Upload
gang
View
217
Download
0
Embed Size (px)
Citation preview
GENERAL GYNECOLOGY
The study of laparoscopic electrosurgical instruments on thermaleffect of uterine tissues
Qian Zhu • Jiaying Ruan • Li Zhang •
Wei Jiang • Hongqian Liu • Gang Shi
Received: 18 May 2011 / Accepted: 30 December 2011 / Published online: 11 January 2012
� Springer-Verlag 2012
Abstract
Purpose It is to compare the thermal damage on myo-
metrium tissue caused by five electrosurgical instruments,
including monopolar forceps, bipolar forceps, PK scalpel,
Ligasure and BiClamp.
Methods Normal myometrium in vitro was collected and
electric coagulation was conducted with five electrosurgi-
cal instruments under corresponding powers. The zones of
thermal injury (ZTI) in each coagulation sites were
examined histologically, while the width and depth of
thermal damage were measured.
Results 1. There were significant differences among vari-
ous groups’ widths of ZTI of myometrium (P \ 0.05). Lig-
asure produced the greatest width of ZTI, and it was
statistically greater than that of PK scalpel and BiClamp
(P \ 0.05). While the widths of ZTI caused by monopolar
and bipolar electrocoagulation lied between that of Ligasure
and PK scalpel, but the differences were of no statistical
significance (P [ 0.05). 2. The depths of ZTI in different
groups were of significant differences (P \ 0.05). Both
monopolar and bipolar forceps had greater depths of ZTI
compared with BiClamp (P \ 0.05) but not had statistical
differences with Ligasure and PK scalpel (P [ 0.05).
Conclusions As for myometrium, the thermal damage is
rather small in the horizontal and vertical directions when
using BiClamp and PK scalpel. Ligasure places larger
range of thermal damage in horizontal direction with little
depth in vertical direction, which is rather safe when acting
on uterine surface. Electrocoagulation was conducted with
monopolar (the power is 55 W) and bipolar forceps (the
power is 40 W) continuously for 3 s, whose thermal
damage range is fairly safe to corpus uteri wall and fundus
uteri.
Keywords Laparoscope � Electrosurgery �Thermal injury
Introduction
On the basis of minimal invasion and high efficiency, lap-
aroscopy has been extensively applied in gynecologic sur-
geries, especially prevailing in the surgery of uterine
fibroids, adenomyoma and other benign lesions, which
obviously benefits from reduction of intraoperative blood
loss, shortened operation time and increased operation
volume. However, the surgery-related complications caused
by various electrosurgical instruments are seldom, involv-
ing intraoperative uterine perforation, postoperative met-
roperitoneal fistula, hysterorrhexis during pregnancy [1–3],
and even some cases of hysterorrhexis during pregnancy
after the enucleation of subserous myoma of uterus [4].
Apart from specified surgical skills, the irrational use of
electrosurgical instruments incurring acute thermal damage
and secondary inflammatory reaction usually risk severe
surgical complications.
Thus, how to identify the optimal one from a large
number of electrosurgical instruments, particularly newly
Q. Zhu � J. Ruan � H. Liu � G. Shi (&)
Department of Obstetrics and Gynecology, West China
Second University Hospital, Sichuan University,
Chengdu 610041, China
e-mail: [email protected]
L. Zhang
The First People’s Hospital of Chengdu, Chengdu,
Sichuan, China
W. Jiang
Department of Pathology, West China Second University
Hospital, Sichuan University, Chengdu 610041, China
123
Arch Gynecol Obstet (2012) 285:1637–1641
DOI 10.1007/s00404-011-2207-0
merged surgical instruments, is strongly required, when
coping with different surgical conditions. However, there are
no reports available to rely on in comparing thermal effects
and damages of uterine tissues caused by different electro-
surgical instruments. This presented study is designed to
compare the thermal damages caused by five common
electrosurgical instruments which act on myometrium tis-
sues, to identify the degree of thermal damage, as well as
providing evidence for the reasonable, high-efficient and
safe clinical use of various electrosurgical instruments.
Materials and methods
Surgical instruments
Following were the surgical instruments used: monopolar
and bipolar curved forceps, Stryker, USA; Gyrus Medical’s
PlasmakineticTM Tissue Management System, Gyrus
Medical’s, GER; closing forceps with 10 mm diameter of
LigasureTM Vessel Sealing System, Valleylab, USA;
VIO300D electrosurgery station BiClamp forceps, ERBE,
GER.
