Upload
others
View
3
Download
0
Embed Size (px)
Citation preview
PacketMicro, Inc. 1900 Wyatt Drive, Suite 9, Santa Clara, CA 95054 www.packetmicro.com
RF Probing With
Anritsu MS46522A VNA
2
Outline
• Why RF Probing Page 3 – 4
• RF Probing Tips Page 5
• RProbe Summary Page 6 - 9
• Tools: Probes, Positioners, and Microscopes Page 10 - 12
• Probe Planarization with Mylar Page 13 - 17
• Probe Planarization with TCS60 Page 18 - 19
• SOLT Calibration Page 20 - 37
www.packetmicro.com
3
Why RF Probing?
• Necessity: Constant shrinking size of circuit
components makes soldering semi-rigid RF cables to test gigahertz circuits impractical.
• Accuracy: RF probes and calibration substrates
allow engineers to perform probe-tip calibration for accurate, repetitive measurements.
• Productivity: Any engineer can do RF
measurements in minutes without being a master in soldering semi-rigid cable
www.packetmicro.com
4
Measurements of RF Low-Pass Filter
RF probe is almost as good as SMA connector!
www.packetmicro.com
SMA-SMA
RProbe - SMA
Soldered Cable - SMA
5
RF Probing Tips
• Good contact of both probe tips with the DUT is essential to accurate calibration and measurements.
• Using a good microscope is very important.
• Affixing a piece of Mylar tape to the PCB can guide the probe planarization process.
www.packetmicro.com
6
RProbes
www.packetmicro.com
Features • High Bandwidth: DC to 12 GHz • Low Insertion Loss: < 3 dB @ 12 GHz • Ruggedness: Beryllium Copper tips • Probe-tip Calibration: TCS 50 Cal Substrate • High Repeatability: No moving part RProbe Part No. • RP-GR-121502 – 12 GHz, 0.2 mm/08 mil pitch • RP-GR-121504 – 12 GHz, 0.4 mm/16 mil pitch • RP-GR-121505 – 12 GHz, 0.5 mm/20 mil pitch • RP-GR-121508 – 12 GHz, 0.8 mm/32 mil pitch • RP-GR-111510 – 12 GHz, 1.0 mm/40 mil pitch • RP-GR-111512 – 12 GHz, 1.2 mm/48 mil pitch
0.5 mm 0.8 mm 1.0 mm
Un-calibrated S21 for 3 probe pitches Un-calibrated S11 for 3 probe pitches
Thru Measurement with Probe-Tip Calibration
Calibrated at SMA connectors
Calibrated at probe tips
www.packetmicro.com 7
Thru Measurement
Calibrated at probe tips
Calibrated at SMA connectors
(MHz)
(MHz) (MHz)
(MHz)
S11 Measurements of a 50Ω Standard
Probe-tip calibration allows accurate, repetitive measurements.
Calibrated at SMA connectors
Calibrated at probe tips
www.packetmicro.com 8
Two-Port Probe-Tip Calibration Setup
Precision TP300 Positioners with RProbes
www.packetmicro.com 9
10
Tools - Accessories
Optical Microscope (~ 90 x magnification)
TCS60 Calibration Substrate
Mylar Tape
USB Digital Microscope (~ 90 x magnification) (Make sure to use a long working range (5 cm @ 90x) microscope!)
• Using a good microscope is essential.
• You might damage the probe if you cannot see the probe tips well.
www.packetmicro.com
11
TP300 Precision Positioners (XYZθ)
www.packetmicro.com
Microprobe on TP300
RProbe on TP300
• Accurate (100 TPI) • Ergonomic • Perfect for Microprobe
and RProbe
12
TP60 Flex Positioners (XYZθ)
(Planarization knob)
www.packetmicro.com
• Accurate (100 TPI) • Coarse and fine adjustment
without tool
13
Probe Planarization with a Mylar Tape
Step 1:
Place the a piece of Mylar tape close to the test pad, so that they have the same reference surface.
Mylar Tape
www.packetmicro.com
RF Probe (tips not touching)
Test Pad
14
Signal tip touches down first
Step 2:
Land the probe tips on the tape and observe the probe-tip footprint. Above image shows that signal tip touches the surface first.
