"" RMIS View/Print Document Cover Sheet""

This document was retrieved from the Boeing ISEARCH System.

Accession #: Dl96069900

Document #: SD-WM-TSAP-005

TitlelDesc: TANK 241 8204 PUSH MODE CORE SAMPLING &ANALYSIS PIAN

Page 2: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352

ENGINEERING CHANGE NOTICE P.Q.1 O I L

2. ECN Category (mark one)

N? 625727 ................ . . . . . . ... . . .. . . . . _ _ _ _ _ :&

Supplemental [I Direct Revision [XI Change ECN [I TemDOraT" I1

5. Project Title/Uo./Work Order No.

Tank 241-5-204 8. Document Numbers Changed by this ECN

(includes sheet no. and rev.)

WHC-SD-WM-TSAP-005, Rev. 0

.. standbi ' I1 Supersedure il Cancel/void [I

6. Bldg./Sys./Fac. No. 7. Approval Designator

N I A Q

N / A N/A

9. Related ECN Ne(s). 10. Related Po NO.

lla. Modification Work llb. Work Package llc. Modification Work Complete NO.

N I A N I A

Cog. Engineer Signature L D a t e

[] Yes (fill out Blk.

[XI Ua (NA Blks. llb, llb)

lld. Restored to Original Condi- tion (Temp. or Standby ECN only) N I A

Cog. Engineer Signature 8. Date

3. Originator's Name, Organization, MSIN. 3a. US0 Required? 4 . Date and Telephone NO.

Leela Sasaki. Charac te r i za t i on [ I Yes [XI NO 10111195 Plans, Coordinat ion and Reports, R2-12, 373-1027

12. Description of Change This ECN i s being generated i n order t o updatejadd a d d i t i o n a l i n fo rma t ion t o t h e e x i s t i n g document

13a. Justification (mark one) Criteria Change [XI Design lmpravement [ I Environmental [I Facility Deactivation [I As- Found [I Facilitate Const [ I Const. ErrorlOmission [ I Design Error/Omission [ I 13b. Justification Details This r e v i s i o n incorporated t h e requirements o f t h e l a t e s t r e v i s i o n o f t h e Safety Screening D a t a Q u a l i t y Object ive (WHC-SD-WM-SP-004, Rev. 2 ) .

RELEASE STAMP 14. Distribvtion (include name, MSIN, and no. of copies) See at tached d i s t r i b u t i o n sheet.

BY wiic OAT' OCT 1 8 1995

A-7900-013-2 (11/94) tEF095

Page 3: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352
Page 4: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352

RELEASE AUTHORIZATION

Document Number: WHC-SD-WM-TSAP-005, REV 1

Document Title: Tank 241-8-204 Push Mode Core Sampling and Analysis P1 an

Release Date: 10/17/95

This document was reviewed following the procedures described in WHC-CM-3-4 and is:

APPROVED FOR PUBLIC RELEASE

WHC Information Release Administration Specialist:

TRADEMARK DISCLAIMER. process, or service by trade name, trademark, manufacturer, or otherwise, does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States Government or any agency thereof or its contractors or subcontractors.

This report has been reproduced from the best available copy. paper copy. this report, contact:

Reference herein to any specific commercial product,

Available in Printed in the United States of America. To obtain copies of

Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352 Telephone: (509) 372-2420; Fax: (509) 376-4989

A-6001-400.2 (09/94) YEF256

Page 5: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352
Page 6: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352
Page 7: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352

WHC-SD WM-TS AP-005 Revision 1

Tank 241=B=204 Push Core

Sampling and Analysis Plan

Prepared for the U.S. Department of Energy Office of Environmental Restoration

and Waste Management

by

Los Alamos Technical Associates 8633 Gage Boulevard

Kennewick, Washington 99336

Page 8: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352

YHC.SD.WR.TSAP.005. REV . 1

TABLE OF CONTENTS

1.0 SAMPLING AND ANALYSIS OBJECTIVES . . . . . . . . 2.0 SAMPLING EVENT REQUIREMENTS . . . . . . . . . . 3.0 LABORATORY ANALYSIS REQUIREMENTS . . . . . . . .

