Upload
sabariah-othman
View
214
Download
0
Embed Size (px)
Citation preview
8/2/2019 STKK2723 - GC
1/51
Some important points to remember!1.) Typical response obtained by chromatography (i.e., a chromatogram):
Where:tR= retention timetM= void timeWb = baseline width of the peak in time units
Wh = half-height width of the peak in time units
8/2/2019 STKK2723 - GC
2/51
2.) Solute Retention:
A solutes retention time or retention volume in chromatography
is directly related to the strength of the solutes interaction withthe mobile and stationary phases.
Retention on a given column pertain to the particulars of thatsystem:
- size of the column- flow rate of the mobile phase
8/2/2019 STKK2723 - GC
3/51
3.) Efficiency:
Efficiency is related experimentally to a solutes peak width.- an efficient system will produce narrow peaks
- narrow peaks
smaller difference in interactions in order to separate twosolutesEfficiency is related theoretically to the various kinetic processes that areinvolved in solute retention and transport in the column
- determine the width or standard deviation (s) of peaks
Wh
Estimate s from peak widths,assuming Gaussian shaped peak:
Wb = 4s
Wh = 2.354s
Dependent on the amount of time that a solute spends in the column (k or tR)
8/2/2019 STKK2723 - GC
4/51
Number of theoretical plates (N): compare efficiencies of a system forsolutes that have different retention times
N = (tR/s
)2
or for a Gaussian shaped peak
N = 16 (tR/Wb)2
N = 5.54 (tR/Wh)2
The larger the value of N is for a column, the better the column will beable to separate two compounds.
- the better the ability to resolve solutes that have small differencesin retention
- N is independent of solute retention- N is dependent on the length of the column
8/2/2019 STKK2723 - GC
5/51
Plate height or height equivalent of a theoretical plate (H or HETP): compareefficiencies of
columns with different lengths:H = L/N
where: L = column lengthN = number of theoretical plates for the column
Note: H simply gives the length of the column that corresponds to one theoreticalplate
H can be also used to relate various chromatographic parameters (e.g., flow rate,particle size, etc.) to the kinetic processes that give rise to peak broadening:
Why Do Bands Spread?a. Eddy diffusionb. Mobile phase mass transfer
c. Stagnant mobile phase mass transferd. Stationary phase mass transfer
e. Longitudinal diffusion
8/2/2019 STKK2723 - GC
6/51
Van Deemter equation: relates flow-rate or linear velocity to H:
H = A + B/m + Cm
where:
m = linear velocity (flow-rate x Vm/L)H = total plate height of the columnA = constant representing eddy diffusion &
mobile phase mass transferB = constant representing longitudinal
diffusionC = constant representing stagnant mobilephase & stationary phase masstransfer
One use of plate height (H) is to relate these kinetic process toband broadening to a parameter of the chromatographic system(e.g., flow-rate).
This relationship is used to predict what the resulting effect wouldbe of varying this parameter on the overall efficiencyof the chromatographic system.
8/2/2019 STKK2723 - GC
7/51
m optimum
Plot of van Deemter equation shows how H changes with the linear velocity (flow-rate) of the mobile phase
Optimum linear velocity (mopt) - where H has a minimum value and thepoint of maximum column efficiency.
mopt is easy to achieve for gas chromatography, but is usually too small forliquid chromatography requiring flow-rates higher than optimal to separate
compounds
8/2/2019 STKK2723 - GC
8/51
4.) Measures of Solute Separation:
separation factor (a) parameter used to describe how well two solutes areseparated by a chromatographic system:
a = k2/k1 k = (tRtM)/tMwhere:
k1 = the capacity factor of the first solutek2 = the capacity factor of the second solute,
with k2~ k
A value of a ~ 1.1 is usually indicative of a good separation
Does not consider the effect of column efficiency or peak widths, only retention.
