SEPPIM S.A.S
Zone industrielle
61500 SEES FRANCE
: +33 (0)2 33 81 21 00 : +33 (0)2 33 28 77 81 Version: CAT-1 May 2009
APPLICATIONS
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
2/123
AA-PROXS/EL80-CAT-1
BIOCHEMISTRY
L
ACID PHOSPHATASE PACI-1 PRO LACTATE LACT-1 PRO
ALBUMIN ALBU-1 CE LDH-L SL Two-reagent procedure LLSLbi-1 PRO
ALP(DEA)SL One-reagent procedure PASLmono-1 CE LDH-P 4+1 SL One reagent procedure LDSL4+1mono-1 CE
ALP(DEA)SL Two-reagent procedure PASLbi-1 CE LDH-P 4+1 SL Two reagent procedure LDSL4+1bi-1 CE
ALP(DEA) PALC-1 PRO LDH-P LDHP-1 PRO
ALT / GPT 4+1 SL One-reagent procedure ALSL4+1mono-1 CE
ALT / GPT 4+1 SL Two-reagent procedure ALSL4+1bi-1 CE
ALT / GPT ALAT-1 PRO MAMYLASE SL AMSL-1 CE MAGNESIUM CALMAGITE One-reagent procedure MAGNmono-1 CE
AST / GOT 4+1 SL One-reagent procedure ASSL4+1mono-1 CE MAGNESIUM CALMAGITE Two-reagent procedure MAGNbi-1 CE
AST / GOT 4+1 SL Two-reagent procedure ASSL4+1bi-1 CE MICROPROTEIN PRTP-1 CE
AST / GOT ASAT-1 PRO
B P
BILIRUBIN TOTAL 4+1 BITO-1 CE PHOSPHORUS PHOS-1 CE
BILIRUBIN DIRECT 4+1 BIDI-1 CE
C T
CALCIUM ARSENAZO CALA-1 CE TOTAL PROTEIN PRTB-1 CE
CALCIUM OCPC One-reagent procedure CALOmono-1 PRO TRIGLYCERIDES MONO SL NE TGMLN-1 CE
CHLORIDE CHLO-1 CE TRIGLYCERIDES TRIG-1 PRO
CHOLESTEROL SL CHSL-1 CE
CHOLESTEROL CHOL-1 PRO
CHOLESTEROL HDL SL 2G HDLL-1 CE U
CHOLESTEROL LDL SL 2G LDLL-1 CE UREA UV One-reagent procedure URSLmono-1 CE
HDL CHOLESTEROL HDLC-1 PRO UREA UV Two-reagent procedure URSLbi-1 CE
CHOLINESTERASE CHES-1 PRO UREA UV URUV-1 PROCK-MB SL One-reagent procedure CMSLmono-1 CE URIC ACID MONO SL AUML-1 CE
CK-MB CKMB-1 PRO URIC ACID SL One-reagent procedure AUSLmono-1 CE
CK NAC SL One-reagent procedure CKSLmono-1 CE URIC ACID SL Two-reagent procedure AUSLbi-1 CE
CK NAC CKNA-1 PRO URIC ACID ACUR-1 PRO
COPPER One-reagent procedure CUIVmono-1 PRO
CREATININE JAFFE One-reagent procedure CRCOmono-1 CE
CREATININE JAFFE Two-reagent procedure CRCObi-1 CE
CREATININE PAP SL Two-reagent procedure CRSLbi-1 CE
G SPECIFIC PROTEINSGAMMA GT SL One-reagent procedure GASLmono-1 CE
GAMMA GT SL Two-reagent procedure GASLbi-1 CE APOA1 IP IAPA-1 CE
GAMMA GT GAGT-1 PRO APOB IP IAPB-1 CE
GLUCOSE HK SL One-reagent procedure GHSLmono-1 CE CRP IP ICRP-1 CE
GLUCOSE HK SL Two-reagent procedure GHSLbi-1 CE HAPTOGLOBIN IP IHAP-1 CE
GLUCOSE PAP SL GPSL-1 CE HbA1c HBAC-1 CE
GLUCOSE PAP GLUP-1 PRO IgA IP IIGA-1 CE
IgG IP IIGG-1 CE
IgM IP IIGM-1 CE
I ALBUMIN IP IMAL-1 CE
IRON FERROZINE FEFR-1 CE OROSOMUCOID IP IORO-1 CE
PREALBUMIN IP IPAL-1 CE
TRANSFERRIN IP ITRF-1 CE
CE = In agreement with 98/78/EC directive, these applications have been validated according to our internal specific procedure (PO-12-105) based on Vassault protocols
1. Vassault, A., et al., Protocole de validation de techniques (Document B, stade 3). Ann. Biol. Clin ., (1986), 44, 686.
2. Vassault, A., et al., Analyses de biologie mdicale: spcifications et normes d'acceptabilit l'usage de la validation des techniques,. Ann. Biol. Clin . (1999), 57, 685.
3 Evaluation of Precision Performance of Quantitat ive Measurement Methods; Approved GuidelineSecond Edition. CLSI (NCCLS) document EP5-A2, vol 24, No. 25, August 2004.
4. Evaluation of the Linearity of the Measurement of Quantitative Procedures: a Statistical Approach; Approved Guideline. CLSI (NCCLS) document EP6-A, vol 23, No. 16, April 2003
5. Method Comparison and Bias estimation Using Patient Samples; Approved GuidelineSecond Edition. CLSI (NCCLS) document EP9-A2, vol 22, No. 19, September 2002
6. Interference Testing in Clinical Chemistry; Approved GuidelineSecond Edition. CLSI (NCCLS) document EP7-A2, vol 25, No. 27, November 2005
PR0 = PROPOSAL: These applications have been established theoretically. It is advised to check it.
2/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
3/123
Ref : PACI-0030 18 x 3 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name PACI Allowed Yes
Code Predilution No 0
Mode Reference values
Wavelength 405 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 0 - 40 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 307 0 106 sec
Sample
Sample blank No Alinearity 10 %normal volume rerun volume* Factor 823.00
20.0 L 15.0 l Absorbance limits
Reagents Absorbance 0.000 - 3.000 Abs
Reagent blank Yes Repeats 3 R. Absorbance 0.000 - 3.000 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) R. Abs. deviation 3.000 Abs
200 L 205 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
* : For information only
ACID PHOSPHATASETOTAL AND PROSTATIC
Testpa
rameters1
Testparameters2
No
ACID PHOSPHATASE
APPLICATION Selectra ProXS / Flexor EL80 PROPOSAL
XXX: Enter data by the user
ACID PHOSPHATASE
Kinetic
XXX
XXX
XXX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-PACI-1
3/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
4/123
Ref : ALBU-0250 12 x 20 mL
ALBU-0600 2 x 125 mL
ALBU-0700 4 x 250 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name ALB Allowed Yes
Code 100 Predilution No 0
Mode Reference values
Wavelength 620 nm 0 Male XXX - XXX g/dL
Units g/dL Female XXX - XXX g/dL
Decimals 2 Pediatric XXX - XXX g/dL
Test repeats 1 Panic XXX - XXX g/dL
Controls
Control 1
Concentration limits 1.5 - 6 g/dL Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 g/dL
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Incubation time 4.5 min 0
Sample
Sample blank No 0 0
normal volume rerun volume* Factor 1.00
3.0 L 2.0 l Absorbance limits
Reagents Absorbance 0.100 - 2.000 Abs
Reagent blank Yes 3 R. Absorbance 0.100 - 0.300 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) 0 9.000 0
360 L 398 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 28 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 XXX 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0
Testp
arameters1
Testparameters2
FAUX
FAUX
0
Calibration
0
FAUX
FAUX
XXX
No
ALBUMIN
ALBUMIN
ALBUMIN
End point monochromatic
FAUX
XXX
XXX
XXX: Enter data by the user
FAUX
FAUX
APPLICATION Selectra ProXS / Flexor EL80
XXX
XXX
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-ALBU-1
4/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
5/123
Ref : ALBU-0250 12 x 20 mL
ALBU-0600 2 x 125 mL
ALBU-0700 4 x 250 mL
Instruction for use :
ALBUMIN
APPLICATION Selectra ProXS / Flexor EL80
0 0
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 1.5 to 6 g/dL.
