User manual
JOFRA DLC-155/158/250/700Copyright 2010 AMETEK Denmark A/S
7/30/2019 User Manual Dlc
2/17
7/30/2019 User Manual Dlc
3/17
127414 01 20-12-2010 3
Table of contents: Page
1.0 General information ......................42.0 Introduction....................................5
3.0 Functionality ..................................6
3.1 Functional description..................6
3.2 Connections ................................6
3.3 Serial number..............................8
4.0 Operation........................................9
4.1 Operation area ............................9
5.0 Maintenance.................................10
5.1 Calibration .................................11
6.0 Technical specifications .............15
7/30/2019 User Manual Dlc
4/17
4 20-12-2010 127414 01
1.0 General information
This manual is only effective for the following
products:
JOFRA DLC-155JOFRA DLC-158JOFRA DLC-250JOFRA DLC-700
The products are manufactured by:
AMETEK Denmark A/S
Gydevang 32-343450 Allerd - Denmark
TEL: +45 48 16 80 00FAX: +45 48 16 80 80
7/30/2019 User Manual Dlc
5/17
127414 01 20-12-2010 5
2.0 Introduction
The JOFRA DLC-155, DLC-158, DLC-250 and DLC-700 probes are designed for fast and traceablecalibration and temperature measuring with JOFRARTC-156/157 B/C, JOFRA RTC-158 B/C, JOFRARTC-250 B/C and JOFRA RTC-700 B/C systemsand are ready for use.
Please read this manual carefully before use, toobtain maximum value of your calibration system.
Warning
Read this manual before use.
Do not use in hazardous area.
Handle carefully.
Never exceed temperature range
7/30/2019 User Manual Dlc
6/17
6 20-12-2010 127414 01
3.0 Functionality
3.1 Functional description
The probes can be used for temperatures in therange :
DLC-155 : -60C to 155C / -76F to 311F
DLC-158 : -60C to 155C / -76F to 311F
DLC-250 : 0C to 250C / 32F to 482F
DLC-700 : 0C to 700C / 32F to 1292F
The JOFRA DLC-155, DLC-158, DLC-250 and DLC-700 probes may be supplied with a certificate for alimited temperature range.
3.2 Connections
The probes are delivered with a connecting cable.The pin-layout is as follows:
Pin 1 : TC V-Pin 2 : TC V+Pin 3 : Memory GND
Pin 4 : Memory I/O
7/30/2019 User Manual Dlc
7/17
127414 01 20-12-2010 7
The figure below is shown from the connector sideof the probe connector.
1
23
4
7/30/2019 User Manual Dlc
8/17
8 20-12-2010 127414 01
3.3 Serial number
The serial number is placed on the probe as shownon the figure below:
Serial number
7/30/2019 User Manual Dlc
9/17
127414 01 20-12-2010 9
4.0 Operation
4.1 Operation area
All the probes are intended for use in areas whichmeet the following:
Ambient temperaturerange : -20C to 70C (-4F to 158F)
Humidity : 0% to 90%
Protection class : IP 50
WarningDo not use in hazardous areas.
7/30/2019 User Manual Dlc
10/17
10 20-12-2010 127414 01
5.0 Maintenance
The probe does not require specific maintenancebefore or after use. The user may carry out thefollowing procedure himself:
Cleaning the probe : Use alcohol or water
and a soft cloth.
Caution
The probe must always be protectedagainst any mechanical damage.
The probe must never be exposed to
mechanical shock effects. Avoid thermal shock
Any bending of the probe may causepermanent damage
7/30/2019 User Manual Dlc
11/17
127414 01 20-12-2010 11
5.1 Calibration
Required equipment:
DLC-probe to be calibrated. RTC-calibrator with DLC-input.
An insert with correct size holes for the DLC-probe and the temperature probe.
For calibrating DLC-155, DLC-158 and DLC-250:an AMETEK 127587 temperature probe(recommended) or similar with short measuringzone is required.
For calibrating DLC-700: a STS-200 A 970temperature probe (recommended) or similar isrequired.
Procedure:
1. Place the DLC-probe in the insert and connectthe probe to the DLC-input of the calibrator.
2. Go to the Sensor setup menu.
3. Set Use load compensation to YES. If thereference probe is connected to the REF input,
set Sensor type to External and SET followsTRUE to No. If no reference probe isconnected, set Sensor type to Internal.
4. Enter the DLC-menu.
5. Read the A, B and C-coefficients of the probeand enter the values in the EXCEL spreadsheet
(127588 DLC-155.XLS, 127884 DLC-158.XLS,127885 DLC-250.XLS or 127589 DLC-700.XLS).
