  • Thermo Scientific Orion VERSA STAR™ pH/ISE/Temperature Module

    Reference Guide

  • Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries.

    AQUAfast, AQUASensors, BOD AutoEZ, ionplus, KNIpHE, LogR, No Cal, ORION, perpHect, PerpHecT, pHISA, pHuture, Pure Water, Sage, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS, and TURBO2 are registered trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries.

    A+, All in One, Aplus, AUTO-BAR, AUTO-CAL, Auto-ID, AUTO-READ, AUTO-STIR, Auto-Test, AutoTration, CISA, digital LogR, DuraProbe, EZ Startup, ISEasy, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, Optimum Results, Orion Dual Star, Orion Star, SAOB, SMART AVERAGING, SMART STABILITY, Star LogR, Star Navigator 21, Stat Face, Triode are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries.

    Guaranteed Success and The Technical Edge are service marks of Thermo Fisher Scientific, Inc. and its subsidiaries.

    © 2011 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific, Inc. and its subsidiaries.

    The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice.

    This publication supersedes all previous publications on this subject.

  • Thermo Scientific Orion VERSA STAR™ pH/ISE/Temperature Module

    Table of Contents

    Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

    Chapter 2 Module Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

    Module Connections and Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Module Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Information From Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

    Chapter 3 Setup Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    Setup Menus for Primary Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Read Type Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6Isopotential Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7Linear Regression Information . . . . . . . . . . . . . . . . . . . . . . . .7Low Level Stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7

    Setup Menu for Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    Chapter 4 pH, RmV and ORP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

    pH Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9RmV Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9ORP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

    Chapter 5 ISE Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

    Chapter 6 Temperature Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

    Chapter 7 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

    pH/mV/RmV/ORP Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15ISE Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

    Chapter 8 Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

    Chapter 9 Incremental Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

    Known Addition and Subtraction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Single Known Addition . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20Double Known Addition . . . . . . . . . . . . . . . . . . . . . . . . . . . .20Single Known Subtraction . . . . . . . . . . . . . . . . . . . . . . . . . .21Double Known Subtraction . . . . . . . . . . . . . . . . . . . . . . . . .21

    Analyte Addition and Subtraction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22To Access Incremental Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

  • Chapter 10 Data Storage and Review . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

    Data Storage Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Measurement Read Type . . . . . . . . . . . . . . . . . . . . . . . . . . .23Datalog, Computer and Printer Export Settings . . . . . . . . . .23

    Datalog Review . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Calibration Review . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

    Chapter 11 Customer Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

    Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Warranty and Registration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25WEEE Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26pH/ISE/Temperature Module Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

    Appendix Advanced Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

    Automatic pH Buffer Recognition Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Testing a pH Electrode for Automatic Buffer Recognition . 33

    pH Electrode Condition Icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Multipoint Calibration Curve Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Automatic Blank (Non-Linear) Correction Feature . . . . . . . . . . . . . . . . . . . . . . . 36Determining an Isopotential Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

  • 1Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Chapter 1 Introduction

    Thank you for your purchase of the Thermo Scientific Orion VERSA STAR™ pH/ISE/Temperature Module.

    This module is for use with the Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter.

    Some key features of this module include:

    • pHrangefrom-2to20withresolutionoptionsof0.001,0.01and0.1

    • ISErangefrom0to19999withresolutionofuptofoursignificantdigits

    • mVandRmVrangesfrom±2000mV

    • Automaticandmanualdataloggingoptions

    • Upto6-pointpHcalibrationwithautomaticrecognitionforUSA/NISTandDINbuffers

    • Upto6-pointISEcalibrationwithadjustableISOpointcapabilities

    • TheabilitytofixpH/ISEcalibrationerrorswithoutacompleterecalibrationthankstocalibrationediting

    • AdvancedISEfeaturessuchasauto-blank,timedendpoint,linearpoint-to-point,non-linearselectable

    auto-blank,low-concentrationrangestability,ISEIDandincrementaltechniques

    Pleasereadthisreferenceguidethoroughly.Anyuseoutsideoftheseinstructionsmayinvalidateyourwarrantyand

    causepermanentdamagetothemeter.

  • This page intentionally left blank.

  • 3Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Chapter 2 Module Overview

    Toconnectthemodulestothemeterandforspecificinformationregardingthemeteroutputs,pleaseseethe

    OrionVERSASTARReferenceGuide.TheReferenceGuideisontheCDincludedwiththemeterandalsoavailable

    at www.thermoscientific.com/water.

    Module Connections and Inputs

    TheOrionVERSASTARmetercanacceptOrioncombinationISEelectrodesaswellasseparatereferenceandsensing

    half-cellelectrodes.ThisallowsforawiderISEmeasurementoffering.Half-cellelectrodesseparatetheionssensing

    fromthereferenceelectrode.Fordifficultsamples,theflexibilityofusingalternatefillsolutionsmaybeanadvantage

    to researchers.

    1. PreparethepHorISEelectrodeandATCprobeforuseasinstructedintheelectrodeuserguides.

    The pin tip reference connector is on the lowest panel of the pH/ISE module.

    2. PlacethepHorISE,ATCprobeandOrionStarstirrerprobe(cat.no.096019,optional)intothe

    electrode arm.

    3. AttachthepHorISEBNCandATCprobetothepH/ISEmetermodule.LowerISEandATCinto

    solutionforconditioningifrequired.Ifusingthestirrerprobe,attachprobetothemeter,placeintothe

    electrodearmandassigntheprobetothechannelinthesetupmenuforinstrumentsettings.

    4. Preparefreshstandardsdailyforcalibration.ForISEtesting,standardsshouldbeatleastonedecade

    (10times)apart.Thestandardsshouldexceedtheexpectedminimumandmaximumsamplerange.

    Serialdilutionisrecommendedusingcalibratedpipettestoaccuratelypreparecalibrationstandards

    from a stock solution.

    5. Gatherreagents,rinsewater,andothersupplies.

    ForISEtesting,ifISAwillbeaddedtothesample,itneedstobeaddedtoallthestandardspriorto

    calibrationtoensureaconsistentdilutionfactor.

    pH/ISE Module

    8pinmini-DINforATC

    BNCconnectionwithshortingcap

    Reference pin socket

  • 4 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Module Maintenance

    Forroutinemaintenance,dustandwipethemodulewithadampcloth.Ifnecessary,warmwaterormildwater-based

    detergentcanbeused.Maintenancecanbeperformedonadaily,weeklyormonthlybasis,asrequiredbyoperating

    environment.Immediatelyremoveanyspilledsubstanceusingthepropercleaningprocedureforthatspilltype.

    Display Information from Module

    ExampleISEMeasurementScreen–SingleChannel

    Dateandtime

    Primaryreading

    Calibrationinformation

    CorrespondingmV value

    Method

    Theinformationonthedisplaycanbecustomized.Enterthesetupmenutomakechanges.1.PressSetup.2.UsethearrowkeystohighlightInst.Settings, press (f3/select) key.3.UsethearrowkeytohighlightDisplayView. Press (f3/select) key.4. Follow the onscreen prompts to choose what you would like to see on the measurement screen.