8/2/2019 STKK2723 - GC
9/51
Resolution (RS) resolution between two peaks is a second measure of howwell two peaks are separated:
RS =
where:tr1, Wb1 = retention time and baseline width for the first eluting peaktr2, Wb2 = retention time and baseline width for the second eluting peak
tr2 tr1
(Wb2 + Wb1)/2
Rs is preferred over a sinceboth retention (tr) and columnefficiency (Wb) are consideredin defining peak separation.
Rs ~1.5 represents baselineresolution, or completeseparation of two neighboringsolutes ideal case.
Rs ~1.0 considered adequatefor most separations.
8/2/2019 STKK2723 - GC
10/51
Gas ChromatographyA.) Introduction:
Gas Chromatography (GC) is a chromatographic technique in whichthe mobile phase is a gas.
GC is currently one of the most popular methods for separating and
analyzing compounds. This is due to its high resolution, low limits ofdetection, speed, accuracy and reproducibility.
GC can be applied to the separation of any compound that is eithernaturally volatile (i.e., readily goes into the gas phase) or can be
converted to a volatile derivative. This makes GC useful in theseparation of a number of small organic and inorganic compounds.
8/2/2019 STKK2723 - GC
11/51
Gas Chromatography
B.) Equipment:
A simple GC system consists of:1. Gas source (with pressure and flow regulators)
2. Injector or sample application system3. Chromatographic column (with oven for temperature
control)4. Detector & computer or recorder
8/2/2019 STKK2723 - GC
12/51
A typical GC system used is shown below (a gas chromatograph)
Carrier gas: He (common), N2, H2Pinlet 10-50 psigFlow = 25-150 mL/min packed columnFlow = 1-25 mL/min open tubular column
Column: 2-50 m coiled stainless steel/glass/TeflonOven: 0-400 C ~ average boiling point of sample
Accurate to
8/2/2019 STKK2723 - GC
13/51
Schematic Diagram of Gas Chromatograph
8/2/2019 STKK2723 - GC
14/51
C.) Mobile Phase:
GC separates solutes based on their different interactions with themobile and stationary phases.
-solutes retention is determined mostly by its vapor pressureand volatility
- solutes retention is controlled by its interaction with thestationary phase
- gas mobile phase has much lower density
decreased chance for interacting with soluteincreased chance that solid or liquid stationary phase
interacts with solute
8/2/2019 STKK2723 - GC
15/51
Carrier gas main purpose of the gas in GC is to move the solutesalong the column, mobile phase is often referred to as carriergas.
Common carrier gas: include He, Ar, H2, N2
8/2/2019 STKK2723 - GC
16/51
Carrier Gas or Mobile phase does not affect solute retention, butdoes affect:
1.) Desired efficiency for the GC System- low molecular weight gases (He, H2) largerdiffusion coefficients
- low molecular weight gases faster, more efficientseparations2.) Stability of column and solutes
- H2 or O2 can react with functional groups on solutesand stationary phase or with surfaces of the injector,connections and detector
3.) Response of the detector- thermal conductor requires H2 or He- other detectors require specific carrier gas
8/2/2019 STKK2723 - GC
17/51
D.) Stationary Phases:
Stationary phase in GC is the main factor determining theselectivity and retentionof solutes.
There are three types of stationary phases used in GC:Solid adsorbentsLiquids coated on solid supportsBonded-phase supports
8/2/2019 STKK2723 - GC
18/51
1.) Gas-solid chromatography (GSC)
- same material is used as both the stationary phase and support material- common adsorbents include: alumina molecular sieves (crystalline
aluminosilicates [zeolites] and clay)silicaactive carbon
Magnified Pores in activated carbon
8/2/2019 STKK2723 - GC
19/51
Gas-solid chromatography (GSC):
advantages:- long column lifetimes- ability to retain and separate some compounds not easily
resolved by other GC methods geometrical isomers permanent gases
disadvantage:- very strong retention of low volatility or polar solutes- catalytic changes that can occur on GSC supports- GSC supports have a range of chemical and physical
environments different strength retention sites non-symmetrical peaks variable retention times
8/2/2019 STKK2723 - GC
20/51
2.) Gas-liquid chromatography (GLC)
- stationary phase is some liquid coated on a solid support
- over 400 liquid stationary phases available for GLC many stationary phases are very similar in terms of
their retention properties
- material range from polymers (polysiloxanes, polyesters,polyethylene glycols) to fluorocarbons, molten salts and liquidcrystals
8/2/2019 STKK2723 - GC
21/51
Based on polarity, of the 400 phases available only 6-12 are needed formost separations. The routinely recommended phases are listed below:
Name
Chemical nature of
polysiloxane
Max.
temp.
