Transmission Electron Microscopy Dr Heath Bagshaw – CMA [email protected]

Embed Size (px)

Citation preview

  • Slide 1

Transmission Electron Microscopy Dr Heath Bagshaw CMA [email protected] Slide 2 What is the difference between SEM and TEM? Scanning Electron Microscope Transmission Electron Microscope In an SEM, electrons are scanned over the surface of the sample. In a TEM, electrons are transmitted through the sample. Slide 3 Electrons pass through a (very thin) sample (i.e. are transmitted) to form an image. Simplistically, In its operation a TEM can be thought of as analogous to a slide projector:- What is a Transmission Electron Microscope (TEM)? slide Objective lens Condenser lens Light source Slide projector Projection screen Fluorescent screen Objective lensCondenser lens Specimen (thin)aperture Electron source Electron beam TEM Slide 4 What is a Transmission Electron Microscope (TEM) cont. In a slide projector white light from the bulb is focussed onto the slide by a condenser lens. This produces an image which is focussed by the objective lens to a point (crossover). The image is then projected onto a screen and magnified by varying the distance from the crossover to the screen. In a TEM the electron beam is focussed on the sample using the condenser lens system. This produces an image which is focussed by the objective lens to a point (crossover). This image is then magnified by a series of projector lenses to vary the size of the image on a fluorescent screen. Changing the current of an electromagnetic lens alters its focal length altering magnification. Slide 5 TEMs have basically the same design independent of manufacturer or age:- Anatomy of a TEM Apertures Viewing Screen Electron Gun Condenser lens system Apertures Sample Viewing Screen a)b)c) Examples of Transmission Electron Microscopes, a)Philips EM420 (1980s), b) Cut through Schematic of Philips CM 200 (1990s) and c) JEOL 2100 (2000s) JEOL 2100 has digital imaging and computer assisted operation. Slide 6 main components of a TEM The TEM can be broken down into a few main components, these are:- The Gun which produces electrons. The condenser system which forms the probe. The sample sample preparation is important, and time consuming. Image formation use of image plane or back focal plane. Projection of the image (magnification). Viewing and recording the image. Lets start by looking at how electrons are produced. Slide 7 Electron Guns Two main types of gun Thermionic and Field Emission. Thermionic sources produce electrons when heated. Field emission sources produce electrons when exposed to an intense electric field. FEGs give much more brightness than thermionic systems. FEGs give a more monochromatic electron source and finer probe (i.e. better resolution). Comparison of the three types of source operating at 100kV Slide 8 The Condenser System The Wehnelt (or 2 nd anode in a FEG) focuses the beam to a crossover which is accelerated down the column. The first condenser de-magnifies the crossover to give a smaller point source this is referred to as C 1 or spot size. The second condenser lens (C2) is used to either converge or spread the beam of illumination on the sample (intensity (Philips) or brightness (JEOL)). A condenser aperture is placed in the beam path to remove electrons far from the optic axis which would reduce resolution. The smaller the aperture the better the resolution, but there is an associated decrease in brightness need to compromise. Slide 9 The Sample Samples are typically 3mm in diameter and