Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
UNIVERSITY OF LATVIA
FACULTY OF PHYSICS AND MATHEMATICS
DEPARTMENT OF OPTOMETRY AND VISION SCIENCE
RETINAL IMAGING AT VARIOUS ILLUMINATIONS
MASTER’S THESIS
Author: Daiga Čerāne
Student ID: dc10014
Supervisors: Ph. D. Pauli Fält
M. Sc. Piotr Bartczak
Assistant Proffesor Pēteris Cikmačs
RĪGA, 2015
ABSTRACT
Master’s thesis is written in Latvian and English, contains 48 pages, 33 figures and 6 tables
and 37 references to literature sources.
Purpose: create a spectrally tunable light source for computer-aided diagnosis of diabetic
lesions and other retinal features.
Methods: A digital micromirror device (DMD) based spectrally tunable light source with a
modified fundus camera and monochromatic camera for recording of images. Images taken of
both healthy human subject and diabetes patients with diabetic retinopathy diagnosis.
Conclusions: Spectrally tunable light source with optimal illuminants is compatible with a
fundus camera for retinal imaging to improve contrast of retinal lesions and features.
Keywords: retina, diabetic retinopathy, fundus camera, spectrally tunable light source, DMD.
CONTENT
ACRONYMS AND ABBREVIATIONS ................................................................................ 1
INTRODUCTION .................................................................................................................... 2
1 LITERATURE REVIEW ................................................................................................. 3
1.1 Fundus of the eye and diabetic retinopathy ................................................................ 3
1.2 History of artificial light sources ................................................................................ 11
1.3 Retinal imaging ............................................................................................................ 14
2 EXPERIMENT ................................................................................................................ 22
2.1 Objective and tasks ...................................................................................................... 22
2.2 Methods ........................................................................................................................ 23
2.3 Subjects ......................................................................................................................... 29
2.4 Results and result analysis .......................................................................................... 31
2.5 Discussion ..................................................................................................................... 40
CONCLUSIONS ..................................................................................................................... 41
ACKNOWLEDGMENTS ...................................................................................................... 42
BIBLIOGRAPHY ................................................................................................................... 43
ATTACHMENT: ANALYZED RETINAL IMAGES ........................................................ 46
1
ACRONYMS AND ABBREVIATIONS
ILM – internal limiting membrane
NFL – nerve fiber layer
GCL – ganglion cell layer
IPL – inner plexiform layer
INL – inner nuclear layer
OPL – outer plexiform layer
DM – diabetes mellitus
IDDM – insulin dependent diabetes mellitus
DR – diabetic retinopathy
NPDR – non-proliferative diabetic retinopathy
PDR – proliferative diabetic retinopathy
AD – Anno Domini
BC – Before Christ
IR – infrared
NIR – near infrared
RGB – a color system that consists of red, green and blue channels, that allows to reproduce
wider array of colors
LED – light emitting diode
LASER – light amplification by stimulated emission of radiation
OD – right eye (oculus dexter from Latin)
OS – left eye (oculus sinister from Latin)
fps – frames per second
NA – not applicable
2
INTRODUCTION
Diabetes is a disorder that in 100% of cases after 20 years show some kind of retinal
lesions. The diabetic retinopathy is leading cause of legal blindness in people aged 20-74.
Therefore constant supervision of an ophthalmologist is required.
Usually color and standard red-free fundus images of diabetes patients are recorded.
However it is found that these images not always showcase the retinal lesions and features at
the highest possible contrast, therefore it is the aim of this study to create an optimized
(spectrally tunable) light source that could aid in computerized diagnostics of such lesions and
features.
The objective of this study was to create a spectrally tunable light source for computer-
aided diagnosis of diabetic lesions and other retinal features.
Tasks:
1. Create a spectrally tunable light source that can be used in setup with fundus camera
optics.
2. Calibrate the spectrally tunable light source for precise retinal imaging.
3. Obtain images of mechanical model of an eye, determine if RGB images can be
recreated.
4. Obtain images of human subjects suffering of diabetic retinopathy and healthy human
subjects. Determine if contrast of optimal illuminations is an improvement over
standard red-free fundus images.
3
1 LITERATURE REVIEW
The literature review shows the basics of eye anatomy, and touches the subject of diabetic
retinopathy. It also shows the history of fundus research, as well as the most commonly used
light sources in fundus imaging in past, as well as currently used illuminations (and filters). To
correctly observe historically available illuminations, a brief history of artificial light sources
is also shown.
1.1 Fundus of the eye and diabetic retinopathy
1.1.1 Basic anatomy of human eye
Basic eye anatomy is separated into two parts: anterior segment: cornea, iris, ciliary body
and lens and the posterior segment that consists of vitreous humor, retina, choroid and optic
nerve head. Choroid, ciliary body and iris make up the uveal tract. The outer segment of the
eye is made of collagen – sclera, corneal limbus and cornea. The inner part of eye – retina,
together with Bruch’s membrane and choroid make up the eye fundus. The structure of the eye
is seen in Fig. 1.1.
Fig. 1.1. Anatomy of eye1
1 http://biology-forums.com/gallery/medium_57728_04_02_13_5_43_26.jpeg
4
1.1.2 Retinal anatomy
The retina is structure located at the back of the eye analogue to camera film. The neural
retina processes photons – incoming visual stimuli and converts it into electrical signal, that is
transported to cortex to be analyzed. All parts of retina except for fovea, ora serrata and optic
disc are made up of 10 retinal layers. The photoreceptive cells in retina are called rods and
cones and are located in the outer part of the retina. Cones are responsible for color vision and
in case of normal color vision there are three different kinds of cones (long, medium and short
wavelength peak sensitivity), rods are responsible for mesopic (twilight) and scotopic (night)
vision. (Fig. 1.2.). (1) (2)
Fig. 1.2. Layers and levels of retina. Layers correspond to the anatomical structure. Levels are
distinguished in relation to fundus photography. (3)
Fovea is a part of retina that has the highest density of photoreceptors – cones. There are
no rods in the central part of the fovea. The center of the foveola (Fig. 1.3.) called umbo
produces the highest possible visual acuity. It consists of tightly packed cones (their size is
exceptionally small compared to the rest of retinal cones, where cones in the fovea are at the
size of 1.5µm, but in the peripheral parts of retina – 6µm)2, external limiting membrane and
pigment epithelium, it is separated from inner nuclear layer by the horizontal oblique fibers of
Henle and between nuclei feature Müllerian fibers. (1)
2 http://webvision.med.utah.edu/book/part-ii-anatomy-and-physiology-of-the-retina/photoreceptors/
5
Fig. 1.3. Structure of retina in the area of foveola (1)
Ora serrata is a part of eye anatomy where vitreous body is fixed to the retina (one being
around the optic disc, and the other being at the ora serrata, that is the extreme periphery of the
eye). At this point the transition of neural part of retina transforms into non-pigmented
epithelium and pars plana of ciliary body, the internal limiting membrane is continued as
basement membrane. The retinal pigmented epithelium and Bruch’s membrane are continued
as pigmented epithelium of ciliary body and its basement membrane. As seen in Fig. 1.4. (1)
Fig. 1.4. Structure of retina in the area of ora serrata (1)
Optic nerve head is the place where the axons of ganglionar cells combine and make up
the optic nerve. As seen in Fig. 1.5., the external limiting membrane combines with retinal
pigment epithelium that is supported by border tissue of Kuhnt. The border tissue continues to
choroidal level, and further the axons fixate the retina to scleral lamina cribrosa. Also, it is the
other fixation place of vitreous body. (1)
6
Fig. 1.5. Structure of retina in the area of optic nerve (1)
1.1.3 Diabetes and diabetic retinopathy
Diabetes Mellitus (DM) is a metabolic disorder that is characterized by elevated blood
glucose level. Elevated blood sugar level may be caused by higher intake of glucose, decreased
cell intake of glucose or lowered blood insulin levels. All these factors make an impact to the
glucose homeostasis. (4) There are two types of diabetes – Type I and Type II, both types are
polygenic in nature – they are caused by genetic predisposition combined with environmental
factors. (5)
Diabetes is the leading factor in the USA for causing legal blindness between the ages of
20 and 74. DM patients are five times more likely to become legally blind than people without
DM. (4)
Type I diabetes patients are diagnosed early – usually before reaching the age of 30. Type
I diabetes mellitus is also called insulin dependent diabetes mellitus (IDDM), the acronym was
dropped in 1995 by American Diabetes Association and from then on has been called Type I
DM. In Type I DM, the production of insulin is almost non-existent, which is due to almost
complete loss of beta cells (always more than 80%), therefore patients with Type I DM are
completely dependent on external injections of insulin, to ensure their survival. The onset of
the disorder is usually triggered by an infection such as chickenpox. At the time of infection, a
virus triggers a response from lymphocytes that destroy pancreatic beta cells, however the
process of beta cell destruction can be also idiopathic. These patients (Type I) usually become
both ketotic and acidotic very easily due to imbalance of sugar and insulin intake. (4) (5)
Type II is an adult onset diabetes. Historically, it’s also called NIDDM – non-insulin
dependent diabetes mellitus. The onset of disorder is steady, and often the early stage Type II
DM is relatively symptomless, making it hard to identify. Upon measurements blood insulin
7
levels are normal or even elevated. The reason for raised blood sugar levels are reduced
sensitivity of insulin receptors on cell membranes. (5) Pancreatic beta cells may not be lost
completely, the damage might be caused by pancreatic or other endocrine gland diseases. (4)
The malfunction of cell’s insulin receptors is often caused by an unhealthy lifestyle: lack of
physical exercise and obesity. (5) Therefore patients with Type II DM usually have increased
blood glucose level, decrease of insulin production and partial or complete insulin resistance.
