Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
Roles and responsibilities in Roles and responsibilities in
traceabilitytraceability
implementationimplementation
Federica BragaFederica Braga
University of Milan Medical School University of Milan Medical School
12th International 12th International ScientificScientific Meeting Meeting
STANDARDIZATION IN LABORATORY MEDICINE STANDARDIZATION IN LABORATORY MEDICINE
AND PATIENT SAFETYAND PATIENT SAFETY
Laboratory measurement paradigm:Laboratory measurement paradigm:
•• Assays that claim to measure the same Assays that claim to measure the same analyteanalyte
should give equivalent measurement results (for should give equivalent measurement results (for
long term and within clinically meaningful limits)long term and within clinically meaningful limits)
•• Measurement results should be independent of:Measurement results should be independent of:
−− TimeTime
−− Location/laboratoryLocation/laboratory
−− Assay systemAssay system
Laboratory results should be Laboratory results should be
equivalent no matter where they equivalent no matter where they
are performedare performed
ISO/EN 17511 ISO/EN 17511 -- MeasurementMeasurement of of quantitiesquantities in in samplessamples of of biologicalbiological originorigin -- MetrologicalMetrological traceabilitytraceability of of
valuesvalues assignedassigned toto calibratorscalibrators and control and control materialsmaterials..
Objective of traceability Objective of traceability
implementation:implementation:
to enable the results obtained by the to enable the results obtained by the
calibrated routine procedure to be calibrated routine procedure to be
expressed in terms of the values obtained at expressed in terms of the values obtained at
the highest available level of the calibration the highest available level of the calibration
hierarchyhierarchy..
EU 98/79/ECEU 98/79/EC--IVD IVD DirectiveDirective
To become To become equivalent for long termequivalent for long term, ,
results must be traceable to higherresults must be traceable to higher--
order referencesorder references
requires manufacturers to ensure requires manufacturers to ensure
traceability of their analytical systems to traceability of their analytical systems to
recognized higher order referencesrecognized higher order references
REGULATION (EU) 2017/746REGULATION (EU) 2017/746 OF THE EUROPEAN PARLIAMENT OF THE EUROPEAN PARLIAMENT
AND OF THE COUNCIL of 5 April 2017 on AND OF THE COUNCIL of 5 April 2017 on in vitro in vitro diagnostic diagnostic
medical devices and medical devices and repealing Directive 98/79/ECrepealing Directive 98/79/EC and and
Commission Decision 2010/227/EUCommission Decision 2010/227/EU
•• Establishment of a calibration hierarchy Establishment of a calibration hierarchy
starting from the unequivocal definition of starting from the unequivocal definition of
the the measurandmeasurand
•• Elimination of measurement biasElimination of measurement bias
•• Adequate estimation of measurement Adequate estimation of measurement
uncertaintyuncertainty
Basic requirements to Basic requirements to
establish traceabilityestablish traceability
Primary Ref. ProcedurePrimary Ref. Procedure
Secondary Ref. Secondary Ref.
ProcedureProcedure
ManufacturerManufacturer’’ss
Internal ProcedureInternal Procedure
EndEnd--useruser’’s Routines Routine
ProcedureProcedure
Primary Reference MaterialPrimary Reference Material
(e.g. pure (e.g. pure analyteanalyte))
Secondary Ref. MaterialSecondary Ref. Material
(matrix(matrix--based)based)
ManufacturerManufacturer’’s Calibrators Calibrator
Routine SampleRoutine Sample
Test ResultTest Result
SI UnitsSI Units
Tra
cea
bility
Tra
cea
bility
*Adapted from ISO 17511
Reference Measurement SystemReference Measurement SystemU
nce
rtain
tyU
nce
rtain
ty
2. Diagnostic manufacturers:
Implement suitable measuring
systems (platform, reagents,
calibrators, controls) fulfilling the
above established goals
2. 2. DiagnosticDiagnostic manufacturersmanufacturers::
ImplementImplement suitablesuitable measuringmeasuring
systemssystems ((platformplatform, , reagentsreagents, ,
calibratorscalibrators, , controlscontrols) ) fulfillingfulfilling the the
aboveabove establishedestablished goalsgoals
PanteghiniPanteghini M,M,
ClinClin ChemChem Lab Lab MedMed
2010;48:72010;48:7
Roles and responsibilities in traceability implementationRoles and responsibilities in traceability implementation
3. End users (clinical laboratories):3. End users (clinical laboratories):
Survey assay and laboratory performance through:Survey assay and laboratory performance through:
-- IQC component I: testing system controls to confirm and verify IQC component I: testing system controls to confirm and verify manufacturermanufacturer’’s declared performance s declared performance
(CE marked (CE marked –– virtually unbiased)virtually unbiased)
-- IQC component IIIQC component II: : estimation of the MEASUREMENT UNCERTAINTY due to the random effeestimation of the MEASUREMENT UNCERTAINTY due to the random effects and cts and
calculation of uncertainty of laboratory measurements (calculation of uncertainty of laboratory measurements (UUcalcal + + UUrandomrandom))
-- EQA: true value in commutable materials for defining measuremenEQA: true value in commutable materials for defining measurement error of laboratoryt error of laboratory
1. Profession (e.g., IFCC, JCTLM):
Define analytical objectives:
reference measurement systems
(traceability chain) and associated
clinically acceptable
uncertainty (fit for purpose)
1. 1. ProfessionProfession (e.g., IFCC, JCTLM):(e.g., IFCC, JCTLM):
Define analytical objectives: Define analytical objectives:
reference measurement systems reference measurement systems
(traceability chain) and associated (traceability chain) and associated
clinically acceptable clinically acceptable
uncertainty (fit for purpose)uncertainty (fit for purpose)
2. Diagnostic manufacturers:
Implement suitable measuring
systems (platform, reagents,
calibrators, controls) fulfilling the
above established goals
2. 2. DiagnosticDiagnostic manufacturersmanufacturers::
ImplementImplement suitablesuitable measuringmeasuring
systemssystems ((platformplatform, , reagentsreagents, ,
calibratorscalibrators, , controlscontrols) ) fulfillingfulfilling the the
aboveabove establishedestablished goalsgoals
PanteghiniPanteghini M,M,
ClinClin ChemChem Lab Lab MedMed
2010;48:72010;48:7
Roles and responsibilities in traceability implementationRoles and responsibilities in traceability implementation