Specimens
Fifteen cases of women in child-bearing period, who had
been conducted hysterectomy because of hysteromyoma,
were brought into, whose normal muscle tissues of uterine
anterior and posterior wall were taken; women who were in
menopause, with complication of adenomyosis and uterine
cancer were excluded, nor the parts which are meaningful
for pathological diagnosis. The study proposal was
approved by the ethical committee of hospital. The study
was explained to all enrolled subjects, and a written
informed consent was obtained from each participant.
Experimental methods
Myometrium tissues were cut into 2 cm 9 2 cm 9 1 cm
tissue blocks, which were randomly distributed to be
appropriately experimentally treated. The distances
between forceps of instruments’ acting on tissues were all
set to be 5 mm, being measured by vernier caliper and
acting sites were marked by prepared Chinese ink. Based
on both the recommended power from the manufacturer’s
specifications and practical clinical application, monopolar
forceps (the electric coagulation power was 55 W) and
bipolar forceps (the electric coagulation power was 40 W)
were set to continuously coagulate for 3 s. The acting
time of PK scalpel (the power was automatically regulated
and controlled), Ligasure (conventional output parameters
of host was regulated to 2–3) and BiClamp forceps (effect
parameter was set to be 1) were regulated and controlled
by intelligent instruments.
The experiments were conducted by the same conduc-
tor, keeping constant contact pressure. After processing,
tissues were maintained at an extended state and were
placed in 10% formalin for fixed conservation. The time
interval from the isolation of all tissues to fixed conser-
vation was controlled to be in 2 h.
After the above processes, tissues were cut off passing
action spots and the section was perpendicular to wound
plane. Then the section was downward and paraffin-
embedded, sliced by 5 lm on section. Conventional HE
staining was carried out.
Outcome measures
Blind trial was adopted to observe pathological sections,
which was conducted by the same pathologist. Optical
microscope with standard micrometer was used to observe
the histopathological changes and measure the scope of
irreversible thermal damage, while the measurement was
accurate to 0.01 mm. Measurement indexes included: (1)
the horizontal width of the zones of thermal injury (ZTI) of
myometrium: the maximum horizontal diameter of damage
zone as the action spot is the center; (2) the vertical depth
of ZTI of myometrium: the maximum radial line which is
vertical to the surface of damage zone.
Statistical methods
Statistical analysis was performed using the SPSS� version
16.0 (SPSS, Chicago, IL). The measurement data are all
showed by �x� s. General Linear Model multivariate
analysis of variance is adopted to compare two sample
means of multi groups and further pairwise comparisons
are inspected by LSD. P = 0.05 (two-sided) was consid-
ered as the limit of significance.
Results
Histopathological features of thermal damage
in myometrium tissues
The morphological changes of thermal damage caused by
different instruments are all acute heat coagulable
necrosis. Under the microscope, membrane of smooth
muscle cells in ZTI disappeared, and endochylema was
homogeneous and concentrated red dyed. Crenas were
scattered in ZTI; nuclear chromatin was darkly stained,
and many nuclei became long flocculent or pyknosis.
In serious denaturation areas, cytoarchitecture could be
seen disappeared, and the whole showed homogeneous
1638 Arch Gynecol Obstet (2012) 285:1637–1641
123
and flake-like, while karyorrhexis and karyolysis were
visible.
The measured parameters of thermal damage in myo-
metrium tissues show that the maximum mean width was
produced by Ligasure, while the maximum mean depth was
caused by monopolar electrocoagulation. The measurement
results of different groups’ thermal damage scopes of
myometrium tissues are shown in Table 1.
Comparison of ZTI width
The comparison of ZTI width of different groups shows an
obvious difference (F = 141.902, P = 0.000), in which
Ligasure group’s thermal damage width is bigger than that
of PK scalpel group, BiClamp group (P \ 0.05), whilst
monopolar and bipolar electrocoagulation group are com-
parable (P [ 0.05) (Fig. 1).
Comparison of ZTI depth
There are differences among the comparison of damage
zones’ depth of myometrium (F = 34.586, P = 0.000),
among which, the depth of ZTI of monopolar electroco-
agulation group and bipolar electrocoagulation group are
greater than that of BiClamp group, and the difference is of
statistical significance (P \ 0.05). Other pairwise com-
parison differences are of no statistical significance
(P [ 0.05) (Fig. 2).