At the time of diagnosing, patients are usually obese (opposite of Type I patients, where patients
are thin), resistant to acidosis and ketosis, as well as diabetic coma. (4) (5)
About 20% of patients with DM experience shifts in the refraction of their eyes. When
patients have high glucose serum levels they experience more myopic refraction that disappears
when blood sugar levels normalize. There are occasions during hypoglycemia when patients
experience more hyperopic refraction, therefore patients experience visual acuity problems with
both their near and far sights. (6)
For DM patients, the risk to develop DR depends on duration of the DM, arterial
hypertension, microalbuminuria, renal failure, pregnancy, dyslipidemia and lifestyle (smoking
etc.) (6), and cataract extraction. In order to prevent the development of DR, strict blood serum
glucose control is required. (4) (7)
Diabetic retinopathy for Type I DM usually doesn’t start in the first 3-5 years after
diagnosis. However statistics show that after 20 years of DM 100% of patients with Type 1 DM
have some form of DR. Furthermore, 50% of these it is proliferative retinopathy. (4) After 30
years of insulin therapy 12% of patients are legally blind. (6) For Type II DM patients the onset
of disease is unknown, therefore statistics are not complete. DR sometimes is the first sign of
diabetes. (4) Statistics show that after 15 years of Type II DM 25% of patients have PDR. After
35 years 67% of patients have PDR. (6) (2)
Pathogenesis of DM include elevated blood sugar level that can alter genetic expression,
therefore altering cellular response, there is free radical formation, neovascularization. (4)
One of the first DM expressions is capillary membrane thickening due to elevated glucose
levels. Over time collagen deposits in the basement membrane of capillaries, pericytes are
destroyed and capillary walls bulge out, which may lead to endothelial cell proliferation. These
bulges are microaneurysms, their formation seen in Fig. 1.6. Microaneurysms are often
accompanied by dot hemorrhages. The differential diagnostics are performed by fluorescein
angiography (FA): dot hemorrhages appear black, but microaneurysms are seen as bright spots,
as their structure is connected to the blood stream. In this stage DR is called NPDR – non-
proliferative diabetic retinopathy. (4) (6) (7) (8)
8
Fig. 1.6. Capillary arms fusing together to form microaneurysm (8)
Retinal hemorrhages are divided in two groups by their location and appearance. Flame-
shaped hemorrhages are located in the retinal nerve fiber layer and are caused by the
hemorrhaging of larger blood vessels – arterioles, therefore hemorrhages appear flame shaped.
Intraretinal hemorrhages are caused by bleeding from the end of capillaries and therefore have
blot-dot shape. As seen in Fig. 1.7. (6) (8)
Fig. 1.7. Blot hemorrhages (on the left) and flame hemorrhages (on the right) (8)
Hard lipid exudates can appear in the later stages of NPDR. Hard exudates are associated
with elevated blood lipid levels, they are intraretinal lipid deposits that are caused by lipoprotein
leakage from blood vessels due to breakdown of endothelial tight junctions, and they are formed
mainly in outer plexiform layer by lipoprotein combination with lipid-filled macrophages. In
fundus images hard exudates appear as white-yellow deposits with well distinguished borders.
Seen on the right image in Fig. 1.7. (4) (6) (8)
Macular edema (swelling) is also associated with NPDR. This symptom is better known
to be exhibited by Type II patients, for Type I DM patients it is usually associated with PDR.
Macular edema makes retina thicker, and there are several classifications: thickening of
9
macular region (thickness around 500 nm), or thickening of retina in the region within one
optical disc diameter from the fovea, or presence of hard exudates, combined with the
thickening of the retina around 500 µm from the fovea. (2) (4) (7)
Other NPDR symptoms are early neovascularization to compensate for nonperfusion.
This is called intraretinal microvascular abnormalities (IRMA). Overtime IRMA can also
include hemorrhages and dilated retinal veins causing beady veins (Fig. 1.8.) If venous beading
is seen in two quadrants in the fundus image, the condition is called severe NPDR. Cotton wool
spots (aka soft exudates or microinfarcts) are also associated with severe NPDR (Fig. 1.8.).
They are caused by capillary nonperfusion areas – retina has edema in those regions, it is
ischemic and it produces angiogenesis factors. (4) (6)
In the cases of mild and moderate NPDR it is found that grid laser photocoagulation
reduces the risk of loss of visual acuity in all cases of macular edema for about 50-60%.
Photocoagulation scars on retina are seen as white/yellow spots, however with time they may
acquire pigmentation. Photocoagulation is also used to treat microaneurysm leakage. (4) (7)
Fig. 1.8. Pre-proliferative DR. Visible – arterial abnormalities, cotton wool spot, IRMA, beaded vein
and capillary closure. (8)
Proliferative diabetic retinopathy (PDR) is caused by severe neovascularization and
formation of fibrous tissue that extends outwards from retina into the vitreous cavity.
Neovascularization is caused by ischemic response from retina – it produces angiogenesis
factors that lead to formation of new blood vessels in ischemic regions in the retina, on the optic
disc and on the iris (iris rubeosis). PDR neovascularization is severe and makes complex, often
looping structures, unlike in NPDR, where the neovascularization is contained to a small region.
PDR neovascularization is often accompanied by fibrotic growth, which can cause vitreous
body detachment, retinal dragging, and tractional retinal detachment. Newly formed blood
10
vessels are easily torn and which causes hemorrhaging in the vitreous body (hemophthalmos).
PDR causes severe visual acuity loss and may lead to legal blindness. (2) (4) (6) (7)
In the cases of severe NPDR and PDR scatter photocoagulation is used to prevent-caused
glaucoma. The complications of this procedure include macular edema, visual field loss and
night blindness. (4) (6)
11
1.2 History of artificial light sources
The history of artificial light sources dates back as far as 400’000BC when first humans
discovered fire. More recognizable forms of actual light sources were primitive lamps that were
found around 13’000BC with animal and vegetable fat ‘lamp’ predecessors, up until 5000BC
that was the only form of artificial light, until some form of bioluminescence was discovered
(fireflies trapped in cages). The next bigger leap for development of artificial light was with
Greek oil-pottery lamps, that underwent a development from open oil lamps to closed off in a
century (600-500BC) (9)
Candles as light source are not strictly restricted to a particular time frame. Many sources
contradict, as the definition of first real candle varies. Some sources call first oil lamps as
candles. (10) However some mention only the date when candles were available as we currently
know them – 400AD. (9) Chinese first candle records date to 259-210BC. In Jewish society
candles date back to 165BC, when candles were used in religious rituals. Beeswax was used in
candle making from 15th century, before that the most popular substance was tallow. In China
it was whale fat, Japan – tree nuts. (11) Candle is also the first illumination used to observe
retina. (12)
In 1645 the magic lantern was discovered – it is the first projector type device that was
invented. 1783/4 was the first larger step in development of oil lamps when Ami Argand created
Argand lamp – based on oil lamps, but one that had hollow circular wick, and a glass cylinder
surrounding it. It produced more light than ordinary oil lamps, and more than several candles,
however it consumed more fuel. In 1790 another oil lamp variation was created – Betty lamp,
where wick was integrated, so that drips could run back to reservoir. (9)
In 1792 William Murdock used heated coal to obtain gas, and used it as illumination for
personal use. 1801 was the year electric arc was discovered by H. Davy, it proved to be high
improvement over other illumination sources for theater and street lamps. (9)
In 1825/6 first lamp using incandescence was developed – a limelight. It showed
properties of being up to 83 times brighter than conventional gas lamps, however it was not the
most convenient of illumination sources, and therefore electric arc lamps gained more
popularity. Limelight lamps found their place in theaters where a lens placed in front of lamp
created the known spotlight. (9) (13)
1853 – Kerosene lamp, still using the same technology as other oil based lamps, however
this one used paraffin as fuel. 1878 Electric candle - Jablochkoff candle that consisted of carbon
rods side by side and separated by inert material. Was never made popular as incandescent lamp
became more popular during that time. The credit of inventing incandescent lamp goes to
12
Edison who patented the incandescent lamp in 1880, as well as Joseph Swan who demonstrated
carbon filament incandescent lamp to public in 1879. In 1885 – Gas mantle by Welsback. (9)
1894 – Discovery of Argon (Ar), nowadays mostly used in incandescent lamps to provide
inert atmosphere. In 1898 – Discovered Neon (Ne) and Xenon (Xe). Following was discovery
of neon colors – neon red – when current flows through neon filled vacuum tube it glows red.