3.1 ANALYSIS SCHEME . . . . . . . . . . . . . 3 . 2 S P E C I F I C METHODS AND ANALYSES . . . . . . 3.3 INSUFFIC IENT SEGMENT RECOVERY . . . . . .

4.0 QUALITY ASSURANCE CONTROL . . . . . . . . . 4.1 LABORATORY OPERATIONS . . . . . . . . 4.2 SAMPLE COLLECTION . . . . . . . . . . 4.3 SAMPLE CUSTODY . . . . . . . . . . . .

. . . . . . . . . 1

. . . . . . . . . 1

. . . . . . . . . 2 . . . . . . . . . 2 . . . . . . . . . 2 . . . . . . . . . 5

. . . . . . . . . 5 . . . . . . . . . 5 . . . . . . . . . 5 . . . . . . . . . 6

5.0 EXCEPTIONS. CLARIFICATIONS. AND ASSUMPTIONS . . . . . . . . . . . . 6 5 . 1 EXCEPTIONS TO DQO REQUIREMENTS . . . . . . . . . . . . . . . . 6 5 . 2 CLARIFICATIONS AND ASSUMPTIONS . . . . . . . . . . . . . . . . 6

6.0 DELIVERABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 6.1 FORMAT I REPORTING . . . . . . . . . . . . . . . . . . . . . . 7 6 . 2 FORMAT 111 REPORTING . . . . . . . . . . . . . . . . . . . . . 8 6.3 FORMAT I V REPORTING . . . . . . . . . . . . . . . . . . . . . 8

7.0 CHANGE CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

8.0 REFERENCES . . . . . . . . . . . . . . . . . . 10

L I S T OF TABLES

1: Tank 8 - 2 0 4 C h e m i c a l . R a d i o l o g i c a l . and P h y s i c a l A n a l y t i c a l R e q u i r e m e n t s . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

ii

Page 9: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352

WHC-SD-HN-TSAP-005, REV. 1

8-204 cm DQO DSC GEA HPGE/MCA I C I C P L i B r LFL RSST SAP SARP TCP TGA T I C TOC WHC

LIST OF ABBREVIATIONS

Tank 241-B-204 cent imeters da ta q u a l i t y o b j e c t i v e d i f f e r e n t i a l scanning ca lo r ime t ry gamma energy ana lys is h igh p u r i t y germanium - m u l t i channel ana lys i s i o n chromatography i n d u c t i v e l y coupled plasma - atomic emission spectroscopy l i t h i u m bromide Lower F1 ammabi 1 i ty L i m i t r e a c t i v e system screening t o o l - ad iaba t i c ca lo r ime t ry Sampling and Analys is Plan Safety and Analys is Report f o r Packaging Tank Charac ter iza t ion Plan thermogravimetr i c ana lys i s t o t a l inorgan ic carbon t o t a l organic carbon Westinghouse Hanford Company

iii

Page 10: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352

WHC-SD-WM-TSAP-005, REV. 1

1.0 SAMPLING AND ANALYSIS OBJECTIVES

This Sampling and Analysis Plan (SAP) will identify characterization objectives pertaining to sample collection, laboratory analytical evaluation, and reporting requirements for two push-mode core samples from tank 241-8-204 (8-204). Quality Objective (Dukelow et al. 1995), This data quality objective (DQO) is described in the Tank Characterization Plan (TCP) for tank 241-8-204 (8-204). This SAP will also .identify procedures and requirements for collecting and characterizing samples from tank 8-204 by the push core sampling method.

It is written in accordance with the Tank Safety Screening D a t a

2.0 SAMPLING EVENT REQUIREMENTS

As of July 31, 1995, tank 8-204 contained 189,000 liters (50,000 gallons) of noncomplexed waste corresponding to a waste depth of 656 cm (258 inches). of sludge with 3,800 liters (1,000 gallon) of supernate and no pumpable liquid remaining (Hanlon 1995).

