Upload
sundeepjaglan
View
215
Download
0
Embed Size (px)
Citation preview
7/31/2019 User Guide spectro
1/6
Application of Hellma TrayCell or Implen Label-
Guard microliter measurement cells in the Eppendorf
BioPhotometer plus for nucleic acid determinations
Martin Armbrecht, Eppendorf AG, Hamburg, Germany
No 031 I BioPhotometer plus
Userguide
The application of microliter measurement cells from Hellma and Implen together with the BioPhotometer plus
allows for the determination of very high DNA and RNA concentrations in extremely small volumes. Due to the
simple handling and the convenient evaluation, this offers a low-cost alternative in comparison with determination
using spectrophotometers, which are also able to perform measurements in this range.
The primary purpose of this Userguide is to optimize the handling of the microliter measurement cells in order to
achieve reproducible results.
Abstract
After isolation and purification of DNA and RNA, it is often
necessary to determine the concentration of the nucleic
acid for downstream applications. This is usually performed
using a spectrophotometer. Since isolated nucleic acids
are mostly highly concentrated and therefore above the
measurement range of the photometer, it is first necessary
to dilute the samples.
The individual dilution steps have to be performed very ac-
curately, otherwise the calculation of the starting concen-
tration may quickly become incorrect. It is also problematic
that the diluted samples mostly cannot be used in down-
stream applications afterwards.
Introduction
A convenient alternative is provided by the possibility of
measuring the samples directly without any additional dilu-
tion using microliter measurement cells in combination with
the BioPhotometer plus. Only a few microliters (0.7 l 5 l)
are sufficient for the measurement. With this method highly
concentrated samples are measured using a very short
light path, e.g., 0.2 mm or 1 mm instead of the usual 10 mm
light path. With a shorter light path at a certain wavelength,
light can pass through the sample even at high concentra-
tions and measurements can reproducibly be performed.
Since the BioPhotometer plus does not require warm-up
times, the measurement cells can directly be used for DNA
or RNA measurements after switching on the device.
7/31/2019 User Guide spectro
2/6
Userguide No 031 | Page 2
For both the 0.2 mm and the 1 mm lid it is recommended to work with an extinction range between 0.1 and 1.5 E.
Table 1 shows concentration ranges for different nucleic acids in comparison to a standard cuvette with a 10 mm light path.
As can be seen in Table 1, when using microliter measurement cells significantly higher concentrations can be determined
than in cuvettes with a 10 mm light path.
Figure 1: Schematic representation of the micro cell from Hellma
(TrayCell) or Implen (LabelGuard).
a) Light, b) Mirror c) Sample, d) Lid, e) Window f) Optical path length.
Concentration range
Table 1: Concentrations of nucleic acids by application of recommended extinction ranges (0.1 1.5 E)
Measuring principle
Figure 1 depicts an overview of the measuring principle.
The light (a) of the photometer is guided into the interior
of the measurement cell through the window (e), it is then
directed upwards via a mirror (b) and is guided through
another window (e) into the sample (c). In the lid the light
is guided via a mirror (b) back to the measurement cell and
transmitted via another mirror out of the cuvette. The
detector of the photometer measures the remaining
amount of light (absorption). The optical path length (f)
that is important for the measurement is derived from the
distance between the mirror and the upper exit window of
the measurement cell and is important for the determina-
tion of the sample concentration. The optical path lengthis determined by the lid: depending on the concentration
and the sample volume, lids with light path of 0.2 or 1 mm
can be used. The microliter measurement cells are always
delivered with both lids (Figure 2a or 2b), with labeling
indicating the usage of the corresponding light path on the
TrayCell lid or LabelGuard [1].
2a) Lid for 0.2 mm light path 2b) Lid for 1 mm light path
Nucleic acid Factor 1 mm lid [ng/l] 0.2 mm lid [ng/l] Cuvette with 10 mm lightpath [ng/l]
dsDNA 50 50 - 750 250 3750 5 75
ssDNA 37 37 - 555 185 2775 3.7 55.5
RNA 40 40 600 200 3000 4 60
Oligo 33 33 - 495 165 2475 3.3 49.5
7/31/2019 User Guide spectro
3/6
Userguide No 031 | Page 3
The microliter measurement cells can be directly applied,
without dilution factors. The prerequisite is the adjustment
of the value for the optical path length of the lid in the
parameter settings at the Biophotometer plus. The settings
will be saved only for the particular method applied. For the
calculation of the concentration the applied light path will
be automatically taken into account.