Blue color – Ar + Hg (1910-1933), 1933 marks discovery of various other neon colors. (9)
1901 – High intensity discharge (HID) lamp was first developed. The first HID lamp was
a mercury lamp, after that followed high and low pressure sodium lamps, as well as metal-
halide lamp. These lamps gave off a line spectrum, some were covered with phosphorus, thus
achieving wider spectrum. The same year Mercury vapor lamp was first discovered. (9) In the
1910s Irving Langmuir developed first gas filled lamp that slowed the tungsten evaporation
from filament and gave off 12-20 lm/W. (9) (14) (15)
In 1926 Edgerton started experiments with Xenon, discovered flashlight. These
experiments mark the beginning of Xenon flash. (16) Xenon flash is used in stroboscope,
flashlight for photography, nowadays – fundus imaging. Xenon lamp SPD comparison can be
viewed in Fig. 1.9.
Fig. 1.9. Xenon flash lamp, halogen lamp and Mercury Xenon lamp SPD comparison3
In 1932 LPS – Low pressure sodium lamps were invented, they were one of the most
energy efficient lamps at the time, modern LPS produces 220 lm/W. In 1937 low pressure gas
3 http://www.photonics.com/Article.aspx?AID=44487
13
discharge lamps were discovered – fluorescent lamps. In 1955 there was an introduction of
dichroic filters into lamps to provide cooler beam, by removing IR radiation from the emitted
light. (9)
In 1960 the first laser was patented. The development towards laser was found earlier,
however due to suspicion of threat towards safety (developer had some activity in left-wing
political activities), the patent was classified. At first there were only red He-Ne lasers. In the
same year halogen and metal halide lamps was patented. (9) (13) (15)
In 1965 there was invention of first LED. While nowadays they can be found having
many different properties (wavelength, power), at first there were only red LEDs. After that
green and amber LEDs were produced. Blue and white LEDs were discovered in the 1990s.
Due to low power consumption and long lives they have been gaining popularity ever since
they were discovered (soon after development started to replace other light sources). Today
there are LEDs of both visible spectrum, NIR and IR. All can be found used in fundus camera
as illumination. In the past several years there have been use of organic material in development
of LEDs, new material is called OLED (organic LED), and even more recently – AMOLED
(active matrix organic light emitting diode). (9) (13)
In 1966 High Pressure Sodium lamps were developed. In 1969 – HMI (as patented by
Osram) mercury medium lamp (HMI stands for Hg – mercury (H), Metal (M) and Halogen (I)).
1994 – Sulfur lamp that was made out of sulfur, combined with argon gas. This Sulfur lamp is
known to produce very bright plasma. (9)
In recent years mostly all artificial light sources have highly developed and new
inventions are fewer. In fundus imaging it can be seen that mostly still Xenon flash lamp is still
used, although it was one of the first light sources in commercial fundus imaging.
14
1.3 Retinal imaging
1.3.1 History of retinal imaging
The history of retinal imaging begun with ability to observe a living human retina with
ophthalmoscopy (1851), before this time any retinal imaging or the fact that the reflected light
through a pupil sometimes appears black or red was not explained. There had been some
research before 1851 with animal eyes (and applications with human eyes) – contribution from
Purkinje (1825) with concave spectacle glasses acting as mirror, that was not recognized until
much later. (12)
Before the actual invention of the ophthalmoscope, there had been attempts at viewing
retina. It could be seen with experiments in 1704, when Mėry noticed retinal blood vessels in
cat eyes, and in 1825 with Purkinje concave spectacle glasses, as well as 1846 Brücke’s
explanation how every eye’s pupil could be illuminated if the viewing axis coincides with
illumination axis (Cumming published the results in 1846). (12)
After introduction of ophthalmoscope by Helmoltz in 1851 there were developments in
ophthalmoscopes in terms of illumination, which were as follows: candle (used by Helmholtz),
oil lamp, Argand gas lamp, and after invention of incandescent light bulb it was used to
illuminate retina. In some cases there have been unorthodox illuminants, such as directed
sunlight. As the ophthalmoscope gained popularity in ophthalmologist society there were
several new designs for ophthalmoscope. (12)
Earliest mention of fundus photography is by Dr. Rosenburg in 1864 using fundus of
unconscious cat. No results were published. It is recorded that Dr. Rosenburg was not successful
in photographing a living human eye. (3) (17)
In 1886 first fundus photographs of human eye in vivo were taken and published by
Jackman and Webster. The viewed area of the fundus was limited (even though mydriatic
agents were applied), details were not easily discernable and the image was distorted by a large
reflection of cornea. Scientists Jackman and Webster also mentioned that the exposure time was
2.5 minutes – a very long time for retinal imaging. Illumination used was albocarbon
(naphthalene) – that is toxic. The Jackman’s and Webster’s published image is seen in Fig. 1.10.
(3) (18)
15
Fig. 1.10. First ever image taken of human eye in vivo. With red arrow area of optic disc is shown, as
well as a major blood vessel. (18)
Espacenet is a patent database that holds in it over 90 million registered patents from
1836. However the first inventions of fundus cameras up until 1941 are not recorded there. (19)
In 1887 L. Howe published an article concerning improvement of recorded film of eye fundus.
He suggested another illumination for retinal observation – an Argand gas lamp. The mentioned
article does not show any images, though the author mentions that the designed setup shows an
improvement over Jackson’s and Webster’s published images in 1886. The same year Barre
wrote of his findings: clear image of optic nerve and optic disc. To his mind it was due to short
camera lens focus (3 in = 7.62 cm), that made the image bright and therefore reducing the
exposure time to 6-10 seconds. (17) (3)
Further development in fundus photography include Gerloff’s 1891 low magnification
photographs and Thorner’s 1903 apparatus (uneven illumination). The first truly discernable
photographs were obtained by Dimmer shown in public and published in period 1899-1907.
Dimmer used carbon arc illumination in his camera. The photographs of these scientist
experiments are seen in the Fig. 1.11. (3)
Fig. 1.11. Fundus photographs of Gerloff (on the left), Thorner (in the middle) and Dimmer (on the
right) (3)
16
It can be seen that fundus photography developed quickly, and in 1926 Zeiss Company
marketed the first commercially available fundus camera. It had only 10 degree field of view,
but already had a color film. The photography took only ½ of second of exposure. Its design
was based on Gulstrand’s principles and was designed by Nonderson. The color photography
of eye fundus at the time was not very popular, as it was very expensive at the time. At the time
of Zeiss camera use the only available illumination was still carbon arc, it continued to be so
up until 1940’s. (3)
With the invention of flash tube in 1946, there was possibility of another illumination
source for fundus cameras, it was first used in ophthalmology in 1949 by Rizutti and
implemented in fundus cameras in 1953. (3)
At this time the first patented fundus cameras appeared and are found in Espacenet to this
date. First three patents recorded are during years 1941 (John A. Clarke – A fundus camera),
1952 (John H. Mcmillin – A fundus camera with included fixation means) and 1959 (Noyori
Kimiharu – A handheld fundus camera), though none include illumination’s used or the
schematics. After that time fundus cameras started to get patented more often and their
distribution over the years can be seen in Fig. 1.14., the total amount of patents during those
years is 498. (19)
The first fundus camera schematics from Jackman and Webster include image of eye and
the light path to and from the eye. Fig. 1.12. However it does not show the precise setup of the
camera. (18) Therefore first full camera schematics that are available from the patent database
are from 1975 ‘for Eye fundus camera with focus setting device’ by M. Isao and K. Yoshimi.
