60
The Effect of Translucency and Background Variations on the Color Difference of CAD/CAM Lithium Disilicate Glass Ceramic Abdulaziz Al Ben Ali, DMD “Thesis submitted in partial fulfillment of the requirements for the degree of Master of Science” “Tufts University School of Dental Medicine” Post-Graduate Prosthodontics Division

The Effect of Translucency and Background Variations on

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: The Effect of Translucency and Background Variations on

The Effect of Translucency and Background

Variations on the Color Difference of

CAD/CAM Lithium Disilicate Glass Ceramic

Abdulaziz Al Ben Ali, DMD

“Thesis submitted in partial fulfillment of the requirements for the degree of

Master of Science”

“Tufts University School of Dental Medicine”

Post-Graduate Prosthodontics Division

Page 2: The Effect of Translucency and Background Variations on

ii

Principal Advisor:

Hiroshi Hirayama, DDS, DMD, MSc, FACP

Thesis Committee Members:

Kiho Kang, DDS, DMD, MSc, FACP

Roya Zandparsa, DDS, DMD, MSc

Matthew Finkelman, PhD

Page 3: The Effect of Translucency and Background Variations on

iii

ABSTRACT Statement of the problem: Replication of natural tooth color is one of the most difficult

challenges in prosthetic dentistry, especially in single tooth restorations in the anterior region and

when the adjacent tooth is discolored or non-vital.

Purpose of the study: The purpose of this study is to compare the effect of translucency and

background variations on the color difference of different shades of CAD/CAM lithium disilicate

glass ceramic.

Materials and methods: A sample size of n=10 was suggested according to a pilot study.

Lithium disilicate glass ceramic cylinders (12 mm in diameter and 13 mm in length) in three

different ceramic shades (BL1, A2 and C3) were fabricated by CAD/CAM technology.

Specimen disks were cut into sections of 1.2 mm thickness and 12 mm diameter. Four different

combinations of translucency and background color were considered among the different shades:

group 1 (HT ND1), group 2 (HT ND4), group 3 (LT ND1), and group 4 (LT ND4). Each

specimen was placed against different backgrounds (ND1, ND4). A spectrophotometer was used

to measure the color difference (ΔE) and ΔLab. Non-parametric tests (Kruskal-Wallis tests) were

used to evaluate the color differences among the tested groups, and follow-up tests (Mann-

Whitney U tests) were used with Bonferroni correction. Furthermore, for each ceramic shade,

high translucency groups were compared to low translucency groups to determine the influence

of translucency on color differences using a non-parametric test (Mann-Whitney U test).

Results: All tested groups displayed statistical significance (P< 0.001). Significant differences

were present among the tested groups of the same shade (P< 0.001). Additionally, the data

revealed that there were significant differences between the high translucency and the low

translucency groups (p < 0.001).

Page 4: The Effect of Translucency and Background Variations on

iv

Conclusion: Within the limitations of this study, it is suggested that the translucency and

background color significantly influenced the lithium disilicate glass ceramic color among the

BL1, A2, and C3 shades. Changing the underlying color from lighter (ND1) to darker (ND4)

resulted in increased color differences (ΔE).

Page 5: The Effect of Translucency and Background Variations on

v

DEDICATION

To my beloved wife and closest friend, Reem Al-Dakheel, who inspired me to follow my

dreams, believed in me and knew I would succeed.

To my parents, Wedad Al-Mukhaizeem and Mubarak Al-Ben Ali, for their unconditional

support, love and their gift of life.

To my dear family and friends who always encouraged me and supported me.

To everyone in my life who motivated me to move forward. I owe you all an unlimited debt of

appreciation.

Page 6: The Effect of Translucency and Background Variations on

vi

ACKNOWLEDGMENTS

First, I am grateful to Allah, The Almighty God, for all the rewards he grants me and for

giving me strength throughout my life. Second, to my home country, Kuwait, and the

Government of the State of Kuwait for supporting me through out my education. Without their

support I would not have been able to achieve my goals.

I owe my deepest gratitude to my principal advisor, Dr. Hiroshi Hirayama, for all the help and

support he offered me to complete this project and for the time contributed. It is his great

knowledge and guidance that helped me accomplish my project.

I would like to express my appreciation to Dr. Matthew Finkelman for his help with the

statistical analysis section and his way of simplifying and delivering the information, which

helped finalize this project.

I would also like to extend my gratefulness to Dr. Ki-Ho Kang for always helping me to think

outside the box. It is his great ideas and critical way of thinking that always inspired me to reach

higher levels.

I am thankful as well to Dr. Roya Zandparsa for her great support, guidance and encouragement

throughout the project. I would also like to thank Mrs. Susan Brown for her hard work in

coordinating me, the committee, and the companies.

Last but not least, a special thanks to Ivoclar Vivadent and Olympus for kindly providing me

with the materials that I used in this study.

Page 7: The Effect of Translucency and Background Variations on

vii

Table of Contents ABSTRACT .................................................................................................................................. iii DEDICATION ............................................................................................................................... v ACKNOWLEDGMENTS ........................................................................................................... vi Index of Tables: .......................................................................................................................... viii Table of Figures: .......................................................................................................................... ix

Introduction: ................................................................................................................................. 2 Background: .................................................................................................................................. 3

Color in dentistry: .................................................................................................................................. 3 Ceramics in dentistry: ............................................................................................................................ 3 All-ceramic restorations: ........................................................................................................................ 4 IPS e.max lithium disilicate glass ceramic: .......................................................................................... 5 Effect of abutment color: ....................................................................................................................... 6 Color determination: .............................................................................................................................. 8

Visual color determination: .................................................................................................................. 8 Instrumental color determination: ........................................................................................................ 9

Color measurement systems: ............................................................................................................... 12 Munsell color system: ........................................................................................................................ 12 CIE color system: ............................................................................................................................... 13

Aim: .............................................................................................................................................. 15

Hypotheses: .................................................................................................................................. 15 Clinical Implication: ................................................................................................................... 15

Materials and methods: .............................................................................................................. 16 Sample size: ........................................................................................................................................... 16 Specimen fabrication: ........................................................................................................................... 16 Finishing: ............................................................................................................................................... 17 Crystallization: ...................................................................................................................................... 18 Glaze Firing: ......................................................................................................................................... 18 Colored background: ........................................................................................................................... 18 Spectrophotometric measurement: ..................................................................................................... 19 Statistical analysis: ............................................................................................................................... 20

Results: ......................................................................................................................................... 21 Discussion: ................................................................................................................................... 24

Limitation of the study: ........................................................................................................................ 27 Future study: ......................................................................................................................................... 27

Conclusions: ................................................................................................................................. 29

Bibliography: ............................................................................................................................... 30

Page 8: The Effect of Translucency and Background Variations on

viii

Index of Tables: TABLE  1-­‐  CRYSTALLIZATION/GLAZING  CYCLE  OF  LITHIUM  DISILICATE  GLASS  CERAMIC  FURNACE  

(PROGRAMAT  P700).  ......................................................................................................................................................................  30  TABLE  2-­‐  LAB  VALUES  OF  REFERENCE  GROUPS  OF  BL1  GROUP,  A2  GROUP,  AND  C3  GROUP.  ..............................  30  TABLE  3-­‐  LAB  VALUES  AND  ΔE  VALUE  OF  BL1  HT  ND4  (MEAN  AND  STANDARD  DEVIATION).  ...........................  30  TABLE  4-­‐  LAB  VALUES  AND  ΔE  VALUE  OF  BL1  LT  ND1  (MEAN  AND  STANDARD  DEVIATION).  ............................  31  TABLE  5-­‐LAB  VALUES  AND  ΔE  VALUE  OF  BL1  LT  ND4  (MEAN  AND  STANDARD  DEVIATION).  .............................  31  TABLE  6-­‐  LAB  VALUES  AND  ΔE  VALUE  OF  A2  HT  ND4  (MEAN  AND  STANDARD  DEVIATION).  ..............................  31  TABLE  7-­‐  LAB  VALUES  AND  ΔE  VALUE  OF  A2  LT  ND1  (MEAN  AND  STANDARD  DEVIATION).  ...............................  32  TABLE  8-­‐  LAB  VALUES  AND  ΔE  VALUE  OF  A2  LT  ND4  (MEAN  AND  STANDARD  DEVIATION).  ...............................  32  TABLE  9-­‐  LAB  VALUES  AND  ΔE  VALUE  OF  C3  HT  ND4  (MEAN  AND  STANDARD  DEVIATION).  ..............................  33  TABLE  10-­‐  LAB  VALUES  AND  ΔE  VALUE  OF  C3  LT  ND1  (MEAN  AND  STANDARD  DEVIATION).  ............................  33  TABLE  11-­‐  LAB  VALUES  AND  ΔE  VALUE  OF  C3  LT  ND4  (MEAN  AND  STANDARD  DEVIATION).  ............................  33  TABLE  12-­‐  KRUSKAL-­‐WALLIS  TEST  RESULTS  AND  DESCRIPTIVE  ANALYSIS  FOR  BL1  TESTED  GROUP  (STD.  

DEV.=STANDARD  DEVIATION  +  IQR=  INTERQUARTILE  RANGE)  .............................................................................  34  TABLE  13-­‐  KRUSKAL-­‐WALLIS  TEST  RESULTS  AND  DESCRIPTIVE  ANALYSIS  FOR  A2  TESTED  GROUP  (STD.  

DEV.=STANDARD  DEVIATION  +  IQR=  INTERQUARTILE  RANGE).  ............................................................................  34  TABLE  14-­‐  KRUSKAL-­‐WALLIS  TEST  RESULTS  AND  DESCRIPTIVE  ANALYSIS  FOR  C3  TESTED  GROUP  (STD.  

DEV.=STANDARD  DEVIATION  +  IQR=  INTERQUARTILE  RANGE).  ............................................................................  34  TABLE  15-­‐  POST-­‐HOC  TEST  (MANN-­‐WHITNEY  U  TEST)  BL1  GROUPS  ...............................................................................  34  TABLE  16-­‐  POST-­‐HOC  TEST  (MANN-­‐WHITNEY  U  TEST)  A2  GROUPS  .................................................................................  35  TABLE  17-­‐  POST-­‐HOC  TEST  (MANN-­‐WHITNEY  U  TEST)  C3  GROUPS.  ................................................................................  35  TABLE  18-­‐  DESCRIPTIVE  ANALYSIS  FOR  BL1  TESTED  GROUP  (STD.  DEV.=STANDARD  DEVIATION  +  IQR=  

INTERQUARTILE  RANGE)  DEPENDENT  VARIABLE  ΔL  ΔA  ΔB.  .....................................................................................  35  TABLE  19-­‐  DESCRIPTIVE  ANALYSIS  FOR  A2  TESTED  GROUP  (STD.  DEV.=STANDARD  DEVIATION  +  IQR=  

INTERQUARTILE  RANGE)  DEPENDENT  VARIABLE  ΔL  ΔA  ΔB.  .....................................................................................  36  TABLE  20  -­‐DESCRIPTIVE  ANALYSES  FOR  C3  TESTED  GROUP  (STD.  DEV.=STANDARD  DEVIATION  +  IQR=  

INTERQUARTILE  RANGE)  DEPENDENT  VARIABLE  ΔL  ΔA  ΔB.  .....................................................................................  36  TABLE  21-­‐  MANN-­‐WHITNEY  U  TEST  RESULTS  FOR  HT  AND  LT  OF  BL1  GROUPS.  DEPENDENT  VARIABLE=  ΔE

 ..................................................................................................................................................................................................................  36  TABLE  22-­‐  MANN-­‐WHITNEY  U  TEST  RESULTS  FOR  HT  AND  LT  OF  A2  GROUPS.  DEPENDENT  VARIABLE=  ΔE

 ..................................................................................................................................................................................................................  37  TABLE  23-­‐  MANN-­‐WHITNEY  U  TEST  RESULTS  FOR  HT  AND  LT  OF  C3  GROUPS.  DEPENDENT  VARIABLE=  ΔE

