If you can't read please download the document
Upload
marion-hunter
View
229
Download
0
Embed Size (px)
DESCRIPTION
Echo IF WE GO TO A BAD HALL , THE SPEAKER ‘S VOICE WILL COME BACK TO YOU WE CANT HEAR ANY THING WHAT DO WE CALL THIS ? Echo
Citation preview
ULTRASOUND THE BASICS
Dr. S. Parthasarathy MD, DA, DNB, Dip Diab.MD ,DCA, Dip software
based statistics, PhD (physiology) FICA Echo IF WE GO TO A BAD HALL
, THE SPEAKER SVOICE WILL COME BACK TO YOU WE CANT HEAR ANY THING
WHAT DO WE CALL THIS ? Echo In ultrasound The same thing The
machine sends waves ,
it echoes from our tissues Comes back to the machine Converted to
digital display for us to see But it is ultrasound waves not sound
Pulse echo principle Frequency 20 KHz man Just above range cats ,
dolphins hear ??
frequencies used clinically lie largely between 3 MHz and 15 MHz.
There are no naturally occurring sources or detectors of such waves
Two main uses Regional anesthesia Vascular access Safety Efficacy
To generate the ultrasound
an electrical current is applied to the crystal (piezo electric)
component within the transducer face. current is then converted to
mechanical (ultrasound) energy and transmitted to the tissues at
very high (megahertz) frequencies. The ultrasound energy produced
then travels through the tissues as pulsed, longitudinal,
mechanical waves In echo Mouth speaks and ears hear the reflected
waves
One machine part does it Transducer One reflector and one receiver
not enough So waves are continuously sent from many transducers !!
ARRAY Phased - Linear curvilinear- arrays Phased Phased arrays are
ideally suited to applications with small access windows, e.g.,
scanning the heart while avoiding the lung field. linear,
curvilinear, and phased
Linear arrays - rectangular field of view suited to applications
with reasonable access and a need to cover a lot of tissue close to
the surface. Curvilinear arrays rely on being able to apply some
pressure to the skin surface in order to make it deform to their
shape. Once done , a very wide field of view can be obtained. A
single probe from 3 13 Mhz
Patient mode Vascular Nerves Small parts 3 - 10 MHz 8- 10
superficial structures 5- 8 medium distance
3 5 deeper structures Sound travels speed m/s 1540 for all scanners
Higher frequency energy is lost to the tissues to a greater extent
than lower frequency energy with progressive tissue penetration.
Irrespective of frequency, lateral and axial resolution always
decrease with increasing tissue depth Lateral perpendicular to the
USG beam Axial parallel to USG beam Selecting transducer
Probes
Frequency Array Foot print or the diameter of the transducer
Attenuation ? Hyperechoic - solids liver stones Hypoechoic
vessels
Continuous echoes 20 frames /second Hypo and hyper echoic Anechoic
and iso echoic Artifacts Something is there but not seen
Some thing is not there but seen Acoustic Enhancement Reverberation
Three types Acoustic Shadowing occurs when structures that are
highly attenuating of ultrasound
energy (such as bone) Nerves !! Near subclavian or enhancement
artifact
Passes through structures of different attenuation The deeper one
may show more reflections and enhanced image Nerves !! Near
subclavian or enhancement artifact Reverberation artifact
Each successive ultrasound pulse emitted from the transducer
produces a temporal echo resulting in a series of parallel lines
both above and below the actual object. commonly seen with highly
attenuating wide-bore needles Beyond the transducer From the
transducer we have to get the image
Which is meaningful !! Put the probe assess the image assess which
is marker side Use gain and depth !! Low gain Brightens everything
to compensate resolution
Medium and high gain Brightens everything to compensate resolution
Time gain compensation Depth Interscalene 2 cm Femoral 4 cm
The resolution will decrease as soon as the depth goes B and M
modes A and B are static C and D are moving B = brightness
M = motion (cardiac) Why use gel ?? Coupling Agent Ultrasound does
not travel through air
Gel provides medium for sound wave into the body by removing air
gap Povidone-iodine, surgical spirit Colour doppler Select doppler
and select site In Colour Doppler mode, blood flow towards the
transducer will be red on screen, flow away from the transducer
will be blue on screen. It must be emphasized that colour does not
necessarily characterize arterial or venous flow. Anechoic lumen,
non-pulsatile, compressible
Anechoic lumen, non-pulsatile, compressible. Valsalva effect,
doppler - continuous flow Short axis Anechoic lumen , pulsatile,
non-compressible. Doppler - pulsatile flow
Long axis Tendon Hyperechoic withbright lines longitudinally or
bright dots at right angles -fibrillary pattern Hypoechoic with
multiple hyperechoic lines muscle Nerve Liver normal and cirrhosis
Normal kidney and kidney stone Save clip , record , patient
,type
Name Age sex Type of scan ---- done Save clip , photo , Review ,
list Caliper This is used to measure a distance (eg kidney
length).
It is used by selecting a starting spot by pressing a kidney key
and using the trackball to measure to a second mark. The distance
between the two marks will then be displayed on screen measured in
cm. This can be used with other functions such as Res/Freeze.
Freeze caliper mouse fix point mouse another point distance will be
shown. Knobology Trackball ( mouse with select buttons) Zoom
magnification of areas of the ultrasound picture.
Looking at Res/Zoomed areas of interest has the advantage of a more
detailed view drawback of less anatomy visible Movements of the
transducer
sliding, tilting, rotating, angling. To create an image orient the
transducer such that the left and right sides are consistent with
the left and right sides of the ultrasound screen, position the
plane of the screen similar to the plane of the transducer Sample
survey of the area Tilting, rotating and angling Slightly give
tilting motion to the transducer get the target to 90*
Anisotropy Rotation from short axis to long axis Angling Gentle
pressure may be applied to one side of the transducer to maintain
full contact along the skin surface in concave or confined fields,
or when attempting to direct the beam under a superficial highly
attenuating structure such as bone. Focus Lateral resolution
better
Different echoic structures better seen Some machines if we adjust
depth focus is automatically changed Beam width decreased Future
developments Two probes in one Three-dimensional probes
. Needle tips which emit an ultrasound signal that is detected by
the probe and visualized as a pulsating spot Summary Ultrasound
echo , array , types frequency , speed, focus
Depth Movements of the transducer Artifacts gain and TGC B and M
modes freeze , caliper , trackball Future developments