Figure 3 shows the necessary adjustment of the Biopho-
tometer plus for the direct usage of the 0,2 mm or 1mm lid.
a)
b)
Figure 3: Adjusting the light path to 0.2 mm (a) or 1 mm (b) (e.g.,
for dsDNA).
Depending on the lid used, the values for the corresponding optical path
length have to be modified in the parameter mode.
To alter the value of the optical path length in the device
settings, choose a method from a method group (e.g., the
method dsDNA in the group DNA). When the arrow in
the display points to the desired method, press param-
eter within the operating field. Press the keys with the
arrows until cuvette is displayed and choose the optical
path length (see Figure 3).
When using the BioPhotometer 6131, the optical path
length of 0.2 mm cannot be directly set on the device.
To use the device without the need of conversion a virtual
50-fold dilution can be entered while using the preset light
path of 10 mm. For this, press dilution and enter 1 l for
sample and 49 l for diluent.
These settings also lead to a correct result.
Settings of the BioPhotometer plus Adjustment with the BioPhotometer 6131
It is recommended to wear gloves during measurements
to avoid contamination of the windows for light entry or
exit. Before each measurement the cuvette should always
be pressed as far as it will go into the cuvette chamber.
Discrepancies in regard to the position of the measure-
ment cell in the cuvette chamber may lead to measurement
differences.
Like for measurements in standard cuvettes with 10 mm
light path, frozen samples should always be completely
thawed and then sufficiently mixed.
Between two measurements the mirror in the lid as well as
the window in the upper side of the measurement cell have
to be cleaned. For this purpose conventional cutton swabs
or usual tissue paper may be used. Remaining debris, lint
or dust should be removed using compressed air. Com-pressed air is available in office supply stores. In case of
stronger contamination the mirror at the inner side of the
lid (Figure 1a, b) and the window in the upper side (Figure
1a, e) have to be cleaned with 60% isopropanol.
Further information regarding cleaning can be obtained
from Hellma (www.hellma-worldwide.com).
Handling and cleaning
7/31/2019 User Guide spectro
4/6
1. Wear gloves during the whole measurement procedure
to avoid contamination of the measurement cell.
2. For your measurement, select a lid with the optical path
length that best corresponds to the concentration range
of your sample (see Table 1).
3. Choose the measurement method (e.g., dsDNA).
4. If needed, change the preset parameter in BioPhoto-
meter plus (see setting of the BioPhotometer plus,
Fig. 3a or Fig. 3b).
5. Determine the zero value (blank) with the liquid,
which was used to dissolve or to dilute your sample.
For this purpose pipette the blank onto the upper
measurement window of the sample (Fig. 4a) or use
the 0.2 mm light path on the mirror in the inner side
of the lid (Fig. 4b).
Userguide No 031 | Page 4
The manufacturers recommend using volumes of 3 - 5 l
for the 1 mm lid and 0.7 - 3 l for the 0.2 mm lid, respec-
tively. When using the 0.2 mm lid together with very small
liquid volumes like, e.g., 1 l, the formation of small air
bubbles within the light path may occur, causing variations
during measurements of the same sample. For very small
volumes below 2 l it is therefore recommended to pipette
Volumes
Figure 4: Loading of the measurement cell at 1 mm (a) or 0.2 mm light
path (b)
Optical
path
length
Figure 5: Entry display before the blank measurement
Figure 6: Blank measurement
a) b)
Measuring Procedure
6. Before the measurement of the blank the preset
parameters like the applied light path are displayed on
the screen (Fig. 5). Place the lid onto the measurement
cell corresponding to the notch on the lid and press the
blank key on the BioPhotometer plus (Fig. 6).
the sample directly onto the mirror in the inner side of the
lid (Fig. 4b), instead of the usual pipetting on the upper
measurement window (Fig. 4a).
