28
+ Basic Ultrasound Physics Dr. A. Ferguson

Ultrasound

Embed Size (px)

Citation preview

Page 1: Ultrasound

+

Basic Ultrasound PhysicsDr. A. Ferguson

Page 2: Ultrasound

+Objectives

Level 1 Knowledge Components

Ultrasound physics, terminology, and safety

Equipment care, ultrasound techniques, and controls

Page 3: Ultrasound

+The transducer

Damper material

Piezoelectric crystal(sends and receives)

Acoustic lens

Impedancematching to skin

CABLE

Page 4: Ultrasound

+The ultrasound wave

Am

plit

ude (

dB

)

Wavelengthl

1 second: cycles/second = frequency(Hz)Clinical use varies from 2.5-20MHz

Velocity m/s1540 m/s approx

Delivered in pulses (bursts)• Length of pulse varies• Frequency of pulse varies

V = fl

Page 5: Ultrasound

+Velocity in tissue

Medium Velocity of US (m/sec)

Air 330

Fat 1450

Water 1480

Soft tissue 1540

Kidney 1560

Blood 1570

Muscle 1580

Bone 4080

Page 6: Ultrasound

+The ultrasound beam

Near zone

Unfocused transducer

Divergenceangle

Side lobes

Beam width

Focal zone

Focused transducer

Page 7: Ultrasound

+Ultrasound beam lobes

Feldman M K et al. Radiographics 2009;29:1179-1189

Page 8: Ultrasound

+ResolutionAxial (along length of beam) – most precise

Smallest resolvable distance = 2 x l Higher frequency = better resolution Independent of depth

Lateral (across beam) Varies with depth Within focal zone may be as good as axial

Elevational (within the slice) Slice might be 3-8mm wide with some probes Strong reflectors at edges may appear in centre

Contrast (shades of gray)

Page 9: Ultrasound

+Ultrasound/tissue interaction

Transducer

Scattering

Reflection

Refraction

Attenuation

Tissue interface

Skin

1 2

3 4

Page 10: Ultrasound

+Scattering

Structures with radius < wavelength scatter US

e.g. RBCs and micro-structures within tissues

Scattering is multidirectional

Only small portion of incident US gets back to probe

Scattering from RBCs contributes to DOPPLER effect

Tissue scattering results in speckled appearance

Page 11: Ultrasound

+Reflection

Critical to image generation

Depends on: Angle of beam relative to tissue Change in acoustic impedance* across

boundary Smooth tissue boundaries act almost as

mirrors Called “specular reflectors” e.g. pleura

* Acoustic impedance = tissue density x US velocity in the tissue

Page 12: Ultrasound

+Acoustic impedance

Medium Acoustic impedance*

Air 0.0004

Lung 0.18

Fat 1.34

Liver 1.65

Blood 1.65

Kidney 1.63

Muscle 1.71

Bone 7.8* x106 Rayls

Page 13: Ultrasound

+Refraction

Waves deflected passing through interface

Can be useful in focusing US waves

Results in artefacts

Page 14: Ultrasound

+Attenuation

Loss of US energy as it passes through tissue

Depends on Attenuation coefficient of tissue Frequency of transducer Distance from transducer Intensity of transmitted US

AIR has a very large attenuation coefficient

Lower frequencies penetrate better than high

Page 15: Ultrasound

+Attenuation values

Medium Half-power distance (cm)

Water 380

Blood 15

Soft-tissue (non-muscle)

1-5

Muscle 0.6-1

Bone 0.2-0.7

Air 0.08

Lung 0.05

Page 16: Ultrasound

+Image artifacts

Poor image quality

Images of structures that are either Not there at all Present in a different location than image

suggests

Lack of visualisation of structures

Images that differ in size or shape from reality

Some artifacts are clinically useful

Page 17: Ultrasound

+Image artifacts Acoustic shadowing

Acoustic enhancement

Refraction artifact

Reverberation artifact

Comet-tail artifact

Mirror-image artifact

Ghosting artifact

Beam-width artifact

Ring-down artifact

Speed-displacement artifact

Page 18: Ultrasound

+Beam-width artifact

Feldman M K et al. Radiographics 2009;29:1179-1189

Grey dot assumed to be in main beam areaArea of interest outside focal zone

Adjust focal zone

Grey dot outside beamArea of interest inside focal zone

Page 19: Ultrasound

+Side-lobe artifact

Feldman M K et al. Radiographics 2009;29:1179-1189

Black dot signal may return from multiple side-lobes resulting in duplication on screen

Page 20: Ultrasound

+Reverberation artifact

Feldman M K et al. Radiographics 2009;29:1179-1189

US bounces back and forth between two strong reflectors

Page 21: Ultrasound

+Ring-down artifact

Feldman M K et al. Radiographics 2009;29:1179-1189

Ring of bubbles with fluid trapped centrally. Fluid vibrations detected as strong signal and displayed as line behind true source.

Page 22: Ultrasound

+Mirror-image artifacts

Feldman M K et al. Radiographics 2009;29:1179-1189

US beam bounces between structure and deeper strong reflector e.g. diaphragm.This means probe receives signals as if from same object on other side of reflector.

Page 23: Ultrasound

+Speed-displacement artifact

Feldman M K et al. Radiographics 2009;29:1179-1189

Part of beam encounters tissue where velocity is much lower than 1540 m/s,e.g. fat. Returning signal appears to come from deeper in body.

Discontinuous diaphragm sign

Page 24: Ultrasound

+Refraction artifact

Feldman M K et al. Radiographics 2009;29:1179-1189

Refraction at an interface between two objects makes the deeper object appear in false location.

Page 25: Ultrasound

+Acoustic shadowing

Feldman M K et al. Radiographics 2009;29:1179-1189

Strong attenuator means weak beam beyond = shadow

Page 26: Ultrasound

+Acoustic enhancement

Feldman M K et al. Radiographics 2009;29:1179-1189

Signal behind weak attenuator is stronger than at same level in adjacent tissues.Gives impression of brighter structures deep to low attenuator =enhancement

Page 27: Ultrasound

+ Probe typesSector Linear array Curved array

Page 28: Ultrasound

+ Use of Gain

GainMin

Max

Near field Far field

Attenuation

Time-gain compensation (TGC)

Pro

cess

ed

Ori

gin

al