Hot Air Oven Sterilizer YCO-010
Horizontal Steam Sterilizer AutoclaveTC Serial
USERS MANUAL
OPERATION / MAINTENANCE
AND PARTS LIST
! Important Information
This manual contains important operation and safety information.
The users must carefully read and understand the contents of this manual prior to use of this equipment in conformance with your local electrical code.
260701.500Horizontal Steam Sterilizer AutoclaveTC SerialUSERS MANUAL
OPERATION / MAINTENANCE
AND PARTS LIST
CONTENT
Safety Information
..1
Alert Boxes
..1
Warnings
..1
Intended use
..1
Technical specifications
..2
Installation
..3
Warning (Protection Recommendation)..3
Preparation
..4
Operation
..5
Control box
..6
Caution and Warning
..7
Maintenance & servicing
..8
Warning
.8
Monthly check list Form..
9-1 / 9-2
Troubleshooting
.10
Direction to Clean, transportation, Storage etc. 11
Diagram of Components
.12
Parts List
.13
Wiring diagram (220V, 380V, 3-phase) ..14
Wiring diagram (220V, 3-phase, automatic) ..15
Wiring diagram (Manual drying, 3-phase) ..16
Wiring diagram (Manual drying, 1-phase) ..17
Safety Information
Your GEMMY Autoclave TC serial has been designed with function, reliability, and safety in mind. It is your responsibility to install it in conformance with your local electrical codes. For your safety operation, please pay attention to the alert boxes throughout the manual.
Alert Boxes
WARNING
Warnings alert you to a possibility of personal injury.
CAUTION
Cautions alert you to a possibility of damage to the equipment
NOTE
Notes alert you to pertinent facts and conditions.
WARNINGSTo avoid electrical shock, always
1 Use a properly grounded electrical outlet of correct voltage and current handling capacity.
2 Disconnect from power supply before maintenance and servicing. Refer to servicing to qualified personnel.
3 In the event that fluid is accidentally spilled into the bath, disconnect grounded plug and turn equipment upside down to avoid fluid contacting the internal components.
To avoid personal injury
1 Ensure that this equipment is connected to electrical service according to local and national electrical codes. Failure to properly connect may create a fire or shock hazard.
2 Always use the gloves and handle to remove the substances out of the chamber
3 WARNINGHOT CHAMBER. The equipment will become hot and remain hot inside the chamber without visual indication for some time after use. Direct body contact is strictly prohibited. 1
INTENDED USEThis Autoclave provides an efficient performance for high pressure steam sterilization of liquid or hard goods. Features present cycle to sterilize media, dressing, instruments, liquid and labware either wrapped or unwrapped & other material that needs sterilization either in clinics or hospitals. Those substances that can produce the chemical toxicity or can become deformed through high pressure and high temperature are prohibited for operation.
Please refer to proper operation method established by technology for products applications or qualified engineers.
Technical specifications
Catalog Number TC-615 TC-612 TC-600 TC-500 TC-459 TC-409 TC-406 TC-339
Chamber (Cm.) 60 x 150 60 x 120 60 x 100 50 x 100 45 x 90 40 x 90 40 x 66 33 x 90
Capacity (Liters) 424 L 339 L 283 L. 196 L. 143 L. 113 L. 83 L. 77 L.
Power consumption 18 KW 18 KW 12 KW 10 KW 7 KW 6 KW 6 KW 5 KW
Ampere
220V 1P 3P 1P 3P 1P 3P 1P 3P 1P 3P 1P 3P 1P 3P 1P 3P 1P 3P
A, A. 82 47 82 47 55 32 45 26 32 18 27 16 23 13 19 10
Working temp 121 132C at 1.2 2.0 kg/cm (1.18 1.96 bar or 17.07 28.45 psi)
Power source 220V or 380V with 1 phase or 3 phase (please specify when ordering)
TC-615, TC-612, TC-500 3-phase only.
