Rev. 3.09 9-1

The Complete EBAT

Colorado Department of Public Health and EnvironmentLaboratory Services Division

EBAT Program

(Evidential Breath Alcohol Test)

Page 2: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-2

1. Appendix 2A Review2. Functions Key3. Stand By Mode4. Warm Up Period5. Diagnostic Tests6. Ready Mode7. Test Sequence

(ACABAIA2ABACA)8. EBAT Print Out

CONTENT

Page 3: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-3

APPENDIX 2AAPPENDIX 2A

Colorado Department of Public Health and Environment Laboratory Services Division Breath Alcohol Testing Program

Approved checklist for Evidential Breath Alcohol Test(s) after upgrade to Intoxilyzer 5000EN software revision 1358.XX, in compliance with the Colorado Board of Heath Rules and Regulations concerning testing for alcohol and other drugs, 5-CCR1005-2, as amended.

SUBJECT: _____________________________________

DATE:_____________________________

Certified operator or instructor conducting the EBAT must initial inside the parentheses to the left of each step and sign in the space provided at the bottom.

( ) 1. Turn power switch on or observe the power switch has been activated. If the EBAT device is in the STANDBY mode, press the START TEST switch.

( ) 2. The subject must remove foreign objects from the nose and mouth including dentures. The subject must be closely and continuously observed for 20 minutes prior to testing to assure no belching, regurgitation or intake of any foreign material by nose or mouth has occurred. If such occurs, another 20 minutes of close and continuous observation must elapse under the same conditions.

Start Time: _______________ Stop Time: _______________

( ) 3. Verify that the external breath tube and simulator vapor tube are both warm.

( ) 4. Observe the simulator temperature is between 33.8 degrees Centigrade and 34.2 degrees Centigrade.

( ) 5. Press the START TEST switch.

( ) 6. Follow the instructions and sequence of events as they appear on the EBAT device display.

( ) 7. Retain all printouts generated by the EBAT device with the DUI packet. (i.e. Error message printouts)

THIS EVIDENTIAL BREATH ALCOHOL TEST WAS CONDUCTED IN ACCORDANCE WITH THE COLORADO BOARD OF HEALTH RULES AND REGULATIONS, 5-CCR1005-2.

Certified Operator or Instructor Conducting Test

Refer to5 CCR 1005-2page 25.

Page 4: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-4

Appendix 2A

1. Turn power button on or observe the power button has been activated. If the EBAT deviceis in the STANDBY mode, press the START TESTbutton.

Page 5: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-5

Appendix 2A

2. The subject must remove foreign objects fromthe nose and mouth including dentures. Thesubject must be closely and continuouslyobserved for 20 minutes prior to testing to assure no belching, regurgitation or intake of any foreignmaterial by nose or mouth has occurred. If suchoccurs, another 20 minutes of close andcontinuous observation must elapse under thesame conditions.

Start Time: Stop Time:

Page 6: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-6

Appendix 2A

3. Verify that the external breath tube andsimulator vapor tube are both warm.

ExternalBreathTube

SimulatorVaporTube

Page 7: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-7

Appendix 2A

4. Observe the simulator temperature is between33.8 degrees Centigrade and 34.2 degreesCentigrade.

Simulator Display

Page 8: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-8

Appendix 2A

5. Press the START TEST button.

Page 9: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-9

Appendix 2A

6. Follow the instructions and sequence of events as they appear on the EBAT device display.

Page 10: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-10

Appendix 2A

7. Retain all printouts generated by the EBAT devicewith the DUI packet. (i.e. Error messageprintouts)

Page 11: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-11

F1: EBAT Printout

Function Keys

Active when the Intoxilyzer 5000EN is in the Ready Mode and the printer buffer still contains the results of the last EBAT performed.

Note: Evidential Breath Alcohol Test (EBAT)

Prints the last EBAT test record.

