English Warranty Statement .............................................................................. 2 Product Registration ............................................................................. 2 Safety Information ................................................................................ 2 Specications ....................................................................................... 2 Dimensions ........................................................................................... 3 Installation Location .............................................................................. 3 Battery Installation and Replacement ................................................... 3 Terminal Identication ........................................................................... 4 General Wiring...................................................................................... 4 System Power Wiring ........................................................................... 5 Pulse Output Wiring with 12 to 24 volt signal ....................................... 5 Pulse Output Wiring with OPTO-input .................................................. 6 Pulse Output Wiring with Closure Switch ............................................. 6 VIEW Mode .......................................................................................... 7 Using the Security Code ....................................................................... 8 Calibration Mode (Code: 5-6-7-8) ......................................................... 8 Edit Mode (Default 8-1-5-0) ............................................................. 9-20 Edit Mode Table of Contents ................................................................ 9 EDIT Mode Overview ...................................................................... 9 Pulse Output Settings ................................................................... 19 Troubleshooting .................................................................................. 21 Ordering Information........................................................................... 24 *4-3150-D.090* 4-8150-3.090 Rev. A 05/17 Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 Integral Rotor-X Paddlewheel The GF Signet AgRotor Totalizer meets or exceeds the most stringent water authority requirements to provide a easy to operate, password protected, tamper-proof housing while providing an accurate easy to read display of irrigation water use in the Agriculture market. The totalizer can be powered by two “AA” lithium batteries (provided) with a 4 year life or can be powered by either a solar panel or direct 12/24 volts DC which will dramatically increase the life time expectancy of the batteries. Features include: Volumetric pulse output - Telemetry compatible Non volatile memory One permanent and one resettable Totalizer Allows for eld calibration Large easy to read display Password protected 4 year battery life Description Table of Contents Product Manual 4-8150-3-P0 4-8150-3 Panel Mount 4-8150-3 Field Mount English

Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

Embed Size (px)

Citation preview

Page 1: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

English

Warranty Statement .............................................................................. 2Product Registration ............................................................................. 2Safety Information ................................................................................ 2Specifi cations ....................................................................................... 2Dimensions ........................................................................................... 3Installation Location .............................................................................. 3Battery Installation and Replacement ................................................... 3Terminal Identifi cation ........................................................................... 4General Wiring...................................................................................... 4System Power Wiring ........................................................................... 5Pulse Output Wiring with 12 to 24 volt signal ....................................... 5Pulse Output Wiring with OPTO-input .................................................. 6Pulse Output Wiring with Closure Switch ............................................. 6VIEW Mode .......................................................................................... 7Using the Security Code ....................................................................... 8Calibration Mode (Code: 5-6-7-8) ......................................................... 8Edit Mode (Default 8-1-5-0) ............................................................. 9-20Edit Mode Table of Contents ................................................................ 9 EDIT Mode Overview ...................................................................... 9 Pulse Output Settings ................................................................... 19Troubleshooting .................................................................................. 21Ordering Information........................................................................... 24

*4-3150-D.090*4-8150-3.090 Rev. A 05/17

Signet 4-8150-3 AgRotor TotalizerSignet 4-8150-3-P0 Integral Rotor-X Paddlewheel

The GF Signet AgRotor Totalizer meets or exceeds the most stringent water authority requirements to provide a easy to operate, password protected, tamper-proof housing while providing an accurate easy to read display of irrigation water use in the Agriculture market.

The totalizer can be powered by two “AA” lithium batteries (provided) with a 4 year life or can be powered by either a solar panel or direct 12/24 volts DC which will dramatically increase the life time expectancy of the batteries.

Features include:

• Volumetric pulse output - Telemetry compatible

• Non volatile memory

• One permanent and one resettable Totalizer

• Allows for fi eld calibration

• Large easy to read display

• Password protected

• 4 year battery life

Description

Table of Contents

Product Manual

4-8150-3-P0

4-8150-3 Panel Mount

4-8150-3 Field Mount

• English

Page 2: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

Caution / Warning / DangerIndicates a potential hazard. Failure to follow all warnings may lead to equipment damage, injury, or death

Personal Protective Equipment (PPE)Always utilize the most appropriate PPE during installation and service of Signet products.

Pressurized System WarningSensor may be under pressure, take caution to vent system prior to installation or removal. Failure to do so may result in equipment damage and/or serious injury.

Refer to your local Georg Fischer Sales offi ce for the most current warranty statement.

All warranty and non-warranty repairs being returned must include a fully completed Service Form and goods must be returned to your local GF Sales offi ce or distributor. Product returned without a Service Form may not be warranty replaced or repaired.

Signet products with limited shelf-life (e.g. pH, ORP, chlorine electrodes, calibration solutions; e.g. pH buffers, turbidity standards or other solutions) are warranted out of box but not warranted against any damage, due to process or application failures (e.g. high temperature, chemical poisoning, dry-out) or mishandling (e.g. broken glass, damaged membrane, freezing and/or extreme temperatures).

Thank you for purchasing the Signet line of Georg Fischer measurement products.If you would like to register your product(s), you can now register online in one of the following ways: • Visit our website www.gfsignet.com.

Under Service and Support click on Product Registration Form

• If this is a pdf manual (digital copy), click here

Warranty Information

Product Registration

Safety Information• Depressurize and vent system prior to

installation/removal.• Confi rm chemical compatibility before use.• Do not exceed maximum temperature/pressure specifi cations.• Wear safety goggles or face shield during

installation/service.• Do not alter product construction.

