Upload
deborah-cameron
View
215
Download
3
Embed Size (px)
Citation preview
Sept 26, 2009Ashley Meister
1
Objectives Case set up Compare cases for set up Positions, effects on patient, risks Fluid replacement, scavenging Suction Ventilator set up
2
Patient Positioning Indications, precautions, complications and procedure for each of the following
patient positions: Supine Prone Lithotomy Beach chair Lateral decubitus Supine/ fracture table
3
General Concepts in positioningsedated/ anesthetized patients should not be
placed in positions they are not comfortable in when they are awake
Compromise between what patient can tolerate structurally and physiologically, and what is required for surgical access
Physiologic instability may be magnified by rapidly moving seriously ill patients
4
PositioningBony prominences can produce ischemic
necrosis of overlying tissue unless proper padding is required
Enhanced by hypothermia and hypotensionCaution particularly with ulnar nerve
5
Supine
Lying horizontallyArm pressure points padded and either tucked to side
or abductedAbduct less than 90 degreesExtend hands ventrally Ensure perfusion to the hand, no skin to metal contact
and no stretch on brachial neurovascular bundleNo compression in the axillaBony contacts at occiput, elbows & heals padded
6
SupineHorizontal supine, minimal changes to
vascular systemIf tipped into trendelenburg or reverse
trendelenburg, effects of gravity on blood flow significant.
Pressures change 2mmHg for each 2.5cm above or below level of the heart
7
SupineReverse trendelenburgBlood pools in legs, decreasing effective circulating volumeDecreased cardiac outputDecreased systemic perfusionPerfusion pressure in brain correspondingly decreased
compared to if measured at level of the heartVentilation dynamics are enhanced
8
SupineTrendelenburgIncreased pressure in cerebral veinsCan increase ICPCongestion around eyes and airwayNegative impact on ventilation
9
SupineRespiratory “Zones of West” shift Diaphragm is pushed cephalad Decreased FRC
10
SupinePregnant uterus rests on great vessels of the
abdomenAortocaval compression- therefore tilt into
Left lateral decubitus position/ left uterine displacement
11
SupineExcessive flexion or extension of the spine in
anesthetized patients who are placed in unique surgical positions may contribute to spinal cord ischemia and catastrophic neurological damage
12
13
Considerations with Prone positioning?
14
Prone Venous pooling in legs, decreased preload and decreased
cardiac output
If pressure is on abdominal viscera, transmitted to veins in spinal canal, causes increased bleeding in spine procedures
Extensive spine procedures in the prone position is associated with post operative visual loss (associated with blood loss, anemia & hypotension)
15
ProneImportance of secure airwayAlways have stretcher outside room in case
airway is lostCongestion of face and airwayCheck eyes & ears carefullyEnsure arms not extended > 90 degrees, and
well padded
16
What would you do?A/W is lost when prone
Key point- prevention
17
1
18
LithotomyGynecologic and urologic proceduresSupine, arms crossed on trunk or extended
laterally on arm boardsFlex lower extremities at hip and kneeBoth limbs simultaneously elevated and
separatedNerve injury possible if hips flexed greater
than 90 degrees
19
LithotomyEnsure padding over lower extremities if
pressure points existCan get hypotension if legs lowered quickly
or decreased effective circulating blood volume
Decreases diaphragmatic excursion and impairs ventilation
Caution with hands and watch BP when leveling table back to neutral
20
LithotomyElevated lower extremity positions may
reduce perfusion pressure in the elevated extremities
conditions for developing compartment syndromes, especially when extremities are elevated for prolonged periods
Maintain perfusion pressure to extremities
21
22
Beach ChairOften intubated as access to airway is
difficultEnsure ETT well secured and stays in place
while moving patient and bedCaution with elevating head of table with
venous pooling and hypotensionCase reports with decreased cerebral