Detection limi tDetermined according to SFBC protocol, the detection limit is equal to 0.03 g/dL.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 1.9%
Level 2 20 Level 2 1.8%
Level 3 20 Level 3 1.8%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific XL on 30 human sera samples.
The sample concentrations were between 1.42 and 5.96 g/dL.
The parameters of linear regression are as follows :
Correlation coefficient (r) :0.9987
Linear regression : y= 1.0334 x - 0.12 g/dL
InterferencesAccording to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin :No significant interference up to 35 mg/dL.
Conjugated bilirubin :No significant interference up to 25 mg/dL.
Haemoglobin: Positive bias from 350 mg/dL on normal human serum and from 50 mg/dL on pathological human serum.
Turbidity :No significant interference up to 600 mg/dL Triglycerides equivalent.
Other compounds may interfere (see product insert).
On board stability (refrigerated position)The reagent is stable 4 weeks on analyser without recalibrating.
3.26
4.68
FAUX
4.69 20
20
Mean (g/dL) NCV%
0.8%
0.7%
0.6%
Mean (g/dL)
2.11
3.30
2.13 20
Modification from the previous version
(05/2009)
AA-PROXS/EL80-ALBU-1
5/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
6/123
Ref : PASL-0400 2 x 62.5 mL
PASL-0420 4 x 62.5 mL
PASL-0500 5 x 125 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name PASL Allowed Yes
Code Predilution No 0
Mode Reference values
Wavelength 405 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 20 - 900 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 50 0 133 sec
Sample
Sample blank No Alinearity 10 %
normal volume rerun volume* Factor 1.00
5.0 L 3.0 l Absorbance limits
Reagents Absorbance 0.200 - 3.000 Abs
Reagent blank Yes 3 R. Absorbance 0.200 - 1.100 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) R. Abs. deviation 3.000 Abs
250 L 252 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 7 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 00 0
0 0
Tes
tparameters1
Testparameters2
FAUX
FAUX
0
Calibration
0
FAUX
XXX
APPLICATION Selectra ProXS / Flexor EL80
FAUX
ALP (DEA) SL
ALP (DEA) SLONE-REAGENT PROCEDURE
XXX
XXX
XXX
XXX: Enter data by the user
Kinetic
FAUX
XXX
No
ALP (DEA) SL
FAUX
FAUX
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-PASLmono-1
6/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
7/123
Ref : PASL-0400 2 x 62.5 mL
PASL-0420 4 x 62.5 mL
PASL-0500 5 x 125 mL
Instruction for use :
APPLICATION Selectra ProXS / Flexor EL80
ALP (DEA) SLONE-REAGENT PROCEDURE
0 0
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 20 to 900 U/L.
Detection limitDetermined according to SFBC protocol, the detection limit is equal to 5 U/L.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 4.9%
Level 2 20 Level 2 2.4%
Level 3 20 Level 3 0.9%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific XL on 30 human sera samples.
The sample concentrations were between 20 and 1041 U/L.
The parameters of linear regression are as follows :
Correlation coefficient (r) :0.9998
Linear regression : y= 1.0284 x + 1.5 U/L
InterferencesAccording to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin :No significant inteference up to 36 mg/dL.
Conjugated bilirubin :No significant interference up to 25 mg/dL.
Hemoglobin: No significant interference up to 500 mg/dL
Turbidity :No significant interference up to 614 mg/dL Triglycerides equivalent.
Other compounds may interfere (see product insert).
On board stability (refrigerated position)
The working reagent is stable 3 weeks on analyser with a new calibration every week.
33
711
FAUX
720 20
20134
Mean (U/L)
37
N
20
137
Mean (U/L)
0.6%
0.6%
1.7%
CV%
Modification from the previous version
(05/2009)
AA-PROXS/EL80-PASLmono-1
7/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
8/123
Ref : PASL-0230 4 x 25 mL
PASL-0400 2 x 62.5 mL
PASL-0420 4 x 62.5 mL
Instruction for use : PASL-0500 5 x 125 mL
General Sample types
Name Normal
Abbreviated name PASL Allowed Yes
Code 120 Predilution No 0
Mode Reference values
Wavelength 405 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 20 - 900 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 50 0 133 sec
Sample
Sample blank No Alinearity 10 %
normal volume rerun volume* Factor 1.00
5.0 L 3.0 l Absorbance limits
Reagents Absorbance 0.200 - 2.000 Abs
Reagent blank Yes 3 R. Absorbance -0.050 - 0.050 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) Substrate depletion 0.100 Abs
200 L 202 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
50 L 50 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 7 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 00 0
0 0
ALP (DEA) SLTWO-REAGENT PROCEDURE
XXX
FAUX
APPLICATION Selectra ProXS / Flexor EL80
XXX
XXX
FAUX
XXX
XXX: Enter data by the user
FAUX
FAUX
FAUX
FAUX
Calibration
0
FAUX
0
FAUX
Tes
tparameters1
Testparameters2
XXX
No
ALP (DEA) SL
ALP (DEA) SL
Kinetic
Modification from the previous version
(05/2009)
AA-PROXS/EL80-PASLbi-1
8/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
9/123
Ref : PASL-0230 4 x 25 mL
PASL-0400 2 x 62.5 mL
PASL-0420 4 x 62.5 mL
Instruction for use : PASL-0500 5 x 125 mL
ALP (DEA) SLTWO-REAGENT PROCEDURE
APPLICATION Selectra ProXS / Flexor EL80
0 0
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 20 to 900 U/L.
Detection limitDetermined according to SFBC protocol, the detection limit is equal to 4 U/L.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 5.5%
Level 2 20 Level 2 1.4%
Level 3 20 Level 3 1.0%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific XL on 30 human sera samples.
The sample concentrations were between 20 and 1041 U/L.
The parameters of linear regression are as follows :
Correlation coefficient (r) :0.9998
Linear regression : y= 1.038 x - 0.08 U/L
InterferencesAccording to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin :No significant inteference up to 36 mg/dL.
Conjugated bilirubin :No significant interference up to 25 mg/dL.
Hemoglobin: No significant interfeence up to 500 mg/dL
Turbidity :No significant interference up to 614 mg/dL Triglycerides equivalent.
Other compounds may interfere (see product insert).
On board stability (refrigerated position)
The reagents are stable 3 weeks on analyser with a new calibration every week.