Signal GND
Mylar indentation caused by signal tip
www.packetmicro.com
15
GND tip touches down first
Step 3:
Adjust the planarization knob on the positioner to lower the GND tip. Above image shows that GND tip touches the surface first.
www.packetmicro.com
Signal GND
Mylar indentation caused by GND tip
16
Both tips touch down simultaneously
Step 4:
Adjust the planarization knob on the positioner to land both probe tips. Above image shows the two probe tips touch the surface evenly.
www.packetmicro.com
Signal GND
Mylar indentation caused by both tips
TCS60 on Plastic Carrier
• Use the Mylar tape on the back of the plastic cap for probe planarization
• Place TCS60 on top the plastic cap and perform SOLT calibration
• Remove the TCS60 and start probing
www.packetmicro.com 17
Use Mylar for Probe Planarization
Use TCS60 for SOLT Calibration
Use Probe Marks on TCS60
Left touching Right touching
Only one tip touches; need adjustment!
18 www.packetmicro.com
Use VNA to Verify Probe Contacts
Both tips touching
• Both tips leave light probe marks
• 180° Phase changed in S11 or S22
Short
www.packetmicro.com 19
2-Port Probe-Tip Calibration
www.packetmicro.com 20
• Reflection calibration (Short, Open, Load calibration for two ports)
• Transmission calibration (Thru calibration)
Thru
G S G S
Open Short
Probe-tip Calibration with MS46522A
1. Load probe-dependent TCS60 cal kit file (e.g. TCS60_08mm.ccf) into VNA
2. Go to “Edit Cal Params” page 3. Select TCS60 cal kit for Test Port 1 and Test Port 2 4. Enter Thru Coefficient 5. Perform SOLT calibration
www.packetmicro.com 21
Load Calkit File
1) Calibration -> Calkit Options -> Load Kit/Charac 2) Browse -> Load TCS60_XXmm.ccf
www.packetmicro.com 22
Edit Cal Params
1) Calibration -> Manual Cal -> 2-Port Cal -> Modify Cal Setup -> Edit Cal Params.
www.packetmicro.com 23
Select Cal Kit for Test Ports
1) Edit Cal Params -> Select TCS60_XXmm (M)
www.packetmicro.com 24
Verify CalKit
1) Verify the TCS60_XXmm coefficients, if necessary
www.packetmicro.com 25
Enter Thru Coefficient
1) Check the TCS60_XXmm (M) is selected on Test Port 1 and Test Port 2
2) Use Thru calculator to enter length
www.packetmicro.com 26
Thru calculator
Enter Thru Coefficient – cont.
1) Enter length (5.8 ps) and dielectric constant (9.8)
www.packetmicro.com 27
Verify CalKit Before Performing SOLT
1) Check the TCS60_XXmm (M) is selected on Test Port 1 and Test Port 2
2) Check Thru is entered
www.packetmicro.com 28
Set S11 to Display Phase Before Cal
1) Response -> S11 2) Display -> Trace Format -> Phase 3) Marker -> Marker 1 (ON)
www.packetmicro.com 29
Probe-tip Calibration - Short
• Calibration -> Manual Cal -> 2-Port Cal -> Port 1 Reflective Devices
• S11 Phase Changes 180° when both tips touching
www.packetmicro.com 30
Short Pattern
Probe
Probe-tip Calibration - Open
• Land the probe tips on “Open” pattern • S11 Phase should not change • Click Open
www.packetmicro.com 31
Open Pattern
Probe-tip Calibration - Load
S11 phase for Load
www.packetmicro.com 32
• Land the probe tips on “Load” pattern • Click Load
Load Pattern
Probe-tip Calibration – Port 2
www.packetmicro.com 33
• Back-> Port 2 Reflective Devices • Click “Open” and VNA automatically displays S22 • Perform SOL for Port 2
Probe-tip Calibration – Thru
www.packetmicro.com 34
• Back-> Thru/Recip. • Make sure both probes make good contact
Both probes make good contact
2-Port Calibration Completed
www.packetmicro.com 35
• Back-> Done • Cal Status becomes “ON”
S21 and S12 should be close to 0 dB
2-Port Measurement of a RF Filter
www.packetmicro.com 36
TDK 2.45 GHz Low Pass Filter (P/N: DEA102500LT-6307A1, Size 0402)
37
Thank You
Contact: [email protected] Office: 408-675-3900
RF Probing with Ease!
Rugged RF probes up to 12 GHz
Probe-tip calibration with TCS60 substrate
Precision TP300 and TP60 Positioners (XYZθ) Don’t forget to use Mylar tape for planarization
www.packetmicro.com