In Fig. 1.13. the structure can be seen consisting of multiple parts however explanation of the
structure is not found. (19)
Fig. 1.12. Schematics of Jackman & Webster fundus camera light path (18)
17
Fig. 1.13. Schematics for fundus camera patented in 1975 by Isao & Yoshimi (19)
Decade 1971-1980 includes first development of zooming features (1980), accuracy of
feature detection (1980), pupil alignment (1976), elimination of undesired light (1978) as well
as a focusing system (1975). (19)
Starting from 1981 there are at least 2 patents a year recorded in the database
corresponding to the search term ‘fundus camera’ (Fig. 1.14.). In the decade 1981-1990 there
are novelty mentions of IR use in cameras, improving the viewing angle, implementation of
automatic and manual focusing, first mentions of non-mydriatic cameras, ammetropia
correction, stereo images, included LED illumination, first mention of implemented barrier
filters, laser beam scanning, tilting mechanisms, creating more uniform illumination and
removal of peripheral aberrations. 1989 is the year scanning laser ophthalmoscopy was first
introduced. The next decade 1991-2000 shows implementation of RGB laser, feature detection
(such as eyelid), eye detection (OD/OS), mention of polarizing plate use, improved precision
and focusing time, first mention of halogen lamp use, improved alignment for peripheral
photography, mentions of digital photography, as well as simplified, lighter, cheaper and higher
quality cameras. (19)
From 2001 until 2015 patents have more mentions of non-mydriatic cameras, as well as
larger percentage of LED and IR, NIR use in cameras, as well as improved alignment,
performance, view of angle, focusing, and higher quality images, that are smaller in size and
cheaper to produce. During those years three dimensional cameras have become more popular.
Novelty during these years includes pattern recognition, focal matching, and correction of
severe ametropia – even severe astigmatism. (19)
18
Fig. 1.14. Fundus camera related patents over the years (19)
In recent years adaptors and alternatives have been appearing. Such as fundus
photography with a mobile phone (2014) (20) or adaptor system for digital single lens reflex
(dSLR) camera. This latter system used a white LED for illumination of retina. (21)
1.3.2 Dyes used in fundus angiography
In 1871 Adolf Baeyer described an organic dye – sodium fluorescein, among many other
dyes, and after ten years it was used for the first time in ophthalmology when Ehrlich used it to
examine flow of aqueous humor. This was followed by Burke in 1910 who examined the fundus
after administration of fluorescein in coffee. Lange and Boyd in 1942 described the circulation
studies of fluorescein by injection of 5% sodium fluorescein in a vein and then observing
various tissue under excitation of purple light. Their paper states that under the magnification
of microscope they can see as the dye stain enters through artery, passes into capillaries, stain
the tissue and then returns to vein. First actual fundus angiogram was that of a cat eye, where
Chao and Flocks in 1959 published findings of circulation time of cat’s retina to be 2 seconds.
The used method was the injection of fluorescein dye in the vein and then the observation of
retina with ophthalmoscope through a cobalt blue filter. That was followed by Mclean and
Maumenee in 1960 when they published the use of angiography to obtain a diagnosis. (3) (22)
The actual technique of fluorescein angiography was published and explained by
Novotny and Alvis in 1961. Their results included actual results with human subjects in vivo.
Their techniques are a standard still used to this day. (3) (22)
Indocyanine green (ICG) dye was first found in 1950s, and approved by FDA (Food and
drug administration) for use in angiography in 1956. First documented phase of fundus ICG
0
5
10
15
20
25
30
1941
1943
1945
1947
1949
1951
1953
1955
1957
1959
1961
1963
1965
1967
1969
1971
1973
1975
1977
1979
1981
1983
1985
1987
1989
1991
1993
1995
1997
1999
2001
2003
2005
2007
2009
2011
2013
2015
Cou
nt
Years
Patent count per year (keyword: 'fundus camera')
19
angiography (ICGA) was described in 1973. (23) In ophthalmology it is primary used for
choroidal circulation documentation, as it operates in NIR region, and excitation light can
permeate deeper tissue. It is also slightly less popular that fluorescein angiography, as it can be
directly observed, without digital means of recording. Nowadays combined FA and ICGA have
become more popular. (23) (24) ICGA is contradicted for patients with renal and liver and/or
renal failure, as well as for patients with iodine and/or shellfish allergy. Both types of
angiographies are not recommended to pregnant females, even though the dyes have not been
proven to be teratogenic. (23)
1.3.3 Filters in fundus photography
Now known barrier filters are used to present several features of the eye fundus, as we
know what the depth of wavelength penetration for retinal layers is. It is known that blue light
is reflected off of nerve fiber layer. In Fig. 1.15. monochromatic image taken with blue filter
shows the nerve fiber path on the retina. Green light is reflected off of the retinal pigment
epithelium. Therefore all structures that are in the path of this light can be shown, as red
structures appear black in this illumination, green filters are used to improve contrast of retinal
blood vessels, shown in Fig. 1.15. with the red-free filter and improved contrast of
microaneurysm (shown with arrow). As the red light passes through retina it can illuminate
structural changes in choroidal level, it is shown in Fig. 1.15. in the third fundus image – arteries
are not well distinguished, however the choroidal nevus is shown clearly. (3)
Fig. 1.15. The depth of wavelength penetration for retinal layers, and their respective monochromatic
images (3)
20
1.3.4 Multispectral fundus imaging
RGB fundus camera carries information from three channels, however the multispectral
imaging of retina allows for more than one wavelength specific channel. This information is
more wavelength specific, therefore more detailed information is available.
In year 2009 there is a patent for Hyperspectral fundus camera based on tunable
wavelength light source and its description includes image recording of multiple wavelengths,
and the ability to analyze retina by morphology and spectral signatures. In 2010 another patent
was published that includes an option of recreating the SPD of a red-free filter without inserting
the filter. (19)
However up until this date there are no commercially available multispectral fundus
cameras. Though the possibilities of creating multispectral fundus imaging have been attempted
on several occasions and have been successful. These articles involve wavelength specific high
power LED illumination combined with interference filters (25), or white LED combined with
Piezo-Actuated Fabry Perot Interferometer (PFPI) (26) or halogen lamp coupled with narrow
band interference filters. (27)
1.3.5 Modern fundus cameras
In modern cameras there are two different kind of setups: for mydriatic and non-mydriatic
pupil. The setup can also be combined – hybrid fundus cameras. The non-mydriatic cameras
have developed highly in the past decade. (19) They are very useful for screening purposes, but
their disadvantages, such as no stereoscopic capability and artifacts from under and over
exposure, doesn’t allow them to completely replace the mydriatic cameras. However one of the
advantages is that the training for use of the camera can be greatly reduced. (28) (29) (30) When
non-mydriatic camera is used with pupillary dilatation, screening results are further improved.
(31) Modern screening capabilities also include mobile phone that is used as a fundus camera,
when used together with a condensing lens (an ophthalmoscopy technique). (20)
Scanning laser ophthalmoscope (SLO) uses various illumination monochromatic lasers
that differ between a brand and properties it is used for. There are SLO’s that use 488 nm or
795nm lasers – for fluorescent angiography and indocyanine green angiography respectively.
(3) There are also SLO’s that have option of blue and IR reflectance, in those cases they are
equipped with multiple lasers (488nm, 514nm, 788nm and 820nm). (32)
All modern cameras that are commercially available are digital cameras. Most of the
cameras have some kind of non-mydriatic photography available, however there are few that
are purely mydriatic cameras. Most of the cameras still use halogen lamp for observation and
Xenon flash lamp for photography, though there are other observation light sources, sometimes
21
in combination of halogen lamp – such as NIR or IR LED light source. Xenon flash lamp is
rarely replaced by other options, such as white LED.45678910
4 http://www.usa.canon.com/cusa/healthcare/products/eyecare/digital_non_mydriatic_retinal_cameras/cr_2#Specifications 5 http://www.zeiss.com/meditec/en_us/products---solutions/ophthalmology-optometry/retina/diagnostics/fundus-imaging/visucam-pro-nm.html#technical-data 6 http://www.centervue.com/product.php?id=637 7 http://www.topconmedical.com/categories/imaging-retinalcameras.htm 8 http://www.kowa-usa.com/Ophthalmic-Diagnostics/Products.php?Product-Line=Retinal-Cameras 9 http://www.csoitalia.it/en/asp/home.asp?prod=1&dbpID=16 10 http://usa.nidek.com/products
22
2 EXPERIMENT
2.1 Objective and tasks
The objective of this study was to create a spectrally tunable light source for computer-
aided diagnosis of diabetic lesions and other retinal features.