3. End users (clinical laboratories):3. End users (clinical laboratories):
Survey assay and laboratory performance through:Survey assay and laboratory performance through:
-- IQC component I: testing system controls to confirm and verify IQC component I: testing system controls to confirm and verify manufacturermanufacturer’’s declared performance s declared performance
(CE marked (CE marked –– virtually unbiased)virtually unbiased)
-- IQC component IIIQC component II: : estimation of the MEASUREMENT UNCERTAINTY due to the random effeestimation of the MEASUREMENT UNCERTAINTY due to the random effects and cts and
calculation of uncertainty of laboratory measurements (calculation of uncertainty of laboratory measurements (UUcalcal + + UUrandomrandom))
-- EQA: true value in commutable materials for defining measuremenEQA: true value in commutable materials for defining measurement error of laboratoryt error of laboratory
1. Profession (e.g., IFCC, JCTLM):
Define analytical objectives:
reference measurement systems
(traceability chain) and associated
clinically acceptable
uncertainty (fit for purpose)
1. 1. ProfessionProfession (e.g., IFCC, JCTLM):(e.g., IFCC, JCTLM):
Define analytical objectives: Define analytical objectives:
reference measurement systems reference measurement systems
(traceability chain) and associated (traceability chain) and associated
clinically acceptable clinically acceptable
uncertainty (fit for purpose)uncertainty (fit for purpose)
1.1. Roles and responsibilities ofRoles and responsibilities of
Professional OrganizationProfessional Organization
••UnequivocalUnequivocal definitiondefinition of the of the
MEASURANDMEASURAND asas the the quantityquantity
subjectsubject toto measurementmeasurement
•• DefinitionDefinition of the of the REFERENCE REFERENCE
MEASUREMENT SYSTEMMEASUREMENT SYSTEM
•• DefinitionDefinition of of ANALYTICAL ANALYTICAL
PERFORMANCE SPECIFICATIONSPERFORMANCE SPECIFICATIONS
MEASURANDMEASURANDMEASURAND
TYPE A MEASURANDSTYPE A MEASURANDS
• Well defined Well defined
compoundscompounds
•• Traceable to SI unitsTraceable to SI units
•• Results are not Results are not
methodmethod--dependentdependent
•• Approx. 65 Approx. 65 analytesanalytes
(metabolites, (metabolites,
electrolytes)electrolytes)
TYPE B MEASURANDSTYPE B MEASURANDS
• OftenOften not well defined not well defined
(heterogeneous mixtures)(heterogeneous mixtures)
•• AnalytesAnalytes can be bound or in can be bound or in
free statefree state
•• Not traceable to SI units, Not traceable to SI units,
but to arbitrary units (e.g. but to arbitrary units (e.g.
WHO International Units)WHO International Units)
•• Immunochemical Immunochemical
procedures show inherent procedures show inherent
variability (different variability (different epitopesepitopes))
•• 400400--600 600 analytesanalytes ((tumourtumour
markers, viral antigens, markers, viral antigens,
clotting factors)clotting factors)
PanteghiniPanteghini M. M. ClinClin BiochemBiochem 2009;42:2362009;42:236
Joint Committee for Traceability in Joint Committee for Traceability in
Laboratory Medicine (JCTLM)Laboratory Medicine (JCTLM)
The WorldThe World’’s only qualitys only quality--assured database of:assured database of:
a)a) Higher Order Reference MaterialsHigher Order Reference Materials
b)b) Higher Order Reference Measurement ProceduresHigher Order Reference Measurement Procedures
c)c) Accredited Laboratory Reference Measurement Services Accredited Laboratory Reference Measurement Services
For use by (primarily):For use by (primarily):
a) IVD industry (a) IVD industry (to assist them in following the EU Directive on compliance and to assist them in following the EU Directive on compliance and
traceability of commercial systemstraceability of commercial systems))
b) Regulators (to verify that results produced by b) Regulators (to verify that results produced by IVDsIVDs are traceable to)are traceable to)
REFERENCE MEASUREMENT SYSTEMREFERENCE MEASUREMENT SYSTEMREFERENCE MEASUREMENT SYSTEM
ANALYTICAL PERFORMANCE
SPECIFICATIONS
ANALYTICAL PERFORMANCE ANALYTICAL PERFORMANCE
SPECIFICATIONSSPECIFICATIONS
Some models are better suited for certain Some models are better suited for certain
measurandsmeasurands than for others. The attention is than for others. The attention is
therefore primarily directed towards the therefore primarily directed towards the
measurandmeasurand and its biological and clinical and its biological and clinical
characteristics.characteristics.
•• Model 1: Based on the effect of analytical Model 1: Based on the effect of analytical
performance on clinical outcomeperformance on clinical outcome
•• Model 2: Based on components of biological Model 2: Based on components of biological
variation of the variation of the measurandmeasurand
•• Model 3: Based on state of the art of the Model 3: Based on state of the art of the
measurement (i.e., the highest level of measurement (i.e., the highest level of
analytical performance technically achievable)analytical performance technically achievable)
The The measurandmeasurand hashas a high a high
homeostatichomeostatic controlcontrol
NeitherNeither centralcentral diagnosticdiagnostic
rolerole nornor sufficientsufficient
homeostatichomeostatic controlcontrolThe The measurandmeasurand has a central has a central
role in diagnosis and role in diagnosis and
monitoring of a specific monitoring of a specific
disease disease
a. Done by direct outcome studies – investigating the impact of analytical performance
of the test on clinical outcome.
b. Done by indirect outcome studies – investigating the impact of analytical
performance of the test on clinical classifications or decisions and thereby on the
probability of patient outcome, e.g., by simulation analysis.
a.a. Done by direct outcome studies Done by direct outcome studies –– investigating the impact of analytical performance investigating the impact of analytical performance
of the test on clinical outcome.of the test on clinical outcome.
b. Done by indirect outcome studies b. Done by indirect outcome studies –– investigating the impact of analytical investigating the impact of analytical
performance of the test on clinical classifications or decisionsperformance of the test on clinical classifications or decisions and thereby on the and thereby on the
probability of patient outcome, e.g., by simulation analysis.probability of patient outcome, e.g., by simulation analysis.