4-8150-3 Signet AgRotor Totalizer

Specifi cations

Specifi cations

GeneralCompatibility .................................. Signet AgRotor Flow SensorInput Frequency ............................ 1 Hz to 400 HzAccuracy ...................................... ±0.5% of reading

EnclosureRating ............................................ NEMA 4X/IP65 (front panel)Dimensions ................................... 1/4 DIN (96 mm x 96 mm x 50 mm)

(3.8 in. x 3.8 in. x 2.0 in.)Case material ................................ PBT resinKeypad material ............................ Sealed 4-key silicon rubber

Display: LCD type4-digit upper line ........................... Flow rate or elapsed time8-digit lower line ............................ Two totalizer options: Permanent Totalizer ................. Life of instrument Resettable Totalizer ................. Resettable at keypadDisplay Contrast ........................... Automatic

ElectricalBattery ........................................... Two 3.6 V Lithium thionyl chloride, AA-size (Use

SAFT LS14500 Lithium Batteries or Equivalent Only)

Input Power .................................. 12 to 24 VDC regulated (28 volts DC Maximum)

Current .......................................... 300 Micro-amps MaximumSensor power output ................... +3.6 VDC @ 20μABattery life ..................................... 4 years nominal @ 50 °C (122 °F)Low Battery indication .................. Battery symbol on LCD display

Pulse OutputMaximum current sink ................. 50 mAMaximum cable length ................ 1000 ftPower requirement ...................... 24 VDC max.

EnvironmentalOperating Temperature ............... -10 °C to 65 °C (14 °F to 149 °F)Storage Temperature ................... -40 °C to 100 °C (-40 °F to 212 °F)Relative Humidity ......................... 0 to 95% Non-condensing

Shipping Weight ........................ 0.5 kg (1.1 lbs)

Standards & Approvals• CE• Manufactured under ISO 9001 for Quality, ISO 14001 for Environmental

Management and OHSAS 18001 for Occupational Health and Safety. Declaration of Conformity according to FCC Part 15 This device complies

with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and, (2) This device must accept any interference received,

including interference that may cause undesired operation.

4-8150-3-P0 Signet Integral PaddlewheelGeneral DataFlow Rate Range ......................... 0.3 to 6 m/s (1 to 20 ft/s)Pipe Size Range .......................... ½ in. to 4 in.Linearity ......................................... ±1% of full rangeRepeatability ................................. ±0.5% full rangeCable Type.................................... 2-conductor twisted pair w/shield (22 AWG)Minimum ReynoldsNumber Required ........................ 4500Retaining Nut Material ................. Black Polypropylene

Wetted MaterialsSensor Body ...............................Glass fi lled PolypropyleneO-Rings .......................................FKM (Std) or EPDM or FFKMPin ...............................................TitaniumRotor ............................................Black PVDF with PTFE Sleeve

Shipping Weight ......................0.454 kg (1 lb)

Pressure/Temperature Ratings12.5 bar (180 psi) max. @ 20 °C (68 °F)

1.7 bar (25 psi) max. @ 90 °C (194 °F)

Standards & Approvals• CE

• Manufactured under ISO 9001 for Quality,ISO 14001 for EnvironmentalManagement and OHSAS 18001 for Occupational Health and Safety.

Declaration of Conformity according to FCC Part 15 This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and, (2) This device must accept any interference received,

including interference that may cause undesired operation.

4-8150-3 Signet AgRotor Totalizer continued

2 4-8150-3 AgRotor Totalizer Manual

Page 3: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

15x I.D. 5x I.D. 20x I.D. 5x I.D.

40x I.D. 5x I.D.25x I.D. 5x I.D. 50x I.D. 5x I.D.

Reducer 90° Elbow

2 x 90° Elbow3 dimensions

2 x 90° Elbow Pump/Valve

Dimensions

Installation LocationRecommend:Lubricate the sensors O-rings prior to installing sensor into tee or fi tting1. Install sensor assembly in the 10 or 2 o'clock position. • Avoids direct sunlight on the totalizer display. • Helps prevent rain water accumulation.

CAUTION

When replacing batteries,remove and replaceone battery at a time

Remove and replacethis battery first.

1

2

3

4

5

6

Note: Lithium BatteriesDispose of properly!

Replace with 3.6 V Lithium battery

+ +

- -

Battery Installation 1. Two 3.6 Volt Lithium thionyl chloride batteries

(7400-0011) are installed in the totalizer.

NOTE: THE AgRotor Totalizer WILL NOT OPERATE WITH STANDARD 1.5 VOLT ALKALINE BATTERIES.USE 3.6 VOLT SAFT LS14500 LITHIUM BATTERIES OR EQUIVALENT ONLY!

2. Remove pull tabs from the batteries to power up the AgRotor Totalizer.

3. Observe polarity! Note that both batteries should face the same direction.

Battery Replacement 1. When the “low battery” indicator appears on the display,

both batteries should be replaced within 90 days.

2. GF recommends Saving Totalizer Data before changing batteries. See VIEW Mode, Save Totalizer Data menu, page 7.

3. Remove and replace battery #1 fi rst, then remove and replace battery #2. This ensures the totalizer values are saved. See Caution decal on battery assembly.

4. Secure the batteries by fastening the hook and loop strap.

Shipping Notice:If the battery pull tabs have been removed, remove the batteries from the totalizer prior to shipping.Dispose of the batteries in accordance with the local regulations.

*

NOTE: The low battery indicator should disappear after 10 seconds. If it remains after installing new batteries, one battery may be reversed or battery #2 was installed before battery #1. Repeat steps 3 and 4 above.

ENTER

Signet AgRotor

Low Battery indicator

Flow Rate (GPM)

Permanent Totalizer

DISPOSE OF EXPENDED BATTERIES PROPERLY!Lithium batteries contain hazardous chemicals.Dispose of batteries according to local regulation.

Battery Installation and Replacement

4-8150-3-P0

96 mm(3.8 in.)

96 mm(3.8 in.)

FRONT VIEW

34.8 mm(1.37 in.)

82 mm(3.2 in.)

92 mm(3.6 in.)

41 mm (1.6 in.)

SIDE VIEW

Field and Panel Mount Panel Mount Field Mount with 8050

SIDE VIEW

64 mm(2.5 in.)

102 mm(4.0 in.)

-X0 or -X1

26.7 mm(1.05 in.)

96 mm(3.8 in.)