 ..................................................................................................................................................................................................................  37  

Page 9: The Effect of Translucency and Background Variations on

ix

Table of Figures: FIGURE  1-­‐  LEFT:  PARTIALLY  CRYSTALLIZED  (HT)  CAD/CAM  LITHIUM  DISILICATE  GLASS  CERAMIC.                                      

RIGHT:  FULLY  CRYSTALLIZED  (HT)  CAD/CAM  LITHIUM  DISILICATE  GLASS  CERAMIC.  ...............................  38  FIGURE  2-­‐  LEFT:  PARTIALLY  CRYSTALLIZED  (LT)  CAD/CAM  LITHIUM  DISILICATE  GLASS  CERAMIC.                                      

RIGHT:  FULLY  CRYSTALLIZED  (LT)  CAD/CAM  LITHIUM  DISILICATE  GLASS  CERAMIC.  ................................  38  FIGURE  3-­‐  MUNSELL  COLOR  SYSTEM.  ...............................................................................................................................................  38  FIGURE  4-­‐  CIE  LAB  L:  LIGHTNESS,  A:  RED  AND  GREEN,  AND  B:  YELLOW  AND  BLUE.  ................................................  38  FIGURE  5-­‐  E4D  CAD/CAM  MACHINE  AND  PRE-­‐CRYSTALLIZED  LITHIUM  DISILICATE  GLASS  CERAMIC.  ..........  39  FIGURE  6-­‐  ISOMET  1000  MACHINE  USED  TO  CUT  THE  CYLINDER  INTO  DISKS  WITH  1.2  MM  THICKNESS.  ...  39  FIGURE  7-­‐  BL1  SHADE  GROUPS.  ...........................................................................................................................................................  39  FIGURE  8-­‐  A2  SHADE  GROUPS.  ..............................................................................................................................................................  40  FIGURE  9-­‐  C3  SHADE  GROUPS.  ..............................................................................................................................................................  40  FIGURE  10-­‐  LEFT:  PRE-­‐CRYSTALLIZED  A2  LT  LITHIUM  DISILICATE  GLASS  CERAMIC  SPECIMEN.                                          

MIDDLE:  PROGRAMAT  P700  FURNACE.                                          RIGHT:  CRYSTALIZED  A2  LT  LITHIUM  DISILICATE  GLASS  CERAMIC  SPECIMEN.  .......................................................................................................................................................  41  

FIGURE  11-­‐  IPS  E.MAX  CERAM  GLAZE  PASTE.  ...............................................................................................................................  41  FIGURE  12-­‐  DIGITAL  CALIPER  (DENTAGUAGE  1)  CONFIRMING  THE  SPECIMEN  THICKNESS  1.2MM.  ...............  41  FIGURE  13-­‐  ND1  AND  ND4  SHADE  IPS  NATURAL  DIE  MATERIAL.  .......................................................................................  41  FIGURE  14  LEFT:  ND4  BACKGROUND  WITH  SPECIMEN  HOLDER.                                        RIGHT:  SPECIMEN,  ND1  AND  

SPECIMEN  HOLDER  IN  THE  DARK  BOX.  ................................................................................................................................  42  FIGURE  15-­‐  SPECTROPHOMETER  (CRYSTALEYE)  WITH  CUSTOM-­‐POSITIONING  JIG  AND  DARK  BOX.  ..............  42  FIGURE  16-­‐  LEFT:  ΔE*  AND  LAB*  OF  THE  TARGET  (SPECIMEN)  AND  THE  REFERENCE  (CONTROL  GROUP)  OF  

GROUP  7  (A2  LT  ND1).                                          RIGHT:  2  MM  IN  DIAMETER  POSITIONED  OVER  THE  MIDDLE  REGION  OF  THE  SPECIMEN.  ..........................................................................................................................................................................  42  

FIGURE  17-­‐  BOXPLOT  ILLUSTRATION  OF  THE  RESULTS  OF  BL1  GROUPS  ΔE*.  .............................................................  43  FIGURE  18-­‐  BOXPLOT  ILLUSTRATION  OF  THE  RESULTS  OF  A2  GROUPS  ΔE*.  ................................................................  43  FIGURE  19-­‐  BOXPLOT  ILLUSTRATION  OF  THE  RESULTS  OF  C3  GROUPS  ΔE*.  ................................................................  43  FIGURE  20-­‐  MANN-­‐WHITNEY  U  TEST  RESULTS  FOR  HT  AND  LT  OF  BL1  GROUPS.  DEPENDENT  VARIABLE=  

ΔE  .............................................................................................................................................................................................................  44  FIGURE  21-­‐  MANN-­‐WHITNEY  U  TEST  RESULTS  FOR  HT  AND  LT  OF  A2  GROUPS.  DEPENDENT  VARIABLE=  ΔE

 ..................................................................................................................................................................................................................  44  FIGURE  22-­‐  MANN-­‐WHITNEY  U  TEST  RESULTS  FOR  HT  AND  LT  OF  C3  GROUPS.  DEPENDENT  VARIABLE=  ΔE

 ..................................................................................................................................................................................................................  44  

Page 10: The Effect of Translucency and Background Variations on

1

The Effect of Translucency and Background

Variations on the Color Difference of

CAD/CAM Lithium Disilicate Glass Ceramic

Page 11: The Effect of Translucency and Background Variations on

2

Introduction:

All-ceramic materials were introduced as metal-free restorations. Ceramics improve

esthetics by allowing light transmission through the restoration and the underlying tooth

structure1-3. One commonly used all-ceramic system is the lithium disilicate glass ceramic, in

which the alumino-silicate glass has lithium disilicate crystals. Lithium disilicate crystals are

needle-like in shape and include about two-thirds of the glass ceramic volume, presenting

outstanding esthetics, high strength, and the ability to be cemented or adhesively bonded4. In

fact, due to the fairly low refractive index of the lithium disilicate crystals, this material can be

very translucent4-6. However, the material can be fabricated in either a pressable or machinable

manner with CAD/CAM technology 7,8.

Although the high translucency of lithium disilicate glass ceramic is advantageous, it might be

challenging in cases of restoring non-vital or discolored teeth. Yet, the influence of different

translucencies on the resulting color has not been tested.

The purpose of this study is to compare the effect of translucency and background

variations on the color difference of different shades of CAD/CAM lithium disilicate glass

ceramic.

Page 12: The Effect of Translucency and Background Variations on

3

Background:

Color in dentistry:

In dentistry, the combination effects of extrinsic and intrinsic colors circumscribe the

tooth color. The extrinsic color is correlated with the absorption of materials onto the surface of

the enamel and dentin, whereas the intrinsic tooth color is connected with the light absorption

and scattering properties of the enamel9-11. Lemire described tooth color as a result of the

reflection of light from the tooth surface combined with light redirected from the dentin that has

undergone some internal refractions and reflections12,13. Moreover, Munsell explained that color

has three dimensions: hue, chroma and value14.

Translucency is one of the vital properties that should also be taken into consideration to

fabricate an esthetic restoration. The latter is defined as the amount to which light is diffused

rather than reflected or absorbed15. High translucency enamel overlaying the dentin results in

greater translucency at both the incisal third and the proximal surfaces of a tooth16. The middle

third of a tooth has a large amount of yellowish dentin, which influences the color of the

covering enamel16. However, the translucent blue-grey enamel often alters the dentin so that the

ending color is a combination of yellow, orange, blue and grey16-21.

Ceramics in dentistry:

In the 18th century, ceramics were first used in dentistry for porcelain dentures21,22. Due

to the brittle nature of this material, it was not until 1962 that the Weinstein brothers invented the

clinically reliable porcelain fused to metal (PFM) crown. This was accomplished by adding

leucite to porcelain preparations, which raised the coefficient of thermal expansion to permit

their merging to certain gold alloys to fabricate full crowns and fixed partial dentures5,23. In

Page 13: The Effect of Translucency and Background Variations on

4

1965, McLean introduced the aluminous porcelain all-ceramic crown, and since then, it has been

improved and employed by both clinicians and technicians16. Currently, there are five methods to

produce all-ceramic crowns: cast and creaming, slip casting, condensation and sintering,

pressing, and computer-aided design / computer-aided manufacturing (CAD / CAM) milling of

pre-crystallized blocks or ceramic blocks. The increased requirement for esthetic restorations has

resulted in the demand for all-ceramic crowns2. The spectrum of dental ceramics is divided into

three fields: predominantly glassy materials, particle-filled glasses, and polycrystalline ceramics.

Highly esthetic dental ceramics are predominantly glassy, while crystalline ceramics are

considered to be higher strength ceramics5.

All-ceramic restorations:

In 1903, the first jacket crown was invented. Since that time, all-ceramic restorative

materials have been under continuous development17,18. To meet esthetic requirements, several

all-ceramic systems have been established. The method of fabrication varies among the systems.

Some systems contain cores, veneered with core material5. The core enhances the strength of the

restoration and may range from opaque to semi-translucent4,16.

Layering the porcelain veneer over ceramic frameworks can achieve inherent beauty and

beneficial light scattering5,24. The veneering material usually has fluorescent properties that

resemble those of the natural teeth25. Usually, veneering porcelains involves a leucite, aluminum

oxide, or glass and crystalline phase of fluoroapatite26. Veneering a zirconium oxide core with

glass, aluminum oxide, or lithium disilicate allows dental technicians to tailor the form and

esthetics of these restorations in terms27. The chipping of the veneering porcelain and/or the

fracture of the coping are the most commonly encountered main clinical complication that results

the in failure of all-ceramic restorations26,28-34.

Page 14: The Effect of Translucency and Background Variations on

5

IPS e.max lithium disilicate glass ceramic:

Several advancements have taken place in combination with lithium disilicate (Li2Si2O5)

glass ceramics35-37. The principal crystal phase is fabricated in the different base glass of the

SiO2–Li2O–K2O–ZnO–P2O5–Al2O3–La2O3 system through heterogeneous nucleation and

crystallization36,38.

In development, an interconnecting microstructure with a crystal content of 460-vol% is

established. The developing product was named lithium disilicate (IPS Empress 2, Ivoclar

Vivadent AG). Additionally, for use in the anterior region, it was introduced to the global market

in 199938-40. A substantial enhancement over IPS Empress 2 was introduced in the SiO2–Li2O–

K2O–Al2O3–ZrO2 system and developed in the production of lithium disilicate glass ceramic

(IPS e.max, Ivoclar Vivadent AG)41,42.

Thorough analysis of the crack propagation in this glass ceramic revealed that the high

fracture toughness, which is 2.3 MPa 0.5 mm, measured by the SEVNB method as a KIC value,

was initiated by crack divergence in the vicinity of the disilicate crystals23. A considerable

amount of energy is lost from the propagating crack while deviating23. This yields an increased

amount of flexural strength (up to 440 MPa) and toughness23.

This material provides prime esthetics, yet has the strength for conventional or adhesive

cementation. In addition to outstanding optical properties, lithium disilicate glass ceramic has a

needle-like crystal configuration that offers excellent durability and strength4,7,23. Furthermore, it

can be conventionally pressed or contemporarily processed through CAD/CAM technology23.

This lithium metasilicate glass ceramic exhibits a single blue color6,23. The blue glass ceramic

undergoes a heat treatment at 850 °C after it has been milled and successfully tried in the

patient’s mouth23. Throughout thermal treatment, it is converted into lithium disilicate glass

Page 15: The Effect of Translucency and Background Variations on

6

ceramic23. Among the tested materials, pressed lithium disilicate glass ceramic demonstrated the

highest light transmission rate42.