IMPRECISION:IMPRECISION: ≤≤ 0.25 CV0.25 CVII (O)(O)
≤≤ 0.5 CV0.5 CVII (D)(D)
≤≤ 0.75 CV0.75 CVI I (M)(M)
BIAS: BIAS: < 0.125 (CV< 0.125 (CVII22 + CV+ CV
GG22))1/21/2 (O)(O)
< 0.25 (CV< 0.25 (CVII22 + CV+ CV
GG22))1/21/2 (D)(D)
< 0.375 (CV< 0.375 (CVII22 + CV+ CV
GG22))1/21/2 (M) (M)
IntraIntra--individualindividual BV: BV: CVCVII%%
InterInter--individualindividual BV: BV: CVCVGG%%
OUTCOME APPROACHOUTCOME APPROACHOUTCOME APPROACH
BIOLOGICAL VARIATION APPROACHBIOLOGICAL VARIATION APPROACHBIOLOGICAL VARIATION APPROACH
<0.75 x CV<0.75 x CVII (Minimum)(Minimum)
<0.50 x CV<0.50 x CVII ((DesirableDesirable))
<0.25 x CV<0.25 x CVII (Optimum)(Optimum)
[Note that these are goals for [Note that these are goals for random variabilityrandom variability, as , as at at
the calibrator level the systematic error (bias), in the calibrator level the systematic error (bias), in
agreement with the metrological traceability theory, agreement with the metrological traceability theory,
must be corrected if present in a non negligible amount]must be corrected if present in a non negligible amount]
<(0.75 x CV<(0.75 x CVII) x 2 (Minimum)) x 2 (Minimum)
<(0.50 x CV<(0.50 x CVII) x 2 () x 2 (DesirableDesirable))
<(0.25 x CV<(0.25 x CVII) x 2 (Optimum)) x 2 (Optimum)
kuU ×=Total budget for uncertainty (GTotal budget for uncertainty (G
UU) should be fulfilled by:) should be fulfilled by:
-- uncertainty of the measuring system employed in the uncertainty of the measuring system employed in the
individual laboratory (random uncertainty) +individual laboratory (random uncertainty) +
-- that accumulated along all the steps of metrological that accumulated along all the steps of metrological
traceability chain andtraceability chain and
-- expanded by a coverage factor (k=2 for a 95%Coverage expanded by a coverage factor (k=2 for a 95%Coverage
Interval)Interval)
APS FOR UNCERTAINTYAPS FOR UNCERTAINTY
APS FOR EXPANDEDAPS FOR EXPANDED
UNCERTAINTYUNCERTAINTY
Measuring system imprecision
Measuring system calibration
uncertainty
Individual laboratory
performance
Measurement
uncertainty
budget
Uncertainty of
references
Measurand definition
Patient result
Un
certa
inty
ClinClin BiochemBiochem 57 (2018) 757 (2018) 7--1111
Uncertainty of measurement must be defined across the entire
traceability chain,
starting with the provider of reference materials,
extending through the IVD manufacturers and their processes for
assignment of calibrator values, and
ultimately to the final result reported to clinicians by end users (i.e.
clinical laboratories).
Measuring system imprecision
Measuring system calibration
uncertainty
Individual laboratory
performance
Measurement
uncertainty
budget
Uncertainty of
references
Measurand definition
Patient result
Un
certa
inty
ClinClin BiochemBiochem 57 (2018) 757 (2018) 7--1111
33%33%
50%50%
100%100%
To obtain that a given measurement can fulfill the established GTo obtain that a given measurement can fulfill the established GUU
when clinical samples are measured, it is necessary to when clinical samples are measured, it is necessary to define how define how
much of the Gmuch of the GUU should be used across the different steps of should be used across the different steps of
metrological traceability chainmetrological traceability chain, by establishing combined , by establishing combined
uncertainty limits at different chain levels as a proportion of uncertainty limits at different chain levels as a proportion of GGUU
Editorial
Foreword
Mauro Panteghini
Opinion Piece
Measurement uncertainty: Friend or foe?
Ilenia Infusino, Mauro Panteghini
Defining permissible limits for the combined uncertainty budget in the implementation of metrological traceability
Federica Braga, Mauro Panteghini
Performance specifications and six sigma theory: Clinical chemistry and industry compared
Wayne P. Oosterhuis, Michel J. Severens
Reviews
What information on measurement uncertainty should be communicated to clinicians, and how?
Mario Plebani, Laura Sciacovelli, Daniela Bernardi, Ada Aita, Giorgia Antonelli, Andrea Padoan
The role of external quality assessment in the verification of in vitro medical diagnostics in the traceability era
Federica Braga, Sara Pasqualetti, Mauro Panteghini
Opinion Piece
Sources and performance criteria of uncertainty of reference measurement procedures
Andrea Mosca, Renata Paleari
Deriving proper measurement uncertainty form Internal Quality Control data: An impossible mission?
Ferruccio Ceriotti
Original Research
Measurement uncertainty in laboratory reports: A tool for improving the interpretation of test results
Andrea Padoan, Laura Sciacovelli, Ada Aita, Giorgia Antonelli, Mario Plebani
Ensuring suitable quality of clinical measurements through design
Anthony Orzechowski, Victoria Petrides, Richard Scopp
Different top-down approaches to estimate measurement uncertainty of whole blood tacrolimus mass concentration values
Raül Rigo-Bonnin, Aurora Blanco-Font, Francesca Canalias
Short Communication
Random uncertainty of photometric determination of hemolysis index on the Abbott Architect c16000 platform
Elena Aloisio, Assunta Carnevale, Sara Pasqualetti, Sarah Birindelli, Alberto Dolci, Mauro Panteghini
Special Issue: Measurement Uncertainty in Medical Laboratories: Friend or Foe?Guest Editors: Federica Braga and Mauro Panteghini
2. Diagnostic manufacturers:
Implement suitable measuring
systems (platform, reagents,
calibrators, controls) fulfilling the
above established goals
2. 2. DiagnosticDiagnostic manufacturersmanufacturers::
ImplementImplement suitablesuitable measuringmeasuring
systemssystems ((platformplatform, , reagentsreagents, ,
calibratorscalibrators, , controlscontrols) ) fulfillingfulfilling the the
aboveabove establishedestablished goalsgoals
PanteghiniPanteghini M,M,
ClinClin ChemChem Lab Lab MedMed
2010;48:72010;48:7
Roles and responsibilities in traceability implementationRoles and responsibilities in traceability implementation
3. End users (clinical laboratories):3. End users (clinical laboratories):
Survey assay and laboratory performance through:Survey assay and laboratory performance through:
-- IQC component I: testing system controls to confirm and verify IQC component I: testing system controls to confirm and verify manufacturermanufacturer’’s declared performance s declared performance
(CE marked (CE marked –– virtually unbiased)virtually unbiased)
-- IQC component IIIQC component II: : estimation of the MEASUREMENT UNCERTAINTY due to the random effeestimation of the MEASUREMENT UNCERTAINTY due to the random effects and cts and
calculation of uncertainty of laboratory measurements (calculation of uncertainty of laboratory measurements (UUcalcal + + UUrandomrandom))
-- EQA: true value in commutable materials for defining measuremenEQA: true value in commutable materials for defining measurement error of laboratoryt error of laboratory
1. Profession (e.g., IFCC, JCTLM):
Define analytical objectives:
reference measurement systems
(traceability chain) and associated
clinically acceptable
uncertainty (fit for purpose)
1. 1. ProfessionProfession (e.g., IFCC, JCTLM):(e.g., IFCC, JCTLM):
Define analytical objectives: Define analytical objectives:
reference measurement systems reference measurement systems
(traceability chain) and associated (traceability chain) and associated
clinically acceptable clinically acceptable
