Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
Sartorius Weighing Seminars
Regulations and Procedures to ensure best Weighing Accuracy in your Laboratory
Modules
2
• Terminology - Avoid errors and understand their daily impact on your
weighing accuracy (Uncertainty)
• Regulations in the pharmaceutical laboratory
• Accurate sample weighing to 0.10% uncertainty according to USP
• Qualification of your laboratory balance
• Monitoring your laboratory balance
Optional:
• Calibration weights - selection, storage and correct handling
• Product presentation Secura, Quintix and Practum
24. September 2013
Qualified Accurate weighing
324. September 2013
Qualified Accurate weighing requires:
• Trained competent personnel
• Defined appropriate equipment and
conditions
• Testing facility designed to minimise
problems
424. September 2013
Qualified Accurate weighing requires:
• Regular calibration of measuring
equipment
• According to accepted methods
• Use of Appropriate test & reference
standards – Calibration Weights
524. September 2013
Qualified weighing requires you to:
• Keep your balance clean
• Position the balance on firm surface
• Level the balance
• Protect the balance from drafts
• Especially air conditioning
• Never position in direct sunlight
• Keep well away from all heat sources
24. September 2013 6
Avoid hazards that impact on your weighing resultsThe correct place of installation?
• Mechanically and climatically stable
− stable temperature and humidity
− better on the first floor than on
upper floors
• No vibrations
− especially low frequencies
< 10Hz should be avoided
− better to be closer to a wall, than
in the middle of the room
724. September 2013
Aircondition Open doors
Open windows
Sunlight
Air draft
unstabletable
Humidity
Hustleand Bustle
Vibrations
Avoid hazards that impact on your weighing resultsThe correct place of installation?
• Avoid installation in the exhaust area
of an air conditioner (disturbances and
temperature)
• Not near to a window, no direct sunlight
− avoid infrared radiation;
no direct light e.g.... a desk lamp
• Quiet place without hustle and bustle
824. September 2013
Aircondition Open doors
Open windows
Sunlight
Air draft
unstabletable
Humidity
Hustleand Bustle
Vibrations
Terminology
• Readability
• Maximum capacity
• Response time
• Repeatability
• Linearity
• Sensitivity drift
• Accuracy
• Uncertainty
24. September 2013 9
Readability
The smallest change in weight
that can be measured
24. September 2013 10
Readability
The smallest change in weight
that can be measured
1124. September 2013
Readability
The smallest change in weight
that can be measured
• 9.99g �10.00g
• Readability = 0.01g (d)
1224. September 2013
Weighing capacity
The maximum amount that can be
weighed out before overloading occurs
24. September 2013 13
Weighing capacity
The maximum amount that can be
weighed out before overloading occurs
8,200 maximum capacity – 5,000 load
= 3,200g remaining weighing capacity
A fill guide prevents overload after taring
1424. September 2013
Stabilisation Time
The time taken for the balance to become
stable after the load is applied
A stable reading is indicated by showing
the „g“
1524. September 2013
Repeatability
Repeatability, or test-retest reliability,
is the variation in measurements taken
by a single person or instrument on the
same item and under the same
conditions.
24. September 2013 16
Repeatability of a balance
24. September 2013 17
± 0.01g
Linearity
Deviation from the theoretical linear slope
24. September 2013 18
± 0.01g
± 0.0001g
Sensitivity DriftDrift associated with temperature change
• ± 1x10-6 per degree Celsius or:
• ± 1 part per million per degrees Celsius
• Relating to a 200g Calibration Weight
• 200.0000g x 10-6
• or
• 200.0000g ÷ 1,000,000
• 0.0002g drift per degree Celsius
24. September 2013 19
Accuracy
• Is the extent to which the readout
approaches the true value of the object.
• Accuracy is subject to local random and
systematic influences therefore cannot
be specified
• Accuracy can only be determined at the
place of use
24. September 2013 20
Uncertainty of measurement
24. September 2013 21
• Any measured value is only meaningful when presented together with it’s
determined uncertainty
• Uncertainty indicates the range above and below the measured value,
in which it will lie
• e.g.
• 20.0003g ± 0.0005g
• with 95% confidence
• Meaning: a 95% certainty the true value is within the range 19.9998g - 20.0008g
Avoid external errorsElectrostatic charges
24. September 2013 22
• Electrostatic charges create an acting force in addition to the weight of the sample
and container.
− This force is not constant. It’s influenced by humidity, the user and handling of
samples and containers.
− The weight readout drifts in one direction. Values are non-repeatable.