Two levels of translucency may be acquired based on the pre-crystallization treatment of

the CAD/CAM ceramic blocks. There is a small number and bigger size of lithium metasilicate

crystals in the pre-crystallized state of the high translucency (HT) material (Fig.1)42, whereas the

low translucency (LT) material contains a larger number of smaller crystals (Fig.2)42. Covered

lithium disilicate crystals (1.5 × 0.8 µm) in a glassy matrix are observed in the HT ceramic

following complete crystallization heat treatment at 850 °C for 10 minutes (Fig.1)42. Highly

soluble lithium phosphate spherical crystals look like spherical pores42. The fully crystallized LT

ceramic has a large number of small (0.8 × 0.2 µm) interconnected lithium disilicate crystals

along with spherical pores, referred to as lithium phosphate crystals (Fig. 2)42.

Effect of abutment color:

When ceramics have higher translucency, more light is able to transmit and scatter. This

means that the underlying abutment has a major impact on the final color43.

However, in clinical conditions, all-ceramic crowns are frequently used in cases such as a

non-vital tooth that has been endodontically treated or a multichromatic abutment44. Thus, it is

crucial to consider the color of the crown as well as the color of the abutment tooth involved45.

The combination of the thickness of the ceramic, the underlying abutment color, and the color of

the cement articulates the optical behavior of ceramic restorations43.

The color of the restoration may be influenced by several aspects, including the ceramic firing

temperature46-48, the number of ceramic firing cycles 46,49,50, surface glaze 51, opaque ceramic

thickness 52-54, ceramic thickness43,49,52,53,55-57, manufacturer50,58-61, metal surface treatment62, and

type of substructure43,62-67.

Page 16: The Effect of Translucency and Background Variations on

7

Previous studies showed that the underlying tooth structure is a principal factor affecting

the look of final ceramic restorations 43,44,57,65,68,69. The color underneath the crown may lead to

shadowing and discoloration of the restoration if an all-ceramic restoration was restored on a

dark underlying tooth structure, for example, a tooth treated with a root canal treated tooth43,44.

Chaiyabutr et al studied the collective outcome of the tooth abutment color, cement color,

and ceramic thickness in the subsequent optical color of a CAD/CAM lithium disilicate glass

ceramic crown44. Their findings suggested that the ΔE values of a CAD/CAM lithium disilicate

glass ceramic crown were significantly affected by the tooth abutment color (P < 0.001), cement

color (P < 0.001), and ceramic thickness (P < 0.001). In addition, significant differences existed

between these three variables (P < 0.001). For example, a dark-colored abutment tooth presented

the greatest ΔE values compared to the other variables. Additionally, there was a significant

decrease in ΔE values when the ceramic thickness was increased (P < 0.01). Furthermore, when

the crowns were cemented using the opaque cement, the ΔE values slightly decreased44.

In a study that assessed the impact of the abutment material on the color of IPS Empress

2 ceramic copings with different thicknesses, they used different degrees of thickness (1, 1.2, 1.4,

1.6, 1.8, or 2 mm) against three different abutments, which were composite, a gold alloy, or a

silver palladium alloy43. Using a colorimeter, color was evaluated according to the CIE LAB

system, and ΔE was calculated43. They found that the abutment material and ceramic thickness

significantly affected the ΔE values43.

Nakamura et al stated that the abutment impacts the ceramic color when the thickness of

ceramic is less than 1.6 mm43, while other studies suggest that the thickness of the ceramic

should be at least 2.0 mm to moderate the effect of the abutment tooth on the overall

color1,44,57,70,71. Furthermore, Heffernan et al studied the effect of core and core-veneering

Page 17: The Effect of Translucency and Background Variations on

8

ceramic thickness on the resultant translucency71,72. According to Lee et al, based on the type of

all-ceramic core material, the layered color of various all-ceramic and veneer combinations was

different, even though the thickness of the layered specimen was set to 1.5 mm24.

Kwiatkowski et al and Ahmad et al discussed that there is an influence of the underlying

tooth structure on the increased light transmission of all-ceramic restorations, whether normal,

discolored, or treated with a post-and-core or a buildup73-75. Moreover, with the recent

technological developments in ceramic, zirconium and fiber-reinforced posts, there are new

possibilities for restorative solutions73,74.

Color determination:

Visual color determination:

In clinical dentistry, the most frequently applied method is visual determination. The

human eye is capable of spotting the minute color variations between two objects and is essential

for color communication76. By using the shade tabs available in the market, the best matching

shade is communicated to the dental laboratory through verbal, graphic and photographic means

77-79. The VITA Classic shade guide system (Vident, Brea, CA) is commonly applied78,79. It

classifies the shades into letter groups according to the tooth hue: A (Orange), B (Yellow), C

(Yellow / Gray), and D (Yellow, Orange or Brown)77-79. Then, to describe chroma and value,

numbers are added to the letters77-79. For example, 1 is less chromatic and higher in value, while

4 is more chromatic and lower in value. Further rational shade guide systems, such as Vitapan

3D-Master, have been introduced to simplify or decrease subjectivity in the process of shade

selection77-79. This shade guide system is based on the sequential determination of the value, the

chroma, and finally, the hue77-79. The human eye is more sensitive to changes in value than subtle

changes in hue15. Usually, the restoration is clinically acceptable when value and chroma are

Page 18: The Effect of Translucency and Background Variations on

9

correct, even if the hue is not precisely matched15. Yet, the visual determination of shade has

been discovered to be unpredictable and subjective. In general, differences in color perception

are present between 25 different subjects: patients are usually more pleased with the shade match

of their PFM restorations than their dentists, especially if the practitioners are prosthodontists80.

Inconsistencies may occur due to one or more of the following: general variables (external light

conditions and background), physiologic variables (such as color blindness), and personal

features (experience, nutrition, fatigue of the human eye, medications, emotions, and age)15,81,82.

Nevertheless, there are limited standardized verbal means to communicate visually assessed

color characteristics83.

Moreover, color matching with shade tabs may be difficult84 for two reasons: because it

is hard to control parameters during the creation of shade tabs, and because tooth shades can vary

compared to the numbers of the shade tabs85,86.

Instrumental color determination:

Instrumental measurements serve as a more objective way to quantify color and they may

enable more precise and uniform color communication. We applied computer analysis to

photographic images to apply the RGB (red, green, blue) color model. The development of high-

tech colorimeters and spectrophotometers has augmented their use in dental color

communication and in dental research9,87-89.

Colorimeter:

The colorimeter represents an easier way of measuring tooth color. Hence, it has been

widely used in the dental field16. This instrument is designed in a way that directly measures the

color, similar to how the human eye identifies it16. Accuracy comparable to spectrophotometers

can be recorded15,16, but only limited data points may be stored15,16. Colorimetric readings have

Page 19: The Effect of Translucency and Background Variations on

10

been compared to spectrophotometer measurements and are believed to be accurate and reliable

for color difference readings77,88,90-93. Colorimetric measurements have also been compared to

human observations, but the results were inconclusive88,90,94. In most of the studies where a

positioning jig has been used, the measurements have been consistent and repeatable93. Although

colorimeters have a numerous advantages, they also exhibit disadvantages. Color may be prone

to errors because these instruments are fabricated to measure flat surfaces. Further, small

aperture colorimeters are disposed to substantial edge-loss effect77. Additionally, the results can

be severely affected by wet and dry conditions77. These systematic errors can adversely affect the

instrument accuracy, and subsequently, the intra-instrument repeatability93.

Spectrophotometer:

The spectrophotometer measures one wavelength at a time from the reflection or the

transmittance of an object. It can measure the visible spectra of both extracted and vital

teeth76,95,96. Very accurate and extensive data can be collected by the spectrophotometer and

translated into a formatted spectral curve15,76,97. Dual spectrometer devices can measure the

entire spectral range simultaneously76,96. Traditionally, spectrophotometers were complex to use

and it was difficult to obtain in vivo tooth color measurements with them90. Recently, however,

easy-use chair-side spectrophotometers have been developed. These spectrophotometers are very

precise and allow the reading of teeth translucency and reflectivity through the use of more

filters. Usually, these machines come with advanced software systems15. A recent study

concluded that the mean ΔE* value for crowns fabricated using the spectrophotometric technique

was significantly lower than the values for the crowns made by the conventional method98.

A number of studies70,76,96,99 that compared the use of different spectrophotometers to the

use of visual shade guides for determination of tooth shade revealed that spectrophotometric

Page 20: The Effect of Translucency and Background Variations on

11

methods bring about higher accuracy and better color match. A recent study100 evaluating visual

shade selection versus a spectrophometer concluded that spectrophotometric shade analysis

seemed to be more reproducible than visual shade determination that resulted in darker

recordings. In another study, a spectrophotometer was compared to three different shade guides;

the former provided more accurate results than visual selection101.

A study by Judeh et al. found that visual shade selection lowered the precision of shade

selection to approximately 31%; this is consistent with previous studies. It is linked to

inconsistencies inherent in color perception by individual person96. Moreover, two studies that

tested dental devices including color shade guides reported a reduction in shade match precision

to nearly 48%94,96,102.

Kim-Pusateri et al. examined the reliability and accuracy of four dental color-matching

devices. The reliability was defined as the consistency of the device in matching the same

specimen while the accuracy was defined as the ability of the device to provide a correct match

for a given specimen103. Three commercial shade guides, Vitapan classical (vident, Bera, CA),

Vitapan 3-D Master (Vident, Brea, CA), and Chromascop (Ivoclar, Vivadent, Schaan,

Liechtenstien) were used103. Ten shade guides from each system and four dental shade-matching

devices, ShadeVision (X-Rite America, Inc, Grand Rapids, MI), VITA Easyshade (Vident, Brea,

CA), ShadeScan (Cynovad, Montreal, Canda), and SpectoShade (MHT Optic Research AG,

Niederhasli, Switzerland) were tested103.

For the reliability evaluation, each shade tab from one Chromascop shade guide (20

shade tabs), one Vitapan Classical shade guide (16 shade tabs), and one Vitapan 3DMaster shade

guide (26 shade tabs) was measured ten separate times by each shade matching instrument103.

Whereas for the accuracy study, each shade tab from thirty shade guides was measured by each

Page 21: The Effect of Translucency and Background Variations on

12

shade-matching instrument103. Most device comparisons had comparable, high reliabilities (over

96%), showing predictable shade values from frequent measurements103. However, noticeable

variability in the devices’ accuracy (67-93%) was observed in most of the device comparisons103.

Odaira et al clinically assessed the spectrophotometer (Crystaleye, Olympus, Tokyo,

Japan)104. They moved the spectrophotometer to modify the position of the tooth to be

captured104. Then, they automatically transmitted the captured images and analyzed them by the

Crystaleye Application Master Software104. To assess the accuracy, Color Analyzing

Spectrophotometer-Iwate Medical University School of Dentistry Type1 (CAS-ID1) and

MultiSpectral Camera system (MSC-2000; Olympus, Tokyo, Japan) were used104. The reliability

of color measurement tests from three color measuring devices showed no significant difference

for the Lab values among the three devices104. The ∆E was between 0.1 and 0.9 and the mean ∆E

was 0.6 +/- 0.3 in the results of the accuracy of repeated color measurements assessment104. The

effect of exterior lighting on color measurements showed the ∆E between the two conditions was

0.9. Performing a t-test, no significant difference was found in the Lab values between the two

conditions104. While evaluating the effect of the examiner on the color measurement, there was

no significant difference in the Lab values between the five examiners. Evaluation of the

reproducibility of tooth color using the Crystaleye Spectrophotometer indicated the mean ∆E

between the target teeth and the fabricated crown as 1.2 +/- 0.4104.

Color measurement systems:

Munsell color system:

In terms of hue, value, and chroma, color may be defined according to the Munsell Color

Space14. Hue represents the color tone and it is the characteristic that allows one to differentiate

among different groups of color14. Value specifies the relative lightness or darkness of a color,

Page 22: The Effect of Translucency and Background Variations on

13

ranging from pure black to pure white14. Chroma is the level of color saturation; it defines the

color strength, intensity or vividness14 (Fig. 3). Television and computer monitors, examples of

emissive media, emit wavelengths that are a mixture of red, green and blue, thereby creating

color. These three colors are considered additive primary colors because almost all of the colors

in the visible spectrum can be produced through their combination. This media is called the RGB

color model14. However, reflective and transmissive media, like photographs or printed

materials, apply a different color model, named CMY (Cyan, Magenta, Yellow)15. In this color

system, the primary colors are those produced by the absorption of one of the RGB wavelengths

and the transmission or reflection of the others15. Cyan is produced when red is absorbed and

blue and green are transmitted or reflected; magenta is produced when green is absorbed and

blue and red are transmitted or reflected; yellow is produced when blue is absorbed and green

and red are transmitted or reflected15. Thus, cyan light is an equal mixture of blue and green,

magenta is an equal mixture of blue and red, and yellow is an equal mixture of green and red15.

