DESIGNED IN CALIFORNIA ASSEMBLED IN CALIFORNIA AND CZECH REPUBLIC 11751 MARKON DRIVE • GARDEN GROVE, CA 92841 • 714.895.4344 • WWW.SENSOREX.COM © Sensorex Corporation. All rights reserved. In the interest of improving and updating its equipment, Sensorex reserves the right to alter specifications to equipment at any time. Sensor Control 2 Software (LabView Version) Installation and Use Instructions

Sensor Control 2 Software (LabView Version) Installation and ......Sensor Control 2 Software (LabView Version) Installation and Use Instructions DESIGED I CALIFORIA ASSEMBLED I CALIFORIA

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

  • DESIGNED IN CALIFORNIAASSEMBLED IN CALIFORNIA AND CZECH REPUBLIC

    11751 MARKON DRIVE • GARDEN GROVE, CA 92841 • 714.895.4344 • WWW.SENSOREX.COM© Sensorex Corporation. All rights reserved. In the interest of improving and updating its equipment,

    Sensorex reserves the right to alter specifications to equipment at any time.

    Sensor Control 2Software (LabView Version)

    Installation and Use Instructions

  • DESIGNED IN CALIFORNIAASSEMBLED IN CALIFORNIA AND CZECH REPUBLIC

    11751 MARKON DRIVE • GARDEN GROVE, CA 92841 • 714.895.4344 • WWW.SENSOREX.COM© Sensorex Corporation. All rights reserved. In the interest of improving and updating its equipment,

    Sensorex reserves the right to alter specifications to equipment at any time.

    Table of ContentsDownload Files and Load Interface ........................................................................ 1Sensor Connection .................................................................................................. 1

    Options for Connecting Sensor to Computer USB Port: ....................................................1Smart Sensor Interface ............................................................................................ 1

    Getting Started .................................................................................................................1Calibrate pH ......................................................................................................................5Calibrate Contacting and Toroidal Conductivity Sensors .................................................6Calibrate Dissolved Oxygen (DO) Sensors .......................................................................7Calibrate Free Chlorine Sensors ........................................................................................8Measurement and Log Tab ...............................................................................................8Advanced Tab ...................................................................................................................10Help Tab ............................................................................................................................10

  • 1

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    Download Files and Load InterfaceNote that your computer must be connected to the Internet during the software installation process, as it must have access to the latest LabView drivers. These instructions only apply to the Microsoft Windows OS.

    1. See link for files:

    2. Unzip files.

    3. Load: Smart_Sensor_User_Interface_D installer. Files will write to C:\Program Files(x86)\Smart_Sensor_User_Interface\D.

    4. Load: Smart_Sensor_User_Interface_D.

    Sensor Connection

    Options for Connecting Sensor to Computer USB Port:

    1. Sensorex CALBOX-MB

    Uses the power from your computer so you will simply match wire colors from sensor cable to CALBOX-MB terminals with corresponding color names.

    2. Modbus485 to USB Converter

    You will also need to connect the sensor to 12VDC power using this connection type. Wiring is as follows:

    Red = 12VDC +Black = GNDWhiteWhite = Modbus AGreen = Modbus B

    Smart Sensor Interface

    Getting Started

    1. Open the Sensorex Smart Sensor Interface. The screen will appear as shown in Figure 1.

  • 2

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    2. Go into Device Manager (in Windows 10, Winkey + S and search “Device Manager”). See Figure 2.

    3. Click on Ports. The USB Serial Port (COMX) will be listed (X is the number after COM listed on your machine).

    Figure 1: Smart Sensor User Interface

    Figure 2: Device Manager

  • 3

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    4. Enter COMX (X will be the number your machine displays with COM) into the Serial COM Port box on the Setup Screen. See Figure 4.