uncertainty (fit for purpose)uncertainty (fit for purpose)
2.2. Roles and responsibilities ofRoles and responsibilities of
Diagnostic manufacturersDiagnostic manufacturers
••Identification of higher order metrological Identification of higher order metrological
REFERENCESREFERENCES
••Definition of a Definition of a CALIBRATION HIERARCHYCALIBRATION HIERARCHY to to
assign traceable values to their system assign traceable values to their system
calibrators calibrators
•• Estimation of combined expanded Estimation of combined expanded
MEASUREMENT UNCERTAINTYMEASUREMENT UNCERTAINTY of calibratorsof calibrators
••Verification that Verification that UNCERTAINTY GOALSUNCERTAINTY GOALS are are
fullfilledfullfilled
EU 98/79/ECEU 98/79/EC--IVD IVD
DirectiveDirective
[[FulfillmentFulfillment of of thethe EU IVD EU IVD Directive andDirective and
REGULATION (EU) 2017/746 RequirementsREGULATION (EU) 2017/746 Requirements]]
1. IDENTIFICATION OF HIGHER ORDER REFERENCES
httpshttps://www.bipm.org/://www.bipm.org/jctlmjctlm//
2. DEFINITION OF A CALIBRATION HIERARCHY TO ASSIGN
TRACEABLE VALUES TO THEIR SYSTEM CALIBRATORS
ERMERM--DA470DA470
U.S. National Reference U.S. National Reference
Preparation no. 12Preparation no. 12--0575C0575C
ERMERM--DA470k/IFCCDA470k/IFCC
ManufacturerManufacturer’’s working calibrators working calibrator
(master lot)(master lot)
ManufacturerManufacturer’’s product calibrators product calibrator
Routine sample resultRoutine sample result
value transfer protocolvalue transfer protocol
ManufacturerManufacturer’’s standing s standing
immunoassayimmunoassay
Commercial immunoassayCommercial immunoassay
value transfer value transfer
protocolprotocol
value transfer protocolvalue transfer protocolDiagnostic manufacturersDiagnostic manufacturers
concentration
coverage interval
B
best
estimate
uncertainty
bias="error"
CRM
assigned
CRM
measured
A
uncertaintyuncertainty
[[AdaptedAdapted fromfrom KallnerKallner A, A,
Scand J Scand J ClinClin & Lab Invest 2010; & Lab Invest 2010;
70(Suppl 242): 34]70(Suppl 242): 34]
3. ESTIMATION OF MEASUREMENT 3. ESTIMATION OF MEASUREMENT
UNCERTAINTY OF CALIBRATORSUNCERTAINTY OF CALIBRATORS
4. VERIFICATION THAT UNCERTAINTY 4. VERIFICATION THAT UNCERTAINTY
GOALS ARE FULLFILLED GOALS ARE FULLFILLED
GUMGUM
Eurachem/
CITAC Guide
Eurachem/
CITAC Guide
CLSI document EP29-ACLSI document EP29-A
PARADIGM SHIFT IN THE THINKINGPARADIGM SHIFT IN THE THINKING
•• IfIf the the manufacturermanufacturer assumesassumes total total responsibilityresponsibility forfor supplyingsupplying productsproducts of of acceptableacceptable qualityquality in in termsterms of of traceabilitytraceability and and uncertaintyuncertainty of the system (of the system (““CE CE markedmarked””), ), itit isisno no longerlonger possiblepossible toto considerconsider separatelyseparately the the componentscomponents of of eacheach measuringmeasuring system (i.e., system (i.e., platformplatform, , reagentsreagents, , calibratorscalibrators and control and control materialsmaterials), ), whichwhich in in termsterms of performance can of performance can onlyonly bebe guaranteedguaranteed and and certifiedcertified byby the the manufacturermanufacturer asas a a wholewhole..
•• AnyAny changechange introducedintroduced byby usersusers or or thirdthirdpartiesparties (e.g., the (e.g., the useuse of of reagentsreagents, , calibratorscalibrators or or control control materialsmaterials fromfrom otherother supplierssuppliers) ) maymaysignificantlysignificantly alter the alter the qualityquality of the of the measuringmeasuringsystem performance, system performance, removingremoving anyanyresponsibilityresponsibility fromfrom the the manufacturermanufacturer and and deprivingdepriving the system (and, the system (and, consequentlyconsequently, the , the producedproduced resultsresults) of the ) of the certificationcertification originallyoriginallyprovidedprovided through CE through CE markingmarking..
2. Diagnostic manufacturers:
Implement suitable measuring
systems (platform, reagents,
calibrators, controls) fulfilling the
above established goals
2. 2. DiagnosticDiagnostic manufacturersmanufacturers::
ImplementImplement suitablesuitable measuringmeasuring
systemssystems ((platformplatform, , reagentsreagents, ,
calibratorscalibrators, , controlscontrols) ) fulfillingfulfilling the the
aboveabove establishedestablished goalsgoals
PanteghiniPanteghini M,M,
ClinClin ChemChem Lab Lab MedMed
2010;48:72010;48:7
Roles and responsibilities in traceability implementationRoles and responsibilities in traceability implementation
3. End users (clinical laboratories):3. End users (clinical laboratories):
Survey assay and laboratory performance through:Survey assay and laboratory performance through:
-- IQC component I: testing system controls to confirm and verify IQC component I: testing system controls to confirm and verify manufacturermanufacturer’’s declared performance s declared performance
(CE marked (CE marked –– virtually unbiased)virtually unbiased)
-- IQC component IIIQC component II: : estimation of the MEASUREMENT UNCERTAINTY due to the random effeestimation of the MEASUREMENT UNCERTAINTY due to the random effects and cts and
calculation of uncertainty of laboratory measurements (calculation of uncertainty of laboratory measurements (UUcalcal + + UUrandomrandom))
-- EQA: true value in commutable materials for defining measuremenEQA: true value in commutable materials for defining measurement error of laboratoryt error of laboratory
1. Profession (e.g., IFCC, JCTLM):
Define analytical objectives:
reference measurement systems
(traceability chain) and associated
clinically acceptable
uncertainty (fit for purpose)
1. 1. ProfessionProfession (e.g., IFCC, JCTLM):(e.g., IFCC, JCTLM):
Define analytical objectives: Define analytical objectives:
reference measurement systems reference measurement systems
(traceability chain) and associated (traceability chain) and associated
clinically acceptable clinically acceptable
uncertainty (fit for purpose)uncertainty (fit for purpose)
3.3. Roles and responsibilities ofRoles and responsibilities of
Clinical laboratoriesClinical laboratories
•• Verification of availability and quality of Verification of availability and quality of
INFORMATION INFORMATION about IVD metrological about IVD metrological
traceability and uncertaintytraceability and uncertainty
•• DAILY SURVEILLANCEDAILY SURVEILLANCE of IVD system of IVD system
traceabilitytraceability
••Estimation of the Estimation of the MEASUREMENT MEASUREMENT
UNCERTAINTYUNCERTAINTY due to the random effects due to the random effects
and calculation of uncertainty of and calculation of uncertainty of
laboratory measurements (laboratory measurements (UUcalcal + + UUrandomrandom))
Post-marketingsurveillance
PostPost--marketingmarketing
surveillancesurveillance
In principle, laboratory users should be able to access the In principle, laboratory users should be able to access the
following:following:
a) an indication of higher order references (materials and/or procedures) used to assign traceable values to calibrators,
b) which internal calibration hierarchy has been applied by the manufacturer, and
c) a detailed description of each step,
d) the expanded combined uncertainty value of commercial calibrators, and
e) which, if any, acceptable limits for uncertainty of calibrators were applied in the validation of the analytical system.