  • 5Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Chapter 3 Setup Menus

    Setup Menus for Primary Measurements

    1. PressSetup (f3) to access the Main Setup Screen.

    2. UsethearrowkeytoscrollacrosstothemodulestoISEchannelyouwishtouseand

    press Select (f3).

    3. Scroll to across to the mode menu and press Select (f3).

    ISE Setup Menu (Default values are in bold in the table below)ElectrodeSerialNumber InputElectrode Type SeetablebelowSampleID Off,Manual,AutoIncrementSignificantDigits 1,2,3,4digitsStability Smart,Fast,Medium,SlowAveraging Off,Automatic Smart Read Type AutoRead,Timed,ContinuousMeasureUnit ppm,Molar,mg/L,%,ppb,NoneIsopotential Value Off,OnLinearRegression Off,OnBlank Correction Off,OnLowLevelStability Off,OnAlarm** Limit,CalDue,SetPoint

    **Limitandsetpointalarmdefaultsetto‘off’. Calduealarmsetto12hours.

    pH Setup Menu (Default values are in bold in the table below)

    ElectrodeID InputSampleID Off,Manual,AutoIncrementResolution 0.1,0.01,0 .001Stability Smart,Fast,Medium,SlowAveraging Off,Automatic SmartRead Type AutoRead,Timed,ContinuousBuffer Group USA,DIN,CustomIsopotential Value Off,On 7 .000LinearRegression Off,OnAlarm** Limit,CalDue,SetPoint

    **Limitandsetpointalarmdefaultsetto‘off’. Calduealarmsetto12hours.

  • 6 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    4. ForISEtesting,selecttheISEelectrodebeingusedinsetup.Ifyourelectrodeisnotlisted,

    usethe“X”electrodewiththevalence(+1,+2,-1,or-2)matchingyourelectrodefor

    accurate results.

    Read Type Information

    Therearethreemainreadtypesavailableasshowninthesetupmenus.

    • Auto-Read – Themeterwilldisplaythemeasurementasitstabilizesandlockandholdthe

    measurementwhenitisstable(TheARiconwilllockonthescreen.)Oncethemeasurementislocked,

    themeterwillautomaticallyexportthemeasurementtothedatalog,ifthedatalogisenabledin

    thesetupmenu,andtoaprinterorcomputer,ifaprinterorcomputerisconnectedtothemeterand

    enabledinthesetupmenu.Totakeanewmeasurement,pressmeasure (esc)/1.

    • Timed – Themeterwillreadanddisplayameasurementatthesettimeintervalthatisprogrammed

    bytheoperator.Therearetwooptions:

    1. Bychecking/selectingthe"Interval"box,atthesettimeinterval,themeterwillautomatically

    exportthemeasurementtothedatalog,ifthedatalogisenabledinthesetupmenu,andto

    aprinterorcomputer,ifaprinterorcomputerisconnectedtothemeterandenabledinthe

    setupmenu.Ifselected,enterthetimeintervalvalueinahours:minutes:secondsformatusing

    the numeric keypad and press f3 (select) key.

    2. Ifthe"Interval"boxisunchecked/notselected,asingletimedmeasurementwillbetaken.

    Themeterwillapplythesamesavingandexportingconditionsaswiththesettimeintervalto

    recordandexportdata,butlockafteronetimedintervalhasexpired.Pressmeasure (esc)/1 to

    resetthecountdownclockandtakeanewreading.

    ISE TypesAg+ I- X- CO2 REDOX(ORP) Cl2 S-2 ClO4 O2Cd+2 NH3 BF-4 K+ X-2 Cu+2 X+ F- SCN-CN- Pb+2 Cl- NO3- Br- NOx Ca+2 Na+ X+2

    ORP, RmV and mV Setup Menus (Default values are in bold in the table below)

    ElectrodeID InputSampleID Off,Manual,AutoIncrementResolution 1.0,0 .1Stability Smart,Fast,Medium,SlowAveraging Off,Automatic SmartRead Type AutoRead,Timed,ContinuousAlarm** Limit,CalDue*,SetPoint

    *Only for ORP and RmV modes. **Limitandsetpointalarmdefaultsetto‘off’. Calduealarmsetto12hours.

  • 7Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    • Continuous – The meter will continuously measure and update the display. This read type is useful

    whenperforminganexperimentthatrequirescontinuousmeasurementstobetaken.Pressthe

    log/print/0printkeytoexportthemeasurementtothedatalog,ifthedatalogisenabledinthesetup

    menu,andtoaprinterorcomputer,ifaprinterorcomputerisconnectedtothemeterandenabledin

    the setup.

    Isopotential Value

    Theslopeofallelectrodeschangeswithtemperature,basedontheNerstequation.Theslopecurvesatthedifferent

    temperaturesintersectattheisopotentialvalue,whichistheconcentrationatwhichthepotentialoftheelectrodedoes

    notvarywithtemperature.ForglasspHelectrodes,thisvalueis7.00.Forotherionselectiveelectrodes(ISEs),ifthe

    isopotentialvalueisknownorcanbemeasuredexperimentally,adjustingthisvaluemakestemperaturecompensation

    foranISEpossible.Temperaturecompensationcanbeperformedforoneandtwopointcalibrations.

    Linear Regression Information

    Thedefaultisoffwhichusesthetraditional,segmentedapproach.Thesegmentedapproachisthemostaccurateand

    precise.MoreinformationcanbefoundintheappendixsectionofthisreferenceguideundertheMultipointCalibration

    Curve Feature.

    Selecting“on”forthisoptionresultsintheuseofasinglebestfitlineforthecalibrationcurve,andcanbebeneficialif

    pHmeasurementsvarygreatly.

    Low Level Stability

    ThisfeatureadjuststhetimingforthecalibrationpointsforlowerlevelISEstandards.

    Setup Menu for Temperature

    AllofthetemperaturemeasurementsettingsarefoundinTemperatureSetupmenuforeachmeasurement.

    *Thisnumberwillchangetomatchthesetupmenuchannel.

    Temperatureprobescanbesetupinthemeasurementmodeofthesampleyouaremeasuring.Ch-1ATCallowsthe

    temperatureprobetobeusedinmorethanonechannelfortemperature-compensatedmeasurements,suchaswhen

    apHelectrodeanda2-cellconductivitycellareinthesamemeasurementsample.TemperatureisacriticalpHand

    conductivitymeasurementasthemeasurementvaluesrequireatemperaturecorrectiontodeterminethereference

    value,often25ºCintheAmericaand20ºCinEurope.

    Temperature Unit ºCorºFselectionTemperature Input and Type Ch-1ATC,Ch-2ATC,Ch-3ATC,Ch-4ATC,MTC

    Temperature Cal Ch-1*ATConlyorAllATCallowsallATCstobe calibratedatonce

  • 8 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    This page intentionally left blank.

  • 9Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Chapter 4 pH, RmV and ORP Calibration

    pH Calibration

    UptosixpHbufferscanbeusedforcalibration.Thiscanbeacombinationofautomatically-recognizedUSA/NISTor

    DINbuffervaluesandcustombuffers.AlwaysusefreshpHbuffersandselectbuffersthatbracketthesamplepHand

    areatleastonepHunitapart.PreparethepHelectrodeaccordingtotheinstructionsintheelectrodeuserguide.