The tank contents were comprised of 185,000 liters (49,000 gallons)

Prior to core sampling, the dome space (below the riser) shall be sampled and analyzed for the presence of flammable gases. The sample shall be taken from within the dome space and the data reported as a percentage of the LFL. The results shall be submitted to the project coordinator within one week of the sampling event. analyzing by gas chromatography/mass spectrometer (GC/MS) or gas-specific monitoring gauges, or gas meter, the necessity for recurring sampling for flammable gas concentration and the frequency of such sampling will be determined by the Flammable Gas Program.

core sampling truck. risers 2 and 7. If a different riser is necessary to meet sampling and analysis requirements, this change must be recorded and approved by the cognizant engineer before sampling. permanent data sheet or recorded directly in the work package.

expected to be taken from tank 8-204. Segment 1 is expected to be 28 cm (11 inches) in depth. should each be 48 cm (19 inches) in depth. The sampling objective is to obtain a vertical profile of the waste; therefore, more or less segments may need to be taken depending on the accuracy of the current waste volume records. For detailed information regarding the sampling activities, refer to work package ES-95-00555. This document contains the operating procedures and the chain-of-custody records for this sampling event.

with deionized water. laboratory hot cell. All collected samples and the field blank shall be shipped to the laboratory following the Load/Transport Sample C a s k ( s ) procedure (TO-080-090). Core samples shall be transported to the laboratory

If the limits are above 25% of the LFL when

above 10% of the LFL when analyzing with a combustible

Tank 8-204 is currently scheduled to be core sampled using a push-mode At least two samples are expected to be taken from

The risers used may be recorded on a

Fourteen segments per core are Segments 2 through 14

One field blank for tank 8-204 shall be obtained by filling a sampler This field blank is to accompany the samples to the

1

Page 11: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352

WHC-SD-WM-TSAP-005, REV. 1

w i t h i n t h r e e calendar days from t h e t ime each segment i s removed from t h e tank.

Occasional ly, l i t h i u m bromide (L iB r ) s o l u t i o n may be used t o a i d i n t h e c o l l e c t i o n o f t h e core samples. I f L i B r s o l u t i o n i s used, Sampling Operations must s t a t e t h i s i n the chain-of-custody form t h a t accompanies t h e sample t o the l abo ra to ry , and must prov ide a L i B r s o l u t i o n b lank t o t h e l abo ra to ry . The L i B r s o l u t i o n b lank s h a l l c o n s i s t o f a conta iner f i l l e d w i t h L i B r s o l u t i o n from t h e same batch o f L i B r s o l u t i o n used du r ing the sampling. This b lank s h a l l be analyzed f o r l i t h i u m and bromide i n order t o determine t h e concentrat ion o f t he t r a c e r a t t he t ime t h e core was taken. Only one L i B r s o l u t i o n b lank per tank i s requ i red i n a d d i t i o n t o the f i e l d / t r i p blank.

3.0 LABORATORY ANALYSIS REQUIREMENTS

3.1 ANALYSIS SCHEME

I n order t o comply w i t h the s a f e t y screening DQO, t he f o l l o w i n g steps s h a l l be performed on each sample:

c Extrude segments c

c c

Analyze d ra inab le 1 i q u i d s and half-segment subsamples from each segment from each core as shown i n Table 1. F i l t e r l i q u i d s p r i o r t o analys is . Archive a t l e a s t 10 mL o f each subsegment o r d ra inab le l i q u i d from each segment.

As a precaut ionary measure, t h e Safety and Analys is Report f o r Packaging (SARP) i n t h e LoadlTransport Sample C a s k ( s ) procedure (TO-080-090) has been reviewed f o r any s a f e t y issues i nvo l ved w i t h t r a n s p o r t a t i o n o f tank 8-204 core samples. vented every 47 days t o re lease any accumulated gas.

inc luded a t t he d i s c r e t i o n o f t he p r o j e c t coord inator when the l a b o r a t o r y i s n o t operat ing a t maximum capaci ty .

du r ing the sample breakdown and analyses s h a l l be documented i n w r i t i n g w i t h j u s t i f i c a t i o n . data r e p o r t .