90°90°

34-8150-3 AgRotor Totalizer Manual

Page 4: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

Wiring Procedure1. Remove 13 mm to 16 mm (0.5 in. to 0.625 in.) of insulation

from the end of the wire to be connected to the transmitter.2. Press the orange terminal lever all the way down with a

small screwdriver to open terminal jaws.3. Insert wire into terminal until it bottoms out.4. Release the orange terminal lever to secure the wire

in place.5. Gently pull on each wire to ensure a good connection.

Wiring TipsDo not route the sensor cable in conduit containing AC power wiring. Electrical noise may interfere with sensor signal. Routing the sensor cable in grounded metal conduit can help prevent electrical noise and mechanical damage.Seal the cable entry points to prevent moisture damage. Only one wire should be inserted into a terminal. Splice double wires outside the terminal.

Recommended: All wiring should only be performed by a certifi ed electrician.

Pull tabs to run sensor off of battery power.

Terminals 7–8: System PowerWhen 12 to 24 VDC is applied to terminals 7–8, the batteries will provide back-up power during power loss. A regulated power supply MUST be used. Maximum current draw is 300 micro-amps.

Terminals 9–10: Pulse OutputPulse Output requires regulated 12 to 24 VDC from an external power supply. Maximum current sink of 50 mA.

Wire the transmitter for all connections with the power off. Keep any Pulse Output devices that are connected to it offl ine at this time.

Sensor Wiring - RotorX and AgRotor

General Wiring

CAUTION: Failure to fully open terminal jaws before removing wire may permanently damage transmitter.

2

1

NO

2

1

YES

2

1

YES

RotorX (Rear View)

8510-P0Flow Sensor

Pulse (+)Pulse (-)PWR (+)PWR (-)

n/an/an/an/a

123456

10987

Splice Black and Shield together outside of terminal

Black

Shield

Red

123456

10987

RotorX (Rear View)

123456

10987

White

Red

Black

4 4-8150-3 AgRotor Totalizer Manual

Page 5: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

AgRotor Totalizer (Rear View)

123456

10987

White

Red

Black

Pull-Up ResistorSee Chart 1

Power Supply12 to 24 VDC

+-

Customer Supplied:Data LoggerSCADA System

Pulse INGround

+-

When 12 to 24 VDC ±10% is applied to terminals 7 and 8, the batteries will provide back-up power during power loss.A regulated power supply MUST be used. Maximum current draw is 300 micro-amps.

Pulse output maximum cable length is 1000 ft. Pulse output maximum current sink of 50 mA. A Pull-Up Resistor MUST be used, choose the correct Pull-Up Resistor using Chart 1 (above). Maximum pull-up voltage is 28 volts at 50 mA maximum. Pulse Output requires regulated 12 to 24 VDC ± 10% from an external power supply. System Power maximum current draw is 300 micro-amps.

Chart 1

Cable LengthResistor (ohms) *

12 Volts 24 Volts

Less than 500 ft 10k 20k

500 ft to 1000 ft 1.2k 2.4k

* For applications with Very High EMI (Noise), use the 500 ft to 1000 ft resistor values.

123456

10987

White

Red

BlackPower Supply12 to 24 VDC

+-

System Power Wiring - Using External Power

Pulse Output Wiring - Volumetric Pulse Output

54-8150-3 AgRotor Totalizer Manual

Page 6: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

AgRotor Totalizer (Rear View)

123456

10987

White

Red

BlackPower Supply12 to 24 VDC

+-

Customer Supplied:Data LoggerSCADA System

Pulse INGround

+-

Pulse output maximum cable length is 1000 ft. Pulse Output requires regulated 12 to 24 VDC ± 10% from an external power supply. System Power maximum current draw is 300 micro-amps.

Pulse output maximum cable length is 1000 ft. Pulse output maximum current sink of 50 mA.A Series Resistor is required if the OPTO-input device does not have one built in. Choose the correct Series Resistor using Chart 2 (above).Pulse Output requires regulated 12 to 24 VDC ± 10% from an external power supply. System Power maximum current draw is 300 micro-amps.

AgRotor Totalizer (Rear View)

123456

10987

White

Red

Black

Series ResistorSee Chart 2 below

Power Supply12 to 24 VDC

+-

Customer Supplied:OPTO-IN Data LoggerOPTO-Isolated Data Logger

Pulse INGround

+-

Chart 2

Voltage Resistor (ohms) *

12 2.2 k

24 4.8 k

* Most OPTO-inputs require about 5mA. If your OPTO-input requires a different mA value, Chart 2Series Resistor values may not apply.

Pulse Output Wiring - Closure Switch

Pulse Output Wiring - OPTO input

6 4-8150-3 AgRotor Totalizer Manual

Page 7: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

Signet AgRotor Totalizer displays fl ow rate in large numerals and totalizer value in smaller numerals. Press or to scroll through VIEW mode.

ENTER

ENTER

ENTER

ENTER

1s

ENTERENTER

1s 1s

ENTER

ENTER

OR

Cancel

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

HOLD

3s

HOLD

3s

Signet AgRotor

Signet AgRotor

Signet AgRotor

Signet AgRotor

Signet AgRotor

Signet AgRotor

Signet AgRotor Signet AgRotor Signet AgRotor Signet AgRotor

Signet AgRotor Signet AgRotor Signet AgRotor

Displays the number of hours since fl ow was last detected.

Any movement of the rotor (fl ow rate greater than zero) will reset the ELAPSED TIME display.

Flow Rate and Permanent Totalizer

Flow Rate and Resettable Totalizer

Default = 00000

Serial Number (Programmable)

No Flow and Elapsed Time

Save Totalizer Data (Recommended before changing batteries) See page 3.

Default = Custom

4-3150-31 = Acre Feet4-3150-32 = Acre Inch4-3150-33 = Gallons x 1000

Totalizer Units

Default = Custom

Flow K-Factor

VIEW Mode

ENTER

Signet AgRotor

74-8150-3 AgRotor Totalizer Manual

Page 8: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

Code: 5-6-7-8

A security code must be entered before settings can be changed or to perform calibration.Default Security Codes: "5-6-7-8" is used to access the Calibration Menu. For details see below. "8-1-5-0" is used to access the Edit Menu. For details see page 9.

