User Guide spectro

Embed Size (px)

Citation preview

  • 7/31/2019 User Guide spectro

    1/6

    Application of Hellma TrayCell or Implen Label-

    Guard microliter measurement cells in the Eppendorf

    BioPhotometer plus for nucleic acid determinations

    Martin Armbrecht, Eppendorf AG, Hamburg, Germany

    No 031 I BioPhotometer plus

    Userguide

    The application of microliter measurement cells from Hellma and Implen together with the BioPhotometer plus

    allows for the determination of very high DNA and RNA concentrations in extremely small volumes. Due to the

    simple handling and the convenient evaluation, this offers a low-cost alternative in comparison with determination

    using spectrophotometers, which are also able to perform measurements in this range.

    The primary purpose of this Userguide is to optimize the handling of the microliter measurement cells in order to

    achieve reproducible results.

    Abstract

    After isolation and purification of DNA and RNA, it is often

    necessary to determine the concentration of the nucleic

    acid for downstream applications. This is usually performed

    using a spectrophotometer. Since isolated nucleic acids

    are mostly highly concentrated and therefore above the

    measurement range of the photometer, it is first necessary

    to dilute the samples.

    The individual dilution steps have to be performed very ac-

    curately, otherwise the calculation of the starting concen-

    tration may quickly become incorrect. It is also problematic

    that the diluted samples mostly cannot be used in down-

    stream applications afterwards.

    Introduction

    A convenient alternative is provided by the possibility of

    measuring the samples directly without any additional dilu-

    tion using microliter measurement cells in combination with

    the BioPhotometer plus. Only a few microliters (0.7 l 5 l)

    are sufficient for the measurement. With this method highly

    concentrated samples are measured using a very short

    light path, e.g., 0.2 mm or 1 mm instead of the usual 10 mm

    light path. With a shorter light path at a certain wavelength,

    light can pass through the sample even at high concentra-

    tions and measurements can reproducibly be performed.

    Since the BioPhotometer plus does not require warm-up

    times, the measurement cells can directly be used for DNA

    or RNA measurements after switching on the device.

  • 7/31/2019 User Guide spectro

    2/6

    Userguide No 031 | Page 2

    For both the 0.2 mm and the 1 mm lid it is recommended to work with an extinction range between 0.1 and 1.5 E.

    Table 1 shows concentration ranges for different nucleic acids in comparison to a standard cuvette with a 10 mm light path.

    As can be seen in Table 1, when using microliter measurement cells significantly higher concentrations can be determined

    than in cuvettes with a 10 mm light path.

    Figure 1: Schematic representation of the micro cell from Hellma

    (TrayCell) or Implen (LabelGuard).

    a) Light, b) Mirror c) Sample, d) Lid, e) Window f) Optical path length.

    Concentration range

    Table 1: Concentrations of nucleic acids by application of recommended extinction ranges (0.1 1.5 E)

    Measuring principle

    Figure 1 depicts an overview of the measuring principle.

    The light (a) of the photometer is guided into the interior

    of the measurement cell through the window (e), it is then

    directed upwards via a mirror (b) and is guided through

    another window (e) into the sample (c). In the lid the light

    is guided via a mirror (b) back to the measurement cell and

    transmitted via another mirror out of the cuvette. The

    detector of the photometer measures the remaining

    amount of light (absorption). The optical path length (f)

    that is important for the measurement is derived from the

    distance between the mirror and the upper exit window of

    the measurement cell and is important for the determina-

    tion of the sample concentration. The optical path lengthis determined by the lid: depending on the concentration

    and the sample volume, lids with light path of 0.2 or 1 mm

    can be used. The microliter measurement cells are always

    delivered with both lids (Figure 2a or 2b), with labeling

    indicating the usage of the corresponding light path on the

    TrayCell lid or LabelGuard [1].

    2a) Lid for 0.2 mm light path 2b) Lid for 1 mm light path

    Nucleic acid Factor 1 mm lid [ng/l] 0.2 mm lid [ng/l] Cuvette with 10 mm lightpath [ng/l]

    dsDNA 50 50 - 750 250 3750 5 75

    ssDNA 37 37 - 555 185 2775 3.7 55.5

    RNA 40 40 600 200 3000 4 60

    Oligo 33 33 - 495 165 2475 3.3 49.5

  • 7/31/2019 User Guide spectro

    3/6

    Userguide No 031 | Page 3

    The microliter measurement cells can be directly applied,

    without dilution factors. The prerequisite is the adjustment

    of the value for the optical path length of the lid in the

    parameter settings at the Biophotometer plus. The settings

    will be saved only for the particular method applied. For the

    calculation of the concentration the applied light path will

    be automatically taken into account.