N
20 34
706
Mean (U/L)
14920
CV%Mean (U/L)
40
712
147
0.4%
0.6%
1.1%
20
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-PASLbi-1
9/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
10/123
Ref : PALC-0030 20 x 3 mL
PALC-0200 12 x 20 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name PALC Allowed Yes
Code Predilution No 0
Mode Reference values
Wavelength 405 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 20 - 700 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 50 0 133 sec
Sample
Sample blank No Alinearity 10 %
normal volume rerun volume* Factor 1.00
5.0 L 3.0 l Absorbance limits
Reagents Absorbance 0.200 - 3.000 Abs
Reagent blank Yes 3 R. Absorbance 0.200 - 1.100 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) R. Abs. deviation 3.000 Abs
250 L 252 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval xxx days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0
Testp
arameters1
Testparameters2
FAUX
FAUX
0
Calibration
FAUX
FAUX
ALP(DEA)
XXX
0
APPLICATION Selectra ProXS / Flexor EL80 PROPOSAL
XXX
ALP(DEA)
Kinetic
FAUX
XXX
XXX
FAUX
XXX: Enter data by the user
FAUX
FAUX
XXX
No
ALP(DEA)
Modification from the previous version
(05/2009)
AA-PROXS/EL80-PALC-1
10/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
11/123
Ref : ALSL-0410 2 x 62.5 mL
ALSL-0430 4 x 62.5 mL
ALSL-0510 5 x 125 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name ALSL Allowed Yes
Code Predilution No 0
Mode Reference values
Wavelength 340 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 15 - 250 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 50 0 159 sec
Sample
Sample blank No Alinearity 20 %
normal volume rerun volume* Factor 1.00
30.0 L 5.0 l Absorbance limits
Reagents Absorbance 0.200 - 3.000 Abs
Reagent blank Yes 3 R. Absorbance 0.900 - 1.900 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) R. Abs. deviation 0.040 Abs
300 L 300 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 14 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 00 0
0 0
ALT / GPT 4+1 SLONE-REAGENT PROCEDURE
XXX
FAUX
APPLICATION Selectra ProXS / Flexor EL80
XXX
XXX
FAUX
XXX
XXX: Enter data by the user
FAUX
FAUX
FAUX
FAUX
Calibration
0
FAUX
0
FAUX
Tes
tparameters1
Testparameters2
XXX
No
ALT / GPT 4+1 SL
ALT / GPT 4+1 SL
Kinetic
Modification from the previous version
(05/2009)
AA-PROXS/EL80-ALSL4+1mono-1
11/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
12/123
Ref : ALSL-0410 2 x 62.5 mL
ALSL-0430 4 x 62.5 mL
ALSL-0510 5 x 125 mL
Instruction for use :
ALT / GPT 4+1 SLONE-REAGENT PROCEDURE
APPLICATION Selectra ProXS / Flexor EL80
0 0
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 15 to 250 U/L.
Detection limitDetermined according to SFBC protocol, the detection limit is equal to 1.5 U/L.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 4.6%
Level 2 20 Level 2 3.0%
Level 3 20 Level 3 2.0%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific E on 30 human sera samples.
The sample concentrations were between 4 and 257 U/L.
The parameters of linear regression are as follows :
Correlation coefficient (r) :0.9978
Linear regression : y= 0.9747 x + 1.6 U/L
InterferencesAccording to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin :Negative bias from 15 mg/dL.
Conjugated bilirubin :Negative bias from 16 mg/dL.
Turbidity :No significant interference up to 600 mg/dL Triglycerides equivalent.
Other compounds may interfere (see product insert).
Note:
Hemolysed sera should not be used since significant haemolysis may increase ALT concentrations
because of high levels of ALT in erythrocytes.
On board stability (refrigerated position)The working reagent is stable 2 weeks on analyser without recalibrating.
N
20 17
202
Mean (U/L)
5020
CV%Mean (U/L)
19
207
47
0.2%
0.6%
1.4%
20
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-ALSL4+1mono-1
12/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
13/123
Ref : ALSL-0250 8 x 25 mL
ALSL-0410 2 x 62.5 mL
ALSL-0430 4 x 62.5 mL
Instruction for use : ALSL-0455 4 x 55 mL
ALSL-0510 5 x 125 mL
General Sample types
Name Normal
Abbreviated name ALSL Allowed Yes
Code 140 Predilution No 0
Mode Reference values
Wavelength 340 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 15 - 250 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 50 0 159 sec
Sample
Sample blank No Alinearity 20 %
normal volume rerun volume* Factor 1.00
30.0 L 5.0 l Absorbance limits
Reagents Absorbance 0.100 - 3.000 Abs
Reagent blank Yes 3 R. Absorbance -0.050 - 0.100 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) Substrate depletion -0.500 Abs
240 L 240 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
60 L 60 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 28 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 00 0
0 0
FAUX
Tes
tparameters1
Testparameters2
FAUX
FAUX
0
Calibration
0
ALT / GPT 4+1 SL
XXX
FAUX
ALT / GPT 4+1 SLTWO-REAGENT PROCEDURE
XXX
XXX: Enter data by the user
Kinetic
XXX
No
ALT / GPT 4+1 SL
FAUX
FAUX
APPLICATION Selectra ProXS / Flexor EL80
FAUX
XXX
XXX
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-ALSL4+1bi-1
13/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
14/123
Ref : ALSL-0250 8 x 25 mL
ALSL-0410 2 x 62.5 mL
ALSL-0430 4 x 62.5 mL
Instruction for use : ALSL-0455 4 x 55 mL
ALSL-0510 5 x 125 mL
ALT / GPT 4+1 SLTWO-REAGENT PROCEDURE
APPLICATION Selectra ProXS / Flexor EL80
0 0
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 15 to 250 U/L.
Detection limitDetermined according to SFBC protocol, the detection limit is equal to 2 U/L.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 3.2%
Level 2 20 Level 2 2.1%
Level 3 20 Level 3 1.2%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific XL on 30 human sera samples.
The sample concentrations were between 7 and 230 U/L.
The parameters of linear regression are as follows :
Correlation coefficient (r) :0.9983
Linear regression : y= 1.0195 x - 0.3 U/L
InterferencesAccording to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin :Negative bias from 15 mg/dL.
Conjugated bilirubin :Negative bias from 16 mg/dL.
Turbidity :No significant interference up to 600 mg/dL Triglycerides equivalent.
Other compounds may interfere (see product insert).
Note:
Hemolysed sera should not be used since significant haemolysis may increase ALT concentrations
because of high levels of ALT in erythrocytes.
On board stability (refrigerated position)The reagents are stable 4 weeks on analyser without recalibrating.
FAUX
211 20
2046
Mean (U/L)
19
Mean (U/L)
20
217
N
20
0.4%
1.2%
2.8%
CV%
53
Modification from the previous version
(05/2009)
AA-PROXS/EL80-ALSL4+1bi-1
14/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
15/123
Ref : ALAT-0200 12 x 20 mL
ALAT-0400 9 x 50 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name ALAT Allowed Yes
Code Predilution No 0
Mode Reference values
Wavelength 340 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 15 - 250 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 77 0 185 sec
Sample
Sample blank No Alinearity 10 %
normal volume rerun volume* Factor 1.00
25.0 L 13.0 l Absorbance limits
Reagents Absorbance 0.200 - 2.500 Abs
Reagent blank Yes 3 R. Absorbance 0.500 - 1.900 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) R. Abs. deviation 0.050 Abs
250 L 262 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval xxx days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0 FAUX
XXX
XXX
FAUX
FAUX
XXX
No
ALT/GPT
FAUX
FAUX
ALT/GPT
XXX
0
APPLICATION Selectra ProXS / Flexor EL80 PROPOSAL
ALT/GPT
Kinetic
XXX: Enter data by the user
XXX
FAUX
Testp
arameters1
Testparameters2
FAUX
FAUX
0
Calibration
Modification from the previous version
(05/2009)
AA-PROXS/EL80-ALAT-1
15/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
16/123
Ref : AMSL-0230 6 x 20 mL
AMSL-0390 1 x 50 mL
AMSL-0400 6 x 50 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name AMSL Allowed Yes
Code 160 Predilution No 0
Mode Reference values
Wavelength 405 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 20 - 2000 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 50 0 159 sec
Sample
Sample blank No Alinearity 10 %
normal volume rerun volume* Factor 1.00
3.0 L 2.0 l Absorbance limits
Reagents Absorbance 0.015 - 2.000 Abs
Reagent blank Yes 3 R. Absorbance 0.015 - 0.800 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) R. Abs. deviation 3.000 Abs
300 L 301 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 28 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0
AMYLASE SL
XXX
FAUX
APPLICATION Selectra ProXS / Flexor EL80
XXX
XXX
FAUX
XXX
XXX: Enter data by the user
FAUX
FAUX
FAUX
FAUX
Calibration
0
FAUX
0
FAUX
Testp
arameters1
Testparameters2
XXX
No
AMYLASE SL
AMYLASE SL
Kinetic
Modification from the previous version
(05/2009)
AA-PROXS/EL80-AMSL-1
16/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
17/123
Ref : AMSL-0230 6 x 20 mL
AMSL-0390 1 x 50 mL
AMSL-0400 6 x 50 mL
Instruction for use :
AMYLASE SL
APPLICATION Selectra ProXS / Flexor EL80
0 0
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 20 to 2000 U/L.