For instructions on how to access Calibration Mode, see Using the Security Code above.

Use or to scroll through Calibration Mode menus. Press + together to store any changes and return to VIEW Mode (See page 13).

Enter security code of 8-1-5-0:

1. Hold ENTER for 2 seconds. The display shows 0000, with the fi rst zero fl ashing.

2. Press twice to scroll the fl ashing zero to 8. 3. Press once to advance the fl ashing character one place to the right. 4. Press once to scroll

the fl ashing zero to 1. 5. Press once to

advance the fl ashing character one place to the right.

6. Press fi ve times to scroll the fl ashing zero to 5. Do not change the last zero.

7. Press the ENTER key. The display now shows the fi rst menu in EDIT Mode (Sn: Serial Number).

• Assign Serial Number• Pipe size (Flow K-Factor)• Timebase (Gallons per seconds, minutes, hours, days)• Flow Decimal Resolution• Total K-Factor units• Perform Auto Calibration

HOLDENTER

2 s

1

ENTER

Signet AgRotor

ENTER

8ENTER

ENTER

ENTER

Signet AgRotor32

ENTERENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor5

ENTER

Signet AgRotor

x5

x2

6 7

4Signet AgRotor

Signet AgRotor

• Reset the resettable totalizer• Manually store totalizer data prior to changing

batteries• Set Speed• Set Sensitivity• Change Security Codes

Security Code 8-1-5-0 must be entered before the following settings can be changed:

Using the Security Code

Calibration Mode

Auto Calibration (Auto CAL) Press to select the Auto Calibration function.See page 13 for instructions.

Current K-Factor in Use Default: 60.00

ENTER

ENTER

Signet AgRotor

Signet AgRotor

Calibration Date Manually entered date of last Auto CAL. Press to edit (only available after successful Auto CAL).See page 14 for instructions.ENTER

Signet AgRotor

8 4-8150-3 AgRotor Totalizer Manual

Page 9: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

Default Code: 8-1-5-0

For instructions on how to access EDIT Mode, see Using the Security Code on page 8.

Use or to scroll through EDIT Mode menus. Press + together to store any changes and return to VIEW Mode (See page 7).

Sn: Serial NumberDefault: 0-0-0-0-0Press to edit. See page 11 for instructions.

Pipe Size and Code (Flow K-Factor)Default: Custom Available pipe sizes: Press to edit. See page 11 for instructions.

Custom Flow K-FactorDefault: 0060.000 Available range: 0000.001 - 9999999.Press to edit. See page 12 for instructions.

Auto Calibration (Auto CAL) Press to select the Auto Calibration function.See page 13 for instructions.

Current K-Factor in UseCurrent K-Factor based on Pipe Size and Code (Flow K-Factor) setting, see page 11. Default: 60.00

Calibration Date Manually entered date of last Auto CAL. Press to edit (only available after successful Auto CAL).See page 13 for instructions.

EDIT Mode Overview ...........................................9-10Serial Number.......................................................... 11Custom Flow K-Factor ............................................. 11Flow K-Factor tables................................................ 12Pipe Size and Code (Flow K-Factor) ....................... 12 Current K-Factor in Use........................................... 13Auto Calibration (Auto CAL) .................................... 13Calibration Date ....................................................... 14Calibration Lock/Unlock ........................................... 14Set Custom Total K-Factor ...................................... 15

Total K-Factor (Totalizer Units) ................................ 15Custom Totalizer Decimal Resolution ...................... 16Timebase ................................................................. 16Flow Decimal Resolution ......................................... 17Speed (Averaging) ................................................... 17Sensitivity ................................................................ 18 Enable Pulse Output................................................ 19Pulse Volume........................................................... 19Test Pulse Output .................................................... 20Changing Security Code 1 (8-1-5-0)........................ 20

EDIT Mode Table of Contents

EDIT Mode Overview

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

94-8150-3 AgRotor Totalizer Manual

Page 10: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

Signet AgRotor

Signet AgRotor

Signet AgRotor

Signet AgRotor

Signet AgRotor

Signet AgRotor

Calibration Lock / Unlock Displays the lock status of the Auto CAL function. If YES, then Auto CAL cannot be performed. If NO, calibration is allowed and will automatically locks for one hour after a successful Auto CAL. Press to edit (to unlock Auto CAL function). See page 14 for instructions.

Total K-Factor (Totalizer Units)Default: based on system part number. See page 15. Available units: 1000 x Gal, Ac Ft, Ac in, CUStPress to edit. See page 15 for instructions.

Custom Total K-FactorDefault: 1.00 Available range: 0000.001 - 9999999. Press to edit. See page 15 for instructions.

Timebase (Flow rate = Gallons per Timebase)Default: Min (minute) Available settings: Sec, Min, Hr, DayPress to edit. See page 16 for instructions.

Custom Totalizer Decimal ResolutionDefault: - - - - - . - - - Available settings: - - - - - - - - . , - - - - - - - . - ,

- - - - - - . - - , - - - - - . - - -Press to edit. See page 16 for instructions.

Flow Decimal ResolutionDefault: - - - . - Available settings: - - . - -, - - - . -, - - - -Press to edit. See page 17 for instructions.

Speed (Averaging)Default: 30 Available settings: 0, 7, 15, 30, 60, 120Press to edit. See page 17 for instructions.

SensitivityDefault: 6 Available range: 1 to 10Press to edit. See page 18 for instructions.

Enable Pulse OutputDefault: no Available range: YES or noPress to Enable/Disable. See page 19 for instructions.

Test Pulse Output (Only visible when Enable Pulse Output = Yes)Temporary turn 'On' (closed) or 'OFF' (open) Pulse Output.Press to activate. See page 20 for instructions.

Pulse Volume (Only visible when Enable Pulse Output = Yes)Default: 100.0 Available range: 00.01 to 9999Press to edit. See page 19 for instructions.

Security Code 1Default: 8-1-5-0Press to edit. See page 20 for instructions.

