MENA3100 - SIMS 1 Secondary ion mass spectrometry Secondary ion mass spectrometry (SIMS) (SIMS) Lasse Vines

Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

1

Secondary ion mass spectrometrySecondary ion mass spectrometry (SIMS)(SIMS)

Lasse

Vines

Page 2: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

2

Secondary ion mass spectrometrySecondary ion mass spectrometry

0 20 40 60 80 1001

10

100

1000

10000

Cou

nts/

sec

Mass (AMU)

Li

OO2

K

Zn

ZnO

ZnO2

Na Cr

1014

1015

1016

1017

1018

1019

1020

1021

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7

PBAs

Atom

ic c

once

ntra

tion

(cm

-3)

Depth (um)

Si07

Ge0.3

Page 3: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

3

ChemicalChemical identificationidentification

Impurity

identification Isotope ratio

64Zn 66Zn67Zn

68Zn

70Zn

Page 4: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

4

Characterization of device structureCharacterization of device structure

1014

1015

1016

1017

1018

1019

1020

1021

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7

PBAs

Atom

ic c

once

ntra

tion

(cm

-3)

Depth (um)

Si07

Ge0.3

Page 5: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

5

Characterization of solar cellCharacterization of solar cell

Characterization•Optimization of processing•Trouble shooting

1,2

1,0

0,8

0,6

0,4

0,2

0,01E16 1E17 1E18 1E19 1E20

Dep

th (µ

m)

P Concentration (cm-3)

Page 6: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

6

Ion Ion imagingimaging

Na distribution

Page 7: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

7

OutlineOutline

Characteristic features–

Comparison with other techniques

Physical processes–

Sputtering

Ionization

SIMS instrumentation–

Types of mass spectrometers

Measurement modes: Mass spectra, Depth profiling, Ion imaging

Examples of applications–

Diffusion studies

Material development

Page 8: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

8

Characteristic featuresCharacteristic features

Quantitative chemical analysis•

High detection sensitivity–

1016

– 1013

atoms/cm3 (ppm-ppb)–

Can measure H

Large dynamic range–

> 5 orders of magnitude

Very high depth resolution–

Resolution of 20 Å

can be obtained

Ion microscopy–

Lateral resolution < 0.5 µm

But,•

Limited to concentration <5%

Samples must be vacuum compatible •

Samples must be partially conductive

Destructive technique

Page 9: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

9

Comparison to other techniquesComparison to other techniques

Page 10: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

10

OutlineOutline

Characteristic features–

Comparison with other techniques

Physical processes–

Sputtering

Ionization

SIMS instrumentation–

Types of mass spectrometers

Measurement modes: Mass spectra, Depth profiling, Ion imaging

Examples of applications–

Diffusion studies

Material development

Page 11: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

11

Ion Ion –– solid interactionsolid interaction

Matrix atom

Impurityatom

Primary ion

Secondary ions are accelerated by an applied sample voltage

Primary beam

Energy is transferred from the energetic primary ions to atoms in the sample. Some of these receive enough energy to escape the sample.

Page 12: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

12

SputteringSputtering

Sigmund P. Theory of Sputtering, Phys. Rev. 184(2), 383 (1969)

Sputtering Yield:number of sputtered atoms per incoming ion

it

in

0

iti E

ESUKES

8/3n 383/1ln5.0 S

Sputtering yield:

50.05for 3

keV5.321

it6/5

tiit

2/13/2t

3/2ititiit

ZZZZK

ZZZZMME

Mi , Zi : Ion mass and atomic numberMt , Zt : Target mass and atomic numberU0 : Surface escape barrier in eVEi : Ion energy

Nuclear stopping cross-section:Sputtering is a multiple collision process involving a cascade of moving target atoms, this cascade may extend over a considerable region inside the target.

Page 13: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

13

SputteringSputtering

Dependence of ion

Page 14: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

14

SputteringSputtering

Example of dependence of target on sputtering yield: (Si1-x

Gex

)

0 20 40 60 80 100

1

2

3

Nor

mal

ized

ion

yiel

d

Ge content (%)

Page 15: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

15

SputteringSputtering

Example of sputtering of polycrystalline Fe surface

The erosion rate is different for the different grains: Sputtering yield vary with the crystal orientation

Page 16: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

16

SputteringSputtering

Example of sputtering yield:

200 µm

0 100 200 300 400 500

-1,0

-0,8

-0,6

-0,4

-0,2

0,0

Dep

th (µ

m)

Width (µm)

Material removed: 1200200 µ3

= 410-8

cm3

21015

atoms

Current: 200 nASputtering time: 700 sec

Incoming ions: 20010-9A

6.241018

ions/C

700 sec = 9x1014

ions

Sputtering Yield = 2.2 atoms/ion

Page 17: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

17

Sputtering Sputtering –– Secondary ionsSecondary ions

Energy distribution of secondary ions

0 20 40 60 80 10010

100

1000

10000

100000

28Si4

28Si3

28Si2

Sec

onda

ry in

tens

ity (a

rb. u

nit)

Energy (eV)

28Si

Page 18: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

18

IonizationIonization

Ion yield/ionization efficiency : The fraction of sputtered ions that becomes ionized.

