37
PacketMicro, Inc. 1900 Wyatt Drive, Suite 9, Santa Clara, CA 95054 www.packetmicro.com RF Probing With Anritsu MS46522A VNA

RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

PacketMicro, Inc. 1900 Wyatt Drive, Suite 9, Santa Clara, CA 95054 www.packetmicro.com

RF Probing With

Anritsu MS46522A VNA

Page 2: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

2

Outline

• Why RF Probing Page 3 – 4

• RF Probing Tips Page 5

• RProbe Summary Page 6 - 9

• Tools: Probes, Positioners, and Microscopes Page 10 - 12

• Probe Planarization with Mylar Page 13 - 17

• Probe Planarization with TCS60 Page 18 - 19

• SOLT Calibration Page 20 - 37

www.packetmicro.com

Page 3: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

3

Why RF Probing?

• Necessity: Constant shrinking size of circuit

components makes soldering semi-rigid RF cables to test gigahertz circuits impractical.

• Accuracy: RF probes and calibration substrates

allow engineers to perform probe-tip calibration for accurate, repetitive measurements.

• Productivity: Any engineer can do RF

measurements in minutes without being a master in soldering semi-rigid cable

www.packetmicro.com

Page 4: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

4

Measurements of RF Low-Pass Filter

RF probe is almost as good as SMA connector!

www.packetmicro.com

SMA-SMA

RProbe - SMA

Soldered Cable - SMA

Page 5: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

5

RF Probing Tips

• Good contact of both probe tips with the DUT is essential to accurate calibration and measurements.

• Using a good microscope is very important.

• Affixing a piece of Mylar tape to the PCB can guide the probe planarization process.

www.packetmicro.com

Page 6: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

6

RProbes

www.packetmicro.com

Features • High Bandwidth: DC to 12 GHz • Low Insertion Loss: < 3 dB @ 12 GHz • Ruggedness: Beryllium Copper tips • Probe-tip Calibration: TCS 50 Cal Substrate • High Repeatability: No moving part RProbe Part No. • RP-GR-121502 – 12 GHz, 0.2 mm/08 mil pitch • RP-GR-121504 – 12 GHz, 0.4 mm/16 mil pitch • RP-GR-121505 – 12 GHz, 0.5 mm/20 mil pitch • RP-GR-121508 – 12 GHz, 0.8 mm/32 mil pitch • RP-GR-111510 – 12 GHz, 1.0 mm/40 mil pitch • RP-GR-111512 – 12 GHz, 1.2 mm/48 mil pitch

0.5 mm 0.8 mm 1.0 mm

Un-calibrated S21 for 3 probe pitches Un-calibrated S11 for 3 probe pitches

Page 7: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Thru Measurement with Probe-Tip Calibration

Calibrated at SMA connectors

Calibrated at probe tips

www.packetmicro.com 7

Thru Measurement

Calibrated at probe tips

Calibrated at SMA connectors

(MHz)

(MHz) (MHz)

(MHz)

Page 8: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

S11 Measurements of a 50Ω Standard

Probe-tip calibration allows accurate, repetitive measurements.

Calibrated at SMA connectors

Calibrated at probe tips

www.packetmicro.com 8

Page 9: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Two-Port Probe-Tip Calibration Setup

Precision TP300 Positioners with RProbes

www.packetmicro.com 9

Page 10: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

10

Tools - Accessories

Optical Microscope (~ 90 x magnification)

TCS60 Calibration Substrate

Mylar Tape

USB Digital Microscope (~ 90 x magnification) (Make sure to use a long working range (5 cm @ 90x) microscope!)

• Using a good microscope is essential.

• You might damage the probe if you cannot see the probe tips well.

www.packetmicro.com

Page 11: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

11

TP300 Precision Positioners (XYZθ)

www.packetmicro.com

Microprobe on TP300

RProbe on TP300

• Accurate (100 TPI) • Ergonomic • Perfect for Microprobe

and RProbe

Page 12: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

12

TP60 Flex Positioners (XYZθ)

(Planarization knob)

www.packetmicro.com

• Accurate (100 TPI) • Coarse and fine adjustment

without tool

Page 13: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

13

Probe Planarization with a Mylar Tape

Step 1:

Place the a piece of Mylar tape close to the test pad, so that they have the same reference surface.

Mylar Tape

www.packetmicro.com

RF Probe (tips not touching)

Test Pad

Page 14: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

14

Signal tip touches down first

Step 2:

Land the probe tips on the tape and observe the probe-tip footprint. Above image shows that signal tip touches the surface first.