    1. Pressf1 (Cal)toperformapHcalibration.

    2. Followscreenpromptsforcalibration.Usefreshbuffers,rinseelectrodewithDIwaterbeforeeach

    reading,blotwithalintfreetissue.

    3. USAandDINbuffersetsareautomaticallyrecognizedbasedofftheselectioninthesetupmenu,

    andthemeterallowscustombufferstobeusedforupto6calibrationpoints.Custombufferoption

    prompts for the nominal values and stores data.

    4. Duringcalibration,pointscanbeeditedorretestedpriortoacceptance.

    5. Informationforthelastcalibrationcanbedisplayedonthemainmeasurementscreen.Refertothis

    referenceguide,Chapter2ModuleOverview,DisplayInformationsectionformoreinformation.

    RmV Calibration

    OnestandardcanbeusedforrelativemVcalibration.Alwaysusefreshstandard.Preparetheelectrodeaccordingto

    theinstructionsintheelectrodeuserguide.

    1. Pressf1 (Cal) toentercalibrationmode.

    2. Followthescreenpromptsforcalibration.

    3. Thecalibrationcanbesavedorprintedwhencomplete.

    4. Informationforthelastcalibrationcanbedisplayedonthemainmeasurementscreen.Refertothis

    referenceguide,Chapter2ModuleOverview,DisplayInformationsectionformoreinformation.

  • 10 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    ORP Calibration

    ThemVreadingofanORPstandardisdependentontheORPelectrodereferencesystem,referencematerial,filling

    solutionandORPstandardtemperature.AnORPelectrodecanbeautomaticallycalibratedtoread420mVrelativeto

    theStandardHydrogenElectrode(EH)at25°CwhentheOrionORPstandard(cat.no.967901or96791),OrionORP

    electrode(cat.no.9678BNWP,9778BNWP,9180BNMDor9179BNMD)andOrion4MKClfillingsolution(cat.no.

    900011)or4MKClgel(inelectrodewithcat.no.9179BNMD)areused.TherawmVreadingofanelectrodewiththe

    sameconditionsis220mVat25°C.

    OnestandardcanbeusedforORPcalibration.Alwaysusefreshstandard.Preparetheelectrodeaccordingtothe

    instructionsintheelectrodeuserguide.

    1. Pressf1 (Cal)toentercalibrationmode.

    2. Followthescreenpromptsforcalibration.

    3. Thecalibrationcanbesavedorprintedwhencomplete.

    4. Informationforthelastcalibrationcanbedisplayedonthemainmeasurementscreen.Refertothis

    referenceguide,Chapter2ModuleOverview,DisplayInformationsectionformoreinformation.

  • 11Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Chapter 5 ISE Calibration

    OnetosixstandardscanbeusedforISEcalibration.Ifmorethanonestandardisused,startwiththelowest

    concentrationfirst.Alwaysusefreshstandards.Selectstandardsthatbracketthesampleconcentrationandarea

    decadeapart.Preparetheionselectiveelectrodeaccordingtotheelectrodeuseguide.

    1. Presschannel key until ISE channel is displayed. Press Cal (f1).Followthescreenpromptsstarting

    with the lowest concentration.

    2. PressEdit (f3)tothecorrectconcentrationvalueofastandardifitisnotautomaticallyrecognized.

    3. Afterthefirstdatapointhasbeenaccepted,pressAccept (2) tocontinuecalibration.

    The meter then offers options of Esc (f1)toleavecalibration,Next (f2)measurenextcalibration

    point,orCal Done (f3).

    4. TheoptionsafterCalDoneispressedarePrint (f1),Cal Edit (f2) or Slope Edit (f3).

    5. TheCalibrationreporttableisdisplayedwiththeconcentration,mVreading,andTemperatureand

    averageslopewhichisupdatedwitheachsuccessivepointmeasured.

    6. Fromthisscreenthenavigationoptionsare Print (f1),Cal Edit (f2) or Slope Edit (f3).Printexports

    datatoprinterorcomputer,Caleditallowsconcentrationtobecorrectedpriortothecompletionof

    thecalibrationbyDelete (f1) or ReMeasure (f3).

    7. PressingtheSlopeEditchangesthescreentitletoSinglePointCal–SlopeEdit.Thisallowsthe

    currentslopelinetobeupdatedwiththepresentelectrodeperformance.

    8. Press Cal Done (f1)whencompleted,pressviewgraph,savethecalibrationinCalLog.

  • 12 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    This page intentionally left blank.

  • 13Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Chapter 6 Temperature Calibration

    TheATCtemperaturedisplayhasarelativeaccuracyof±0.1°C.ATCprobes(bothseparateandthosebuiltinto

    electrodes)havevaryingtemperatureaccuracies,usually±0.5°Cto±2°C.Use this function only if necessary.

    Sincethetemperatureoffsetcalculatedduringthecalibrationisappliedtoallfuturetemperaturemeasurements,

    recalibrateifadifferentATCprobeisused.Fortemperaturecalibration,theprobeneedstobeconnectedtothemeter

    andthecalibrationsolutionshouldhaveaknown,stabletemperature.ItisrecommendedthattwoNIST-traceable

    thermometersbeusedtomeasureandverifysolutiontemperature.

    1. Pressf3 (setup) key.

    2. Usethearrowkeystohighlightthechannelthatthetemperatureprobeisconnectedto.

    Press f3 (select).

    3. UsethearrowkeystohighlightTemperatureandpressf3 (select).

    4. UsethearrowkeystohighlighttheappropriateTemperatureCaloptionandpressf3 (select).

    5. Follow the onscreen prompts.

  • 14 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    This page intentionally left blank.

  • 15Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Chapter 7 Measurement

    pH/mV/RmV/ORP Measurement

    Afterasuccessfulcalibration,thesamplescanbetested.

    1. Rinsetheelectrodewithdeionizedwater.Blotdrywithlint-freetissue.

    2. Placeelectrode/ATCprobeinfirstsampleandstirgently.

    3. IfthemeterisinAutoReadmode(meterdefault)pressmeasure (esc) (1) key. If the meter is in

    continuousortimedreadmode,themeterwillimmediatelystarttakingreadings.

    4. IfinAutoReadmodeanddatalogisenabled(default),thereadingwillautomatically

    bestoredwhenthe“AR”appears.Ifincontinuousreadmodeanddatalogisenabled,

    press log/print (0) keytostoreintothemeter’smemory.

    Note: Toenableordisabledatalogging,presssetupandselectInst.Settings.

    ISE Measurement

    1. Preparethesampleasinstructedintheelectrodeuserguide.

    2. FollowdirectionscarefullyonISAandothersampleadditions.

    3. Rinsetheprobe(s)andplaceintothesample.IfthemeterisinAutoReadmode(meterdefault)press

    measure (esc) (1)key.Ifthemeterisincontinuousortimedreadmode,

    themeterwillimmediatelystarttakingreadings.