perfusion
23
Lateral DecubitusTurned onto one side(left side down = left lateral decubitus
position)Place an axillary roll just under chest to take
pressure off axillary neurovascular bundleV/Q mismatch may occur, particularly with
co-existing pulmonary diseaseCaution with pressure to eyes & ears
24
25
Fracture TableFor repair of fractured femurPelvis is retained in place by a vertical pole at
perineum with the foot of the injured extremity fixed to a mobile rest
Traction is applied between the foot and pelvis
Perineal crush injury possible
26
Setting up the caseAssist with surgical draping, while
maintaining the integrity of the sterile fieldAvoid walking between or crossing over
sterile fields
27
Setting up the casePrepare, in consultation with the
anaesthesiologist, medication needs for general and regional anesthesia
28
Emergency DrugsSelection and preparation of medications, checked and
labelled for usage as appropriateFor every case:Succinylcholine 20 mg/ml 10mL syringeAtropine 0.4mg/ml- 0.6 mg/ml vials, 1mL syringeEphedrine 5mg/ml (50mg vial/ 10cc)Phenylephrine 100mcg/ml (10mg/100cc)
29
Equipment to PrepareLocal Sedation RegionalNeuraxial – spinal/ epidural/ CSEGeneral
30
CAS monitors Required: Pulse oximeter Apparatus to measure blood pressure, either directly or
noninvasively Electrocardiography Capnography, when endotracheal tubes or laryngeal masks are
inserted. Agent-specific anesthetic gas monitor, when inhalation
anesthetic agents are used.
31
CAS monitorsExclusively available for each patient: Apparatus to measure temperature Peripheral nerve stimulator, when
neuromuscular blocking drugs are used Stethoscope - either precordial, esophageal
or paratracheal Appropriate lighting to visualize an exposed
portion of the patient.
32
CAS monitorsImmediately available: Spirometer for measurement of tidal volume.
33
Preparation for Local/ standbyStandard CAS standard monitors in useAnesthesia available to provide sedationLocal anesthetic as per surgeon (watch
doses)Have emergency drugs available
34
Preparation for Sedation
CAS monitorsEmergency drugs available, IV, oxygenUseful to monitor capnographyMany drugs can be used to provide sedationMidazolamFentanylRemifentanilKetamine
35
Preparation for Regional CAS monitors Emergency Drugs available, iv, oxygen Again, variety of drugs may be used Midazolam Fentanyl ketamine Titrate to effect
36
Preparation for RegionalNeuromuscular stimulator, electrodes - ultrasound
availableSurface electrode - dressing if catheter Skin prep - local anesthetic for
skin Local anesthetic for skin infiltration - glovesLocal anesthetic for nerve block Nerve stimulating needle for block
37
Regional Setup
38
Preparing for Spinal/EpiduralCAS monitors, iv, oxygen may be requiredEmergency drugs available -
skin prepPrepackaged trays - traysLocal anesthetic/ opiod for injection - local
anestheticReady to assist with patient positioning
39
Preparing for General Anesthesia CAS monitors iv fluids Machine checked Other lines as necessary Emergency drugs ready ( Drugs for case ready ) Any other lines, procedures, equipment ready if anticipated
40
Preparing for General Anesthesia SuctionOxygen LaryngoscopeETTStyletConsider Airway and location of A/W backup
equipment
41
How to manage emergenciesAnaphylaxis
42
Emergency Situation- AnaphylaxisABC’sFluid resuscitationLarge bore iv access availableEpinephrine titrated to response start at 10 mcg, escalate dose as required, 50-100mcg if hypotensive, 1mg ACLS dose
43
Emergency Situation- AnaphylaxisH1 blocker Benadryl 50mgCorticosteroid Hydrocortisone 50-100mgStop inciting allergen exposure
44
How to manage emergenciesCardiovascular collapse
45
Emergency Situations- Cardiovascular EventsABC’sACLS protocolsResponses dictated by clinical scenarioCrash cart availableEnsure CPR started
46
How to manage emergenciesIncreased ICP
47
Emergency Situations- Increased ICP
Head of bed 30 degrees elevatedEnsure adequate cerebral venous drainageGeneral goals:Avoid hypoxemiaAvoid hypotension/ maintain cerebral perfusion CPP= MAP - ICPAvoid abrupt hypertension