CIE color system:

With the advanced technologies, data logging machines have been developed to precisely

define color, most frequently by the CIE L*a*b* color space system (Commission Internationale

d’Eclairage, International Commission on Illumination). In 1976, this color system was

introduced. It agrees with the acknowledged theory of color sensitivity based on three separate

color receptors of red, green and blue105. In the CIE Lab color space, the color space is uniform

with evenly perceived color differences. There are three axes, L*, a*, b* in this three-

dimensional color space. The main advantage of this system is that differences in color can be

stated in units that can be linked to clinical significance and visual perception. The L* color

coordinate ranges from 0 to 100 and it characterizes lightness; the a* color coordinate ranges

Page 23: The Effect of Translucency and Background Variations on

14

from -90 to 70 and it characterizes greenness on the positive axis and redness on the negative;

the b* color coordinate ranges from -80 to 100 and it characterizes yellowness (positive b*) and

blueness (negative b*)16,70,106 (Fig. 4). The differences within the specimens are measured in

ΔL*, Δa* and Δb*, and their combination is described by ΔE*, determined by different

equations: ΔE* = [(ΔL*)2+ (Δa*)2+(Δb*)2]1/215.

Page 24: The Effect of Translucency and Background Variations on

15

Aim:

To evaluate the effect of translucency and background variations on the color difference

of CAD/CAM lithium disilicate glass ceramic among different shades.

Hypotheses: 1. High translucency CAD/CAM lithium disilicate glass ceramic will exhibit more color

difference than low translucency CAD/CAM lithium disilicate glass ceramic of the same

shade.

2. Natural die 4 shade group will have more color difference than natural die 1 shade group.

Clinical Implication: The knowledge learned from this study may assist the practitioner in predicting the final color of

lithium disilicate glass ceramic restoration in the case of a multi-colored abutment by

determining how translucency and background color will affects the overall color of CAD/CAM

lithium disilicate glass ceramic.

Page 25: The Effect of Translucency and Background Variations on

16

Materials and methods:

Variables:

Translucency and background color were the two variables were tested in the study.

Translucency was divided into High Translucency (HT) or Low Translucency (LT) while

background color was divided into Natural Die (ND, Ivoclar Vivadent, Schaan, Liechtenstein)

(ND1 and ND4) among different ceramic shades (BL1, A2, or C3 shade).

Sample size:

Power calculations were performed using nQuery Advisor (version 7.0). Assuming a

variance of means of 0.535 for translucency, a variance of means of 1.215 for background, and a

standard deviation of 1.1 (values obtained from pilot study) a sample size of n=10 per group was

adequate to obtain a type I error rate of ∝= 0.05, a power of 98% for translucency, and a power

of 99% for background.

Specimen fabrication:

The specimens were designed by using a software on CAD/CAM (E4D, D4D

Technologies, LLC, Richardson, TX) to make a digital mold with 12 mm in diameter and 13 mm

length lithium disilicate glass ceramic cylinder (IPS e.max CAD, Ivocler Vivadent, Schaan,

Liechtenstein) in three different shades BL1, A2 and C3. Cylinders were fabricated with lithium

disilicate glass ceramic blocks and they were milled using CAD/CAM E4D according to

manufacturer’s recommendations (Fig. 5). The lithium disilicate glass ceramic cylinders were cut

in the pre-crystallized stage using a low speed (275 rpm) diamond disc (Isomet 1000, Buehler,

Page 26: The Effect of Translucency and Background Variations on

17

Lake Bluff, IL). The specimen disks’ thicknesses were standardized to 1.2 mm (Fig. 6), which

represents the thickness needed for middle area1.

Each ceramic shade was divided into two main groups according to translucency and two

subgroups according to background color. A total of twelve groups were derived as follows:

BL1 Shade (Fig.7) High Translucent group (HT): Group 1: HT ND1 shade (reference). Group 2: HT ND4 shade. Low Translucent group (LT): Group 3: LT ND1 shade. Group 4: LT ND4 shade.

A2 Shade (Fig.8) High Translucent group (HT): Group 5: HT ND1 shade (reference). Group 6: HT ND4 shade. Low Translucent group (LT): Group 7: LT ND1 shade. Group 8: LT ND4 shade. C3 Shade (Fig.9) High Translucent group (HT): Group 9: HT ND1 shade (reference). Group 10: HT ND4 shade. Low Translucent group (LT): Group 11: LT ND1 shade. Group 12: LT ND4 shade.

Finishing:

A green stone (Dura-Green Stone TC2, SHP, Shofu, Kyoto, Japan) was used to smooth

out the roughness around all specimens’ margins caused by the cutter machine. The thickness of

specimens was checked using a digital caliper (Dentaguage 1, Erskine Dental, Marina Del Rey,

CA).

Page 27: The Effect of Translucency and Background Variations on

18

Crystallization:

Crystallization was carried out in a ceramic furnace (Programat P700 furnace, Ivoclar

Vivadent, Schaan, Liechtenstein) following the sequence described in Table 1 (Fig. 10).

Glaze Firing:

While glaze firing can be performed simultaneously with the crystallization procedure, in

this study, it was conducted in a separate step. This ensured the completion of the crystallization

step in accordance with the manufacturer’s recommendation. The glazing material (IPS e.max

Ceram Glaze Paste) (Fig. 11) was applied on a layer of approximately 0.01 mm on one side of

each specimen using a medium flat brush (no.G4, Ivoclar Vivadent, Schaan, Liechtenstein). The

glaze firing was conducted on a honeycombed firing tray using Programat P700 furnace in the

same cycle as described in (Table.1); a digital caliper was used to ensure the specimens’

thickness had not deviated from 1.2 mm (Fig. 12).

Colored background:

For the fabrication of the ND1 and ND4 background, a pink baseplate wax (Kerr

Manufacturing Co, Romulus, MI) was used (14 mm X 27 mm X 3 mm); the dimensions were

adjusted according to the dimension of the specimen holder in the dark box. Light cured urethane

dimethacrylate (TRIAD™ Colorless, Visible Light Cure Material; Dentsply International Inc.,

York, PA) was used to fabricate a mold. ND1 and ND4 backgrounds were injected into the mold

and cured in a halogen spectral range between 400 and 500 nm (TRIAD™ 2000; Dentsply

International Inc., York, PA) to fabricate ND1 and ND4 backgrounds107 (Fig. 13,14).

Page 28: The Effect of Translucency and Background Variations on

19

Spectrophotometric measurement:

The specimens were measured by a spectrophotometer (Crystaleye, Olympus, Tokyo,

Japan). Each specimen, along with the tested background, was placed in a specimen holder

inside a black box which served to eliminate the impact of external light. Using an auto-

polymerizing resin (Ivolen, Ivoclar Vivadent, Schaan Liechtenstein), a custom-positioning jig

was made to achieve measurement repeatability and accuracy. The distance between the

specimen surface and the tip of the spectrophotometer was approximately 4 mm and the angle

was 45 degrees, in accordance with the manufacturer’s recommendations (Fig. 15). Prior to color

measurements, all specimens were cleaned and visually checked for dust.

The specimen to be measured was placed in the center of the display screen. As the disks

were combined, a droplet of distilled water was positioned between them (refraction index close

to 1.7)52. This was performed to enhance the optical contact during the spectrophotometric

measurement which served to minimize the loss of light through the margins of the specimens

(known as edge-loss)70. Over the middle region, we located a standardization area of 2 mm in

diameter. This was determined by the fabrication of a custom-made template that fits the

computer screen. During all of the measurements, the custom-made template was used (Fig. 16).

In addition, before each measurement, the spectrophotometer was recalibrated with a calibration

plate. HT ND1 group was chosen to be the reference group among different shades44. The

reference group specimens were measured as mentioned previously. The means of Lab* were

calculated (Table.2) and the data were then used to calculate the ΔE and ΔLab.

Five measurements per specimen over both background colors (ND1 and ND4) were

recorded, in CIE (CIE Lab*) coordinates, to minimize the effect of any possible misreadings.

The capture time was 0.2 seconds and the reflectance values were from 400-700 nm with 1-nm

Page 29: The Effect of Translucency and Background Variations on

20

intervals for each pixel. The data were obtained in CIE Lab color system by calculating ∆E*

through the specimens over ND1 and ND4 backgrounds. Color difference between the

specimens was calculated using the equation: ∆E* = [(∆L*) 2+(∆a*) 2+(∆b*) 2]1/215 (Tables.3-

11).

Statistical analysis:

Descriptive statistics (mean, standard deviations, median, interquartile range) were

computed for each of the nine groups and each outcome. Data analyses were conducted in IBM-

SPSS version 20. Levene’s test showed significant difference among the tested group, in other

words, there was no equality of variance and one of the ANOVA assumptions was violated.

Because of this, within each ceramic shade all groups were compared using a non-parametric test

(Kruskal-Wallis test). A post-hoc (Mann-Whitney U) test was used with the Bonferroni

correction to the α=0.05/3 to detect any significant difference between the groups within the

same ceramic shade. Furthermore, in each ceramic shade high translucency groups were

compared to low translucency groups to determine the influence of translucency on color

difference. Because the histograms showed the data were not normally distributed, a non-

parametric test (Mann-Whitney U test) was used among each ceramic shades.

Page 30: The Effect of Translucency and Background Variations on

21

Results:

A summary of the descriptive statistical analysis of the results showing the mean color

difference (ΔE) and the standard deviation for all the groups is shown in (Tables.12-14). For

illustration, box plots of the values for ΔE of the tested groups are shown in (Fig.17-18). The

means and standard deviations calculated for the BL1 shade were described as following (Mean

± Std. Dev.) N: group 2 (HT ND4) specimens is (11.30 ± 0.29) N; group 3 (LT ND1) specimens

is (4.36 ± 0.18) N; group 4 (LT ND4) specimens is (4.93 ± 0.14). While A2 shade N; group 6

(HT ND4) specimens is (8.74 ± 0.23) N; group 7 (LT ND1) specimens is (1.23 ± 0.13) N; group

8 (LT ND4) specimens is (4.63 ± 0.13). Finally, C3 shade were described as following N; group

10 (HT ND4) specimens is (6.74 ± 0.17) N; group 11 (LT ND1) specimens is (1.26 ± 0.14) N;

group 12 (LT ND4) specimens is (4.66 ± 0.25).

After computed of the descriptive analysis, a non-parametric test (Kruskal-Wallis test) was

carried out for each ceramic shade. The test revealed a significant difference between the groups

for each shade (p< 0.001). Statistical significance was set at alpha = 0.05. Looking at the box

plot in (Fig. 17-18) and (Tables 12-14), the data demonstrate a statistical significance difference

among the groups (P-value < 0.001).

Three follow-up tests (Mann-Whitney U tests) were conducted to detect any significant

differences between the groups within the same shade. The Bonferroni correction was considered

in this case, and the α of the test was adjusted to be equal to 0.05/3 = 0.017. The statistical results

of this test are summarized in Tables 15-17 and illustrated in a box plot (Fig. 17-18). The

Page 31: The Effect of Translucency and Background Variations on

22

outcome presented a statistically significant (p < 0.001) difference between the groups within

the same shade.