• Static electricity occurs on:
− substances or containers with low electrical conductivity
− large surfaces (plastic or glass containers, filters)
• The reasons are:
− internal friction in powders, external friction
− transfer of electrical charges by persons
− problem increases with decreasing humidity (<40%)
Avoid errorsElimination of electrostatic Charges
Internal solutions:
• Built-in ionizer inside the draft shield of
semi-micro, analytical and 1mg-balances
External solutions
• Ionizer cube YIB01
• Stat Pen YSTP01 for small sample
containers and membrane filters
2324. September 2013
Avoid errorsMagnetic and magnetisable samples
The problem:
Magnetic materials in samples or containers, such as
nickel, iron, steel, generate force fields that act against
the permanent magnet of the weighing system.
• Weight values are stable, but non-repeatable.
• Different values are displayed depending on theposition of the sample on the weighing pan.
How to avoid:
• Do not weigh with magnetic stirrer inside the vessel
• Perform demagnetisation before weighing. Use a nonmagnetic object to distance the sample from theweighing pan.
• Use Mu-metal (81% Ni and 19% Fe) foil as shielding.
• Use below balance weighing
24. September 2013 24
Avoid errorsNon-level balance
One of the worst errors that occurs during
weighing is caused when the balance is
out of level.
24. September 2013 25
αααα
400µm
Avoid errorsNon-level balance
Example:
If the balance is inclined only 400µm
(Thickness of a business card) on one side
an incorrect measurement is produced:
• That means, the mass of 200g is displayed 0.37mg too low
• The resulting error is greater than the allowable tolerance of a 200g ± 0.3mg E2 calibration weight for an analytical balance.
Error calculation:
md = m · cos α = 200g · cos (0,11°)
= 199,99963g
α = arctan (0.4mm/200mm)
∆m = -0.00037g
24. September 2013 26
αααα
400µm
Avoid errors – due tochanges in regional gravity, influenced by changes in elevation andlatitude
• Ground floor 200.00000g
• First floor 199.99974g
• Difference 0.00026g
24. September 2013 27
Avoid errors:Balance warm-up period
Balances need to be warmed up
(after new installation)
• Precision Balances (d >1 mg)approx. 30 min
• Analytical Balances (d> 0.1 mg)approx. 4 h
• Semimicro Balances (d> 0.01 mg)approx. 12 h
• Ultramicro/Micro Balances (d> 0.001 mg)approx. 24 h
To avoid subsequent warm up periods :
• do not disconnect from power
• Always leave balance in the standby-mode
24. September 2013 28
Avoid errorsCritical samples
Hygroscopic substances:
• Moisture absorption of powders
Efflorescent substances
• evaporation from the sample
• both cause a real weight change
• different results when weighing the
sample at different times
• weighing value is drifting
How to avoid?
• use narrow-necked or covered vessels
• stabilise sample humidity in a desiccator
• work quickly24. September 2013 29
Avoid errorsNon acclimatised samples and tare containers
Sample containers are very often not
sufficiently acclimatised:
• temperature differences betweenweighing chamber and sample/samplevessels
• weighing results are varying
• weighing value is drifting:
− sample too cold: “weight increasing”
− sample too warm: “weight decreasing”
How to avoid?
• acclimatisation of sample/samplecontainers
• place weighing containers and sample inside weighing chamber beforeweighing
24. September 2013 30
Avoid errorsHandling of samples, vessels and calibration weights
• handling the sample with hands raises
temperature and causes “gain-weight”
effects
• this temperature effect can still be
recognised after 10 minutes
• fingerprints falsify the real mass approx.
50 to 100µg
• additionally fingerprints can be
hygroscopic
• using the hand inside the draft shield
causes additional turbulence and
temperature effects
• weighing values are different, not stable
and drifting
24. September 2013 31
Avoid errorsHandling of samples, vessels and calibration weights
How to avoid?
• never handle the sample without
tweezers or some other suitable gripping
device
• always use gloves
• avoid working inside the draft shield, use
a spatula which is long enough to keep
the hand outside
24. September 2013 32
Modules
33
• Terminology - Avoid errors and understand their daily impact on your weighing
accuracy (Uncertainty)
• Regulations in the pharmaceutical laboratory
• Accurate sample weighing to 0.10% uncertainty according to USP
• Qualification of your laboratory balance
• Monitoring your laboratory balance
Optional:
• Calibration weights - selection, storage and correct handling
• Product presentation Secura, Quintix and Practum
24. September 2013
Following Regulations in the pharmaceutical laboratoryWhat should you expect from your balance supplier
• Manufacturers expertise
• Reliable, highly precise products
• Adaptible to meet pharmaceutical
regulations
• Global service and support
• Feel safe factor
24. September 2013 34
Following Regulations in the pharmaceutical laboratoryWhat can you expect from Sartorius
Help and advice from experts in all
weighing related questions
Local weighing specialist - Worlwide
This mainly includes the expertise of the
practical handling of the pharmaceutical
related standards and regulations.