48
Emergency Situations- Increased ICPPharmacologic measures to lower ICPModerate hyperventilation pCO2 30-35,
(short term)Mannitol 0.5-1g/kg through 50 micron filterLasix 0.5mg/kg
49
How to manage emergenciesMalignant Hyperthermia
50
How to manage emergenciesMalignant hyperthermiaHypermetabolic disorder of skeletal muscle Intracellular hypercalcemia in muscle activates
metabolic pathwaysEnergy depletion, acidosis, membrane destruction, cell
deathHeritable component, not invariably present by family
historyHallmark- hypercarbia, tachycardia, tachypnea,
hyperthermia, rigidity, arrhythmias, hyperkalemia, renal failure, DIC, death
51
Emergency Situations- Malignant Hyperthermia ABC’s Ensure MH crisis issued - MH cartStop triggering agents - hyperventilate,
100% O2, Finish case ASAP high flowsDantrolene 2.5mg/kg, repeat q5min prn until
10mg/kg (20mg mix with 60ml sterile H2O
52
Emergency Situations- Malignant HyperthermiaArterial line- blood work and blood gassesBegin cooling patientTreat supportivelyPatient will need ongoing treatment in ICU
53
Determine case requirements for suction; such as: Airway suction Gastric suction Thoracic suction Surgical suction Post-surgical wound drainage
54
SuctionAirwayHave suction ready as part of any inductionAttached to bronchoscopy portGastricBowel obstructions- low intermittent suction
55
Cell SaverCell saverIntraoperative blood salvageAnticoagulate salvaged blood as it leaves the
surgical fieldSeparates rbc’s from other components and
debrisWashes the rbc’s for return to patient
56
Cell SaverUseful for procedures with anticipated
significant blood lossReduce the use of autologous rbc transfusionContraindications:infection - malignant cellsContamination with urine, bowel contents,
amniotic fluid
57
Cell saverComplicationsDilutional coagulopathyReinfusion of contaminants- fat, leukocytes,
red blood cell stroma, air, free hemoglobin, heparin, bacteria, debris from surgical field
58
The Anesthesia Machine
59
High Intermediate
Low Pressure Circuit
Anesthesia Workstation
High pressure circuitCylinders including N2O, O2 & AirO2 2200psi -> 50 psiN20 750 psi -> 50 psiDecreased through pressure regulators
60
High Pressure SystemReceives gasses from the high
pressure E cylinders attached to the back of the anesthesia machine (2200 psig for O2, 745 psig for N2O)
Consists of:Hanger Yolk (reserve gas cylinder
holder)Check valve (prevent reverse flow of
gas)Cylinder Pressure Indicator (Gauge)Pressure Reducing Device (Regulator)
Usually not used, unless pipeline gas supply is off6161
61
E Size Compressed Gas CylindersCylinder Cylinder CharacteristicCharacteristicss
OxygeOxygenn
Nitrous Nitrous OxideOxide
Carbon Carbon DioxideDioxide
AirAir
ColorColor White White (green)(green)
BlueBlue GrayGray Black/Black/White White (yellow)(yellow)
StateState GasGas Liquid and Liquid and gasgas
Liquid and Liquid and gasgas
GasGas
Contents (L)Contents (L) 625625 15901590 15901590 625625
Empty Weight Empty Weight (kg)(kg)
5.905.90 5.905.90 5.905.90 5.905.90
Full Weight Full Weight (kg)(kg)
6.766.76 8.808.80 8.908.90
Pressure Full Pressure Full (psig)(psig)
20002000 750750 838838 1800180062
Example ½ full E cylinder, 30 L gas, at 10 L/min, approximately 30 min of oxygen available
Hanger Yolk orients and supports
the cylinder, providing a gas-tight seal and ensuring a unidirectional gas flow into the machine
Index pins: Pin Index Safety System (PISS) is gas specific prevents accidental rearrangement of cylinders (e.g.. switching O2 and N2O)
63
Anesthesia WorkstationIntermediate pressure circuit
Includes pipeline O2 and N20 at 50-55psiExtends to flow control valves
64
Intermediate Pressure SystemReceives gasses from the
regulator or the hospital pipeline at pressures of 40-55 psig
Consists of: Pipeline inlet connections Pipeline pressure indicators Piping Gas power outlet Master switch Oxygen pressure failure
devices Oxygen flush Additional reducing devices Flow control valves
65
Pipeline Inlet ConnectionsN2O and O2, usually have
air and suction tooInlets are non-
interchangeable due to specific threading as per the Diameter Index Safety System (DISS)