The Lab* values for descriptive statistics, with means and standard deviation, are reported for

each group and for each of the coordinates. The results showing the mean ΔL* values and the

standard deviation for all the groups are shown in Tables 18-20. The means and standard

deviations calculated for the BL1 shade were described as the following (Mean ± Std. Dev.) N:

group 2 (HT ND4) specimens is (7.49 ± 0.18) N; group 3 (LT ND1) specimens is (2.1 ± 0.17) N;

group 4 (LT ND4) specimens is (3.42 ± 0.15). While A2 shade N; group 6 (HT ND4) specimens

is (4.74 ± 0.33) N; group 7 (LT ND1) specimens is (0.39 ± 0.16) N; group 8 (LT ND4)

specimens is (2.25 ± 0.62). Finally, C3 shade were described as following N; group 10 (HT

ND4) specimens is (4.83 ± 0.23) N; group 11 (LT ND1) specimens is (0.84 ± 0.27) N; group 12

(LT ND4) specimens is (3.62 ± 0.16).

The mean Δa* values and the standard deviation for all the groups are shown in Tables 18-20.

The means and standard deviations calculated for the BL1 shade were described as follows

(Mean ± Std. Dev.) N: group 2 (HT ND4) specimens is (3.2 ± 0.25) N; group 3 (LT ND1)

specimens is (0.33 ± 0.06) N; group 4 (LT ND4) specimens is (2.48 ± 0.14). While A2 shade N;

group 6 (HT ND4) specimens is (4.97 ± 0.17) N; group 7 (LT ND1) specimens is (0.85 ± 0.06)

N; group 8 (LT ND4) specimens is (3.64 ± 0.04). Finally, those of C3 shade were described as

the following N; group 10 (HT ND4) specimens is (4.09 ± 0.18) N; group 11 (LT ND1)

specimens is (0.29 ± 0.14) N; group 12 (LT ND4) specimens is (2.88 ± 0.26).

Page 32: The Effect of Translucency and Background Variations on

23

The results showing the mean Δb* values and the standard deviation for all the groups are shown

in Tables 18-20. The means and standard deviations calculated for the BL1 shade were described

as the following (Mean ± Std. Dev.) N: group 2 (HT ND4) specimens is (7.43 ± 0.39) N; group 3

(LT ND1) specimens is (3.75 ± 0.21) N; group 4 (LT ND4) specimens is (1.35 ± 0.18). While

A2 shade N; group 6 (HT ND4) specimens is (5.39 ± 0.22) N; group 7 (LT ND1) specimens is

(0.75 ± 0.25) N; group 8 (LT ND4) specimens is (1.57 ± 0.53). Finally, those of C3 shade were

described as the following N; group 10 (HT ND4) specimens is (2.17 ± 0.80) N; group 11 (LT

ND1) specimens is (0.83 ± 0.25) N; group 12 (LT ND4) specimens is (0.34 ± 0.22).

Furthermore, non-parametric tests (Mann-Whitney U tests) were carried out in each ceramic

shade group, and the data revealed that there were significant differences between the high

translucent and the low translucent groups (p < 0.001). The HT showed more ΔE than the LT

groups, and these findings were observed in all ceramic shades (Tables 21-23, Fig. 20-22).

Page 33: The Effect of Translucency and Background Variations on

24

Discussion:

In the past few years, lithium disilicate glass ceramic has been used more

often in clinical practice. This material provides optimum esthetics as well as

increased strength, which enables adhesive or conventional cementation6,42. Lithium

disilicate glass ceramic has a needle-like crystal structure that offers excellent

durability and strength, as well as exceptional optical properties4,7,23. Although the

variation in the translucency of the lithium disilicate glass ceramic is considered to be

an advantage, it has a negative effect on the resulting color when the underlying tooth

structure is dark.

Earlier research has demonstrated that the underlying tooth structure has a

major impact on the appearance of definitive ceramic restorations43,44,65,68,69.

However, the influence of different translucencies on the resulting color has not been

tested. Thus, the main focus of this study was to assess the effect of translucency and

background variations on the color difference of different shades of CAD/CAM

lithium disilicate glass ceramic. Translucency and background color were the two

variables that were tested in the study. Translucency was divided into high or low,

while background was divided into ND1 and ND4 among the different ceramic

shades (BL1, A2, or C3 shade). The hypothesis of this study was of two parts: first,

the high translucency CAD/CAM lithium disilicate glass ceramic will have more

color differences than the low translucency ceramic of the same shade; second, the

ND4 group will have more color differences than the ND1 group. The reason behind

the interest in different ceramic shades as conditioned variables in this study was to

define the influence of translucency variations of the same shade and consistency

among the different ceramic shades.

Page 34: The Effect of Translucency and Background Variations on

25

Using the CIELAB system, the color difference (ΔE) was calculated. These

coordinates were taken from a spectrophotometer’s spectral reflectance

measurements. They give a numerical description of the color position in a 3-

dimensional color space70,108. CIELAB units were equally spaced in relation to visual

sensitivity, so that the spectral measurements can be associated with subjective

observations; this is an advantage over the Munsell system (hue, chroma, and

value)70,108.

The results of this study showed that the color difference (ΔE) of a CAD/CAM

lithium disilicate glass ceramic is affected by the level of translucency and the

background color. Background findings of the study were in agreement with earlier

studies in the literature12,17,19,44. High translucent groups and darker background

groups showed increased ΔE values compared with other groups. The highest ΔE

values were noticed in group 2 (BL1 HT ND4), while the lowest ΔE values were

found in both group 7 (A2 LT ND1) and group 11 (C3 LT ND1).

In BL1 groups, the results of this study demonstrated that color difference

(ΔE) interaction was significant among the three groups (Table 15). However, the

most affected group was shown to be group 2 (BL1 HT ND4) and the lowest was

group 3 (BL1 LT ND1) (Table 12). The possible reason for high ΔE values may be

due to the combined effect of high translucency and darker background in the tested

group. Chaiyabtur et al showed a similar finding, as the color of lithium disilicate

glass ceramic was affected by many factors, including the underlying tooth color44.

Additionally, the same finding was observed in the A2 groups and the C3 groups.

However, the effect of translucency and background color was smaller when

compared with BL1 groups (Tables 13, 14). A possible explanation is that an increase

Page 35: The Effect of Translucency and Background Variations on

26

in the chromaticity of the A2 and C3 shades has the potential to reduce the color

effect of the underlying tooth structure.

Based on our current knowledge and literature search, a standard reference of

∆Lab* was not available to facilitate color matching. However, a further analysis of

the Lab* values was performed, with the evaluation of L* values, a* values, and b*

values among tested groups, to achieve a better understanding of the color

differences109.

In this experiment, the ΔL* (lightness) value was greater when HT with an

ND4 background was used among the different ceramic shades (Tables 18-29). This

may be due to the optical properties of the material itself. The ceramic is an optical

mixture of a glass matrix and the lithium disilicate crystalline phase, which allows the

light to pass through a material and reflects the color of the background3,68. The Δa*

and Δb* values, which represent the hue and chroma of color, differed for specimens

against the ND1 compared to the ND4 background, proposing that yellow and red

coloration shows through the ND4 background,43 and this difference was noticed the

most in the BL1 group (Table 18). Furthermore, the reason could be the nature of the

BL1 shade – it possesses less chroma than the other shades. The a* and b* values had

no consistency within the tested groups, and previous studies presented the same

finding110,111. Thus, further studies need to be conducted for a better evaluation.

According to the human perception of color, the color difference (∆E) is

noticed. Fifty percent of people visually notice color difference when ∆E is greater

than 1 ∆E unit58,83. However, under uncontrolled clinical conditions, minor

differences in color may be undetected, because average color differences below 3.7

are believed to be a match in the oral environment88,112. According to Douglas and

Brewer, thresholds for acceptability are at ∆E=1.7 in control environment113. In the

Page 36: The Effect of Translucency and Background Variations on

27

current study, by evaluating ∆E values of LT ND1 groups in A2 and C3 groups, ∆E

was less than 1.7. This indicates that the translucency had a smaller effect than the

ND1 background in the A2 and C3 groups in clinical conditions.

The study data demonstrated that there were significant differences between

the high translucency and the low translucency groups, and this finding was observed

in the three ceramic shades. The HT groups showed more ΔE than the LT groups,

suggesting that the microstructure of HT groups presented bigger lithium disilicate

crystals compared to LT groups. Thus, the light transmission rate is high, which

allows more light to pass through and reflect the background color (Fig. 20-22).

Limitations of the study:

Using one all-ceramic system with one thickness was one of the limitations of this

study. Furthermore, three ceramic shades and two background colors were used in the

study. Thus, more ceramic shades and background color variations need to be

considered. Based on our literature search, only a few studies addressed the optical

properties of lithium disilicate glass ceramic. Additionally, the literature was lacking

in studies that evaluated the effect of translucency. Thus, a comparison of the results

with previous research could not be performed. In addition, based on current

knowledge, a standard Lab* for all shades is not available.

Future study:

Using different backgrounds and utilizing different ceramic fabrication

techniques (i.e., the pressed technique) may provide us with more knowledge and

insight into the influence of translucency and background colors on ceramics. In

addition, measuring the color difference of low translucency A and C shade groups in

different thicknesses could provide valuable information. Future studies should also

consider the tooth form, as the light reflection will differ between the flat surface and

Page 37: The Effect of Translucency and Background Variations on

28

the convex or concave surfaces. The testing different thicknesses and different

opacities is also required. Furthermore, comparing the optical properties of

CAD/CAM lithium disilicate glass ceramic to different all-ceramic systems would be

beneficial in expanding our knowledge in this field.

Page 38: The Effect of Translucency and Background Variations on

29

Conclusions: Within the limitations of this study, the following conclusions can be drawn:

1. The translucency and background color significantly influenced lithium disilicate glass

ceramic color difference among BL1, A2, and C3 ceramic shades.

2. Changing the underlying color from lighter (ND1) to darker (ND4) resulted in increased

color difference (ΔE).

Page 39: The Effect of Translucency and Background Variations on

30

Table 1- Crystallization/Glazing cycle of lithium disilicate glass ceramic furnace (Programat P700).

Firing temperature (F)

Holding time (min)

Vacuum 1 Stage 1 (F) Stage 2 (F)

Vacuum 2 Stage 1 (F) Stage 2 (F)

Long-term cooling (F)

840 7:00 550 1022

820 1508

700

Stand-by temperature (F)

Closing time (min)

Heating rate (F/min)

Firing temperature (F)

Holding Temperature (min)

Heating rate (F/min)

403 6:00 90 820 0:10 30 Table 2- Lab values of reference groups of BL1 group, A2 group, and C3 group.

Reference Group L a b BL1 HT ND1 81.2 (.03) -0.4 (.08) 5.2 (.16) A2 HT ND1 74.3 (.07) 0.1 (.02) 17.1 (.14) C3 HT ND1 69.4 (.1) 1.4 (.3) 21.9(.1)

Table 3- Lab values and ΔE value of BL1 HT ND4 (mean and standard deviation).

BL1 HT ND4 L a b ΔE 1 73.74 3.84 12.94 11.59 2 73.96 3.7 12.43 11.05 3 73.83 3.67 12.02 10.86 4 73.92 3.59 12.11 10.83 5 73.75 3.6 12.98 11.51 6 73.87 3.7 12.82 11.37 7 73.29 3.24 12.21 11.21 8 73.81 3.71 12.94 11.49 9 73.74 3.8 12.92 11.56 10 73.74 3.84 12.94 11.59 Mean (SD) 73.76 (0.18) 3.66 (0.17) 12.63 (0.39) 11.30 (0.29)

Page 40: The Effect of Translucency and Background Variations on

31

Table 4- Lab values and ΔE value of BL1 LT ND1 (mean and standard deviation).

BL1 LT ND1 L a b ΔE 1 79.1 -0.69 8.84 4.21 2 79.28 -0.82 8.57 4.53 3 79.39 -0.82 9.3 4.51 4 79.22 -0.71 9.07 4.37 5 79.21 -0.72 9.01 4.32 6 79.18 -0.66 8.89 4.23 7 79.29 -0.79 8.83 4.13 8 78.81 -0.79 9.11 4.61 9 78.89 -0.63 9.14 4.59 10 79.15 -0.69 8.75 4.12 Mean (SD) 79.15 (0.17) -0.73 (0.06) 8.95 (0.21) 4.36 (0.18) Table 5-Lab values and ΔE value of BL1 LT ND4 (mean and standard deviation).