    4. IfinAutoReadmodeanddatalogisenabled(default),thereadingwillautomaticallybestoredwhen

    the“AR”appears.Ifincontinuousreadmodeanddatalogisenabled,presslog/print (0) key to store

    intothemeter’smemory.

    Note:Toenableordisabledatalogging,presssetupandselectInst.Settings.

  • 16 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    This page intentionally left blank.

  • 17Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Chapter 8 Methods

    Themeterstoresuptotenmethodsperchannelwiththelastcalibrationdatastoredforeachmethod.

    Usethechannelsetuptoaccessandmodifymethods.

    1. Pressf3 (setup) key.

    2. Usethearrowkeystohighlighttheappropriatechannel.Pressf3 (select).

    3. UsethearrowkeystohighlightMethodandpressf3 (select).

    4. Themethodscanbeloaded,copied,savedandeditedbyfollowingtheonscreenprompts.

  • 18 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    This page intentionally left blank.

  • 19Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Chapter 9 Incremental Techniques

    Known Addition and Subtraction

    TheOrionVERSASTARmeterhaspresetmethodsforsingleknownaddition,singleknownsubtraction,doubleknown

    additionanddoubleknownsubtraction.

    • Whenperformingincrementaltechniques,thesampleconcentrationmustbeknownwithinan

    orderofmagnitudesotheincrementmaybecorrectlychosenandthesampleandstandardmustbe

    volumetrically determined.

    • Itisveryimportantthatanycomplexingagentispresentinexcess(atleast50to100times)ornot

    atall.Theratiooffreetocomplexedionsmustremainconstantovertheaddition.Anindicationthat

    acomplexingagentispresent,butnotingreatenoughquantitytomaintainaconstantfreeionto

    complexedionratio,isanabnormallyhighorlowslope.Ifthisisasuspectedproblem,addanexcess

    amountofcomplexingordecomplexingagenttotheoriginalsolutioninthebeaker.

    • Electrodeinterferencesshouldbeataminimum,sincetheeffectoftheinterferencemightchange

    astheconcentrationoftheionofinterestchanges.Thissituationmayalsoresultinhighorlow

    electrodeslopes.Consulttheelectrodeuserguideforinformationonreducingoreliminatingsome

    electrode interferences.

    • Knownadditionisausefulmethodformeasuringdilutesamples,occasionalsamplesorsamplesthat

    containanexcessofcomplexingagent.Knownadditioncanalsobeusedtoverifytheresultsofa

    direct measurement.

    • Knownsubtractionisusefulwhenmeasuringionsforwhichstablestandardsdonotexist.Known

    subtractioncanalsobeusedasaquickversionofatitration.Whenperformingaknownsubtraction,

    itisnecessarytoknowthestoichiometricratiobetweenthestandardandsample.

    • Doubleincrementaltechniquescalculatetheelectrodeslopeandsampleconcentration

    simultaneously.Theelectrodeslopeisdetermineddirectlyinthesample,whichresultsingreater

    accuracyforsampleswithcomplexmatricesandgreateranalysisspeed.

  • 20 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Single Known Addition

    A standard of known concentration is added to a sample of unknown concentration. The total concentration should

    approximatelydoubleupontheadditionofthestandard,thereforesampleconcentrationshouldbeknownwithinan

    orderofmagnitude.Refertothefollowingtabletochoosethecorrectstandardconcentrationandvolumeofaddition.

    • Refertotheappropriateelectrodeuserguideforinstructionsforpreparingtheelectrode,required

    solutions(ionicstrengthadjusters,standards,etc)andanyspecialrequirements.Determinethe

    electrodeslopeasdirectedintheelectrodeuserguide.

    • Ensurethattheelectrode,ATCprobeandreferenceelectrode(ifapplicable)areconnectedtothe

    meter inputs that correspond with the selected channel.

    Double Known Addition

    Formostdoubleknownadditionanalysis,itisrecommendedthat:

    • Thesamplevolumeshouldbe100mL.

    • Theconcentrationofthestandardaddedshouldbe100timestheexpectedsampleconcentration.

    • Calculatethevolumeofstandardrequiredtohalvetheinitialsampleconcentrationanddivideby

    six.Fivesixthsofthevolumewillbeaddedinthefirstadditionandtheremainderinthesecond.If

    available,consultaprintoutfromapreviousanalysisofasimilarsample.Note:theproposedvaluesfor

    constantmVwillbesmallerthanthoserecommendedforknownaddition.Thisisanefforttominimize

    mechanceofover"subtracting"thesample(overshootingtheendpoint).

    Refertotheappropriateelectrodeuserguideforinstructionsforpreparingtheelectrode,requiredsolutions

    (ionicstrengthadjusters,standards,etc)andanyspecialrequirements.

    Ensurethattheelectrode,ATCprobeandreferenceelectrode(ifapplicable)areconnectedtothemeterinputsthat

    correspond with the selected channel.

    Amount of standard to be added per 100 mL of sample

    Standard concentration compared to expected sample concentration

    1mL 100timesmoreconcentrated5mL 20timesmoreconcentrated10mL 10timesmoreconcentrated

  • 21Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Single Known Subtraction

    Inknownsubtraction,eachadditionofstandardsubtractsanincrementofthespeciestobemeasured.Enterthe

    correctreactionratioofthespeciestobemeasuredwiththestandardtobeadded.Forexample,asulfide(S-2)sample

    istobesubtractedbytheadditionofsilver(Ag+).Twosilverionsarerequiredtosubtracteachsulfideionandthe

    reactionratiois0.5.

    Sampleconcentrationshouldbeknownwithinanorderofmagnitudeandthetotalsampleconcentrationshould

    beapproximatelyhalveduponadditionofthestandard.Refertothefollowingtabletochoosethecorrectstandard

    concentration and volume of addition.

    Refertotheappropriateelectrodeuserguideforinstructionsforpreparingtheelectrode,requiredsolutions(ionic

    strengthadjusters,standards,etc)andanyspecialrequirements.Determinetheelectrodeslopeasdirectedinthe

    electrodeuserguide.

    Ensurethattheelectrode,ATCprobeandreferenceelectrode(ifapplicable)areconnectedtothemeterinputsthat

    correspond with the selected channel.

    Double Known Subtraction

    Ensurethatthesubtractionabilityofthechosenstandardconcentrationandvolumeaddeddoesnotexceedthe

    concentration of sample.

    Formostdoubleknownsubtractionanalysis,itisrecommendedthat:

    • Thesamplevolumeshouldbe100mL.

    • Theconcentrationofthestandardaddedshouldbe50timestheexpectedsampleconcentration.

    • Calculatethevolumeofstandardrequiredtohalvetheinitialsampleconcentrationanddivideby

    six.Fivesixthsofthevolumewillbeaddedinthefirstadditionandtheremainderinthesecond.If

    available,consultaprintoutfromapreviousanalysisofasimilarsample.Note:theproposedvaluesfor

    constantmVwillbesmallerthanthoserecommendedforknownaddition.Thisisanefforttominimize

    mechanceofover"subtracting"thesample(overshootingtheendpoint).