    Figure 3: USB Serial Port (COM3)

    Figure 4: Enter COMX Value in Setup Screen > Serial COM Port

  • 4

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    5. Once this is completed, click the white arrow underneath the toolbar. The arrow will change to a black arrow. The program will then be running. See Figure 5.

    Figure 5: Click White Arrow - Black Arrow Appear and Program Runs

    6. Wait... loading sensor information... will appear in the Wait Message box. See Figure 6.

    Figure 6: Wait Message

  • 5

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    7. If the sensor is recognized, the message will say OK... and the window will disappear.

    8. Next, click on the Calibration tab at the top of the screen. The screen will appear as shown in Figure 7 for pH. The following sections provide specific calibration instructions. Choose the one relevant to your application.

    Figure 7: Calibration Tab

    Calibrate pH

    1. Place electrode in pH4 or low calibration buffer.

    2. Enter the value of the low buffer into the Enter Low Cal Value box in the Sensor Calibration window. See Figure 8 .

    Figure 8: Enter Value of Low Buffer in Enter Low Cal Value (Sensor Calibration Window)

  • 6

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    3. Click Low Cal Point button.

    4. A message will flash in the Cal Message window: When reading is stable, click off the Low Cal Point button.

    5. Click off the Low Cal Point button.

    6. The Cal Message window will say it is averaging the reading before displaying: Low Point Calibration Finished.

    7. Rinse electrode in deionized or distilled water.

    8. Place electrode in High Calibration buffer (pH 7.00 or pH 10 are most common).

    9. Repeat Steps 4 -6.

    10. The Cal Message window will show: High Point Calibration finished. Click Save button to save the calibration data into the sensor memory. Click the Save button.

    11. The Cal Message window will show: Saving Calibration Data into Sensor Memory.

    12. The next message will show: Calibration data saved and Read back the calibration log.

    13. The saved data will then be entered into the table on the bottom of the screen in the Recent Calibration Data section’s top position, Latest Time. Note that the software will only save the last three calibration data sets.

    Calibrate Contacting and Toroidal Conductivity Sensors

    1. Place sensor in air to calibrate low calibration buffer reading (0us).

    2. Enter the value of the low buffer into the window.

    3. Click the Low Cal Point button.

    4. The following message will flash in the Cal Message window: When reading is stable, click off the Low Cal Point button.

    5. Click off the Low Cal Point button.

    6. The Cal Message window will say it is averaging the reading, then: Low Point Calibration Finished.

    7. Rinse electrode in deionized or distilled water.

    8. Place electrode in High Calibration solution.

  • 7

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    9. Repeat Steps 4-6 above.

    10. The Cal Message window will show: High Point Calibration finished. Click Save button to save the calibration data into the sensor memory. Click the Save button.

    11. The Cal Message window will display: Saving Calibration Data into Sensor memory.

    12. The next message will say: Calibration data saved and Read back the calibration log.

    13. The saved data will then be entered into the table on the bottom of the Recent Calibration Data screen in the top position labeled Latest Time. Note that the software will only save the last 3 calibration data sets.

    Calibrate Dissolved Oxygen (DO) Sensors

    1. Place sensor in Zero Solution to calibrate low calibration buffer reading (

  • 8

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    Calibrate Free Chlorine Sensors

    1. Place sensor in air to calibrate low calibration buffer reading (0 ppm).

    2. Enter the value of the low buffer into the window (0 ppm).

    3. Click the Low Cal Point button.

    4. The following message will flash in the Cal Message window: When reading is stable, click off the Low Cal Point button.

    5. Click off the Low Cal Point button.

    6. The Cal Message window will say it is averaging the reading, then Low Point Calibration Finished.

    7. Rinse the sensor in deionized or distilled water. Dry off.

    8. Pour calibration solution 20ppm into a beaker or cup. Insert sensor for high point calibration value.

    9. Repeat Steps 4 -6 above.

    10. The Cal Message window will show: High Point Calibration finished. Click Save button to save the calibration data into the sensor memory. Click the Save button.