For core samples from t a n k E-204, t h e sh ipp ing conta iner be

Oppor tun i s t i c analyses as de f i ned i n K r i s t o f z s k i (1995a) w i l l be

Any decis ions, observations, o r dev ia t i ons made t o t h i s work p lan o r

The r e p o r t i n g formats f o r analyses are contained i n Table 1. These dec is ions and observat ions s h a l l a l so be repor ted i n the

3.2 SPECIFIC METHODS AND ANALYSES

The analyses i n Table 1 t o be performed on tank 8-204 core samples are based on the s a f e t y screening DQO referenced i n Sect ion 1.0. The l a b o r a t o r y procedure numbers, which s h a l l be used f o r t he analyses, are inc luded i n the t a b l e .

2

Page 12: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352
Page 13: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352

WHC-SD-WM-TSAP-005, REV. 1

Table 1: 8-204 Chemical, Radio log ica l and Physical A n a l y t i c a l Requirements

' S-LEV LiQ-liquid taken fmm the segment level, FB- feld blank

' PR-precision, ACaccuracy, ea-each. smpl-sample, DUPduplicate. SPWMSD-spike and matrix spike duplicate, AB-analytical batch. PBpreparation blank, NIA-not applicable, mtm-matrix 'Except as noted, units for notification limits and expected range are those listed in the "units" column.

d-direct, 1-fusion dissolution. bacid dissolution, w-water dissolution

Dry weight basis. Direct liquid samples may be diluted in acid or water to adjust to pmper sample size andlor pH.

'This analyses required if TOC by persulfate is less than 3 wt%. 'Tracer or canief may be used in piace of a spike and reauks corrected for recovery.

'ORSST performed only if DSC exceeds notifcation iiml. The RSST method. yet to be proceduralized. may be found in WHCSD-WM-TP-104. " Corrected from weight basis to volumetric basis assuming a liquid density of 1 .O glmL. "Perform this analysis if the DSC iimit Is exceeded.

Either serial dilutions or mat* spikes will be p r f w m d .

4

Page 14: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352

WHC-SD-Wn-TSAP-005, REV. I

3.3 INSUFFICIENT SEGMENT RECOVERY

If the amount of material recovered from samples taken from a tank is insufficient to perform the analyses requested in the respective SAP and permit a minimum 10 mL archive per sample, the laboratory shall notify the Tank Cognizant Engineer within one working day. prioritization of the analyses maybe provided to the laboratories. Any analyses prescribed by the SAP, but not performed, shall be identified in the appropriate data report with justification for non-performance.

At that time, a

4.0 QUALITY ASSURANCE CONTROL

4.1 LABORATORY OPERATIONS

Laboratories performing analysis in support of this Tank Sampling and Analysis Plan shall have approved and implemented quality assurance project plans. These QA plans shall meet the Hanford Analytical Services Quality Assurance Plan (DOE 1995) minimum requirements as the baseline for laboratory quality systems. Quality requirements for conducting Characterization Project sampling and analysis are described in TWRS Characterization Program Quality Assurance Program Plan (Whelan 1994), Fiscal Year 1995 Tank Waste Remediation System Tank Waste Analysis Plan (Haller 1994), and in this SAP. Characterization Project sampling and analysis shall be conducted in conformance with these quality assurance requirements.

Sample quality control (duplicates, spikes, standards) are identified in Table 1. If no criteria are provided, the performing laboratory shall perform to its quality assurance plan.

4.2 SAMPLE COLLECTION

Before sampling can be performed on a tank, available risers must be identified for use in the sampling event. The selected risers must be inspected and prepared to confirm their ability to be used in sampling. Safety hazards must be identified and special precautions must be made if needed. If deemed necessary by the sampling and tank cognizant engineers, video surveillance should be performed to identify any potential problems that may occur during the sampling event.

Samples are to be taken from a tank and shipped t o the performing laboratory by Sampling Operations in accordance with the respective work package. That work package shall also initiate the chain-of-custody for the samples. Approved procedure TO-080-090 ("Load/Transport Sample Cask(s) ") is to be used during the sampling event. number before being shipped to the performing laboratory. The sampling team is responsible for documenting any problems and procedural changes affecting the validity of the sample in a field notebook. enter this information in the comment section of the chain-of-custody form for addition to the data reports.

