Upload
lael
View
21
Download
0
Embed Size (px)
DESCRIPTION
Some considerations on Cryogenic system for 9-cell SC cavities test at IHEP. Accelerator research center Shaopeng Li The Third IHEP-KEK 1.3GHz SC Technology Collaboration Meeting Dec. 7-8, 2010, IHEP, Beijing. contents :. Horizontal test stand of BEPCII 500MHz SC cavities - PowerPoint PPT Presentation
Citation preview
Some considerations on Cryogenic system for 9-cell SC
cavities test at IHEP
Accelerator research centerShaopeng Li
The Third IHEP-KEK 1.3GHz SC Technology Collaboration Meeting
Dec. 7-8, 2010, IHEP, Beijing
contents : Horizontal test stand of BEPCII 500MHz
SC cavities Some considerations on cryogenic
system for 2K 9-cell cavities test at IHEP
Layout of BEPCII cryogenic system
Cryogenic hall and tank farm
Facilities
He compressor systems, VFDs, air compressor, air drier, ORS, cooling-water system, spare power engine, gas tank farm, control system
Tank farm Tank farm
Two 130m3 ,1.45Mpa Two 130m3 ,0.73Mpa One 30m3 ,0.08Mpa One 5m3 ,6Mpa One 10m3 ,0.73Mpa
SRF Cavity(On line)
SRF Cavity test stand
Multi-channel transfer line
SRF quench line
Waveguide
LHe dewar SRF valve box
500W refrigerator
LN2 transfer line
BEPCII second colliding hall
heater
UPS
PLC rack
Horizontal test stand for BEPCII 500MHz SC cavities
Horizontal
test stand
500MHz
SC cavityOperation position
Some considerations on 9-cell SC cavities test
The feature of vertical test of 9-cell SC cavity• Short term test, low power test
• Open cycle and Close cycle system• Open cycle :normal test cryostat, but recovery and purify system ;• Close cycle: special test cryostat with tight leakage
• Depressurized pumping system ;• Off-line or on-line liquifier/refrigerator
The feature of horizontal test of 9-cell SC cavity• Continuous and long term running
• Close cycle system, high power test
• Depressurized pumping system ;• On-line liquifier/refrigerator
Flow diagram of Vertical test
Pure storage tank
Pure storage tank
80K purifier
LHe mobile D
ewar Vertical test stand
pump Gas bag Recovery compressor
Impure storage tank
Pure storage tank
Liquifier/refrigerator Vertical test stand
pump
Open cycle
Close cycle
Liquifier Storage dewar
Helium liquified plant
Flow diagram of vertical/horizontal test
Recovery&purity system
Pumping unit Heat exchangeHeater
2K ValveBox
4K Valve Box
2000LDewar
500W4.5K
refrigeratorMain Comp.
Horizontal test stand
Vertical teststand
Normal methods to 2K
Case 1 : Directly to depressurize saturated vapor pressure of helium
Case 2 : Depressurization + JT ; Case 3 : Depressurization +JT + Hex ;
Case 1 : Directly to depressurize Directly to depressurize the saturated va
por pressure of LHe in test cryostat by pumping system and then the low temperature can be produced
Vertical testcryostat
Pressure ( KPa ) 24.04 5.33 3.13 1.64
Temperatuer( K )
3 2.2 2 1.8
Pumping system
Total heat loss Q=100W
1.8K ,16mbar
During the SC cavities test, the LHe of test cryostat will be consumed and LHe level will be lowered ;
The heat loss of [email protected] will consume 1.08L/h
To assume total heat load 100W , the mass flow rate will be LHe 4.37g/s and LHe consumption is 108L/h
Case 2 : Depressurization + JT The LHe throttled directly to test dewar from storage dewar ; Isoenthalpy thrrotle h1=h2 =h2Gx+h2L(1-x) , h1 is enthalpy
value at 4.2K ; h2G and h2L are enthalpy at 1.8K LHe and GHe ; x is helium fluid quality ;
Energy equation : mh2+Q=mh4 , h4 is enthalpy at 1.8K GHe , m is mass flow rate
The produce of liquid helium is about 61% from 4.2K to 1.8K
If total heat loss is 100W , the total mass flow rate of helium is around 7g/s
3
Verticaltest dewar
Toatl heat loss Q=100W
pump 2000L Dewar
Saturated LHe4.2K , 1.0bar
1.8K ,16mbar
1
24
JT
T-S diagram
Case 3 :Depressurization +JT+ heatexchange
The flow pass through the heatexchange to be cooled by return cold GHe and then it will be throttled to more low temperature by JT ;
3
Vertical Test de
warTotal heat loss
Q=100W
Pump 2000L dewar
Saturated LHe4.2K , 1.0bar
1.8K ,16mbar
1
5
4JT
6
T 5
( K )4.2 3 2.2
1-x 61% 80% 87.7%
m(g/s) 7.01 5.38 4.88
2
Comparison of three methods
Case 1 : No JT and HEX, Ease operation and simple. But it has to be interrupted when
the LHe level lower to the value of limitation and then refilling LHe again.