    Refertotheappropriateelectrodeuserguideforinstructionsforpreparingtheelectrode,requiredsolutions(ionic

    strengthadjusters,standards,etc)andanyspecialrequirements.

    Ensurethattheelectrode,ATCprobeandreferenceelectrode(ifapplicable)areconnectedtothemeterinputsthat

    correspond with the selected channel.

    Amount of standard to be added per 100 mL of sample

    Standard concentration compared to expected sample concentration

    1mL 50timesmoreconcentrated5mL 10timesmoreconcentrated10mL 5 times more concentrated

  • 22 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Analyte Addition and Subtraction

    InAnalyteAddition,thefirstmeasurementisinaknownstandardsolutioncontainingthesametypeofanalytein

    interest as the sample. The second measurement follows the addition of the unknown solution. The concentration of

    the unknown is then calculated.

    InAnalyteSubtraction,thefirstmeasurementusesknownstandardsolutioncontainingthetypeofionsthattheISE

    canmeasureandthatwillcomplexwiththeanalyteininterestintheunknownsample.Thesecondmeasurement

    follows the addition of the unknown sample that reduces the concentration of the ions in the measurement solution.

    The concentration of the analyte in interest in the unknown is then calculated.

    • Isopotentialvalueneedstobeknownandentered.

    • Blankcorrectionisnotapplied.

    • Theslopevalueat25°Cneedstobeknownandentered.

    To Access Incremental Techniques

    1. SetthemeasurementmodetoISE.

    2. Pressf3 (setup) key.

    3. UsethearrowkeystohighlighttheISEchannel.Pressf3 (select).

    4. UsethearrowkeystohighlightIncrementalTechniquesandpressf3 (select).

    5. Follow the onscreen prompts.

  • 23Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Chapter 10 Data Storage and Review

    Data Storage Settings

    Measurement Read Type

    The measurement read type determines when the meter sends measurements to outputs as selected in the setup menu

    and when properly connected to the meter. These are set individually for each channel. Measurement Read Type is set

    perchannelinthesetupmenu.Pleasereferchapter3SetupMenusinthisreferenceguideformoreinformation.

    Datalog, Computer and Printer Export Settings

    OrionVERSASTARMetershavea2000-datalogcapability.Tomakeyourselection:

    1. Frommeasurementmode,presssetup/3.

    2. PressthearrowkeystohighlightInst.Settings.Presssetup/3.

    3. Usethearrowkeys,f2 (page) key and f3 (select) key to:a. set the communication protocol

    b.turnonprinting

    c. select the print format

    d.turnondatalogging

    Formoreinformationonthesesettings,refertotheOrionVERSASTARReferenceGuide,Chapter3InstrumentSettings

    toturnonDataLogtoenabledatastorage.

    Datalog Review

    Thisoptionallowsforstoreddatatobe:

    • reviewed

    • printedindividually,inagrouporall

    • deleted

    • analyzedwithgeneralstatisticalinformation

    Thedatalogwillreflectascreencaptureofalltheinformationatthetimeofdatalogging.

    Toreviewdataloginformation

    • Presslog view (9) key

    • Pressf3 (select) key

    • Followpromptstoreviewinformation

  • 24 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Calibration Review

    Thisfeatureallowsforreviewandprintingofeachofthelast10calibrationsforeachparameterofthemeter.

    Theinformationwillmatchtheinformationrecordedforthatcalibration.

    Toreviewcalibrationinformation

    1. Presslog view (9) key.

    2. Pressright arrow (6) keytoselectCalLoginformationanddown arrow (8)tohighlightthe

    desired parameter.

    3. Press f3 (select) key and follow the prompts.

  • 25Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Chapter 11 Customer Services

    Troubleshooting Guide

    Assistance

    Aftertroubleshootingallcomponentsofyourmeasurementsystem,contactTechnicalSupport.WithintheUnitedStatescall1.800.225.1480andoutsidetheUnitedStatescall978.232.6000orfax978.232.6031.InEurope,theMiddleEastandAfrica,contactyourlocalauthorizeddealer.Forthemostcurrentcontactinformation,orthelatestapplicationandtechnicalresourcesforThermoScientificOrionproducts,visitwww.thermoscientific.com/water.

    Warranty and Registration

    Toregisteryournewmeterandforthemostcurrentwarrantyinformation,visitwww.thermoscientific.com/water.

    WEEE Compliance

    ThisproductisrequiredtocomplywiththeEuropeanUnion’sWasteElectrical&ElectronicEquipment(WEEE)Directive2002/96/EC.Itismarkedwiththesymbolabove. ThermoFisherScientifichascontractedwithoneormorerecycling/disposalcompaniesineachEUMemberStateandthisproductshouldbedisposedoforrecycledthroughthem.Furtherinformationoncompliancewiththesedirectives,therecyclersinyourcountry,andinformationonThermoScientificOrionproductsthatmayassistthedetectionofsubstancessubjecttotheRoHSDirectiveareavailableatwww.thermoscientific.com.

    Error Recommended Actions

    Measurement is flashing9999andoverrangeorunderrange.

    Measurementisoutsidetheallowablemeasurementrange.Checkelectrodeconnectionandsettingsinthesetupmenu.Cleantheelectrodeaccordingtotheelectrodeuserguideandrecalibratetheelectrodewithnewbuffersorstandards. Iftheerrorcontinues,performthemeterselftestinthediagnosticsmenu.

    The measurement freezesandwill notchange.

    ThemeterisinAutoReadmode.(ARiconappearssolidonthedisplay.)Press measure (esc)/1keytotakeanewreadingorgotosetuptochangethereadtype.

  • 26 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Declaration of Conformity

    Manufacturer: Thermo Fisher Scientific Inc .

    Address: AyerRajahCrescent Blk55#04-16/24 Singapore139949 Singapore

    Hereby declares that the following products:Orion VERSA STAR Advanced Electrochemistry Benchtop Meters with Modules are rated 100to240VAC,50/60Hz,0.5A.

    Equipment Class:

    Measurement,controlandlaboratoryOrionStarA-seriesmetersareEMCClassA

    Conforms with the following directives and standards:

    EN61326-1:2006 Electromagnetic Compatibility (EMC Directive) Electricalequipmentformeasurement, controlandlaboratoryuse-EMCrequirements

    EN61010-1:2001 Safety StandardsUL61010-1:2004 Safetyrequirementsforelectricalequipmentformeasurement,CAN/CSA C22 .2 No . 61010-1-04 controlandlaboratoryuse-generalrequirements

    CheowKwangChan QA/RegulatoryManager

    PlaceandDateofIssue: 12December,2011 Singapore

  • 27Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    pH/ISE Module (VSTAR-ISE) SpecificationspH Range -2.000to20.000

    Resolution 0.1,0.01,0.001Relative Accuracy ±0.002CalibrationPoints Upto6CalibrationEditing YesAdjustableISOPoint Yes

    mV/RmV Range ±2000.0mVResolution 0.1Relative Accuracy ±0.2mVor±0.05%ofreadingwhicheverisgreaterEH ORP Mode Yes

    ISE Range 0to19999Resolution Upto4significantdigitsRelative Accuracy ±0.2mVor±0.05%ofreadingwhicheverisgreaterUnits ppm,M,mg/L,%,ppb,noneCalibrationPoints Upto6CalibrationEditing YesAdjustableISOPoint Yes

    CalibrationFeatures Timedendpoint,linearpointtopoint,non-linearselectableauto-blank,lowconcentrationrangestability

    IncrementalTechniquesSingleknownaddition&subtraction, doubleknownaddition&subtraction, singleanalyteaddition&subtraction

    Advanced Features Displayselectrodetype,temperaturecompensationTemperature Range -5to105°C,23to221ºF

    Resolution 0.1Relative Accuracy ±0.1OffsetCalibration 1point

    Inputs pH/ISE:BNC,referencepin ATC:8-pinmini-DIN

    pH/ISE/Temperature Module Specifications

    NOTE:Wereservetherighttomakeimprovements.Specificationssubjecttochangewithoutnotice.