Discussion
Monopolar and bipolar forceps are frequently used
electrosurgical instruments in laparoscopic surgery. Sur-
geons always judge the degree of tissues’ thermal dam-
age according to its blanching, crenation, stiffening and
other manifestations, but cannot accurately estimate tis-
sues’ in-depth thermal damage range. The power and
process time of monopolar and bipolar forceps vary in
different literatures. According to the manufacturer’s
instructions, the recommended electric coagulation power
of monopolar and bipolar forceps was 30–70 W in
general surgery. Furthermore, it is frequently observed in
clinical practice and our previous study that excellent
hemostatic effect was always obtained through mono-
polar (55 W) and bipolar (40 W) forceps’ continuous 3-s
electrocoagulation. On this basis, we found that mono-
polar and bipolar forceps’ thermal damage depths in
vertical direction are remarkably more than the other
three instruments. When it has to be used, the surgeon
needs to be alert so as to avoid complications because
the thermal damage caused by the tip of forceps in
vertical direction may affect neighboring important tis-
sues. While in horizontal direction, the width of thermal
damage caused by monopolar and bipolar forceps is
smaller than that of Ligasure, but the difference is
without statistical significance. It is analyzed to be
connected with the Ligasure’s longer acting time and
Table 1 The width and depth
of ZTI of myometrium (�x� s)Groups Cases of tissues (n) Width (mm) Depth (mm)
Monopolar electrocoagulation group (ME) 9 4.57 ± 0.65 2.79 ± 1.11
Bipolar electrocoagulation group (BE) 10 4.29 ± 1.02 2.67 ± 1.26
PK scalpel group (PK) 10 3.90 ± 1.05 1.77 ± 0.95
Ligasure group (LS) 8 5.08 ± 1.28 2.16 ± 1.37
BiClamp forceps group (BC) 7 3.82 ± 1.39 1.49 ± 0.95
4.574.29
3.9
5.08
3.82
0
1
2
3
4
5
6
ME BE PK LS BC
wid
th m
eans
of Z
TI(
mm
)
ME
BE
PK
LS
BC
Fig. 1 The histogram of width means of ZTI of myometrium
2.792.67
1.77
2.16
1.49
0
0.5
1
1.5
2
2.5
3
ME BE PK LS BC
dept
h m
eans
of Z
TI(
mm
)
ME
BE
PK
LS
BC
Fig. 2 The histogram of depth means of ZTI of myometrium
Arch Gynecol Obstet (2012) 285:1637–1641 1639
123
greater power output. Compared with conventional
bipolar electrocoagulation, Ligasure’s major function can
automatically adjust the output of power according to the
clamped tissues’ thickness or density, and close the
blood vessels, ligaments and tissue bundles that are
smaller than 7 mm, whose zonula occludens can bear
three times of normal human systolic blood pressure. It
is proved by experiments that this permanently closed
zonula occludens fibrosis can withstand the tension up to
900 mmHg, as same as that of ligation and vessel clamp
which are remarkably higher than that of Harmonic
scalpel and conventional bipolar electrocoagulation
[5–7]. When the power of high-frequency electric scalpel
is set, it will not vary with the changes of tissue
impedance, and its durative action may increase tissue
temperature markedly. Carbonization occurs, and it is
inseparably adhered with electrodes, so that it is easy to
tear eschar down, resulting in the bleeding of the he-
mostatic sites again. Therefore, for monopolar and
bipolar forceps, to achieve the same hemostatic effect as
Ligasure, acting time should be lengthened or electro-
coagulation repeated. But researches have shown that
heat energy produced by high-frequency electric current
is in direct proportion to acting time. With the length-
ening of cumulative acting time, the extent of thermal
damage extends outward [8]. Hence, when using mono-
polar and bipolar forceps, the electric coagulation should
be strictly controlled. Based on the previous studies on
the thermal effect caused by monopolar and bipolar
electric scalpels’ electrocoagulation, it has been indicated
that the unexpected surrounding thermal necrosis of
bipolar electric coagulation, involving bladder, ureter,
rectum and many other tissues, is less than that of
monopolar electric coagulation [9, 10]. The finding in
this study is consistent with others. However, the extent
of thermal damage between monopolar and bipolar are
similar, which may be related with the short acting time
set in this research and rather great tissue resistance,
which generates smaller thermal damage.
