20
THE BIOFIDELITY OF HYBRID III DUMMIES Dimitrios Kall ieris , Andreas Rietti, Rainer Mattern Institute of Legal Medicine University of Heidelberg ABSTRACT The Hybrid I I I dummy is the test device used to ascertain compl iance with the dynamic crash test requirements of FMVSS 208. lnjury criteria derived from measurements at designated regions of the dummy body must be met and therefore the Hybrid I I I must display satisfactory biofidelity . In order to investigate biofidel ity, crash simulations were peormed using human cadavers and Hybrid III dummies. Frontal collisions were simulated at an impact velocity of 50 km/h and a mean car deceleration of between 1 4 and 1 6 g. The surrogates were protected with either a 3- point belt or an air bag - knee bolster system. Two belt systems of different stiffnesses were used, i.e 6% and 1 6% elongation. Assessment of Hybrid III biofidel ity was based on the kinemat ic behaviour and thoracic responses. In the 3-point belt tests, head-neck flexion of the two test subject groups differed in magnitude and phase. Greater flexion was obseed with the cadavers compared to the dummy and maximum Hybrid I I I neck flexion occurr·ed earlier. With both 3-point belts, 6% and 1 6% elongation, the Hybrid III experienced higher acceleration maxima and 3ms values at the sternum ( in x- direction) and at Th1 (resultant). The resultant acceleration at Th1 2 was in agreement with the cadaver. X-direction r ib accelerations were measured at the medium axil lar line of the 8th rib (left and right) in cadavers and 5th rib (left and right) in dumm ies, and d id not show clear differences between the two test subject groups. Thoracic contours and deflections in the horizontal plane were measured with ehest bands. lnter-subject and inter-test variabil ity was observed for thoracic contours and deformations. Chest deflection was generally lower for the dummy than for the cadaver. The results show that in frontal col lision simu lations the biofidelity of the dummy neck and thorax should be improved to allow for more valid assessment of occupant protection criteria. - 135 -

THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

  • Upload
    doanque

  • View
    213

  • Download
    0

Embed Size (px)

Citation preview

Page 1: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

THE B IOFIDELITY OF HYBRID I I I D U M M I ES

Dimitrios Kallieris , Andreas Rizzetti, Rainer Mattern

Institute of Legal Medicine University of Heidelberg

ABSTRACT

The Hybrid I I I dummy is the test device used to ascertain compliance with the dynamic crash test requirements of FMVSS 208. lnjury criteria derived from measurements at designated regions of the dummy body must be met and therefore the Hybrid I I I must display satisfactory biofidelity . In order to investigate biofidelity, crash simulations were performed using human cadavers and Hybrid I I I dummies. Frontal collisions were simulated at an impact velocity of 50 km/h and a mean car deceleration of between 1 4 and 1 6 g. The surrogates were protected with either a 3- point belt or an air bag - knee bolster system. Two belt systems of different stiffnesses were used, i .e 6% and 1 6% elongation. Assessment of Hybrid I I I biofidelity was based on the kinematic behaviour and thoracic responses.

In the 3-point belt tests, head-neck flexion of the two test subject groups differed in magnitude and phase. Greater flexion was observed with the cadavers compared to the dummy and maximum Hybrid I I I neck flexion occurr·ed earlier. With both 3-point belts, 6% and 1 6% elongation, the Hybrid I I I experienced higher acceleration maxima and 3ms values at the sternum (in x- direction) and at Th1 (resultant). The resultant acceleration at Th1 2 was in agreement with the cadaver. X-direction rib accelerations were measured at the medium axillar line of the 8th rib (left and right) in cadavers and 5th rib (left and right) in dummies, and did not show clear differences between the two test subject groups.

Thoracic contours and deflections in the horizontal plane were measured with ehest bands. lnter-subject and inter-test variability was observed for thoracic contours and deformations. Chest deflection was generally lower for the dummy than for the cadaver.