Tasks:
5. Create a spectrally tunable light source that can be used in setup with fundus camera
optics.
6. Calibrate the spectrally tunable light source for precise retinal imaging.
7. Obtain images of mechanical model of an eye, determine if RGB images can be
recreated.
8. Obtain images of human subjects suffering of diabetic retinopathy and healthy human
subjects. Determine if contrast of optimal illuminations is an improvement over
standard red-free fundus images.
23
2.2 Methods
2.2.1 Components of spectrally tunable light source
For calibration purposes a spectrometer Hamamatsu PMA-12 Photonic Multichannel
C10027-01 was used. This spectrometer works also as optical detector. The black correction is
done by a shutter closing within the spectrometer, therefore is precise. Wavelength sensitivity
from 200 nm to 950 nm. Exposure time from 19ms to 64s. Can be controlled by manufacturers
provided PMA software.11
Digital micromirror device (DMD) used in the experiment setup is manufactured by
Texas Instruments. Micromirror array consist of 1024 columns and 768 rows of micromirrors.
Each mirror can be controlled to be in an ‘on’ or an ‘off’ position by being rotated on its
torsional hinges by 12 degrees in either direction (Fig. 2.*d). These settings are controlled by a
custom made computer program called DMD/Hamamatsu scan – v2.0. The software controls
both DMD and records reading from used Hamamatsu spectrometer. The interface of the
software is shown in Fig. 2.*c.
DMD is controlled by patterns – monochromatic BMP files in the size of 1024x768 pixels
(px). Each micromirror by this program can be either activated (a black pixel) or deactivated
(white pixel). The program controls allow user to set different exposure times, and multiple
repeats of measurements and recording of real-time SPD measurements to a text file (using
Hamamatsu spetrometer). The input for the program is a folder that has one or more BMP files
in its directory. Using custom software it was possible to load all the selected patterns to the
DMD’s buffer, and if needed, use those again.
Fig. 2.*d DMD mirrors/pixels in on (on the right) and off state (on the left).12
11 http://www.hamamatsu.com/resources/pdf/sys/SDSS0008E_PMA12.pdf 12 Texas instruments DMD 101: Introduction to Digital Micromirror Device (DMD) Technology, Texas Instruments Incorporated, 2013
24
Fig. 2.*c Interface of DMD/Hamamatsu scan – v2.0
The optimal light source specta that were used for DMD pattern generation were created
during previous research by P. Fält. (33) These optimal light sources were designed to improve
the visibility and contrast of diabetic retinopathy lesions and other retinal features, such as
macula and to improve contrast between arteries and veins. The scientific computation software
MATLAB (The Mathworks, Inc., USA) was used to generate patterns for DMD input. For this
thesis, the optimal illumination patterns were used to improve contrast of hard exudates, dot-
blot hemorrhages, as well as macula and artery/vein feature contrast enhancement.
The light source used was Thorlabs HPLS-30-04, manufactured by Thorlabs, Inc., USA.
HPLS-30-04 is a high power light source that has both UV, NIR and visible spectrum
components. The luminous flux of the light source – 2800 lm, and correlated color temperature
6500K. The focal point of light source is 8.36 mm (measured from tip of cone). (34) The
measured SPD of the light source is shown in Fig. 2.*g.
Fig. 2.*g SPD of Thorlabs HPLS-30-04 and a red-free filter simulated by the DMD light source
25
In the setup a Thorlabs GR50-0305 ruled diffraction grating was used. Thorlabs, Inc.
USA. This diffraction grating has 300 lines/mm.
The liquid light guide used to transmit light from the spectrally tunable light source to the
fundus camera was a LLG0338-4 manufactured by Thorlabs, Inc., USA. This liquid light guide
provides over 70% transmission of light in the visible range, and lower transmissions in near
UV and near IR region – from 340 nm to 800 nm. The core diameter 5 mm. Liquid light guide
operates well in long term, if the temperature is kept within range from -5C to +35C.13
2.2.2 Spectrally tunable light source
Schematics of created the spectrally tunable light source are seen in Fig.2.*a. It consists
of a plasma light source that produces bright white light. Light source was chosen so that even
after spectral component removal by the DMD, the intensity of the illumination is still high
enough to illuminate the retina. Following the light source is a slit (optimal width – 600µm).
The slit is followed by a diffraction grating and the spectrally dispersed light is focused on the
digital micromirror array. Using the micromirror array, light with desired spectral power
distribution can be created and guided to a fundus camera by a liquid light guide. As the
particular light source also has UV spectral components, a UV filter was included in the setup,
due to IR spectral components heat sinks and a fan was used near the slit.
Fig. 2.*a Optical schematics of the created spectrally tunable light source
13 http://www.thorlabs.de/thorcat/19800/LLG0338-4-SpecSheet.pdf
26
Fig. 2.*h shows one loaded pattern (a) and measured output (d). If the pattern is loaded
in the DMD (illuminated as Fig. 2.*, then all the micromirrors corresponding to black pixels,
will be turned in the desired direction (towards the liquid light guide), but the micromirrors
corresponding to white pixels will tilt to reflect the light towards a light trap (as shown in Fig.
2.*h (c)).
Fig. 2.*h Pattern loaded in DMD (a), the spectrum as illuminating the DMD (b), the open
micromirrors (c) and measured output of optimal light (d)
When a particular pattern set is loaded in the DMD/Hamamatsu scan – v2.0 program, the
DMD follows through the input settings – all the loaded patterns are displayed with the set
exposure time (minimum exposure time 19ms), and repeated as many times as indicated. The
last displayed pattern stays unchanged until other settings are loaded.
2.2.3 Calibration of spectrally tunable light source
Calibration of the spectrally tunable light source was done to ensure that the output
patterns would correspond with the determined SPD’s of the algorithms. Calibration of the
wavelength was achieved with 5px and 1px columns (one by one columns changed from shorter
wavelengths to longer wavelengths). Optimal exposure time was found beforehand, so as to
achieve maximum illumination and not to get overexposed data). The Hamamatsu sensor
b a
c d
27
located at the end of the liquid light guide registered the spectral output, while the DMD
micromirrors changed.
Calibration of the intensity was achieved by activating the rows (first row activated in the
middle, then adding a row on top, next time – on the bottom, until all mirrors were activated).
Optimal exposure time was found with maximum intensity – when all the micromirrors were
in active position. As with the wavelength calibration, spectral data was collected with
Hamamatsu spectrometer.
Slit width determines output illumination power (the wider the slit, the higher the
illumination power), and calibration results for wavelength bandwidth (the wider the slit, the
broader bands). It was found in calibration (measuring from slit width of 100 µm to 600µm in
steps of 100 µm), that the wavelength bandwidth is acceptable in for slit width of 600 µm, and
considering that the best illumination power (Fig. 2*o) was achieved at these settings, this slit
width would be used.
Fig. 2.*o Slit width calibration for intensity at slit width 200 µm (on the left) and 600 µm (on the
right)
2.2.4 Fundus camera
The fundus camera optics used are of the fundus camera Topcon TRC-50V, manufactured
in 1983, by Tokyo Optical. Fundus camera had the native illumination and photography light
sources removed from optical path, the same with native photography film camera and relay
lens. Modified fundus camera parts were covered to prevent dust and other foreign particle
entrance. Existing barrier and excitation filters were locked in empty position. Fundus camera’s
astigmatism and ametropia corrective lenses were left for additional focusing and ametropia
correction means. Fundus camera optics schematics are shown in Fig. 2.*b.
28
Fig. 2.*b Schematics of the modified fundus camera used in the experiment (pattern of camera taken
from (19))
Primary calibration means and measurements were performed at the University of Eastern
Finland using a mechanical model of a human eye (Carl Zeiss Meditec, Germany). These
measurements were performed with the above mentioned fundus camera system and spectrally
tunable light source and coupling the system with a Hamamatsu ORCA Flash4.0 LT
monochrome camera. Its properties include low noise, large field of view, fast speed (up to 30
fps). The camera was controlled with native Hamamatsu provided program. For image
recording it is possible to adjust exposure time (auto-exposure also possible), offset and
binning. Possibility to control gain was not optional. Output images were saved to previously
determined location on a computer hard drive in TIFF format (format records the settings
individual to the image taken, such as camera name, date of creation, bit depth, binning,
exposure time, offset and size of image.