Default Code: 8-1-5-0

For instructions on how to access EDIT Mode, see Using the Security Code on page 8.

Use or to scroll through EDIT Mode menus. Press + together to store any changes and return to VIEW Mode (See page 7).

Edit Mode Overview continued

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

Signet AgRotor

Signet AgRotor

Signet AgRotor

Signet AgRotor

Signet AgRotor

Signet AgRotor

ENTER

Signet AgRotor

10 4-8150-3 AgRotor Totalizer Manual

Page 11: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

A fi ve digit serial number (0-9) can be assigned to the AgRotor -3. Default: 0-0-0-0-0

Example: Change 0-0-0-0-0 serial number to 1-9-0-0-0. 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

The fi rst menu item will be displayed (Sn - Serial Number). 2. Press once to edit the Serial Number. The fi rst '0' will begin fl ashing.

3. Press once to advance the fl ashing character one place to the right: '0'.

4. Press once to change the fl ashing '0' to 9.

5. Press ENTER to fi nish editing and save the changed Serial Number.• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

Assigning Serial Number (Sn)

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

ENTER ENTER ENTER

Signet AgRotor Signet AgRotor Signet AgRotor

1 2 3 4 5

This menu is only available when CUSt (custom) is selected for Pipe Size and Code (Flow K-Factor). See page 15.Available range: 0000.001 - 9999999. Default: 0060.000

The K-Factor is the number of pulses generated by the fl ow sensor for each measure of water that moves past the sensor. Your fl ow sensor manual contains K-Factor data in terms of U.S. gallons and liters.Locate the K-Factor that matches your pipe size material. If necessary, you can convert the K-Factor into other units of measure. The minimum K-Factor value is 0.001, maximum is 999999.

Example: Change 0060.000 to 0100.000 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press once to display the Pipe Size and Code (K-Factor) menu. Select CUSt (see page 15).

2. Press once to display the Custom Flow K-Factor menu.

3. Press once to edit the Custom Flow K-Factor. The '0' (zero) on the left will begin fl ashing.

4. Press once to advance the fl ashing character one place to the right.

5. Press once to scroll the fl ashing '0' to '1'.

6. Press once to advance the fl ashing character one place to the right.

7. Press four times to scroll the fl ashing '6' to '0'.

8. Press ENTER to fi nish editing and save the changed Custom Flow K-Factor.

• The display has returned to EDIT Mode, use or to scroll through settings.

• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

ENTER

Signet AgRotor

1 42 3

5 86 7

x4

Set Custom Flow K-Factor

114-8150-3 AgRotor Totalizer Manual

Page 12: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

AgRotor Totalizer supports four common pipe sizes, each with multiple code settings, and a custom setting. Default: 8 in. 04

Each Pipe Size and Code combination has a pre-programmed K-Factor value. The CUSt (custom) setting is programmed in the next menu. See page 11.

The K-Factor is the ratio of pulses per unit of fl ow. This ratio represents the number of pulses generated by the fl ow sensor for each measure of water that moves past the sensor. See 4-3150-3X manual or contact distributer/factory..

Example: Change 8 in 04 (K-Factor) to 10 in 01 (K-Factor) 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER. Press once to display the Pipe Size and

Code (K-Factor) menu.

2. Press once to edit. '8 in 04' will begin fl ashing.

3. Press eight times to scroll to the '10 in 01' setting.

4. Press ENTER to fi nish editing and save the changed Pipe Size and Code.

• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

Changing Pipe Size and Code (Flow K-Factor)

ENTER

ENTER ENTER ENTER

Signet AgRotor

ENTER

Signet AgRotor Signet AgRotorSignet AgRotor

1 2 3 4

x4

PIPESIZE(IN.)

FITTING 515/8510-XX

U.S. GAL LITERS

SCH 80 PVC TEES FOR SCH 80 PVC PIPE½ 4-MPV8T005 480.19 126.87¾ 4-MPV8T007 257.72 68.0901 4-MPV8T010 174.67 46.1481¼ 4-MPV8T012 83.390 22.0321½ 4-MPV8T015 58.580 15.4772 4-MPV8T020 32.480 8.5812SCH 80 PVC SADDLES FOR SCH 80 PVC PIPE2 4-PV8S020 32.480 8.58122½ 4-PV8S025 21.833 5.76833 4-PV8S030 13.541 3.57754 4-PV8S040 7.6258 2.0147

PIPESIZE(IN.)

FITTING 515/8510-XX

U.S. GAL LITERS

SCH 80 IRON SADDLES ON SCH 80 PIPE2 4-IR8S020 32.360 8.54952½ 4-IR8S025 22.220 5.87053 4-IR8S030 13.420 3.54564 4-IR8S040 7.6600 2.0238SCH 80 IRON SADDLE ON SCH 40 PIPE2 4-IR8S020 26.820 7.08592½ 4-IR8S025 18.800 4.96703 4-IR8S030 11.990 3.16784 4-IR8S040 6.8500 1.8098SCH 80 PVC SADDLE ON SCH 40 PVC PIPE2 4-PV8S020 27.350 7.22592½ 4-PV8S025 18.874 4.98663 4-PV8S030 12.638 3.33894 4-PV8S040 6.7282 1.7776

A K-Factor is the number of pulses a sensor will generate for each engineering unit of fl uid which passes the sensor. K-factors for water are listed below in U.S. gallons and liters. For example, in a 1-inch PVC pipe, the Rotor-X Paddlewheel generates 174.67 pulses per gallon of water passing the rotor. K-factors are listed for pipes up to 8 inches.

K-Factors

12 4-8150-3 AgRotor Totalizer Manual

Page 13: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

This menu displays:

• Flow K-Factor value currently used to determine fl ow rate. Or• Custom Flow K-Factor (above). Or• Automatic Flow K-factor resulting from Auto CAL (calibration) feature (page 13).

The K-Factor is the ratio of pulses per unit of fl ow. This ratio represents the number of pulses generated by the fl ow sensor for each measure of water that moves past the sensor.

NOTE: This value cannot be changed from this menu. See above to change Flow K-Factor.