Page 12: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-12

F2: Print Simulator Log

Function Keys

When depressed, it clears the printer buffer. The F1 key will no longer print the last EBAT due to the printer buffer being cleared.

Prints the current simulator log.

Active when the Intoxilyzer 5000EN is in the Ready Mode.

Page 13: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-13

F5: Data Entry Return Key

Function Keys

Is active only when in a data entry mode.

Returns the Intoxilyzer 5000EN to the Ready Mode when in a data entry operation.

Page 14: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-14

F6: Subject Non-Compliance Key

Function Keys

Aborts the Test Sequence. Is active only from the end of the 2nd Air Blank until the end of the 2nd Breath Sample.

Used only when the subject belches, regurgitates, or intakes a foreign material into their nose or mouth before the successful completion of the second breath test.

Not to be used as a Refusal!

Page 15: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-15

Stand By Mode

The purpose the Stand By mode is to extend thelife of critical electronic components in the Intoxilyzer5000EN.

The power to the heater tapes is reduced during the Stand By mode. This allows the Sample Chamber, ExternalBreath Tube, Internal Breath Tube, and Heated SimulatorTube to warm up quickly to their proper operating temperatures within ten minutes when the Start/Test Switch is depressed.

The Stand By mode is automatically entered whenever the instrument is in-active for 120 minutes.

Page 16: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-16

Stand By Mode

Three indications that the instrument is in the Stand By Mode are:

The Intoxilyzer 5000EN display is blank.

The red LED below the Power On button is lit.

The Guth 2100 Simulator displays ‘IDLE’.

Page 17: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-17

Stand By Mode

To bring the Intoxilyzer 5000EN out of the Stand By modeand into the Ready mode, depress the Start/Test Switch.

The instrument will display the following message:

And perform a short air purge

NOT READY

Page 18: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-18

Stand By Mode

The instrument will now display:

Message repeats

WARM UP PERIOD

XX:XX MST/MDT

Current Time

DATE XX/XX/XXXX

Current Date

Page 19: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-19

Stand By Mode

The ‘WARM UP’ period allows the Intoxilyzer 5000EN and Guth 2100 Simulator to warm up to their proper operating temperatures.

After the Warm Up period, the instrument will perform selfdiagnostic tests. These tests will ensure the Intoxilyzer 5000EN is ready to accurately perform Evidential Breath Alcohol tests.

Once all these tests have passed, the instrument will enterthe Ready mode.

Page 20: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-20

Diagnostic Tests

Checks the temperature of the Sample Chamber toensure it is between 45C and 54C.

If this test fails, the instrument will pause theDiagnostic Tests and display the current

temperatureof the Sample Chamber. Once the temperature

is correct, the Diagnostic Tests will continue.

TEMP CHECK

During the diagnostic tests, numerous checks areperformed to verify the proper operator of theIntoxilyzer 5000EN. Some of these tests are:

Page 21: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-21

Diagnostic Tests

Tests the interface between the instrument and printer.

Printer error information is displayed if detected. This error information will be displayed until the

erroris corrected (i.e. Printer offline, Out of Paper).

PRINTER CHECK

Indicates all tests have passed.

DIAGNOSTIC OK

Page 22: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-22

Ready Mode

The above messages will continue to repeat until the Start Test Switch is depressed or

after 120 minutes of inactivity and the

instrument will go into the Stand By Mode.

