Giorgia Greco, Alexander Boychenko, Remco Swart, Thermo Fisher Scientific, Germering, Germany APPLICATION NOTE Key Words 1.0 mm i.d. column, food safety, micro flow, sensitivity, TSQ Endura Triple Quadrupole MS, Vanquish Horizon UHPLC System Goal Demonstrate the performance of the Thermo Scientific Vanquish Horizon UHPLC system for the LC-MS analysis of pesticides using 1.0 mm i.d. columns for improved sensitivity and reduced solvent usage. Introduction UHPLC-MS applications are generally developed with 2.1 mm i.d. columns at relatively high flow rates, in most cases, between 0.3 and 1.0 mL/min. In the last few years, interest in increasing sensitivity and reducing solvent consumption has grown. In this regard, the use of 1.0 mm i.d. columns instead of 2.1 mm i.d. columns presents several advantages. First, microflow UHPLC applications based on 1.0 mm i.d. columns give higher sensitivity when compared to normal flow UHPLC. 1 In addition, it allows the use of less mobile phase and reduces waste, providing a cost savings and a diminished environmental impact of the toxic solvent. Robust LC-MS analysis of pesticides with 1.0 mm i.d. column using the Vanquish Horizon UHPLC system No. 1160 When scaling down a method, the theoretical increase of sensitivity can be calculated as a ratio, in which each of the respective internal diameter values are squared. Thus, scaling down a method from a 2.1 mm i.d. column to a 1.0 mm i.d. column would lead to a sensitivity increase of 2.1 2 /1.0 2 = 4.4 for concentration sensitive detectors, if the same sample amount was loaded onto the column. 2 In many aspects, a mass spectrometric detector with an electrospray ionization (ESI) source behaves as a concentration sensitive detector for LC- MS applications. However, when UHPLC methods with 1.0 mm i.d. columns are coupled to MS detection, the sensitivity increase can be substantially different from the theoretical value. 2 Several aspects, such as the specific physical chemical properties of each compound, ionization efficiency, and emitter ID, can influence the sensitivity. It is evident that this approach is highly attractive for all trace quantitative applications, such as the analysis of pesticide residues in food that must achieve demanding limits of detection (LODs). Another important aspect that has to be considered for successful applications with 1.0 mm i.d. columns is the performance of the chromatographic system at low flow rates. Pulsation-free pumps at low flow rates are required for optimal gradient performance. At the same time, the system fluidic has to be optimized to minimize the extra column volume and prevent unwanted diffusive phenomenon that can decrease the efficiency of the method and result in peak broadening.

Robust LC-MS Analysis of Pesticides with 1.0mm i.d. Column ... · Peak width – The peak width was measured at half maximum (full peak width at half maximum, FWHM). The system provided

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Robust LC-MS Analysis of Pesticides with 1.0mm i.d. Column ... · Peak width – The peak width was measured at half maximum (full peak width at half maximum, FWHM). The system provided

Giorgia Greco, Alexander Boychenko, Remco Swart, Thermo Fisher Scientific, Germering, Germany

APPLICATION NOTE

Key Words1.0 mm i.d. column, food safety, micro flow, sensitivity, TSQ Endura Triple Quadrupole MS, Vanquish Horizon UHPLC System

Goal Demonstrate the performance of the Thermo Scientific™ Vanquish™ Horizon UHPLC system for the LC-MS analysis of pesticides using 1.0 mm i.d. columns for improved sensitivity and reduced solvent usage.

IntroductionUHPLC-MS applications are generally developed with 2.1 mm i.d. columns at relatively high flow rates, in most cases, between 0.3 and 1.0 mL/min. In the last few years, interest in increasing sensitivity and reducing solvent consumption has grown. In this regard, the use of 1.0 mm i.d. columns instead of 2.1 mm i.d. columns presents several advantages. First, microflow UHPLC applications based on 1.0 mm i.d. columns give higher sensitivity when compared to normal flow UHPLC.1

In addition, it allows the use of less mobile phase and reduces waste, providing a cost savings and a diminished environmental impact of the toxic solvent.