The results show that in frontal collision simulations the biofidelity of the dummy neck and thorax should be improved to allow for more valid assessment of occupant protection criteria.

- 135 -

Page 2: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

D U M M I E S A R E U S E D as test devices to ascertain compliance with dynamic crash test requirements of FMVSS 208 for frontal impact protection and the FMVSS 2 1 4 for side impact protection. In these tests the biofidelity of the prescribed dummies is assumed to be appropriate. At the University of Heidelberg, I nstitute for Legal Medicine, a large number of human cadaver tests have been performed during the last 20 years. The authors have investigated the biofidelity of side impact dummies (Klaus et al 1 984 and Kallieris et al 1 992), and in the 1 9SOs frontal collision simulations using cadavers and Hybrid II dummies were performed to investigate dummy biofidelity. During frontal collisions at an impact velocity of 50 km/h and a car deceleration of 20 g with belt restrained Hybrid I I dummy and cadaver occupants, it was observed that the neck bending angle of the cadaver was clearly greater than the dummy and the maximum angle occurred earlier with the dummy.

In the 1 990s Hybrid I I I dummies have been tested and used in controlled collisions. The results of padded and rigid Hybrid I I I dummy and cadaver head impacts at between 20 and 22 km/h were similar to the Hybrid I I tests. lt was observed that the maximum cadaver head/neck extension was about 30 % greater than Hybrid I I I extension. The maximum bending angle occurred earlier with the dummy than the cadaver. The same observations were made for rear­end collisions at an impact velocity of 25 km/h.

The aim of the paper is to investigate Hybrid I I I dummy biofidel ity, focussing on the.kinematic responses and ehest deflections in identical frontal crash sim ulations.

METHODS

Test subjects

The test subjects were unembalmed human cadavers and Hybrid I I I dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table 1. The cadavers were matched anthropometrically to the Hybrid I I I dummy.

Test Equipment

The tests were performed on the University of Heidelberg's deceleration sied. In the tests the front part of the passenger compartment of a mid-sized car was mounted on the sied. Each subject was positioned in the driver's seat and restrained by a 3-point belt. A belt with an elongation of either 6% or 1 6% was used, furthermore the subject was protected with a driver side air bag-knee

- 136 -

Page 3: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

bolster system. Figura 1 illustrates the experimental configuration. Frontal collisions were simulated with an impact velocity of 48 km/h and a trapezoidal deceleration pulse with an average value of between 1 4 and 1 7 g.

Table 1 : Test - Matrix

Run No. II 8 [years]

Age

Sex

C 1 m 34

04 m

C2 m 29

C3 f 52

05 m

C4 m 31

es m 25

C6 m 38

Instrumentation

Body mass [kg]

75

75

75

71

75

68

68

75

70

74

79

Body length [cm]

1 76

1 76

1 76

1 80

1 76

1 84

1 68

1 76

1 70

1 85

1 74

In each test the subject's thorax was instrumented with a 1 2-accelerometer array

(Robbins et al 1 976, Eppinger et al 1 978). The configurations were:

Cadaver: Upper/lower sternum: x-direction, 4th rib left/right (medium axillar line): y-direction, 8th rib left/right (medium axillar line): x-direction, Th1 : 3-axial, Th1 2: 3-axial.

Dummy: Upper/lower sternum: x-direction, 2nd rib left/right (medium axillar line): y-direction; 5th rib left/right (medium axillar line): x-direction, Thorax c.g.: 3-axial, thoracic vertebral column, level 6th rib: 3-axial.

- 137 -

Page 4: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

The 4th and 8th rib of the eadaver eorrespond to the 2nd and 5th rib of the dummy. To measure thoraeie eontours during dynamie loading two ehest bands were used. The first was mounted at the level of the eadaver's 4th rib (dummy 2nd rib) and the seeond at the level of the eadaver's 8th rib (dummy 5th rib} (Eppinger 1 989).