BL1 LT ND4 L a b ΔE 1 77.88 3.14 6.3 5.02 2 77.82 3.04 6.25 4.98 3 77.97 2.71 6.66 4.75 4 77.88 2.73 6.63 4.82 5 77.94 2.87 6.85 4.94 6 78.09 2.75 6.63 4.69 7 77.75 2.94 6.5 5.01 8 77.55 3 6.43 5.18 9 77.74 2.83 6.66 4.99 10 77.74 2.8 6.6 4.96 Mean (SD) 77.83 (0.15) 2.88 (0.14) 6.55 (0.18) 4.93 (0.14) Table 6- Lab values and ΔE value of A2 HT ND4 (mean and standard deviation).

A2 HT ND4 L a B ΔE 1 69.09 5.3 22.38 8.99 2 69.07 5.26 22.45 9.02 3 69.74 5.43 22.71 8.91 4 69.13 5.16 22.52 8.97 5 69.71 5.31 22.63 8.81 6 69.81 5.23 22.13 8.4 7 69.81 5.06 22.54 8.55 8 69.88 5.04 22.57 8.52 9 69.7 4.83 22.99 8.77 10 69.76 5.05 22.38 8.47 Mean (SD) 69.57 (0.33) 5.16 (0.17) 22.53 (0.22) 8.74 (0.23)

Page 41: The Effect of Translucency and Background Variations on

32

Table 7- Lab values and ΔE value of A2 LT ND1 (mean and standard deviation).

A2 LT ND1 L a b ΔE 1 74.97 0.98 17.67 1.16 2 74.91 0.97 17.55 1.07 3 74.88 1.07 17.63 1.16 4 74.08 0.96 17.9 1.11 5 74.8 1.12 17.79 1.24 6 74.02 1.08 17.84 1.17 7 74.04 1.04 18.18 1.37 8 74.05 1.11 18.21 1.43 9 74 1 18.3 1.45 10 74.06 1.09 17.9 1.2 Mean (SD) 74.38 (0.44) 1.04 (0.06) 17.89 (0.25) 1.23 (0.13)

Table 8- Lab values and ΔE value of A2 LT ND4 (mean and standard deviation).

A2 LT ND4 L a b ΔE 1 72.76 3.86 19.36 4.56 2 72.76 3.82 19.45 4.58 3 72.81 3.82 19.18 4.42 4 71.57 3.82 18.33 4.7 5 71.52 3.91 18.27 4.79 6 71.58 3.84 18.38 4.73 7 71.55 3.85 18.44 4.76 8 71.64 3.86 18.21 4.66 9 71.56 3.88 18.18 4.74 10 72.79 3.73 19.31 4.42 Mean (SD) 72.05 (0.62) 3.83 (0.04) 18.71 (0.53) 4.63 (0.13)

Page 42: The Effect of Translucency and Background Variations on

33

Table 9- Lab values and ΔE value of C3 HT ND4 (mean and standard deviation).

C3 HT ND4 L a b ΔE 1 64.82 5.55 24.72 6.79 2 64.65 5.77 24.81 7.07 3 64.23 5.56 23.21 6.76 4 64.54 5.48 23.47 6.53 5 64.43 5.42 23.86 6.68 6 64.16 5.08 22.51 6.43 7 64.77 5.66 24.49 6.79 8 64.71 5.61 24.53 6.82 9 64.74 5.43 24.8 6.8 10 64.75 5.55 24.65 6.8 Mean (SD) 64.58 (0.23) 5.511 (0.18) 24.10 (0.8) 6.74 (0.17) Table 10- Lab values and ΔE value of C3 LT ND1 (mean and standard deviation).

C3 LT ND1 L a b ΔE 1 68.43 0.99 21.73 1.08 2 68.15 1.18 22.78 1.54 3 68.31 1.47 22.83 1.42 4 68.27 1.07 22.67 1.4 5 68.81 0.97 22.84 1.17 6 68.58 1.28 22.7 1.13 7 68.84 1.04 22.9 1.18 8 68.54 1.3 22.78 1.22 9 68.97 1.06 23.1 1.3 10 68.8 1.02 22.91 1.22 Mean (SD) 68.57 (0.27) 1.13 (0.16) 22.72 (0.36) 1.26 (0.14) Table 11- Lab values and ΔE value of C3 LT ND4 (mean and standard deviation).

C3 LT ND4 L a b ΔE 1 65.75 4.55 22.56 4.86 2 65.74 4.58 22.4 4.88 3 65.82 4.61 22.56 4.85 4 65.81 4.57 22.5 4.83 5 65.99 4.08 21.99 4.34 6 65.94 4.03 22.27 4.37 7 65.59 4.34 21.98 4.82 8 65.77 4.07 21.69 4.52 9 65.95 3.94 22.32 4.31 10 65.46 4.32 21.82 4.91 Mean (SD) 65.78 (0.16) 4.30 (0.26) 22.2 (0.31) 4.66 (0.25)

Page 43: The Effect of Translucency and Background Variations on

34

Table 12- Kruskal-Wallis Test results and descriptive analysis for BL1 tested group (Std. Dev.=Standard Deviation + IQR= Interquartile Range)

Tested Group

ΔE

BL1 Mean Std. Dev. Median IQR P-value (Sig.)

HT ND4 11.30 0.29 11.43 0.56 < 0.001 LT ND1 4.36 0.18 4.34 0.36 < 0.001 LT ND4 4.93 0.14 4.97 0.21 < 0.001 Table 13- Kruskal-Wallis Test results and descriptive analysis for A2 tested group (Std. Dev.=Standard Deviation + IQR= Interquartile Range).

Tested Group

ΔE

A2 Mean Std. Dev. Median IQR P-value (Sig.)

HT ND4 8.74 0.23 8.79 0.47 < 0.001 LT ND1 1.23 0.13 1.18 0.24 < 0.001 LT ND4 4.63 0.13 4.68 0.22 < 0.001 Table 14- Kruskal-Wallis Test results and descriptive analysis for C3 tested group (Std. Dev.=Standard Deviation + IQR= Interquartile Range).

Tested Group

ΔE

C3 Mean Std. Dev. Median IQR P-value (Sig.)

HT ND4 6.74 0.17 6.79 0.16 < 0.001 LT ND1 1.26 0.14 1.22 0.24 < 0.001 LT ND4 4.66 0.25 4.82 0.50 < 0.001

Table 15- Post-hoc test (Mann-Whitney U Test) BL1 groups

BL1 Group Mean Rank (n=10)

P-Value (Sig.)

HT ND4 15.50 < 0.001 LT ND1 5.50 < 0.001 LT ND4 15.50 < 0.001 LT ND1 5.50 < 0.001 HT ND4 15.50 < 0.001 LT ND4 5.50 < 0.001

Page 44: The Effect of Translucency and Background Variations on

35

Table 16- Post-hoc test (Mann-Whitney U Test) A2 groups

A2 Group Mean Rank (n=10)

P-Value (Sig.)

HT ND4 15.50 < 0.001 LT ND1 5.50 < 0.001 LT ND4 15.50 < 0.001 LT ND1 5.50 < 0.001 HT ND4 15.50 < 0.001 LT ND4 5.50 < 0.001

Table 17- Post-hoc test (Mann-Whitney U Test) C3 groups.

C3 Group Mean Rank (n=10)

P-Value (Sig.)

HT ND4 15.50 < 0.001 LT ND1 5.50 < 0.001 LT ND4 15.50 < 0.001 LT ND1 5.50 < 0.001 HT ND4 15.50 < 0.001 LT ND4 5.50 < 0.001

Table 18- Descriptive analysis for BL1 tested group (Std. Dev.=Standard Deviation + IQR= Interquartile Range) Dependent variable ΔL Δa Δb.

BL1 Group Mean Std. Dev. Median IQR ΔL

HT ND4 7.49 0.18 7.48 0.14 LT ND1 2.10 0.17 2.06 0.24 LT ND4 3.42 0.15 3.42 0.21

Δa HT ND4 3.20 0.25 3.28 0.31 LT ND1 0.33 0.06 .31 0.12 LT ND4 2.48 0.14 2.45 0.27

Δb HT ND4 7.43 0.39 7.67 0.76 LT ND1 3.75 0.21 3.75 0.31 LT ND4 1.35 0.18 1.41 0.26

Page 45: The Effect of Translucency and Background Variations on

36

Table 19- Descriptive analysis for A2 tested group (Std. Dev.=Standard Deviation + IQR= Interquartile Range) Dependent variable ΔL Δa Δb.

A2 Group Mean Std. Dev. Median IQR ΔL

HT ND4 4.74 0.33 4.58 0.69 LT ND1 .39 0.16 0.30 0.32 LT ND4 2.25 0.62 2.70 1.21

Δa HT ND4 4.97 0.17 5.0 0.26 LT ND1 0.85 0.06 0.86 0.12 LT ND4 3.64 0.04 3.65 0.04

Δb HT ND4 5.39 0.22 5.39 0.27 LT ND1 0.75 0.25 0.73 0.53 LT ND4 1.57 0.53 1.27 1.07

Table 20 -Descriptive analyses for C3 tested group (Std. Dev.=Standard Deviation + IQR= Interquartile Range) Dependent variable ΔL Δa Δb.

C3 Group Mean Std. Dev. Median IQR ΔL

HT ND4 4.83 0.23 4.73 0.38 LT ND1 0.84 0.27 0.85 0.52 LT ND4 3.62 0.16 3.62 0.24

Δa HT ND4 4.09 0.18 4.13 0.20 LT ND1 0.29 0.14 0.35 0.27 LT ND4 2.88 0.26 2.91 0.51

Δb HT ND4 2.17 0.80 2.58 1.34 LT ND1 0.83 0.25 0.87 0.21 LT ND4 0.34 0.22 0.36 0.49

Table 21- Mann-Whitney U test results for HT and LT of BL1 groups. Dependent variable= ΔE

Tested Group

ΔE

BL1 Mean Rank (n=20)

P-Value (Sig.)

HT 15.50 < 0.001

LT 5.50 < 0.001

Page 46: The Effect of Translucency and Background Variations on

37

Table 22- Mann-Whitney U test results for HT and LT of A2 groups. Dependent variable= ΔE

Tested Group

ΔE

A2 Mean Rank (n=20)

P-Value (Sig.)

HT 15.50 < 0.001

LT 5.50 < 0.001

Table 23- Mann-Whitney U test results for HT and LT of C3 groups. Dependent variable= ΔE

Tested Group

ΔE

C3 Mean Rank (n=20)

P-Value (Sig.)

HT 15.50 < 0.001

LT 5.50 < 0.001

Page 47: The Effect of Translucency and Background Variations on

38

Figure 1- Left: Partially crystallized (HT) CAD/CAM lithium disilicate glass ceramic. Right: Fully crystallized (HT) CAD/CAM lithium disilicate glass ceramic.

Figure 2- Left: Partially crystallized (LT) CAD/CAM lithium disilicate glass ceramic. Right: Fully crystallized (LT) CAD/CAM lithium disilicate glass ceramic.

Figure 3- Munsell color system.

Figure 4- CIE LAB L: Lightness, a: Red and Green, and b: Yellow and Blue.

Page 48: The Effect of Translucency and Background Variations on

39

Figure 5- E4D CAD/CAM machine and pre-crystallized lithium disilicate glass ceramic.

Figure 6- Isomet 1000 machine used to cut the cylinder into disks with 1.2 mm thickness.

Figure 7- BL1 shade groups.

BL1 shade Group

High Translucency

(HT)

HT ND1 (reference) HT ND4

Low Translucency

(LT)

LT ND1 LT ND4

Page 49: The Effect of Translucency and Background Variations on

40

Figure 8- A2 shade groups.