Plus: help in selecting the correct
balance based on your weighing needs
24. September 2013 35
Regulations in the pharmaceutical laboratory andWeighing related regulations
24. September 2013 36
Which weighing related regulations are relevant in pharmaceutical Laboratories
• Quality Control
• GLP/GMP
• FDA
• ISO 17025 (government lab, accredited)
• Legal Metrology - verified balances in Europe subject to local differences
• R&D
Regulations in the pharmaceutical laboratoryGLP/GMP
GLP
• in labs that work on preclinical studies
GMP
• in Labs that work in the clinical phase
and beyond (includes QC)
General demands:
• control of inspection, measuring and test
equipment (e.g. calibrations)
• instrument qualification (IQ, OQ, PQ)
• documentation of measured values
• standardised procedures (SOP)
• Four eyes principle
24. September 2013 37
Regulations in the pharmaceutical laboratoryFDA
FDA –Food and Drug Administration
• General Demands
− process validation
− 21CFR Part 11 electronic data handling
− electronic signatures
− audit trails
• Special Demand:
− USP (United States Pharmacopeia)
− determination of the Operating Range
24. September 2013 38
Regulations in the pharmaceutical laboratoryLegal Metrology
• Legally required and controlled quality
assurance system to protect consumers
• acc. international guidelines
(e.g. OIML R76)
• translated into national law
• globally large differences
24. September 2013 39
Regulations in the pharmaceutical laboratoryWhere must a verified balance be used in Europe?
24. September 2013 40
• determination of mass for commercial transactions;
• determination of mass for the calculation of a toll, tariff, tax, bonus, penalty,
remuneration, indemnity or similar type of payment;
• determination of mass for the application of laws or regulations; expert opinion
given in court proceedings;
• determination of mass in the practice of medicine for weighing patients for the
purposes of monitoring, diagnosis and medical treatment;
• determination of mass for making up medicines on prescription in a pharmacy and
determination of mass in analyses carried out in medical and pharmaceutical
laboratories;
• determination of price on the basis of mass for the purposes of direct sales to the
public and the making-up of pre-packages
Verified balances are marked with M
Renewing is subject to local regulations
Regulations in the pharmaceutical laboratory
24. September 2013 41
... a lot of different regulations with different demands
BUT
... there is one similarity :
Not all regualtions are not defined specifically for balances –
specifics given in general terms (so as to fit all lab measuring equipment)
Regulations in the pharmaceutical laboratoryMeasurement and test equipment acc. DIN ISO 10 012
24. September 2013 42
Target
• To ensure that measurements are made with the intended accuracy
• To specify the features of a system which ensures that measuring equipment
meets the requirements for its use.
Regulations in the pharmaceutical laboratoryRequirements for measuring and test equipment acc. DIN ISO 10 012
24. September 2013 43
• The required metrological characteristics must be established, e.g. accuracy,
stability, range and resolution.
• Accuracy: Essentially, the accuracy of measuring equipment should be preferably
1/10 of the permissible error in your process. However, for economic reasons, an
accuracy of 1/3 of this error is still acceptable.
• An additional requirement is to establish and maintain a documented system for
managing, confirmation and use of measuring equipment, including measurement
standards.
Regulations in the pharmaceutical laboratoryRequirements for measuring and test equipment acc. DIN ISO 10 012
24. September 2013 44
• When measurements are performed, all significant identified uncertaintiesin the measurement process, including those contributed by personnel, procedures and
environment, must be taken into account.
• Calibrations and adjustments are to be performed at prescribed
intervals using measuring and test equipment, including measurement standards, that
has been cross-checked against other standard in an officially recognised procedure to
demonstrate and confirm the traceability of this equipment to the standard.
• Written procedures (Standard Operating Procedures) must be
provided and used for all important aspects concerning the use of measuring and test
equipment.
Page 1
United States Pharmacopeia – Chapter 41 New Release
Have you heard about
the new USP?
24. September 2013
...it is easier than you think!...it is easier than you think!...it is easier than you think!...it is easier than you think!
US Pharmacopeia Chapter 41 „Balances“ – New Release USP39Operational Range
USP no longer mentions „Minimum
Sample Weight“
New:
User defines the „Opeartional Range“
24. September 2013 47
Displayed W
eighing Value
OperationalRange
Operational
range
Displayed W
eighing Value
200g
US Pharmacopeia Chapter 41 “Balances” – it is easier than you think!