Ion yield can generally not be predicted theoretically.•

Ion yield can vary by several orders of magnitude depending on element and chemistry of the sputtered surface.

Oxygen on the surface will increase positive ion yield•

Cesium on the surface will increase negative ion yield

Page 19: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

19

IonizationIonizationNegative secondary

Positive secondary

vAC

vEC i

/exp YieldIon Negative

/exp YieldIon Positive

EiA

C±: Constantsv: velocity perpendicular to surface: work function

(Cs)

(O)

Page 20: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

20

IonizationIonization

64 66 68 70100

101

102

103

104

105

106

107

Sec

onda

ry In

tens

ity(c

ps)

M/q (AMU)

Negative modePositive mode

Mass

spectrum

of ZnO, Zn peaks.64Zn

(48.6%)66Zn

(27.9%)

67Zn

(4.1%)

68Zn

(18.8%)

70Zn

(0.6%)

Page 21: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

21

IonizationIonization

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

0 20 40 60 80 100

Nor

mal

ized

P- - y

ield

Ge concentration (%)

Phosphorus in Si1-x Gex

Limits the quantification procedure:–

SIMS is mainly a tool for measuring small concentrations in a given matrix

Page 22: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

22

General YieldGeneral Yield

Measured intensity It for a specific target atom

TCYII ttPt

IP : Primary ion current Y : Sputtering yield

(number of sputtered particles per impinging primary ion)

[Ct ]: Concentration of species t t : Secondary ion formation and

survival probability (ionization efficiency)

T: Instrument transmission function

t

is highly dependent on species and matrix

Page 23: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

23

OutlineOutline

Characteristic features–

Comparison with other techniques

Physical processes–

Sputtering

Ionization

SIMS instrumentation–

Types of mass spectrometers

Measurement modes: Mass spectra, Depth profiling, Ion imaging

Examples of applications–

Diffusion studies

Material development

Page 24: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

24

The SIMS instrumentThe SIMS instrument

Focused

ion beamMass

Spectrometer

Page 25: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

25

The SIMS instrumentThe SIMS instrument

Instruments are usually classified by the type of mass spectrometer:–

Time of Flight•

Simultaneous detection of many elements

High transmission•

Measures large molecules

Quadrupole•

Low impact energy

Magnetic Sector•

High mass resolution

High transmission•

Low detection limit

Page 26: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

26

Time of Time of FlightFlight SIMSSIMS

Page 27: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

27

QuadrupoleQuadrupole SIMSSIMS

Page 28: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

28

Magnetic sector Magnetic sector -- mass mass spectrometer spectrometer

sample

detector

ion source

electrostatic sector analyser

magnetic sector analyser

Primary beam

Seco

ndar

y be

amre rm

E0B

Lorenz’ force: BvEF qq

Centripetal force:rr

mv rF2

2

e0 r

mvqE

Electrostaticsector analyser

2

mrmvqvB

Magneticsector analyser

2

0e

m

ErBr

qm

Page 29: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

29

Secondary ion mass spectrometrySecondary ion mass spectrometry

2m

0 e

Brmq E r

sample

detector

ion source

electrostatic sector analyser

magnetic sector analyser

Primary beam

Seco

ndar

y be

am

re rm

E0B

0 100 200 300 400 500 6001

10

100

1000

10000

Inte

nsity

(cou

nts/

sec)

Sputter time (sec)

Depth profile

0 20 40 60 80 1001

10

100

1000

10000

Cou

nts/

sec

Mass (AMU)

Mass spectrum

Ion image

20 µm

Page 30: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

30

InstrumentationInstrumentationion sources

sample chamber

electrostatic

sector

analysermagnetic

sector

analyser

detectors

Page 31: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

31

MassMass spectrumspectrum

0 20 40 60 80 1001

10

100

1000

10000

Cou

nts/

sec

Mass (AMU)

Li

OO2

K

Zn

ZnO

ZnO2

Na Cr

Mass

spectrum

of a ZnO-sample

with

traces of Li, Na, K, and Cr.