For measurements with human subjects a Retiga-4000DC monochrome camera was used
instead of the Hamamatsu camera. The camera has a high sensitivity for low-light applications.
The resolution of the camera is 2048x2048 px. Multiple binning options available, as well as
possible gain control. Camera records image in TIFF format, including information about
settings used during image capture (such as – exposure time, binning, gain and bit depth).14
2.2.5 Safety standards
It was found that the maximum illumination irradiance for white light at the plane of
cornea was 20 mW/cm2. ANSI safety standards allow for 12 minutes of continuous retinal
exposure. (35)
14 http://www.qimaging.com/products/datasheets/Retiga4000dc.pdf
29
2.3 Subjects
Written consent of all the participants in this study was acquired prior to the fundus
imaging. Trials followed the ethical tenets of the Declaration of Helsinki.
In the experiment there were two subjects with pre-existing diagnosis of diabetic
retinopathy and a control subject that had no previous retinal condition diagnosed. Both patients
have records at the Tampere University Hospital. The patients’ RGB fundus images can be seen
in Fig. 2.*e (subject 1) and Fig. 2*f (subject 2) respectively.
Fig.*e 1st subject’s fundus images of the right and left eye (taken 16.03.2015. at Tampere University
Hospital)
Fig.*f 2nd subject’s fundus images of the right and left eye (taken 16.03.2015. at Tampere University
Hospital)
30
Both patients eye fundus have dot hemorrhages, microaneurysms and hard exudates. 2nd
patient also has laser photocoagulation scars. DR lesions seen in Fig. 2.*e and Fig. 2.*f.
Control subject having no prior retinal condition had no available RGB images to display.
Red-free images of control subject were acquired during the experiment with the newly created
spectrally tunable light source. Images with red-free filter of both DR patients were also
acquired for comparison.
Prior to the experiment intra ocular pressure of all subjects was recorded with ICARE
tonometer (sterile, measures pressure without anesthetic). The dilation of subject’s pupils was
acquired with Oftan Tropicamide 5mg/ml, Saten Oy, Finland. After the application of
Tropicamide the intra ocular pressure was monitored periodically.
31
2.4 Results and result analysis
Due to involuntary and voluntary eye movements participants were asked to fixate gaze
on an external fixation means. For all subjects images were taken with spectrally tunable light
and with a red-free filter for contrast comparison. Then images were filtered – overexposed, out
of focus and images without red-free area matches were not analyzed. Due to use of older
fundus camera, the obtained images were never completely uniformly illuminated, therefore for
analysis of retinal features a selected area of image was analyzed (Fig. 2.*i).
Fig. 2.*i image on the left has uneven illumination around the edges and a ring of artefact (green
arrow), image on the left has been cut to avoid the artefacts
Result analysis was done using MATLAB R2014a Student Version. All taken fundus
photographs were normalized against exposure time. Normalization was done by dividing each
pixel of an image with the exposure value and then scaling the value from 0 to 1. Output of this
was monochromatic image where black pixel corresponded to value 0 and white pixel
corresponded to value 1.
Contrast calculation was done before any other image enhancements were performed
(brightness not modified), the only modification to these images was normalization. For
contrast calculation the Michelson’s contrast formula was used (36):
𝐶𝑀 =𝐿𝑚𝑎𝑥 − 𝐿𝑚𝑖𝑛𝐿𝑚𝑎𝑧 + 𝐿𝑚𝑖𝑛
Contrast calculation values were obtained by calculating average brightness of area in the
size of 3x3px.
32
2.4.1 Patterns and SPD’s
Pattern of optimal illuminant for artery and vein detection is seen in Fig. 2*h (a), and its
corresponding SPD, in Fig. 2.*h (d). The patterns used for dot hemorrhaging, hard exudates
and macula can be seen in Fig. 2.*p. Their spectral power distributions and the reference spectra
from P. Fält work (33) can be seen in Fig. 2*q. It can be observed that the actual spectra doesn’t
completely match the reference spectra, however it is due to the sharp peaks necessary for the
reference spectra.
Fig. 2.*p Patterns used for loading in DMD to obtain the optimal spectra (from left: dot hemorrhaging,
hard exudates, macula)
Fig. 2*q Reference spectra (red line) and actual spectra (blue line) for (a) arteries and veins, (b) dot
hemorrhaging, (c) hard exudates and (d) macula
b a
c d
33
2.4.2 Results of a mechanical eye model
First measurements in University of Eastern Finland were attempted with a mechanical
model of eye (Zeiss). Measurement results were recorded with Hamamatsu ORCA camera.
Measurement results do not correspond with a human eye, as the RGB image of the mechanical
eye indicates (Fig. 2.*k) – color of retina, macula, optic nerve and blood vessels does not match
a living human eye. RGB image was reconstructed using MATLAB, the illumination for each
channel were created by the spectrally tunable light source with optimal SPD’s to recreate
natural white light source.
Fig. 2*k. Reconstructed RGB image of mechanical eye model used in experiment
Each of the used illuminations (optimal for artery and vein distinction, blot bleeding, hard
exudate and macula) was used to take images of the mechanical eye model. Images were taken
with automatic exposure time of Hamamatsu ORCA camera (manufacturer provided software)
and binning of 1x1, 2x2 and 4x4 for ten times each time. The result can be seen in Table 2.*z
Table 2.*z.
The result of multiple auto-exposure data of camera at different illuminations
Binning 1x1 2x2 4x4
average exposure time, ms
standard error, ms
average exposure time, ms
standard error, ms
average exposure time, ms
standard error, ms
Arteries and veins 1658,69 2,98 409,51 0,57 101,48 0,08 Blot bleeds 2579,81 3,15 629,01 0,96 155,51 0,12 Hard exudates 1862,96 2,04 459,55 0,50 133,86 0,10 Macula 1588,60 1,96 392,96 0,91 97,34 0,11
34
Considering that the images were taken from a consistent target – a mechanical model of
an eye, it can be concluded that the cameras was relatively stable (Table 2.*z). This indicated
that for the optimal retinal illumination to retain good image quality and not over-tax patients,
optimal binning would be either 2x2 or 4x4 for camera Hamamatsu ORCA.
The actual images or mechanical eye model were normalized and are shown in Fig. 2.*j,
showing images of all above described illuminations.
Fig. 2.*j Mechanical eye model images taken with optimal (a) artery/vein, (b) blot bleed, (c) hard
exudate and (d) macula illuminations
All images show a circular artifact – Fig. 2.*i, image taken with a human subject with
Retiga 4000 DC, as well as the reconstruction of RGB (Fig. 2.*k) and the monochromatic
images in Fig. 2.*j – taken with Hamamatsu ORCA camera. This indicates that the artifact is
due to fundus camera optics, not the camera’s used.
2.4.3 Results of human retina
In some cases for artery and vein analysis the same region of arteries and veins could not
be matched to the same region (control subject). It can be seen in Fig. 2.*l, that below the optic
nerve arteries and veins are not discernable as they fall into unevenly illuminated part of retina,
and therefore those could not be analyzed. For artery and vein analysis in case of optimal
illuminants also region bellow optic nerve head was taken.
b a
c d
35
Fig. 2.*l Red-free illuminations of control subjects left eye
Control subjects images taken with optimal illuminants for artery and vein contrast can
be seen in Fig. 2.*m.
Fig. 2.*m Optimal illumination images for artery and vein contrast improvement for control
subject
The contrast values that were calculated for each region with Michelson’s formula (found
above) for the case of control subject for optimal illuminant for artery and vein distinction can
be found in Table 2.*y. It can be concluded that in case of optimal illumination for control
subject artery and vein contrast was vastly improved, if area of interest was evenly illuminated.
Artery and vein contrast comparison was also done for subject 1 and subject 2. The
contrast comparison result can be viewed in Table 2.*x, the images of selected areas can be
viewed in attachment No. 1. For these subject 1, the artery and vein contrast was improved,
however in case of 2nd subject, the improvement was not seen.
36
Table 2.*y
Contrast comparison for red free filter and optimal illuminant in case of arteries and veins for control
subject
Red-free filter contrast Optimal illuminant contrast
1. Above optic disc 11.53% 8.06% Below optic disc NA 32.91%
2. Above optic disc 1.14% 6.68% Below optic disc NA 24.62%
3. Above optic disc 5.97% Below optic disc NA
Table 2.*x
Contrast comparison for red free filter and optimal illuminant in case of arteries and veins for subject 1
and subject 2
Subject 2 Subject 1
Red-free filter contrast
Optimal illuminant contrast
Red-free filter contrast
Optimal illuminant contrast
1. Above disc 19.97% 21.25%
8.07% 22.32% Below disc 38.29% 22.50%
2.