Detection limi tDetermined according to SFBC protocol, the detection limit is equal to 2 U/L.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 2.0%
Level 2 20 Level 2 1.1%
Level 3 20 Level 3 2.3%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific XL on 30 human sera samples.
The sample concentrations were between 20 and 2193 U/L.
The parameters of linear regression are as follows :
Correlation coefficient (r) :0.9997
Linear regression : y= 1.0354 x - 0.6 U/L
InterferencesAccording to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin :Negative bias from17 mg/dL on pathological human.
No significant interference up to 35 mg/dL on normal human sera.
Conjugated bilirubin :Negative bias from 21mg/dL on pathological human.
No significant interference up to 25 mg/dL on normal human sera.
Haemoglobin: No significant interference up to 400 mg/dL.
Turbidity :No significant interference up to 600 mg/dL Triglycerides equivalent.
Other compounds may interfere (see product insert).
On board stability (refrigerated position)The reagent is stable 4 weeks on analyser without recalibrating.
N
20 52
404
Mean (U/L)
15620
CV%Mean (U/L)
52
403
154
0.6%
0.7%
1.2%
20
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-AMSL-1
17/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
18/123
Ref : ASSL-0410 2 x 62.5 mL
ASSL-0430 4 x 62.5 mL
ASSL-0510 5 x 125 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name ASSL Allowed Yes
Code Predilution No 0
Mode Reference values
Wavelength 340 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 10 - 250 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 50 0 159 sec
Sample
Sample blank No Alinearity 20 %
normal volume rerun volume* Factor 1.00
30.0 L 5.0 l Absorbance limits
Reagents Absorbance 0.800 - 3.000 Abs
Reagent blank Yes 3 R. Absorbance 0.500 - 1.900 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) R. Abs. deviation 0.040 Abs
300 L 300 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 14 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 00 0
0 0
FAUX
Tes
tparameters1
Testparameters2
FAUX
FAUX
0
Calibration
0
AST / GOT 4+1 SL
XXX
FAUX
AST / GOT 4+1 SLONE-REAGENT PROCEDURE
XXX
XXX: Enter data by the user
Kinetic
XXX
No
AST / GOT 4+1 SL
FAUX
FAUX
APPLICATION Selectra ProXS / Flexor EL80
FAUX
XXX
XXX
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-ASSL4+1mono-1
18/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
19/123
Ref : ASSL-0410 2 x 62.5 mL
ASSL-0430 4 x 62.5 mL
ASSL-0510 5 x 125 mL
Instruction for use :
AST / GOT 4+1 SLONE-REAGENT PROCEDURE
APPLICATION Selectra ProXS / Flexor EL80
0 0
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 10 to 250 U/L.
Detection limitDetermined according to SFBC protocol, the detection limit is equal to 1.5 U/L.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 2.0%
Level 2 20 Level 2 2.5%
Level 3 20 Level 3 1.5%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific E on 30 human sera samples.
The parameters of linear regression are as follows :
Correlation coefficient (r) :0.9984
Linear regression : y= 1.0289 x - 0.6 U/L
InterferencesAccording to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin :Negative bias from 15 mg/dL.
Conjugated bilirubin :Negative bias from 16 mg/dL.
Turbidity :No significant interference up to 600 mg/dL Triglycerides equivalent.
Other compounds may interfere (see product insert).
Note:
Hemolysed sera should not be used since significant haemolysis may increase AST concentrations
because of high levels of AST in erythrocytes.
On board stability (refrigerated position)
The working reagent is stable 2 weeks on analyser without recalibrating.
FAUX
182 20
2054
Mean (U/L)
19
Mean (U/L)
20
193
N
20
0.3%
2.3%
2.1%
CV%
44
Modification from the previous version
(05/2009)
AA-PROXS/EL80-ASSL4+1mono-1
19/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
20/123
Ref : ASSL-0250 8 x 25 mL
ASSL-0410 2 x 62.5 mL
ASSL-0430 4 x 62.5 mL
Instruction for use : ASSL-0455 4 x 55 mL
ASSL-0510 5 x 125 mL
General Sample types
Name Normal
Abbreviated name ASSL Allowed Yes
Code 180 Predilution No 0
Mode Reference values
Wavelength 340 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 10 - 250 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 50 0 159 sec
Sample
Sample blank No Alinearity 20 %
normal volume rerun volume* Factor 1.00
30.0 L 5.0 l Absorbance limits
Reagents Absorbance 0.100 - 3.000 Abs
Reagent blank Yes 3 R. Absorbance -0.050 - 0.100 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) Substrate depletion -0.500 Abs
240 L 240 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
60 L 60 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 28 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 00 0
0 0
AST / GOT 4+1 SLTWO-REAGENT PROCEDURE
XXX
FAUX
APPLICATION Selectra ProXS / Flexor EL80
XXX
XXX
FAUX
XXX
XXX: Enter data by the user
FAUX
FAUX
FAUX
FAUX
Calibration
0
FAUX
0
FAUX
Tes
tparameters1
Testparameters2
XXX
No
AST / GOT 4+1 SL
AST / GOT 4+1 SL
Kinetic
Modification from the previous version
(05/2009)
AA-PROXS/EL80-ASSL4+1bi-1
20/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
21/123
Ref : ASSL-0250 8 x 25 mL
ASSL-0410 2 x 62.5 mL
ASSL-0430 4 x 62.5 mL
Instruction for use : ASSL-0455 4 x 55 mL
ASSL-0510 5 x 125 mL
AST / GOT 4+1 SLTWO-REAGENT PROCEDURE
APPLICATION Selectra ProXS / Flexor EL80
0 0
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 10 to 250 U/L.
Detection limitDetermined according to SFBC protocol, the detection limit is equal to 2.5 U/L.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 2.7%
Level 2 20 Level 2 2.9%
Level 3 20 Level 3 2.0%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific XL on 31 human sera samples.
The sample concentrations were between 14 and 242 U/L.
The parameters of linear regression are as follows :
Correlation coefficient (r) :0.9945
Linear regression : y= 1.0297 x - 0.8 U/L
InterferencesAccording to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin :Negative bias from 15 mg/dL.
Conjugated bilirubin :Negative bias from 16 mg/dL.
Turbidity :No significant interference up to 600 mg/dL Triglycerides equivalent.
Other compounds may interfere (see product insert).
Note:
Hemolysed sera should not be used since significant haemolysis may increase AST concentrations
because of high levels of AST in erythrocytes.
On board stability (refrigerated position)The reagents are stable 4 weeks on analyser without recalibrating.