... it is all about measuring uncertainty (accuracy)!
0.1%
• USP dictates a maximum uncertainty error of 0.10%
measuringuncertaintyerror (rel)
sample weight
• as smaller the sample, as larger the effect of the measuring uncertainty error (relatively)
• a balance always has a measuring uncertainty error
X mgmaxcapacity
Operating Range
not allowed
• this demand specifies the usable weighing range (operating range) of the balance
0g 1g 10g 100g
0%
1%
US Pharmacopeia Chapter 41 “Balances” – Why?
• FDA inspects customers on USP regulations
• USP Chapter 41 is mandatory
• Minimum accuracy of 0.10% for important
weighings is dictated
24. September 2013
Please note:
• Near its zero point, a balance is very inaccurate.
• The higher you go in a balance’s weighing range,
the more accurate it becomes (relatively)
US Pharmacopeia Chapter 41 “Balances” – New Release
• From the 1st of December 2013 on, the new
USP release will be available
• After 20 years, USP Chapter 41 will be
significantly changed
• e.g. renamed from Weights and Balances
to Balances
• It does not reference minimum sample
weight anymore, it defines Repeatability and
Accuracy, which specifies the Operating
Range
24. September 2013
US Pharmacopeia Chapter 41 “Balances” – New Release
• Defines the starting point of the
operating range of the balance
• Calculation from 10 measurements
• 2 x Standard Deviation / nominal value
≤ 0.10 %
• The starting point of the operational
weighing range
• When SD < 0.41d, it must be replaced
by 0.41 digits
24. September 2013
%10.0W
SD 2≤
1000**2min SDOR =
dOR 8201000*41.0*2min ==
OR = Operating RangeSD = Standard Deviationd = DigitW = Weight
1. Repeatability:
US Pharmacopeia Chapter 41 “Balances” – Repeatability
September 24, 2013
What does this mean?
1. The new USP does not refer to minimum sample weight anymore. It specifies the
operating range. The operating range is specified by the capacity and a “starting point”
(formally known as minimum sample weight) – so this starting point needs to be defined
according to the Repeatability determination routine.
0 g 100 g 200 g displayed weight value
weight on balance
100 g
200 g
max capacity, e.g. 220 g
starting pointwith min accuracy of 0.10%
Please note:
The higher you get in the
weighing range of a
balance, the more
accurate (%) it becomes!
US Pharmacopeia Chapter 41 “Balances” – Repeatability
What does this mean?
2. The determination parameters are modified:
old: 3 x SD (three times standard deviation)
new: 2 x SD (two times standard deviation) - 30%
old: 0.1 % (which means 0.149%)
new: 0.10% (not possible to enlarge) + 30 %
• So the result will be in a similar range
• No significant changes expected!
September 24, 2013
US Pharmacopeia Chapter 41 “Balances” – Repeatability
What does this mean?
3. The absolute minimum is defined new
old: 1000 d (acc. Sartorius interpretation)
new: 820 d (2 x 0.41d x 1000 = 2 x 410d)
• This means that the maximum operating range for a Sartorius balance goes from
820d to max capacity:
Balance Type Readability Operating Range
optimal typical*
MS*2.7S 0.1 µg 0.082 mg – 2 g 1 mg – 2 g
MS*6.6S* 1 µg 0.82 mg – 6 g 2 mg – 6 g
MS*225S* 10 µg 8.2 mg – 220 g 20 mg – 220 g
MS*224S* 0.1 mg 82 mg – 220 g 100 mg – 220 g
MS*623S* 1 mg 820 mg – 620 g 900 mg – 620 g
September 24, 2013
Please note:
d = Digit
A digit is one displayed
step on a balance
****based on Sartorius’ experience
US Pharmacopeia Chapter 41 “Balances” – Accuracy
• Defines the test weight and the weight
class
• Weight must be within 5% to 100% of
capacity
• Deviation of measured value compared to
weight value ≤ 0.10% of weight value
• The measuring uncertainty of the weight
must be 1/3 of 0.10%
September 24, 2013
WtVm %10.0≤
%10.0*3/1≤Mw
Wt = Test WeightMw= measuring uncertainty of weightVm = Mw - Wt
2. Accuracy:
US Pharmacopeia Chapter 41 “Balances” – Accuracy
What does this mean?
1. The USP describes the quality of the weights, used for this test
e.g. 100 g Accuracy ≤ 0.1 g 99.9000 ... 100.1000 g
Measuring uncertainty ≤ 0.03 g
• Calibrated weights must be used to be able to show the uncertainty.