7/30/2019 User Manual Dlc
12/17
12 20-12-2010 127414 01
The calibration temperatures are:Sensor DLC-155 DLC-155 DLC-700
Calibrator RTC-156 RTC-157 RTC-700
-30C/-22F -45C/-49F 33C/91.4F-20C/-4F -20C/-4F 200C/392F
0C/32F 0C/32F 350C/662F
50C/122F 50C/122F 500C/932F
100C/212F 100C/212F 660C/1220F
*Recommendedcalibrationtemperatures
155C/311F 155C/311F 700C/1292F
Lower
measuring point
Bottom of
insert
Bottom of
insert
Bottom of
insertUppermeasuring point
60 mm /2.4 frombottom
60 mm /2.4 frombottom
80 mm / 3.1from bottom
Curve fitwarning limit
0.01C/0.018F
0.01C /0.018F
0.05C /0.09F
* Note : The temperatures must be entered in Celcius.
7/30/2019 User Manual Dlc
13/17
127414 01 20-12-2010 13
Sensor DLC-158 DLC-250
Calibrator RTC-158 RTC-250
-22C/-72F 28C/82F
-10C/-4F 50C/122F0C/32F 100C/212F
50C/122F 150C/302F
100C/212F 200C/392F
*Recommendedcalibrationtemperatures
155C/311F 250C/482F
Lowermeasuring point
Bottom ofinsert
Bottom ofinsert
Uppermeasuring point
60 mm /2.4 frombottom
60 mm /2.4 frombottom
Curve fitwarning limit
0.01C/0.018F
0.01C /0.018F
* Note : The temperatures must be entered in Celcius.
The following procedure is repeated for eachtemperature:
1. Set the SET temperature.
2. Place the reference temperature probe in thebottom of the insert.Wait until the temperature have stabilized and theDLC-reading is 0.00C. If the DLC-readingcontinuously shows a positive value at -30C(RTC-156) / -45C (RTC-157) / -22C (RTC-158),increase the SET-temperature in steps of 1C
until the DLC-reading displays 0.00C.
7/30/2019 User Manual Dlc
14/17
14 20-12-2010 127414 01
3. Record the temperature of the reference probe inthe excel spreadsheet.
4. Move the reference probe to the upper position
60 mm (RTC-156/157/158/250) or 80 mm (RTC-700).
5. Wait until the temperature have stabilized and theDLC-reading is 0.00C.
6. Record the temperature of the reference probeand the DLC-probe in the excel spreadsheet.
7. Go to next SET-temperature.
After the last STEP is completed:
8. Check that all values in the calculated curve fitare less than the values specified in the table. Ifthis is OK, the new coefficients can be
downloaded using the CON050 software.
7/30/2019 User Manual Dlc
15/17
127414 01 20-12-2010 15
6.0 Technical specifications
Probe specifications DLC-155:
Sensor type : Differential TC type T
Probe length : 177 mm (6.96 in)
Temperature range : -60C to 155C / -76F to311F
Diameter : OD3 mmDepth : 140 mm (5.51 in)
Media compatibility : AISI316
Connection : Redel plug with build inmemory is standard
Probe specifications DLC-158:
Sensor type : Differential TC type T
Probe length : 192 mm (7.56 in)
Temperature range : -60C to 155C / -76F to
311F
Diameter : OD3 mm
Depth : 140 mm (5.51 in)
Media compatibility : AISI316
Connection : Redel plug with build inmemory is standard
7/30/2019 User Manual Dlc
16/17
16 20-12-2010 127414 01
Probe specifications DLC-250:
Sensor type : Differential TC type T
Probe length : 192 mm (7.56 in)
Temperature range : 0C to 250C / 32F to482F
Diameter : OD3 mm
Depth : 140 mm (5.51 in)
Media compatibility : AISI316
Connection : Redel plug with build inmemory is standard
Probe specifications DLC-700:
Sensor type : Differential TC type N
Probe length : 177 mm (6.96 in)
Temperature range : 0C to 700C / 32F to1292F
Diameter : OD4 mm
Depth : 200 mm (7.9 in)
Media compatibility : INCONEL 600
Connection : Redel plug with build inmemory is standard
7/30/2019 User Manual Dlc
17/17
127414 01 20-12-2010 17
Certificate:
If the DLC probes are supplied with a certificate, thecalibration is carried out as recommended below
according to the ITS 90 temperature scale.
The DLC-155/158 probe is as standard calibrated inthe range :
45C to 155C (-49F to 311F).
The DLC-250 probe is as standard calibrated in therange :
28C to 250C (82F to 482F).
The DLC-700 probe is as standard calibrated in therange :
33C to 700C (91.4F to 1292F).