N
20 20
206
Mean (U/L)
4620
CV%Mean (U/L)
21
208
46
0.8%
0.9%
1.7%
20
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-ASSL4+1bi-1
21/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
22/123
Ref : ASAT-0200 12 x 20 mL
ASAT-0400 9 x 50 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name ASAT Allowed Yes
Code Predilution No 0
Mode Reference values
Wavelength 340 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 15 - 250 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 77 0 185 sec
Sample
Sample blank No Alinearity 10 %
normal volume rerun volume* Factor 1.00
25.0 L 13.0 l Absorbance limits
Reagents Absorbance 0.200 - 2.500 Abs
Reagent blank Yes 3 R. Absorbance 0.900 - 1.900 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) R. Abs. deviation 0.050 Abs
250 L 262 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval xxx days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0
Testp
arameters1
Testparameters2
FAUX
FAUX
0
Calibration
FAUX
FAUX
AST/GOT
XXX
0
APPLICATION Selectra ProXS / Flexor EL80 PROPOSAL
XXX
AST/GOT
Kinetic
FAUX
XXX
XXX
FAUX
XXX: Enter data by the user
FAUX
FAUX
XXX
No
AST/GOT
Modification from the previous version
(05/2009)
AA-PROXS/EL80-ASAT-1
22/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
23/123
Ref : BITO-0600 2 x 125 mL
BITD-0600 2 x 125 mL
BITO-0250 8 x 25 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name BITO Allowed Yes
Code 220 Predilution No 0
Mode Reference values
Wavelength 546 nm 0 Male XXX - XXX mg/dL
Units mg/dL Female XXX - XXX mg/dL
Decimals 2 Pediatric XXX - XXX mg/dL
Test repeats 1 Panic XXX - XXX mg/dL
Controls
Control 1
Concentration limits 0.3 - 20 mg/dL Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 mg/dL
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Incubation time 6.5 min 0
Sample
Sample blank No 0 0
normal volume rerun volume* Factor 1.00
30.0 L 10.0 l Absorbance limits
Reagents Absorbance -0.010 - 2.500 Abs
Reagent blank Yes 3 R. Absorbance -0.010 - 0.050 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) 0 9.000 0
240 L 240 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
60 L 60 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 14 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 00 0
0 0
BILIRUBIN TOTAL 4+1
XXX
FAUX
APPLICATION Selectra ProXS / Flexor EL80
XXX
XXX
FAUX
XXX
XXX: Enter data by the user
FAUX
FAUX
FAUX
FAUX
Calibration
0
FAUX
0
FAUX
Tes
tparameters1
Testparameters2
XXX
No
BILIRUBIN TOTAL 4+1
BILIRUBIN TOTAL 4+1
End point monochromatic
Modification from the previous version
(05/2009)
AA-PROXS/EL80-BITO-1
23/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
24/123
Ref : BITO-0600 2 x 125 mL
BITD-0600 2 x 125 mL
BITO-0250 8 x 25 mL
Instruction for use :
BILIRUBIN TOTAL 4+1
APPLICATION Selectra ProXS / Flexor EL80
0 0
PREPARATION OF REAGENTS:Position R1: Reagent 1 (Ready to use)
Position R3: Reagent 2 (Ready to use)
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 0.3 to 20 mg/dL.
Detection limit
Determined according to SFBC protocol, the detection limit is equal to 0.04 mg/dL.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 2.4%
Level 2 20 Level 2 3.1%
Level 3 20 Level 3 3.0%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific E on 30 human sera samples.
The sample concentrations were between 0.17 and 28.82 mg/dL.
The parameters of linear regression are as follows :
Correlation coefficient (r) :0.9992
Linear regression : y= 1.0084 x + 0.034 mg/dL
Interferences
According to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Turbidity :No significant interference up to 600 mg/dL Triglycerides equivalent.
Hemoglobin: No significant interference up to 500 mg/dL.
Ascorbic acid: No significant interference up to 20 mg/dL.
Other compounds may interfere (see product insert).
On board stability (refrigerated position)The reagents are stable 4 weeks on analyser with a new calibration every 2 weeks.
N
20 0.77
9.46
Mean (mg/dL)
2.7820
CV%Mean (mg/dL)
0.82
9.39
2.91
0.5%
0.4%
2.1%
20
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-BITO-1
24/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
25/123
Ref : BITD-0600 2 x 125 mL
BIDI-0600 2 x 125 mL
BIDI-0250 8 x 25 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name BIDI Allowed Yes
Code 200 Predilution No 0
Mode Reference values
Wavelength 546 nm 0 Male XXX - XXX mg/dL
Units mg/dL Female XXX - XXX mg/dL
Decimals 2 Pediatric XXX - XXX mg/dL
Test repeats 1 Panic XXX - XXX mg/dL
Controls
Control 1
Concentration limits 0.15 - 18 mg/dL Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 mg/dL
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Incubation time 6.5 min 0
Sample
Sample blank No 0 0
normal volume rerun volume* Factor 1.00
30.0 L 30.0 l Absorbance limits
Reagents Absorbance -0.010 - 2.500 Abs
Reagent blank Yes 3 R. Absorbance -0.010 - 0.010 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) 0 9.000 0
240 L 240 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
60 L 60 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 14 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 00 0
0 0
Tes
tparameters1
Testparameters2
FAUX
FAUX
0
Calibration
0
FAUX
BILIRUBIN DIRECT 4+1
BILIRUBIN DIRECT 4+1
XXX
XXX
APPLICATION Selectra ProXS / Flexor EL80
FAUX
XXX
BILIRUBIN DIRECT 4+1
FAUX
FAUX
XXX: Enter data by the user
FAUX
End point monochromatic
XXX
No
FAUX
XXX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-BIDI-1
25/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
26/123
Ref : BITD-0600 2 x 125 mL
BIDI-0600 2 x 125 mL
BIDI-0250 8 x 25 mL
Instruction for use :
BILIRUBIN DIRECT 4+1
APPLICATION Selectra ProXS / Flexor EL80
0 0
PREPARATION OF REAGENTS:Position R1: Reagent 1 (Ready to use)
Position R3: Reagent 2 (Ready to use)
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 0.15 to 18 mg/dL.
Detection limit
Determined according to SFBC protocol, the detection limit is equal to 0.02 mg/dL.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 3.5%
Level 2 20 Level 2 1.9%
Level 3 20 Level 3 2.7%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific E on 30 human sera samples.
The sample concentrations were between 0.15 and 18.4 mg/dL.
The parameters of linear regression are as follows :
Correlation coefficient (r) :0.9994
Linear regression : y= 1.0127 x - 0.014 mg/dL
Interferences
According to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Turbidity :No significant interference up to 600 mg/dL Triglycerides equivalent.
Hemoglobin: Negative bias from 100 mg/dL.
Ascorbic acid: Positive bias from 2 mg/dL.
Other compounds may interfere (see product insert).
On board stability (refrigerated position)The reagents are stable 4 weeks on analyser with a new calibration every 2 weeks.