Fig. 1 : Test eonfiguration

Kinematics

Three h igh-speed eameras (frame rate 1 000 frames/s), two on-board and one stationary, were used to doeument the impaet dynamies.

Autopsy - lnjury Severity

A full autopsy with a detailed investigation of the vertebral eolumn was performed on eaeh eadaver and injuries were eoded aeeording to AIS 1 990.

Data Analysis

The data were reeorded in analog format and were digitised at 1 0000

samples per seeond. Eaeh signal was filtered with a ehannel class 1 80 digital Butterworth Filter. The thoraeie deformation eontours were eomputed using the RBAND-PC Software from NHTSA. A program evaluated the deformation of the front ehest wall in relation to the fixed vertebral eolumn.

To evaluate the Viseous Criterion (VC) at the level of the 4th and 8th rib, the deformation veloeity was defined through differentiation of deformation with respeet to time. The deformation veloeity time-histories were by means of spline interpolation. For the evaluation of ehest eompression, the ehest depth was used (instantaneous deformation/ehest depth, Lau and Viano 1 986).

- 1 38 -

Page 5: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

VC(t) = 2ÖD(t) /8 t* D(t) I D

where D(t): Deformation of the ehest D: ehest depth measured by the ehest band

Biofidelity

Biofidelity is a deseription of the extent to whieh passive human-like meehanieal behaviour is simulated in a meehanieal surrogate. Biofidelity ean be assessed through eomparison of the kinematie and meehanieal responses of a meehanieal surrogate to a human volunteer or eadaver. A valid eomparison ean be made when the test speeimens have the same anthropometry and are subjeet to the same impaet eonditions, e.g. impaet direetion, impaet severity, restraint system and belt geometry. Meehanieal responses ean be eompared most effeetively when the same measurement methods are used for eaeh test speeimen.

RESULTS

A total of 1 1 frontal eollision tests were undertaken using a 50th. pereentile Hybrid I I I dummy and six eadavers. The eadavers were matehed anthropometrieally to the Hybrid I I I dummy. The mean pereentage differenees between the eadavers and dummy for body length and mass were 0.5% and 4.5% respeetively.

Kinematics

Head/Neek Flexion

Figures 2 and 3 show sequential photographs of a belt restrained eadaver and Hybrid I I I dummy under the same impaet eonditions, i.e. impaet veloeity of 48 km/h, mean deeeleration of 1 6 g, and 3-point belt with 6% elongation. The global kinematies of the Hybrid I I I dummy and eadaver are eomparable, however there are temporal and spaeial differenees in head/neek motion. After a translatory movement of about 40 ms duration post-impaet, a head rotation oeeurs eaused by the restraining effeet of the belt. The magnitude of head/neek flexion is greater with the eadaver and maximum flexion oeeurs later than observed with the Hybrid I I I dummy. Comparisons between dummy and eadaver head/neek flexion for tests using the two different belts are shown in figure 4.

- 139 -

Page 6: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

- 140 -

.

cft. <O <tS II c: c: 0 "- · -

Q) ... > CU CU C>

"'C c: <tS ..Q 0 Q)

"'C C> Q) c: c: · -

· - ..c <tS ..c "- Q) Ci) ::: � .=: Q) ..c <tS

. c: 0 ..... ·-0 .� (/) -..c: 0 a. � <tS CU "- ... C> c: 0 0

+"' "-0 .....

..c: C> � <O <tS .,.... +:i "'

c: ..c: Q) .......... :J E c- � Q) ""'

(f) � C\I

. O>

· -

LL

Page 7: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

1 - „

-

Fig

. 3:

Se

qu

ent

ial

ph

oto

gra

ph

s o

f a

be

lte

d r

es

tra

ine

d H

yb

rid

III d

um

my

in

a 4

8 k

m/h

, 16

g fr

on

tal c

olli

sio

n.

Se

it w

eb

bin

g e

lon

ga

tio

n =

6°/o

.

Page 8: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

1 ....