Case 2 : No heatexchange , lower efficiency. But it can maintain and extend the test time
Case 3 : Complex , higher efficiency. It’s normal to be used in cryogenic system of accelerator.
Calculation of pump speed
Temperature K
Density of saturated liquid
g/l
Latent heat J/g
Density of saturated vapor
g/l
Pressure of saturated vapor Pa
Density at pump inlet at 300K
g/l
1.8 145.42 22.977 0.45469 1638.4 2.6291E-03
2.0 145.65 23.046 0.7936 3129.3 5.0215E-03
300K/1bara / / 0.1604 / /
Temperature K
Heat load W
Mass flow g/s
Volume flow m3/h
Volume flow at outlet
at 300K@1bara m3/h
Volume flow at pump inlet
at 300K m3/h
1.8 20 0.870 6.892 19.536 1191.879
2.0 20 0.868 3.937 19.477 622.162
1.8 50 2.176 17.229 48.840 2979.697
2.0 50 2.170 9.842 48.694 1555.405
1.8 80 3.482 27.567 78.144 4767.515
2.0 80 3.471 15.747 77.910 2488.648
1.8 100 4.352 34.458 97.680 5959.394
2.0 100 4.339 19.684 97.387 3110.810
Performance curve of Edwards EH1200 Roots pump
1mbar is the limitation of vacuum of
Mechanical pump
1000m3/h
Equivalent heat load at 1.5K for 1.8K design
Temperature K
Heat load WMass flow g
/s
Volume flow at 300K/1638.4Pa
m3/h
Equvalent mass flow at 300K/47
1.54Pa g/s
Equvalent heat load at 1.5K W
1.8
10 0.435 595.939 0.125 2.783 3.6 times
20 0.870 1191.878 0.250 5.566
50 2.176 2979.697 0.626 13.917
100 4.352 5959.394 1.253 27.834
Heat power to room temperature for return pipe
Temperature KEnthalpy of cold GH
e J/gPressure of saturated
vapor Pa
Enthalpy of 300K at saturated vapor
pressure J/g
Difference value of Enthalpy J/g
1.8 24.201 1638.4 1573.2 1548.999
2.0 25.041 3129.3 1573.2 1548.159
Temperature K Heat load W Mass flow g/s Heat power W
1.8 20 0.870 1348.30
2.0 20 0.868 1343.54
1.8 50 2.176 3370.76
2.0 50 2.170 3358.85
1.8 80 3.482 5393.22
2.0 80 3.471 5374.15
1.8 100 4.352 6741.52
2.0 100 4.339 6717.69
Recovery and purify system
VACUUM
COMPRESSOR 4.5g/s
INTEGRATED TUBE PURE HELIUM 4000Nm3
INTEGRATED TUBE IMPURE HELIUM 4000Nm3
GHe BAG 50-100m3
10-20 m3
LN2 TOWERPURIFIER
4.5g/s
AMBIENT HEATER 300m3/h
ROOTS PUMPS STATION
LHe OUTLET TO REFRIGERATOR
LHe INLET FROM REFRIGERATOR
EACH ROOTS PUMP PUMPING SPEED 500-1000m3/h TO HELIUM
TEST STATION
1.8K SYSTEM
Recovery & purify systen
LN2 INLET
GN2 OUTLET
Summary
The pump speed will be choosed according to the heat loss of 20W at 1.8K, that is around the capacity of 1000m3/h
The capacity of recovery and purify system are to be thought with the heat loss of 100W at 1.8K and the capacity will be 100m3/h or 4.5g/s
The design and calculation in detail for 2K test system are still underway
Tanks for your attention !