  • 28 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Ordering Information

    CML # DescriptionVSTAR00 Orion VERSA STAR Benchtop Meter

    -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR10 Orion VERSA STAR pH Benchtop Meter Set -OrionVERSASTARmeterwithpH/temperaturemodule -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR12 Orion VERSA STAR pH Benchtop Meter Kit -OrionVERSASTARmeterwithpH/temperaturemodule -8302BNUMDOrionROSSTriodepH/ATCprobe -096019OrionStarstirrerprobe -810199ROSSsolutionkit(475mLeachofpH4,7and10buffers; storagesolution;cleaningsolution;andpHelectrodestoragebottle) -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR20 Orion VERSA STAR Conductivity Benchtop Meter Set -OrionVERSASTARmeter -OrionVERSASTARconductivity/temperaturemodule -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR22 Orion VERSA STAR Conductivity Benchtop Meter Kit -OrionVERSASTARmeter -OrionVERSASTARconductivity/temperaturemodule -013005MDOrionDuraProbeconductivitycell,K=0.475 -011007conductivitystandard,1413µS,5x60mL -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR30 Orion VERSA STAR RDO/Dissolved Oxygen Benchtop Meter Set -OrionVERSASTARmeter -OrionVERSASTARdissolvedoxygen/temperaturemodule -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR32 Orion VERSA STAR RDO/Dissolved Oxygen Benchtop Meter Kit -OrionVERSASTARmeter -OrionVERSASTARdissolvedoxygen/temperaturemodule -083005MDOrionpolarographicDOprobe -080017calibrationsleeveforDOprobe -080513DOprobemaintenancekit -970802BODfunnel/stirrer -080360BODadapter -Electrodearmwithredesignedholder -Universalpoweradapter

  • 29Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    CML # DescriptionVSTAR40A Orion VERSA STAR pH/ISE Benchtop Meter Set

    -OrionVERSASTARmeter -OrionVERSASTARpH/ISE/temperaturemodule -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR40A2 Orion VERSA STAR pH/ISE Benchtop Meter Kit -OrionVERSASTARmeter -OrionVERSASTARpH/ISE/temperaturemodule -8102BNUWPOrionROSSUltrapHelectrode -096019OrionStarstirrerprobe -927007MDATCprobe,stainlesssteel -ROSSsolutionkit(475mLeachofpH4,7and10buffers;storagesolution; cleaningsolution;andpHelectrodestoragebottle) -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR40B Orion VERSA STAR Benchtop Meter with Two pH/ISE Modules -OrionVERSASTARmeter -TwoOrionVERSASTARpH/ISE/temperaturemodules -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR40B2 Orion VERSA STAR Benchtop Meter with Two pH/ISE Modules Kit -OrionVERSASTARmeter -TwoOrionVERSASTARpH/ISE/temperaturemodules -8102BNUWPOrionROSSUltrapHelectrode -096019OrionStarstirrerprobe -927007MDATCprobe,stainlesssteel -ROSSsolutionkit(475mLeachofpH4,7and10buffers;storagesolution; cleaningsolution;andpHelectrodestoragebottle) -Twoelectrodearmswithredesignedholders -Universalpoweradapter

    VSTAR50 Orion VERSA STAR pH/Conductivity Benchtop Meter Set -OrionVERSASTARmeter -OrionVERSASTARpH/temperaturemodule -OrionVERSASTARconductivity/temperaturemodule -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR52 Orion VERSA STAR pH/Conductivity Benchtop Meter Kit -OrionVERSASTARmeterwithpH/temperaturemodule -OrionVERSASTARconductivity/temperaturemodule -8157BNUMDOrionROSSUltraTriodepH/ATCprobe -013005MDOrionDuraProbeconductivitycell,K=0.475 -810199ROSSsolutionkit(475mLeachofpH4,7and10buffers; storagesolution;cleaningsolution;andpHelectrodestoragebottle) -011007conductivitystandard,1413µS,5x60mL -Electrodearmwithredesignedholder -Universalpoweradapter

  • 30 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    CML # DescriptionVSTAR80 Orion VERSA STAR pH with LogR Benchtop Meter Set

    -OrionVERSASTARmeter -OrionVERSASTARpH/temperaturewithLogRtechnologymodule -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR82 Orion VERSA STAR pH Benchtop Meter Kit -OrionVERSASTARmeter -OrionVERSASTARpH/temperaturewithLogRtechnologymodule -8172BNXOrionROSSSure-FlowpHelectrode -927007MDATCprobe,stainlesssteel -ROSSsolutionkit(475mLeachofpH4,7and10buffers;storagesolution; cleaningsolution;andpHelectrodestoragebottle) -810007ROSSfillsolution,5x60mL -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR90 Orion VERSA STAR pH/ISE/Conductivity/RDO/Dissolved Oxygen Benchtop Multiparameter Meter Set -OrionVERSASTARmeter -OrionVERSASTARpH/ISE/temperaturemodule -OrionVERSASTARconductivity/temperaturemodule -OrionVERSASTARdissolvedoxygen/temperaturemodule -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR91 Orion VERSA STAR pH/ISE/Conductivity/RDO/Dissolved Oxygen Benchtop Multiparameter Meter Set -OrionVERSASTARmeter -TwoOrionVERSASTARpH/ISE/temperaturemodules -OrionVERSASTARconductivity/temperaturemodule -OrionVERSASTARdissolvedoxygen/temperaturemodule -Two096019OrionStarstirrerprobes -Twoelectrodearmswithredesignedholders -Universalpoweradapter

  • 31Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    CML # DescriptionVSTAR92 Orion VERSA STAR pH/ISE/Conductivity/RDO/Dissolved Oxygen

    Benchtop Multiparameter Meter Set -OrionVERSASTARmeter -OrionVERSASTARpH/ISE/temperaturemodule -OrionVERSASTARconductivity/temperaturemodule -OrionVERSASTARdissolvedoxygen/temperaturemodule -8157BNUMDOrionROSSUltraTriodepH/ATCprobe -013005MDOrionDuraProbeconductivitycell,K=0.475 -083005MDOrionpolarographicDOprobe -080017calibrationsleeveforDOprobe -080513DOprobemaintenancekit -970802BODfunnel/stirrer -080360BODadapter -810199ROSSsolutionkit(475mLeachofpH4,7and10buffers; storagesolution;cleaningsolution;andpHelectrodestoragebottle) -011007conductivitystandard,1413µS,5x60mL -Electrodearmwithredesignedholder -Universalpoweradapter