    Figure 3 shows the necessary adjustment of the Biopho-

    tometer plus for the direct usage of the 0,2 mm or 1mm lid.

    a)

    b)

    Figure 3: Adjusting the light path to 0.2 mm (a) or 1 mm (b) (e.g.,

    for dsDNA).

    Depending on the lid used, the values for the corresponding optical path

    length have to be modified in the parameter mode.

    To alter the value of the optical path length in the device

    settings, choose a method from a method group (e.g., the

    method dsDNA in the group DNA). When the arrow in

    the display points to the desired method, press param-

    eter within the operating field. Press the keys with the

    arrows until cuvette is displayed and choose the optical

    path length (see Figure 3).

    When using the BioPhotometer 6131, the optical path

    length of 0.2 mm cannot be directly set on the device.

    To use the device without the need of conversion a virtual

    50-fold dilution can be entered while using the preset light

    path of 10 mm. For this, press dilution and enter 1 l for

    sample and 49 l for diluent.

    These settings also lead to a correct result.

    Settings of the BioPhotometer plus Adjustment with the BioPhotometer 6131

    It is recommended to wear gloves during measurements

    to avoid contamination of the windows for light entry or

    exit. Before each measurement the cuvette should always

    be pressed as far as it will go into the cuvette chamber.

    Discrepancies in regard to the position of the measure-

    ment cell in the cuvette chamber may lead to measurement

    differences.

    Like for measurements in standard cuvettes with 10 mm

    light path, frozen samples should always be completely

    thawed and then sufficiently mixed.

    Between two measurements the mirror in the lid as well as

    the window in the upper side of the measurement cell have

    to be cleaned. For this purpose conventional cutton swabs

    or usual tissue paper may be used. Remaining debris, lint

    or dust should be removed using compressed air. Com-pressed air is available in office supply stores. In case of

    stronger contamination the mirror at the inner side of the

    lid (Figure 1a, b) and the window in the upper side (Figure

    1a, e) have to be cleaned with 60% isopropanol.

    Further information regarding cleaning can be obtained

    from Hellma (www.hellma-worldwide.com).

    Handling and cleaning

  • 7/31/2019 User Guide spectro

    4/6

    1. Wear gloves during the whole measurement procedure

    to avoid contamination of the measurement cell.

    2. For your measurement, select a lid with the optical path

    length that best corresponds to the concentration range

    of your sample (see Table 1).

    3. Choose the measurement method (e.g., dsDNA).

    4. If needed, change the preset parameter in BioPhoto-

    meter plus (see setting of the BioPhotometer plus,

    Fig. 3a or Fig. 3b).

    5. Determine the zero value (blank) with the liquid,

    which was used to dissolve or to dilute your sample.

    For this purpose pipette the blank onto the upper

    measurement window of the sample (Fig. 4a) or use

    the 0.2 mm light path on the mirror in the inner side

    of the lid (Fig. 4b).

    Userguide No 031 | Page 4

    The manufacturers recommend using volumes of 3 - 5 l

    for the 1 mm lid and 0.7 - 3 l for the 0.2 mm lid, respec-

    tively. When using the 0.2 mm lid together with very small

    liquid volumes like, e.g., 1 l, the formation of small air

    bubbles within the light path may occur, causing variations

    during measurements of the same sample. For very small

    volumes below 2 l it is therefore recommended to pipette

    Volumes

    Figure 4: Loading of the measurement cell at 1 mm (a) or 0.2 mm light

    path (b)

    Optical

    path

    length

    Figure 5: Entry display before the blank measurement

    Figure 6: Blank measurement

    a) b)

    Measuring Procedure

    6. Before the measurement of the blank the preset

    parameters like the applied light path are displayed on

    the screen (Fig. 5). Place the lid onto the measurement

    cell corresponding to the notch on the lid and press the

    blank key on the BioPhotometer plus (Fig. 6).

    the sample directly onto the mirror in the inner side of the

    lid (Fig. 4b), instead of the usual pipetting on the upper

    measurement window (Fig. 4a).