CMI INC INTOXILYZER - ALCOHOL ANALYZERCO MODEL 5000EN --- PUSH BUTTON TO

START TEST

PUSH BUTTON

HH=MM MST/MDT

DATE MM/DD/YY

Scrolling Flashing

Page 23: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-23

Data Entry

ENTRY MESSAGE MAX CHAR TYPE FORMAT REQUIRED PRINTED STORED NOTES1 SUB LAST NAME = 15 Alpha/Numeric None Yes Yes Yes2 SUB FIRST NAME = 15 Alpha/Numeric None Yes Yes Yes3 SUB MIDDLE NAME = 15 Alpha/Numeric None Yes Yes Yes4 SUB DOB = 6 Numeric MMDDYY Yes Yes Yes5 SUB SEX = 1 Alpha/Numeric M or F Yes Yes Yes6 ZIP CODE = 5 Numeric XXXXX No No Yes7 SSN = 11 Numeric XXX-XX-XXXX No No Yes8 STATE OF ISSUE = 2 Alpha None No No Yes

8A SUB DRIV LIC = 15 Alpha/Numeric XX-XXX-XXXX No No Yes Colorado 15 Alpha/Numeric XX-XXX-XXXX No No Yes Other

9 CRASH Y/N 1 Alpha Y or N Yes No Yes9A INJURY Y/N 1 Alpha Y or N Yes No Yes10 ARREST OFFICER = 20 Alpha None Yes Yes Yes11 AGENCY = 15 Alpha None Yes Yes Yes New Entry12 OPER NAME = 20 Alpha None Yes Yes Yes13 AGENCY = 15 Alpha None Yes Yes Yes New Entry14 OPER CERT DATE = 6 Numeric MMDDYY Yes Yes Yes See Notes15 CASE NO. = 15 Alpha/Numeric None No Yes Yes New Entry16 REVIEW DATA? Y/N 1 Alpha Y or N Yes Yes No

Page 24: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-24

Test SequenceA C A B A I A 2 A B A C A

The test sequence of the Intoxilyzer 5000EN is comprisedof a series of steps or functions that ensure that when aerror free print out is obtained the result reported isreliable and accurate.

This is accomplished by performing numerous internaltests checking the operation of the Intoxilyzer. If theinstrument detects an operation or result that is improper,it will generate the appropriate error message and abortthe test sequence.

Page 25: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-25

Test SequenceA C A B A I A 2 A B A C A

There are five functions that are performed during thetest sequence. Each function executes specific operationsas follows:

A – Air BlankC – Calibration CheckB – Breath TestI – Internal Standard Test2 – Two minute wait period.

Most of these functions are performs two or more timesduring the sequence.

Page 26: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-26

Test Sequence

Air Blank

A C A B A I A 2 A B A C A

There are seven Air Blanks performed during the testsequence. The primary purpose is to circulate room airthrough the sample chamber to remove any contaminatesthat may be present. This guarantees the sample chamberis clean prior to a Calibration Check, Breath Test, orInternal Standard Test.

Page 27: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-27

Test Sequence

1st Air Blank (continued).

XX:XX MST/MDT

DATE XX/XX/XXXX

AIR BLANK

A C A B A I A 2 A B A C A

AIR BLANK

During an Air Blank, the instrument will display thefollowing:

Page 28: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-28

Test Sequence

1st Air Blank (continued).

A C A B A I A 2 A B A C A

AIR BLANK .000

Once complete, the instrument will display the resultof the Air Blank.

Page 29: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-29

Test Sequence

CAL CheckThere are two calibration checks (CAL Check) performedduring the test sequence. The primary function is to ensurethe instrument is within calibration limits (± 10%).

This is accomplished by re-circulating the vapor from theGuth 2100 Simulator through the sample chamber and backto the simulator.

The re-circulation allows the utilization of the standardsimulator solution for up to 96 subject tests.

A C A B A I A 2 A B A C A

Page 30: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-30

Test Sequence

CAL Check

A C A B A I A 2 A B A C A

. . . . . . .

CAL. CHECK

CAL. CHECK .###

During a CAL Check, the instrument will display:

At the completion of the second CAL Check, the instrumentwill compare the results of the two CAL Checks to ensurethey are within 10% of each other.

Page 31: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-31

Test Sequence

Breath Test

A C A B A I A 2 A B A C A

The primary function of a Breath Test is to accurately determine the BrAC of the subject.

This is accomplished by the Beer-Lambert law.