Sampling Operations should transport each sample collected to the performing laboratory within one working day of removing the sample from the

5

Samples shall be identified by a unique

Sampling Operations shall

Page 15: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352

WHC-SD-WM-TSAP-005, REV. 1

tank, bu t must t ranspor t each segment o r sample w i t h i n th ree calendar days. Sampling Operations i s responsib le f o r v e r b a l l y n o t i f y i n g t h e WHC 2 2 2 4 Laboratory (373-2435) a t l e a s t 24 hours i n advance o f an expected shipment.

Supervisor:

Date o f Sampling:

Shipment No.:

4.3 SAMPLE CUSTODY

Sample No.:

Time o f Sampling:

S e r i a l No.:

The chain-of-custody form i s i n i t i a t e d by the sampling team as descr ibed i n t h e work package. Tank Sample Seal.

Samples are shipped i n a cask and sealed w i t h a Waste

L WASTE TANK SAMPLE SEAL

The sealed and labe led samples are shipped t o the l abo ra to ry along w i t h the chain-of-custody form. The r e c e i p t and con t ro l o f samples i n t h e WHC 222- S Laboratory are descr ibed i n l abo ra to ry procedure LO-090-101.

5.0 EXCEPTIONS, CLARIFICATIONS, AND ASSUMPTIONS

5.1 EXCEPTIONS TO DQO REQUIREMENTS

ana lys is i s requested as a secondary ana ly te on a quarter-segment l e v e l . However, i t i s no t poss ib le t o change an already prepared homogenized half-segment t o a quarter-segment. Safe ty Screening Program, i t has been decided t h a t cyanide would be r u n on half-segments i n c o r r e l a t i o n w i t h a l l o ther analyses requested by the sa fe ty screening DQO.

I n the sa fe ty screening DQO (Dukelow e t a l . 1995), t o t a l cyanide

Therefore, a f t e r d iscuss ions w i t h the

5 . 2 CLARIFICATIONS AND ASSUMPTIONS

A number o f c l a r i f i c a t i o n s and assumptions r e l a t i n g t o t h e n o t i f i c a t i o n l i m i t s o r dec i s ion thresholds i d e n t i f i e d i n the app l i cab le DQO e f f o r t s need t o be made w i t h respect t o the analyses i n Table 1. discussed below:

b

Each o f these issues i s

Any exothermic r e a c t i o n ( i n c a l / g o r J /g) determined by d i f f e r e n t i a l scanning ca lo r ime t ry (DSC) must be repor ted on a d ry .we igh t bas is as shown i n equat ion (1) us ing t h e weight percent water determined from thermograv imetr ic ana lys is (TGA).

6

Page 16: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352

WHC-SD-HI-TSAP-005, REV. 1

NOTE: A l a r g e e r r o r i n the DSC value may r e s u l t when conver t ing samples conta in ing g rea te r than 90 percent water t o a d r y weight bas is . However, t h i s conversion i s s t i l l requ i red .

b The sa fe ty screening DQO (Dukelow, e t a l . 1995) requ i res t h a t add i t i ona l analyses be performed i f t o t a l a lpha a c t i v i t y i s g rea te r than 1 g/L. To ta l a lpha i s measured i n pC i /g r a t h e r than g/L. n o t i f i c a t i o n l i m i t f o r t o t a l a lpha i n t o a number more r e a d i l y usable by the labora tory , it was assumed t h a t a l l alpha decay o r i g i n a t e s from Pu-239. equat ion ( 2 ) :

To conver t t he

The n o t i f i c a t i o n l i m i t may then be ca l cu la ted as shown i n

NOTE: Samples measured f o r t o t a l alpha s h a l l a l so be measured f o r dens i ty . The n o t i f i c a t i o n l i m i t s h a l l be 41 pCi /g unless the measured dens i t y exceeds 1.5 g/mL when the n o t i f i c a t i o n l i m i t w i l l be ad justed according t o equat ion 2.