Robust LC-MS analysis of pesticides with 1.0 mm i.d. column using the Vanquish Horizon UHPLC system

No

. 1160

When scaling down a method, the theoretical increase of sensitivity can be calculated as a ratio, in which each of the respective internal diameter values are squared. Thus, scaling down a method from a 2.1 mm i.d. column to a 1.0 mm i.d. column would lead to a sensitivity increase of 2.12/1.02 = 4.4 for concentration sensitive detectors, if the same sample amount was loaded onto the column.2 In many aspects, a mass spectrometric detector with an electrospray ionization (ESI) source behaves as a concentration sensitive detector for LC-MS applications. However, when UHPLC methods with 1.0 mm i.d. columns are coupled to MS detection, the sensitivity increase can be substantially different from the theoretical value.2 Several aspects, such as the specific physical chemical properties of each compound, ionization efficiency, and emitter ID, can influence the sensitivity. It is evident that this approach is highly attractive for all trace quantitative applications, such as the analysis of pesticide residues in food that must achieve demanding limits of detection (LODs).

Another important aspect that has to be considered for successful applications with 1.0 mm i.d. columns is the performance of the chromatographic system at low flow rates. Pulsation-free pumps at low flow rates are required for optimal gradient performance. At the same time, the system fluidic has to be optimized to minimize the extra column volume and prevent unwanted diffusive phenomenon that can decrease the efficiency of the method and result in peak broadening.

Page 2: Robust LC-MS Analysis of Pesticides with 1.0mm i.d. Column ... · Peak width – The peak width was measured at half maximum (full peak width at half maximum, FWHM). The system provided

Parameter Setting

Method A Method B Method C

Column: Thermo Scientific™ Acclaim™ PepMap™ RSLC 150 x 1.0 mm, 2.0 µm

Thermo Scientific™ Hypersil GOLD™ 100 x 2.1 mm, 1.9 µm

Thermo Scientific Hypersil GOLD 100 x 1.0 mm, 1.9 µm

Mobile Phases: A) Water/methanol (98:2, v/v, %) with 5 mM ammonium formate and formic acid 0.1% (v/v, %)

B) Water/methanol (2:98, v/v, %) with 5 mM ammonium formate and formic acid 0.1% (v/v, %)

Temperature: 60 °C, Forced Air Mode 40 °C, Still Air Mode 40 °C, Still Air Mode

Injection Volume: 0.25–1 μL 2 μL 2 μL

Gradient: 0.00–0.10 min: 20% B; 0.10–0.67 min: 20–60% B; 0.67–3.00 min: 60–100% B; 3.00–5.00 min: 100% B; 5.00–5.17 min: 100–20% B; 5.17–6.00 min: 20% B

0.0–4.0 min: 20–95% B; 4.0–4.2 min: 95% B; 4.2–4.6 min: 95-20% B; 4.6–7.0 min: 20% B

0.0–4.0 min: 20–95% B; 4.0–4.2 min: 95% B; 4.2–4.6 min: 95-20% B; 4.6–7.0 min: 20% B

Flow Rate: 0.200 mL/min 0.450 mL/min 0.100 mL/min

In this study, we test the performance of the Vanquish Horizon UHPLC system coupled to a Thermo Scientific™ TSQ Endura™ triple quadrupole mass spectrometer with 1.0 mm i.d. columns for the analysis of pesticides at ppb and sub-ppb levels.

ExperimentalSample preparationFood samples, such as strawberry and leek, were homogenized (10 g) and weighed into a Thermo Scientific™ HyperSep™ QuEChERS extraction tube (P/N 60105-216). After the addition of 10 mL of acetonitrile, the tube was shaken for 10 min and centrifuged at 5000 rpm for 5 min. Pesticide stock solutions were prepared in water and matrix extracts. Working neat solution and matrix-fortified samples were obtained by dilution in the corresponding solvent or matrix to get the final concentration of 1 or 10 μg/L (1 or 10 ppb).