N

1

du

mm

y

ca

da

ve

r

6°/o

elo

ng

ati

on

16°/o

elo

ng

ati

on

du

mm

y

ca

da

ve

r

Fig

. 4:

Sti

ll p

ho

tog

rap

hs

of

the

du

mm

y a

nd

a c

ad

av

er

du

rin

g f

ron

tal

cra

sh

te

st

sh

ow

ing

he

ad

/ n

ec

k f

lex

ion

un

de

r th

e i

nfl

ue

nc

e o

f tw

o d

iffe

ren

t b

elts

(6o/o

an

d 1

6°/o e

lon

ga

tio

n)

Page 9: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

Mechanical Responses

Shoulder Belt Force

Figure 5 shows shoulder belt force time-histories for one cadaver and two Hybrid I I I dummy tests. D ifferences exist between the shape of the time histories for the cadaver and the dummy, however the force duration is comparable between the two subjects. Clear differences exist between the magnitudes of the measured forces. Maximum belt forces of 1 0 kN were measured in the two identical dummy tests, while the maximum shoulder belt force of 7.3 kN in the cadaver test was about 30% less. The cadaver's body mass and length, 71 kg and 1 80 cm respectively, were comparable to the Hybrid I I I anthropometry.

Trunk Accelerations

Acceleration time-histories from the 12 accelerometer thoracic array were evaluated for each test. Comparisons of signals from identical measuring locations for cadaver and dummy were made and found to be partially in agreement. However, they show clear differences for both subjects in signal magnitude, duration and shape. The best agreement between cadaver and dummy thoracic accelerations was at Th1 2. At the remaining measurement locations a greater variability was present.

Examples of acceleration time-histories for the resultant at the Th 1 and the upper sternum in the X-direction are shown in figures 6 and 7. Clear differences were observed according to duration, shape, and acceleration magnitude for both subjects used. Th 1 acceleration maxima for the dummy were 35% to 40% higher than the cadaver (figure 6). Similar observations were made for the measuring location 'upper sternum'. The dummy maxima were 38% to 40% higher than those of the cadaver (figure 7).

Acceleration maxima and 3ms values are presented in table 2. Maxima and 3ms values for dummy and cadaver accelerations show a concurrence within the normal scatter of biodynamic tests, however between the two subjects clear differences exist. Furthermore the acceleration magnitudes depended on the used restraint system. In general: accelerations at Th1 2 were in agreement for dummy and cadaver; clear differences were found at the 8th left and right ribs for air bag-knee bolster tests; and, differences were observed with sternal accelerations during the tests with 3-point belts of either 6% or 1 6% elongation (table 2).

- 143 -

Page 10: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

- g

1 0 k N

9

8

7 ca da ver

6

5

4

3

2

1

0

0 30 60 90 120 1 50 ms

Fig . 5: Shoulder belt force time-histories for the dummy and a cadaver

60

50

40

30

20

1 0

0 0

5 1 .6 g 47.0 g

r...., I r

I I

/ I

30 60

dummy

/ J '-' \.- cadaver "J \ \

\ \ J' \ '\

\ "

90 120 150 ms

Fig. 6: Th1 . acceleration time-histories for the dummy and a cadaver

- 144 -

Page 11: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

g

60

50

40

30

20

1 0

- 1 0

30.8 g

ca da ver

5 1 ·8 g dummy ,.___ 49.5 g

i ' ! . 1 I I . , 1 , , 1 , , . 1 1 1 1 1 i

1

Fig. 7: Acceleration time-histories measured at the upper sternum for the dummy and a cadaver

- 145 -

Page 12: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

Table 2: Thorax acceleration measurements of the dummy and cadaver

Run

No.

01

02

03

C1

04

C2

C3

05

C4

C5

C6

Impact Dece- up. stern. low. stern. 8. rib right 8. rib left res. TH1 res. TH1 2

velocity leration max. 3 ms max. 3 ms max. 3 ms max. 3 ms max. 3 ms max.

