Operatin
g M
an
ual
Operatin
g M
an
ual Thermo bath & cool block
DBD-H SERIES
Operating Manual Operating Manual
1. PRODUCTS DESCRIPTION2. USER CONSIDERATION3. USER OPERATION4. ACCESSORIES5. MAINTENANCE AND CLEANING & FAULTY RUNNING6. WARRANTY
CAUTION-Before attempting to operate the unit,
read this instruction manual carefully
CAUTION-HIGH TEMPERATURE
"Don't touch the block while it is in operation and
until the temperature of the block goes down after
turning it off"
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529
MRC. 9.18
Difference of bath
P.I.D temperature controller
- Easy to control and check setting temp.
MRC Blocks are much thicker than others
- Prevent current temp. from sudden down when a tube is put into a tube hole. Do not need put water or mineral oil into the gab between
hole and tube. The shape of hole is same with tubes.
Temperature Calibration
- Able to calibrate the different temperature between setting and current temp. Safe device.
A. Electrical fuse
B. If current temp, does not reach the setting temp, for 99hour 59min
ALM Lamp on temperature controller would be blink.
C. This books are used in up to 150˚(DBD-H1,DBD-H2,DBD-H4),
95˚ (DBD-C1). If the current temp. is over 150.5˚(DBD-H1,
DBD-H2,DBD-H4), 95.5˚(DBD-C1). ALM lamp would be blink
D. If the current temp, keep over 150˚C continually, the heater would be turned out automatically.
1
PR
OD
UC
T D
ES
CR
IPT
ION
Introduction
HEAT BLOCK
- One block- Ambient +5 ~150˚C- Dimension - external : W205XD263XH163
DBD-H1
DBD-6400- One block- 0 ~120˚C- Dimension - external : W215XD327XH215
COOL BLOCK
- Two block- Ambient +5 ~150˚C- Dimension - external : W205XD360XH163
DBD-H2
- Four block- Ambient +5 ~150˚C- Dimension - external : W335XD360XH163
DBD-H4
Description
AC100 - 240V(*220 - 230V : CE cert)
50/60Hz
Ambient +5˚C ~ 150˚C
Model
Temp. Range
Temp. Accuracy
Temp. Timer
Wattage
Power
DBD-H1 DBD-H2
±0.2˚C
99hour 59min
Heating method
Cooling method
DBD-H4
310W 610W 1200W
Frequency
BACK SIDE
MAIN BODY
TEMPERATURE CONTROLLER
STOP
RUN ●
TIME ●
HEAT ●
ALM ●
PV
SV
˚C
RUNMODE
STOP
Current temperature
Setting temperature
MODE key
UP keyDOWN keyMOVE key
RUN/STOP key
RUN Lamp(orange)TIME Lamp(green)
HEAT Lamp(red)ALARM Lamp(red)
Specification
PR
OD
UC
T D
ES
CR
IPT
ION
2
Dimensions(mm)
Weight
Capacity 1 Block2 Block1 Block
11.5kg (without block)6kg (without block)3kg (without block)
W205 x D360 x H163 W335 x D360 x H163W205 x D260 x H163
MaterialEnergy plate an sample block : Aluminuminside and insulating side : Stainless steel
Cartridge heater
- Peltier
0˚C ~ 120˚C
Receptacle Specification
Voltage / Hz
Serial Number
RUN ●TIME ●
HEAT ●ALM ●
PV
SV
˚C
RUNSTOP MODE
Power switch
Inner chamber
Ventilation hole
Bottom foot
Temperaturecontroller
DBD-C1
4 Block
10kg (without block)
W215 x D327 x H215
360W
V011
110V
220V
Place the unit on a stable, flat and cleansurface and check if the four rubber legs are touched firmly on the surface.
Check if the voltage indicated on appliance corresponds to the local mains voltage before you connect the appliance.
Unplug the appliance : - If problems occur during the heating and rotating process.- Before you start cleaning the unit.- Never use the appliance if the plug, the cord or the appliance it self is damaged
Do not place this instruments beside sunnyplace or heater. If the main body is higherthan room temperature, It could affects theinside setting temperature.
Be careful not to drop any chemical liquid or water into the hole for blocks, because it could be reason of causing unstable setting temperature.