Standard accessory Stainless steel tray x 1 pc
Auto water filling Optional
Automatic drying Optional
Vacuum drying Optional ( or called Pre-vacuum)
Safety door lock
Activated by pressure Optional
Safety devices Double walled construction insulated by glass fiber; safety valve; Steam Trap;
Electronic low water level detector; over-heat alarm and shutout
MOTE
Model TC-615 and TC-612 are rated in 18KW or 15 KW. Please specify when ordering.
AMPERE conversion formula
1-phase 3-phase
Wattage / Voltage / 1.732 = Amp for 3-phase
2INSTALLATION
Always use lift truck to deliver this autoclave.
Remove nails and steel packing strips by the proper tools. Remove starting from the top piece firstly, then the surrounding sides. Always handle with care.
WARNING: Heavy Weight, Handle With Care.
Use the lifter to deliver autoclave after removing the packaging material.
Place the equipment onto a flat and sturdy place leaving at least 5 10cm away from the wall.
Do not tilt this equipment. It will deform the legs of autoclave probably.
Do not place the equipment in a draft, sunlight or near a place of equipment, which emits heat as well as electromagnetic conduction emission.
Recycle the packaging materials.
Check the electrical specification label located on back of this equipment. Make sure the power specifications must conform to your local electrical code.
Plug the power cord into a properly grounded outlet.
To avoid virus infection, isolate this autoclave from other equipment.
WARNING
This equipment contains glass fiber insulation, which can produce respiratory fibers and dust when handled. These fibers or dusts can cause irritation and can aggravate pre-existing respiratory disease. The insulating materials are located between the inner chamber and outer body. There is no significant risk of exposure to dust or respirable fibers resulting from operation of this equipment under normal conditions. However, there may be a risk of exposure to respirable dusts or fibers when repairing or maintaining the insulating material, or when otherwise disturbing the material in a manner, which causes release of dust of fibers there from. Use sufficient natural or mechanical ventilation to keep dust level to below PEL/TLV/WEG (workplace exposure guideline) use dust collection when tearing out.
Protection Recommendation:
Use or wear the following when servicing:
Goggles
Gloves
Respirator, filter
Ventilation
Others that you can imagine to keep apart form dust.
3PREPARATION1. Recycle the packaging materials, but do not place them around the equipment.
2. Check the specification label, and make sure that the electrical data of the control box must be conformed to your local code or not.
3. This autoclave is the HI-Voltage equipment. For safety sake, plug power cord into the household like outlet is strictly prohibited.
4. User is requested to prepare a no-fuse breaker. The breaker specifications must be equal to higher than the rating current, ampere and voltage of autoclave. Connect the power cord with this no-fuse breaker.
When the power cord is 1-phase/3-wire
Connect 2 terminals with breaker, disregard + or . Let the third wire be grounding as depicted figure B. When the power source is 3-phase/4-wire
Connect the 3-terminal with breaker, disregard + or . Let the 4th wire be grounding as depicted figure A.
5. For functions of automatic Water Filling and Vacuum Drying, inspect the motor spinning direction, it must be clockwise direction. If not, exchange assembly of any 2 terminals in breaker, then the motor will work in the correct direction.
6. Adjust the level of autoclave by adjusting its 4 feet. Vibration or unstable is strictly prohibited.
7. Prepare the distilled water or soft water for sterilization. Hard water or tap water will create the scale, which will clog the piping system and also damage the heating element and sensor, so as to reduce the operation lifetime.
8. Clean the operation zone, set and divide into 2 zones Sterilized and Non-sterilized. It can prevent infection by this way.
9. The qualified or licensed operator is required for operation.
10. Repeat and inspect the above steps, once the above-regulated steps are proved excellent, this autoclave can start to be tested and operated.
4-wire Grounding
3-wire Grounding
A
B
4
OPERATION
A) Water Filling
1 Open the Water Filling valve ( # 7)
2 Open the Exhaust Valve ( # 5)
3 Fill water into the water cup ( # 6)
4 Fill water up to the level where the red line is marked
This procedure is not required if this unit is functioning Automatic water filling ..