9.35
FAUX
Mean (mg/dL)Mean (mg/dL) N
9.76 20
20
0.7320
2.63
0.85
0.8%
0.6%
0.7%
CV%
2.77
Modification from the previous version
(05/2009)
AA-PROXS/EL80-BIDI-1
26/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
27/123
Ref : CALA-0250 12 x 20 mL
CALA-0600 2 x 125 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name CALA Allowed Yes
Code 240 Predilution No 0
Mode Reference values
Wavelength 660 nm 0 Male XXX - XXX mg/dL
Units mg/dL Female XXX - XXX mg/dL
Decimals 1 Pediatric XXX - XXX mg/dL
Test repeats 1 Panic XXX - XXX mg/dL
Controls
Control 1
Concentration limits 5 - 15 mg/dL Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 mg/dL
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Incubation time 4.5 min 0
Sample
Sample blank No 0 0
normal volume rerun volume* Factor 1.00
6.0 L 6.0 l Absorbance limits
Reagents Absorbance 0.100 - 2.500 Abs
Reagent blank Yes 3 R. Absorbance 0.050 - 1.000 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) 0 9.000 0
300 L 300 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 28 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0
CALCIUM ARSENAZO
XXX
FAUX
APPLICATION Selectra ProXS / Flexor EL80
XXX
XXX
FAUX
XXX
XXX: Enter data by the user
FAUX
FAUX
FAUX
FAUX
Calibration
0
FAUX
0
FAUX
Testp
arameters1
Testparameters2
XXX
No
CALCIUM ARSENAZO
CALCIUM ARSENAZO
End point monochromatic
Modification from the previous version
(05/2009)
AA-PROXS/EL80-CALA-1
27/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
28/123
Ref : CALA-0250 12 x 20 mL
CALA-0600 2 x 125 mL
Instruction for use :
CALCIUM ARSENAZO
APPLICATION Selectra ProXS / Flexor EL80
0 0
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 5 to 15 mg/dL.
Detection limi tDetermined according to SFBC protocol, the detection limit is equal to 0.03 mg/dL.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 1.5%
Level 2 20 Level 2 1.5%
Level 3 20 Level 3 0.7%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific XL on 30 human sera samples.
The sample concentrations were between 4.4 and 14.7 mg/dL.
The parameters of linear regression are as follows :
Correlation coefficient (r) :0.9987
Linear regression : y= 1.0311 x - 0.06 mg/dL
InterferencesAccording to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin :Positive bias from 3 mg/dL.
Conjugated bilirubin :No significant interference up to 25 mg/dL.
Haemoglobin: Positive bias from 450 mg/dL.
Turbidity :Positive bias from 400 mg/dL Triglycerides equivalent.
Other compounds may interfere (see product insert).
On board stability (refrigerated position)The reagent is stable 4 weeks on analyser without recalibrating.
N
20 7.1
12.3
Mean (mg/dL)
9.420
CV%Mean (mg/dL)
7.0
12.5
9.4
0.4%
0.8%
1.0%
20
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-CALA-1
28/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
29/123
Ref : CALO-0600 2 x 125 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name CALO Allowed Yes
Code Predilution No 0
Mode Reference values
Wavelength 578 nm 620 nm Male XXX - XXX mg/dL
Units mg/dL Female XXX - XXX mg/dL
Decimals 1 Pediatric XXX - XXX mg/dL
Test repeats 1 Panic XXX - XXX mg/dL
Controls
Control 1
Concentration limits 5 - 13.5 mg/dL Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 mg/dL
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Incubation time 4.5 min 0
Sample
Sample blank No 0 0
normal volume rerun volume* Factor 1.00
3.0 L 2.0 l Absorbance limits
Reagents Absorbance -0.100 - 3.000 Abs
Reagent blank Yes 3 R. Absorbance -0.100 - 1.200 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) 0 9.000 0
300 L 301 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval xxx days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0
Testp
arameters1
Testparameters2
FAUX
FAUX
0
Calibration
0
FAUX
FAUX
CALCIUM OCPCONE-REAGENT PROCEDURE
XXX
XXX
XXX
No
CALCIUM OCPC
CALCIUM OCPC
End point bichromatic
FAUX
FAUX
XXX
APPLICATION Selectra ProXS / Flexor EL80 PROPOSAL
XXX
FAUX
XXX: Enter data by the user
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-CALOmono-1
29/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
30/123
Ref : CHLO-0250 12 x 20 mL
CHLO-0600 2 x 125 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name CHLO Allowed Yes
Code 260 Predilution No 0
Mode Reference values
Wavelength 505 nm 0 Male XXX - XXX mEq/L
Units mEq/L Female XXX - XXX mEq/L
Decimals 1 Pediatric XXX - XXX mEq/L
Test repeats 1 Panic XXX - XXX mEq/L
Controls
Control 1
Concentration limits 50 - 130 mEq/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 mEq/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Incubation time 4.5 min 0
Sample
Sample blank No 0 0
normal volume rerun volume* Factor 1.00
3.0 L 2.0 l Absorbance limits
Reagents Absorbance 0.000 - 0.800 Abs
Reagent blank Yes 3 R. Absorbance 0.000 - 0.200 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) 0 9.000 0
300 L 301 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 7 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 XXX 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0
APPLICATION Selectra ProXS / Flexor EL80
XXX
XXX
FAUX
XXX: Enter data by the user
FAUX
FAUX
CHLORIDE
CHLORIDE
End point monochromatic
FAUX
XXX
XXX
XXX
No
CHLORIDE
Testp
arameters1
Testparameters2
FAUX
FAUX
0
Calibration
0
FAUX
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-CHLO-1
30/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
31/123
Ref : CHLO-0250 12 x 20 mL
CHLO-0600 2 x 125 mL
Instruction for use :
APPLICATION Selectra ProXS / Flexor EL80
CHLORIDE
0 0
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 50 to 130 mEq/L.
Detection limi tDetermined according to SFBC protocol, the detection limit is equal to 1 mEq/L.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 1.6%
Level 2 20 Level 2 1.5%
Level 3 20 Level 3 1.6%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific XL on 30 human sera samples.
The sample concentrations were between 48 and 128.7 mEq/L.
The parameters of linear regression are as follows :
Correlation coefficient (r) :0.9993
Linear regression : y= 0.9994 x + 0.07 mEq/L
InterferencesAccording to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin :No significant interference up to 35 mg/dL on normal human sera.
Positive bias from 23 mg/dL on pathological human sera.
Conjugated bilirubin :No significant interference up to 25 mg/dL.
Haemoglobin: Positive bias from 250 mg/dL
Turbidity :Positive bias from 300 mg/dL Triglycerides equivalent.
Other compounds may interfere (see product insert).
On board stability (refrigerated position)The reagent is stable 4 weeks on analyser with a new calibration every week.
80.6 20
0.3%
0.4%
0.5%
Mean (mEq/L)
82.5
107.1
Mean (mEq/L) NCV%
112.3
117.4
FAUX
113.8 19
20
Modification from the previous version
(05/2009)
AA-PROXS/EL80-CHLO-1
31/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
32/123
Ref : 12 x 20 mL
6 x 45 mL
1 x 100 mL
6 x 100 mL
Instruction for use : 4 x 250 mL
6 x 100 mL +STD
4 x 250 mL +STD
General Sample types
Name Normal
Abbreviated name CHSL Allowed Yes
Code 280 Predilution No 0
Mode Reference values
Wavelength 505 nm 0 Male XXX - XXX mg/dL
Units mg/dL Female XXX - XXX mg/dL
Decimals 0 Pediatric XXX - XXX mg/dL
Test repeats 1 Panic XXX - XXX mg/dL
Controls
Control 1
Concentration limits 20 - 600 mg/dL Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 mg/dL
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Incubation time 4.5 min 0
Sample
Sample blank No 0 0normal volume rerun volume* Factor 1.00
3.0 L 3.0 l Absorbance limits
Reagents Absorbance 0.000 - 2.000 Abs
Reagent blank Yes 3 R. Absorbance 0.000 - 0.150 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) 0 9.000 0
300 L 300 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator PredilutionName Auto predilution No
Number of standard 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 28 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0
XXX
CHSL-0250
CHSL-0455
CHSL-0490
CHSL-0707
CHSL-0507
CHSL-0700
CHSL-0500
APPLICATION Selectra ProXS / Flexor EL80
CHOLESTEROL SL
XXX
FAUX
XXX
XXX: Enter data by the user
FAUX
FAUX
CHOLESTEROL SL
FAUX
0
FAUX
FAUX
FAUX
Calibration
0
FAUX
CHOLESTEROL SL
End point monochromatic
XXX
Testpa
rameters1
Testparameters2
XXX
No
Modification from the previous version
(05/2009)
AA-PROXS/EL80-CHSL-1
32/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
33/123
Ref : 12 x 20 mL
6 x 45 mL
1 x 100 mL
6 x 100 mL
Instruction for use : 4 x 250 mL
6 x 100 mL +STD
4 x 250 mL +STD
CHSL-0250
CHSL-0455
CHSL-0490
CHSL-0707
CHSL-0507
CHSL-0700
CHSL-0500
APPLICATION Selectra ProXS / Flexor EL80
CHOLESTEROL SL
0 0
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 20 to 600 mg/dL.