• No need to use higher class weights, F1 or F2 is absolutely enough!
24. September 2013
US Pharmacopeia Chapter 41 “Balances” – New Release
Additional Information:
• The USP now defines that customers must use “calibrated balances”.
• There are different ways to calibrate a balance.
• A well accepted method is according to ISO 17025
• Sartorius offers the service to perform ISO 17025 calibrations.
24. September 2013
US Pharmacopeia Chapter 41 “Balances” – Summary
24. September 2013
Chapter 41 Old New
Scope of validity
Applies only to assays of substances Applies to all sample weights
Determination of the minimum sample weight
Determination of the operating range
Repeatability (around the starting point)
Repeatability tolerance 0.1(49)% 0.10%
Expansion factor K 3 2
Number of weighing values for the repeatability test
10 10
Acceptance criterion 3 x std. dev./m ≤ 0.1% 2 x std. dev./m ≤ 0.10%
Smallest possible sample weight/lowest starting point
1,000 d 820 d
Accuracy
Tolerance None specified ≤ 0.10%
Test weight None specifiedBetween 5% and 100% of the balance’s capacity
US Pharmacopeia Chapter 41 “Balances” – Summary
24. September 2013
Chapter 41 Old New
Scope of validity
Applies only to assays of substances Applies to all sample weights
Determination of the minimum sample weight
Determination of the operating range
Repeatability (around the starting point)
Repeatability tolerance 0.1(49)% 0.10%
Expansion factor K 3 2
Number of weighing values for the repeatability test
10 10
Acceptance criterion 3 x std. dev./m ≤ 0.1% 2 x std. dev./m ≤ 0.10%
Smallest possible sample weight/lowest starting point 1,000 d 820 dAccuracy
Tolerance None specified ≤ 0.10%
Test weight None specifiedBetween 5% and 100% of the balance’s capacity
24. September 2013 60
0.01000g
0.1000g
1.000g
0.00820g
0.0820g
0.820g
What we mean by Digits
USP Chapter 1251 “Weighing on an Analytical Balance”
Chapter 1251 is “only” a recommendation, not mandatory!
It describes methods on:
• Balance qualification and calibration
• Risk analysis for balances
• Description of applications (e.g. differential weighing)
Argumentation:
• The recommendations in 1251 is only a suggestion.
• There are other approaches available which are even better!
24. September 2013
Accurate sample weight according to USP My sample weight is 10 mg. Which balance should I use?
• 10 mg is the absolute limit for a 5
decimal balance in “perfect” conditions
• The minimum sample weight is based
on the reproducibility of a balance and
is significantly influenced by ambient
conditions at the place of installation.
• Specifying minimum weight to the
absolute limit it is not recommended
since performance is not guaranteed
over a balances complete life time.
Recommendation:
Use a microbalance with a readability
of 1 µg.
24. September 2013 62
Accurate sample weight 0.10% according to USP Typical* Sample Weights achieved by Sartorius Balances
Readability/Balance Type
0.1µg
1 µg
2 µg
0.01mg
0.1 mg
1 mg
* typical = based on Sartorius experience
Smallest Sample Weight
1 mg
2 – 5 mg
5 – 15 mg
15 – 100 mg
100 mg
1 g
24. September 2013 63
Modules
64
• Terminology - Avoid errors and understand their daily impact on your weighing
accuracy (Uncertainty)
• Regulations in the pharmaceutical laboratory
• Accurate sample weighing to 0.10% uncertainty according to USP
• Qualification of your laboratory balance
• Monitoring your laboratory balance
Optional:
• Calibration weights - selection, storage and correct handling
• Product presentation Secura, Quintix and Practum
24. September 2013
Equipment QualificationValidation - Qualification
24. September 2013 65
• Validation is the documented evidence that a process fulfills the defined needs in
practical use.
• Qualification is the evidence that the equipment is able to fulfill it´s part in this
validated process
• Therefore a balance is never validated – it is qualified to be used in a validated
process
Equipment QualificationAny equipment qualification needs 4 steps
24. September 2013 66
DQ = Design Qualification
IQ = Installation Qualification
OQ = Operational Qualification
PQ = Performance Qualification
Technical specification forequipment (specification meet the user requirements)
Equipment is delivered and installed in accordance with the qualified design
Test of operational functions
Test of specified application
Ongoing performance tests and preventive maintenance
Before
Purchasing
At
installation
After
installation
At regular
intervals
Equipment QualificationDesign Qualification (DQ)
24. September 2013 67
• DQ is the documented collection of activities that define the functional and
operational specifications of the instrument and criteria for selection of the vendor,
based on the intended purpose of the instrument
• It documents the design of the system and include :
− Functional specification
− Technical / performance specification for equipment
• A thoroughly executed DQ ensures that:
− Specification meet the user requirements
− Design is compliant with GMP and other regulatory requirements
− Details all equipment that must be ordered
Equipment QualificationInstallation Qualification (IQ)
24. September 2013 68
• IQ is the documented collection of activities necessary to establish that an
instrument is delivered as specified, is properly installed in the selected
environment, and that this environment is suitable for the instrument
• Parts of the IQ
− Comparison of the equipment received with the purchase order
− Equipment check for any damage.