Each inlet must contain a check valve to prevent reverse flow (similar to the cylinder yolk)
66
Low Pressure SystemExtends from the flow control valves to the
common gas outletConsists of:
Flow metersVaporizer mounting deviceCheck valveCommon gas outlet
67
Anesthesia Workstation
Cylinder supply source is back-up if pipeline supply fails
Fail-safe valve located downstream from N2O supply sources
Interface between O2 & N20 with proportioning system
Prevent delivery of hypoxic gas mixtures
68
Anesthesia WorkstationHigh priority alarm- if O2 supply pressure is less than a
critical pressure (<30psi)Regulated flow enters low pressure circuit with
adjustments in flowmetersGas mixture travels through a common manifold directed to
vaporizerPrecise amounts of inhaled anesthetics added, controlled
by dial flow
69
Anesthesia WorkstationFresh gas flow with added anesthetic vapor travel to
common gas outletDatex-Ohmeda have one-way check valves between
vaporizer and common gas outletPrevent back flow into the vaporizer during PPVMinimize effects of downstream intermittent pressure
fluctuations on inhaled anesthetic concentrationsOne-way check valve influences preoperative leak test
70
Pipeline Supply SourceCritical errors have occurred if incorrect
supply attached to machinesPipeline inlet fittings are gas specific with
threaded fittingsDiameter Index Safety System (DISS)If pipeline crossover suspected: turn on
back-up O2 cylinder Pipeline supply must then be disconnected
71
Cylinder Supply SourceE cylindersPin Index Safety SystemPressure reducing valve downstreamIf not turned off, will be preferentially usedVolume of gas remaining in the cylinder is
proportional to cylinder pressure
72
Oxygen supply pressure failure safety deviceDesigned to not allow hypoxic mixture
deliveryAlarm sounds if oxygen pressure fallsO2 linked with delivery of other gasses to be
oxygen dependentIf O2 pressure falls, other gas delivery falls
73
FlowmetersIndicator float position is where upward force
from gas flow equals downward force on float from gravity
O2 flow knob physically different from other gas knobs
N2O and O2 interfaced mechanically/ pneumatically, maximum 3:1 ratio
Oxygen flowmeter located downstream from other flowmeters in case of a leak
74
Limitations of Proportioning SystemsMachines equipped with proportioning
systems can still deliver a hypoxic mixture under the following conditions: Wrong supply gas Defective pneumatics or mechanics (e.g.. The Link-25
depends on a properly functioning second stage regulator) Leak downstream (e.g.. Broken oxygen flow tube) Inert gas administration: Proportioning systems generally
link only N2O and O2In general, an oxygen analyzer is the only
machine safety device that can detect these problems (gas sampling done at the Y-piece of the patient circuit)
75
Oxygen Flush ValveDirect communication with high pressure and
low pressure circuitEnters circuit downstream from vaporizers
and from machine outlet check valve100% O2 at 35-75 L/min (50 psi)Potential problems: barotrauma, decreasing
volatile anesthetic concentration, awareness
76
Oxygen Flush Valve (O2+)Receives O2 from pipeline
inlet or cylinder reducing device and directs high, unmetered flow directly to the common gas outlet (downstream of the vaporizer)
Machine standard requires that the flow be between 35 and 75 L/min
The ability to provide jet ventilation via the O2 flush valve is presence of a check valve between the vaporizer and the O2 flush valve (otherwise some flow would be wasted retrograde)
77
Vaporizers Instrument designed to change a liquid anesthetic agent
into its vapor and add a controlled amount of this vapor to the fresh gas flow
Important that each volatile anesthetic is in type specific vaporizer
Physical properties of volatile anesthetics If incorrectly filled with inappropriate anesthetic, resulting
output drastically changes
78
VaporizersVariable bypass- regulating anesthetic agent
outputConcentration control dial determines ratio
of flow through the bypass chamber and enters the vaporizer inlet