By means of real time detection of tissues’ resistance or
the current changes of power supply, smart bipolar system
feeds back to control components, to regulate the output
power for achieving better effect. Currently, the working
principles of the clinically frequently used PK scalpel,
Ligasure and BiClamp are similar, which can automatically
regulate output power and acting time. This research
indicates that Ligasure’s thermal damage width and depth
are greater than that of PK scalpel and BiClamp. After
analyzing the reasons, it owes to the various output powers
of different instruments, so that their effects are varied.
Tansatit et al. [11] found that the average burst pressure of
closing carotid of fresh body by Ligasure is higher than that
of BiClamp, and before the acting sites are completely
dried, BiClamp has automatically stopped energy output,
which indicates that its solidification effect of blood vessel
is worse than Ligasure. Oussoultzoglou et al. conducted
perspective study on the severity of hypocalcemia after
total thyroidectomy, then they found that the postoperative
severity of hypocalcemia of the patients who were operated
with BiClamp was lighter than who were with Ligasure
[12]. It indirectly suggests that BiClamp has less output
energy and smaller thermal damage range, which is con-
sistent with this research results. In the animal experiments
carried out by Carbonell et al., the burst pressure of Lig-
asure’s closing 2–3 mm diameter vessels was indifferent
from that of PK scalpel, but obviously higher when refer-
ring to 4–7 mm vessels. It implies that the hemostatic
effect of PK scalpel is weaker than Ligasure for the vessels
whose diameters are [3 mm [13]. In this research, the
thermal damage ranges of PK scalpel and BiClamp in
horizontal and vertical directions are rather small and
advantageous in safety, but the differences of its hemostatic
effect with Ligasure require further study.
Uterus is a muscular organ with cavity and thick wall,
whose different parts own varied thickness. When judging
the damage degree of myometrium in laparoscopic surgery,
it should be considered by combining uterine morphology.
Some scholars put forward that the depth of thermal
necrosis is inadvisable to exceed 20% of myometrium
thickness in uterine electrosurgery [14]. Duffy et al. [15]
measured the thickness of uterine wall during in vitro
experiments, acquiring that the average thickness of fundus
uteri was 1.4 cm, corpus uteri antetheca 1.8 cm, paries
posterior 1.9 cm, isthmus 1.3 cm, while the thinnest part
was just 0.7 cm, and the average thickness of cornua uteri
which was 0.5 cm from the entrance of oviduct was
0.6 cm. With the results of this study, we deduce that
monopolar forceps whose power is set to be 55 W and
bipolar clamp whose power is set to be 40 W continuously
electric coagulate for 3 s, while PK scalpel, Ligasure and
BiClamp are used under recommended setting, of which
the thermal damage range is relatively safe to uterine wall
and fundus uteri but beyond security scope when applied in
isthmus and cornua uteri. Cobellis enrolled 15 women who
underwent uterine myomectomy previously, and observed
uterine scar in the cesarean section. It was found that most
of the scars’ thickness of abdominal incision cases was in
accordance with the surrounding tissues. The scars of those
who had been conducted laparoscopic surgery possess
great tension and uneven edges, being thinner than the
surrounding tissues [16]. The structure of scar tissues dif-
fers from myometrium, without effective extention during
pregnancy. There is the risk of hysterorrhexis. Therefore,
when operating on the uterus of those who desires for
fertility, the time of electrocoagulation should be
strictly controlled, avoiding excessive coagulation. It is
1640 Arch Gynecol Obstet (2012) 285:1637–1641
123
noteworthy that the widths of ZTI are greater than the
depths, so what should be prevented is repeated or con-
tinuous electrocoagulating on section of myomectomy,
precluding that thermal damage exceeds security scope or
even extends through the full thickness of uterine wall.
Conclusion
The current finding suggests the use of BiClamp and PK
scalpel in uterine involved laparoscopy, especially for
those fertility conserved surgery, due to its relatively small
thermal damage on uterine tissues both in horizontal and
vertical directions. Ligasure’s thermal damage affects more
widely in horizontal direction but with very small depth
vertically, which is comparatively safe for superficial per-
formance on uterus, e.g. removal of subserous myoma.