6.0 DELIVERABLES

A l l analyses w i l l be repor ted as Format I , 111, o r I V as i nd i ca ted i n Table 1. Add i t i ona l in fo rmat ion regard ing r e p o r t i n g formats i s g iven i n (Schre iber 1994a).

6 .1 FORHAT I REPORTING

Table 1 conta ins the n o t i f i c a t i o n l i m i t s f o r each analy te. Any r e s u l t s exceeding t h e i r n o t i f i c a t i o n l i m i t s s h a l l be repor ted v i a telephone by the 222-S Laboratory S h i f t Manager t o t h e East Tank Farm Operations S h i f t Manager as soon as the da ta are obtained and reviewed ( K r i s t o f z s k i 1995b). verbal n o t i f i c a t i o n must be fo l lowed w i t h i n one working day by w r i t t e n communication t o the Safety Screening Representative, Process Contro l , and the Charac ter iza t ion P ro jec t O f f i c e documenting t h e observat ions (Schrei ber 1994b). between t h e per forming labo ra to ry and the contacts above by a r e v i s i o n t o t h i s document, o r by a memorandum o f understanding.

Th is

Add i t i ona l analyses f o r v e r i f i c a t i o n purposes may be cont rac ted

Page 17: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352
Page 18: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352
Page 19: RMIS View/Print Document Cover Sheet/67531/metadc678591/m2/1/high_re… · Westinghouse Hanford Company - Document Control Services P.O. Box 1970, Mailstop H6-08, Richland, WA 99352

WHC-SD-WR-TSAP-005, REV. I

8.0 REFERENCES

Bobrowski, S. F., S. J. Harr is , and J. D. Schwab, 1994, "Standard E lec t ron i c Format S p e c i f i c a t i o n f o r Tank Charac ter iza t ion Database Loader, Version 2.0" , P a c i f i c Northwest Laboratory, Richland, Washington (Transmit ted by PNL L e t t e r TCD-94-002, June 16, 1994, S. F. Bobrowski, PNL, t o C. S. Hal 1 er , WHC) .

DOE, 1994, Hanford A n a l y t i c a l Serv ices Q u a l i t y Assurance Plan, DOE/RL-94-55, Rev. 0, U.S. Department o f Energy, Richland F i e l d O f f i ce , Richland, Washington.

Screening Data Quality Object ive, WHC-SD-WM-SP-004, Rev. 2, Westinghouse Hanford Company, R ich l and, Washington.

Hanlon, B. M., 1995, Waste Tank Summary f o r Month Ending July 31, 1995, WHC-EP-0182-88, Westinghouse Hanford Company, Richland, Washington.

Dukelow, G. T., J. W. Hunt, H. Babad, and J. E. Meacham, 1995, Tank Safe ty

Homi, C . S., 1995, F i s c a l Year 1996 Tank Waste Ana lys is Plan, WHC-SD-WM-PLN-101, Rev. 0, Westinghouse Hanford Company, Richland, Washington.

K r i s t o f z s k i , J. G . , 1995a, L e t t e r t o J. H. Baldwin, e t a l . , "D i rec t i ons f o r 'Oppor tun i s t i c Analyses, "' 75310-95-103, dated September 13, 1995.

K r i s t o f z s k i , J. G., 1995b, L e t t e r t o G. B. G r i f f i n , "Reestablishment o f Immediate N o t i f i c a t i o n System", 75310-95-102, dated September 8, 1995.

Schreiber, R. D., 1994a, L e t t e r t o C. S. Ha l l e r , "Revised I n t e r i m Tank Charac ter iza t ion Plan Guidance:, 7E720-94-121, dated May 13, 1994.

Schreiber, R. D., 1994b, L e t t e r t o J. G. K r i s t o f z s k i , "Po in t o f Con tac t /D is t r i bu t i on L i s t " , 7E720-94-141, dated October 11, 1994.

Whelan, T. E., 1994, TWRS Charac ter iza t ion Program Q u a l i t y Assurance Program Plan, WHC-SD-WM-QAPP-025, Rev. 0, Westinghouse Hanford Company, R ich l and, Washington.

10