Figure 9- C3 shade groups.

A2 shade Group

High Translucency

(HT)

HT ND1 (reference) HT ND4

Low Translucency

(LT)

LT ND1 LT ND4

C3 shade Group

High Translucency

(HT)

HT ND1 (reference) HT ND4

Low Translucency

(LT)

LT ND1 LT ND4

Page 50: The Effect of Translucency and Background Variations on

41

Figure 10- Left: Pre-crystallized A2 LT lithium disilicate glass ceramic specimen. Middle: Programat P700 furnace. Right: Crystalized A2 LT lithium disilicate glass ceramic specimen.

Figure 11- IPS e.max Ceram Glaze Paste.

Figure 12- Digital caliper (Dentaguage 1) confirming the specimen thickness 1.2mm.

Figure 13- ND1 and ND4 shade IPS Natural die material.

Page 51: The Effect of Translucency and Background Variations on

42

Figure 14 Left: ND4 background with specimen holder. Right: Specimen, ND1 and specimen holder in the dark box.

Figure 15- Spectrophometer (Crystaleye) with custom-positioning jig and dark box.

Figure 16- Left: ΔE* and Lab* of the target (specimen) and the reference (control group) of group 7 (A2 LT ND1). Right: 2 mm in diameter positioned over the middle region of the specimen.

Page 52: The Effect of Translucency and Background Variations on

43

Figure 17- Boxplot illustration of the results of BL1 groups ΔE*.

Figure 18- Boxplot illustration of the results of A2 groups ΔE*.

Figure 19- Boxplot illustration of the results of C3 groups ΔE*.

Page 53: The Effect of Translucency and Background Variations on

44

Figure 20- Mann-Whitney U test results for HT and LT of BL1 groups. Dependent variable= ΔE

Figure 21- Mann-Whitney U test results for HT and LT of A2 groups. Dependent variable= ΔE

Figure 22- Mann-Whitney U test results for HT and LT of C3 groups. Dependent variable= ΔE

Page 54: The Effect of Translucency and Background Variations on

45

Bibliography:

1. Douglas RD, and M Przybylska. Predicting porcelain thickness required for dental shade matches. The Journal Of Prosthetic Dentistry 1999;82(2):143-149.

2. Denry IL. Recent Advances in Ceramics for Dentistry. Critical Reviews in Oral Biology & Medicine. January 1, 1996 1996;7(2):134-143.

3. Holloway JA MR. The effect of core translucency on the aesthetics of all-ceramic restorations. Pract Periodontics Aesthet Dent. 1997;9(5):567-574.

4. McLaren EA CP. Ceramics in Dentistry—Part I: Classes of Materials. Inside Dentistry. 2009(Oct 2009):94-104.

5. Kelly JR. Dental ceramics: current thinking and trends. Dent Clin North Am. Apr 2004;48(2):viii, 513-530.

6. Della Bona A, Kelly JR. The clinical success of all-ceramic restorations. J Am Dent Assoc. Sep 2008;139 Suppl:8S-13S.

7. Holand W, Ritzberger C, Apel E, et al. Formation and crystal growth of needle-like fluoroapatite in functional glass-ceramics. Journal of Materials Chemistry. 2008;18(12):1318-1332.

8. Holand W, Schweiger M, Watzke R, Peschke A, Kappert H. Ceramics as biomaterials for dental restoration. Expert review of medical devices. Nov 2008;5(6):729-745.

9. Joiner A. Tooth colour: a review of the literature. J Dent. 2004;32 Suppl 1:3-12.

10. ten Bosch JJ, Coops JC. Tooth color and reflectance as related to light scattering and enamel hardness. J Dent Res. Jan 1995;74(1):374-380.

11. Watts A, Addy M. Tooth discolouration and staining: a review of the literature. British dental journal. Mar 24 2001;190(6):309-316.

12. The Academy of Prosthodontuics: Glossary of prosthodontic Terms. J Prosthet Dent. 1994;71(1):41-112.

13. Lemire P.A BB. Color in Dentistry JM Ney Co. 1975. 14. Munsell AH, Farnum RB. A Color Notation: An Illustrated System Defining

All Colors and Their Relations 1923 (Large Print Edition): Kessinger Publishing; 1923.

15. Chu SJ, Devigus A, Mieleszko AJ. Fundamentals of color: shade matching and communication in esthetic dentistry: Quintessence Pub. Co.; 2004.

16. McLean JW, Dentistry LSUMCSo. The science and art of dental ceramics: a collection of monographs: Louisiana State University School of Dentistry Continuing Education Program; 1974.

17. O'brien WJ, W. M. Johnston, and F. Fanian. Double-layer Color Effects in Porcelain Systems. Journal of Dental Research 1985;64(6):940.

18. Sorensen SE. Spectrometric Analysis of the Influence of Metal Substrates on the Color of Metal-Ceramic Restorations. Journal of Dental Research 1985;64(1):74.

19. Sorensen JA, and T J Torres. Improved color matching of metal-ceramic restorations. Part III: Innovations in porcelain application. The Journal Of Prosthetic Dentistry 1988;59(1):1-7.

Page 55: The Effect of Translucency and Background Variations on

46

20. Sperling HG, Anthony A. Wright, and Stephen L. Mills. Color vision following intense green light exposure: Data and a model. Vision Research 1991;31(10):1797-1812.

21. Ring ME. Dentistry, an illistrated history. 1985. 22. Jones DW. Development of dental ceramics. An historical perspective.

1985;29(4):621-644. 23. Hench LL, Day DE, Höland W, Rheinberger VM. Glass and Medicine.

International Journal of Applied Glass Science. 2010;1(1):104-117. 24. Lee Y-K, Cha H-S, Ahn J-S. Layered color of all-ceramic core and veneer

ceramics. The Journal of Prosthetic Dentistry. 2007;97(5):279-286. 25. Monsénégo G, Burdairon G, Clerjaud B. Fluorescence of dental porcelain. The

Journal of Prosthetic Dentistry. 1993;69(1):106-113. 26. Conrad HJ, Seong W-J, Pesun IJ. Current ceramic materials and systems with

clinical recommendations: A systematic review. The Journal of Prosthetic Dentistry. 2007;98(5):389-404.

27. Aboushelib MN, de Jager N, Kleverlaan CJ, Feilzer AJ. Microtensile bond strength of different components of core veneered all-ceramic restorations. Dental materials: official publication of the Academy of Dental Materials. Oct 2005;21(10):984-991.

28. Sjogren G, Lantto R, Granberg A, Sundstrom BO, Tillberg A. Clinical examination of leucite-reinforced glass-ceramic crowns (Empress) in general practice: a retrospective study. The International journal of prosthodontics. Mar-Apr 1999;12(2):122-128.

29. Fradeani M, Redemagni M. An 11-year clinical evaluation of leucite-reinforced glass-ceramic crowns: a retrospective study. Quintessence Int. Jul-Aug 2002;33(7):503-510.

30. Marquardt P, Strub JR. Survival rates of IPS empress 2 all-ceramic crowns and fixed partial dentures: results of a 5-year prospective clinical study. Quintessence Int. Apr 2006;37(4):253-259.

31. Otto T, De Nisco S. Computer-aided direct ceramic restorations: a 10-year prospective clinical study of Cerec CAD/CAM inlays and onlays. The International journal of prosthodontics. Mar-Apr 2002;15(2):122-128.

32. Malament KA, Socransky SS. Survival of Dicor glass-ceramic dental restorations over 14 years: Part I. Survival of Dicor complete coverage restorations and effect of internal surface acid etching, tooth position, gender, and age. J Prosthet Dent. Jan 1999;81(1):23-32.

33. Olsson KG, Furst B, Andersson B, Carlsson GE. A long-term retrospective and clinical follow-up study of In-Ceram Alumina FPDs. The International journal of prosthodontics. Mar-Apr 2003;16(2):150-156.

34. Wolfart S, Bohlsen F, Wegner SM, Kern M. A preliminary prospective evaluation of all-ceramic crown-retained and inlay-retained fixed partial dentures. The International journal of prosthodontics. Nov-Dec 2005;18(6):497-505.

35. Freiman SW, Hench LL. Effect of Crystallization on the Mechanical Properties of Li2O-SiO2 Glass-Ceramics. Journal of the American Ceramic Society. 1972;55(2):86-90.

36. Bengisu M, Brow RK, White JE. Interfacial reactions between lithium silicate glass-ceramics and Ni-based superalloys and the effect of heat treatment at elevated temperatures. Journal of Materials Science. 2004;39(2):605-618.

Page 56: The Effect of Translucency and Background Variations on

47

37. Burgner LL, Lucas P, Weinberg MC, Soares Jr PC, Zanotto ED. On the persistence of metastable crystal phases in lithium disilicate glass. Journal of Non-Crystalline Solids. 2000;274(1–3):188-194.

38. Headley TJ, Loehman RE. Crystallization of a Glass-Ceramic by Epitaxial Growth. Journal of the American Ceramic Society. 1984;67(9):620-625.

39. Deubener J. Compositional onset of homogeneous nucleation in (Li, Na) disilicate glasses. Journal of Non-Crystalline Solids. 2000;274(1–3):195-201.

40. Rüssel C, Keding R. A new explanation for the induction period observed during nucleation of lithium disilicate glass. Journal of Non-Crystalline Solids. 2003;328(1–3):174-182.

41. Apel E, Deubener J, Bernard A, et al. Phenomena and mechanisms of crack propagation in glass-ceramics. Journal of the Mechanical Behavior of Biomedical Materials. 2008;1(4):313-325.

42. Denry I, Holloway J. Ceramics for Dental Applications: A Review. Materials. 2010;3(1):351-368.

43. Nakamura T, Saito O, Fuyikawa J, Ishigaki S. Influence of abutment substrate and ceramic thickness on the colour of heat-pressed ceramic crowns. Journal of Oral Rehabilitation. 2002;29(9):805-809.

44. Chaiyabutr Y, Kois JC, Lebeau D, Nunokawa G. Effect of abutment tooth color, cement color, and ceramic thickness on the resulting optical color of a CAD/CAM glass-ceramic lithium disilicate-reinforced crown. J Prosthet Dent. Feb 2011;105(2):83-90.

45. Mutobe Y, Maruyama, T. & Kataoka, S. In harmony with nature: esthetic restoration of a nonvital tooth with IPS-Empress all-ceramic material. Quintessence of Dental Technology. 1997:20, 83. .

46. Hammad IA, Stein RS. A qualitative study for the bond and color of ceramometals. Part II. J Prosthet Dent. Feb 1991;65(2):169-179.

47. Lund PS, Piotrowski TJ. Color changes of porcelain surface colorants resulting from firing. The International journal of prosthodontics. Jan-Feb 1992;5(1):22-27.

48. Rosenstiel SF, Johnston WM. The effects of manipulative variables on the color of ceramic metal restorations. J Prosthet Dent. Sep 1988;60(3):297-303.

49. Jorgenson MW, Goodkind RJ. Spectrophotometric study of five porcelain shades relative to the dimensions of color, porcelain thickness, and repeated firings. J Prosthet Dent. Jul 1979;42(1):96-105.

50. O'Brien WJ, Kay KS, Boenke KM, Groh CL. Sources of color variation on firing porcelain. Dental materials: official publication of the Academy of Dental Materials. Jul 1991;7(3):170-173.

51. Kim IJ, Lee YK, Lim BS, Kim CW. Effect of surface topography on the color of dental porcelain. Journal of Materials Science: Materials in Medicine. 2003;14(5):405-409.

52. Dozić A, Kleverlaan CJ, Meegdes M, van der Zel J, Feilzer AJ. The influence of porcelain layer thickness on the final shade of ceramic restorations. The Journal of Prosthetic Dentistry. 2003;90(6):563-570.

53. Terada Y, Maeyama S, Hirayasu R. The influence of different thicknesses of dentin porcelain on the color reflected from thin opaque porcelain fused to metal. The International journal of prosthodontics. 1989;2(4):352-356.