Monopolar (the power is 55 W) and bipolar forceps (the
power is 40 W) continuously electrocoagulate for 3 s, as
well as PK scalpel, Ligasure and BiClamp, are relatively
safe when coping with uterine wall and fundus. However,
they may be at risk when acting on uterine isthmus and
cornua.
In this study, with certain power setting, the various
thermal damages on myometrium led by five common
electrosurgical instruments were outlined through a histo-
logical way. In brief, our findings provide evidence for
gynecologists to choose appropriate instruments in various
uterine laparoscopy. Nevertheless, there is limitation in this
study because of ex vivo tissue lack of blood supply, which
may not accurately reflect the in vivo situation. Further
research is needed in the thermal damage caused by the
same instrument under different powers and different act-
ing time.
Conflict of interest None.
References
1. Dubuisson JB, Fauconnier A, Deffarges JV et al (2000) Preg-
nancy outcome and deliveries following laparoscopic myomec-
tomy. Hum Reprod 15(4):869–873
2. Hockstein S (2000) Spontaneous uterine rupture in the early third
trimester after laparoscopically assisted myomectomy. A case
report. J Reprod Med 45(2):139–141
3. Lieng M, Istre O, Langebrekke A (2004) Uterine rupture after
laparoscopic myomectomy. J Am Assoc Gynecol Laparosc 11(1):
92–93
4. Pelosi MA 3rd, Pelosi MA (1997) Spontaneous uterine rupture at
thirty-three weeks subsequent to previous superficial laparoscopic
myomectomy. Am J Obstet Gynecol 177(6):1547–1549
5. Landman J, Kerbl K, Rehman J et al (2003) Evaluation of a
vessel sealing system, bipolar electrosurgery, harmonic scalpel,
titanium clips, endoscopic gastrointestinal anastomosis vascular
staples and sutures for arterial and venous ligation in a porcine
model. J Urol 169(2):697–700
6. Harold KL, Pollinger H, Matthews BD et al (2003) Comparison
of ultrasonic energy, bipolar thermal energy, and vascular clips
for the hemostasis of small-, medium-, and large-sized arteries.
Surg Endosc 17(8):1228–1230
7. Heniford BT, Matthews BD, Sing RF et al (2001) Initial results
with an electrothermal bipolar vessel sealer. Surg Endosc 15(8):
799–801
8. Harrison JD, Morris DL (1991) Does bipolar electrocoagulation
time affect vessel weld strength? Gut 32(2):188–190
9. Tucker RD, Kramolowsky EV, Platz CE (1990) In vivo effect of
5 French bipolar and monopolar electrosurgical probes on the
porcine bladder. Urol Res 18(4):291–294
10. Tulikangas PK, Smith T, Falcone T et al (2001) Gross and his-
tologic characteristics of laparoscopic injuries with four different
energy sources. Fertil Steril 75(4):806–810
11. Tansatit T, Kulvitit Y, Jindatip D et al (2006) Comparison of
economic bipolar vessel sealer and biclamp for the hemostasis of
large-sized cadaver arteries. J Med Assoc Thai 89(3):169–173
12. Oussoultzoglou E, Panaro F, Rosso E et al (2008) Use of
BiClamp decreased the severity of hypocalcemia after total thy-
roidectomy compared with LigaSure: a prospective study. World
J Surg 32(9):1968–1973
13. Carbonell AM, Joels CS, Kercher KW et al (2003) A comparison
of laparoscopic bipolar vessel sealing devices in the hemostasis of
small-, medium-, and large-sized arteries. J Laparoendosc Adv
Surg Tech A 13(6):377–380
14. Onbargi LC, Hayden R, Valle RF et al (1993) Effects of power
and electrical current density variations in an in vitro endometrial
ablation model. Obstet Gynecol 82(6):912–918
15. Duffy S, Reid PC, Sharp F (1992) In vivo studies of uterine
electrosurgery. Br J Obstet Gynaecol 99(7):579–582
16. Cobellis L, Pecori E, Cobellis G (2004) Comparison of intramural
myomectomy scar after laparotomy or laparoscopy. Int J Gynecol
Obstet 84(1):87–88
Arch Gynecol Obstet (2012) 285:1637–1641 1641
123