The MPGD-based photon detectors for the COMPASS RICH-1 upgrade Stefano Levorato – INFN TRIESTE on behalf of COMPASS THGEM group: Alessandria, Aveiro, Budapest, Freiburg, Kolkata, Liberec, Prague, Torino, Trieste. Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 1

The MPGD-based photon detectors for the COMPASS RICH …thgem.ts.infn.it/PapersTalks/S_Levorato_IEEE2017-Atlanta... · Why MPGD-based photon detectors and a changein the ... 19 CsI

Embed Size (px)

Citation preview

The MPGD-based photon detectors for the COMPASS

RICH-1 upgradeStefano Levorato – INFN TRIESTE

on behalf of COMPASS THGEM group:Alessandria, Aveiro, Budapest, Freiburg, Kolkata, Liberec, Prague, Torino, Trieste.

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 1

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 2

Introduction: Outline

• Introduction: COMPASS RICH-1 @ CERN

• Why MPGD-based photon detectors and a change in the Cherenkov photo detection technology

• The architecture of the MPGD-based detector

• Construction, quality control and assembly

• Detector commissioning

• Performance hints

Outline

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 3

Introduction: COMPASS RICH-1 @ CERN

3

Central Cathodes n of ph. @ β=1

COMPASS Spectrometer dedicated to h physics

COMPASS RICH-1 upgraded in 2006with MAPMT in the most inner centralregion MAPMTs coupled to lenstelescopes 1.4 m2

MWPCs+CsI (from RD26): successful but performance limitations, in particular for the 4 central chambers

COmmon Muon Proton Apparatus Structure Spectroscopy Fixed target experiment at CERN, SPS50 mt. long spectrometer

The photo detector system of compass RICH-1 (upper part)

Number of photons for the very top and very bottom 60x60 cm2

MWPC:• On average lower than

the other PC 𝑵𝑵𝒑𝒑𝒑𝒑 =𝟏𝟏𝟏𝟏

• Slow decreasing trend 𝑵𝑵𝒑𝒑𝒑𝒑 vs year

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 4

Introduction: Outline

• Introduction: COMPASS RICH-1 @ CERN

• Why MPGD-based photon detectors and a change in the Cherenkov photo detection technology

• The architecture of the MPGD-based detector

• Construction, quality control and assembly

• Detector commissioning

• Performance hints

Outline

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 5

Cherenkov light detection, MWPC with CsI photocathode: the nearly worldwide used architecture and its limitation

CsI coating

MWPCs + CsIMWPCs with CsI photocathode, the limitations

Severe recovery time (~ 1 d) after adetector discharge Ion accumulation at the photocathode

Feedback pulses

Ion and photons feedback from themultiplication process

Ageing (QE reduction) after integrating afew mC / cm2

Ion bombardment of thephotocathode

Low gain: a few times 104(effectivegain: <1/2) “slow”detector

To overcome the limitations:

Less critical architecturesuppress the PHOTON & ION feedbackuse intrinsically faster detectors

MPGDs

RD26development

Reduced wire-cathode gap because of :

Fast RICH (fast ion collection) Reduced MIP signalReduced cluster sizeControl photon feedback spread

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 6

Introduction: Outline

• Introduction: COMPASS RICH-1 @ CERN

• Why MPGD-based photon detectors and a change in the Cherenkov photo detection technology

• The architecture of the MPGD-based detector

• Construction, quality control and assembly

• Detector commissioning

• Performance hints

Outline

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 7

The MPGD hybrid approach: THGEMs and MicroMegas

• One 300 mm x 600 mm Bulk Micromegas detector• Two layers of THGEMs (300 mm x 600 mm) in

staggered configuration

Signal read out via capacitive coupling pad readout and APV25 F/E boards

Two modules are put side by side to build a 600 mm x 600 mm detector

To simplify the construction requirements amodular architecture has been adoptedwhere one “module” consists of:

THGEM 1

THGEM 2

Hybrid detector concept

IBF reduction: approx. 3%Charge splitting processesLarger Gas Gain

8mmx8mm pad size0.5 mm pad spacing

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 8

Introduction: Outline

• Introduction: COMPASS RICH-1 @ CERN

• Why MPGD-based photon detectors and a change in the Cherenkov photo detection technology

• The architecture of the MPGD-based detector

• Construction, quality control and assembly

• Detector commissioning

• Performance hints

Outline

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 9

THGEM characterization and studies, IBF reduction with MicroMegas

THGEM characterization

Photoelectron extraction

IBF (Ion Back Flow) suppressionCherenkov light detection in TB

100 μm rim

no rim

Time resultion~7 ns

UV light scanvs E_driftPhoton yield (blue)