McReynolds constants
x y z m sSE-30 Dimethyl 350 14 53 44 64 41
Dexsil300 Carborane-dimethyl 450 43 64 111 151 101
OV-17 50% Phenyl methyl 375 119 158 162 243 202
OV-210 50% Trifluoropropyl 270 146 238 358 468 310
OV-225 25% Cyanopropyl-
25% phenyl
250 238 369 338 492 386
Silar-SCP 50% Cyanopropyl-
50% phenyl
275 319 495 446 637 531
SP-2340 75% Cyanopropyl 275 520 757 659 942 804
OV-275 Dicyanoallyl 250 629 872 763 1106 849
McReynolds constants based on retention of 5 standard probe analytesBenzene, n-butanol, 2-pentanone, nitropropanone, pyridine
Higher the
numberthe highertheabsorption.
8/2/2019 STKK2723 - GC
22/51
Preparing a stationary phase for GLC:
- slurry of the desired liquid phase and solvent is madewith a solid support
solid support is usually diatomaceous
earth (fossilized shells of ancient aquaticalgae (diatoms), silica-based material)
- solvent is evaporated off, coating the liquid stationaryphase on the support
- the resulting material is then packed into the column
8/2/2019 STKK2723 - GC
23/51
disadvantage:- liquid may slowly bleedoff with time
especially if high temperatures are used contribute to background change characteristics of the column with
time
8/2/2019 STKK2723 - GC
24/51
3.) Bonded-Phase Gas chromatography
- covalently attach stationary phase to the solid supportmaterial
- avoids column bleeding in GLC- bonded phases are prepared by reacting the desiredphase with the surface of a silica-based support
reactions form an Si-O-Si bond between the stationary phaseand support
or
reactions form an Si-C-C-Si bond between the stationary phaseand support
8/2/2019 STKK2723 - GC
25/51
- many bonded phases exist, but most separations can be formed with thefollowing commonly recommended bonded-phases:
Dimethylpolysiloxane Methyl(phenyl)polysiloxane
Polyethylene glycol (Carbowax 20M) Trifluoropropylpolysiloxane Cyanopropylpolysiloxane
advantages:- more stable than coated liquid phases- can be placed on support with thinner and more uniform
thickness than liquid phases
Si
CH3
CH3
O
n
Si
CH3
CH3
O
n
Si
C6H5
C6H5
O
m
C CHO O
H
H
H
H
H
n
8/2/2019 STKK2723 - GC
26/51
E.) Support Material:
There are two main types of supports used in GC:Packed columns
large sample capacity preparative work
Capillary (open-tubular) columns higher efficiency smaller sample size
analytical applications
8/2/2019 STKK2723 - GC
27/51
8/2/2019 STKK2723 - GC
28/51
F.) Elution Methods:A common problem to all chromatographic techniques is that in any onesample there may be many solutes present, each retained by thecolumn to a different degree:
Best separation and limitsof detection are usuallyobtained with solutes withk values of 2-10
Difficult to find onecondition that elutes allsolutes in this k rangegeneral elution problem
Gradient elution - change column condition with time which changes
retention of solutes to overcome general elution problem
8/2/2019 STKK2723 - GC
29/51
Temperature Programming changing the temperature on the column with time tosimulate gradient elution in GC since a solutes retention in GC is related to its
volatility.