Since the work presented here is always performed with
very small volumes, the pipetting steps should be very
accurate. Of course, this is also true for pipetting the blank
solutions.
7/31/2019 User Guide spectro
5/6
7. The result of the blank measurement should be 0
(see Figure 7).
8. Next, remove all liquid remnants from the inner side of
the lid (Fig. 8a) using cutton swabs or tissue paper
(Fig. 8b).
9. Remove debris, lint or dust from the mirror (Fig. 9a) and
the measurement window (Fig. 9b) using a compressed
air sprayer, if necessary.
Userguide No 031 | Page 5
Figure 7: Result of the blank measurement
Figure 10: Pressing the Sample key
Figure 11: Example of a result
Figure 8: Cleaning of the upper measurement window (a) and the
inner side of the lid (b)
a) b)
Figure 9: Removing debris, lint or dust from the inner side of the lid (a)
and the upper measuring window (b) with compressed air
a) b)
10. Apply your sample as shown in Fig. 4 for the blank
measurement and press the Sample key (Fig. 10).The result (Fig. 11) will be displayed automatically.
The applied light path will be automatically included.
11. (Optional) If necessary the sample can be recovered
with a pipette from the measurement window on top.
12. Next, remove all liquid remnants from the inner side of
the lid using a cutton swabs or a tissue paper (Fig. 8b).
13. If needed, clean both areas additionally with 60%
isopropanol.
7/31/2019 User Guide spectro
6/6
Userguide No 031 | Page 6
References
[1] C. Voolstra, A. Jungnickel, L. Borrmann, R. Kirchner, A. Huber
Spectrophotometric quantification of nucleic acids - LabelGuardTM allows quantification of sample amounts in submicroliter
range with commercial spectrophotometers. Application Note Implen GmbH, www.implen.com
eppendorfi
saregisteredtrademark.
Allrightsreserved,includinggraphicsandimages.
A
U03139020/GB1/0309/0T/KW
Ordering Information Eppendorf
Product Order no. International Order no. North America
BioPhotometer plus 230 V / 50-60 Hz 6132 000.008 952000006
Thermoprinter DPU 414, incl. mains adapter and printer cable 230 V 6131 011.006 952010140
Thermo paper, 5 rolls 0013 021.566 952010409
Ordering Information Hellma GmbH & Co. KG
Product Order no. International
Hellma TrayCell, 68.5 mm, 8.5 mm, (incl. lid for layer thickness
of 0.2 mm and 1 mm)
105.800-UVS Z 8.5 mm
Lid for TrayCell, optical path length of 1.0 mm 665.703
Lid for TrayCell, optical path length of 0.2 mm 665.704
Ordering Information - Implen GmbH
Product Order no. International
LabelGuard Microliter Cell for measurement, pack with 0.2 and 1 mm
path length lid, 0.7 - 5 l sample volume
LG100UV-G
LabelGuard Microliter Cell for measurement, pack with 1 mm path length,
3-5 l sample volume
LG100UV
LabelGuard Microliter Cell for measurement, pack with 0.2 mm path
length, 0.7 - 4 l sample volume
LG101UV
Implen is a registered trademark of Implen GmbH
LableguardTM is protected trademark of Implen GmbH
Hellma is registered trademark of Hellma GmbH & Co. KG
TrayCellTM is protected trademark of Hellma GmbH & Co. KG
Your local d istr ibutor: www.eppendorf .com/worldwide
Eppendorf AG 22331 Hamburg Germany Tel: +49 40 538 01-0 Fax: +49 40 538 01-556 E-mail: [email protected]
Eppendorf North America, Inc. One Cantiague Road P. O. Box 1019 Westbury, N.Y. 11590-0207 USA
Tel: +1 516 334 7500 Toll free phone: +1 800 645 3050 Fax: +1 516 334 7506 E-mail: [email protected]
Application Support Europe, International: Tel: +49 1803 666 789 E-mail: [email protected]
North America: Tel: +1 800 645 3050 ext. 2258 E-mail: [email protected]
Asia Pacific: Tel: +60 3 8023 6869 E-mail: support_asiapacif [email protected]