Gas channeled through the vaporizing chamber flows over the liquid anesthetic and becomes saturated with vapor
Flow over- method of vaporization
79
VaporizersTemperature compensated- maintains a
constant vaporizer output over a wide range of operating temperatures
Agent specificIf vaporizer is overfilled or tilted, liquid
anesthetic can spill into the bypass chamberFinal concentration of inhaled anesthetic is
the ratio of the flow of the inhaled anesthetic to the total gas flow
80
Generic Bypass VaporizerFlow from the
flowmeters enters the inlet of the vaporizer
The function of the concentration control valve is to regulate the amount of flow through the bypass and vaporizing chambers
Splitting Ratio = flow though vaporizing chamber/flow through
bypass chamber
Examples include the Tec 3, Tec 4, Tec 5 and the Drager 19.1
81
Vaporizers- safety featuresAgent- specific, keyed filling devicesOverfilling minimized because the filler port
is located at the maximum safe liquid levelFirmly secured to a vaporizer manifoldInterlock system to prevent administration of
>1 anesthetic agent
82
Desflurane’s Tec 6 VaporizerBecause of physical properties of Desflurane, supplying it
in a conventional vaporizer would lead to excessive cooling of the vaporizer
Vapor pressure is much higher than other volatile anesthetics
Much less potent (higher MAC)Would vaporize many more volumes of Desflurane than
other agentsTec 6 electrically heated and vaporized
83
Tec-6 VaporizerElectronically heated
and pressurized to achieve controlled vaporization of desflurane
2 independent circuits (fresh gas and vaporizer)
Vaporizer output is controlled by adjusting the concentration control valve (R2)
Pressure in the two limbs is equalized by the pressure regulating valve
84
Desflurane’s Tec 6 VaporizerEssentially a dual gas blenderBy controlling the dial, the operator controls
a variable restrictor valve
85
The Circuit: Circle SystemSo-called because the
components are arranged in a circular manner
Arrangement is variable, but to prevent re-breathing of CO2, the following rules must be followed: Unidirectional valves
between the patient and the reservoir bag
Fresh-gas-flow cannot enter the circuit between the expiratory valve and the patient
Adjustable pressure-limiting valve (APL) cannot be located between the patient and the inspiratory valve
86
Circle Breathing System Prevents rebreathing of CO2 by use of CO2 absorbents
Allows partial rebreathing of other exhaled gasses Components: Fresh gas inflow source - CO2 absorbent Inspiratory and expiratory unidirectional valves - reservoir
bag Adjustable Pressure Limiting (APL) valve - Y-piece
connector
87
Circle Breathing SystemUnidirectional flow Maintenance of relatively stabile inspired gas
concentrationsConservation of respiratory moisture and
heatPrevention of OR pollutionDisadvantage is- many possible sites for
misconnections and leaks
88
The Adjustable Pressure Limiting (APL) Valve User adjustable valve that releases
gases to the scavenging system and is intended to provide control of the pressure in the breathing system
Increased pressure in the breathing system (from patient) pushes the diaphragm off its seat venting the excess gas into the scavenging system
The control knob controls the position of the diaphragm
Bag-mask Ventilation: Valve is usually left partially open. During inspiration the bag is squeezed pushing gas into the inspiratory limb until the pressure relief is reached, opening the APL valve. At this point the additional volume the patient receives is determined by the relative resistances to flow exerted by the patient and the APL valve
Mechanical Ventilation: The APL valve is excluded from the circuit when the selector switch is changed from manual to automatic ventilation
89
CO2 absorber2 clear plastic canisters arranged in seriesSoda lime, Baralyme and calcium hydroxide limeSoda lime- calcium hydroxide, water, sodium hydroxide and
potassium hydroxide, silicaCO2 + H2O <-> H2CO3H2CO3 + 2NaOH (KOH) <-> Na2CO3 (K2CO3)+2H20 +
heatNa2CO3 (K2CO3) + Ca(OH)2 <->CaCO3 +2NaOH (KOH)
90