B) Sterilizing substances
5 Unlock the door Remove scale or dirt inside the chamber if any.
6 Place the substances, which are being sterilized, onto the tray of chamber
7 Close door, lock water fill valve and exhaust valve.
C) Set Time
8 Sterilization Time
TC-61560 minutes (including TC-612, TC-600)
TC-50060 minutes
TC-45945 minutes
9 Drying Time10 ~15 minutes For automatic Drying Function only.
Another 20 minutes more is required for textile, clothing and fabrics substances.
D) Heating
10 Switch power to ON position by pressing the key START.
11 It starts to heat up in the exterior chamber up to pressure 1.8kg/cm12 The pressure will be delivered to the interior chamber automatically at 1.8kg/cm.
13 The timer will start to work when the pressure reaches 1.8kg/cm.
5
E) Exhaust
14 When the Set time of sterilization has elapsed, the buzzer will alert you.
15 The steam will be exhausted automatically from the interior chamber activated by the electrical solenoid valve.
16 The door can be opened only until the pressure of interior chamber has returned to ZERO ( 0 ).
F) Drying
17 Manual Drying function When the pressure has returned to ZERO ( 0 ),
Open the Dry Valve ( # 18) and also open door by just 10 15cm space.
Set time again for 10 15 minutes (or 20 minutes more for textile fabric, clothing)
Completion when the set time has elapsed, and shut out power supply automatically.
18 Automatic Drying Function
When the pressure has returned to ZERO ( 0 ), the drying time will start to count automatically.
Once the Drying set time has passed, the buzzer will alert you automatically, and then shut out the power supply automatically.
Completion
19 Inspect the pressure gauge, it must return to 0, then open the door and remove the sterilized substance from the chamber.
CONTROL BOX
Automatic Drying Manual Drying
Power Lamp Power LampHeating Lamp Sterilization Lamp
SterilizationCompletion Steam switch Water Level
Outer to inner chamber Indicator Lamp
Low water level Sterilization TIMER Indicator Switch
Sterilization
Timer (drying)
6CAUTION and WARNNING
This autoclave operates its pressure within range of 1.2kg/cm 1.8kg/cm which has been calibrated in house before delivery. Unless otherwise the qualified or licensed engineers, users own adjustment or calibration is strictly prohibited. GEMMY does not ensure the responsibility if user violates the above regulation.
Sterilization Pressure1.2kg/cm 1.8kg/cmSterilization temperature 121C 132C
Inspect and make sure that the water level must be set properly.
WARNINGThe equipment will become hot and remain hot without visual and audible indication for some time after use, hand or body touch is strictly prohibited. Proper use of this equipment is required to avoid physical harm. Improper use can cause personal injury and/or tissue damage. This equipment should not be used unless the operator has been instructed by a physician or other qualified health care professional.
WARNING Do not open the door until the pressure returned to ZERO (0).
Before use of this autoclave, make sure that the pressure must start from ZERO, otherwise, we will not ensure the responsibility for any possible damage and personal injury.
Be sure to disconnect from power supply and refer to the licensed engineers or qualified personnel when installing this equipment.
Inspect the door gasket always and replace the new one when needed.
Inspect the filter always and change the new one when needed.
Refer to the check list to test the PRESSURE and SAFETY VALVE monthly or any time when it is required.
This autoclave has been well inspected and tested in house before shipment. User is requested not to remove parts or recalibrate the parameter of this equipment under surveillance by the licensed engineers.
7
MAINTENANCE & SERVICING
NOTE
To avoid the electrical shock, always disconnect from power supply prior to maintenance and servicing.
People other than the qualified personnel or licensed engineers are strictly prohibited from maintenance & servicing.
Always inspect the inner chamber and clean the dirt and any residue.
Inspect and test the SAFETY VALVE periodically
Pull the Safety Valve up to make sure if the steam can be released, and record in the check list. If not, it indicates the problem of the valve, replace the new one.
Drain the water out of the chamber monthly as stated in the OPERATION. Distilled water or soft water is recommended.
Always keep the autoclave from the wall by 20cm more.