Page 32: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

32

MassMass InterferenceInterference

11,5 12,0 12,5 13,0 13,5

100

101

102

103

104

105

106

Sec

onda

ry in

tens

ity (c

ps)

Mass (AMU)

Mass

spectrum of graphite

91.6%

8.4%

Page 33: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

33

Mass interferenceMass interference

Several ions/ionic molecules have similar mass to charge ratios:

10B -

30Si3+ Monitor 11B

75As - 29Si30Si16O

Page 34: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

34

Energy selectionEnergy selectionelectrostatic sector analyser

Seco

ndar

y be

am

re

E0

2

e0 r

mvqE

Increasing kinetic energy

Energy selection slit

Ejection energy (eV)lo

g (

ion

in

ten

sity

)

75As

29Si30Si16O

0 50 100

Page 35: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

35

MassMass interferenceinterference

Several

ions/ionic

molecules

have similar mass

to charge

ratios:

10B -

30Si3+ Monitor 11B

75As - 29Si30Si16O Energy selection

31P - 30Si1H

Page 36: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

36

HighHigh massmass resolutionresolution

30,85 30,90 30,95 31,00 31,05 31,10

100

1000

10000

Inte

nsity

(cou

nts/

sec)

Mass (AMU)

magnetic sector analyser

rm

B

2

mrmvqvB

Exit slit

Discriminating

between

31P and 30Si1H:

M(31P) = 30.973761M(30Si1H) =30.98160

30,85 30,90 30,95 31,00 31,05 31,10

100

1000

10000

Inte

nsity

(cou

nts/

sec)

Mass (AMU)

31P

30Si1H

Page 37: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

37

Mass interferenceMass interference

Several ions/ionic molecules have similar mass to charge ratios:

10B -

30Si3+ Monitor 11B

75As - 29Si30Si16O Energy selection

31P - 30Si1H High mass resolution

Page 38: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

38

SIMS SIMS –– depth profilingdepth profiling

0 100 200 300 400 500 6001

10

100

1000

10000

Inte

nsity

(cou

nts/

sec)

Sputter time (sec)

Primary beam

Page 39: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

39

0 100 200 300 400 500 600 700

1

10

100

Inte

nsity

(cou

nts/

sec)

Sputter time (sec)

CalibrationCalibration of of depthdepth profilesprofiles

”Raw”

phosphorus

profile

0 100 200 300 400 500

-1,0

-0,8

-0,6

-0,4

-0,2

0,0

Dep

th (µ

m)

Width (µm)

Depth calibration

Sputter time: 700 sec Depth: 9310 Å

Erosion rate:13,3 Å/sec

Page 40: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

40

0 100 200 300 400 500 600 700

1

10

100

Inte

nsity

(cou

nts/

sec)

Sputter time (sec)

CalibrationCalibration of of depthdepth profilesprofiles

TCYII ttPt

”Raw”

phosphorus

profile

t1 CS

S: Sensitivity

factor

0 100 200 300 400 500 6000,1

1

10

100

1000

10000

Inte

nsity

(cou

nts/

sec)

Sputter time (sec)

Concentration calibration

Ion implanted sample:P dose 1e15 P/cm2

sensitivity factor:Relate the intensity to atomic concentration

xxIS

d Dose

Sensitivity factor:1 count/sec = 3,41015

P/cm3

Page 41: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

41

Calibration of depth profilesCalibration of depth profiles

0,0 0,2 0,4 0,6 0,81E14

1E15

1E16

1E17

1E18

P co

ncen

tratio

n (c

m-3)

Depth (µm)0 100 200 300 400 500 600 700

1

10

100

Inte

nsity

(cou

nts/

sec)

Sputter time (sec)

”Raw”

phosphorus profile

Calibratedphosphorus profile

Erosion rate:13,3 Å/sec

Sensitivity factor:1 count/sec = 3,41015

P/cm3

Page 42: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

42

Ion imagingIon imaging

Distribution of given atoms at the surface

Primary

beam

Secondary beam to detector

Intensity recorded as a function of primary beam position

Sample Surface

Page 43: Secondary ion mass spectrometry · Sputter time (sec) Depth profile. 020 40 60 80 100 1 10 100 1000 10000. Counts/sec Mass (AMU) Mass spectrum. Ion image. 20 µm. MENA3100 - SIMS

MEN

A3100 -

SIM

S

43

Ion Ion imagingimaging