a) an indication of higher order references (materials and/or procedures) used to assign traceable values to calibrators,
b) which internal calibration hierarchy has been applied by the manufacturer, and
c) a detailed description of each step,
d) the expanded combined uncertainty value of commercial calibrators, and
e) which, if any, acceptable limits for uncertainty of calibrators were applied in the validation of the analytical system.
Braga F & Panteghini M, Clin Chim Acta 2014;432:55Braga F & Panteghini M, Clin Chim Acta 2014;432:55
(ideally all this information should be available in (ideally all this information should be available in
the assay or calibrator package inserts)the assay or calibrator package inserts)
VERIFICATION OF AVAILABILITY AND QUALITY OF INFORMATIONVERIFICATION OF AVAILABILITY AND QUALITY OF INFORMATION
Currently, the full information about
calibration is usually not available
Currently, the full information about
calibration is usually not available
Some Some organisationsorganisations are are frequentlyfrequently
mentionedmentioned ((oftenoften withoutwithout explanationexplanation): ):
usedused asas a a ““trustedtrusted brandbrand””
•• NIST, IRMM, IFCC, CLSI (NIST, IRMM, IFCC, CLSI (protocolsprotocols))
GCGC--IDMS @ NISTIDMS @ NIST
NIST SRM 917NIST SRM 917
ManufacturerManufacturer’’s s
internal procedureinternal procedure
Commercial Commercial
systemsystem
NIST SRM 965NIST SRM 965
(glucose in human serum)(glucose in human serum)
CommercialCommercial
calibratorcalibrator
PatientPatient’’s sample resultss sample results
GCGC--IDMS IDMS [accredited reference laboratory][accredited reference laboratory]
ManufacturerManufacturer’’s internals internal
procedureprocedure
CommercialCommercial
systemsystem
CommercialCommercial
calibratorcalibrator
PatientPatient’’s sample resultss sample results
ComparisonComparison onon
biologicalbiological samplessamples
AA BB
CC DD NIST SRM 917NIST SRM 917
ManufacturerManufacturer’’s s
internal procedureinternal procedure
NIST SRM 917NIST SRM 917
CDC CDC HexokinaseHexokinase[accredited reference laboratory][accredited reference laboratory]
ManufacturerManufacturer’’s internals internal
procedureprocedure
CommercialCommercial
systemsystem
CommercialCommercial
calibratorcalibrator
PatientPatient’’s sample resultss sample results
ComparisonComparison onon
biologicalbiological samplessamples
NIST SRM 917NIST SRM 917
Commercial Commercial
systemsystem
CommercialCommercial
calibratorcalibrator
PatientPatient’’s sample resultss sample results
Types of metrological chains that can be used to implement the tTypes of metrological chains that can be used to implement the traceability of blood glucose results*raceability of blood glucose results*
Braga F & Panteghini M, Clin Chim Acta 2014;432:55Braga F & Panteghini M, Clin Chim Acta 2014;432:55
*all*all JCTLM JCTLM recognizedrecognized
Chain A = 1.45% vs. Chain C = 3.26%
Are the Are the measuringmeasuring systemssystems commerciallycommercially availableavailable forfor glucoseglucose determinationdetermination ableable
toto achieveachieve the the desirabledesirable limitlimit forfor combinedcombined uncertaintyuncertainty in a in a clinicalclinical settingsetting??
uurefref
(u(u22refref ++ uu22
calcal))½½
((uu22ref ref + + uu22
calcal + + uu22randomrandom))½½
System imprecision
System calibration
uncertainty
Individual lab
performance
(IQC safety margin)
Measurement
uncertainty
budget
Uncertainty of
references
Measurand definition
Patient result
5.4%5.4% desirabledesirable
Braga F & Panteghini M, Clin Chim Acta 2014;432:55Braga F & Panteghini M, Clin Chim Acta 2014;432:55
Pasqualetti S, Braga F, Panteghini M, Pasqualetti S, Braga F, Panteghini M, Clin Biochem 2017; 50:587Clin Biochem 2017; 50:587--594.594.
L M H
•• Measured in triplicate in 3 Measured in triplicate in 3
consecutive daysconsecutive days byby participatingparticipating
laboratorieslaboratories
59
.9 U
/L5
9.9
U/L
18
6.9
U/L
18
6.9
U/L
40
1.5
U/L
40
1.5
U/L
•• ALP target values ALP target values
assigned with assigned with IFCC RMPIFCC RMP
The importance of the postThe importance of the post--marketing surveillancemarketing surveillance
EQA experimentEQA experiment
(performed 5(performed 5 years later years later the availability of the IFCC RMPthe availability of the IFCC RMP) )
to evaluate the level of ALP standardizationto evaluate the level of ALP standardization
The ability to meet or not the
desirable APS for bias is clearly
dependent on the measuring
system used.
The ability to meet or not the
desirable APS for bias is clearly
dependent on the measuring
system used.
Schumann G, et al.
Clin Chem Lab Med 2011;49:1439-46
Schumann G, et al.Schumann G, et al.