Instrumentation• Thermo Scientific Vanquish Horizon UHPLC system

including:

– System Base Vanquish Horizon (P/N VH-S01-A)

– Binary Pump H (P/N VH-P10-A)

– Split Sampler HT (P/N VH-A10-A)

– Column Compartment H (P/N VH-C10-A)

• Thermo Scientific TSQ Endura triple quadrupole mass spectrometer with Thermo Scientific™ Ion Max NG API source

• Thermo Scientific™ Vanquish™ MS connection kit (P/N 6720.0405)

LC conditions for three methods are listed in Table 1 and the MS conditions are listed in Table 2. Table 3 provides information on capillaries and fittings used to optimize extra-column volume.

Table 1. LC conditions.

Page 3: Robust LC-MS Analysis of Pesticides with 1.0mm i.d. Column ... · Peak width – The peak width was measured at half maximum (full peak width at half maximum, FWHM). The system provided

Table 2. MS conditions.

Table 3. Optimized extra-column volume.

Parameter Setting

Method A Method B Method C

Ionization Conditions: HESI HESI HESI

Polarity: Positive/Negative switching Positive Positive

Acquisition Mode: SRM SRM SRM

Sheath Gas Flow Rate: 25 units 40 units 20 units

Aux Gas Flow Rate: 9 units 20 units 7 units

Spray Voltage Positive Ion: 3500 V 3200 V 3200 V

Spray Voltage Negative Ion: 2500 V - -

Ion Transfer Tube Temp.: 300 °C 350 °C 350 °C

Vaporizer Temp.: 300 °C 400 °C 200 °C

CID Gas: 2 mTorr 2 mTorr 2 mTorr

Cycle Time: 0.52 s 0.05 s 0.05 s

Product Part Number

Capillary from the Column to the Grounding Union

350 mm x 50 µm i.d.; Thermo Scientific™ nanoViper™ capillary 6041.5540

Grounding Union Zero dead-volume Thermo Scientific™ Viper™ union 6040.2304

Grounding Union-Needle Insert 150 mm x 50 µm i.d.; nanoViper capillary 6041.5124

Needle Insert Needle insert, electrospray, low flow 80000-60152

Data acquisition and processingData were acquired with Thermo Scientific™ SII for Xcalibur™ software 1.2 and processed with Thermo Scientific™ Chromeleon™ Chromatography Data System (CDS) software 7.2 SR4.

Results and discussionOptimization of the system configurationTo reduce the extra-column volume, all the fluidic path form the outlet of the column to the mass spectrometer was optimized using fused silica capillaries with 50 μm inner diameter that are smaller than the standard configuration. A needle insert for low flow that has a smaller inner diameter was used to reduce post column peak broadening.3

Analytical figures of merit for Vanquish Horizon UHPLC system with 1.0 mm i.d. columnA mixture containing 255 pesticides was analyzed with the Vanquish Horizon UHPLC system coupled to a TSQ Endura triple quadrupole mass spectrometer with a 1.0 mm i.d. column and extra-column volume optimized as described above (Figure 1), according to method A. The maximum system pressure was 1030 bar. The analysis was performed at four calibration levels, acquiring six replicates for each calibration level. In detail, a 5 ppb solution was analyzed at four different injection volumes (0.05, 0.1, 0.2, and 1 µL volume injections, corresponding respectively to 0.25, 0.5, 1, and 5 pg on column, or alternately to 0.25, 0.5, 1, and 5 ppb concentration).

Page 4: Robust LC-MS Analysis of Pesticides with 1.0mm i.d. Column ... · Peak width – The peak width was measured at half maximum (full peak width at half maximum, FWHM). The system provided

Figure 1. Extracted ion chromatograms of 1 ppb calibration standard mixture containing 255 pesticides separated on 1 mm i.d. column.

Figure 3. Limits of detection (LODs) based on the signal-to-noise ratio (S/N) showed S/N > 3 for almost all the compound at the lowest calibration level of 0.25 ppb. Results are presented as a percentage of the compounds with S/N >3 at the concentration level described in legend.

Figure 2. Calibration plots for six generic pesticides. Linearity, measured as squared correlation coefficient (R2), showed R2 > 0.95 for most of the compounds. Results are presented as percentage of the compounds with R2 in the ranges described in the legend.