WARNING
HOT CHAMBER. The chamber and the exposed piping and components will become hot and remain hot without visual indication for some time after use. Direct contact will create personnel injury. Always use the handle and gloves to remove the hot substances from chamber to avoid personnel injury.
Do not open the door only until the chamber pressure has returned to ZERO.
WARNING
This equipment contains glass fiber insulation, which can produce respiratory fibers and dust when handled. These fibers or dusts can cause irritation and can aggravate pre-existing respiratory disease. The insulating materials are located between the inner chamber and outer body. There is no significant risk of exposure to dust or respirable fibers resulting from operation of this equipment under normal conditions. However, there may be a risk of exposure to respirable dusts or fibers when repairing or maintaining the insulating material, or when otherwise disturbing the material in a manner, which causes release of dust of fibers there from. Use sufficient natural or mechanical ventilation to keep dust level to below PEL/TLV/WEG (workplace exposure guideline) use dust collection when tearing out.
The pressure has been set in house, inspect the control box and make sure if the pressure is fallen in the normal position. Do not recalibrate the pressure without the licensed engineers.
8
MONTHLY CHECK LIST FORM MODEL Date of Inspection Document No.
Inspection objects Inspection items & methods Inspection result
BODY Housing 1Is housing & its corresponding clamp parts deformed, corroded or cracked?
2Is insulated material come off, peel off or fractured?
3Is the surface rusty?
4If the joints of piping line cracked or leaked or loose?
Grounding 1If grounding parts corroded, damaged or cracked?
2If the fastener loose?
Door & Fastener Door panel &
Flange If Door panel & fixed fittings deformed, corroded, loose, scaled or damaged?
Fastener If door gasket & fasteners loose, damaged, come off or abraded?
Piping & ValvePipes, flange1If pipes abraded, damaged or deformed?
2If junction damaged, cracked or creviced?
3If door packing, fasteners damaged or corroded?
Valves 1If valve box, base & bar deformed smooth, deformed or corroded?
2If valve bar positioned or bent?
3If flange packing leaked, corroded?
4If valve rotation normally?
Accessories 1Pressure Gauge1Is scale accurate & clear?
2If the corresponding pipes damaged, deformed or corroded?
2Temperature gauge 1Is scale accurate & clear?
2If the corresponding pipes damaged, deformed or corroded?
3safety valve1If valve damaged, deformed, cracked or leaked?
2If the stop valve stays opened?
3Can it pull up smoothly?
4Automatic alarm systemIf its signal and set value normal?
5Emergency stop valve 1If operation normal?
2If deformed, corroded, cracked?
6Emergency power source1If operation normal?
2if deformed, corroded, cracked?
Leg 1Structure 1If structure loose or shambled?
2If welding cracked or loose?
3If fasteners loose, cracked?
2Basic installation If any crack, corrosion or cracks?
3leg & tableIs it stable or shambled?
Page 1
9-1
Defects finding, analyze the possible cause
Assess the hazard (severity & possibility)
Assessment result & improvement measures
Review the applicability of improvement measures
Remarks
1Inspection Resultmarks OK V Fair NG X Fair and NG results need to be improved and corrected completely.
2Inspection according to your local law and regulation.
3Keep and maintain the inspection record for at least 3 years.
Comments and sign from the related departments
Chief of Safety &
Sanitary Department Safety & Sanitary
Department
Chief of operation
Department Inspector
License Number of Inspector (engineer)
9-2
TROUBLESHOOTING
Problem Possible Corrective action
Door cant be unlocked Pressure activated Cant open after completion
Lock cant be positioned Open when pressure back to zero
Open the dry valve and try again
Hit the door control bar slightly.
Defect of water supply- Water filling cup- Water filling valve
- Drain valve- Dirt on water level sensor - Check if cup if clogged. - Close valve and open again. - Replace the new one
- Clean the dirt and scale on it.
Pressure exceeds 2.2kg
and continue rising.
Overheating temperature - Safety valve - Pressure control
- Pressure gauge
- Test safety valve to release steam - Check if the setup pressure falls
into the normal area or not.
- Check if the gauge is working normally or not.