ClinClin ChemChem Lab Lab MedMed 2011;49:14392011;49:1439--4646
The importance of the postThe importance of the post--marketing surveillancemarketing surveillance
CompanyCompany PlatformPlatformPrinciple of Principle of
commercial methodcommercial methodCalibratorCalibrator
Declared standard Declared standard
uncertaintyuncertainty
HigherHigher--order reference order reference
employedemployed
Abbott Abbott ArchitectArchitect pp--NPP/AMPNPP/AMP Calibration factorCalibration factor NANA IFCC reference method (2011)IFCC reference method (2011)
pp--NPP/AMPNPP/AMP Calibration factorCalibration factor NANA Molar extinction coefficientMolar extinction coefficient
Beckman Beckman AUAU IFCC (1983)IFCC (1983) System calibratorSystem calibrator 6.00%6.00% Beckman Coulter Master CalibratorBeckman Coulter Master Calibrator
pp--NPP/DEANPP/DEA System calibratorSystem calibrator NANA Beckman Coulter Master CalibratorBeckman Coulter Master Calibrator
SynchronSynchron pp--NPP/AMPNPP/AMP Enzyme Enzyme ValidatorValidator level 1level 1 6.22%6.22% IFCC reference method (2011)IFCC reference method (2011)
Enzyme Enzyme ValidatorValidator level 2level 2 1.86%1.86% IFCC reference method (2011)IFCC reference method (2011)
pp--NPP/AMPNPP/AMP Enzyme Enzyme ValidatorValidator level 1level 1 3.64%3.64% DGKC standard methodDGKC standard method
Enzyme Enzyme ValidatorValidator level 2level 2 1.27%1.27% DGKC standard methodDGKC standard method
Roche Roche CobasCobas cc IFCC gen. 2IFCC gen. 2 C.f.a.sC.f.a.s.. 0.59%0.59% IFCC reference method (1983)IFCC reference method (1983)
IntegraIntegra IFCC gen. 2IFCC gen. 2 C.f.a.sC.f.a.s 1.22%1.22% IFCC reference method (1983)IFCC reference method (1983)
ModularModular IFCC liquidIFCC liquid C.f.a.sC.f.a.s 1.65%1.65% IFCC reference method (1983)IFCC reference method (1983)
SiemensSiemens Dimension VistaDimension Vista pp--NPP/AMPNPP/AMP ALPI calibratorALPI calibrator 4.51%4.51%cc IFCC reference method (2011)IFCC reference method (2011)
AdviaAdvia pp--NPP/AMPNPP/AMP Chemistry calibrator control 1Chemistry calibrator control 1 3.70%3.70% IFCC reference method (2011)IFCC reference method (2011)
Chemistry calibrator control 2Chemistry calibrator control 2 1.00%1.00% IFCC reference method (2011)IFCC reference method (2011)
pp--NPP/DEANPP/DEA Chemistry calibrator control 1Chemistry calibrator control 1 1.40%1.40%cc Molar extinction coefficientMolar extinction coefficient
Chemistry calibrator control 2Chemistry calibrator control 2 1.30%1.30%cc Molar extinction coefficientMolar extinction coefficient
• The availability of an internationally agreed reference
measurement system does not automatically mean that
the traceability to it is implemented!
• If a post-marketing surveillance is lacking, alternatives
that do not comply with the EU Directive can remain
undisturbed on the market.
• The availability of an internationally agreed reference The availability of an internationally agreed reference
measurement system does not automatically mean that measurement system does not automatically mean that
the traceability to it is implemented!the traceability to it is implemented!
•• If a postIf a post--marketing surveillance is lacking, alternatives marketing surveillance is lacking, alternatives
that do not comply with the EU Directive can remain that do not comply with the EU Directive can remain
undisturbed on the market.undisturbed on the market.
Roche recalibrated Roche recalibrated
its analytical its analytical
methods for the methods for the
ALP measurement ALP measurement
according to 2011 according to 2011
IFCC RMP!IFCC RMP!
Trueness evaluation and verification of interTrueness evaluation and verification of inter--assay agreement of 11 serum assay agreement of 11 serum
immunoglobulin A measuring systems: implications for medical decimmunoglobulin A measuring systems: implications for medical decisionsisions
Braga F, Infusino I, Frusciante E, Ceriotti F, Panteghini M [under review]
Evaluating the status of standardization of available Evaluating the status of standardization of available IgAIgA measuring systems to measuring systems to
investigate if a common lower reference limit can be implementedinvestigate if a common lower reference limit can be implemented to identify to identify
an an IgAIgA deficiencydeficiency, , especially for pediatric populationespecially for pediatric population
• Trueness evaluationERM-DA470k/IFCCCertified value: 1.80 g/L
Expanded uncertainty (U) (k =2): 0.05 g/L
[immediately shipped to centres]
• Inter-assay agreement18 SERUM POOLS- 5 ≤0.7 g/L
- 10 between 0.7 to 4.0 g/L
- 3 ≥ 4.0 g/L
[stored at -80°C until the distribution to centres]
AimAim
Laboratories were requested to perform analysis within two days Laboratories were requested to perform analysis within two days of delivery, in a single of delivery, in a single
analytical run, by applying the manufactureranalytical run, by applying the manufacturer’’s recommended instructions and strictly s recommended instructions and strictly
following a specific analysis sequencefollowing a specific analysis sequence
ResultsResults
• Commutability assessment of ERM-DA470k/IFCC
1. CLSI 1. CLSI guidelineguideline EP30EP30--AA
[based on the pair-wise comparison of sets of IgA results on patient pools by the 11 MS]
DemingDeming regressionregression
95% 95% predictionprediction
intervalsintervals
2. IFCC 2. IFCC recommendationsrecommendations
ReferenceReference
materialmaterial
Results confirmed the acceptable commutability of Results confirmed the acceptable commutability of
ERMERM--DA470k/IFCC for all MS we evaluated, DA470k/IFCC for all MS we evaluated,
suggesting that this material was appropriate for suggesting that this material was appropriate for
estimating the trueness of the procedures.estimating the trueness of the procedures.
[CLSI. Document EP30-A Wayne (PA): CLSI; 2010]
-0.60
-0.40
-0.20
0.00
0.20
0.40
0.60
0 1 2 3 4 5
Mean conc both (x+y)/2 (g/L)
Bia
s ln
(AU
48
0)-
ln(A
rch
ite
ct
c4
00
0)_
ReferenceReference materialmaterial
Minimum Minimum
quality level quality level
for bias from for bias from
BVBV
[Nilsson G, et al Clin Chem 2018;64:455–64]
ResultsResults
• Trueness evaluation of IgA MS
Significantly
overestimated
Significantly
underestimated
ExcludedExcluded
fromfrom the the
followingfollowing
data data
elaborationelaboration
ResultsResults
• Verification of agreement among
IgA MS
PassingPassing--BablokBablok regression analysis and average percentage bias of measured regression analysis and average percentage bias of measured
IgAIgA values for patient pools vs. median values for patient pools vs. median IgAIgA concentrations for MS. concentrations for MS.