CO2 AbsorberpH indicator added to assess absorbentChanges to violet color when pH of the
absorbent decreases as a result of CO2 absorption
Indicates absorptive capacity of material has been consumed
91
Scavenging System Collection and subsequent removal of waste anesthetic gases
from the operating room Minimizes OR pollution by removing excess gasses National Institute for Occupational Safety and Health (NIOSH)
standards 2ppm for halogenated agent alone 25 ppm for N2O Halogenated with N20 0.5 ppm
92
Scavenging Systems Scavenging Interface:
Protects the breathing circuit or ventilator from excessive positive or negative pressure. There are 2 kinds of scavenging interfaces: Open: Contains no valves and is
open to the atmosphere allowing both positive and negative pressure relief
Closed: Communicates with the atmosphere through valves
Gas Disposal Assembly: This assembly ultimately eliminates the waste gas. There are 2 kinds of gas disposal assemblies: Passive: Uses the pressure of
the waste gas itself to produce flow through the system
Active: Uses a central vacuum to induce flow in the system, moving the waste gas along. A negative pressure relief valve is mandatory (in addition to positive pressure relief) 93
Scavenging System
Adds to OR pollution:Failure to turn off gas flow at end of casePoorly fitting masks, flushing the circuitFilling vaporizersOther circuit types which are difficult to
scavenge
94
Scavenging SystemActive or passiveActive- uses central evacuation system to
eliminate waste gasesPassive- pressure of waste gas itself produces
flowWaste anesthetic gases are vented through
the APL valve or through the ventilator relief valve
95
Scavenging SystemPotential problems:Obstruction- excessive positive pressure in
the breathing circuit and barotraumaExcessive vacuum- negative pressures within
the breathing circuit
96
Generic Ascending Bellows VentilatorBellows physically separates
the driving gas circuit from the patient gas circuit
During the inspiratory phase the driving gas enters the bellow chamber resulting in: Compression of bellows
delivering the anesthetic gases within the bellows to the patient
Closure of the overflow valve, preventing anesthetic gas from escaping into the scavenging system
During the expiratory phase the driving gas exits the bellows chamber. Exhaled gas fills the bellows Excess gas opens the
overflow valve (PEEP of 2-3 cmH2O) allowing scavenging of excess gases to occur
97
Machine CheckAnesthesia Apparatus Checkout Recommendations,
FDA. 1993. Categories of check:
• Emergency ventilation equipment - high pressure system
• Low-Pressure system - low pressure system
• Scavenging system - breathing system
• Monitors - final position
• Manual and automatic ventilation system
• Final Position
98
Checking Anesthesia Machines
99
Preoperative Checklist- High Pressure SystemCheck O2 cylinder supply -open cylinder and verify at least ½ full
-close cylinderCheck central Pipeline Supplies
- check connections and pipeline gages
100
Preoperative Checklist- Low Pressure SystemCheck initial status of low pressure system
- close flow control valves and turn vaporizers off - check fill level and tighten vaporizer’s filler cap
• Perform Leak Check - machine master switch and flow control valves OFF - attach suction bulb to common gas outlet - squeeze bulb until fully collapsed - verify bulb fully collapsed > 10 seconds - check same for each vaporizer
101
Low Pressure Circuit Leak TestChecks the integrity of the anesthesia machine from the
flow control valves to the common outlet (e.g. leaks at flow tubes, O-rings, vaporizer)
Two types of leak test (depending on presence or absence of check valve) Oxygen Flush Positive-Pressure Leak Test: Only used in machines
without check valves; basically just pressurize the low pressure circuit using the O2+ flush valve and look for leak
Negative Pressure Leak Test: Used in machines with or without check valves (i.e. Ohmeda). Attach suction bulb to common gas outlet, squeeze repeatedly until fully collapsed and ensure that it remains collapsed for 10 seconds. Will detect leaks as small as 30 ml/min.