0.00 1.00 2.00 3.00 4.00 5.00 -20,000

0

50,000

100,000

150,000

200,000

min

counts

The validation and robustness of the method was evaluated based on the following parameters:

Linearity – Linearity of the system response was evaluated plotting the compound areas against the nominal concentration. Calibration plots showed excellent linearity (R2 > 0.95) for most of the compounds (Figure 2).

LOD – Limits of detection (LODs) were calculated based on the signal-to-noise ratio (S/N). Almost all the compounds showed S/N > 3 at the lowest calibration level of 0.25 ppb (Figure 3). These data highlight the excellent sensitivity of the method.

98%

2%

0.25 ppb

0.50 ppb

Area precision – Area precision was evaluated at 0.50 ppb level and calculated as RSD% of six replicates. More than half of the compounds were detected at this low concentration with good area RSD below 25% (Figure 4). At 1 ppb level, almost 80% of the compounds were detected with area RSD < 25%.

65% 12%

23% RSD < 25%

25% < RSD < 30%

RSD > 30%

85%

9% R2 > 0.95

0.90 < R2 < 0.95 0.80 < R2 < 0.90

8% Figure 4. Area precision (RSD% of six replicates) calculated at 0.50 ppb level. Results are presented as a percentage of the compounds with area precision in the ranges described in the legend.

LOD (S/N > 3)

Area Precision

Linearity (R2)

Page 5: Robust LC-MS Analysis of Pesticides with 1.0mm i.d. Column ... · Peak width – The peak width was measured at half maximum (full peak width at half maximum, FWHM). The system provided

0.000

0.005

0.010

0.015

0.020

0.025

0.030

0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5

Rete

ntio

n Ti

me

SD (m

in)

Pesticide Retention Time (min)

Retention Time Precision

0.3 s

Retention time precision – The system provided excellent retention time precision over all the gradient with retention time SD < 0.01 min (= 0.6 s) (Figure 5). Retention time precision is of utmost importance in target SRM quantitation experiments, giving the confidence that the peaks of interest run after run will not shift out of the acquisition window.

Figure 5. Retention time precision evaluated as retention time SD of six replicates at 0.50 ppb level.

Peak width – The peak width was measured at half maximum (full peak width at half maximum, FWHM). The system provided very narrow peaks with FWHM below 2 seconds for the majority of the compounds (Figure 6). Narrow peaks bring the benefit of having less compound co-elution in short gradient methods, with consequently less matrix effect for a better sensitivity. In these experiments 10–15 data points were acquired per peak.

0.0

1.0

2.0

3.0

4.0

0 0.5 1 1.5 2 2.5 3 3.5 4 4.5

FWH

M (s

)

Pesticide Retention Time (min)

Peak Width

Figure 6. FWHM of pesticides at 0.50 ppb level. Values displayed are average of six replicates.

Comparison of multi-pesticide methods with 1.0 and 2.1 mm i.d. columnsMethod A, based on the use of a 1.0 mm i.d. column operated at a flow rate of 200 μL/min, was compared with a previously developed method based on a 2.1 mm i.d. column operated at flow rate of 900 μL/min (linear velocity approximately 6 mm/s). The two methods have the same linear velocity and were applied for the analysis of 98 pesticides in strawberry matrix at 1 ppb level. (For more information on the method based on 2.1 mm i.d. column, please refer to Application Note 1138.4) The data showed that the sensitivity achieved with the 1.0 mm i.d. column, although it is compound dependent, is around two times higher in comparison with the 2.1 mm i.d. column (Figure 7).