User consideration
US
ER
CO
NS
IDE
RA
TIO
N
Environmental conditions- Indoor use.- Environment temperature 5˚C to 40˚C.- Maximum relative humidity 80% for temperatures up to 31˚C decreasing linearly to 50% relative humidity at 40˚C.- Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage.- Other supply voltage fluctuations as stated by the manufacturer. - Transient overvoltages according to installation categories.- Installation category : CATⅡ - Pollution degree 2- IPXO (ordinary)
3
Power switch is in the "on" position and operate it after 3 seconds for stable using of product.
RUN ●TIME ●
HEAT ●ALM ●
PV
SV
˚C
RUNSTOP MODE
RUN ●TIME ●
HEAT ●ALM ●
PV
SV
˚C
RUNSTOP MODE
During use this instrument, Please do not touch the block. Hand could be burnt.
RUN ●TIME ●
HEAT ●ALM ●
PV
SV
˚C
RUNMODE
Do not need put water or mineral oil into the gab between holes and tubes. Because the shape ofhole and tube is same.* Never use PEG withthis instrument, PEG could peel off the paint of main body.
RUN ●TIME ●
HEAT ●ALM ●
PV
SV
˚C
RUNSTOP MODE
For stable temperature, the heating plate and the bottom of block should be clean before using the instrument.
Heating plate
RUN ●TIME ●
HEAT ●ALM ●
PV
SV
˚C
RUNMODE
For precise test, Do not fill upreagent in the tube, keep theline inside the hole.
When you move this instruments, Please be careful not to drop.Because it is designed by very sensitive parts. Some impact couldbe cause fatal defects.
RUN ●
TIME ●
HEAT ●
ALM ●
PV
SV
˚C
RUNSTOP
MODE
43
21
2
V022
1
This section summarizes operation procedures for MRC`s thermo bath controlling the temperature and the rotate.
Check the name rating plate for the correct voltage, then plug the power cord into thereceptacle on back of the instrument.
Plug the power cordinto the wall receptacle.
User operation
US
ER
OP
ER
AT
ION
* Temp. setting
Turn the power ON/OFF control button to position'ON'
When switch turn on, (the type of sensor)is displayed on PV and (software vision)isdisplayed on SV for1second.
After 1second,the current insidetemperature isdisplayed on PV andthe setting temperatureis displayed on SV.(This is NORMAL MODE)
In order to set thedesired TEMPERATURE,press MODE key for1.5seconds thenis displayed on PV and the rightmost number blinkon SV then stop pressingMODE key. Unless stop pressing MODE key, the setting return to NORMALMODE
p
When press MOVE Key, the rightmost number blink to moves in left. temperature range can be set per 0.1˚C. for example to set 45˚C, press MOVE key till the rightmost number blink to moves in the second placeof right.
※ Never touch an electrical socket or the power cord with wet hands.
4
3 4Put a block to the hole
Before place a block, check the bottom of block and insideheating plate clean.
RUN ●TIME ●
HEAT ●ALM ● PV
SV
˚C
RUNSTOP MODE
RUN ●TIME ●
HEAT ●ALM ● PV
SV
˚C
RUNSTOP MODE
RUN ●TIME ●
HEAT ●ALM ● PV
SV
˚C
RUNMODE
3
2
5
US
ER
OP
ER
AT
ION
STOP
RUN ●
TIME ●
HEAT ●
ALM ●
PV
SV
˚C
RUNMODE
STOP
pWhen press UP key,the displayed figureincreases one by onethen press UP keycontinuously till thedisplayed figure reachesat 5.
STOP
RUN ●
TIME ●
HEAT ●
ALM ●
PV
SV
˚C
RUNMODE
STOP
The figure shouldbe displayed at 45.0 to use MOVE,UP and DOWN key, press MODE keyto complete the set.
* Timer & BEEP setting
STOP
RUN ●
TIME ●
HEAT ●
ALM ●
PV
SV
˚C
RUNMODE
STOP
0000
Press MODE key onceagain after the settingof the TEMPERATURE then is displayed on PVand is displayed on SV. TIME range canbe set to 99 Hours59minutes. Two positionsof the front are hour and next two positionsare minute. The method of TIMER setting is thesame as the method ofTEMPERATURE setting.
I e I e
STOP
RUN ●
TIME ●
HEAT ●
ALM ●
PV
SV
˚C
RUNMODE
STOP
0000Eep
PPress MODE key onceagain after TIME settingthen is displayed on PV and 0000 is displayed on SV. sound time can be set to9999times.The method of BEEPsound time setting is the same as the method ofTEMPERATURE setting and TIMER setting.