− Documentation check for completeness
− Installation of hardware
− Initial cleaning
− SQ-min-Identification
Good to know:
Responsibility for IQ lies, as the whole EQ, with the user but activities can be
supported and carried out by the vendor.
Equipment QualificationOperational Qualification (OQ)
24. September 2013 69
• OQ is the documented collection of activities necessary to demonstrate that an
instrument will function according to its operational specification in the selected
environment
• OQ tests verify that the instrument
− meets manufacturer or users specification
− operating within established limits and tolerances
• Parts of the OQ
− Basic function check and check if the instrument powers up after it is switched on
− Calibration
− Operator training
Equipment QualificationPerformance Qualification (PQ)
24. September 2013 70
• PQ is the documented collection of activities necessary to demonstrate that an
instrument consistently performs under ‘full operational‘ conditions, and is
appropriate for the intended use
• PQ tests using production materials
• Tests include a condition or set of conditions encompassing upper and lower
operating limits
• PQ is ongoing
Good to know:
PQ is described as a separate activity, for balances it is appropriate to perform it in
conjunction with OQ
IQ/OQ-Service
• IQ/OQ documentation is based on a
generic risk analysis
• Documentation consists of the IQ and
OQ planning and of protocols for the IQ
and OQ tests
• Documentation supports GLP, GMP and
FDA guidelines
• Accredited calibration certificates for our
balances
• All documentation could easily be
integrated in existing QM systems
24. September 2013 71
Equipment Qualification
24. September 2013 72
One main question remains:
• What should be tested during a IQ|OQ?
− Is it important to test the right colour of the equipment housing?
− Is it important to test if the right power supply was delivered?
− Is it important to issue a calibration certificate?
• How is the scope of test defined?
Failure mode and effects analysis (FMEA)
24. September 2013 73
• Sartorius IQ|OQ Documents are based on a generic risk analysis
• Potential risks that are connected with the installation and operation of a laboratory
balances where evaluated with an FMEA .
• Equipment and application specific risks had not been taken into account.
Modules
74
• Terminology - Avoid errors and understand their daily impact on your weighing
accuracy (Uncertainty)
• Regulations in the pharmaceutical laboratory
• Accurate sample weighing to 0.10% uncertainty according to USP
• Qualification of your laboratory balance
• Monitoring your laboratory balance
Optional:
• Calibration weights - selection, storage and correct handling
• Product presentation Secura, Quintix and Practum
24. September 2013
Monitoring the balance
24. September 2013 75
Two things are important in daily work with the balance in a pharmaceutical lab:
• Monitoring the test equipment –
making sure that the balance always is working within the defined tolerance limits
• Avoiding weighing errors caused by the user, the environmental conditions and the
sample or sample container influences
Routine testing for constant safety – test equipment monitoring
24. September 2013 76
Every Lab, which is accredited according to ISO/IEC 17025 or works according to the
GLP/GMP guidelines must regularly calibrate its testing equipment, and adjust as
necessary.
• Minimize the risk of erroneous measurements
• Fulfill the required process accuracy at any time
But, when and how often should calibration/adjustment be carried out on regular
basis?
• The length of the inspection interval depends on the risk of weighing, that means:
The customer has to clarify “how critical is the weighing value in the complete
process”
• Based on that evaluation „warning levels“ and „action levels“ are determined
Page 5
Monitoring test equipment– how often ?