0

1

2

3

4

5

Are

a R

atio

(

1.0

mm

/2.1

mm

)

Acep

hate

Ac

etam

iprid

Am

etry

n Am

inoc

arb

Anilo

fos

Aram

ite

Atra

zine

Az

amet

hiph

os

Bend

ioca

rb

Beno

dani

l Be

nzox

imat

e Br

odifa

coum

Br

omox

ynil

pos

Bupr

ofez

in

Buta

fena

cil 4

92

Car

bend

azim

C

arbe

tam

ide

Car

bofu

ran

Chl

orm

equa

t C

hlor

otol

uron

C

loth

iani

din

Cro

toxy

phos

C

yflu

fena

mid

Des

met

hyl-p

irim

icar

b D

esm

etry

n D

icro

toph

os

Dife

naco

um

Dife

noco

nazo

le

Dim

efur

on

Dim

ethe

nam

id

dim

etho

ate

Dod

emor

ph

Epox

icon

azol

e Es

proc

arb

Ethi

ofen

carb

-sul

fone

Et

hiof

enca

rb-s

ulfo

xide

Et

hipr

ole

Etox

azol

e Fe

nam

idon

e Fe

nam

ipho

s Fe

naza

quin

Fe

nobu

carb

Fe

nthi

on-s

ulfo

xide

Fe

nuro

n Fl

ufen

oxur

on

Fluo

pyra

m

Fost

hiaz

ate

Hal

oxyf

op-m

ethy

l H

exaz

inon

e Im

azet

hapy

r Ip

rova

licar

b Is

oxab

en

Lena

cil

Mal

aoxo

n M

efen

acet

M

epiq

uat c

hlor

ide

Mep

roni

l M

etaz

achl

or

Met

cona

zole

M

ethi

ocar

b M

ethi

ocar

b-su

lfoxi

de

Met

hom

yl

Met

sulfu

ron-

met

hyl

Mon

ocro

toph

os

Nite

npyr

am

Oxa

dixy

l O

xam

yl+N

a Pi

pero

nyl b

utox

ide

Pipe

roph

os

Pirim

ipho

s-m

ethy

l Pr

ochl

oraz

Pr

ofen

opho

s Pr

omec

arb

Prom

eton

Pr

opam

ocar

b Pr

opic

onaz

ole

Pros

ulfo

carb

Py

roxs

ulam

Q

uino

xyfe

n Sc

hrad

an

Tebu

cona

zole

Te

bufe

npyr

ad

Tebu

thiu

ron

Terb

umet

on

Tetra

cona

zole

Th

iabe

ndaz

ole

Thia

met

hoxa

m

Thio

dica

rb

Tral

koxy

dim

Tr

iazo

phos

Tr

ichl

orfo

n Tr

icyc

lazo

le

Vam

idot

hion

Zo

xam

ide

Dem

eton

-S-m

ethy

l sul

fone

Figure 7. Analysis of 98 pesticides spiked in strawberry matrix at 1 ppb level. Comparison between the analysis performed with 1.0 mm i.d. column operated at flow rate of 200 μL/min and with a 2.1 mm i.d. column at 900 μL/min. For each compound, the area value was the average of six replicates with area RSD < 20%.

Page 6: Robust LC-MS Analysis of Pesticides with 1.0mm i.d. Column ... · Peak width – The peak width was measured at half maximum (full peak width at half maximum, FWHM). The system provided

System performance at 100 μL/minTo evaluate the performance of the Vanquish Horizon UHPLC system at a lower flow rate, a mixture of seven pesticides at 1 ppb level was analyzed in gradient mode with a 2.1 mm i.d. column operated at flow rate of 450 μL/min (method B). The method was then transferred to a 1.0 mm i.d. column operated at flow rate of 100 μL/min to keep constant the linear velocity at approximately 3 mm/s (method C).

Area and retention time precisionIn both cases, the Vanquish Horizon UHPLC system provides very good retention time precision and generally RSD is below 0.1% (Figure 8). The results obtained at flow rates as low as 100 μL/min were comparable to those collected at higher flow rates of 450 μL/min, highlighting the great performance of the pump even at a lower flow rate. Similarly, comparable area reproducibility was observed with the two methods (Figure 9).

0.00%

0.10%

0.20%

0.30%

0.40%

metham

idopho

s

fenuro

n

carbofu

ran

metaza

chlor

metalax

yl

fluom

eturon

bosca

lid Re

tent

ion

Tim

e RS

D%

2.1 mm 1.0 mm

0.0%

0.4%

0.8%

1.2%

1.6%

fenuro

n

carbofu

ran

metaza

chlor

metalax

yl

fluom

eturon

bosca

lid

Area

RSD

%

2.1 mm 1.0 mm

metham

idophos

Figure 8. Retention time precision (RSD%) for seven pesticides at 1 ppb level analyzed in gradient mode with 2.1 mm i.d. column operated at flow rate of 450 μL/min and 1.0 mm i.d. column operated at flow rate of 100 μL/min.