Temperature cannot riseSteam trap is the major part to create the overheating temp. If the temp cannot rise continuously, do not dismantle the part. Use a screwdriver to hit this part slightly to remedy the problem. Otherwise, call a licensed engineer or vendor for the necessary action.
Cant sterilize completely Heater short circuit
Pressure cant fall into 1.2-1.8 kg/cm2, or temperature less than 120C
Cloggy sterilization valve or dry valve -
Use the distilled water or pure water or replace the electric solenoid valve.
Others Please refer to us.
CAUTION Troubleshooting is strictly limited to the qualified personnel and licensed engineers.
Disconnect from power supply prior to troubleshooting.
Troubleshooting can be performed only until the temperature is chamber comes back to the ambient value.
10Direction to Clean
Disconnect from power supply prior to cleaning.
Use a moist cleaning cloth to clean the exterior case and interior chamber
Dry the equipment thoroughly
Do not immerse the equipment in water
Wear gloves when clean the equipment.
Remove shelves and immerse in water and clean completely. Dry shelves thoroughly then place onto chamber.
Never use benzine or paint thinner for cleaning.
Transportation
Fragile, Heavy Weight, Handle With Care
Use no Hook
Do not drop the equipment
Do not place the equipment upside down.
Do not vibrate and bump equipment during transportation.
Storage
Do not place equipment in a draft, sunlight or near a piece of equipment, which emits heat as well as electromagnetic conduction emission.
Switch power to OFF position after use of this equipment
Do not place or store equipment where it can fall easily.
This equipment shall be stored under the condition of room temperature and humidity.
Conditions for Transportation & Storage
Temperature
10C 40C
Humidity
40 % 90%
11
Parts List
Code Description Manual drying Automatic drying Automatic drying
(Safety Pressure activated door lock)
01 Pressure gauge, inner chamber
02 Pressure gauge, outer chamber
03 Safety valve, pressure release
04 Sterilization valve X X
Solenoid valve, sterilization X
05 Steam Exhaust Valve
06 Cup, water Filling
07 Water Filling Valve
08 Water Level Tubing
09 Sensor, water level
10 Water Drainage Valve
11 Inlet, steam exhaust
12 Inlet, water drainage
13 Inlet, trap
14 Heater
15 Temperature gauge, inner chamber
16 Y type Filter
17 Trap
18 Steam Exhaust Valve, inner chamber
19 Stainless Steel Tray
20 Door Gasket
21 Front Door
22 Handlers
23 Bar, door lock
24 Base, hinges
25 Hinges
26 Legs
27 Feet
28 Arm Yoke
29 Solenoid, Drying X
30 Safety Door Lock, pressure X X
31 Control Bar, safety door lock X X
32 Auto Water Filling System X Option Option
33 Vacuum Drying System X Option Option
13
One Year Limited Warranty
Your GEMMY product is guaranteed to be free from defects in materials and workmanship for one ( 1 ) year under normal use from the date of purchase.
This WARRNATY does not apply to any product damaged by accident, misuse, abuse, neglect, improper line voltage, drop, fire, flood or if the products were altered or repaired by anyone other than the qualified service personnel.
The liability of Gemmy Industrial Corp. is limited to repair or replacement and under no circumstances shall GEMMY be liable for any collateral consequential damages or loss. This guarantee specifically excludes the expendables and consumables.
All warranty claims must be directed to the distributors or agents authorized by Gemmy Industrial Corp. responsible for the sale of this equipment. The users are responsible for shipping expense.
Please print
Company Name
Address
Phone No.
Fax No.
E mail
Date of purchase
Product model
Serial No.
Distributor
Manufacturer
GEMMY EUROPE
ISO9001
Gemmy Industrial Corp.rue de Rotterdam
ISO13485
Taipei, Taiwan
No. 23, Boite
GMP
4000 LIEGE
CE
BELGIQUE
30A
( (
30A AMP.
( ( (
Water FillingSterilizing substancesSet Time
Heating upSterilizationSteam Exhaust
DryingCompletion
ON
OFF
Timer