    VSTAR93 Orion VERSA STAR pH/ISE/Conductivity/RDO/Dissolved Oxygen Ultimate Benchtop Multiparameter Meter Set -OrionVERSASTARmeter -OrionVERSASTARpH/ISE/temperaturemodule -OrionVERSASTARconductivity/temperaturemodule -OrionVERSASTARdissolvedoxygen/temperaturemodule -8157BNUMDOrionROSSUltraTriodepH/ATCprobe -013005MDOrionDuraProbeconductivitycell,K=0.475 -083005MDOrionpolarographicDOprobe -Two096019OrionStarstirrerprobes -080017calibrationsleeveforDOprobe -080513DOprobemaintenancekit -970802BODfunnel/stirrer -080360BODadapter -810199ROSSsolutionkit(475mLeachofpH4,7and10buffers; storagesolution;cleaningsolution;andpHelectrodestoragebottle) -011007conductivitystandard,1413µS,5x60mL -Twoelectrodearmswithredesignedholders -Universalpoweradapter

  • 32 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    AccessoriesCML # DescriptionVSTAR-PH Orion VERSA STAR pH/temperature moduleVSTAR-ISE Orion VERSA STAR pH/ISE/temperature moduleVSTAR-CND Orion VERSA STAR conductivity/temperature moduleVSTAR-RD OrionVERSASTARRDO/DO/temperaturemoduleVSTAR-LR OrionVERSASTARpH/temperaturewithLogRtechnologymoduleSTARA-BEA BenchtopelectrodearmforOrionStarA-seriesmetersSTARA-HB FreestandingbaseforusewithOrionStarA-seriesbenchtopelectrodearmSTARA-PWR Replacement universal power adapter for Orion VERSA STAR meters096019 OrionStarstirrerprobe8102BNUWP OrionROSSUltrapHelectrode,refillable,glassbody8172BNWP OrionROSSSure-FlowpHelectrode,glassbody8156BNUWP OrionROSSUltrarefillablepHelectrode,epoxybody8157BNUMD OrionROSSTriodepH/ATCelectrode,refillable,epoxybody8302BNUMD OrionROSSTriodepH/ATCelectrode,refillable,glassbody8172BNX OrionROSSSure-FlowpHelectrode,glassbody8611BNWP OrionROSSsodiumcombinationISE927007MD OrionATCprobe,stainlesssteelbody927005MD OrionATCprobe,epoxybody013005MD OrionDuraProbeconductivitycell,4-cell,K=0.475013016MD Orionconductivitycell,flowthrough,K=0.1083005MD Oriondissolvedoxygenprobe,polarographic086030MD Orionauto-stirDO/BODprobe,polarographic087010MD OrionRDOopticalDOprobe,3mcable810199 OrionROSSall-in-onepHbufferkit910104 OrionpH4.01buffer,475mL910107 OrionpH7.00buffer,475mL910110 OrionpH4.01buffer,475mL011008 Orionconductivitystandard,100µS/cm,5x60mLbottles011007 Orionconductivitystandard,1413µS/cm,5x60mLbottles011006 Orionconductivitystandard,12.9mS/cm,5x60mLbottles1010001 Orionconductivitycalibrationresistorkit9512HPBNWP Orionhigh-performanceammoniaISE9512BNWP Orion standard ammonia ISE951007 Ammoniastandard,1000ppm,475mL951210 Low-levelammoniaISA951211 AmmoniaISA,475mL951209 FillsolutionforOrionhigh-performanceammoniaISE,60mL951202 FillsolutionforstandardammoniaISE,60mL951213 AmmoniaISEstoragesolution9609BNWP OrionfluoridecombinationISE040906 Fluoridestandard,1ppmwithTISABII,475mL040907 Fluoridestandard,2ppmwithTISABII,475mL040908 Fluoridestandard,10ppmwithTISABII,475mL940909 TISABII,1gallon8611BNWP OrionROSSsodiumcombinationISE

  • 33Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Appendix Advanced Features

    Automatic pH Buffer Recognition Feature

    TheOrionVERSASTARMeterwithpH/ISE/TemperaturemodulesystemsarecapableofautomaticallyrecognizingpH1.68,4.01,7.00,10.01and12.46buffersorpH1.68,4.01,6.86,and9.18buffersduringapHcalibration,dependingonthepHbuffersetthatisselectedinthesetupmenu.Duringacalibration,themeterusestheselectedbuffersetandtherawmVreadingofthepHelectrodeinthebuffertorecognizeanddisplaythebuffervalueatthemeasuredtemperature.TherawmVreadingofthepHelectrodeinthebuffermustbeabout±30mVfromthetheoreticalmVreadingofthebufferinorderforthemetertoautomaticallyrecognizethebuffer.

    USA pH Buffer Set DIN pH Buffer SetBuffer mV Range Buffer mV Range

    1.68 +285to+345mV 1.68 +285to+345mV4.01 +207to+147mV 4.01 +207to+147mV7.00 -30to+30mV 6.86 +38to-22mV10.01 -207to-147mV 9.18 -99to-159mV12.46 -293to-353mV

    Testing a pH Electrode for Automatic Buffer Recognition

    ToverifythattherawmVreadingofthepHelectrodeinuseis±30mVfromthetheoreticalmVreadingofthebuffer,andthereforeverifythatthepHelectrodeinuseiscapableofperformingautomaticbufferrecognition,performthefollowingprocedure.

    1. SetthemeasurementmodetomV.PreparethepHelectrodeaccordingtotheinstructionsinthepHelectrodeuserguide.

    2. RinsethepHelectrodewithdistilledwater,blotitdrywithalint-freetissueandplacethepHelectrodeintoapH4.01bufferatapproximately25°C.

    3. WaitforthemeasurementtostabilizeandrecordthemVvalueofthepH4.01bufferwhenthemeterindicatesthatthemeasurementisstable.ThereadtypeselectedinthesetupmenuwilldeterminehowthemVmeasurementsaredisplayedbythemeter.

    4. RemovethepHelectrodefromthepH4.01buffer.

    5. RinsethepHelectrodewithdistilledwater,blotitdrywithalint-freetissueandplacethepHelectrodeintoapH7.00bufferatapproximately25°C.

    6. WaitforthemeasurementtostabilizeandrecordthemVvalueofthepH7.00bufferwhenthemeterindicatesthatthemeasurementisstable.

    7. ThemVreadingofthepHelectrodeinpH4.01buffershouldbeintherangeof+207to+147mVandthemVreadingofthepHelectrodeinpH7.00buffershouldbeintherangeof-30to+30mV.IfthemVreadingsofthepHelectrodeareinthecorrectranges,thepHelectrodeiscapableofperformingautomaticbufferrecognition.IfthemVreadingsofthepHelectrodearenotinthecorrectranges,thevaluesofthepHbufferswillneedtobemanuallyenteredduringapHcalibration.