    11. The Cal Message window will display: Saving Calibration Data into Sensor memory.

    12. The next messages will display: Calibration data saved and Read back the calibration log.

    13. The saved data will then be entered into the table on the bottom of the Recent Calibration Data screen in the top position labeled Latest Time. Note that the software will only save the last three calibration data sets.

    Measurement and Log Tab

    1. Click on the Measurement and Log tab. The screen should appear as shown in Figure 9.

  • 9

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    Figure 9: Measurement and Log Tab

    2. Click the Start/Stop Measurement button.

    3. In Plot Display Selection (seen to the right of Start/Stop Measurement), select Probe Value.

    4. The data will display on the graph. The X and Y axes will auto-range.

    5. Record the data by:

    a. Create a folder destination in the C:\ drive with whatever name deemed appropriate (for example, Test Folder).

    b. Click the Open Folder icon. See Figure 10. Choose the folder labeled Test.

    c. Type a filename into the File Path Control input box. See Figure 10.

    Figure 10: Record Measurement and Log Data

  • 10

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    Advanced Tab

    The Advanced tab contains information to read and write the following parameters:

    • Modbus ID Setup: 240 is the default address.• Communication Baud Rate: 19200 is the default value.• Communication Serial Format: 8n1 is the default value.• Operation Mode:

    – Free Chlorine Range – Dissolved Oxygen Range – 4-20mA Scale Setup – Reset Button

    Help Tab

    The Help tab covers operating tips and how-to instructions for the Setup, Calibration, Measurement and Log, and Advanced tabs. See Figure 12.

    Figure 11: Advanced Tab

  • 11

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    Setup

    1. The factory default setup for every Smart Sensor:

    • Modbus ID: 240• Baud Rate:• Parity: None• Data Bit: 8• Stop Bits: 1

    2. Under Serial COM Port, use the pull-down menu to select Refresh COM Port/VISA. Select the COM port that the

    sensor has been plugged into.

    3. Click the white arrow on the left corner (found under Edit) to run the program. The other two icons on the right of the Run icon are Circulate and Run and Stop. See Figure 13.

    Figure 12: Help Tab

    Figure 13: White Arrow Changes to Black Arrow

  • 12

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    4. If the screen shows the previous default setup values, then the right COM port has been selected. Otherwise, select a different COM port and repeat Step 3.

    5. You can choose to add a label to your sensor (maximum 12 characters) for identification. Move the cursor into the user label slot, click it, and type in the desired label. Then click on the Write button. The Read User Label space will show your label.

    Calibration

    1. To calibrate the sensor, you will need two standard calibration solutions corresponding to the measurement. Follow the onscreen messages in the Cal Message box to conduct the calibration.

    2. After the two-point calibration is complete, click the Save Calibration button. The previous three calibration values and time (YYYYMMDDHHMM format) will be updated with the newest calibration values.

    3. To conduct the temperature calibration, enter the reference standard value in Celsius. Click Write Temperature for the sensor to accept the temperature calibration value.

    Measurement and Log

    1. Click the Start/Stop Measurement button to display the measurement.

    2. Select Plot Display Selection for the chart to display in one of the three parameters.

    3. Create the folder on the computer before logging. Move the cursor into the file name slot and enter the file name.

    4. When the measurement is running, click the Start/Stop Record button to start logging data.

    5. The default value is one second/sample. This may be changed as needed.

    6. Move the cursor to the scale number of the plot chart axis. Change the value as needed. See Figure 14.

  • 13

    Sensor Control 2 Software (LabView Version) Installation and Use Instructions

    Form: ??????-Rev A [Rev: 2020-07-07] © 2020 Sensorex Corporation

    7. When you are ready to stop the measurement, click Start/Stop Measurement. Be advised that the user must stop the measurement before doing any other task. Modbus allows for only one command at a time.

    Figure 14: Plot Chart Axis