[km'h] [g] [g] [g] [g] [g] [g] [g] [g] [g] [g] [g] [g]

47 1 6 50 40 55 48 44 39 55 52 43 41 47

48 1 6 45 40 50 48 42 39 63 60 45 38 ---

48 1 7 49 44 53 46 54 45 49 47 53 49 55

48 1 6 31 24 40 26 40 34 32 23 31 28 50

48 1 4 56 49 60 51 42 40 61 59 40 39 50

49 1 5 49 30 39 35 32 25 47 45 48 37 32

48 1 4 40 33 35 24 42 37 57 54 35 29 49

48 1 6 52 52 64 47 49 47 56 5 1 57 51 46

47 1 6 33 32 52 48 29 28 34 31 38 36 58

49 1 7 54 5 1 44 38 35 33 42 40 56 5 1 49

47 1 7 53 50 74 65 28 27 37 35 48 46 52

Chest Deformation and Viscous Criterion (VC)

Figures 8 and 9 show time-histories for: the evaluated deformation, deformation velocity and VC. For both subjects the presented time-histories were derived from the ehest band positioned at the level of the 4th rib during tests with a 3-point belt of 1 6% elongation. The duration of the deformation amounted to about 1 60 ms for both the cadaver and dummy. However, the shape, magnitude and time between impact and deformation maximum were different. A maximum deformation of 6 cm was reached for the cadaver 73 ms post-impact, compared to a maximum deformation of 3.3 cm at 47 ms with the dummy. Maximum deformation velocities were observed in the same time region as the deformation maxima and were of similar magnitudes, 2. 1 4 m/s and 2.32 m/s for the cadaver and dummy respectively. The deformation velocity time-history profile and the time at which the maxima were reached differ. The VC value of the cadaver was more than double that of the dummy.

- 146 -

3 ms [g]

43

---

52

46

50

30

47

44

49

46

50

Page 13: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

D U [ CM ) 2 0 1--...._-;,;:.;--:::..:::-------- --------------------------

- 2 - 4

- 6

-- - - - - - ---- - --- - - - - - - - - - - - - - - - - - -- - - - - - - - - - - - - - - - - - - - - - - --- - - - - - - - - - - - - - - - -

-3.34i cm i i !

- 8 L.-.-��.......-.......-,....,....i!,...........,r-r--T.......,."'"T"""r..,.....,.."T""T"",-r"-r-T"".,...,-r-r-i"""T!-.--r"""T""T(""T,...,.....,..,.-r-r-"r"T"""T"' 0 2 0 4 0 6 0 8 0 1 0 0 1 2 0 1 6 0 1 8 0 2 0 0 T [ M S ]

V D E F [ M/S ] �����������������������-----, 3 2

0 - 1 - 2 - 3

vc [ M/S ] 0 . 6 0 . 4 0 . 2 0 . 0

- 0 . 2 - 0 . 4

- 0 . 6

- - - - - - - - - - - - - - - - - -- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

-2.32 m/s

0 2 0 4 0 6 0 8 0 1 0 0 1 2 0 1 4 0 1 E O 1 8 0 T [ MS l

0.2 m/s

0 2 0 4 0 6 0 8 0 1 0 0 1 2 0 1 4 0 1 6 0 1 8 0 r [ MS ]

Fig. 8: Chest deformation, deformation velocity and VC time-histories measured from a belt restrained

(1 6°/o) Hybrid I I I dummy at the level of the 2nd. rib.