Signal GND

Mylar indentation caused by signal tip

www.packetmicro.com

Page 15: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

15

GND tip touches down first

Step 3:

Adjust the planarization knob on the positioner to lower the GND tip. Above image shows that GND tip touches the surface first.

www.packetmicro.com

Signal GND

Mylar indentation caused by GND tip

Page 16: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

16

Both tips touch down simultaneously

Step 4:

Adjust the planarization knob on the positioner to land both probe tips. Above image shows the two probe tips touch the surface evenly.

www.packetmicro.com

Signal GND

Mylar indentation caused by both tips

Page 17: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

TCS60 on Plastic Carrier

• Use the Mylar tape on the back of the plastic cap for probe planarization

• Place TCS60 on top the plastic cap and perform SOLT calibration

• Remove the TCS60 and start probing

www.packetmicro.com 17

Use Mylar for Probe Planarization

Use TCS60 for SOLT Calibration

Page 18: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Use Probe Marks on TCS60

Left touching Right touching

Only one tip touches; need adjustment!

18 www.packetmicro.com

Page 19: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Use VNA to Verify Probe Contacts

Both tips touching

• Both tips leave light probe marks

• 180° Phase changed in S11 or S22

Short

www.packetmicro.com 19

Page 20: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

2-Port Probe-Tip Calibration

www.packetmicro.com 20

• Reflection calibration (Short, Open, Load calibration for two ports)

• Transmission calibration (Thru calibration)

Thru

G S G S

Open Short

Page 21: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Probe-tip Calibration with MS46522A

1. Load probe-dependent TCS60 cal kit file (e.g. TCS60_08mm.ccf) into VNA

2. Go to “Edit Cal Params” page 3. Select TCS60 cal kit for Test Port 1 and Test Port 2 4. Enter Thru Coefficient 5. Perform SOLT calibration

www.packetmicro.com 21

Page 22: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Load Calkit File

1) Calibration -> Calkit Options -> Load Kit/Charac 2) Browse -> Load TCS60_XXmm.ccf

www.packetmicro.com 22

Page 23: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Edit Cal Params

1) Calibration -> Manual Cal -> 2-Port Cal -> Modify Cal Setup -> Edit Cal Params.

www.packetmicro.com 23

Page 24: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Select Cal Kit for Test Ports

1) Edit Cal Params -> Select TCS60_XXmm (M)

www.packetmicro.com 24

Page 25: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Verify CalKit

1) Verify the TCS60_XXmm coefficients, if necessary

www.packetmicro.com 25

Page 26: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Enter Thru Coefficient

1) Check the TCS60_XXmm (M) is selected on Test Port 1 and Test Port 2

2) Use Thru calculator to enter length

www.packetmicro.com 26

Thru calculator

Page 27: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Enter Thru Coefficient – cont.

1) Enter length (5.8 ps) and dielectric constant (9.8)

www.packetmicro.com 27

Page 28: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Verify CalKit Before Performing SOLT

1) Check the TCS60_XXmm (M) is selected on Test Port 1 and Test Port 2

2) Check Thru is entered

www.packetmicro.com 28

Page 29: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Set S11 to Display Phase Before Cal

1) Response -> S11 2) Display -> Trace Format -> Phase 3) Marker -> Marker 1 (ON)

www.packetmicro.com 29

Page 30: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Probe-tip Calibration - Short

• Calibration -> Manual Cal -> 2-Port Cal -> Port 1 Reflective Devices

• S11 Phase Changes 180° when both tips touching

www.packetmicro.com 30

Short Pattern

Probe

Page 31: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Probe-tip Calibration - Open

• Land the probe tips on “Open” pattern • S11 Phase should not change • Click Open

www.packetmicro.com 31

Open Pattern

Page 32: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Probe-tip Calibration - Load

S11 phase for Load

www.packetmicro.com 32

• Land the probe tips on “Load” pattern • Click Load

Load Pattern

Page 33: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Probe-tip Calibration – Port 2

www.packetmicro.com 33

• Back-> Port 2 Reflective Devices • Click “Open” and VNA automatically displays S22 • Perform SOL for Port 2

Page 34: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

Probe-tip Calibration – Thru

www.packetmicro.com 34

• Back-> Thru/Recip. • Make sure both probes make good contact

Both probes make good contact

Page 35: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

2-Port Calibration Completed

www.packetmicro.com 35

• Back-> Done • Cal Status becomes “ON”

S21 and S12 should be close to 0 dB

Page 36: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

2-Port Measurement of a RF Filter

www.packetmicro.com 36

TDK 2.45 GHz Low Pass Filter (P/N: DEA102500LT-6307A1, Size 0402)

Page 37: RF Probing - PacketMicro Inc.€¦ · RF Probing Tips •Good contact of both probe tips with the DUT is essential to accurate calibration and measurements. •Using a good microscope

37

Thank You

Contact: [email protected] Office: 408-675-3900

RF Probing with Ease!

Rugged RF probes up to 12 GHz

Probe-tip calibration with TCS60 substrate

Precision TP300 and TP60 Positioners (XYZθ) Don’t forget to use Mylar tape for planarization

www.packetmicro.com