View Current Flow K-Factor

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

The Auto CAL feature allows the AgRotor Totalizer to be adjusted to match the fl ow rate of an external reference meter.*Maximum calibration allowed is 10% of the current Flow K-Factor setting. See page 12.Recommended for best results:

• Flow in the pipe should be as stable as possible for best results• If the displayed fl ow rate is erratic, set Speed (see page 17) to 120 seconds for the Auto CAL procedure.• The Timebase (Sec, Min, Hour, Day) on the reference meter and the AgRotor Totalizer must be the same.

Example: AgRotor Totalizer displays 60 GPM, while an external reference displays a true fl ow rate of 65 GPM.Auto CAL can be used to adjust the AgRotor fl ow K-Factor automatically to correctly display 65 GPM.

1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER. Press three times to display the Auto CAL menu.

NOTE: If "LOC" is displayed on the Auto CAL menu, it must fi rst be unlocked before use. See page 14. 2. Press once to select the Auto CAL function. The current fl ow rate will begin fl ashing. For maximum accuracy, wait at least 2

minutes before proceeding to the next step. 3. Press once to change the fl ow rate.

"Auto CAL" will be replaced with 'Set Flo' and the fi rst digit of the current fl ow rate will begin fl ashing.

4. Press once to advance the fl ashing character one place to the right: '0'.

5. Press fi ve times to change the '0' to '5'.

6. Press ENTER to complete the automatic calibration process. A new Flow K-Factor will be displayed (fi rst digit fl ashing). This K-Factor is based on the change in fl ow rate.

a. Press + together to Cancel the Auto CAL. 'StoPPEd' will be displayed and then return to EDIT Mode.

b. 'Err Set Flo' will be displayed if the new calculated K-Factor is more than 10% of the original Flow K-Factor setting. Contact GF for more assistance. Verify the fl ow rate and restart the Auto CAL procedure from step 1.

7. Press ENTER to fi nish editing and save the Auto CAL K-Factor.

• The display has returned to EDIT Mode, use or to scroll through settings.

• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER ENTER

1

4

2

5

3

6

ENTER

Signet AgRotor6b

ENTER

6a

7

OR

OR

TO CANCEL:

+

Signet AgRotor

Signet AgRotor

x5

Auto Cal (Flow K-Factor Calibration)

134-8150-3 AgRotor Totalizer Manual

Page 14: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

Example: Change 01-01-15 to 02-01-15 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press four times to display the Calibration Date menu.

2. Press once to edit Calibration Date. The left-most digit '0' will begin fl ashing.

3. Press once to advance the fl ashing character one place to the right: '1'.

4. Press once to scroll the fl ashing '1' to '2'.

5. Press ENTER to fi nish editing and save the changed Calibration Date.

• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

Example: Change no to YES (Lock Auto CAL). 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press fi ve times to display the Calibration Lock / Unlock menu.

2. Press once to edit Calibration Lock / Unlock. 'no' will begin fl ashing.

3. Press once to scroll to the 'YES' setting.

4. Press ENTER to fi nish editing and save the changed Calibration Lock / Unlock.

5. The Auto CAL menu will now display 'LOC' and Auto CAL cannot be performed.

• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

This menu displays the date of the last successful Auto CAL. The AgRotor Totalizer does not have a real time clock, so the date must be entered manually. This date can only be edited after a successful Auto CAL, see page 13. Default: 01-01-15

NOTE: This value can not be changed if the Auto CAL function is Locked (LOC). See below.

This menu displays the lock status and allows manual locking and unlocking of the Auto CAL function. If YES is displayed, then Auto CAL cannot be performed. Default: no

NOTE: The Auto CAL function automatically locks for one hour after a successful Auto CAL.

ENTER

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor1 4 52 3

ENTER

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor1 4 52 3

Calibration Date

Calibration Lock/Unlock

14 4-8150-3 AgRotor Totalizer Manual

Page 15: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

ENTER

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor1 42 3

AgRotor Totalizer can display total volume of water (totalizer) as Gallons x 1000, Acre Inch, Acre Feet, or Custom.

WARNING: It is not permitted to use the custom (CUSt) Total K-Factor setting for measuring irrigation water.

Default:

Example: Change 1000 GAL to CUSt (Custom K-Factor). 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press six times to display the Total K-Factor menu.

2. Press once to edit Total K-Factor. '1000 GAL' will begin fl ashing.

3. Press once to scroll to the 'CUSt' setting.

4. Press ENTER to fi nish editing and save the changed Total K-Factor.• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

Total K-Factor Ag Rotor Part No.

Acre Feet Ac Ft 4-3150-31

Acre Inch Ac in 4-3150-32

Gallons x 1000 1000 GAL 4-3150-33

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

1 42 3

x2ENTER

Signet AgRotor5

This menu is only available when CUSt (custom) is selected for Total K-Factor (Totalizer Units). See above.Available range: 0000.001 - 9999999. Default: 1000000.

The TOTAL K-Factor is a multiple of the FLOW K-Factor. Use it to program the incremental count size of the totalizer.

Example: If the fl ow RATE registers in GALLONS per minute, the totalizer may be set to 1 (factory standard), so it counts in 1 gallon increments, or it may be set to 1000, so it counts in 1 gallon X 1,000 increments. By converting the fl ow K-Factor, the totalizer can also be set to count in other engineering units, See below for additional information about Total K-Factor adjustments.

Example: Change 1000000. to 3000000. 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press six times to display the Total K-Factor menu. Select CUSt as Total K-Factor (see above).

2. Press once to display the Custom Total K-Factor menu.

3. Press once to edit the Custom Total K-Factor. The '1' on the left will begin fl ashing.

4. Press twice to scroll the fl ashing '1' to '3'.

5. Press ENTER to fi nish editing and save the changed Custom Total K-Factor.• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

Total K-Factor (Totalizer Units)

Set Custom Total K-Factor

154-8150-3 AgRotor Totalizer Manual

Page 16: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

This menu is only available when CUSt (custom) is selected for Total K-Factor (Totalizer Units). See page 15.Available range: - - - - - - - - . , - - - - - - - . - , - - - - - - . - - , - - - - - . - - - Default: - - - - - - - . - NOTE: The decimal point does not auto-range (self adjust). The totalizer will roll over when greater than 99999999,

regardless of where the decimal point is located.