- 147 -

2 0 0

2 0 0

Page 14: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

1 O E F ( C M )

2

0

- 2

- 4

- 6

- 8

0 2 0

V D E F ( M/ S )

3 2

0

- l

! ! - - - - - - - - - - - - - - - - -�- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

-5.99, cm

4 0 6 0 8 0 1 0 0

T ( M S ]

1 2 0 1 4 0 1 6 0 1 8 0 2 0 0

- 2 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

v c ( M / S )

0 . 6

0 . 4 0 2 0 0

- 0 . 2

- 0 . 4

-0 2 0

-2. 1 4 m/s

4 0 6 0

0.46 m/s

8 0 1 0 0 1 2 0 1 4 0 1 6 0 1 8 0 2 0 0

T ( M S )

- 0 . 6

L,--r---r-r--r-r--r-r...,....,...-.-.--i--r--r-T"".,.....,.....,.-,-�.--.-.-r-i....,-,r-T""""1-r-T-r-r-r-r...,..........,.......-r-r"T""T"-i--r--i--r-T-r' 0 2 0 4 0 6 0 8 0 l 0 0 1 2 0 1 4 0 1 6 0 1 8 0

T [ M S )

Fig . 9 : Chest deformation, deformation velocity and VC time-histories measured from a belt restrained

( 1 6o/o) cadaver at the level of the 4th. rib.

- 148 -

2 0 0

Page 15: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

Figures 1 0 and 1 1 show the same time-histories as figures 8 and 9, but at the level of the 8th rib. The deformation durations for both subjects were shorter and the deformation time-histories had different profiles. The cadaver shows higher deformation velocity values and much higher VC values, 2.2m/s and 0.3 m/s respectively, in comparison to 1 .44 m/s and 0. 1 2 m/s respectively for the dummy.

Thoracic deformation velocities and VC are presented in table 3. According to the restraint system used, deformations between 2 cm and 7 cm for the cadaver were observed via the ehest band method. The dummy usually showed smaller deformation and VC values than the cadaver, while the deformation velocity values are comparable. With air bag tests some cadaver values were higher than in belt restraint tests.

Table 3: Chest deformation values - VC values for dummy and cadaver

Run deformation [cm] defo. velocity VC [m/s] defo

[m/s] standard

No. upper lower upper lower upper lower [cm]

01 --- --- --- --- --- --- 2.2

02 . --- --- --- --- --- --- 2.8

03 4.7 5.0 2.9 2.2 0.25 0.26 ---

C1 5.4 2.7 --- --- --- --- ---

04 3.3 3.2 2.3 1 .4 0.20 0. 1 2 ---

C2 --- 3.8 --- 1 .4 --- 0. 1 4 ---

C3 6 4.8 2.2 2.2 0.46 0.30 ---

05 --- 2.0 --- --- --- --- 1 .3

C4 2.0 3,2 --- --- --- --- ---

es 2.1 3.8 0.8 1 .8 0.05 0.21 ---

C6 3.5 7.2 3.7 9.4 0.74 1 .5 ---

- 149 -

Page 16: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

D E F [ CM ) 2

0

- 2

- 4

- 6

- 8

0 2 0

V D E F ( M/S ) 3 2

o -:i ---- 1

- 2

i 1 - - - - - - - - - - - - - - - - -r- - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

1 1 • I �,..,..__�: """'=- -:-::':--==-::':----- - - - - -- - - - - - - - - - ----- - - - -- ------- - --3. 1 7 pm

4 0 6 0

- 1 .44 m/s

i 1

8 0 1 0 0

T [ M S ] 1 2 0 1 4 0 1 6 0 1 8 0 2 0 0

- 3 '-T-r-.-r--r-r--.-.-,...-,-,...-,-...-.-,..-,.-..-r-ir-r-r--r-i--r-r.....-r-r-r-i--r-,...-,--r-r-r-r-r-r-ir-r-ir-r-r--r-i--r-r-r-r� 0 2 0 4 0 6 0 8 0 1 0 0 1 2 0 1 4 0 1 6 0 1 8 0 2 0 0

T ( M S ) v c ( M/ S ] ,.---������������������������

0 . 6

0 . 4

-0. 1 2 m/s 0 . 2

0 . 0

--------------�===----�====�--------------------------- 0 . 2

- 0 . 4

- 0 . 6

0 2 0 4 0 6 0 8 0 1 0 0

T ( M S ] 1 2 0 1 4 0 1 6 0 1 8 0 2 0 0

Fig. 1 O : Chest deformation, deformation velocity and

VC time-histories measured from a belt restrained ( 1 6°/o) Hybrid I I I dummy at the level of the 5th. rib.