102
Preoperative Checklist- Low Pressure SystemTurn on Machine Master SwitchTest flowmeters - adjust flow off all gasses checking for
smooth operation of floats and undamaged flow tubes
- attempt to create a hypoxic N2O/O2 mixture and verify correct changes in flow
103
Preoperative Checklist- Scavenging SystemAdjust and check scavenging system - ensure proper connections between
scavenging system and APL valve and ventilator relief valve
- adjust waste gas vacuum - fully open APL valve and occlude Y-piece
104
Preoperative Checklist- Scavenging System - with minimum flow, allow scavenger
reservoir bag to collapse completely and verify that absorber pressure gauge reads zero
- with O2 flush activated, allow scavenger reservoir bag to distend full, and verify that absorber pressure gauge reads <10 cm H2O
105
Preoperative Checklist- Breathing System Calibrate O2 monitor - ensure monitor reads 21% on room air
- verify low O2 alarm is enabled and functioning
- reinstall sensor in circuit and flush breathing system with O2
- verify that monitor now reads > 90%
106
Preoperative Checklist- Breathing SystemCheck Initial Status of Breathing System
- set switch to “bag” mode- check that circuit is complete,
undamaged and unobstructed- verify that CO2 absorbent is adequate- install breathing circuit accessory
equipment to be used during case (HME)
107
Preoperative Checklist- Breathing System Perform Leak Check of the Breathing System
- Set all gas flows to zero- Close APL valve and occlude Y-piece- Pressurize breathing system to 30 cmH2O
with O2 flush- Ensure that pressure remains fixed >
10seconds- Open APL valve and ensure pressure
decreases
108
Preoperative Checklist- Manual and Automatic Ventilation Systems Test Ventilation systems and unidirectional valves
-place a second breathing bag on Y-piece-switch on automatic ventilation-turn ventilator on and fill bellows and breathing
bag with O2 flush-set O2 flow to minimum, other gasses off
109
Preoperative Checklist- Manual and Automatic Ventilation Systems
- verify that during inspiration bellows deliver appropriate TV and that during expiration bellows fill completely
- set fresh gas flow to approximately 5 L/min
-Verify ventilator bellows and simulated lungs fill and empty appropriately without sustained pressure and end expiration
-Check for proper functioning of unidirectional valves
110
Preoperative Checklist- Manual and Automatic Ventilation Systems
-switch to bag/APL mode- Ventilate manually and assure
inflation and deflation of artificial lungs and appropriate feel of system resistance and compliance
- Remove second breathing bag from Y-piece
111
Preoperative Checklist- MonitorsCheck, calibrate and/or set alarm limits of all
monitors- Capnometry- O2 analyzer- Pressure monitor with high and low
A/W pressure alarms- Pulse oximeter- Respiratory volume monitor
112
Preoperative Checklist- Final Position of MachineCheck final status of machine
- vaporizers off - APL valve open
- selector switch to “bag”- all flowmeters to zero/minimum- patient suction level adequate- breathing system ready to use
113
Oxygen Analyzer Calibrationonly machine safety device that evaluates the
integrity of the the low-pressure circuit continuously
Other machine safety devices are upstream from flow control valves
Expose O2 concentration sensing element to room air for calibration to 21%
114
The Virtual Anesthesia Machinehttp://vam.anest.ufl.edu/
115