    Since the work presented here is always performed with

    very small volumes, the pipetting steps should be very

    accurate. Of course, this is also true for pipetting the blank

    solutions.

  • 7/31/2019 User Guide spectro

    5/6

    7. The result of the blank measurement should be 0

    (see Figure 7).

    8. Next, remove all liquid remnants from the inner side of

    the lid (Fig. 8a) using cutton swabs or tissue paper

    (Fig. 8b).

    9. Remove debris, lint or dust from the mirror (Fig. 9a) and

    the measurement window (Fig. 9b) using a compressed

    air sprayer, if necessary.

    Userguide No 031 | Page 5

    Figure 7: Result of the blank measurement

    Figure 10: Pressing the Sample key

    Figure 11: Example of a result

    Figure 8: Cleaning of the upper measurement window (a) and the

    inner side of the lid (b)

    a) b)

    Figure 9: Removing debris, lint or dust from the inner side of the lid (a)

    and the upper measuring window (b) with compressed air

    a) b)

    10. Apply your sample as shown in Fig. 4 for the blank

    measurement and press the Sample key (Fig. 10).The result (Fig. 11) will be displayed automatically.

    The applied light path will be automatically included.

    11. (Optional) If necessary the sample can be recovered

    with a pipette from the measurement window on top.

    12. Next, remove all liquid remnants from the inner side of

    the lid using a cutton swabs or a tissue paper (Fig. 8b).

    13. If needed, clean both areas additionally with 60%

    isopropanol.

  • 7/31/2019 User Guide spectro

    6/6

    Userguide No 031 | Page 6

    References

    [1] C. Voolstra, A. Jungnickel, L. Borrmann, R. Kirchner, A. Huber

    Spectrophotometric quantification of nucleic acids - LabelGuardTM allows quantification of sample amounts in submicroliter

    range with commercial spectrophotometers. Application Note Implen GmbH, www.implen.com

    eppendorfi

    saregisteredtrademark.

    Allrightsreserved,includinggraphicsandimages.

    A

    U03139020/GB1/0309/0T/KW

    Ordering Information Eppendorf

    Product Order no. International Order no. North America

    BioPhotometer plus 230 V / 50-60 Hz 6132 000.008 952000006

    Thermoprinter DPU 414, incl. mains adapter and printer cable 230 V 6131 011.006 952010140

    Thermo paper, 5 rolls 0013 021.566 952010409

    Ordering Information Hellma GmbH & Co. KG

    Product Order no. International

    Hellma TrayCell, 68.5 mm, 8.5 mm, (incl. lid for layer thickness

    of 0.2 mm and 1 mm)

    105.800-UVS Z 8.5 mm

    Lid for TrayCell, optical path length of 1.0 mm 665.703

    Lid for TrayCell, optical path length of 0.2 mm 665.704

    Ordering Information - Implen GmbH

    Product Order no. International

    LabelGuard Microliter Cell for measurement, pack with 0.2 and 1 mm

    path length lid, 0.7 - 5 l sample volume

    LG100UV-G

    LabelGuard Microliter Cell for measurement, pack with 1 mm path length,

    3-5 l sample volume

    LG100UV

    LabelGuard Microliter Cell for measurement, pack with 0.2 mm path

    length, 0.7 - 4 l sample volume

    LG101UV

    Implen is a registered trademark of Implen GmbH

    LableguardTM is protected trademark of Implen GmbH

    Hellma is registered trademark of Hellma GmbH & Co. KG

    TrayCellTM is protected trademark of Hellma GmbH & Co. KG

    Your local d istr ibutor: www.eppendorf .com/worldwide

    Eppendorf AG 22331 Hamburg Germany Tel: +49 40 538 01-0 Fax: +49 40 538 01-556 E-mail: [email protected]

    Eppendorf North America, Inc. One Cantiague Road P. O. Box 1019 Westbury, N.Y. 11590-0207 USA

    Tel: +1 516 334 7500 Toll free phone: +1 800 645 3050 Fax: +1 516 334 7506 E-mail: [email protected]

    Application Support Europe, International: Tel: +49 1803 666 789 E-mail: [email protected]

    North America: Tel: +1 800 645 3050 ext. 2258 E-mail: [email protected]

    Asia Pacific: Tel: +60 3 8023 6869 E-mail: support_asiapacif [email protected]