not fulfilling
desirable bias
(±9.1%)
not fulfilling
minimum bias
(±13.6%)
All methods median, g/L
0 1 2 3 4 5
Arc
hitect
c400
0 -
All
meth
ods m
ed
ian, %
-100
-50
0
50
100
All methods median, g/L
0 1 2 3 4 5
AU
48
0 -
All
me
tho
ds m
ed
ian
, %
-100
-50
0
50
100
All methods median, g/L
0 1 2 3 4 5
Imm
age
800
- A
ll m
eth
od
s m
edia
n,
%
-100
-50
0
50
100
All methods median, g/L
0 1 2 3 4 5
Optilit
e -
All
meth
ods m
ed
ian,
%
-100
-50
0
50
100
All methods median, g/L
0 1 2 3 4 5
Ro
ch
e C
OB
AS
c7
02
- A
ll m
eth
od
s m
ed
ian
, %
-100
-50
0
50
100
All methods median, g/L
0 1 2 3 4 5
Min
dra
y B
S4
80
- A
ll m
eth
od
s m
edia
n,
%
-100
-50
0
50
100
All methods median, g/L
0 1 2 3 4 5
Vis
ta -
All
me
thod
s m
edia
n,
%
-100
-50
0
50
100
All methods median, g/L
0 1 2 3 4 5
Ta
uru
s -
All
me
tho
ds m
ed
ian
, %
-100
-50
0
50
100
All methods median, g/L
0 1 2 3 4 5
ILA
B 6
50
- A
ll m
eth
od
s m
ed
ian,
%
-100
-50
0
50
100
All methods median, g/L
0 1 2 3 4 5
Advia
- A
ll m
eth
ods m
edia
n, %
-100
-50
0
50
100
PercentagePercentage differencedifference plotsplots
In In generalgeneral, the , the situationsituation clearlyclearly worsenedworsened whenwhen the the
analysisanalysis focusedfocused on the on the lowlow IgAIgA concentrationsconcentrations
typicaltypical of of childrenchildren
ResultsResults
• Measurement uncertainty of IgA MS on patient pools
For suitable clinical application of For suitable clinical application of IgAIgA measurements, the expanded measurement measurements, the expanded measurement
uncertainty at the patient sample level should remain within uncertainty at the patient sample level should remain within ±±2.7%, 2.7%, ±±5.4%, and 5.4%, and ±±8.1%8.1%
for the optimum, desirable, and minimum quality levels, respectifor the optimum, desirable, and minimum quality levels, respectivelyvely..
ConclusionConclusion
Although the harmonization among marketed Although the harmonization among marketed IgAIgA MS MS
is substantially good, is substantially good, the implementation of the implementation of
traceability to higher order references is still far from traceability to higher order references is still far from
perfectperfect. Most of the MS evaluated in this study . Most of the MS evaluated in this study
showed unacceptable standardization of showed unacceptable standardization of IgAIgA
measurements, especially at concentrations measurements, especially at concentrations ≤≤0.4 g/L. 0.4 g/L.
The definition of traceable reference intervals in The definition of traceable reference intervals in
pediatric population should consider this analytical pediatric population should consider this analytical
background.background.
Daily surveillance of IVD system traceability Daily surveillance of IVD system traceability
1.1. Verification of the consistency of declared performance Verification of the consistency of declared performance
during routine operations performed in accordance with the during routine operations performed in accordance with the
manufacturermanufacturer’’s instructionss instructions
(checking of system alignment by (checking of system alignment by IQC component IIQC component I))
2.2. Participation to appropriately structured EQASParticipation to appropriately structured EQAS
((““that meet metrological criteriathat meet metrological criteria””))
Braga F & Panteghini M, Braga F & Panteghini M,
Clin Chim Acta 2014;432:55Clin Chim Acta 2014;432:55
IVDIVD
TraceabilityTraceability
SurveillanceSurveillance
Daily surveillance of IVD system traceability Daily surveillance of IVD system traceability
1.1. Verification of the consistency of declared performance Verification of the consistency of declared performance
during routine operations performed in accordance with the during routine operations performed in accordance with the
manufacturermanufacturer’’s instructionss instructions
(checking of system alignment by (checking of system alignment by IQC component IIQC component I))
2.2. Participation to appropriately structured EQASParticipation to appropriately structured EQAS
((““that meet metrological criteriathat meet metrological criteria””))
Braga F & Panteghini M, Braga F & Panteghini M,
Clin Chim Acta 2014;432:55Clin Chim Acta 2014;432:55
IVDIVD
TraceabilityTraceability
SurveillanceSurveillance
Analytical quality
of measurement
Check alignment Imprecision
qualify
Reliability of the analytical system
AnalyticalAnalytical QualityQuality Control in the Traceability EraControl in the Traceability Era
PanteghiniPanteghini M, M, ClinClin ChemChem Lab Lab MedMed 2010;48:72010;48:7
Braga F & Panteghini M, Clin Chim Acta 2014;432:55Braga F & Panteghini M, Clin Chim Acta 2014;432:55
10-M
ay
25-M
ay
28-J
un
13-Jul
28-J
ul
11-A
ug
27-Aug
11-Sep
26-Sep
13-O
ct
24
+3SD
+2SD
+1SD
19
29
34
Monitoring the reliability of the measuring system through IQC: Monitoring the reliability of the measuring system through IQC:
Component I. Check alignment (Component I. Check alignment (““system traceabilitysystem traceability””))A
cce
pta
ble
Acc
ep
tab
lera
ng
era
ng
eo
f co
ntr
ol
ma
teri
al
of
con
tro
l m
ate
ria
l
This program checks whether in the course of an analytical run tThis program checks whether in the course of an analytical run the performance of a he performance of a
measuring system complies with the set goals, represented by themeasuring system complies with the set goals, represented by the acceptable acceptable
ranges of control materials. ranges of control materials.
Clinical laboratories must verify the consistency of declared peClinical laboratories must verify the consistency of declared performance during routine operations rformance during routine operations
performed in accordance with the manufacturerperformed in accordance with the manufacturer’’s instructions, by s instructions, by checking that values of control checking that values of control
materials provided by the manufacturer as component of the measumaterials provided by the manufacturer as component of the measuring system are in the ring system are in the
established control rangeestablished control range, with no clinically significant changes in the assumed traceabl, with no clinically significant changes in the assumed traceable results.e results.