Figure 10. Peak area reproducibility of seven pesticides spiked at 10 ppb level in leek extract analyzed with 1.0 mm i.d. column operated at flow rate of 100 μL/min.

Figure 9. Area precision (RSD%) for seven pesticides at 1 ppb level analyzed in gradient mode with 2.1 mm i.d. column operated at flow rate of 450 μL/min and 1.0 mm i.d. column operated at flow rate of 100 μL/min.

Long-term robustnessThe robustness of the method at 100 μL/min was evaluated. For this purpose, the system was stressed by consecutive injections of leek extracts for two days (over 200 injections). The area response showed excellent stability for all the duration of the analysis, with area RSD below 2.5% (Figure 10). The system with the 1.0 mm i.d. column operated at flow rate of 100 μL/min is robust and capable of long, demanding analyses.

0

10,000

20,000

30,000

40,000

50,000

60,000

70,000

80,000

90,000

0 50 100 150 200 Ar

ea (c

ount

s*m

in)

Injections

methamidophos

fenuron

carbofuran

metazachlor methalaxyl

fluometuron

boscalid

Page 7: Robust LC-MS Analysis of Pesticides with 1.0mm i.d. Column ... · Peak width – The peak width was measured at half maximum (full peak width at half maximum, FWHM). The system provided

Find out more at thermofisher.com/liquidchromatography

For Research Use Only. Not for use in diagnostic procedures. © 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. AN72162-EN 1016

ConclusionThis application note highlights the excellent performance of Vanquish Horizon UHPLC system with 1.0 mm i.d. columns for robust and sensitive UHPLC-MS/MS quantifications. The results showed the following:

• The pump of the Vanquish Horizon UHPLC system delivers excellent performance using 2.1 mm, as well as 1.0 mm, i.d. columns.

• The wide flow-pressure footprint of the Vanquish Horizon UHPLC allows it to operate 1 mm i.d. columns at high linear velocity/flow rate and achieve high-throughput separations of complex samples.

• The use of 1.0 mm i.d. columns increases MS sensitivity for pesticides analysis, allowing the detection of compounds at sub-ppb level.

• UHPLC methods at low flow rates are fast and robust, and at the same time allow considerable solvent saving (around four times less solvent is required for operating 1.0 mm i.d. columns in comparison with 2.1 mm i.d. columns).

References1. Uclés Moreno, A.; Herrera López, S.; Reichert, B.; Lozano Fernández, A.; Hernando

Guil, M.D.; Fernández-Alba, A.R.: Anal Chem. 2015, 87, 1018–1025.

2. Ledigo-Quigley, C.; Smith, N.W.; Mallet, D.: J. Chromatogr. A 2002, 976, 11–18.

3. Boychenko, A.; Meding, S.; Samonig, M.; Swart, R.: Thermo Fisher Scientific Poster Note PN-64787-EN-0616S: Sensitive, fast and robust quantification of antibodies in complex matrices by capillary flow UHPLC and high-resolution accurate-mass MS, [Online] http://tools.thermofisher.com/content/sfs/posters/PN-64787-Antibodies-Capillary-Flow-UHPLC-HRAM-ASMS2016-PN64787-EN.pdf (accessed July 2016)

4. Greco, G.; Martins, C.P.B.; Bousova, K.; Swart, R.: Thermo Fisher Scientific Application Note 1138, AN-71729-EN-0815S: Fast and Ultrafast LC-MS/MS Methods for Robust and Reliable Analysis of Pesticides in Food Using the Vanquish UHPLC System, [Online] https://tools.thermofisher.com/content/sfs/brochures/AN-1138-LC-MS-Vanquish-Pesticides-Food-AN71729-EN.pdf (accessed July 2016)