c. o
r its
aff
iliat
es