& Charged Particles (red) vs Drift Field

IBF suppression (<3%)introducing a MM stage

Hybrid architecture

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 10

THGEM: quality assessment

Measurement of the raw material thickness before the THGEM production, accepted:± 15 µm ↔ gain uniformity σ < 7%

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 11

THGEM: quality assessment

Measurement of the raw material thickness before the THGEM production, accepted:± 15 µm ↔ gain uniformity σ < 7%

THGEM polishing with an “ad hoc” protocol setup by usincluding backing:>90% break-down limit obtained

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 12

THGEM and Micromegas: quality assessment

Measurement of the raw material thickness before the THGEM production, accepted:± 15 µm ↔ gain uniformity σ < 7%

THGEM polishing with an “ad hoc” protocol setup by usincluding backing:>90% break-down limit obtained

X-ray THGEM test to accessgain uniformity (<7%) and spark behaviour

55Fe source spectra

X-ray MM test to access integrity and gain uniformity (<5%)

MICROMEGAS

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 13

Micromegas: characterization and studies, the discrete element approach

1 Single pad scheme:

Blue pad at HV via individual pad resistor at the PCB rear surface

Red pad: signal induced by RC coupling

Resistor arraysConnector 8+1 pin

Test of the (4 x 2) 30 x 60 cm2 MMs[in total: 1.4 m2, 19040 pads]:

-2 pads with shorts-1 pad: no read-out connection

3 bad pads out of 19040 before installation

“buried pad”

“surface anode” pad

“Z drilling controlled via” planarity issue

“Via closure” leakage issue

The COMPASS RICH-1 approach

Mesh at GroundPads HV segmentation

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 14

Finale detector: the mounting phase

mechanics

Wire planes

THGEM staggering

detector layers

Glueing the support pillars

Automatized glueing

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 15

THGEM: CsI coating at CERN

Turbopump

THGEM

THGEM box

4 evaporators

piston

evaporators

QE measurement

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 16

THGEM: CsI coating at CERN, QE measurements

19 CsI evaporations performed in 2015 – 2016 on 15 pieces: 13 THGEMs, 1 dummy THGEM,and 1 reference piece (best from previous coatings) on gold coated substrate

11 coated THGEMs available, 8 used + 3 spares

THGEM number evaporation date at 60 degrees at 25 degreesThick GEM 319 1/18/2016 2.36 2.44Thick GEM 307 1/25/2016 2.65 2.47Thick GEM 407 2/2/2016 2.14 2.47Thick GEM 418 2/8/2016 2.79 2.98Thick GEM 410 2/15/2016 2.86 3.14Thick GEM 429 2/22/2016 2.75 2.74Thick GEM 334 2/29/2016 2.77 3.00Thick GEM 421 re-coating 3/10/2016 2.61 2.83

Reference piece 7/4/2016 3.98 3.76

QE is the result of a surface scan(12 x 9 grid, 108 measurements)

Good uniformity, in the exampleσQE / <QE> = 3%

QE measurements indicate

<THGEM QE> =0.73 x Ref. pieceQE

in agreement with expectations

THGEM optical opacity = 0.77

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 17

Introduction: Outline

• Introduction: COMPASS RICH-1 @ CERN

• Why MPGD-based photon detectors and a change in the Cherenkov photo detection technology

• The architecture of the MPGD-based detector

• Construction, quality control and assembly

• Detector commissioning

• Performance hints

Outline

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 18

Final mounting on the RICH-1 detector

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 19

The High Voltage Control System

In total 136 HV channels

with correlated values

Hardware, commercial by CAENCustom HV control system

Custom-made (C++, wxWidgets) Compliant with COMPASS DCS (slow control) “OwnScale” to fine-tune for gain uniformity V, I measured and logged at 1 Hz Autodecrease HV if needed (too high spark-rate) User interaction via GUI Correction wrt P/T to preserve gain stability

Gain stability vs P, T: G = G(V, T/P) Enhanced in a multistagedetector ∆T = 1oC ∆G ≈ 12% ∆P = 5 mbar ∆G ≈ 18 %

THE WAY OUT: Compensate T/P variations by V

Gain stability better than 10%

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 20

THGEM and MICROMEGAS what have we learned so far:

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 21

The final detector: noise figure

MWPCs (0.2 pF): <σ> ~ 700 e-

Hybrids (4 pF): <σ> ~ 900 e-

APV 25 FE chipApproximately 70 kch.