‘Absorption is directly proportional to concentration.’

The following basic formula is used for this calculation:

X – Y = Alcohol Concentration (BrAC)

Were:

X = Previsious Air Blank value (IR Concentration).Y = Current sample chamber value (IR Concentration).

Page 32: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-32

Test Sequence

Breath Test (continued)

A C A B A I A 2 A B A C A

. . . . .

. .

PLEASE BLOW/R INTO MOUTHPIECEUNTIL TONE STOPS

PLEASE BLOW/R

The instrument will display the following at the start of abreath test.

Page 33: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-33

Test Sequence

Breath Test (continued)

A C A B A I A 2 A B A C A

When the instrument displays ‘Please Blow/R’, the subjecthas approximately three minutes to provide a valid breathsample. If the subject does not provide a valid breathsample, a ‘Deficient Sample’ error message will be generated and the test sequence will be aborted.

As long as the instrument displays ‘Please Blow/R’, theoperator can initiate a ‘Refusal’ if the subject refuses tofollow their instructions (“R” + <ENTER> on keyboard).

Once the subject starts blowing, the instrument will display‘Please Blow’ and the refusal option can no longer be performed.

Page 34: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-34

Test Sequence

Breath Test (continued)

Valid Breath Sample: There are four requirements for a successful breath sample. They are:

1. Flow Rate

2. Time

3. Volume

4. Slope

A C A B A I A 2 A B A C A

Page 35: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-35

Test Sequence

Breath Test (continued)

1. Flow Rate: The subject must blow into the mouth piece at a rate of ≥ .15L / second. This minimum flow rate must be maintained until all four requirements are met.

Valid Breath Sample

A C A B A I A 2 A B A C A

2. Time: The subject must blow, at an acceptable rate, for a minimum of 1 second.

Page 36: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-36

Test Sequence

Breath Test (continued)

Valid Breath Sample

3. Volume: The subject must provide, at an acceptable rate, a minimum of 1.1 Liters of breath sample.

A C A B A I A 2 A B A C A

4. Slope: Once the first three requirement are met, the Intoxilyzer will determine a level slope of the breath sample within the sample chamber. A level slope is when the BrAC inside the sample chamber no longer increases by more than 3.2% per second or deceases by more than 1.6% per second.

Page 37: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-37

Typical Breath Sample:

0 1 2 3 4 5 6 7 8 9 10

Seconds

BrAC

Rising Slope during the first 4 seconds and the

first three requirements are met.

Slope levels out at approximately 4

seconds

Level Slope Detected

Subject Stops Blowing

Test SequenceA C A B A I A 2 A B A C A

Page 38: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-38

Test SequenceA C A B A I A 2 A B A C A

Mouth Alcohol:

TIME

0 1 2 3 4 5 6 7 8 9 10

Positive Slope during the first 3 seconds.

The first 3 requirements are

met.

Slope peaks at approximately 3 seconds

and should level out.

Slope then start to continually drop from the peak at a rate

of ≥ 5%.

BrAC

Page 39: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-39

Test Sequence

Breath Test (continued)Instrument checks for interferent conditions.

( Acetone, Toluene, and other common alcohol.)

NOTE: If an interferent condition is detected, the instrument will display INTERF DETECTED. After a few seconds, if the instrument displays SUBTRACTED, it indicates to the operator that the interferent was Acetone.

Acetone has similar effects on the human brain as does Ethanol. A diabetic with an improper blood sugar level can naturally produce Acetone in their body, and in some cases exhibit the same mannerism as an individual who is intoxicated.

The subject needs immediate medical assistance!

A C A B A I A 2 A B A C A

Page 40: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-40

Test Sequence

Breath Test (continued)

SUBJECT TEST . - - -

The instrument will not display the results of thebreath test.

This helps in ensuring the subject will providea second valid breath sample later in the testsequence.