12.33% 3. 8.17%
2nd subjects right eye was observed to determine contrast enhancement in case of dot
hemorrhages. It can be concluded that dot hemorrhage contrast is improved in case of optimal
illuminant. The images can be seen in Fig. 2.*n, the contrast results in Table 2.*w.
Fig. 2.*n 2nd subject’s dot hemorrhaging images in case of red-free (on the left) and optimal (on the
right) illuminations
37
Table 2.*w
Contrast comparison for red free filter and optimal illuminant in case of dot hemorrhages for subject 2
Red-free filter contrast Optimal illuminant contrast Hemorrhaging on the left 15.12% 18.48% Hemorrhaging on the right 15.43% 19.85%
Images of 1st subject’s left eye were analyzed for contrast comparison for hard exudates.
Fig. 2.*r shows a red-free images and an images of hard exudates for optimal illuminant. The
contrast results for this subject shown in Table 2.*v.
Fig. 2.*r 1st subject’s (a-b) red-free image with three and two hard exudates respectively (for analysis)
and (c-d) optimal illuminant images with three hard exudates, each for analysis
Table 2.*v
Contrast comparison for red free filter and optimal illuminant in case of hard exudates for subject 1
Red-free filter contrast Optimal illuminant contrast Hard exudate location Fig. 2.*r(a) Fig. 2.*r(b) Fig. 2.*r(c) Fig. 2.*r(d) top 36.74% NA 35.59% 87.30% middle 35.70% 43.40% 52.97% 47.17% bottom 46.5% 44.73% 50.29% 37.39%
b a
c d
38
Table 2.*v results show that in case of narrow band illuminants focusing has become
more important, as some features are better focused in one image Fig. 2.*r (d), the top exudate
is most focused, and therefore has the highest contrast. In general for red-free illumination
contrast is moderate, but for optimal illuminants it is possible to achieve higher contrast values,
if image is focused properly.
For control subject the contrast of macula was calculated with both red-free illumination
and optimal illuminant. For right eye, the circular artifact didn’t impact the calculations of
contrast, however for left eye, the artifact covered one edge of macular area, and therefore
contrast calculations were obtained by the opposite edge.
In Fig. 2*s the OD/OS image of red-free and optimal illuminants can be seen, the rest of
the images are included in the 1st attachment.
Fig. 2.*s Fig. 2.*r Control subject’s (a-b) OD and (c-d) OS eye images. (a,c) Red-free images and
(b,d) optimal illuminant images for macula enhancement
b a
c d
39
Table 2.*u shows the contrast differences between red-free and optimal illuminants in
control subject. All cases show that the contrast of macula is improved.
Table 2.*u
Contrast comparison for red free filter and optimal illuminant in case of macula enhancement for control
subject (green cells show contrast of images in Fig. 2*s)
Right Eye Left Eye Red-free filter
contrast Optimal illuminant contrast
Red-free filter contrast
Optimal illuminant contrast
1. 20.28% 43.26% 28.19% 30.71% 2. 11.31% 37.62% 16.17% 3. 27.26%
40
2.5 Discussion
The results of this thesis show that narrow-band illuminations for retinal imaging include
detailed spectral information about the fundus of the eye that improve computer aided
diagnostics, however the implementation of such setup is not easy. Light sources have to have
an output of narrow band illumination that is spectrally tunable and short exposure time.
Most of the research on the subject of spectral retinal imaging, arrive at the conclusion
that brighter illumination is required, as the output narrow-band illumination is quite low, if
compared with input broadband illumination. The spectrally tunable light source created in this
study has high power output illumination that is enough to illuminate the retina even at narrower
bands, however the shorter and longer wavelengths (below 400 nm and above 650 nm) have
lower intensity output.
There have been many other different approaches to this field of study, such as narrow
band filters in front of a broad band illumination (different configurations, such as a wheel or
lath (27)). As well, there have been spectral light sources created by combining high powered
multiple wavelength LED illumination with narrow band interference filters, therefore the
spectral images can be recorded, but the light source is not completely spectrally tunable. (25)
It is mentioned that liquid crystal tunable filter (LCTF) is too slow, when multiple
wavelengths are required. (25) Another way to record specific wavelengths is a computer
tomographic imaging spectrometer that records both spatial and spectral information to a single
file. (37)
More recent study of spectral imaging include a LED light source and Piezo actuated
Fabry-Perot tunable light source that showed good results, however authors mention that
brighter light output would be needed for actual retinal imaging. (26)
41
CONCLUSIONS
The spectrally tunable light source has been constructed and the used optimal
illuminations provide improved contrast, therefore improving possibilities of potential means
of computer aided diagnostics of retinal lesions and feature detection.
Tasks:
1. Spectrally tunable light source has been constructed. The light source can be connected
to a modified fundus camera with a liquid light guide.
2. Calibration of light source has been successful. The parameters of setup have been
modified and the SPD’s of achieved illuminants are closely corresponding with target
illuminant SPD’s.
3. Images of mechanical model of the eye have been obtained. The optimal exposure time
and binning for a specific camera are acquired. Using the spectrally tunable light source
illumination for three separate channels (blue, green and red), allowed to create RGB
image of the mechanical model of the eye.
4. Images of a healthy human subject for retinal feature detection and diabetic retinopathy
patient images for retinal lesions have been obtained. In most cases the contrast of
optimal illuminants were an improvement over red-free filter images. In cases where
contrast wasn’t significantly improved, images were not focused on the lesion or feature,
therefore indicating that the focus of narrow-band illumination requires constant auto-
focusing.
42
ACKNOWLEDGMENTS
This study was funded by Academy of Finland, decision no. 259530.
I thank both of my supervisors from University of Eastern Finland Pauli Fält and Piotr
Bartczak for help in writing this thesis, doing the experiments and processing the data. As well
as for inviting me to participate in the ongoing research in their University.
I thank Niko Penttinen at University of Eastern Finland for creating a software for
controlling the spectrally tunable light source.
I thank Pasi Ylitepsa and Elina Hietanen for their assistance with image recording at their
department of Ophthalmology in University of Tampere.
I also thank my supervisor from University of Latvia – Pēteris Cikmačs for support in
writing this thesis.
Also I would like to thank prof. Māris Ozoliņš for help with terminology and translation
of the Thesis.
I thank for support in writing this thesis to my family, especially those who supported me
asking to read and re-read my thesis before it was completed.
43
BIBLIOGRAPHY
1. Schubert, H. D. Structure and Function of the Neural Retina. [book auth.] M. Yanoff and
J. S. Duker. Ophthalmology. 3rd ed. s.l. : Mosby Elsevier, 2009, p. 1528.
2. Ryan, S. J. Retina. 4th ed. s.l. : Mosby Elsevier, 2006. p. 2731.
3. Saine, P. J., Tyler, M. E. Ophthalmic Photography. 2nd ed. s.l. : Butterworth-Heinemann,
2002. p. 398.
4. Muchnick, B. Clinical Medicine in Optometric Practice. 2nd ed. s.l. : Mosby Elsevier,
2008. p. 398.
5. Walls, L. L. What the Practicing Optometrist Should Know About Diabetes. Optometry
Today. 05 02, 2002, pp. 42-45.
6. Laganovska, G., Klindžāne, M., Rasa, I. Tīklenes pārmaiņas diabēta gadījumā. [grāmatas
aut.] G. Laganovska. Acu slimības. Rīga : SIA "Nacionālais apgāds", 2008, lpp. 356.
7. Schlote, T., Rohrbach, J., Grueb, M., Mielke, J. Pocket Atlas of Ophthalmology. s.l. :
Georg Thieme Verlag, 2006. p. 258.
8. Kanski, J. J. Clinical Ophthalmology. 6th ed. s.l. : Butterworth-Heinemann, 2007. p. 931.
9. Williams, B. A History of Light and Lighting. Professional Lighting Resources. [Online]
1990-1999. [Cited: 05 04, 2015.] http://www.mts.net/~william5/history/hol.htm.
10. Baker, N., Brandt, L., Zakrewski, C. Smith College History of Science. [Online] [Cited:
05 2015, 04.] http://www.smith.edu/hsc/museum/ancient_inventions/candles2.html.
11. Tradition, History & Development of Candles. Waxmelters. [Online] [Cited: 05 04,
2015.]