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 22

Introduction: Outline

• Introduction: COMPASS RICH-1 @ CERN

• Why MPGD-based photon detectors and a change in the Cherenkov photo detection technology

• The architecture of the MPGD-based detector

• Construction, quality control and assembly

• Detector commissioning

• Performance hints

Outline

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 23

The Cherenkov photon signal detection

(A0<5 ADC units, 0.2<A1/A2<0.8)

All sectors provide the same time response

1 µs

Selecting good hit candidates Clusterization to separate charged particle tracks

MWPC (old detector)

APV saturationphotons

Charged Tracks

Noise

Charged particle ionization suppression by strong reversed bias

Hybrid MPGD (novel detector)

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 24

The Cherenkov photon signal detection

(A0<5 ADC units, 0.2<A1/A2<0.8)

All sectors provide the same time response

1 µs

Selecting good hit candidates Clusterization to separate charged particle tracks

MWPC (old detector)

APV saturationphotons

Charged Tracks

Noise

Charged particle ionization suppression by strong reversed bias

Hybrid MPGD (novel detector)

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 25

Preliminary results of the new detectors

Correlation between photons and trajectories

For reference:

θ (β = 1) = 52.5 mrad

Ring centre (calc.)

p = 3.8 GeV/cθ = 38 mrad p = 3.5 GeV/c

θ = 34 mradp = 7.8 GeV/cθ = 49 mrad

p = 8.4 GeV/cθ = 49.5 mrad

p = 4.8 GeV/cθ = 43.5 mrad

σ = 1.9 mrad

Residual distribution for individual photons (preliminary π-sample):

θ calculated−θ photon

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 26

Conclusion, remarks

Four hybrid PDs covering 1.4 m2 were built, tested and mounted on COMPASS RICH and successfullyoperated during 2016 and 2017 run.

COMPASS RICH-1 is the first RICH where single photon detection is accomplished by MPGDs

Characterization is ongoing!

QUESTIONS ?

Thanks for your attention!

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 27

Conclusion, remarks

SPARES

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 28

Hybrids and MWPC

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 29

Photon signal

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 30

Field shaping near the detector borders

large field values at the chamber edges and on the guard wires

Field shaping electrodes in the isolating material protections of the chamber frames

isolating material (Tufnol 6F/45) protection

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 31

P/T correction, the method

10500

11500

12500

13500

14500

9:00:00 AM 1:00:00 PM 5:00:00 PM 9:00:00 PM 1:00:00 AM 5:00:00 AM 9:00:00 AM 1:00:00 PM 5:00:00 PM 9:00:00 PM 1:00:00 AM 5:00:00 AM

Effec

tive G

ain

Time[HH:MM:SS]

Effective Gain Vs Time: With Correction (26/06 to 28/06)

16000

17000

18000

19000

20000

21000

22000

23000

24000

10:00:00 AM 2:00:00 PM 6:00:00 PM 10:00:00 PM 2:00:00 AM 6:00:00 AM 10:00:00 AM 2:00:00 PM 6:00:00 PM

Effe

ctive

Gain

Time[HH:MM:SS]

Effective Gain Vs Time For 2 days (30/06 and 01/07)

With Correction Without Correction

GAIN MEASUREMENTSALTERNATING

W, W/O CORRECTION

Correction of Voltage 𝑓𝑓 𝑃𝑃,𝑇𝑇 LabVIEW based system fully automated + logging

p, T sensor at gas input and output

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 32

IBF measurements

Trieste home-built picoammeters

The result of the direct measurement: 3% nicely matches the expectation

Stefano Levorato – INFN Trieste – 2017 Nuclear Science Symposium – Atlanta, Georgia -USA 33

CsI as photon detector, handling the VUV domain

COMPASS RICH-1, gas transparency-gas cleaning by on-line filters,-separate functions:

-Cu catalyst, ~ 400C for O2-5A molecular sieve, ~ 100C for H2O

transmission through 1,87 m, corresponding to:

H2O: ~1 ppm,

O2: ~3 ppm

typical

tran

smis

sion

00.

20.

40.

60.

81

160 170 180 190wavelength (nm)

200 210

CsI QE

quartz

(n-1) r.m.s (assuming Frank and Tamm):

30×10−6 46×10−6

PHENIX HBD CsI + GEMs

CsI gasous sensors usedin several Cherenkov detectors

Ref

ract

iein

dex