ISOTHERMALColumn temp. 120oC
Programmed temp.(30oC to 180oC) (5o/min)
Comparison of a GC separation using isothermal conditions and temperatureprogramming is shown below
Temperature programming is usually done either in a stepwise change, a linear change or acombination of several linear changes. A single linear change or ramp is the most common
8/2/2019 STKK2723 - GC
30/51
G.) GC Detectors:
The following devices are common types of GC detectors:
1. Thermal Conductivity Detector (TCD)2. Flame Ionization Detector (FID)3. Nitrogen-phosphorus Detector
4. Electron Capture Detector (ECD)5. Mass Spectrometers (discussed later in the course)
The choice of detector will depend on the analyte andhow the GC method is being used (i.e., analytical or
preparative scale)
8/2/2019 STKK2723 - GC
31/51
1.) Thermal Conductivity Detector (TCD)
- katherometer or hot-wire detector- first universal detector developed for GC
Process
- measures a bulk property of the mobile phase leaving thecolumn.
- measures ability to conduct heat away from a hot-wire (i.e.,thermal conductivity)
- thermal conductivity changes with presence of othercomponents in the mobile phase
8/2/2019 STKK2723 - GC
32/51
Design- based on electronic circuit known as a Wheatstone bridge.- circuit consists of an arrangement of four resistors with a fixed current applied to
them.- thermal conductivity changes with presence of other components in the mobile
phase.- the voltage between points (+) and (-) will be zero as long as the resistances in the
different arms of the circuit are properly balanced
as solute emerge from column:change in thermal conductivity change in amount of heat removed fromresistor change in resistors temperature and resistance change in voltagedifference between points (+) and (-).
-one resistor in contact withmobile phase leavingcolumn
-another in contact withreference stream of puremobile phase
8/2/2019 STKK2723 - GC
33/51
Considerations
- mobile phase must have very different thermal conductivitythen solutes being separated.
- most compounds separated in GC have thermal conductivityof about 1-4X10-5.
- H2 and He are carrier gases with significantly differentthermal conductivity values.- H2 reacts with metal oxides present on the resistors, so not
used
8/2/2019 STKK2723 - GC
34/51
advantages:- truly universal detector
applicable to the detection of any compound in GC- non-destructive
useful for detecting compounds frompreparative-scale columns
useful in combination with other types of GC
detectors
disadvantage:- detect mobile phase impurities- sensitive to changes in flow-rates
- limit of detection ~ 10-7 M
much higher then other GC detectors
8/2/2019 STKK2723 - GC
35/51
2.) Flame Ionization Detector (FID)
- most common type of GC detector- universal detector capable of measuring the presence of almost any
organic and many inorganic compound
Process- measures the production of ions when
a solute is burned in a flame.
- ions are collected at an electrode tocreate a current
8/2/2019 STKK2723 - GC
36/51
advantages:
- universal detector for organicsdoesnt respond to common inorganic
compounds
- mobile phase impurities not detected- carrier gases not detected- limit of detection: FID is 1000x better than TCD- linear and dynamic range better than TCD
disadvantage:
- destructive detector
8/2/2019 STKK2723 - GC
37/51
3.) Nitrogen-Phosphorus Detector (NPD)
- used for detecting nitrogen- or phosphorus containing compounds
- also known as alkali flame ionization detector or thermionic detector
Process- same basic principal as FID- measures production of ions when a solute
is burned in a flame- ions are collected at an electrode to
create a current- contains a small amount of alkali metal
vapor in the flame- enhances the formation of ions from
nitrogen- and phosphorus- containing compounds
Alkali Bead
8/2/2019 STKK2723 - GC
38/51
3.) Nitrogen-Phosphorus Detector (NPD)
advantages:- useful for environmental testing
detection of organophosphate pesticides- useful for drug analysis
determination of amine-containing or basicdrugs
- Like FID, does not detect common mobile phase impurities
or carrier gases- limit of detection: NPD is 500x better than FID in detecting
nitrogen and phosphorus- containing compounds- NPD more sensitive to other heterocompounds, such assulfur-, halogen-, and arsenic- containing molecules
disadvantage:- destructive detector- NPD is less sensitive to organic compounds compared to
FID
4 ) El C D (ECD)
8/2/2019 STKK2723 - GC
39/51
4.) Electron Capture Detector (ECD)
- radiation-based detector- selective for compounds containing
electronegative atoms, such as halogens
Process- based on the capture of electrons byelectronegative atoms in a molecule
- electrons are produced by ionization of thecarrier gas with a radioactive source3H or 63Ni
- in absence of solute, steady stream ofthese electrons is produced
- electrons go to collector electrode where
they produce a current- compounds with electronegative atoms
capture electrons, reducing current
8/2/2019 STKK2723 - GC
40/51
advantages:
- useful for environmental testingdetection of chlorinated pesticides or herbicidesdetection of polynuclear aromatic carcinogens
detection of organometallic compounds
- selective for halogen- (I, Br, Cl, F), nitro-, and sulfur-containing compounds
- detects polynuclear aromatic compounds, anhydridesand conjugated carbonyl compounds
8/2/2019 STKK2723 - GC
41/51
Gas Chromatography Application
8/2/2019 STKK2723 - GC
42/51
Gas Chromatography can only be used to
analyze volatile components (i.e. compoundsthat can be volatilized at 250oC without
decomposition).