Detection limitDetermined according to SFBC protocol, the detection limit is equal to 2 mg/dL.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 1.4%
Level 2 20 Level 2 2.1%
Level 3 20 Level 3 2.2%
CorrelationA comparative study has been performed between Selectra ProXS / Flexor EL80 and Vital Scientific E on 31 human sera samples.
The sample concentrations were between 9 and 626 mg/dL.
The parameters of linear regression are as follows :Correlation coefficient (r) :0.9993
Linear regression : y= 1.0351 x - 4 mg/dL
InterferencesAccording to SFBC recommendations some studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin :Negative bias from 30 mg/dL.
Conjugated bilirubin :No significant interference up to 25 mg/dL.
Haemoglobin: No significant interference up to 500 mg/dL.
Turbidity :No significant interference up to 600 mg/dL Triglycerides equivalent.
Other compounds may interfere (see product insert).
On board stability (refrigerated position)The reagent is stable 4 weeks on analyser without recalibrating.
N
20
1.3%
1.6%
114
292
Mean (mg/dL)
2.4%
CV%
187
20
20188
Mean (mg/dL)
115
296
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-CHSL-1
33/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
34/123
Ref : CHOL-0220 12 x 20 mL
CHOL-0420 9 x 50 mL
CHOL-0520 6 x 100 mL
Instruction for use : CHOL-0720 4 x 250 mL
General Sample types
Name Normal
Abbreviated name CHOL Allowed Yes
Code Predilution No 0
Mode Reference values
Wavelength 505 nm 0 Male XXX - XXX mg/dL
Units mg/dL Female XXX - XXX mg/dL
Decimals 0 Pediatric XXX - XXX mg/dL
Test repeats 1 Panic XXX - XXX mg/dL
Controls
Control 1
Concentration limits 20 - 500 mg/dL Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 mg/dL
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Incubation time 4.5 min 0
Sample
Sample blank No 0 0
normal volume rerun volume* Factor 1.00
3.0 L 2.0 l Absorbance limits
Reagents Absorbance -0.100 - 1.500 Abs
Reagent blank Yes 3 R. Absorbance -0.100 - 0.300 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) 0 9.000 0
300 L 301 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval xxx days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0
FAUX
FAUX
CHOLESTEROL
FAUX
FAUX
0
FAUX
FAUX
CHOLESTEROL
XXX
XXX: Enter data by the user
XXX
XXX
No
End point monochromatic
APPLICATION Selectra ProXS / Flexor EL80 PROPOSAL
CHOLESTEROL
Testp
arameters1
Testparameters2
XXX
XXX
Calibration
0
FAUX
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-CHOL-1
34/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
35/123
Ref : HDLL-0230 4 x 28 mL
HDLL-0380 1 x 80 mL
Instruction for use : HDLL-0390 3 x 80 mL
General Sample types
Name Normal
Abbreviated name HDLL Allowed Yes
Code 340 Predilution No 0
Mode Reference values
Wavelength 578 nm 0 Male XXX - XXX mg/dL
Units mg/dL Female XXX - XXX mg/dL
Decimals 1 Pediatric XXX - XXX mg/dL
Test repeats 1 Panic XXX - XXX mg/dL
Controls
Control 1
Concentration limits 5.0 - 200.0 mg/dL Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 mg/dL
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Point one, two -3 0 236 sec
Sample
Sample blank No 0 0
normal volume rerun volume* Factor 1.00
2.4 L 2.4 l Absorbance limits
Reagents Absorbance -0.010 - 1.000 Abs
Reagent blank Yes 3 R. Absorbance -0.010 - 0.010 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) Substrate depletion 0.000 Abs
240 L 240 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
80 L 80 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval XXX days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0
Testp
arameters1
Testparameters2
FAUX
FAUX
0
Calibration
0
FAUX
CHOLESTEROL HDL SL 2G
CHOLESTEROL HDL SL 2G
XXX: Enter data by the user
Two point
APPLICATION Selectra ProXS / Flexor EL80
FAUX
XXX
XXXXXX
No
CHOLESTEROL HDL SL 2G
XXX
FAUX
FAUX
XXX
FAUX
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-HDLL-1
35/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
36/123
Ref : HDLL-0230 4 x 28 mL
HDLL-0380 1 x 80 mL
Instruction for use : HDLL-0390 3 x 80 mL
CHOLESTEROL HDL SL 2G
APPLICATION Selectra ProXS / Flexor EL80
0 0
PERFORMANCE DATA
Analytical rangeDetermined according to EP6-A protocol of CLSI
The reagent is linear from 5 to 200 mg/dL.
Detection limi t
Determined according to SFBC protocol, the detection limit is equal to 1.0 mg/dL.
Precision
Determined according to EP5-A2 protocol of CLSI
N
Level 1 80
Level 2 80
Level 3 80
Correlation
Determined according to EP9-A2 protocol of CLSI.A comparative study has been performed against a concurrent system on 83 human samples.
The sample concentrations were between 9.3 and 151 mg/dL.
The parameters of linear regression are as follows :
Correlation coefficient (r): 0.996
Linear regression : y= 1.141 x - 4.0 mg/dL
InterferencesAccording to SFBC recommendations and EP7-A2 protocol of CLSI, studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin: No significant interference up to 616 mol/L.
Conjugated bilirubin: No significant interference up to 427 mol/L.
Turbidity: No significant interference up to 500 mg/dL Triglycerides equivalent.
Hemoglobine: No significant interference up to 500 mg/dL.Other compounds may interfere. (see product insert)
On board stability (refrigerated position)The reagents are stable 3 weeks on analyser. A new calibration is required when the quality control result are outside
expected specification. To increase the calibration stability, it is recommended to close the vials when they are not used.
87.3
2.8%55.6
Within-run
30.8
Total
CV% CV%
FAUX
3.0%
0.7%
Mean (mg/dL)
1.4% 3.3%
1.4%
Modification from the previous version
(05/2009)
AA-PROXS/EL80-HDLL-1
36/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
37/123
Ref : LDLL-0230 4 x 28 mL
LDLL-0380 1 x 80 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name LDLL Allowed Yes
Code 360 Predilution No 0
Mode Reference values
Wavelength 578 nm 0 Male XXX - XXX mg/dL
Units mg/dL Female XXX - XXX mg/dL
Decimals 1 Pediatric XXX - XXX mg/dL
Test repeats 1 Panic XXX - XXX mg/dL
Controls
Control 1
Concentration limits 15 - 700.0 mg/dL Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 mg/dL
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Point one, two -3 0 236 sec
Sample
Sample blank No 0 0
normal volume rerun volume* Factor 1.00
2.4 L 2.4 l Absorbance limits
Reagents Absorbance -0.010 - 2.000 Abs
Reagent blank Yes 3 R. Absorbance -0.010 - 0.020 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) Substrate depletion 0.000 Abs
240 L 240 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
80 L 80 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval 21 days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0
XXX
FAUX
FAUX
CHOLESTEROL LDL SL 2G
Two point
FAUX
CHOLESTEROL LDL SL 2G
XXX
XXX: Enter data by the user
FAUX
0
FAUX
0
FAUX
FAUX
Calibration
APPLICATION Selectra ProXS / Flexor EL80
Testp
arameters1
Testparameters2
XXX
FAUX
XXX
No
CHOLESTEROL LDL SL 2G
XXX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-LDLL-1
37/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
38/123
Ref : LDLL-0230 4 x 28 mL
LDLL-0380 1 x 80 mL
Instruction for use :
CHOLESTEROL LDL SL 2G
APPLICATION Selectra ProXS / Flexor EL80
0 0
PERFORMANCE DATA
Analytical rangeDetermined according to EP6-A protocol of CLSI
The reagent is linear from 15 to 700 mg/dL.