- 1 50 -

Page 17: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

O E F ( C M ) 2 0

- 2

- 4

- 6

�---- _ _ _ _ _ _ _ _ __ _ _ 1 _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

1 l i j 1 • !

-4.8 11 cm

1 - 8 '-T--.-....--r-�"T"""T""""T"'""r"'""T"'""r"'"......,-....,-r-r-ir1--r-t-r-r-,-,-..,.......�"T"""T"""....-r-.,-.,-.,-.,-....,-r-r-i,-....,,....,-,.-,-,.-r-T�

V O E F [ M/ S ) 3 2

0

- 1 - 2 - 3

v c [ M/ S ] 0 . 6

0 . 4

0 . 2

0 . 0

- 0 . 2

- 0 . 4

- 0 . 6

0

0

0

2 0 4 0 6 0 8 0

-2. 1 8 m/s

2 0 4 0 6 0 8 0

· -0.30 m/s

2 0 4 0 6 0 8 0

1 0 0 1 2 0

T [ M S ]

1 0 0 1 2 0

T [ M S ]

1 0 0 1 2 0

T [ M S )

1 4 0 1 6 0

1 4 0 1 6 0

1 4 0 1 6 0

1 8 0 2 0 0

1 8 0 2 0 0

1 8 0 2 0 0

Fig. 1 1 : Chest deformation, deformation velocity and

VC time-histories measured from a belt restrained ( 1 6°/o) cadaver at the level of the 8th. rib.

- 151 -

Page 18: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

Medical findings

No rib fractures were found when the driver air bag - knee bolster restraint system was used (subjects' age 25 to 38 years). In tests with the 3-point belt restraint ( 1 6% elongation) sternum and rib fractures (AIS 2) were observed in two cadavers (age: 29 and 52 years). In one case (age: 34 years) no rib fractures were found when a 3-point belt with 6 % elongation was used. Except for one laceration of the ligamentur.n flavum at Th 1!Th2 (C3, AIS 2), no vertebral column injuries were observed in the investigated cadavers.

DISCUSSION

Frontal collisions with restrained Hybrid I I I dummy and cadavers under identical crash conditions and restraint systems were performed. The kinematic behaviour of both test subjects compare well globally, however cadaver head-neck flexion is greater. A possible reason for the greater neck flexibil ity observed in cadavers is that the upper thoracic vertebral column is included in the bending process. This may also account for injuries observed at the transition zone between the thoracic and cervical spines. The authors observed previously that the Hybrid I I I dummy neck stiffness was greater in comparison to a cadaver's during rigid and padded frontal head impacts at an impact speed of 20 and 22 km/h. This was also observed in 25 km/h rear-end collisions.

Measured thoracic accelerations demonstrated a similar behaviour. While the acceleration magnitudes and profiles over time at the measuring location Th 1 2 were in agreement between the two subjects, greater differences were found for accelerations measured at the level of Th1 and sternum. In these cases accelerations of the dummy's thorax were higher than those of the cadaver, which suggests that the Hybrid III trunk and trunk articulations may be stiffer than the cadaver.

Shoulder belt forces depend upon the body mass and tissue deformability in frontal collisions. Clear differences between the dummy and cadavers were observed even for anthropometrically matched pairs. Higher belt forces were observed for the dummy, as body masses were similar this also indicates that the dummy thorax is stiffer than in a cadaver.

Finally, thoracic deformations were measured using ehest bands. They were greater with the cadaver's thorax, and in anthropometrically matched pairs twice that of Hybrid I I I deformation. This observation was also made by Cesari et al. ( 1 990, 1 994) and Backaitis et al. ( 1 986) who used cadavers and

- 152 -

Page 19: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

volunteers, but loaded the subject's thorax in a different configuration than the authors.