A C A B A I A 2 A B A C A

Page 41: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-41

Test Sequence

Breath Test (continued)

PLEASE REMOVE SPIT TRAP ANDDEPRESS START TEST SWITCH

REMOVE SPIT TRAP

Scrolling

Flashing

Operator must comply with the above instrument message within approximately 1 minute and 10 seconds. If not, a ‘Step Expired’ error message will occur and the

test will be aborted.

A C A B A I A 2 A B A C A

Page 42: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-42

Test Sequence

Breath Test (continued)

A C A B A I A 2 A B A C A

At the completion of the 2nd Breath Test, the two will becompared to ensure they are within 0.02 of each other.

If they are outside this limit, a ‘NO 02’ agreement error will be generated.

Page 43: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-43

Test Sequence

Internal Standard TestINTERNAL STD

INTERNAL PASS

1. This is a check of the instrument’s calibration and linearity using internal electronics values that equate to BrAC values of 0.100, 0.200, and 0.300. The results of all three Internal Standards must be within +/- 5% of the target values.

2. The addition of this step brackets the subject’s 1st and 2nd breath test with an instrument calibration and linearity check.

3. Will reduce challenges to the breath test accuracy by demonstrating the instrument was working properly after and before the subject provided the two breath samples.

4. Same Internal Standards test that is performed during the Diagnostic Checks.

A C A B A I A 2 A B A C A

Page 44: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-44

Test Sequence

2 Minute Wait Period.

A C A B A I A 2 A B A C A

WAIT 2 MINUTES

WAIT 01.56

WAIT 00.00

Display countsdown to zero

Allows the subject’s lungs to equilibrate to the 1:2100 ratio based on Henry’s Law.

1 second tone

Page 45: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-45

Ready Mode

CMI INC INTOXILYZER - ALCOHOL ANALYZERCO MODEL 5000EN --- PUSK BUTTON TO

START TEST

PUSH BUTTON

HH=MM MST/MDT

DATE MM/DD/YY

Scrolling Flashing

The above messages will continue to repeat until the Start Test Switch is depressed or

after 120 minutes of inactivity and the

instrument will go into the Stand By Mode.

Page 46: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-46

EBAT PRINT OUT

Header

Data Entry

TestSequence

Signature

Page 47: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-47

EBAT PRINT OUT

Header

Serial Number providedby instrument.

Page 48: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-48

EBAT PRINT OUT

Data Entry

SEQUENCE NUMBER: 00277SUB NAME = DOE, JOHN, QSUB DOB = 11/24//90ZIP CODE = 99999SUB DRIV LIC = CO / 999-99-9999OPER NAME = SMITH

INTOX. RE-CERT DATE = 04/18/07

COPY NO X OF XX

CASE NUMBER = 01724SIM SOL NO = 11-11-11SUB SEX = MSSN = 999-99-9999

AGENCY = CDPHEAGENCY = CDPHEOPER CERT DATE = 12/11/07

Sequence Number, Intox Re-Cert Date, Copy No and Sim Sol No are provided by the instrument.

Page 49: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-4910-49

EBAT PRINT OUT

Test Sequence

Displays the test sequence alongwith the results of each step.Includes simulator temperaturesand breath volumes.

Times of the completion ofeach step are also printed.

Reported value equals the lesserof the two independent breathsamples.

Page 50: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-50

EBAT PRINT OUT

Signature

Operator must sign the signature line of the EBAT printout

Page 51: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-51

THE COMPLETED EBAT

1. Complete and sign Appendix 2A.

2. Sign EBAT printout.

Page 52: Rev. 3.099-1 The Complete EBAT Colorado Department of Public Health and Environment Laboratory Services Division EBAT Program (Evidential Breath Alcohol

Rev. 3.09 9-52

1. Appendix 2A Review2. Functions Key3. Stand By Mode4. Warm Up Period5. Diagnostic Tests6. Ready Mode7. Test Sequence

(ACABAIA2ABACA)8. EBAT Print Out