Detection limi t
Determined according to SFBC protocol, the detection limit is equal to 1.1 mg/dL.
Precision
Determined according to EP5-A2 protocol of CLSI
N
Level 1 80
Level 2 80
Level 3 80
Correlation
Determined according to EP9-A2 protocol of CLSI.A comparative study has been performed against a concurrent system on 60 human samples.
The sample concentrations were between 14.7 and 352.1 mg/dL.
The parameters of linear regression are as follows :
Correlation coefficient (r) : 0.985
Linear regression : y= 0.937 x + 9.4 mg/dL
InterferencesAccording to SFBC recommendations and EP7-A2 protocol of CLSI, studies have been performed to determine the level of interference
from different compounds:
Unconjugated bilirubin: No significant interference up to 513 mol/L.
Conjugated bilirubin: No significant interference up to 359 mol/L.
Turbidity: No significant interference up to 600 mg/dL Triglycerides equivalent.
Hemoglobine: No significant interference up to 500 mg/dL.Other compounds may interfere (see product insert)
On board stability (refrigerated position)The reagents are stable 3 weeks on analyser without recalibrating.
2.3%
161.7 1.9% 3.3%
121.4 1.3%
101.9
Mean (mg/dL)
Within-run
1.8% 3.4%
Total
CV% CV%
FAUX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-LDLL-1
38/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
39/123
Ref : HDLC-0060 3 x 10 mL
HDLC-0520 3 x 100 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name HDLC Allowed Yes
Code Predilution No 0
Mode Reference values
Wavelength 505 nm 0 Male XXX - XXX mg/dL
Units mg/dL Female XXX - XXX mg/dL
Decimals 0 Pediatric XXX - XXX mg/dL
Test repeats 1 Panic XXX - XXX mg/dL
Controls
Control 1
Concentration limits 0 - 200 mg/dL Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 mg/dL
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Incubation time 4.5 min 0
Sample
Sample blank No 0 0
normal volume rerun volume* Factor 1.00
15.0 L 10.0 l Absorbance limits
Reagents Absorbance 0.100 - 2.500 Abs
Reagent blank Yes 3 R. Absorbance 0.000 - 2.000 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) 0 9.000 0
300 L 305 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval xxx days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0
Testp
arameters1
Testparameters2
FAUX
FAUX
0
Calibration
0
End point monochromatic
FAUX
XXX
XXX
XXX: Enter data by the user
FAUX
XXX
No
HDL CHOLESTEROL
FAUX
XXX
FAUX
XXX
HDL CHOLESTEROL
FAUX
FAUX
APPLICATION Selectra ProXS / Flexor EL80 PROPOSAL
HDL CHOLESTEROL
Modification from the previous version
(05/2009)
AA-PROXS/EL80-HDLC-1
39/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
40/123
Ref : CHES-0053 16 x 3 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name CHES Allowed Yes
Code Predilution No 0
Mode Reference values
Wavelength 405 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 1200 - 8000 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 24 0 106 sec
Sample
Sample blank No Alinearity 10 %
normal volume rerun volume* Factor 1.00
2.4 L 2.4 l Absorbance limits
Reagents Absorbance 0.000 - 3.000 Abs
Reagent blank Yes 3 R. Absorbance 0.050 - 0.500 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) Substrate depletion 3.000 Abs
360 L 360 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
12 L 12 l 0 9.000 0
Calibrator Predilution
Name Auto predilution No
Number of standards 1 0
Calibration accepted - 0 0
Repeats 3 Cut-off
Interval xxx days Cut-off No
Curvefit 0
0 0 0
0 0 0 9.000 9.000
Auto accept calibration Yes 0 9.000 9.000
N Used Concentration
#1 Standard #1 xxx 0
0 0
0 0
0 0 * : For information only0 0
0 0
0 0
0 0 FAUX
XXX
XXX
FAUX
FAUX
XXX
No
CHOLINESTERASE
FAUX
FAUX
CHOLINESTERASE
XXX
0
APPLICATION Selectra ProXS / Flexor EL80 PROPOSAL
CHOLINESTERASE
Kinetic
XXX: Enter data by the user
XXX
FAUX
Testp
arameters1
Testparameters2
FAUX
FAUX
0
Calibration
Modification from the previous version
(05/2009)
AA-PROXS/EL80-CHES-1
40/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
41/123
Ref : CMSL-0230 4 x 25 mL
CMSL-0410 2 x 62.5 mL
CMSL-0430 4 x 62.5 mL
Instruction for use :
General Sample types
Name Normal
Abbreviated name CMSL Allowed Yes
Code 400 Predilution No 0
Mode Reference values
Wavelength 340 nm 0 Male XXX - XXX U/L
Units U/L Female XXX - XXX U/L
Decimals 0 Pediatric XXX - XXX U/L
Test repeats 1 Panic XXX - XXX U/L
Controls
Control 1
Concentration limits 3 - 600 U/L Control 2
Calibrator name Control 3
Correlation
Prozone check Correlation factor 1.000
Correlation offset 0.000 U/L
General Measurement
Name Slope Blank No 0 0
Predilution No 0 Delay, min. time 130 0 159 sec
Sample
Sample blank No Alinearity 10 %
normal volume rerun volume* Factor 7949.00
10.0 L 5.0 l Absorbance limits
Reagents Absorbance 0.000 - 1.500 Abs
Reagent blank Yes Repeats 3 R. Absorbance 0.000 - 1.000 Abs
normal volume rerun volume*
Reagent addition at - 2.25 minutes (R1) R. Abs. deviation 3.000 Abs
250 L 255 l
Reagent addition at 2.90 minutes (R2) 0
0 L 0 l 0 9.000 0 9.000 0
Reagent addition at 4.70 minutes (R3) 0 9.000 0 9.000 0
0 L 0 l 0 9.000 0
* : For information only
Testp
arameters1
Testparameters2
CK-MB SL
Kinetic
No
CK-MB SL
XXX
APPLICATION Selectra ProXS / Flexor EL80
CK-MB SLONE-REAGENT PROCEDURE
XXX: Enter data by the user
XXX
XXX
Modification from the previous version
(05/2009)
AA-PROXS/EL80-CMSLmono-1
41/123
7/24/2019 Selectra Pro XS, Flexor EL80 Manual Configuracin de Reactivos
42/123
Ref : CMSL-0230 4 x 25 mL
CMSL-0410 2 x 62.5 mL
CMSL-0430 4 x 62.5 mL
Instruction for use :
APPLICATION Selectra ProXS / Flexor EL80
CK-MB SLONE-REAGENT PROCEDURE
PERFORMANCE DATA
Analytical rangeThe reagent is linear from 10 to 600 U/L.
Detection limi tDetermined according to SFBC protocol, the detection limit is equal to 3 U/L.
Precision
Within run reproducibility Between run reproducibility
N CV%
Level 1 20 Level 1 4.7%
Level 2 20 Level 2 3.7%
Level 3 20 Level 3 2.7%
CorrelationA comparative study has been performed