Example: Change - - - - - . - - - to - - - - - - - . - 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press six times to display the Total K-Factor menu. Select CUSt as Total K-Factor (page 15).

2. Press twice to display the Totalizer Decimal Resolution menu.

3. Press once to edit the Totalizer Decimal Resolution menu. The '.' will begin fl ashing and move to the far right.

4. Press once to scroll the fl ashing '.' one place to the left.

5. Press ENTER to fi nish editing and save the changed Totalizer Decimal Resolution.

• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

ENTER

x2ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor1 42 3

AgRotor Totalizer timebase can be changed to display the fl ow rate as Gallons per Second, Minute, Hour, or Day. Default: Min (Minute)

Example: Change Min (Minute) to Day. 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press six times to display the Timebase menu.

2. Press once to edit Timebase. 'Min' will begin fl ashing and the other options (Sec, Hr, Day) will display.

3. Press twice to scroll to the 'Day' setting.

4. Press ENTER to fi nish editing and save the changed Timebase.

• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

1 42 3

x2ENTER

Signet AgRotor5

Change Custom Totalizer Decimal Resolution

Changing Timebase

16 4-8150-3 AgRotor Totalizer Manual

Page 17: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

ENTER

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor1 42 3

Flow rate decimal resolution can be changed to display hundredths ( - - . - - ), tenths ( - - - . - ), or whole numbers ( - - - - . ). Default: - - - . - (tenths)

The selected decimal resolution will auto-range to the next level of resolution:

Decimal Resolution Display Range Auto-Range Actualfl ow rate Will display as

Whole Numbers ( - - - - . ) 0000. - 9999. n/a 10.55 11Tenths ( - - - . - ) 000.0 - 999.9 1000. - 9999. 10.55 10.6Hundredths ( - - . - - ) 00.00 - 99.99 100.1 - 999.9, 1000. - 9999. 10.55 10.55

Example: Change - - - . - (tenths) to - - . - - (hundredths). 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press fi ve times to display the Flow Decimal Resolution menu.

2. Press once to edit Flow Decimal Resolution. The '.' in - - - . - will begin fl ashing.

3. Press once to move the '.' one place to the left: '- - . - -'

4. Press ENTER to fi nish editing and save the changed Flow Decimal Resolution.

• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

ENTER

x2ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor1 42 3

Speed smooths out fl uctuations in fl ow rate by averaging readings over: 0, 7, 15, 30, 60, or 120 seconds. Default: 30Fluctuations may be caused by inadequate straight pipe runs after pumps, values, and elbows in the application.

Averaging Speed Application conditionsFaster 0 s to 30 s Well-established, Stable FlowSlower 60 s to 120 s Unstable Flow

NOTE: While the SPEED setting helps to smooth out the fl uctuations caused by piping conditions, it also causes a delay in showing actual changes in fl ow rate. The SENSITIVITY function (see page 18) is designed to help offset this effect.

Example: Change Speed (Averaging) 30 to 120. 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press four times to display the Speed (Averaging) menu.

2. Press once to edit the Speed (Averaging). '30' will begin fl ashing.

3. Press twice to scroll to the '120' setting.

4. Press ENTER to fi nish editing and save the changed Speed (Averaging).• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

Changing Flow Decimal Resolution

Speed (Averaging)

174-8150-3 AgRotor Totalizer Manual

Page 18: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

ENTER

x3ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor1 42 3

Sensitivity determines how the AgRotor Totalizer responds to surges in fl ow rate. Sensitivity "overrides" Speed (Averaging) long enough to display the change in fl ow rate. Speed (Averaging) smooths out fl uctuations in fl ow rate by averaging readings over the selected number of seconds (see page 17).

The result of Sensitivity is a smooth fl ow display and a quick response to large shifts in fl ow rate. Available Sensitivity range: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10. Default: 5NOTE: Sensitivity is ineffective if Speed (Averaging) is set to zero (seconds)

Example: Change Sensitivity from 5 to 8. 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press three times to display the Sensitivity menu.

2. Press once to edit Sensitivity. '5' will begin fl ashing.

3. Press three times to scroll to the '8' setting.

4. Press ENTER to fi nish editing and save the changed Sensitivity.

• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

10 s 20 s 30 s 40 s 50 s 60 s 70 sTime

Velo

city

No SPEED, no SENSITIVITY

SPEED only

SPEED and SENSITIVITY

With SPEED set to 0 (zero) and with SENSITIVITY set to 1, the fl ow rate may be very unstable.This line represents the actual output of the fl ow sensor as it responds to unstable fl ow conditions in the pipe.

With SPEED set to 60 seconds and SENSITIVITY set to 1 the fl ow rate is stabilized, but a sharp change in fl ow rate is not represented for 60 seconds or longer.

With SPEED at 60 seconds and SENSITIVITY set to 6, the fl ow rate is stabilized, while the sudden shift in fl ow is refl ected very quickly.

(dotted green line)

(solid blue line)

(dashed red line)

Sensitivity

18 4-8150-3 AgRotor Totalizer Manual

Page 19: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

ENTER

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor1 42 3

This menu Enables or Disables the Pulse Output option. For Pulse Output wiring, see pages 5 and 6.Default: no (off)

Example: Change no to YES (enable Pulse Output). 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press twice times to display the Enable Pulse Output menu.

2. Press once to edit Enable Pulse Output. 'no' will begin fl ashing.

3. Press once to change to 'YES'.

4. Press ENTER to fi nish editing and save the changed Enable Pulse Output status.

• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

This menu is only available after Enabling Pulse Output. See above. This value represents the number of volume units required to generate a pulse output. Default: 100.0 (units per pulse output)

Example: Change 100.0 to 200.0 (200 units per pulse output). 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press twice to display the Enable Pulse Output menu. Select 'YES' to Enable Pulse Output (see above).