  • 34 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    pH Electrode Condition Icon

    TheelectrodeconditioniconindicatestheperformanceofthepHelectrode,basedonthelastsavedcalibration andelectrodemeasurementstability.

    Icon pH Electrode Definition of Icon

    Electrodeconditionisgoodandtheelectrodeslopeis90to110%.

    Electrodeconditionisfairandtheelectrodeslopeis80to120%.

    Electrodeconditionisbadandtheelectrodeslopeislessthan80%orgreater than120%.

    Consulttheelectrodeuserguideforinstructionsonhowtoclean,conditionandtroubleshootthe electrode.

  • 35Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Multipoint Calibration Curve Feature

    TheOrionVERSASTARMeterwithpH/ISE/Temperaturemodulesystemsutilizeapoint-to-pointschemeformultipointcalibrationsinthepHandISEmodes.Inthistechnique,straightlinesegmentsaredrawnbetweeneachsuccessivepairofcalibrationpoints.SeparateslopeandE0valuesarecalculatedusingtheadjacentpairofdatapoints.Forexample,inafivepointcalibration,fourslopesandfiveE0valuesarecalculatedtocorrespondtothefourlinesegmentsthatconnectthedata.pHorconcentrationvaluesforsamplesarecalculatedaccordingtothesegmentthattheirpotentialvaluesfallinto.Thoseaboveorbelowtherangeofcalibrationarecalculatedbyextrapolationfromthetoporbottommostsegment.

    Thedisplayedslopeisanaverageslopeoftheseparatesegments.Theslopemaybeoutsidetherangenormallyexpectedforanelectrode.Forexample,iftheslopeofonesegmentislow,thentheaveragewillbelowerthantheoretical,eventhoughtheresponsemaybeNernstianovermostoftherange.

    Thefigurebelowisanexampleofatypicalcalibrationcurveusingthismethod.

  • 36 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Automatic Blank (Non-linear) Correction Feature

    TheOrionVERSASTARMeterwithpH/ISE/Temperaturemodulesystemshaveafeatureknownasautomaticblankcorrection.Inthistechnique,themeterdecideswhetherblankcorrectionisthebestmeasurementstrategybyanalyzingtheelectroderesponseduringamultipointcalibration.Aseparateblankdoesnothavetoberun.Graphically,automaticblankcorrectionisequivalenttodrawingasmoothcurvethroughthelowestthreepointsofthemultipointcalibrationandextrapolatingtozeroconcentrationbasedontheassumptionofNernstianelectrodebehavior.

    Multipointcalibrationsatthelowerlimitofdetectionaredesirablewhentheresponseofanionselectiveelectrodeisnon-linearandcannotbecharacterizedwithaoneortwopointcalibration.Thisisusuallyseenasalowelectrodeslope.Generally,theelectrodeisbehavinginaNernstianmannerbuttheeffectofablankisbeingobserved.Seethefigurebelow.

    Thisblankmaybeatruereagentblank,tracesofanalyteioninthereagentsoritmaybethe“mud”valueoftheelectrode.Itcouldalsobeaninterferenceinthereagentsthatbecomesapparentatlowlevelsofanalyteion,oritcouldbeanycombinationoftheseeffects.TheexpandedversionoftheNernstequationtraditionallyusedforblankcorrectionis as follows:

    E = Eo + S * log (C + b) wherebistheblank

    Inamulti-pointcalibration,asetofequationsisgeneratedandtherelationshipbetweenthemevaluated.Forexampletheequationsgeneratedinathreepointcalibrationwouldbeasfollows:

    E1 = Eo + S * log (C1 + b)

    E2 = Eo + S x log (C2 + b)

    E3 = Eo + S x log (C3 + b)

    ThemeterevaluatestherelationshipsbetweenthethreepotentialsE1,E2andE3andthethreeconcentrationsC1,C2 andC3.Iftherelationshipsindicatethatblankcorrectionisdesired,ablankwillautomaticallybecalculatedandthe non-linearitywillbecorrectedforinaNernstianmanner.Iftheappropriateconditionsarenotmet,theblankissetatzeroandeachsegmentofthemultipointcalibrationistreatedindependently.

  • 37Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Whenallthreeofthefollowingconditionsaremet,blankcorrectionisinvoked.

    1. Theconcentrationofthefirststandardiszero,ortheslopeoftheelectrodebetweenthefirstandsecondstandardsislessthantheslopebetweenthesecondandthird.

    2. Potentialdifferencesbetweenpointsaresignificant.Forexample,E3-E1>10mV

    3. Theblankcorrectionalgorithmconvergesatreasonableblankandslopevalues.Conditionsinsteps1and2preventfailuretoconvergeinmostsituations.However,theslopewillattainanyvaluenecessaryforconvergencetoacalculatedblankvalueof3xC3.

    Whenblankcorrectionisimplemented,theslopevaluefortheelectrodemaybeoutsidetherangeofvaluesnormallyconsideredacceptableduringanordinarycalibration.Forbestresults,calibrationstandardsshouldbecloseinrangetotheexpectedsampleconcentrationsandshouldbrackettheexpectedsampleconcentration.Whenconditions1,2and3arenotmet,thecalibrationdataishandledbythemultipointcalibrationmethoddescribedearlier.

    Incalibrationswithmorethanthreepoints,acombinationofmethodsisutilized.Automaticblankcorrectionisusedifthelowestthreepointssatisfythecriteriaandmultipointcalibrationisusedfortheotherpoints.Theslopedisplayedaftercalibrationisanaverageoftheslopecalculatedintheautomaticblankcorrectionalgorithmandtheslope(s)foreachadditionalsegment.

    Determining an Isopotential Value

    1. Prepareseveralstandardswithconcentrationrangesoverthemeasuringrangeoftheelectrode.

    2. Measurethemillivoltvalueofthestandardsatroomtemperature,about20°Cto25°C.

    3. Measurethemillivoltvalueofeachstandardat75°C.

    4. Measurethemillivoltvalueofeachstandardat10°C.

    5. Onsemi-logarithmicgraphpaper,plottheconcentrationvaluesonthelogaxisversusthemillivoltvaluesonthelinearaxis.

    6. Thelineswillintersectattheisopotentialvalue.Readtheconcentrationfromthegraphforthispointfromthelogaxis.Thisistheisopotentialvalue.

  • 38 Thermo Scientific Orion VERSA STAR pH/ISE/Temperature Module

    Notes

  • This page intentionally left blank.

  • 9 0 0 1 : 2 0 0 8

    Water Analysis Instruments

    North America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 [email protected]

    NetherlandsTel: (31) 033-2463887 [email protected]

    India Tel: (91) [email protected]

    Japan Tel: (81) [email protected]

    ChinaTel: (86) 21-68654588 [email protected]

    Singapore Tel: (65) [email protected]

    Australia Tel: (613) 9757-4300in Australia (1300) [email protected]/water

    © 2011 Thermo Fisher Scientific Inc. All rights reserved.

    68X592004 Rev. A 12-11