TRACEABILITY and UNCERTAINTY5 Roche/Hitachi – Preciset® Lp(a) Gen. 2
Cat. No. 05852641 190
Roche Diagnostics GmbH January 2013
Routine Method
Roche/cobas c systems
Reference Method Reference Material Selected Measurement
Procedure
Assigned Value Uncertainty 1 Unit
LPA2Lipoprotein (a)
immunoturbidimetric
Tina-Quant, Gen.2
SRM2B IFCC/WHO2,3,4
Manufacture’s selected measurement procedure
Level 1: 11.1
Level 2: 21.6
Level 3: 60.0
Level 4: 158
Level 5: 211
1.81
4.33
7.47
8.49
7.93
nmol/L
1 This uncertainty (expanded uncertainty; k = 2) was calculated in accordance with the “Guide to the expression of uncertainty in measurement” (GUM:1993). For the estimation of the single standard uncertainties normal distribution of the measurement results is assumed; the level of confidence of the expanded uncertainty is about 95% with the coverage factor k=2 (analogue to the 2sd of the standard uncertainty).
2 SRM = Standard Reference Material3 IFCC = International Federationof clinical chemistry and laboratory medicine4 WHO = World Health Organisation5 Unless otherwise noted, the assigned values and uncertainties are valid for Roche Hitachi Modular P provided that the method is available on that analyzer.