2. Press once to display the Pulse Volume menu.

3. Press once to edit the Pulse Volume quantity. The '1' on the left will begin fl ashing.

4. Press once to scroll the fl ashing '1' to '2'.

5. Press ENTER to fi nish editing and save the changed Pulse Volume.

• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

ENTER

ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor42 3

ENTER

Signet AgRotor5

ENTER

Signet AgRotor1

Enable Pulse Output

Pulse Volume

194-8150-3 AgRotor Totalizer Manual

Page 20: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

The security code prevents unauthorized tampering with calibration and operation settings in the Totalizer.

Example: Change security code of 8-1-5-0 to 0-1-0-0. 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press once to display the Security Code 1 menu. 2. Press once to edit Security Code 1. The fi rst digit ('8') will begin fl ashing. 3. Press twice to scroll the fl ashing '8' to '0'. 4. Press twice to advance the fl ashing character two places to the right: '5'. 5. Press fi ve times to scroll the fl ashing '5' to '0'. 6. Press ENTER to fi nish editing and save the changed Security Code 1.

• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

Recommended: Record and store your security code in a safe place!

x2 x2

ENTER

x5

ENTER

Signet AgRotor1

ENTER

Signet AgRotor2

ENTER

Signet AgRotor3

ENTER

4FLO-WISE SC-3

ENTER

5FLO-WISE SC-3

ENTER

6FLO-WISE SC-3

This menu is only available after Enabling Pulse Output. See page 19. Allows the Pulse Output to be turned 'On' (closed) or 'OFF' (open) for testing purposes. Normal Pulse Output is suspended while testing and up to 60 normal pulses will be saved for later output.

Testing Pulse Output: 1. Hold ENTER for 2 seconds, enter Security Code 8-1-5-0, and press ENTER.

Press twice to display the Enable Pulse Output menu. Select 'YES' to Enable Pulse Output (see page 19).

2. Press twice to display the Test Pulse Output function.

3. Press once to activate Test Pulse Output function. 'OFF' will fl ash on the display, indicating the Pulse Output is off (open).

4. Press once to change the fl ashing 'OFF' to 'On'. The fl ashing 'On' indicates the Pulse Output is on (closed).

5. Press ENTER to deactivate the Test Pulse Output function.

• The display has returned to EDIT Mode, use or to scroll through settings.• Press + together to exit EDIT Mode. "Storing" will display, then return to VIEW Mode (operating mode).

For instructions on how to access EDIT mode, see Using the Security Code on page 8.

ENTER

x2ENTER

Signet AgRotor

ENTER

Signet AgRotor

ENTER

Signet AgRotor42 3

ENTER

Signet AgRotor5

ENTER

Signet AgRotor1

Test Pulse Output

Changing Security Code 1 (8-1-5-0)

20 4-8150-3 AgRotor Totalizer Manual

Page 21: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

Display Condition Probable Cause Suggested Solutions

Batteries are dead or missingReplace both batteries. (See page 4).Use Saft LS14500 Lithium Batteries or equivalent ONLY.

Pull tabs have not been removed. Remove plastic pull tabs protecting the batteries.

Totalizer has not received a signal from the fl ow sensor for the displayed number of hours. (ex: three hours)

1. There is no fl ow in the pipe.2. Flow sensor is not turning due to blockage or damage.3. Sensor wiring is loose or incorrect.

Displayed Flow Rate is unstable.

Usually caused by inadequate straight pipe run upstream of sensor.

1. Correct piping layout to provide more straight pipe upstream of sensor.

2. Set the SPEED to higher setting to average out the fl uctuations caused by piping conditions. (Contact GF representative)

Both batteries are too depleted to safely store settings.

Replace battery #1, then replace battery #2. (See page 4)

Use Saft LS14500 Lithium Batteries or equivalent ONLY.

Troubleshooting

214-8150-3 AgRotor Totalizer Manual

Page 22: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

Notes

22 4-8150-3 AgRotor Totalizer Manual

Page 23: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

Notes

234-8150-3 AgRotor Totalizer Manual

Page 24: Signet 4-8150-3 AgRotor Totalizer Signet 4-8150-3-P0 ... · PDF file1/4 DIN (96 mm x 96 mm x 50 mm) ... 4-8150-3 AgRotor Totalizer Manual 3. Wiring Procedure 1. Remove 13 mm to 16

Georg Fischer Signet LLC, 3401 Aero Jet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090For the most up-to date information, please refer to our website at www.gfsignet.com

4-8150-3.090 Rev. A 05/17 English © Georg Fischer Signet LLC 2017

Ordering Information

Totalizers and Replacement PartsMfr. Part No. Code Description4-8150-3* 159 050 422 Irrigation Flow Totalizer, Field Mount

4-8150-3P* Special Order** Irrigation Flow Totalizer, Panel Mount

4-8510-P0 Special Order** Replacement Paddlewheel Sensor

3-8050 159 000 184 Universal Remote Mount w/ Conduit Base

7400-0011 159 000 935 3.6V Lithium Battery for 8150 (two required)

7310-2024 159 973 005 24 VDC power supply, 24W, 1.0A

*NOTE: Use Saft LS14500 Lithium Batteries or equivalent ONLY.**NOTE: Contact factory for support/quote.

Part No. Code Description4-MPV8T005F Contact Factory ½ in. PVC Tee4-MPV8T007F Contact Factory ¾ in. PVC Tee4-MPV8T010F Contact Factory 1 in. PVC Tee4-MPV8T012F Contact Factory 1¼ in. PVC Tee4-MPV8T015F Contact Factory 1½ in. PVC Tee4-MPV8T020F Contact Factory 2 in. PVC Tee4-PV8S020 Contact Factory 2 in. PVDC Saddle4-PV8S025 Contact Factory 2 ½in. PVDC Saddle4-PV8S030 Contact Factory 3 in. PVDC Saddle4-PV8S040 Contact Factory 4 in. PVDC Saddle

Tees and Saddles for 4-8150-3-P0