Simple rule:• Daily, weekly or monthly in case the weight values are not critical• If weighing values are critical in the whole process, than
calibration should be done before every series of measurement
Perhaps Better:• Dynamic verification and adjustment of intervals based on
calibration results and defined warning and action thresholds
Page 6
Monitoring test equipment
Dynamic verification of calibration intervals
Time
Mea
suring Value
Upper intervention limit
Target Value
Upper warning limit
Lower intervention limit
Lower warning limit
Time
Mea
suring Value
Upper intervention limit
Target Value
Upper warning limit
Lower intervention limit
Lower warning limit
Increase IntervalIncrease Interval
Time
Measuring Value
Upper intervention limit
Target Value
Upper warning limit
Lower intervention limit
Lower warning limit
Time
Measuring Value
Upper intervention limit
Target Value
Upper warning limit
Lower intervention limit
Lower warning limit
Decrease IntervalDecrease Interval
Time
Mea
suring Value
Upper intervention limit
Target Value
Upper warning limit
Lower intervention limit
Lower warning limit
Time
Mea
suring Value
Upper intervention limit
Target Value
Upper warning limit
Lower intervention limit
Lower warning limit
case 1: increase intervals
case 3: call service
case 2: decrease intervals
Monitoring test equipmentHow many and which weights to use ?
Often calibration is done only at one point
of the whole weighing range.
Better:
• Calibration should be performed at three
prominent points of the characteristic
balance curve
• Near the minimum weight, to fulfill the
regulatory requirements of USP
• In the working range, where the balance
is finally used; typically this is 10 to 20%
• Near the maximum load, because at that
point the absolute error is the highest
24. September 2013 79
Monitoring test equipmentThe balance is monitoring itself - isoCAL
• Sartorius Cubis and Secura series
balances are designed for
Pharmaceutical labs and include
functions, which make monitoring easier
for the user
• Automatic calibration/adjustment function
isoCAL
• isoCAL maintains the accuracy of the
balance within an narrow tolerance band
without testing efforts
24. September 2013 80
It corrects:
• deviations regarding to the sensitivity
of the balance
• temperature influence
(temperature limits can be set)
• compensates the influence of gravitation
(location, differences in altitude)
• air pressure, air density and humidity
fluctuations (but it‘s not a compensation)
• aging effects of mechanic and electronic
24. September 2013 81
Monitoring test equipmentThe balance is monitoring itself - isoCAL
One of the worst errors that occurs during
weighing is caused when the balance is
unleveled.
24. September 2013 82
αααα
400µm
Monitoring test equipmentThe balance is monitoring itself – Q-Level
24. September 2013 83
αααα
400µm
Monitoring test equipmentThe balance is monitoring itself – Q-Level
Example:
If the balance is inclined only 400µm ( thickness of a
businesscard) on one side an incorrect measurement is
produced:
md = m · cos α = 200g · cos (0,11°) = 199,99963g
α = arctan (0.4mm/200mm)
∆m = -0.00037g
That means, the mass of 200g is displayed 0.37mg too
less
The resulting error is greater than the allowable
tolerance of a 200g ± 0.3mg E2 calibration weight for
an analytical balance.
The automatic motorised leveling featured
on some balances controls that the
balance is correctly leveled and performs
the time
consuming alignment of the balance.
It is activated by pressing a start key or
fully automatically in conjunction with the
isoCAL function
24. September 2013 84
Monitoring test equipmentThe balance is monitoring itself – Q-Level
Modules
24. September 2013 85
• Qualified weighing
• Terminology
• Avoid errors – and understand their daily impact on your weighing accuracy
• Uncertainty determination
• Regulations in the pharmaceutical laboratory
• Minimum sample weight according to USP
• Qualification of your laboratory balance
• Monitoring your laboratory balance
• Calibration and servicing of your laboratory balances
• Calibration weights - selection, storage and correct handling
• Health and safety during weighing
Calibration and servicing of your laboratory balances
Would you hang up your life on bungee
which was not
regularly checked?
• How can you be sure that the bungee
has the right length for your weight
• Would you trust a balance that is not
calibrated
24. September 2013 86
Calibration and servicing of your laboratory balancesMaintenance
24. September 2013 87
• Qualification is not a one time event.
• The user needs to demonstrate and document with regular testing and regular
preventive maintenance that an instrument consistently performs and is appropriate
for the intended use
• Maintenance and test procedures should be in place to detect and fix problems
before they can have a negative impact on analytical data.
Calibration and servicing of your laboratory balancesCalibration
24. September 2013 88
• Who has the responsibility to define the maintenance frequency?
• What is the right maintenance frequency?
• Who defines the required maintenance work?
• How should maintenance activities be documented?
The auditor is only interested in two things
PROTOCOL
How and
Where is it
documented
PLAN
Why and
How are
things done
Calibration and servicing of your laboratory balancesCalibration vs. Adjustment
24. September 2013 89
Calibration:
• Calibration in metrological meaning is a measuring process for the determination
and documentation of the deviation of a measuring instrument, incorporated by a
reference standard.