PlatformPlatform
CalibratorsCalibratorsReagentsReagents
Control material(s)Control material(s)
PlatformPlatform
CalibratorsCalibratorsReagentsReagents
PlatformPlatform
CalibratorsCalibratorsReagentsReagents
Control material(s)Control material(s)
Acceptance/rejection of
the analytical run in
“real time”
Acceptance/rejection of Acceptance/rejection of
the analytical run in the analytical run in
““real timereal time””
Internal Quality Control
(Component I)
Internal Quality Control Internal Quality Control
(Component I)(Component I)
Testing alignmentTesting alignment
[[““system traceabilitysystem traceability””]]
Any “out of control” signal must be made available with
sufficient time to allow immediate corrective actions
to bring again the situation under control (virtually
“unbiased”) and before reports related to the samples
analyzed in the affected analytical run are issued.
Any Any ““out of controlout of control”” signal must be made available with signal must be made available with
sufficient time to allow immediate corrective actions sufficient time to allow immediate corrective actions
to bring again the situation under control (virtually to bring again the situation under control (virtually
““unbiasedunbiased””) and before reports related to the samples ) and before reports related to the samples
analyzed in the affected analytical run are issued.analyzed in the affected analytical run are issued.
Daily surveillance of IVD system traceability Daily surveillance of IVD system traceability
1.1. Verification of the consistency of declared performance Verification of the consistency of declared performance
during routine operations performed in accordance with the during routine operations performed in accordance with the
manufacturermanufacturer’’s instructionss instructions
(checking of system alignment by (checking of system alignment by IQC component IIQC component I))
2.2. Participation to appropriately structured EQASParticipation to appropriately structured EQAS
((““that meet metrological criteriathat meet metrological criteria””))
Braga F & Panteghini M, Clin Braga F & Panteghini M, Clin
Chim Acta 2014;432:55Chim Acta 2014;432:55
IVDIVD
TraceabilityTraceability
SurveillanceSurveillance
Analytical quality
of measurement
Check alignment Imprecision
qualify
Reliability of the analytical system
AnalyticalAnalytical QualityQuality Control in the Traceability EraControl in the Traceability Era
PanteghiniPanteghini M, M, ClinClin ChemChem Lab Lab MedMed 2010;48:72010;48:7
Braga F & Panteghini M, Clin Chim Acta 2014;432:55Braga F & Panteghini M, Clin Chim Acta 2014;432:55
Requirements for the applicability of EQAS results in the Requirements for the applicability of EQAS results in the
evaluation of the performance of participating laboratories evaluation of the performance of participating laboratories
in terms of traceability of their measurementsin terms of traceability of their measurements
FeatureFeature AimAim
EQAS materials valueEQAS materials value--assigned with assigned with
reference procedures by an accredited ref. reference procedures by an accredited ref.
laboratorylaboratory
To check traceability of commercial To check traceability of commercial
system to reference systemsystem to reference system
Proved commutability of EQAS materialsProved commutability of EQAS materials To allow transferability of To allow transferability of
participating laboratory performance participating laboratory performance
to the measurement of clinical to the measurement of clinical
samplessamples
Definition and use of the clinically Definition and use of the clinically
allowable measurement errorallowable measurement error
To verify the suitability of laboratory To verify the suitability of laboratory
measurements in clinical settingmeasurements in clinical setting
PanteghiniPanteghini M, M, ClinClin ChemChem Lab Lab MedMed 2010;48:7 2010;48:7
InfusinoInfusino I I etet al., al., ClinClin ChemChem Lab Lab MedMed 2010;48:3012010;48:301
Braga F & Braga F & PanteghiniPanteghini M. M. ClinClin ChemChem Lab Lab MedMed 2013;51:17192013;51:1719
Braga F & Panteghini M, Clin Chim Acta 2014;432:55Braga F & Panteghini M, Clin Chim Acta 2014;432:55
[Ferraro S, Braga F, Panteghini M. Clin Chem Lab Med 2016;54:523]
Analytical quality
of measurement
Check alignment Imprecision
qualify
Reliability of the analytical system
AnalyticalAnalytical QualityQuality Control in the Traceability EraControl in the Traceability Era
PanteghiniPanteghini M, M, ClinClin ChemChem Lab Lab MedMed 2010;48:72010;48:7
Braga F & Panteghini M, Clin Chim Acta 2014;432:55Braga F & Panteghini M, Clin Chim Acta 2014;432:55
Internal Quality Control
(Component II)
Internal Quality Control Internal Quality Control
(Component II)(Component II)
Testing the uncertainty Testing the uncertainty
due to the random effectsdue to the random effects
((““imprecisionimprecision””))
System stability at
medium/long term
System stability at System stability at
medium/long termmedium/long term
This program provides, through
mechanisms of retrospective
evaluation, data useful to the
knowledge of variability of
measuring system and of its use by
the individual laboratory.
This program provides, through This program provides, through
mechanisms of retrospective mechanisms of retrospective
evaluation, data useful to the evaluation, data useful to the
knowledge of variability of knowledge of variability of
measuring system and of its use by measuring system and of its use by
the individual laboratory.the individual laboratory.
System
Reagent lots
Laboratory
Monitoring the reliability of the measuring system through IQC: Monitoring the reliability of the measuring system through IQC:
Component II. Evaluate the system + individual lab imprecisionComponent II. Evaluate the system + individual lab imprecision
System imprecision
System calibration
(combined) uncertainty
Individual lab
performance
(IQC safety margin)
Measurement
uncertainty
budget
CardiacCardiac troponintroponin T high sensitiveT high sensitiveMonthlyMonthly monitoringmonitoring of of imprecisionimprecision byby IQC materialIQC material
0
1
2
3
4
5
6
7
8
Jan
Feb
Mar
Apr
May
Jun
Jul
Aug
Sep
Imp
recis
ion
, C
V%
Requirements for IQC material
(Component II)
RequirementsRequirements forfor IQC material IQC material
((ComponentComponent II)II)
RequirementRequirement CommentComment
MatrixedMatrixed material from a thirdmaterial from a third--
party independent source party independent source
should be used (e.g., freshshould be used (e.g., fresh--
frozen pool)frozen pool)
Material must be different from the Material must be different from the
system control material used for system control material used for
checking alignmentchecking alignment
Specimens closely resembling Specimens closely resembling
authentic patient samples authentic patient samples
(commutability)(commutability)
Commercial nonCommercial non--commutable controls commutable controls
may provide a different impression of may provide a different impression of
imprecision performanceimprecision performance
Specimens of concentrations Specimens of concentrations
appropriate to the clinical appropriate to the clinical
application of the application of the analyteanalyte
When clinical decision cutWhen clinical decision cut--points are points are
employed, samples around these employed, samples around these
concentrations should preferentially be concentrations should preferentially be
selected selected
F. BragaF. Braga
http://http://users.unimi.itusers.unimi.it//cirmecirme/home//home/index.phpindex.php
Research Centre for Metrological Traceability in Laboratory Medicine (CIRME),
University of Milan, Milan, Italy