12. A Brief History of the Ophthalmoscope. CR Keeler, The Royal College of
Ophthalmologists. s.l. : Optometry in Practice, 2003, Vol. 4, pp. 137-145.
13. Csele, M. Fundamentals of Light Sources and Lasers. s.l. : John Wiley & Sons, Inc.,
2004. p. 344.
14. Lamp inventors 1880-1940. The National Museum of American History. [Online]
[Cited: 05 05, 2015.] http://americanhistory.si.edu/lighting/bios/langmuir.htm.
15. Smith, F. G., King, T. A., Wilkins, D. Optics and Photonics. 2nd ed. s.l. : John Wiley
& Sons, Ltd., 2007. p. 506.
16. Harold E. Edgerton. Encyclopædia Britannica. [Online] [Cited: 05 05, 2015.]
http://www.britannica.com/EBchecked/topic/179106/Harold-E-Edgerton.
17. Photography of the Interior of the Eye. Howe, L. s.l. : Transactions of the American
Ophtalmological Society, 1887, Vol. 4, pp. 568-571.
18. Jackman, W. T., Webster, J. D. On Photographing the retina of the living eye.
Philadelphia Photographer. 1886, 265, pp. 340-341.
44
19. Espacnet Patent Search. European Patent Office. [Online] [Cited: 05 05, 2015.]
http://worldwide.espacenet.com/advancedSearch?locale=en_EP.
20. Fundus imaging with mobile phone: A review of techniques. Shanmugam, M. P.,
Divyansh, M., Madhukumar, R., Ramanjulu, R. J., Reddy, S., Y., Rodrigues, G. 9, s.l. :
Indian Journal of Ophthalmology, 2014, Vol. 62, pp. 960-962.
21. Reflex-free Digital Fundus Photography Using a Simple and Portable Camera Adaptor
System. A Viable Alternative. Pirie, C. G., Pizzirani, D., Pizzirani, S. 4, s.l. : Journal of
Visual Communications in Medicine, 2011, Vol. 34, pp. 146-155.
22. Landmarks in the Historical Development of Fluorescein Angiography. Saine, P. J. 1,
s.l. : Journal of Ophthalmic Photography, 1993, Vol. 15, pp. 17-22.
23. Yilmaz, T. The Basics of Indocyanine Green Angiography. s.l. : Association of
Technical Personel in Ophthalmology.
24. A Review of Indocyanine Green Fluorescent Imaging in Surgery. Alander, J. T.,
Kaartinen, I., Laakso, A., Pätilä, T., Spillmann, T., Tuchin, V. V., Venermo, M., Välisuo,
P. s.l. : International Journal of Biomedical Imaging, 2012, pp. 1-26.
25. Multispectral imaging of the ocular fundus using light emitting diode illumination.
Everdell, N. L., Styles, I. B., Cacagni, A., Gibson, J., Hebden, J., Claridge, E. s.l. :
Review of Scientific Instruments, 2010, Vol. 81, pp. 093706-1 - 093706-9.
26. Development of Tunable Fabry-Perot Spectral Camera and Light Source for Medical
Applications. Kaarre, M., Kivi, S., Panouillot, P. E., Saari, H., Mäkynen, J., Sorri, I.,
Juuti, M. s.l. : 3rd International Topical Meeting on Optical Sensing and Artificial
Vision, 2013, pp. 231-237.
27. Extending diabetic retinopathy imaging from color to spectra. Fält, P., Hiltunen, J.,
Hauta-Kasari, M., Sorri, I., Kalesnykiene, V., Uusitalo, H. s.l. : Proceedings of SCIA'09 -
16th Scandinavian Conference on Image Analysis, Lecture Note in Computer Science,
2009, Vol. 5575, pp. 149-158.
28. Webb, J. A. Digital fundus camera offers multiple functions. Optometry Times. 2009,
Vol. 1, 8, p. 12.
29. Screening coverage for diabetic retinopathy using a three-field digital non-mydriatic
fundus camera. Soto-Pedre, E., Hernaez-Ortega, M. C. s.l. : Primary Care Diabetes 2,
2008, pp. 141-146.
30. Meszaros, L. Why non-mydriatic cameras will not replace dilated fundus exams.
Optometry Times. June 2012, p. 28.
31. Systematic screening for diabetic retinopathy with a digital fundus camera following
pupillary dilatation in a university diabetes department. Deb-Joardar, N., Germain, N.,
45
Thuret, G., Garcin, A.-F., Manoli, P., Defreyn, A., Gain, P., Estour, B. s.l. : Diabetic
Medicine, 2007, Vol. 24, pp. 303-307.
32. High resolution fundus imaging by confocal scanning laser ophthalmoscopy in the
mouse. Paques, M., Simonutti, M., Roux, M. J., Picaud, S., Levavasseur, E., Bellman, C.,
Sahel, J. s.l. : Vision Research, 2006, Vol. 46, pp. 1336-1345.
33. Spectral imaging of the human retina and computationally determined optimal
illuminants for diabetic retinopathy lesion detection. Fält, P., Hiltunen, J., Hauta-Kasari,
M., Sorri. I., Kalesnykiene, V., Pietilä, J., Uusitalo, H. 3, s.l. : Journal of Imaging Science
and Technology, 2011, Vol. 55, pp. 030509-1 - 030509-10.
34. Thorlabs. [Online] 03 18, 2011. [Cited: 05 15, 2015.]
http://www.thorlabs.de/thorcat/19700/HPLS-30-04-Manual.pdf.
35. Maximum permissible exposures for ocular safety (ANSI 2000), with emphasis on
ophthalmic devices. Delori, F. C., Webb, R. H., Sliney, D. H. s.l. : Journal of the Optical
Society of America, 2007, Vol. 24, pp. 1250-1265.
36. In Search of a Contrast Metric: Matching the Perceived Contrast of Gabor Patches at
Different Phases and Bandwidths. Peli, E. 23, s.l. : Vision Research, 1997, Vol. 37, pp.
3217-3224.
37. Snapshot hyperspectral imaging in ophthalmology. Johnson, W. R., Wilson, D. W.,
Fink, W., Humayun, M., Bearman, G. 1, s.l. : Journal of Biomedical Optics, 2007, Vol.
12, pp. 014036-1 - 014036-7.
38. Quantitative analysis of multi-spectral fundus images. Styles, I. B., Calcagni, A.,
Claridge, E., Orihuela-Espina, F., Gibson, J. M. s.l. : Medical Image Analysis, 2006, Vol.
10, pp. 578-597.
46
ATTACHMENT: ANALYZED RETINAL IMAGES
Artery and vein comparison for red free filter on the left and optimal illuminant on the right
(subject 2, OS):
Artery and vein images for red free filter (subject 1, OS):
Artery and vein image for optimal illuminant (subject 1, OS):
47
Macula enhancement red free (on the left) and optimal illuminant (on the right) images (control
subject, OD):
Macula images for ref-free filter (control subject, OS):
48
Maģistra darbs “Tīklenes attēlu iegūšana ar dažādiem apgaismojumiem” izstrādāts Latvijas
Universitātes Fizikas un matemātikas fakultātē.
Master’s Thesis “Retinal imaging at various illuminations” was written in University of Latvia,
in faculty of Physics and Mathematics.
Ar savu parakstu apliecinu, ka pētījums veikts patstāvīgi, izmantoti tikai tajā norādītie
informācijas avoti un iesniegtā darba elektroniskā kopija atbilst izdrukai.
With my signature I confirm, that research has been done independently, all the used
bibliography has been shown and the submitted electronic copy of research corresponds with
printed version.
Autore/Author: _____________________________ Daiga Čerāne 25.05.2015.
Rekomendēju/nerekomendēju darbu aizstāvēšanai
I recommend/do not recommend thesis for defense
Vadītājs/Supervisor: Ph.D Pauli Fält _____________________________________________
Vadītājs/Supervisor: M.Sc. Piotr Bartczak _________________________________________
Rekomendēju/nerekomendēju darbu aizstāvēšanai
Vadītājs: Docents Pēteris Cikmačs _______________________________________________
Recenzents/Reviewer: Dr. Skaidrīte Purviņa
Darbs iesniegts Optometrijas un redzes zinātnes nodaļā ______________________________
Thesis has been submitted to Department of Optometry and Vision Science
Dekāna pilnvarotā persona/Dean’s authorized person: metodiķe Dzintra Holsta
_______________
Darbs aizstāvēts Valsts pārbaudījuma komisijas sēdē
______________. protokla Nr. ________________
Komisijas sekretārs: docents, Dr.fiz, Pēteris Cikmačs _______________________________