• The frequency of calibration should depend on the frequency of use, and the
importance of the measuring instrument in the manufacturing process
• Calibration does not make any changes within the balance
Adjustment:
• Correction of the measurement error to within an allowable level
Calibration and servicing of your laboratory balancesTraceability
Traceability:
A calibration is traceable if the used
standard is related to a national or
international standards by a uninterrupted
chain of comparisons all having stated
uncertainties.
24. September 2013 90
Calibration and servicing of your laboratory balancesAccreditation
24. September 2013 91
The trust placed in calibration certificates depends on the competence of the
calibration laboratory. The accreditation of a calibration laboratory is the verification
of a independent accreditation body that the calibration is done with technical
competence and with an internationally comparative standard.
Accreditation:
• Ensures traceability to a national standard
• Documents measurement results, deviations and uncertainty of measurement
• Creates trust in the quality and safety of products and services.
Modules
92
• Terminology - Avoid errors and understand their daily impact on your weighing
accuracy (Uncertainty)
• Regulations in the pharmaceutical laboratory
• Accurate sample weighing to 0.10% uncertainty according to USP
• Qualification of your laboratory balance
• Monitoring your laboratory balance
Optional:
• Calibration weights - selection, storage and correct handling
• Product presentation Secura, Quintix and Practum
24. September 2013
Calibration weightsselection, storage and correct handling
24. September 2013 93
Calibration Weights
24. September 2013 94
• Calibration weights
• International Measurement Standards
• The International Prototype
• Design
• Class of weights
• Tolerences
• Correct handling and storage of weights
Calibration weights - selection, storage and correct handlingWeight Units
24. September 2013 95
Unit Unit Symbol Relation to Base
Unit
Nanogram ng ÷ µg 1000
Microgram µg ÷ mg 1000
Milligram mg ÷ g 1000
Gram g ÷ 1000
Kilogram kg Base Unit
Ton t x 1000
Calibration weights - selection, storage and correct handling Metrology and standards - all scientifically defined?
24. September 2013 96
m meter
s second
A Ampere
K Kelvin
cd candela
kg kilogram
Length
Time
Electric Current
Temperature
Luminous Intensity
Mass
Calibration weights - selection, storage and correct handling The international prototype kilogram is defined as:
• The mass of the international prototype
which is held at the BIPM in Paris
• Cylinder of Platinum Iridium - 90%
Platinum 10% Iridium 39mm diameter
and density 21.5gcm-3
24. September 2013 97
Calibration weights - selection, storage and correct handling All certified weights are connected to the international 1 kg prototype
24. September 2013 98
BIPM
international
NIST
USA
LNE
France
NMI
e.g. Netherlands
PTB
Germany
No. 52, 1,000 000 125 kg +/- 28 µg
90% platinum10% iridium
1 kg internationalprototype
National Metrological Institutes
Calibration weights - selection, storage and correct handling OIML
Organisation International de
Metrology Legale specifies:
• Design,
• Material,
• Surface,
• Quality,
• Tolerances
• Permissible Markings
24. September 2013 99
Calibration weights - selection, storage and correct handling Classes of calibration weights
24. September 2013 100
E1 Solid Stainless Steel No markings 1kg +/- 0.5mg
E2 Solid stainless Steel No markings 1kg +/- 1.5mg
F1 Stainless Steel with adjustment cavity 1kg +/- 5.0mg
Calibration weights - selection, storage and correct handling Shape of weights
24. September 2013 101
• Construction and shape defined
• Designed to minimise surface area
• Convenient to handle
• Especially so for the “milligram”
weights
• Convenient to handle
• Minimum risk of physical damage
• F1 leaf weights
• E2/E1 wire weights
Calibration weights - selection, storage and correct handling Handling and care of weights
store in boxes supplied
24. September 2013 102
suitable for clean room conditions
Calibration weights - selection, storage and correct handling Handling and care of weights
Use lifting devices:
• Plastic forks or tipped forceps –
keep them clean!
• Cotton or chamois gloves
24. September 2013 103
Handling and care of weights
• Remove dust with (camel hair) brush
• Never touch weights
• Fingerprints are acidic and greasy
• Polish your weights with cotton or
chamois
• Ethyl Alcohol is to remove absorption
layer or grease - during recalibration
• Weights become heavier with deposits
• Weights become lighter with wear and
tear
• Regular calibration and cleaning is
recommended – After 3 performed
recertifications it is possible to give
recommendations for future calibration
intervals
24. September 2013 104
Calibration of weights
24. September 2013 105
• Class E1 & E2 Weights can´t be
adjusted but only calibrated.
• The calibration certificate of a weight
states the uncertainty of the weight.
Thank you very much foryour kind attention