EepP
EepP
STOP
RUN ●
TIME ●
HEAT ●
ALM ●
PV
SV
˚C
RUNMODE
STOP
Press MODE key onceagain after the settingof BEEP sound timethen the screen returnto the NORMAL MODE.
5
User operation
6
1
5
43
21
US
ER
OP
ER
AT
ION
STOP
RUN ●
TIME ●
HEAT ●
ALM ●
PV
SV
˚C
RUNMODE
STOP
After finishing the setting,press RUN key for 1.5seconds. Then RUN lamp is on and temperature rises in beginning. If the TIMER has been set, TIME lamp will be on till to reach the setting TEMPERATURE. TIMERwill start when the temperature has reached the setting TEMPERATURE.※ Unless the TIMER is sets, TIME lamp is not on.
* RUN & Check
STOP
RUN ●
TIME ●
HEAT ●
ALM ●
PV
SV
˚C
RUNMODE
STOP
HEAT lamp let you know that heater normallywhile the unit is running.
STOP
RUN ●
TIME ●
HEAT ●
ALM ●
PV
SV
˚C
RUNMODE
STOP
If the TIMER is set andyou want to see theset value while the unitis running, press UP key once then SV will be changedsetting TEMPERATUREto TIMER.
STOP
RUN ●
TIME ●
HEAT ●
ALM ●
PV
SV
˚C
RUNMODE
STOP
45.0
After the setting TIMERhas passed, is displayed on SV andthe setting BEEP soundbeeps for the settingBEEP sound time.Then press stop key BEEPsound will be stoppedand will be return toNORMAL MODE.
STOP
RUN ●
TIME ●
HEAT ●
ALM ●
PV
SV
˚C
RUNMODE
STOP
The existing setting TEMPERATURE, TIMER and BEEP sound time are kept in memory. Therefore,Pressing the RUN key 1.5 seconds before the unitwill operate the same way as.If you want to change thesetting TEMPERATURE, TIMER and BEEP sound time,please start to reset all of thesetting value.
6
User operation
Accessories
AC
CE
SS
OR
IES
7
NON-STANDARDS CAN BE MADE BY ORDER
- DIMENSION (mm) : W129 X D100 X H55
0.5ml centrifuge tube : 63holesAB100-0563
1.5ml centrifuge tube : 48holesAB100-1548
96PCR plate : 1ea or8strip tube : 12ea
AB100-961.5ml centrifuge tube : 24holesAB100-1524 (with Rack)
50ml conical tube : 8holesAB100-508
15ml conical tube : 16holesAB100-1516
TM
JD-90
Cover
CVH4-80 CVH2-80 CVH1-80 CVC1-80
CVH4-F CVH2-F CVH1-F CVC1-F
- DIMENSION (mm) : W129 X D100 X H78
Faulty running
Always unplug the instrument before you start cleaning it.
Never immerse the appliance in water.
This equipment is maintenance-freeClean the surfaces of the heated bath at regular interval- water containing a surfactant detergent additive.- Isopropyl alcohol
.
Never use the instrument if the plug, the cord or the instrument itself is damaged.
Maintenance and Cleaning
MA
INT
EN
AN
CE
AN
D C
LE
AN
ING
& F
AU
LT
Y R
UN
NIN
G
1. The thermometer and setting temperature is different - Thermometer may not correspond each certification authorities.
2. When you turn on the power switch, the lights don’t work. - You can find the fuse box for spare one in the power module as the below drawings.
If the equipment have other defects, It should be informed to an authorized servicing agent.
Pull the fuse case using a small driver or other tool Push the fuse case
Find the fuse case on the power socket Take the new fuseRemove used fuse
Spare fuse
8
Warranty
1 YEAR WARRANTY
Limited Warranty :
This equipment has been tested all the functions and parts for proper operationbefore shipment.Please read this manual carefully. If youhave any questions or problems, checkwith your dealer or write us direct.This equipment is warranted for a period of 1year from date of purchase againstdefects in workmanship and material only.If a system component should ever fail to work perfectly. It can be repaired or replaced by our after-sales service withoutdifficulty.
Model#:
Serial# :
Date Purchased :
Dealer Name :
City and Country :
WA
RR
AN
TY