54. Terada Y, Sakai T, Hirayasu R. The masking ability of an opaque porcelain: a spectrophotometric study. The International journal of prosthodontics. 1989;2(3):259-264.

Page 57: The Effect of Translucency and Background Variations on

48

55. Dagg H, O'Connell B, Claffey N, Byrne D, Gorman C. The influence of some different factors on the accuracy of shade selection. Journal of Oral Rehabilitation. 2004;31(9):900-904.

56. Jacobs SH, Goodacre CJ, Moore BK, Dykema RW. Effect of porcelain thickness and type of metal-ceramic alloy on color. The Journal of Prosthetic Dentistry. 1987;57(2):138-145.

57. Vichi A, Ferrari M, Davidson CL. Influence of ceramic and cement thickness on the masking of various types of opaque posts. Journal of Prosthetic Dentistry. 2000;83(3):412-417.

58. Seghi RR, Johnston WM, O'Brien WJ. Spectrophotometric analysis of color differences between porcelain systems. The Journal of Prosthetic Dentistry. 1986;56(1):35-40.

59. Groh CL, O'Brien WJ, Boenke KM. Differences in color between fired porcelain and shade guides. The International journal of prosthodontics. 1992;5(6):510-514.

60. Barghi N, Pedrero JAF, Bosch RR. Efects of batch variation on shade of dental porcelain. The Journal of Prosthetic Dentistry. 1985;54(5):625-627.

61. Rinke S, Hüls A, Kettler MJ. Colorimetric analysis as a means of quality control for dental ceramic materials. The European journal of prosthodontics and restorative dentistry. 1996;4(3):105-110.

62. O'Neal SJ, Leinfelder KF, Lemons JE, Jamison HC. Effect of metal surfacing on the color characteristics of porcelain veneer. Dental Materials. 1987;3(3):97-101.

63. Barghi N, Richardson JT. A study of various factors influencing shade of bonded porcelain. The Journal of Prosthetic Dentistry. 1978;39(3):282-284.

64. Crispin BJ, Seghi RR, Globe H. Effect of different metal ceramic alloys on the color of opaque and dentin porcelain. The Journal of Prosthetic Dentistry. 1991;65(3):351-356.

65. Koutayas SO, Kakaboura A, Hussein A, Strub JR. Colorimetric evaluation of the influence of five different restorative materials on the color of veneered densely sintered alumina. Journal of Esthetic and Restorative Dentistry. 2003;15(6):353-360.

66. Kourtis SG, Tripodakis AP, Doukoudakis AA. Spectrophotometric evaluation of the optical influence of different metal alloys and porcelains in the metal-ceramic complex. Journal of Prosthetic Dentistry. 2004;92(5):477-485.

67. Stavridakis MM, Dent DM, Papazoglou E, Seghi RR, Johnston WM, Brantley WA. Effect of different high-palladium metal-ceramic alloys on the color of opaque and dentin porcelain. Journal of Prosthetic Dentistry. 2004;92(2):170-178.

68. Chu FCS, Chow TW, Chai J. Contrast ratios and masking ability of three types of ceramic veneers. Journal of Prosthetic Dentistry. 2007;98(5):359-364.

69. Li Q, Yu H, Wang YN. Spectrophotometric evaluation of the optical influence of core build-up composites on all-ceramic materials. Dental Materials. 2009;25(2):158-165.

70. Dozic A, Kleverlaan CJ, Meegdes M, van der Zel J, Feilzer AJ. The influence of porcelain layer thickness on the final shade of ceramic restorations. J Prosthet Dent. Dec 2003;90(6):563-570.

Page 58: The Effect of Translucency and Background Variations on

49

71. Heffernan MJ, Aquilino SA, Diaz-Arnold AM, Haselton DR, Stanford CM, Vargas MA. Relative translucency of six all-ceramic systems. Part I: Core materials. The Journal of Prosthetic Dentistry. 2002;88(1):4-9.

72. Heffernan MJ, Aquilino SA, Diaz-Arnold AM, Haselton DR, Stanford CM, Vargas MA. Relative translucency of six all-ceramic systems. Part II: Core and veneer materials. Journal of Prosthetic Dentistry. 2002;88(1):10-15.

73. Kwiatkowski S, Geller W. A preliminary consideration of the glass-ceramic dowel post and core. The International journal of prosthodontics. 1989;2(1):51-55.

74. Ahmad I. Yttrium-Partially Stabilized Zirconium Dioxide Posts: An Approach to Restoring Coronally Compromised Nonvital Teeth. International Journal of Periodontics and Restorative Dentistry. 1998;18(5):455-465.

75. Raptis NV, Michalakis KX, Hirayama H. Optical behavior of current ceramic systems. International Journal of Periodontics and Restorative Dentistry. 2006;26(1):31-41.

76. Paul S, Peter A, Pietrobon N, Hammerle CH. Visual and spectrophotometric shade analysis of human teeth. J Dent Res. Aug 2002;81(8):578-582.

77. van der Burgt TP, ten Bosch JJ, Borsboom PC, Kortsmit WJ. A comparison of new and conventional methods for quantification of tooth color. J Prosthet Dent. Feb 1990;63(2):155-162.

78. Sorensen JA, Torres TJ. Improved color matching of metal-ceramic restorations. Part I: A systematic method for shade determination. J Prosthet Dent. Aug 1987;58(2):133-139.

79. Sorensen JA, Torres TJ. Improved color matching of metal ceramic restorations. Part II: Procedures for visual communication. J Prosthet Dent. Dec 1987;58(6):669-677.

80. Al-Wahadni A, Ajlouni R, Al-Omari Q, Cobb D, Dawson D. Shade-match perception of porcelain-fused-to-metal restorations: a comparison between dentist and patient. J Am Dent Assoc. Sep 2002;133(9):1220-1225; quiz 1260-1221.

81. McPhee E. Light and color in dentistry. Part I--Nature and perception. The Journal Of The Michigan Dental Association. 1978 60(11):565-572.

82. Billmeyer FW, Saltzman M. Principles of color technology: Wiley; 1981. 83. Seghi RR, Johnston WM, O'Brien WJ. Performance assessment of

colorimetric devices on dental porcelains. J Dent Res. Dec 1989;68(12):1755-1759.

84. Sproull RC. Color matching in dentistry. II. Practical applications of the organization of color. J Prosthet Dent. May 1973;29(5):556-566.

85. Grah CL, O'Brien, W. J., & Boenke, K. M.. Differences in Color Between Fired Porcelain and Shade Guides. International Journal of Prosthodontics. 1992;5(6):510.

86. Wee AG, Monaghan P, Johnston WM. Variation in color between intended matched shade and fabricated shade of dental porcelain. J Prosthet Dent. Jun 2002;87(6):657-666.

87. al SRe. Spectrophotometeric analysis of colour differences between porcelain systems. J prosthet Den 1986(56):35-40.

88. Johnston WM, Kao EC. Assessment of appearance match by visual observation and clinical colorimetry. J Dent Res. May 1989;68(5):819-822.

Page 59: The Effect of Translucency and Background Variations on

50

89. Brewer JD, Glennon JS, Garlapo DA. Spectrophotometric analysis of a nongreening, metal-fusing porcelain. J Prosthet Dent. May 1991;65(5):634-641.

90. Tung FF, Goldstein GR, Jang S, Hittelman E. The repeatability of an intraoral dental colorimeter. J Prosthet Dent. Dec 2002;88(6):585-590.

91. Goldstein GR, Schmitt GW. Repeatability of a specially designed intraoral colorimeter. J Prosthet Dent. Jun 1993;69(6):616-619.

92. O'Brien WJ, Nelson D, Lorey RE. The assessment of Chroma sensitivity to porcelain pigments. J Prosthet Dent. Jan 1983;49(1):63-66.

93. Douglas RD. Precision of in vivo colorimetric assessments of teeth. J Prosthet Dent. May 1997;77(5):464-470.

94. Okubo SR, Kanawati A, Richards MW, Childress S. Evaluation of visual and instrument shade matching. J Prosthet Dent. Dec 1998;80(6):642-648.

95. Russell MD, Gulfraz M, Moss BW. In vivo measurement of colour changes in natural teeth. J Oral Rehabil. Sep 2000;27(9):786-792.

96. Judeh A, Al-Wahadni A. A comparison between conventional visual and spectrophotometric methods for shade selection. Quintessence Int. Oct 2009;40(9):e69-79.

97. Ishikawa-Nagai S, Sato RR, Shiraishi A, Ishibashi K. Using a computer color-matching system in color reproduction of porcelain restorations. Part 3: A newly developed spectrophotometer designed for clinical application. The International journal of prosthodontics. Jan-Feb 1994;7(1):50-55.

98. Da Silva JD, Park SE, Weber HP, Ishikawa-Nagai S. Clinical performance of a newly developed spectrophotometric system on tooth color reproduction. J Prosthet Dent. May 2008;99(5):361-368.

99. Paul SJ, Peter A, Rodoni L, Pietrobon N. Conventional visual vs spectrophotometric shade taking for porcelain-fused-to-metal crowns: a clinical comparison. The International journal of periodontics & restorative dentistry. Jun 2004;24(3):222-231.

100. Derdilopoulou FV, Zantner C, Neumann K, Kielbassa AM. Evaluation of visual and spectrophotometric shade analyses: a clinical comparison of 3758 teeth. The International journal of prosthodontics. Jul-Aug 2007;20(4):414-416.

101. Fani G, Vichi A, Davidson CL. Spectrophotometric and visual shade measurements of human teeth using three shade guides. American journal of dentistry. Jun 2007;20(3):142-146.

102. Maupome G, Pretty IA. A closer look at diagnosis in clinical dental practice: part 4. Effectiveness of nonradiographic diagnostic procedures and devices in dental practice. J Can Dent Assoc. Jul-Aug 2004;70(7):470-474.

103. Kim-Pusateri S, Brewer JD, Davis EL, Wee AG. Reliability and accuracy of four dental shade-matching devices. The Journal of Prosthetic Dentistry. 2009;101(3):193-199.

104. Odaira C, Itoh S, Ishibashi K. Clinical evaluation of a dental color analysis system: the Crystaleye Spectrophotometer(R). Journal of prosthodontic research. Oct 2011;55(4):199-205.

105. Shinomori K, B. E. Schefrin, and J. S. Werner. Age-related changes in wavelength discrimination. OPTICAL SOCIETY OF AMERICA 2001;A(18):310-318.

106. O'Brien WJ, Hemmendinger H, Boenke KM, Linger JB, Groh CL. Color distribution of three regions of extracted human teeth. Dental materials:

Page 60: The Effect of Translucency and Background Variations on

51

official publication of the Academy of Dental Materials. May 1997;13(3):179-185.

107. Azer SS, Rosenstiel SF, Seghi RR, Johnston WM. Effect of substrate shades on the color of ceramic laminate veneers. The Journal of Prosthetic Dentistry. 2011;106(3):179-183.

108. Wyszecki G, Stiles WS. Color science: concepts and methods, quantitative data and formulae: Wiley; 1982.

109. Paniz G, Kim Y, Abualsaud H, Hirayama H. Influence of framework design on the cervical color of metal ceramic crowns. The Journal of Prosthetic Dentistry. 2011;106(5):310-318.

110. Lee Y-K, Yu B, Lim H-N. Lightness, chroma, and hue distributions of a shade guide as measured by a spectroradiometer. The Journal of Prosthetic Dentistry. 2010;104(3):173-181.

111. Tashkandi E. Consistency in color parameters of a commonly used shade guide. The Saudi Dental Journal. 2010;22(1):7-11.

112. Ruyter IE, Nilner K, Möller B. Color stability of dental composite resin materials for crown and bridge veneers. Dental Materials. 1987;3(5):246-251.

113. Douglas RD, Brewer JD. Acceptability of shade differences in metal ceramic crowns. The Journal of Prosthetic Dentistry. 1998;79(3):254-260.