With regards to ehest biofidelity, Horsch et al. ( 1 99 1 ) observed that Hybrid I I I it was satisfactory in compression during blunt sternal impacts, but 'stiff' during belt loading. lndeed Hybrid I I I ehest calibration is performed via a blunt frontal impact. The biofidelity of the Hybrid I I I for blunt sternal impacts was not confirmed in the presented driver air bag tests. In agreement with Cesari et al. ( 1 990), it was observed that the VC values of the cadaver were twice as high as those calculated from the dummy's responses.

CONCLUSIONS

1 ) The global kinematic behaviour of the Hybrid I I I dummy in frontal collisions restrained with a 3-point belt is comparable with those of the cadaver. However, the magnitude of cadaver head-neck frontal flexion is clearly greater.

2) By observing the acceleration and deformation measurements it appears that the dummy's thorax is stiffer than the cadaver.

REFERENCES

Backaitis S, St-Laurent A ( 1 986), 'Chest Deflection Characteristics of Volunteers and Hybrid I I I Dummies', SAE Paper No.861 884, Proceedings of the

30th. Stapp Gar Crash Conference, p. 157-1 66

Cesari D, Bouquet R ( 1 990), 'Behaviour of Human Surrogates Thorax under Seit Loading', SAE Paper No. 90231 0, Proceedings of the 34th. Stapp Gar

Crash Conference, p. 73-81

Cesari D, Bouquet R ( 1 994), 'Comparison of Hybrid I I I and Human Cadaver Thorax Deformations Loaded by a Thoracic Seit', SAE Paper No. 942209, Proceedings of the 38th. Stapp Gar Crash Conference, p. 65-76

Eppinger RH, Augustyn K, Robbins DH (1 978), 'Development of a Promising Universal Thoracic Trauma Prediction Methodology', SAE Paper No. 780891 , Proceedings of the 22th. Stapp Gar Crash Conference, p. 209-268

Eppinger RH, ( 1 989), 'On the Development of a Deformation Measurement System and lts Application Toward Developing Mechanically Based lnjury Indices', SAE Paper No. 892426, Proceedings of the 33th. Stapp Gar Crash

Conference, p 2 1 -28

- 1 53 -

Page 20: THE BIOFIDELITY OF HYBRID III DUMMIES - IRCOBI · dummies. Six cadaver and 5 dummy tests were performed. The most important anthropometric data of the test subjects are shown in table

Harsch JD, Melvin JW, Viano DC, Mertz HJ ( 1 991 ), 'Thoracic lnjury Assessment of Belt Restraint Systems Based on Hybrid I I I Chest Compression', SAE Paper No. 912895, Proceedings of the 35th. Stapp Gar Crash Conference,

p. 85- 1 08

Kallieris D , Mattem R, Mclntosh A, Boggasch F, (1 992), 'The Biofidelity of Eurosid 1 and Biosid', SAE Paper No. 92251 8, Proceedings of the 36th. Stapp

Gar Crash Conference, p. 1 03- 1 1 7

Klaus G , Sinnhuber R, Hoffmann G , Kallieris D, Mattem R, ( 1 984), 'Side Impact- A Comparison Between Dummies and Cadavers . Correlations Between Cadaver Loads and lnjury Severity', SAE Paper No.841 655, Proceedings of the 28th. Stapp Gar Crash Conference, p. 237-259

Lau 1 , Viano D,(1 986), 'The Viscous Criterion-Bases and Application of an lnjury Severity Index for Soft Tissue', SAE Paper No. 861 882, Proceedings of

the 30th. Stapp Gar Crash Conference, p. 1 23- 142

Robbins OH, Melvin JW, Stalnaker AL, (1 976), 'The Prediction of Thoracic Impact lnjuries', SAE Paper No. 760822, Proceedings of the 20th. Stapp Gar

Crash Conference, p. 699-729

- 154 -