Non-volatile components have to bederivatized before GC analysis.
Retention Times
8/2/2019 STKK2723 - GC
43/51
Response
GC Retention Time on SE-30
Unknown compound
RT= 4 min on SE-30
Response
GC Retention Time on SE-30
Hexane
RT= 4.0 min on SE-30
Retention Times
8/2/2019 STKK2723 - GC
44/51
TENTATIVE IDENTIFICATION OF UNKNOWN COMPOUNDSResponse
GC Retention Time on Carbowax-20 (min)
Mixture of known compounds
Hexane
OctaneDecane
1.6 min = RT
Response
Unknown compound may be Hexane
1.6 min = RT
Retention Time on Carbowax-20 (min)
SEMI QUANTITATIVE ANALYSIS OF FATTY ACIDS
8/2/2019 STKK2723 - GC
45/51
SEMI- QUANTITATIVE ANALYSIS OF FATTY ACIDS
C
C
C
Detector Response
Retention Time
14
16
18
Peak Area (cm )
Sample Concentration (mg/ml)
2
4
6
8
10
0.5 1.0 1.5 2.0 2.5 3.0
2
The content % of C fatty acids =C
C + C + C
= the content % of C fatty acids14
14
8/2/2019 STKK2723 - GC
46/51
GLC ADVANTAGES
1. Very good separation
2. Time (analysis is short)
3. Small sample is needed - ml4. Good detection system
5. Can be used for both qualitative
and quantitative analysis
8/2/2019 STKK2723 - GC
47/51
DISADVANTAGES OF GASCHROMATOGRAPHY
Material has to be volatilized at 250oCwithout decomposition. If not the analyte hasto be derivatized.
e.g. Fatty acids Methyl esters
Phenols Silanized
Sugars Silanized
8/2/2019 STKK2723 - GC
48/51
R C OH CH3OH H2SO4
O
R C O CH3
O
CH2 O C R
CH O C R
CH2 O C R
O
O
O
CH3OH
O
R C O CH3
CH3ONa
Fatty Acids Methylester
Reflux
+ 3
Volatile in Gas
Chromatography
Volatile in Gas
Chromatography
+ +
Gas Chromatogram of Methyl Esters of Fatty
8/2/2019 STKK2723 - GC
49/51
Gas Chromatogram of Methyl Esters of Fatty
Acids
Derivation of Glucose with Trimethylchlorosilan
8/2/2019 STKK2723 - GC
50/51
OH
O
OH
OHHO
CH2OH
1
23
45
6
+ Si
CH3
CH3
CH3
5Cl
O-Si(CH3)3
O
O-Si(CH3)3
O-Si(CH3)3(CH3)3-Si-O
CH2O-Si(CH3)3
1
23
45
6
5HCl+
Derivation of Glucose with Trimethylchlorosilan
Glucose Trimethylchlorosilane
8/2/2019 STKK2723 - GC
51/51
Effects of Derivation
1. Time consumption
2. Side reaction
3. Loss of sample